K Number
K172758

Validate with FDA (Live)

Date Cleared
2018-03-07

(175 days)

Product Code
Regulation Number
876.4300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Polypectomy Snare is used endoscopically in the removal and/or and cauterization of diminutive polyps, sessile polyps, pedunculated polyps and tissue from within the gastrointestinal tract.

Device Description

The Polypectomy Snare consists of a flexible wire cable (core wire) and loop which can be extended and retracted from the Snare's flexible outer sheath using a threering handle. The flexible cable and loop can also be rotated 360° using the rotation actuator on the handle. The inner diameter of the sheath is PTFE to provide minimal friction during extension, rotation, and retraction of the loop from the sheath. When passed through an endoscope and activated, the Snare delivers a monopolar electrical current to cut and cauterize tissue with the loop.

AI/ML Overview

The document provides a 510(k) summary for the "Rotatable Snares, Non-Rotatable Snares" device (Polypectomy Snare). This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than independent clinical performance studies. Therefore, much of the requested information regarding AI device performance, sample sizes for test and training sets, expert qualifications, adjudication methods, and MRMC studies is not applicable to this submission type and the provided documentation.

However, the document does contain information about performance testing to ensure the device meets specified criteria.

Here's a summary of the available information based on your request:

1. Table of Acceptance Criteria and Reported Device Performance

The submission states that "In-vitro Testing has been performed and all components, subassemblies, and/or full devices met the required specifications for the completed tests." A summary of the tests performed is provided, implying that the device successfully met the acceptance criteria for these tests. The specific quantitative acceptance criteria for each test are not detailed in this summary document, but the device is reported to have met them.

Test PerformedReported Device Performance
Working LengthMet required specifications
Tortuous Path FunctionalityMet required specifications
Rotation CapabilityMet required specifications
Handle to Core Wire Tensile StrengthMet required specifications
Electrical ResistanceMet required specifications
Electrical SafetyMet required specifications
Snare ActuationMet required specifications
EO residual and ECH residual (after sterilization)Met criteria defined in ISO 11135 and ISO 10993-7
Shelf-life (3 years)Validated in accelerated testing (ASTM F1980-16)
PackagingMet requirements of ISO 11607-1 and ISO 11607-2
Biocompatibility (cytotoxicity, sensitization, intracutaneous irritation, system toxicity pyrogen, endotoxin)Demonstrated biocompatibility (in accordance with ISO 10993-1)
Electromagnetic Compatibility and Electrical SafetyPerformed in accordance to AAMI/ANSI ES 60601-1 and AAMI/ANSI/IEC 60601-2-2

2. Sample size used for the test set and the data provenance

  • Sample Size: The document does not specify the sample sizes used for each of the in-vitro performance tests. It only states that "all components, subassemblies, and/or full devices" underwent testing.
  • Data Provenance: The tests are in-vitro, meaning they were performed in a lab setting. The country of origin of the data is not explicitly stated for each test, but the manufacturer is based in Hangzhou, Zhejiang, China. The data is retrospective in the sense that the tests were completed before the 510(k) submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • This information is not applicable. The performance testing described is in-vitro engineering and biocompatibility testing, not clinical studies requiring expert ground truth for diagnostic or prognostic purposes.

4. Adjudication method for the test set

  • This information is not applicable. The tests performed are objective measurements against predefined specifications, not subjective assessments requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This information is not applicable. The device is an endoscopic accessory (polypectomy snare), not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was conducted.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • This information is not applicable. The device is a physical medical instrument, not an algorithm, and therefore there is no standalone algorithm performance to evaluate.

7. The type of ground truth used

  • The "ground truth" for the performance tests outlined (e.g., working length, tensile strength, electrical resistance, biocompatibility, sterilization residuals) is defined by established engineering specifications, international standards (ISO, ASTM, AAMI/ANSI), and regulatory guidance, rather than expert consensus, pathology, or outcomes data. For example, ISO 11135 defines acceptable EO residual levels.

8. The sample size for the training set

  • This information is not applicable. The device is a physical medical instrument, not an AI model, and therefore there is no training set.

9. How the ground truth for the training set was established

  • This information is not applicable. As there is no training set for an AI model, there is no ground truth to establish for it.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

March 7, 2018

Zhejiang Chuangxiang Medical Technology Co., LTD. Lucius Long Quality Supervisor 301B, No.22, XinYan Road Hangzhou, Zhejiang 311100 China

Re: K172758

Trade/Device Name: Rotatable Snares, Non-Rotatable Snares Regulation Number: 21 CFR§ 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: II Product Code: FDI Dated: January 4, 2018 Received: January 4, 2018

Dear Lucius Long:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/6 description: The image shows the name "Benjamin R. Fisher -S" in a large, bold font. The text is black and appears to be centered on a white background. The letters are clearly legible, and the overall impression is clean and professional.

