K Number
K172866

Validate with FDA (Live)

Device Name
NeuroCap
Manufacturer
Date Cleared
2018-04-27

(219 days)

Product Code
Regulation Number
882.1320
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NeuroCap (model DEC18) is an Electroencephalogram (EEG) electrode array intended for use in ER (emergency room), ICU (intensive care unit) and OR (operating room) for recording of STAT EEGs in patients of 18 years of age and older.

Device Description

The NeuroCap (model DEC18) is an Electroencephalogram (EEG) electrode array intended for use in ER (emergency room), ICU (intensive care unit) and OR (operating room) for recording of STAT EEGs in patients of 18 years of age and older. NeuroCap is intended for use with amplifiers of medical signals equipped with appropriate plug or by special adapter will not be included in subject device). NeuroCap is a disposable electrodes system made of polycarbonate based film with applied conductive paths and sensors. Each sensor is equipped with a sponge with a conductive gel for providing sensor contact with the skin surface. NeuroCap is fixed on the patient by means of adhesive tape. NeuroCap could be connected to amplifier directly or via a special adapter.

The NeuroCap disposable electrode system is 16 channels. The headband is non-sterile and disposable for single patient use and designed to be used with the Neuro EEG device (K173460) for EEG acquisition and recording. NeuroCap is intended for prescription use in healthcare facilities, or clinical research environments.

AI/ML Overview

The NeuroCap (model DEC18) is an Electroencephalogram (EEG) electrode array. The provided text outlines a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed study with specific acceptance criteria and detailed device performance metrics typically found in clinical efficacy trials for AI/ML devices.

However, based on the information provided, we can infer the "acceptance criteria" and "device performance" in the context of the 510(k) submission, which are primarily related to safety and performance through lab bench testing and comparison to a predicate device.

Here's an analysis based on the document:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of quantitative acceptance criteria and corresponding device performance values as would be typical for an AI/ML algorithm's diagnostic accuracy. Instead, the "acceptance criteria" are implied by compliance with established standards and the "device performance" is demonstrated through successful completion of these tests and a showing of substantial equivalence to the predicate.

Acceptance Criteria (Implied)Reported Device Performance
Biocompatibility: Conformance with ISO 10993-5 and ISO 10993-10Device passed biocompatibility tests according to ISO 10993-5 and ISO 10993-10 standards.
Risk Management: Conformance with ISO 14971:2012Device passed risk management procedures according to ISO 14971:2012 standards.
Performance: Conformance with ANSI/AAMI EC12:2000Device passed performance tests according to ANSI/AAMI EC12:2000 Disposable ECG electrodes standards.
Substantial Equivalence: To predicate device (StatNet)The NeuroCap's technological characteristics, features, specifications, materials, mode of operation, and intended use are substantially equivalent to the predicate device (StatNet). Differences in physical dimensions (size of caps, surface area, weight, package dimensions) were deemed to have no impact on safety and effect.

2. Sample Size Used for the Test Set and Data Provenance

The document describes lab bench testing, biocompatibility testing, and risk management assessment. These types of tests do not typically use "test sets" in the same way an AI/ML algorithm evaluation would (i.e., a dataset of patient cases).

  • Sample Size for Test Set: Not applicable in the context of a "test set" for an algorithm. The testing described (biocompatibility, performance, risk management) would have involved samples of the device itself and potentially biological samples for biocompatibility, but not a patient-level data test set.
  • Data Provenance: Not applicable in the context of patient data provenance. The testing was conducted in laboratories to assess the physical and biological properties of the device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: Not applicable. The ground truth for these types of engineering and biocompatibility tests is established by adhering to published standards (e.g., ISO, ANSI/AAMI) and engineering specifications, not by expert consensus on clinical cases.
  • Qualifications of Experts: N/A. The "experts" involved would be qualified lab technicians, engineers, and quality assurance personnel performing the tests according to the standards.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. Adjudication methods like 2+1 or 3+1 are used for establishing ground truth in clinical datasets, particularly for image interpretation or diagnosis. The tests mentioned are for device performance and safety, not for diagnostic accuracy evaluated against human expert consensus.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • MRMC Study: No. The document does not mention any MRMC study. The NeuroCap is an EEG electrode array, a hardware device, not an AI/ML diagnostic software. The evaluation focuses on the safety and performance of the electrodes themselves.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • Standalone Performance: Not applicable. The NeuroCap is a hardware device (electrodes), not an algorithm. Its performance is inherent to its physical and electrical properties, not algorithmic output, and it is intended to be used with an EEG device and interpreted by human clinicians.

7. The Type of Ground Truth Used

  • Type of Ground Truth: The ground truth for the evaluations conducted (biocompatibility, risk management, performance) is based on established industry standards and regulatory requirements (e.g., ISO 10993, ISO 14971, ANSI/AAMI EC12). For substantial equivalence, the "ground truth" is the established characteristics and performance of the legally marketed predicate device.

8. The Sample Size for the Training Set

  • Training Set Sample Size: Not applicable. The NeuroCap is a hardware device; it does not involve machine learning algorithms that require a "training set."

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth for Training Set: Not applicable, as there is no training set for a hardware device like the NeuroCap.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

April 27, 2018

Memory MD INC % Cassie Lee Manager Guangzhou GLOMED Biological Technology Co., Ltd. Suite 306, Kecheng Mansion, No.121 Science Road Guangzhou Science Park Guangzhou, Guangdong, 510006 Cn

Re: K172866

Trade/Device Name: NeuroCap Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: February 7, 2018 Received: February 12, 2018

Dear Cassie Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Vivek J. Pinto -S

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172866

Device Name NeuroCap

Indications for Use (Describe)

The NeuroCap (model DEC18) is an Electroencephalogram (EEG) electrode array intended for use in ER (emergency room), ICU (intensive care unit) and OR (operating of STAT EEGs in patients of 18 years of age and older.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.

