K Number
K172837

Validate with FDA (Live)

Date Cleared
2018-11-26

(433 days)

Product Code
Regulation Number
882.5890
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

PMS (Mode 1)

It is intended to stimulate healthy muscles in order to improve and facilitate muscle performance.

TENS (Mode 2)

To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back of the neck, arm, leg, and foot due to strain from exercise or normal household work activities by applying current to stimulate nerve.

Device Description

Physiotherapy instrument has 2 operation modes and one channel, after insert the electrode wire of the electrode pads, just put the electrode pads on the same side of your body, for shoulder, waist, back, back of the neck, arm, leg or foot (depends on your need, please consult your doctor before using if need), which can give certain electrical pulse through 2 pieces of electrode pads on the skin.

The electronic stimulatory module has the operating elements of ON/OFF Switch, Display screen, Mode Selection key, Intensity enhancement/weakening keys, language switch button and time selection keys. The LCD display screen can show selected mode, output intensity of body and/or sole, time remaining of an application mode, language display systemy.

The device is equipped with accessories of electrode pads, electrode wire, adapter.

The electrode wire is used to connect the pads to the main unit; the adapter wire is used to connect the adapter to the device.

AI/ML Overview

The provided document is a 510(k) summary for a Low-frequency Multi-function physiotherapy instrument. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a performance study with acceptance criteria for a new AI/ML-driven medical device.

Therefore, the requested information (acceptance criteria table, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, training set sample size, and ground truth establishment for the training set) is not present in this document.

This document describes a traditional medical device (Physiotherapy instrument) and its submission for FDA clearance via a 510(k) pathway. The test summary mentions "lab bench testing" for electrical safety, electromagnetic compatibility, biocompatibility, usability, software verification and validation, and waveform testing, all according to established IEC and ISO standards and FDA guidance documents. This type of submission relies on demonstrating that the new device is as safe and effective as a legally marketed predicate device, primarily through engineering and bench testing, rather than extensive clinical studies or performance against specific AI/ML-related metrics.

In summary, this document does not contain the information required to answer the prompt, as it pertains to the clearance of a non-AI/ML medical device through predicate equivalence, not a performance study of an AI/ML device against predefined acceptance criteria.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out as "U.S. Food & Drug Administration".

November 26, 2018

Shenzhen OSTO Technology Co., Ltd. % Cecilia Ceng Manager Guangzhou GLOMED Biological Technology Co., Ltd. Suite 306, Kecheng Mansion, No.121 Science Road Guangzhou Science Park Guangzhou. 510006 Cn

Re: K172837

Trade/Device Name: Low-frequency Multi-function physiotherapy instrument Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH, NGX Dated: October 17, 2018 Received: October 26, 2018

Dear Cecilia Ceng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Timothy A. Marjenin -S

Carlos L. Peña, PhD, MS For Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172837

Device Name

Physiotherapy instrument (Models: AST- 2011A, AST- 2011B, and AST-802B)

Indications for Use (Describe)

PMS (Mode 1)

It is intended to stimulate healthy muscles in order to improve and facilitate muscle performance.

TENS (Mode 2)

To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back of the neck, arm, leg, and foot due to strain from exercise or normal household work activities by applying current to stimulate nerve.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Chapter 5. 510(k) Summary

510(k) Summary

This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.

1. Submitter's Information

510(k) Owner's Name: Shenzhen OSTO Technology Co., Ltd. Establishment Registration Number: Applying Address: No.43 Longfeng Road, Xinsheng Community, Longgang Street Longgang District, Shenzhen City Guangdong Province, CHINA Tel: +86-755-29769546 Fax: +86-755-29769540 Contact Person: Li Yang (General Manager) Email: annaosto@163.com

Application Correspondent:

Contact Person: Ms. Cassie Lee Guangzhou GLOMED Biological Technology Co., Ltd. Address: Suite 306, Kecheng Mansion, No.121 Science Road, Guangzhou Science Park, Guangzhou 510663, China Tel: +86-20-61099984 Email: regulatory@glomed-info.com

2. Subject Device Information

Type of 510(k): Traditional Trade/Device Name: Physiotherapy instrument, Models: AST-2011A, AST-2011B, AST-802B Common Name: Electronic Simulator Classification Name: 1. Stimulator. Nerve. Transcutaneous. Over-The-Counter (NUH) 2. Stimulator, Muscle, Powered, For Muscle Conditioning (NGX) Review Panel: Neurology Product Code: NUH, NGX Regulation Number: 21 CFR 882.5890, 890.5850 Regulatory Class: II

