K Number
K172685

Validate with FDA (Live)

Device Name
LiverMultiScan
Date Cleared
2017-11-21

(76 days)

Product Code
Regulation Number
892.1000
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

LiverMultiScan (LMSv2) is indicated for use as a magnetic device software application for noninvasive liver evaluation that enables the generation, display and review of 2D magnetic resonance medical image data and pixel maps for MR relaxation times.

LiverMultiScan (LMSv2) is designed to utilize DICOM 3.0 compliant magnetic resonance image datasets, acquired from compatible MR Systems, to display the internal structure of the abdomen including the liver. Other physical parameters derived from the images may also be produced.

LiverMultiScan (LMSv2) provides a number of quantification tools, such as Region of Interest (ROI) placements, to be used for the assessment of regions of an image to quantify liver tissue characteristics, including the determination of triglyceride fat fraction in the liver, T2* and iron-corrected T1 measurements.

These images and the physical parameters derived from the images, when interpreted by a trained clinician, yield information that may assist in diagnosis.

Device Description

LiverMultiScan (LMSv2) is a standalone software application for displaying 2D Magnetic Resonance medical image data acquired from compatible MR Scanners. LiverMultiScan runs on a general purpose workstation with a colour monitor, keyboard and mouse.

LiverMultiScan (LMSv2) is designed to allow the review of DICOM 3.0 compliant datasets stored on the workstation and the operator may also create, display, print, store and distribute reports resulting from interpretation of the datasets.

LiverMultiScan (LMSv2) allows the display and comparison of combinations of magnetic resonance images and provides a number of tools for the quantification of magnetic resonance images, including the determination of triglyceride fat fraction in the liver, T2* and iron-corrected T1 measurements.

LiverMultiScan (LMSv2) provides a number of tools, such as circular region of interest placements, to be used for the assessment of regions of an image to support a clinical workflow.

LiverMultiScan (LMSv2) allows the operator to create relaxometry parameter maps of the abdomen which can be used by clinicians to help determine different tissue characteristics to support a clinical workflow. Examples of such workflows include, but are not limited to, the evaluation of the presence or absence of liver fat.

LiverMultiScan (LMSv2) is intended to be used by trained operators. Reports generated by trained operators are intended for use by interpreting clinicians, including, but not limited to radiologists, gastroenterologists, and hepatologists.

LiverMultiScan (LMSv2) is an aid to diagnosis. When interpreted by a trained clinician, the results provide information, which may be used as an input into existing clinical procedures and diagnostic workflows.

LiverMultiScan (LMSv2) offers the following.

  • Advanced visualisation of MR data.
  • Processing of MR data to quantify tissue characteristics including MR relaxivity constants such as T2*, T1, ironcorrected T1 (cT1) and triglyceride fat fraction (expressed as liver fat percentage).
  • Circular region of interest statistics.
  • Snapshot of images to include in a report.
  • Report to include region statistics, snapshot images and user-entered text.
  • Export of snapshot images to report.
AI/ML Overview

Here's a summary of the acceptance criteria and study details for LiverMultiScan (LMSv2) based on the provided FDA 510(k) summary:

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria are framed in terms of accuracy, repeatability, reproducibility, and equivalence to the predicate device (LMSv1) for the measurements of T1 (corrected T1 or cT1 for in vivo), T2*, and Proton Density Fat Fraction (PDFF, also referred to as triglyceride fat fraction).