1. Submitter's Information

510(k) Owner's Name: MemoryMD INC Establishment Registration Number: Applying Address: 205 E 42nd Street, 14th Floor New York, New York 10017 USA Phone: +1 (215)341-6373 Fax: +7 (8634) 614-323 Contact Person: Abdus-Salaam Muwwakkil, Chief Quality Officer E-mail: abdus@memorymd.com

Application Correspondent:

Application Correspondent: Contact Person: Ms. Cassie Lee Guangzhou GLOMED Biological Technology Co., Ltd. Address: Suite 306, Kecheng Mansion, No.121 Science Road, Guangzhou Science Park, Guangzhou 510663, China Tel: +86-20-61099984 Email: regulatory@glomed-info.com

2. Subject Device Information

Trade Name: NeuroCap Model: DEC18 Common Name: Cutaneous electrode Disposable Classification name: Cutaneous electrode Disposable Review Panel: Neurology Product Code: GXY Regulation Class: 2 Regulation Number: 882.1320

3. Predicate Device Information

510(K) Number: K092828 Company Name: HydroDot, Inc. Trade/Device Name: StatNet Regulation Number: 882.1320 Requlatory Class: 2 Product Code: GXY

4. Device Description

The NeuroCap (model DEC18) is an Electroencephalogram (EEG) electrode array intended for use in ER (emergency room), ICU (intensive care unit) and OR (operating room) for recording of STAT EEGs in

{4}------------------------------------------------

Sponsor:Memory MD INC
Subject Device:NeuroCap, Model: DEC18
Document Name:FDA 510(k) Submission Report

patients of 18 years of age and older. NeuroCap is intended for use with amplifiers of medical signals equipped with appropriate plug or by special adapter will not be included in subject device). NeuroCap is a disposable electrodes system made of polycarbonate based film with applied conductive paths and sensors. Each sensor is equipped with a sponge with a conductive gel for providing sensor contact with the skin surface. NeuroCap is fixed on the patient by means of adhesive tape. NeuroCap could be connected to amplifier directly or via a special adapter.

The NeuroCap disposable electrode system is 16 channels. The headband is non-sterile and disposable for single patient use and designed to be used with the Neuro EEG device (K173460) for EEG acquisition and recording. NeuroCap is intended for prescription use in healthcare facilities, or clinical research environments.

5. Intended Use / Indications for Use

The NeuroCap (model DEC18) is an Electroencephalogram (EEG) electrode array intended for use in ER (emergency room), ICU (intensive care unit) and OR (operating room) for recording of STAT EEGs in patients of 18 years of age and older.

6. Test Summary

The proposed NeuroCap has been evaluated the safety and performance by lab bench testing as following:

  • � Biocompatibility test according to ISO 10993-5 and ISO 10993-10 standards
  • � Risk management according to ISO 14971:2012 standards
  • � Performance test according to ANSI/AAMI EC12:2000 Disposable ECG electrodes standards

7. Comparison to predicate device and conclusion

The technological characteristics, features, specifications, materials, mode of operation, and intended use of the proposed NeuroCap is substantially equivalent to the predicate devices quoted below. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.

Elements ofComparisonSubject DevicePredicate DeviceRemark
Device Nameand ModelNeuroCap (Model: DEC18)StatNet (model: 3100, 3110)--
510(k) NumberApplyingK092828--
Product CodeGXYGXYSE

{5}------------------------------------------------

Sponsor:Memory MD INC
Subject Device:NeuroCap, Model: DEC18
Document Name:FDA 510(k) Submission Report
Intended Use /Indications forUseThe NeuroCap (modelDEC18) is anElectroencephalogram(EEG) electrode arrayintended for use in ER(emergency room), ICU(intensive care unit) and OR(operating room) forrecording of STAT EEGs inpatients of 18 years of ageand older.An Electroencephalogram(EEG) electrode array intendedfor use in the ER, ICU and ORfor recording of STAT EEGs inpatients of 18 years of age andolder.SE
PatientpopulationAdults only (>18 years old)Adults only (>18 years old)SE
UserTrained TechnicianTrained TechnicianSE
Number ofelectodes1818SE
ElectrodematerialAg/AgClAg/AgClSE
Type ofconductingmediumConductive gelConductive gelSE
Sizes of EEG -caps (by headcircumference)S (47-53 cm)M (50-56 cm)L (56-62 cm)M (50-56)L (56-62)SENote 1
Surface area ofeach electrode1.13 cm²1.56 cm²SENote 1
Work durationafter thepackageopening4 hours4 hoursSE
Storage life18 months18 monthsSE
Packagedimensionsless 490x460x15 mmless 520x290x15 mmSENote 1
Net Weightless 25 gramsless 40 gramsSENote 1
Gross Weightless 200 gramsless 80 gramsSENote 1

{6}------------------------------------------------

  • SE means " substantially equivalent".

Comparison in Detail(s):

Note 1: The "Sizes of EEG-caps", "Surface area of each electrode", "Net Weight", "Gross Weight", "Package Dimensions" have little difference between the subject device and the predicate one. They are all the appearance specifications and complying with the ISO 10993 biocompatibility requirements and ISO14971:2012 harm control application to medical device. Therefore, the differences between the two have no impacts on the safety and effect.

Finial Conclusion:

The subject product the NeuroCap is Substantial Equivalent to the predicate product.

  1. Date of the summary prepared: April 27, 2018

§ 882.1320 Cutaneous electrode.

(a)
Identification. A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.(b)
Classification. Class II (performance standards).