K Number
K172682

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-10-06

(30 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Edge is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic system in all general purpose diagnostic procedures. Not to be used for mammography.

Device Description

Edge digital flat panel X-ray detector is based on flat-panel technology. This radiographic image detector and processing unit consists of a scintillator coupled to an a-Si TFT sensor. This device needs to be integrated with a radiographic imaging system. It can be utilized to capture and digitalize X-ray images for radiographic diagnosis.

The RAW files can be further processed as DICOM compatible image files by separate console SW (K160579 / XmaruView V1 and XmaruPACS/ Rayence Co.,Ltd.) for a radiographic diagnosis and analysis.

AI/ML Overview

The provided text describes the 510(k) submission for the "Edge Digital Flat Panel X-ray Detector" and its substantial equivalence to a predicate device, the "1717SCC_140µm". The performance testing described primarily focuses on demonstrating this equivalence through non-clinical (bench) testing and a clinical consideration report.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of "acceptance criteria" for the device's performance in the typical sense of a target value that must be met for approval. Instead, the acceptance criteria for the study proving substantial equivalence appears to be based on the similarity of performance characteristics to the predicate device.

The performance characteristics used for comparison are:

CharacteristicAcceptance Criteria (Implicit: Similar/Not Inferior to Predicate)Reported Device Performance (Edge)Predicate Device Performance (1717SCC_140µm)Remarks
Intended UseSameDigital imaging solution for general radiographic system for human anatomy; replace film or screen-based systems; not for mammography.Digital imaging solution for general radiographic system for human anatomy; replace film or screen-based systems; not for mammography.Same
Detector TypeSameAmorphous Silicon, TFTAmorphous Silicon, TFTSame
ScintillatorSameCsI:TlCsI:TlSame
Imaging AreaSame17 x 17 inches (3072 x 3072 matrix)140 type : 3072 x 3072Same
Pixel MatrixSame3072 x 30723072 x 3072Same
Pixel PitchSame140 µm140 µmSame
ResolutionSimilar3.9 lp/mm3.9 lp/mmSame
A/D conversionSame14 / 16 bit14 / 16 bitSame
Preview TimeSame or better≤2≤2Same
MTF (@1 lp/mm)Similar / Not inferior59.1 (%)58.2 (%)Similar (Edge marginally better)
DQESimilar / Not inferior76 (%)74 (%)Similar (Edge marginally better)
Data OutputSameRAW (convertible to DICOM 3.0)RAW (convertible to DICOM 3.0)Same
Imaging SoftwareSameXmaru View 1 / Xmaru PACS (Version 2.0.0)Xmaru View 1 / Xmaru PACS (Version 2.0.0)Same
DimensionsSame460 x 460 x 15.5 mm460 x 460 x 15.5 mmSame
WeightSame4 kg4 kgSame
ApplicationSameGeneral Radiology system or Portable systemGeneral Radiology system or Portable systemSame
Clinical Image QualitySuperior or equivalent to predicateSuperior to predicate in spatial and soft tissue contrast resolution.-Clinical consideration by expert

2. Sample size used for the test set and the data provenance

The document states: "To further demonstrate the substantial equivalency of two devices, clinical images are taken from both subject devices and reviewed by a licensed US radiologist to render an expert opinion."

  • Sample Size: Not explicitly stated as a number of images or patients. It mentions "sample radiographs of similar age group and anatomical structures". This suggests a limited sample size rather than a large clinical trial.
  • Data Provenance:
    • Country of Origin: US (implied by "licensed US radiologist").
    • Retrospective or Prospective: Not explicitly stated, but the description "clinical images are taken from both subject devices and reviewed" and "sample radiographs... taken" suggests these were specifically acquired for this comparison, implying a prospective data collection for the purpose of the study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Number of Experts: "a licensed US radiologist" - One (1) expert.
  • Qualifications: "licensed US radiologist". No specific years of experience or subspecialty are provided.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Adjudication Method: None mentioned. The single radiologist's opinion appears to be the sole basis for the clinical image quality assessment ("reviewed by a licensed US radiologist to render an expert opinion").