Phantom Study Performance:

phantom metricAcceptance Criteria (Performance metric)Reported Device Performance (95% CI Limits of Agreement)
Accuracy
T1Consistent with literature-reported underestimation19-25% lower than ground truth (consistent with MOLLI techniques)
T2*Accurate over expected physiological range+/- 2ms
PDFF < 30%Accurate over expected physiological range+/- 3%
PDFF ≥ 30%Accurate over expected physiological range+/- 21%
RepeatabilityHighly repeatable
T1+/- 10ms+/- 10ms
T2*+/- 1.7ms+/- 1.7ms
PDFF < 30%-2.5 to 1 %-2.5 to 1 %
PDFF ≥ 30%+/- 5%+/- 5%
ReproducibilityReproducible between systems (at same field strength for T2*)
T1-34 to 27ms-34 to 27ms
T2*+/- 4ms+/- 4ms (between systems at same field strength)
PDFF < 30%+/- 4%+/- 4%
PDFF ≥ 30%+/- 10%+/- 10% (between systems)
Equivalence to LMSv1
T1Within 1ms of predicate-0.5 to 0.4ms
T2*Within 1ms of predicate-0.4 to 0.2ms
PDFF < 30%Within 1% of predicate-0.6 to 0.4%
PDFF ≥ 30%Within 2% of predicate-1.6 to 2%

In Vivo Study Performance:

Volunteer metricAcceptance Criteria (Performance metric)Reported Device Performance (95% CI Limits of Agreement)
RepeatabilityHighly repeatable
cT1+/- 60ms+/- 60ms
T2*+/- 7ms+/- 7ms
PDFF+/- 1%+/- 1%
ReproducibilityReproducible between systems (at same field strength for T2*)
cT1+/- 120ms+/- 120ms
T2*+/- 10ms+/- 10ms (between systems at same field strength)
PDFF+/- 2%+/- 2% (between systems)
Intra-operator VariationWell within proscribed acceptance criteria
cT1+/- 18ms+/- 18ms
T2*+/- 3ms+/- 3ms
PDFF+/- 1%+/- 1%
Inter-operator VariationMinor additional variation compared to intra-operator
cT1+/- 25ms+/- 25ms
T2*+/- 4ms+/- 4ms
PDFF+/- 1.1%+/- 1.1%
Equivalence to LMSv1
cT1Within 40ms of predicate-14.8 to 40ms
T2*Within 1ms of predicate-0.5 to 0.8ms
PDFFNegligible difference-0.4 to 0.2%

Conclusion: The document states that "all acceptance criteria were met," and LMSv2 is concluded to be substantially equivalent to LMSv1.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size:
    • In Vivo: The document mentions "in vivo volunteer data" but does not explicitly state the number of volunteers used for the test set.
    • Phantom: The document mentions "phantom scans" but does not explicitly state the number of phantoms or scans.
    • Substantial Equivalence (In Vivo): The document mentions "in vivo measurements" show specific equivalence values, implying a test set was used, but the size is not specified.
    • Substantial Equivalence (Phantom): Similarly, "phantom measurements" are cited for equivalence, but the sample size is not specified.
  • Data Provenance: The document does not specify the country of origin of the data nor whether the studies were retrospective or prospective. It only mentions "volunteer scans" and "phantom scans."

3. Number of Experts and Qualifications for Ground Truth

The document does not provide details on the number of experts, their specific qualifications, or their role in establishing ground truth for the test set. The device is intended to be used by trained operators, and reports interpreted by trained clinicians (radiologists, gastroenterologists, hepatologists).

4. Adjudication Method

The document does not describe any specific adjudication method (e.g., 2+1, 3+1) used for the test set. It does mention "intra-operator" and "inter-operator" variation, indicating that multiple readings by the same operator and different operators were performed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No explicit MRMC comparative effectiveness study is described where human readers' improvement with AI vs. without AI assistance is quantified. The evaluation focuses on the performance of the software itself and the variability introduced by operators using the software.

6. Standalone Performance Study

Yes, a standalone performance study was done for the algorithm (LiverMultiScan v2). The entire performance data presented in the tables (accuracy, repeatability, reproducibility, equivalence) represents the algorithm's performance. The "intra-operator" and "inter-operator" studies also demonstrate the variability introduced by the human-in-the-loop when using the standalone software.