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172758

Device Name Rotatable Snares Non-Rotatable Snares

Indications for Use (Describe)

The Polypectomy Snare is used endoscopically in the removal and/or and cauterization of diminutive polyps, sessile polyps, pedunculated polyps and tissue from within the gastrointestinal tract.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains a logo with a geometric shape on the left, transitioning from blue to green. To the right of the shape, there is text in blue that appears to be Chinese characters. Below the Chinese characters, the word "MEDNOVA" is written, with "MED" in blue and "NOVA" in green, matching the color gradient of the geometric shape.

Section 2 510(k) Summary( 21CFR 807.92)

1. Submitter's information

Name: Zhejiang Chuangxiang Medical Technology Co., LTD.

Address: 301B, No.22, XinYan Road Yuhang Economic And Technological Development Zone Hangzhou Zhejiang China

Contact person: Lucius.Long Telephone: 86-571-89167088 Fax: 86-571-89167086

2. Device

Trade Name: Rotatable Snares, Non-Rotatable Snares Common Name: Polypectomy Snare Classification name: Endoscopic electrosurgical unit and accessories Regulation class: 2 Regulation number:876.4300 Panel: Gastroenterology/Urology Product code: FDI

3. Predicative device

510(k) Number: K160637, Product Name: Rotatable Snares

4. Device description

The Polypectomy Snare consists of a flexible wire cable (core wire) and loop which can be extended and retracted from the Snare's flexible outer sheath using a threering handle. The flexible cable and loop can also be rotated 360° using the rotation actuator on the handle. The inner diameter of the sheath is PTFE to provide minimal friction during extension, rotation, and retraction of the loop from the sheath. When passed through an endoscope and activated, the Snare delivers a monopolar electrical current to cut and cauterize tissue with the loop.

For All models information in this submission, please refer to below table 1.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains a logo with text. On the left side, there is a geometric shape with a gradient of green and blue. To the right of the shape, there is the text "MEDNOVA" in green and blue. Above the text, there are three Chinese characters in blue.

Table 1 List of models in this submission

ModelPart NumberMax. OD(mm)≤ D (mm)Snare LoopOpening Width(mm)B +10%-20%Effective Length(mm)L (mm) ±100
EOvalMD-E-SNC181024Φ1.8102400
MD-E-SNC181524Φ1.8152400
MD-E-SNC181018Φ1.8101800
MD-E-SNC181518Φ1.8151800
EOvalMD-E-SNC241024Φ2.4102400
MD-E-SNC241524Φ2.4152400
MD-E-SNC241018Φ2.4101800
MD-E-SNC241518Φ2.4151800
MD-E-SNC242516Φ2.4251600
MD-E-SNC242524Φ2.4252400
MD-E-SNC243224Φ2.4322400
MD-E-SNC242518Φ2.4251800
Standard(C)MD-E-SNC243218Φ2.4321800
MD-H-SNC241024Φ2.4102400
MD-H-SNC241524Φ2.4152400
HHexagonalMD-H-SNC242224Φ2.4222400
MD-H-SNC243224Φ2.4322400
MD-H-SNC241018Φ2.4101800
MD-H-SNC241518Φ2.4151800
HHexagonalMD-H-SNC242218Φ2.4221800
MD-H-SNC243218Φ2.4321800
MD-C-SNC241524Φ2.4152400
MD-C-SNC242524Φ2.4252400
CCrescentMD-C-SNC243224Φ2.4322400
MD-C-SNC241518Φ2.4151800
MD-C-SNC242518Φ2.4251800
MD-C-SNC243218Φ2.4321800
RMD-R-SNC243224$Φ$ 2.4322400
RoundMD-R-SNC243218$Φ$ 2.4321800
EOvalMD-E-SNR181024$Φ$ 1.8102400
MD-E-SNR181524$Φ$ 1.8152400
MD-E-SNR181018$Φ$ 1.8101800
MD-E-SNR181518$Φ$ 1.8151800
MD-E-SNR241024$Φ$ 2.4102400
MD-E-SNR241524$Φ$ 2.4152400
MD-E-SNR241018$Φ$ 2.4101800
MD-E-SNR241518$Φ$ 2.4151800
MD-E-SNR242516$Φ$ 2.4251600
MD-E-SNR242524$Φ$ 2.4252400
MD-E-SNR243224$Φ$ 2.4322400
MD-E-SNR242518$Φ$ 2.4251800
MD-E-SNR243218$Φ$ 2.4321800
Rotational(R)HHexagonalMD-H-SNR241024$Φ$ 2.4102400
MD-H-SNR241524$Φ$ 2.415
MD-H-SNR242224$Φ$ 2.422
MD-H-SNR243224$Φ$ 2.432
MD-H-SNR241018$Φ$ 2.410
MD-H-SNR241518$Φ$ 2.415
MD-H-SNR242218$Φ$ 2.422
MD-H-SNR243218$Φ$ 2.432
CCrescentMD-C-SNR241524$Φ$ 2.4152400
MD-C-SNR242524$Φ$ 2.425
MD-C-SNR243224$Φ$ 2.432
MD-C-SNR241518$Φ$ 2.415
MD-C-SNR242518$Φ$ 2.425
MD-C-SNR243218$Φ$ 2.432
RMD-R-SNR243224$Φ$ 2.4322400
RoundMD-R-SNR243218$Φ$ 2.4321800