3. Predicate Device Information

Sponser: Shenzhen OSTO Technology Company Limited Trade Name: Health Expert Electronic Stimulator Common Name: Electronic Stimulator Classification name: Stimulator, Nerve, Transcutaneous, Muscle, Powered, For Muscle Conditioning, Over-The-Counter 510(k)number: K133929 Review Panel: Neurology, Physical Medicine Product Code: NUH, NGX Regulation Number: 882.5890, 890.5850

{4}------------------------------------------------

Requlation Class: II

4. Device Description

Physiotherapy instrument has 2 operation modes and one channel, after insert the electrode wire of the electrode pads, just put the electrode pads on the same side of your body, for shoulder, waist, back, back of the neck, arm, leg or foot (depends on your need, please consult your doctor before using if need), which can give certain electrical pulse through 2 pieces of electrode pads on the skin.

The electronic stimulatory module has the operating elements of ON/OFF Switch, Display screen, Mode Selection key, Intensity enhancement/weakening keys, language switch button and time selection keys. The LCD display screen can show selected mode, output intensity of body and/or sole, time remaining of an application mode, language display systemy.

The device is equipped with accessories of electrode pads, electrode wire, adapter.

The electrode wire is used to connect the pads to the main unit; the adapter wire is used to connect the adapter to the device.

5. Intended Use / Indications for Use

PMS (Mode 1)

It is intended to stimulate healthy muscles in order to improve and facilitate muscle performance. TENS (Mode 2)

To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, back of the neck, arm, leg, and foot due to strain from exercise or normal household work activities by applying current to stimulate nerve.

6. Test Summary

Physiotherapy instrument, Models: AST-2011B, AST-2011B, AST-802B has been evaluated the safety and performance by lab bench testing as following:

  • Electrical safety test according to IEC 60601-1, IEC 60601-1-11 and IEC 60601-2-10 standards �
  • � Electromagnetic compatibility test according to IEC 60601-1-2 standard
  • � Biocompatibility test according to ISO 10993-5 and ISO 10993-10 standards
  • � Usability test according to IEC 62366 standard
  • � Software verification and validation test according to the requirements of the FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"
  • � The waveform test has also been conducted to verify the output specifications of the device according to "Guidance for Transcutaneous Electrical Nerve Stimulator for Pain Relief Intended for Over the Counter Use and Guidance for Powered Muscle Stimulator for Muscle Conditioning".

7. Comparison to predicate device and conclusion

The technological characteristics, features, specifications, mode of operation, and intended use of Electronic Muscle Stimulator is substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.

Comparison in Detail(s):