7. Type of Ground Truth Used

  • Phantom Studies: For phantom studies, the ground truth for T1 was based on "literature-reported underestimation of ground truth T1 using MOLLI techniques." For T2* and PDFF, the phantoms likely had known, controlled values for these parameters, which is implied by the term "ground truth" and the quantitative comparisons.
  • In Vivo Studies: The document does not explicitly state how ground truth was established for "in vivo volunteer data." Given the measurements are for MRI parameters (cT1, T2*, PDFF), the "ground truth" for repeatability and reproducibility would likely be the measurements themselves when repeated under identical or varied conditions, rather than an external gold standard like pathology, unless otherwise specified. For accuracy, a reference standard might be implied, but it's not detailed.

8. Sample Size for the Training Set

The document does not specify the sample size used for the training set.

9. How Ground Truth for the Training Set was Established

The document does not describe how ground truth was established for the training set. It focuses on the validation and verification against user needs and intended use, and comparison to a predicate device, rather than the development or training of the algorithm itself.

{0}------------------------------------------------

November 21, 2017

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Perspectum Diagnostics Ltd. % Jaco Jacobs Chief Compliance Officer 23-38 Hythe Bridge Street Oxford. Oxfordshire OX1 2ET UNITED KINGDOM

Re: K172685

Trade/Device Name: LiverMultiScan Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: September 1, 2017 Received: September 6, 2017

Dear Jaco Jacobs:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice

(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Sincerely,

Michael D. O'Hara

For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172685

Device Name LiverMultiScan (LMSv2)

Indications for Use (Describe)

LiverMultiScan (LMSv2) is indicated for use as a magnetic device software application for noninvasive liver evaluation that enables the generation, display and review of 2D magnetic resonance medical image data and pixel maps for MR relaxation times.

LiverMultiScan (LMSv2) is designed to utilize DICOM 3.0 compliant magnetic resonance image datasets, acquired from compatible MR Systems, to display the internal structure of the abdomen including the liver. Other physical parameters derived from the images may also be produced.

LiverMultiScan (LMSy2) provides a number of quantification tools, such as Region of Interest (ROI) placements, to be used for the assessment of regions of an image to quantify liver tissue characteristics, including the determination of triglyceride fat fraction in the liver, T2* and iron-corrected T1 measurements.

These images and the physical parameters derived from the images, when interpreted by a trained clinician, yield information that may assist in diagnosis.

Type of Use (Select one or both, as applicable)
✖ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92.

1. Submitter

Name:Perspectum Diagnostics Ltd
Address:Perspectum Diagnostics Ltd23-38 Hythe Bridge StreetOxfordOX1 2ETUnited Kingdom
Contact Person:Dr. Jaco Jacobs
Date Prepared:1st September 2017
2. Subject DeviceName of Device:LiverMultiScan (LMSv2)
Classification Name:Magnetic Resonance Diagnostic Device
Regulation:21 CFR 892.1000
Regulatory Class:Class II
Product Classification Code:LNH
3. Predicate DeviceName of predicate device:LiverMultiScan (LMSv1)
Predicate Manufacturer:Mirada Medical LtdOxford Centre for InnovationNew RoadOxfordOX1 1BYUnited Kingdom
Predicate Trade Name:LiverMultiScan

4. Reference Device

No reference devices were used in this submission.

5. Device Description

LiverMultiScan (LMSv2) is a standalone software application for displaying 2D Magnetic Resonance medical image data acquired from compatible MR Scanners. LiverMultiScan runs on a general purpose workstation with a colour monitor, keyboard and mouse.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains the logo for Perspectum Diagnostics. The word "Perspectum" is written in a bold, dark gray font, with the word "Diagnostics" written in a lighter gray font underneath. To the right of the text is a circular graphic with four colored sections: blue, yellow, green, and pink. The pink section is split in half by a white line.