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No, an MRMC study was not conducted. This study's purpose was to demonstrate substantial equivalence between two X-ray detectors, not to evaluate AI assistance for human readers. The device itself is an X-ray detector, not an AI algorithm for improving reader performance.
  • Effect Size: Not applicable as no such study was performed or needed for this device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Standalone Performance: The device is an X-ray detector. Its performance is standalone in terms of image acquisition (MTF, DQE, etc.). The "clinical consideration" involved human review of the images produced by the device, but this was to compare the image quality of the device to its predicate, not to assess an AI's diagnostic performance without a human. The non-clinical tests (MTF, DQE) are indeed "standalone" measures of the detector's physical performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • The term "ground truth" isn't explicitly used in the context of diagnostic findings. For the non-clinical tests (MTF, DQE), the "ground truth" is established by adherence to IEC 62220-1 standards for objective physical performance measurements.
  • For the "clinical consideration" part, the "ground truth" for image quality comparison was the expert opinion of a single licensed US radiologist. This is a subjective assessment of image characteristics ("spatial and soft tissue contrast resolution are superior"). There's no mention of a diagnostic "ground truth" (e.g., presence/absence of disease confirmed by pathology or other means) being established for the clinical images used for review. The focus was on comparing the quality of the images produced by the devices, not on the diagnostic accuracy of those images for specific conditions.

8. The sample size for the training set

  • This information is not applicable and not provided. The "Edge Digital Flat Panel X-ray Detector" is an imaging hardware device, not an AI algorithm that requires a training set. The clinical consideration involved comparing images produced by two hardware devices.

9. How the ground truth for the training set was established

  • This information is not applicable and not provided as the device is not an AI algorithm requiring a training set.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 6, 2017

OSKO, INC. % Mr. Dave Kim Medical Device Regulatory Affairs Mtech Group 8310 Buffalo Speedway HOUSTON TX 77025

Re: K172682

Trade/Device Name: Edge Digital Flat Panel X-ray Detector Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: August 29, 2017 Received: September 6, 2017

Dear Mr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D. O'Hara For

Robert Ochs, Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K172682

Device Name Edge Digital Flat Panel X-ray Detector

Indications for Use (Describe)

Edge is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic system in all general purpose diagnostic procedures. Not to be used for mammography.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

1. Traditional 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date 510K summary prepared: September 29, 2017 Submitter's Name, address, telephone number, a contact person:

Submitter's Name :OSKO, INC.
Submitter's Address:8085 NW 90th Street, Miami, Florida 33166, USA
Submitter's Telephone:+1-305-599-7161
Contact person:Vivian Garcia / RA Manager / +1-305-599-7161
Official Correspondent:Dave Kim (davekim@mtech-inc.net) (U.S. Agent)
Address:8310 Buffalo Speedway, Houston, TX 77025
Telephone:+713-467-2607
Fax:+713-464-8880

Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known:

Trade/proprietary name:Edge
Common Name: Digital Flat Panel X-ray Detector
Classification Name: 21CFR892.1680 / Stationary x-ray system
Product Code: MQB
Classification: Class II
Predicate Device :
Manufacturer: Rayence Co., Ltd
Device: 1717SCC_140µm
510(k) Number: K171420
Classification Name: 21CFR892.1680 / Stationary x-ray system
Product Code: MQB

{4}------------------------------------------------

2. Device Description :

Edge digital flat panel X-ray detector is based on flat-panel technology. This radiographic image detector and processing unit consists of a scintillator coupled to an a-Si TFT sensor. This device needs to be integrated with a radiographic imaging system. It can be utilized to capture and digitalize X-ray images for radiographic diagnosis.

The RAW files can be further processed as DICOM compatible image files by separate console SW (K160579 / XmaruView V1 and XmaruPACS/ Rayence Co.,Ltd.) for a radiographic diagnosis and analysis.

3. Indication for use :

Edge is indicated for digital imaging solution designed for general radiographic system for human It is intended to replace film or screen based radiographic systems in all general anatomy. purpose diagnostic procedures. Not to be used for mammography.

4. Summary of the Design Control Risk Management:

The Edge digital X-ray detector is a modification of 1717SCC 140 um (K171420) which was developed for the purpose of retrofitting the stationary a film X-ray system. Edge is connected to a PC with an ethernet cable.

The risks and the hazardous impact of the device modification were analyzed with FMEA method. The specific risk control and protective measures to mitigate the risks from the modification were reviewed and implemented in the new product design phase. The overall assessment concluded that all risks and hazardous conditions identified arising from the design change were successfully mitigated and accepted.

The Edge digital solid state X-ray detector uses the same panel and components that the predicate device, 1717SCC_140 µm (K171420) and that no changes were necessary to eiter the hardware or firmware of the device.