Elements ofComparisonSubject DevicePredicate DeviceVerdict
Elements ofComparisonSubject DevicePredicate DeviceVerdict
CompanyShenzhen OSTOTechnology Co., LtdShenzhen OSTO TechnologyCo., Ltd--
Trade NamePhysiotherapy instrumentHealth Expert ElectronicStimulator--
ModelAST-2011A, AST-2011B,AST-802BAST-300C, AST-300D--
ClassificationNameStimulator, Nerve,Transcutaneous, Over-The-CounterStimulator, Nerve,Transcutaneous, Over-The-Counter--
510(k) NumberApplyingK133929--
Product CodeNUH, NGXNUH, NGXSE
Intended Use /Indications for UsePMS (Mode 1)It is intended to stimulatehealthy muscles in orderto improve and facilitatemuscle performance.TENS (Mode 2)To be used for temporaryrelief of pain associatedwith sore and achingmuscles in the shoulder,waist,back, back of the neck,arm, leg, and foot due tostrain from exercise ornormal household workactivitiesby applying current tostimulate nerve.PMS (Mode 18)It is intended to stimulatehealthy muscles in order toimprove and facilitate muscleperformance.TENS (Mode 925)To be used for temporaryrelief of pain associated withsore and aching muscles inthe shoulder, waist,back, back of the neck, arm,leg, and foot due to strainfrom exercise or normalhousehold work activitiesby applying current tostimulate nerve.SE
Power SourceAdaptor Input: 100-240Vac, 50-60Hz, 0.1AOutput: 5Vdc, 1AUnit Input: 5Vdc, 1AAdaptor Input: 100-240Vac,50-60Hz, 0.1AOutput: 5Vdc, 1AUnit Input: 5Vdc, 1ASE
-Method of LineCurrent IsolationType BF Applied PartType BF Applied PartSE
PatientLeakageCurrentNormal Condition:AC: 54.5μA, DC: 0.5μASingle Fault Condition:AC:120.0μA, DC: 0.6μANormal Condition:AC: 54.5μA, DC: 0.5μASingle Fault Condition:AC:120.0μA, DC: 0.6μASE
Number of OutputModes225SENote 2
Number ofOutput Channels:12SENote 2
Elements ofComparisonSubject DevicePredicate DeviceVerdict
- Synchronous orAlternating?N/ASynchronousSENote 2
- Method ofChannel IsolationN/AVoltage TransformIsolation"BODY▼" and"BODY▼" buttons forbody channel,"SOLE▲" and"SOLE▼" buttons forfeet channelSENote 2
Regulated Currentor RegulatedVoltage?Voltage ControlVoltage ControlSE
Software/Firmware/MicroprocessorControl?YesYesSE
AutomaticOverload Trip?NoNoSE
Automatic No-LoadTrip?NoNoSE
Automatic ShutOff?YesYesSE
Patient OverrideControl?NoNoSE
IndicatorDisplay:- On/OffStatus?YesYesSE
- LowBattery?NoNoSE
Voltage/CurrentLevel?YesYesSE
Timer Range(minutes)5-15 min25 minSENote 3
Compliance withVoluntaryStandards?NoNoSE
Elements ofComparisonSubject DevicePredicate DeviceVerdict
Compliance* with21 CFR 898?YesYesSE
Main Unit Weight105 g2 kgSENote 1
Main UnitDimensionAST-802B: 156.6 x 74.5 x17.8 mmAST-2011A: 157.03 X72.87 X 23.89 mmAST-2011B: 156.6 x 74.5x 17.8 mm428 x 428.8 x 185mmSENote 1
Housing Materialsand ConstructionMain unit: ABS plasticMain unit: ABS plasticSE
Output Specification
WaveformPulsed, symmetric,biphasicPulsed, symmetric,biphasicSE
ShapeRectangular, withinterphase intervalRectangular, withinterphase intervalSE
Maximum OutputVoltage44V±10% @ 500Ω80V±10% @ 2KΩ112V±10% @ 10KΩ44V±10% @ 500Ω80V±10% @ 2KΩ112V±10% @ 10KΩSE
Maximum OutputCurrent88mA±10% @ 500Ω40mA±10% @ 2KΩ11.2mA±10% @ 10KΩ88mA±10% @ 500Ω40mA±10% @ 2KΩ11.2mA±10% @ 10KΩSE
Pulse Duration120µs120µsSE
Pulse frequency77.3Hz77.3HzSE
Symmetricalphases?YesYesSE
Phase Duration12.94 ms12.94 msSE
Net Charge (mCper pulse)0µC @ 500ΩMethod: Balancedwaveform0µC @ 500ΩMethod: Balanced waveformSE
Maximum PhaseCharge, (mC)12.78 µC @ 500Ω12.78 µC @ 500ΩSE
Maximum CurrentDensity, (mA/cm²)0.235 mA/cm² @ 500Ω0.235 mA/cm² @ 500ΩSE
Maximum PowerDensity, (W/cm²)1.38 mW/cm² @ 500Ω1.38 mW/cm² @ 500ΩSE
StimulationparametersSee attached table 1See attached table 1SE
Elements ofComparisonSubject DevicePredicate DeviceVerdict
BurstModePulses perburst400400SE
Bursts persecond0.630.63SE
Burstduration(seconds)2.5882.588SE
Duty Cycle[Line (b) xLine (c)]1.631.63SE
ON Time (seconds)0.6s0.6sSE
OFF Time(seconds)0.6s0.6sSE
Additional Features
Environment foroperationTemperature: 5 ~ 45Humidity: 15% - 90%Atmospheric pressurerange of 700 hPa to 1 060hPaTemperature: 5 ~ 45° CHumidity: 20 ~ 65% RHSENote 4
Environment forstorageTemperature: -25 - +70°CHumidity: up to 90%Atmospheric pressurerange of 700 hPa to 1 060hPaTemperature: 0 ~ 45° C,Humidity: 10 ~ 90% RHElectrode Pad: 10~20° CSENote 4

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

Table1
Output Specification
Pulse Duration and PulseCycle
Note: "T" means Pulse Cycle.Note: "T" means Pulse Cycle.
Cycle ofMode 1N/A
MaxMin
Intensity
Peak Voltage (V) (0.5KΩ)44 V0.63
Peak Voltage (V) (2KΩ)80 V2.48
Peak Voltage (V) (10K Ω)112 V11.80
Cycle ofMode 2200T200T
IntensityMaxMinMaxMin
Peak Voltage (V) (0.5KΩ)44 V0.6344 V0.63
Peak Voltage (V) (2KΩ)80 V2.4880 V2.48