LiverMultiScan (LMSv2) is designed to allow the review of DICOM 3.0 compliant datasets stored on the workstation and the operator may also create, display, print, store and distribute reports resulting from interpretation of the datasets.

LiverMultiScan (LMSv2) allows the display and comparison of combinations of magnetic resonance images and provides a number of tools for the quantification of magnetic resonance images, including the determination of triglyceride fat fraction in the liver, T2* and iron-corrected T1 measurements.

LiverMultiScan (LMSv2) provides a number of tools, such as circular region of interest placements, to be used for the assessment of regions of an image to support a clinical workflow.

LiverMultiScan (LMSv2) allows the operator to create relaxometry parameter maps of the abdomen which can be used by clinicians to help determine different tissue characteristics to support a clinical workflow. Examples of such workflows include, but are not limited to, the evaluation of the presence or absence of liver fat.

LiverMultiScan (LMSv2) is intended to be used by trained operators. Reports generated by trained operators are intended for use by interpreting clinicians, including, but not limited to radiologists, gastroenterologists, and hepatologists.

LiverMultiScan (LMSv2) is an aid to diagnosis. When interpreted by a trained clinician, the results provide information, which may be used as an input into existing clinical procedures and diagnostic workflows.

LiverMultiScan (LMSv2) offers the following.

  • ു Advanced visualisation of MR data.
  • C Processing of MR data to quantify tissue characteristics including MR relaxivity constants such as T2*, T1, ironcorrected T1 (cT1) and triglyceride fat fraction (expressed as liver fat percentage).
  • Circular region of interest statistics.
  • Snapshot of images to include in a report.
  • Report to include region statistics, snapshot images and user-entered text.
  • Export of snapshot images to report.

6. Indications for Use

"LiverMultiScan (LMSv2) is indicated for use as a magnetic resonance diagnostic device software application for noninvasive liver evaluation that enables the generation, display and review of 2D magnetic resonance medical image data and pixel maps for MR relaxation times.

LiverMultiScan (LMSv2) is designed to utilize DICOM 3.0 compliant magnetic resonance image datasets, acquired from compatible MR Systems, to display the internal structure of the abdomen including the liver. Other physical parameters derived from the images may also be produced.

LiverMultiScan (LMSv2) provides a number of quantification tools, such as Region of Interest (ROI) placements, to be used for the assessment of regions of an image to quantify liver tissue characteristics, including the determination of triglyceride fat fraction in the liver, T2* and iron-corrected T1 measurements.

These images and the physical parameters derived from the images, when interpreted by a trained clinician, yield information that may assist in diagnosis."

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Perspectum Diagnostics. The word "Perspectum" is written in a bold, dark gray font. Below it, the word "Diagnostics" is written in a lighter gray font. To the right of the text is a circular logo with four colored sections: yellow, blue, green, and pink.

7. Comparison to Predicate

The following characteristics were compared between the subject device and the predicate device in order to demonstrate substantial equivalence.

Overview

The subject device is substantially equivalent to the predicate device, both regulated under regulation 21 CFR 892.1000, Magnetic Resonance Diagnostic Device.