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image contains a logo for a company called MEDNOVA. The logo consists of a green and blue shield with a white geometric shape inside. To the right of the shield is the company name, with "MED" in green and "NOVA" in blue. Above the company name are two Chinese characters in blue.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for MEDNOVA. The logo consists of a green and blue shield-like shape on the left, with the text "MEDNOVA" in green below it. To the right of the shield is the company name in Chinese, written in blue.

5. Indication for use

The Polypectomy Snare is used endoscopically in the removal and/or and cauterization of diminutive polyps, sessile polyps, pedunculated polyps and tissue from within the gastrointestinal tract.

6. Technological Characteristics

The Polypectomy Snare incorporates substantially equivalent device materials, design, configuration, packaging fundamental technology, manufacturing processes, sterilization process and intended use as those featured in the Boston Scientific Corporation predicate devices.

Comparison to predicate Devices:

ItemProposed devicePredicate deviceComparison toPredicateDevices
Device namePolypectomySnareRotatable SnaresSimiliar
510(k)numberNA, newsubmissionK160637Different 510(k)number
ManufacturerZhejiangChuangxiang medicaltechnology Co., Ltd.Boston ScientificCorporationDifferentmanufacturer
ProductCodeFDIFDI, FGXSame
RegulationNo.876.4300876.4300Same
Class22Same
SuppliedSterileYesYesSame
ConfigurationOval, Hexagonal,Crescent, RoundOval, Hexagonal,Crescent, RoundSame
Minimalworkingchannel2.8mm2.8mmSame
Sheath OD2.4mm2.4mmSame
WorkingLength1600mm,1800mm,2400mm2400mmsimilar
Loop width10mm, 15mm, 22mm10mm, 11mm, 13mm, 15mmsimilar

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for MEDNOVA. The logo consists of a geometric shape in shades of green on the left, followed by the text "MEDNOVA" in green. To the right of the logo is Chinese text in blue.

mm,25mm,32mm,mm,20mm,25mm,27mm,30mm,33mm
Indicationsfor UseThe PolypectomySnare is usedendoscopically in theremoval and/orcauterization ofdiminutive polyps,sessile polyps,pedunculated polypsand tissue from withinthe gastrointestinaltract.The Rotatable Snare isused endoscopically in theremoval and/orcauterization of diminutivepolyps, sessile polyps,pedunculated polyps andtissue from within thegastrointestinal tract.Same
Single UseYesYesSame
PackagingSingle-use EOsterilized pouch withone device per pouchSingle-use EO sterilizedpouch with one device perpouchSame

7. Performance data

In-vitro Testing has been performed and all components, subassemblies, and/or full devices met the required specifications for the completed tests. A summary of the test results has been provided for Working Length, Tortuous Path Functionality, Rotation Capability, Handle to Core Wire Tensile Strength, Electrical Resistance, Electrical Safety and Snare Actuation.

The EO residual and ECH residual were measured after sterilization of the device to meet the criteria defined in ISO 11135 Second edition 2014 and ISO 10993-7 Second Edition 2008-10-15.

The shelf-life for three years had been validated in accelerated testing according to ASTM F1980-16 (2016) and the requirements on packaging for terminally sterilized medical device per ISO 11607-1 First Edition 2006-04-15 and ISO 11607-2 First Edition 2006-04-15 are also met. The testing successfully demonstrated essential performance is achieved before and after the shelf life test.

Biocompatibility testing was performed in accordance with the FDA Guidance," Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" issued on June 26, 2016. The cytotoxicity, sensitization, intracutaneous irritation, system toxicity pyrogen and endotoxin tests were performed to demonstrate the biocompatibility of the device.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for MEDNOVA. The logo consists of a blue and green hexagon shape on the left, with the text "MEDNOVA" in green and blue to the right of the shape. Above the text "MEDNOVA" are three blue Chinese characters.

Tests on Electromagnetic Compatibility and Electrical Safety were performed in accordance to requirements per AAMI/ANSI ES 60601-1:2005/(R)2012 and AAMI/ANSI/IEC 60601-2-2:2009.

8. Clinical Test Conclusion

No clinical study is included in this submission.

9. Conclusions

Chuangxiang medical has demonstrated that the proposed Polypectomy Snare are substantially equivalent to Boston Scientific Corporation's currently marketed Rotatable Snares(K160637).

§ 876.4300 Endoscopic electrosurgical unit and accessories.

(a)
Identification. An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.(b)
Classification. Class II (performance standards).