{10}------------------------------------------------

Peak Voltage (V) (10K Ω)112 V11.80112 V11.80
Cycle ofMode 3Image: Sub-mode 5 X 1 1s Sub-mode 6 X 2
IntensityN/AMaxMin
Peak Voltage (V) (0.5KΩ)44 V0.63
Peak Voltage (V) (2KΩ)80 V2.48
Peak Voltage (V) (10K Ω)112 V11.80
Cycle ofMode 4Image: Sub-mode 6 X 1 0.5s Sub-mode 17 X 10
IntensityN/AMaxMin
Peak Voltage (V) (0.5KΩ)44 V0.63
Peak Voltage (V) (2KΩ)80 V2.48
Peak Voltage (V) (10K Ω)112 V11.80
Cycle ofMode 5Image: Sub-mode 9 X 1 1s Sub-mode 15 X 10
IntensityN/AMaxMin
Peak Voltage (V) (0.5KΩ)44 V0.63
Peak Voltage (V) (2KΩ)80 V2.48
Cycle ofMode 6N/AImage: Sub-mode 15 X 1 and Sub-mode 16 X 1 with 0.5s
IntensityN/AMaxMin
Peak Voltage (V) (0.5KΩ)N/A44 V0.63
Peak Voltage (V) (2KΩ)N/A80 V2.48
Peak Voltage (V) (10K Ω)N/A112 V11.80
Cycle ofMode 7N/AImage: Sub-mode 5 X 1 and Sub-mode 6 X 1 with 1s
IntensityN/AMaxMin
Peak Voltage (V) (0.5KΩ)N/A44 V0.63
Peak Voltage (V) (2KΩ)N/A80 V2.48
Peak Voltage (V) (10K Ω)N/A112 V11.80
Cycle ofMode 8N/AImage: Sub-mode 18 X 5 and Sub-mode 17 X 5 with 0.05s
IntensityN/AMaxMin
Peak Voltage (V) (0.5KΩ)N/A44 V0.63
Peak Voltage (V) (2KΩ)N/A80 V2.48
Peak Voltage (V) (10K Ω)N/A112 V11.80
Cycle ofMode 9N/AImage: Sub-mode 11 X 5 and Sub-mode 15 X 5 with 0.3s
IntensityMaxMin
Peak Voltage (V) (0.5KΩ)N/A44 V0.63
Peak Voltage (V) (2KΩ)80 V2.48
Peak Voltage (V) (10K Ω)112 V11.80
Cycle ofMode 10Image: Sub-mode 8 X 1, Sub-mode 14 X 10, 0.1s
IntensityMaxMin
Peak Voltage (V) (0.5KΩ)N/A44 V0.63
Peak Voltage (V) (2KΩ)80 V2.48
Peak Voltage (V) (10K Ω)112 V11.80
Cycle ofMode 11Image: Sub-mode 5 X 5, Sub-mode 11 X 5, 1s
IntensityMaxMin
Peak Voltage (V) (0.5KΩ)N/A44 V0.63
Peak Voltage (V) (2KΩ)80 V2.48
Peak Voltage (V) (10K Ω)112 V11.80
Cycle ofMode 12Image: Sub-mode 2 X 1, Sub-mode 6 X 20, 0.5s
IntensityMaxMin
Peak Voltage (V) (0.5KΩ)44 V0.63
Peak Voltage (V) (2KΩ)N/A80 V2.48
Peak Voltage (V) (10K Ω)112 V11.80
Cycle ofMode 13Image: Sub-mode 11 X 1 Sub-mode 12 X 1 0.3s
IntensityMaxMin
Peak Voltage (V) (0.5KΩ)N/A44 V0.63
Peak Voltage (V) (2KΩ)80 V2.48
Peak Voltage (V) (10K Ω)112 V11.80
Cycle ofMode 14Image: 200T 0.5s
IntensityN/AMaxMin
Peak Voltage (V) (0.5KΩ)44 V0.63
Peak Voltage (V) (2KΩ)80 V2.48
Peak Voltage (V) (10K Ω)112 V11.80
Cycle ofMode 15N/AImage: 15T 0.5s
IntensityMaxMin
Peak Voltage (V) (0.5KΩ)44 V0.63
Peak Voltage (V) (2KΩ)80 V2.48
Peak Voltage (V) (10K Ω)112 V11.80
Cycle ofMode 16N/AImage: 200T 0.25s
IntensityN/AMaxMin
Peak Voltage (V) (0.5KΩ)44 V0.63
Peak Voltage (V) (2KΩ)80 V2.48
Peak Voltage (V) (10K Ω)112 V11.80
Cycle ofMode 17N/AImage: 15T 0.3s
IntensityN/AMaxMin
Peak Voltage (V) (0.5KΩ)44 V0.63
Peak Voltage (V) (2KΩ)80 V2.48
Peak Voltage (V) (10K Ω)112 V11.80
Cycle ofMode 18N/AImage: 200T 0.1s
IntensityN/AMaxMin
Peak Voltage (V) (0.5KΩ)44 V0.63
Peak Voltage (V) (2KΩ)80 V2.48
Peak Voltage (V) (10K Ω)112 V11.80
Cycle ofMode 19N/AImage: 15T 0.15s
IntensityMaxMin
Peak Voltage (V) (0.5KΩ)44 V0.63
Peak Voltage (V) (2KΩ)80 V2.48
Peak Voltage (V) (10K Ω)112 V11.80
Cycle ofMode 20N/AImage: 100T 0.5s
IntensityMaxMin
Peak Voltage (V) (0.5KΩ)44 V0.63
Peak Voltage (V) (2KΩ)80 V2.48
Peak Voltage (V) (10K Ω)112 V11.80
Cycle ofMode 21Image: 30T 0.05sImage: 30T 0.05s
IntensityMaxMinMaxMin
Peak Voltage (V) (0.5KΩ)44 V0.6344 V0.63
Peak Voltage (V) (2KΩ)80 V2.4880 V2.48
Peak Voltage (V) (10K Ω)112 V11.80112 V11.80
Cycle ofMode 22N/AImage: [100T 0.25s]
IntensityN/AMaxMin
Peak Voltage (V) (0.5KΩ)N/A44 V0.63
Peak Voltage (V) (2KΩ)N/A80 V2.48
Peak Voltage (V) (10K Ω)N/A112 V11.80
Cycle ofMode 23N/AImage: [30T 0.3s]
IntensityN/AMaxMin
Peak Voltage (V) (0.5KΩ)N/A44 V0.63
Peak Voltage (V) (2KΩ)N/A80 V2.48
Peak Voltage (V) (10K Ω)N/A112 V11.80
Cycle ofMode 24N/AImage: [100T 0.1s]
IntensityN/AMaxMin
Peak Voltage (V) (0.5KΩ)N/A44 V0.63
Peak Voltage (V) (2KΩ)N/A80 V2.48
Peak Voltage (V) (10K Ω)112 V11.80
Cycle ofMode 25N/AImage: [Pulse Diagram]
IntensityMaxMin
Peak Voltage (V) (0.5KΩ)44 V0.63
Peak Voltage (V) (2KΩ)80 V2.48
Peak Voltage (V) (10K Ω)112 V11.80