  • C The indications for use and intended uses of both the subject device and predicate device are equivalent.
  • The subject device and predicate device both support multi-slice MR data acquired using the specific acquisition protocols, from supported MR Systems, to acquire the input data.
  • The subject and predicate devices include software applications which utilise MR data to visualise and enable quantification of physiological characteristics in the liver to provide measurements which may be used to aid diagnosis.
  • C The subject and predicate devices are both standalone software applications to facilitate the import and visualization of MR data sets.
  • 3 Both the subject and predicate devices are designed to run on general-purpose computers.
  • G Both the subject device and the predicate device include applications to facilitate the import and visualization of MR data sets and include tools to enable the manipulation of the views and to enable the quantification and analysis of tissue characteristics in the liver from the MR data.
  • a The subject and predicate devices enable the quantification of analysis of tissue characteristics in the liver from the MR data.
  • The subject and predicate devices use the same algorithmic specifications to support quantitative analysis.
  • G The subject and predicate devices both support region of interest measurements derived from MR images and parametric maps of tissue characteristics.
  • The subject and predicate devices facilitate the creation of a medical report containing the images and analysis output derived from quantification of liver tissue parameters.
  • G The subject and predicate device reports all include tabular display of quantification statistics, parametric map images and include normal range references.
  • © Performance testing demonstrates that the subject device performs at least as safely and effectively as the proposed predicate.
CharacteristicLMSv2 (Subject)LMSv1 (Predicate)
510(k)Not knownK143020
Regulation21 CFR 892.100021 CFR 892.1000
Regulatory ClassClass IIClass II
Product CodeLNHLNH
ManufacturerPerspectum Diagnostics LtdMirada Medical Ltd
Intended Use andIndications for UseLiverMultiScan is indicated for use asmagnetic resonance diagnostic devicesoftware application for non-invasive liverevaluation that enables the generation,display and review of 2D magnetic resonancemedical image data and pixel maps for MRrelaxation times.LMS is indicated for use as magneticresonance diagnostic device softwareapplication for non-invasive liver evaluationthat enables the generation, display andreview of 2D magnetic resonance medicalimage data and pixel maps for MR relaxationtimes.
LiverMultiScan is designed to utilize DICOM3.0 compliant magnetic resonance imagedatasets, acquired from compatible MRSystems, to display the internal structure ofthe abdomen including the liver. Otherphysical parameters derived from the imagesmay also be produced.LiverMultiScan provides a number ofquantification tools, such as Region ofInterest (ROI) placements, to be used for theassessment of regions of an image toquantify liver tissue characteristics, includingthe determination of triglyceride fat fractionin the liver, T2* and iron-corrected T1measurements.LMS is designed to utilize DICOM 3.0compliant magnetic resonance imagedatasets, acquired from Siemens MAGNETOMSkyra MR Scanners, to display the internalstructure of the abdomen including the liver.Other physical parameters derived from theimages may also be produced.LMS provides a number of quantification toolssuch as rulers and region of interest to beused for the assessment of regions of animage to support existing clinical workflows.
These images and the physical parametersderived from the images, when interpretedby a trained clinician, yield information thatmay assist in diagnosis.These images and the physical parametersderived from the images, when interpreted bya trained physician, yield information that mayassist in diagnosis.
Target PopulationPatients suitable to undergo an MRI scan andnot contra-indicated for MRI.Patients suitable to undergo an MRI scan andnot contra-indicated for MRI.
Device UserTrained operatorTrained clinician
Report UserInterpreting clinicianInterpreting clinician
AnatomicalLocationAbdomen, LiverAbdomen, Liver
EnergyConsiderationsSoftware only application. The device, astandalone software application, does notdeliver or depend on energy delivered to orfrom patients.Software only application. The device, astandalone software application, does notdeliver or depend on energy delivered to orfrom patients.
Design: PurposeStandalone software application to facilitatethe import and visualization of MR data setsencompassing the abdomen, including theliver with functionality independent of theMRI equipment vendor.Software application intended to display andvisualize 2D multi-slice, spin-echo MR datasets encompassing the abdomen. The usermay process, and review DICOM 3.0compliant datasets within the system and/oracross computer networks.Standalone software application to facilitatethe import and visualization of MR data setsencompassing the abdomen, including theliver with functionality independent of theMRI equipment vendor.Software application intended to display andvisualize 2D multi-slice, spin-echo MR datasets encompassing the abdomen. The usermay process, and review DICOM 3.0 compliantdatasets within the system and/or acrosscomputer networks.
Design: MRRelaxometryT1, iron-corrected T1 (cT1) and T2* mapping.T1, iron-corrected T1 (cT1) and T2* mapping.
Design: Liver FatQuantificationUtilizes MR images that exploit thedifference in resonance frequenciesbetween hydrogen nuclei in water andtriglyceride fat using the three point DIXONmethod.Utilizes MR images that exploit the differencein resonance frequencies between hydrogennuclei in water and triglyceride fat using thethree point DIXON method.
Design: Parametric MapsIron corrected T1 (cT1), T2* and triglyceride fat (also known as Proton Density Fat Fraction (PDFF)) parametric maps are supported.Iron corrected T1 (cT1), T2* and triglyceride fat (also known as Proton Density Fat Fraction (PDFF)) parametric maps are supported.
Design: VisualisationIron corrected T1 (cT1) displayed using LMS colourmap, designed to have maximum contrast on liver parenchymal tissue.Iron corrected T1 (cT1) displayed using LMS grayscale colourmap.
Design: Region of Interest (ROI)Median and interquartile range measurements created from a cross sectional slice of liver tissue. For each parametric map, statistics from multiple Regions of Interest (ROIs) – potentially placed across multiple slices – are summarised.Mean, median, standard deviation, interquartile range measurements created from a cross sectional slice of liver tissue.
Design: Supported ModalitiesDICOM 3.0 compliant MR data from GRE Echo and MOLLI acquisition protocols.DICOM 3.0 compliant MR data from GRE Echo and MOLLI acquisition protocols.
Design: ReportingLiver quantification images and analysis collated in a report.Liver quantification images and analysis collated in a report.
PerformanceValidated with phantom scans, synthetic raw data and volunteer scans covering a range of physiological values for cT1, T2* and PDFF.Validated with volunteer and phantom scans.
Supported SystemsValidated across all supported manufacturers and field strengths.Validated on Siemens Skyra (3T).
StandardsIEC 62304, DICOM 3.0, ISO 14971, ISO 13485IEC 62304, DICOM
Operating SystemMac OSMicrosoft Windows
MaterialsNot applicable, standalone software.Not applicable, standalone software.
BiocompatibilityNot applicable, standalone software.Not applicable, standalone software.
SterilityNot applicable, standalone software.Not applicable, standalone software.
Electrical SafetyNot applicable, standalone software.Not applicable, standalone software.
Mechanical SafetyNot applicable, standalone software.Not applicable, standalone software.
Chemical SafetyNot applicable, standalone software.Not applicable, standalone software.
Thermal SafetyNot applicable, standalone software.Not applicable, standalone software.
Radiation SafetyNot applicable, standalone software.Not applicable, standalone software.

Device Comparison Table

{6}------------------------------------------------

Image /page/6/Picture/4 description: The image shows the logo for Perspectum Diagnostics. The word "Perspectum" is written in a bold, dark gray font. Below it, the word "Diagnostics" is written in a lighter gray font. To the right of the words is a colorful circular logo with yellow, blue, green, and pink colors.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Perspectum Diagnostics. The word "Perspectum" is in bold, dark gray font, and the word "Diagnostics" is in a lighter gray font and is located underneath the word "Perspectum". To the right of the words is a colorful circular logo that is divided into four sections: yellow, green, blue, and pink.

In conclusion, the subject device does not result in any new potential safety risk and performs in accordance with its intended use as well as comparatively with the intended use of the chosen predicate.

8. Performance Data

Introduction

Substantial equivalence performance testing will consist of a head-to-head evaluation of the subject and predicate devices' ability to measure (corrected) T1, T2* and PDFF using phantoms as well as in-vivo volunteer data.

For in-vivo performance data please see

For each of the comparisons, the following will be calculated:

  • ട Mean difference in T1/ T2* and PDFF (magnitude of the bias)
  • Standard deviation
  • ල Bland-Altman 95% limits of agreement

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for Perspectum Diagnostics. The word "Perspectum" is written in bold, dark gray letters. Below it, the word "Diagnostics" is written in a lighter shade of gray. To the right of the text is a circular graphic with four colored sections: yellow, green, blue, and pink.