{11}------------------------------------------------

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

{16}------------------------------------------------

{17}------------------------------------------------

{18}------------------------------------------------

Note 1 (Weight, Dimension):

These data would be different for different devices because the internal circuit design and components choosing are different, which make them have similar difference on weight, dimensions. But weight and dimensions won't affect the safety and effectiveness of the device so it can deem as the substantially equivalence.

Note 2 (Number of Output Modes):

Although the "Number of Output Modes" of subject device are different from the predicate devices, thev are all comply with IEC 60601-1 and IEC 60601-2-10 requirements. So, the differences of the function specifications will not raise any safety or effectiveness issue.

Note 3 (Time Range):

The design of the time range is basing on the intended use. And according to the output specification comparing with the predicated devices, we set the default treatment time is 15min and the user could adjust the levels which could meet the requirements in the energy aspect. Thus, the is the same as predicated ones.

Note 4 (Environment for Operating and storage):

These data are to specify the condition of working, transportation and storage, for different device, the internal circuit design and components choosing are different, so they also have similar difference on working or storage conditions. But the temperature. RH are approximate and complied with the normal conditions, so it can be deemed as the substantially equivalence.

Final conclusion:

The nonclinical tests and comparation demonstrate that the subject device "Physiotherapy instrument model: AST-2011A, AST-2011B, AST-802B" is as safe, as effective, and performs as well as the predicate devices "Health Expert Electronic Stimulator AST-300C, AST-300D". So the subject device is Substantial Equivalent to predicate device K133929.

8. Date of the summary prepared: October 17. 2018

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).