ි Slope of bias

Phantom

The performance of LMSv2, as summarised in the tables below demonstrates that:

  • o The MOLLI-based T1 measurement produced by LMSv2 is consistent with the literature-reported underestimation of ground truth T1 using MOLLI techniques 1,2.
  • LiverMultiScan (LMSv2) measurement of T2* is accurate over the expected physiological range of values.
  • o LiverMultiScan (LMSv2) measurement of PDFF is accurate over the expected physiological range of values.
PhantommetricAccuracy95% CI Limits of AgreementPerformance metric
T119-25% lower than ground truth
T2*+/- 2ms
PDFF < 30%+/- 3%
PDFF ≥ 30%+/- 21%
  • LiverMultiScan (LMSv2) measurement of T1 is highly repeatable.
  • LiverMultiScan (LMSv2) measurement of T1 is both repeatable within the same system and reproducible between systems.
  • LiverMultiScan (LMSv2) measurement of T2* is highly repeatable.
  • o LiverMultiScan (LMSv2) measurement of T2* is both repeatable within the same system reproducible between systems at the same field strength.
  • LiverMultiScan (LMSv2) measurement of PDFF is highly repeatable.
  • o LiverMultiScan (LMSv2) measurement of PDFF is reproducible between systems.
PhantomRepeatability95% Cl Limits of AgreementReproducibility95% Cl Limits of Agreement
metricPerformance metricPerformance metric
T1+/- 10ms-34 to 27ms
T2*+/- 1.7ms+/- 4ms
PDFF < 30%-2.5 to 1 %+/- 4%
PDFF ≥ 30%+/- 5%+/-10 %

1 Piechnik et al. (2010) Journal of Cardiovascular Magnetic Resonance 12:69

2 Puntmann et al. (2013) Journal of Cardiovascular Magnetic Resonance 15(Suppl 1):E18

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image contains the logo for Perspectum Diagnostics. The word "Perspectum" is in bold, black font, and the word "Diagnostics" is in a smaller, gray font underneath. To the right of the words is a circular logo with four different colored sections: blue, yellow, green, and pink. There is a trademark symbol in the upper right corner of the logo.

In Vivo

The in vivo performance testing of LMSv2, as summarised in the table below demonstrates that:

  • LiverMultiScan (LMSv2) measurement of cT1 is highly repeatable. o
  • o LiverMultiScan (LMSv2) measurement of cT1 is reproducible between systems.
  • LiverMultiScan (LMSv2) measurement of T2* is highly repeatable.
  • o LiverMultiScan (LMSv2) measurement of T2* is reproducible between systems at the same field strength.
  • LiverMultiScan (LMSv2) measurement of PDFF is highly repeatable.
  • LiverMultiScan (LMSv2) measurement of PDFF is reproducible between systems.
VolunteermetricRepeatability95% Cl Limits of AgreementPerformance metricReproducibility95% CI Limits of AgreementPerformance metric
cT1+/- 60ms+/- 120ms
T2*+/- 7ms+/- 10ms
PDFF+/- 1%+/- 2%

As demonstrated in the table below, the variation introduced by operator reading (use of LMSv2) is well within the proscribed acceptance criteria. There is only minor additional variation introduced by having two readers read the same study compared to the same reader performing a second analysis blinded.

VolunteermetricIntra-operator95% CI Limits of AgreementPerformance metricInter-operator95% CI Limits of AgreementPerformance metric
cT1+/- 18ms+/- 25ms
T2*+/- 3ms+/- 4ms
PDFF+/- 1%+/- 1.1%

Substantial Equivalence

The substantial equivalence testing of LMSv2 versus its predicate device, LMSv1, is summarised in the tables below.

Phantom measurements show that LMSv2 produces T1 and T2* values within 1ms of the predicate (negligible difference). PDFF is within 1% of the predicate where PDFF < 30%, and within 2% at higher percentages.

MetricPhantom Equivalence95% CI Limits of Agreement
-----------------------------------------------------------

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the logo for Perspectum Diagnostics. The word "Perspectum" is written in a bold, sans-serif font in black. Below it, the word "Diagnostics" is written in a smaller, lighter gray font. To the right of the text is a circular graphic with four colored sections: yellow, green, blue, and pink. The graphic is divided horizontally, with the top half showing the yellow, green, and blue sections, and the bottom half showing the pink section.

Performance metric
T1-0.5 to 0.4ms
T2*-0.4 to 0.2ms
PDFF < 30%-0.6 to 0.4%
PDFF ≥ 30%-1.6 to 2%

In vivo measurements show that cT1 values produced by LMSv2 lie within 40ms of the predicate device, with T2* values within 1ms. There is negligible difference in the measurement of PDFF between LMSv2 and the predicate.

MetricIn vivo Equivalence95% CI Limits of Agreement
Performance metric
cT1-14.8 to 40ms
T2*-0.5 to 0.8ms
PDFF-0.4 to 0.2%

In summary, as all acceptance criteria were met, we conclude that LMSv2 is substantially equivalent to its predicate, LMSv1.

Summarv

In conclusion, performance testing demonstrates that LiverMultiScan (LMSv2) is substantially equivalent to, and performs at least as safely and effectively as the listed predicate device. In-vivo and phantom clinical performance criteria established were successfully met during the performance evaluation.

LiverMultiScan (LMSv2) meets requirements for safety and effectiveness and does not introduce any new potential safety risks.

9. Biocompatibility Testing

Not applicable as the LiverMultiScan (LMSv2) device is standalone software.

10. Electrical Safety and Electromagnetic Compatibility (EMC)

Not applicable as the LiverMultiScan (LMSv2) device is standalone software.

11. Software Verification and Validation Testing

LiverMultiScan is validated and verified against its user needs and intended use by the successful execution of planned software verification and validation testing included in this submission. The results of verification and validation testing demonstrate that LiverMultiScan meets the user needs and requirements of the device, which are demonstrated to be substantially equivalent to those of the listed predicate device.

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the logo for Perspectum Diagnostics. The word "Perspectum" is written in a bold, dark gray font, while the word "Diagnostics" is written in a lighter gray font and is positioned below the first word. To the right of the words is a circular logo with four different colored sections: yellow, blue, green, and pink.

Verification and Validation for LiverMultiScan has been carried out in compliance with the requirements of ISO 13485:2003, 21 CFR Part 820 and in adherence to the DICOM standard.

In conclusion, software verification and validation testing demonstrates that LiverMultiScan is substantially equivalent to, and performs at least as safely and effectively as the listed predicate device. LiverMultiScan meets requirements for safety and effectiveness and does not introduce any new potential safety risks.

12. Mechanical and Acoustic Testing

Not applicable as the LiverMultiScan (LMSv2) device is standalone software.

13. Animal Study

Animal performance testing was not required to demonstrate safety and effectiveness of the device.

15. Benchtop Performance Data

Please see Phantom and Substantial Equivalence sections under Performance Data.

14. Clinical Studies

Please see In Vivo and Substantial Equivalence sections under Performance Data.

15. Conclusion

In conclusion, Perspectum Diagnostics Ltd is of the opinion that:

  • 1 the subject device LiverMultiScan (LMSv2) performs in accordance with its intended use as well as comparatively with the intended use of the chosen predicate, LiverMultiScan (LMSv1)
  • G the subject device LiverMultiScan (LMSv2) is as safe and effective as the predicate device LiverMultiScan (LMSv1);
  • C the subject device LiverMultiScan (LMSv2) does not raise new questions of safety and effectiveness; and
  • the subject device LiverMultiScan (LMSv2) is substantially equivalent to the currently marketed predicate device LiverMultiScan (LMSv1).

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.