5. Summary of the technological characteristics of the device compared to the predicate device:

The Edge digital solid state X-ray detector described in this 510(k) has the same indications for use and similar technical characteristics as its predicate device, 1717SCC_140 um (K171420) flat panel detector, of Rayence Co., Ltd

The Edge digital solid state X-ray detector described in this 510(k) uses the same hardware, firmware and imaging software Xmaru View 1 and Xmaru PACS (K160579), without any change, that are used by the predicate device, 1717SCC_140 um.

{5}------------------------------------------------

Xmaru View 1 FDA 510(k) information is as follows:
------------------------------------------------------
ItemDeviceClassificationNameDevice name510(k)numberApplicant
FDA510(k)System, ImageProcessing,RadiologicalXmaru View V1 andXmaru PACS, Medicalimage processingsoftwareK160579Rayence Co.,Ltd.

5.1 Scintillator layer

*scintillator layer. ( * scintillator : a phosphor that produces scintillations)

ProposedPredicate
CsI(Cesium Iodide)Edge1717SCC_140 \u03bcm

5.2 Dimensions and Weight

ProposedEdgePredicate1717SCC_140 $ \mu $ m
DetectorModel nameEdge1717SCC_140 $\mu$ m
W x L x H460 X 460 X 15.5460 X 460 X 15.5
Weight4 kg4 kg
PowersupplyModel nameRS1717RS1717
W x L x H188 X 92 X 41.5188 X 92 X 41.5
Weight0.5 kg0.5 kg

5.3 Power Requirements (Power supply)

ProposedEdgePredicate1717SCC_140 μm
PowersupplyModel nameRS1717RS1717
W x L x HAC 100 - 240 V~,50/60HzAC 100 - 240 V~,50/60Hz

{6}------------------------------------------------

ModelManufactureSpecification
32kW40kW50kW
CMP 200Communications &Power IndustrieskVp40-12540-150
mA10-40010-50010-630
EDITOR HFe 501Rontgenwerk BochumkVp40-150
mA10-630
UD150L-40E/40FShimadzukVp40-150
mA@100 kVp- 500(320)
mA@80 kVp- 630(400)
PXR-321BPoskom Co.,Ltd.kVp125/150
mA500

5.4 Recommended Generator Specification

Image /page/6/Picture/3 description: The image shows a yellow caution sign. The sign is triangular in shape and has a black border. Inside the triangle is a black exclamation point. Below the triangle is the word "CAUTION" in black, block letters.

To our best knowledge, the detector is compatible with the X-ray . generators with the specifications described above.

▪ If you have questions regarding the compatibility issue for other generators which are not listed above, please contact your Rayence representative. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

This information will be included in user/installation documents for the end user.

5.5 Comparison table

CharacteristicProposedEdgePredicate1717SCC_140 $\mu$ mRemarks
510(k)numberK172682K171420
IntendedUseEdge X-ray detectors, 140um, areindicated for digital imagingsolution designed for generalradiographic system for humananatomy. It is intended to replacefilm or screen based radiographicsystems in all general purposediagnostic procedures. Not to beused for mammography.1717SCC_127um and1717SCC_140um, areindicated for digital imagingsolution designed for generalradiographic system forhuman anatomy. It is intendedto replace film or screenbased radiographic systems inall general purpose diagnosticprocedures. Not to be used formammography.Same
ManufacturerOSKO, INC.Rayence Co., Ltd.

{7}------------------------------------------------

DetectorTypeAmorphous Silicon, TFTAmorphous Silicon, TFTSame
ScintillatorCsI:TlCsI:TlSame
ImagingArea17 x 17 inches140 type : 3072 x 3072Same
Pixel matrix3072 x 30723072 x 3072Same
Pixel pitch140 μm140 μmSame
Resolution3.9 lp/mm3.9 lp/mmSame
A/Dconversion14 / 16 bit14 / 16 bitSame
Previewtime≤2≤2Same
MTF (@11p/mm)59.1 (%)58.2 (%)Similar
DQE76 (%)*Result value from Non-ClinicalReport74 (%)*Result value from Non-Clinical ReportSimilar
Data outputRAW*The RAW files are convertibleinto DICOM 3.0 by console S/WRAW*The RAW files areconvertible into DICOM 3.0by console S/WSame
ImagingSoftwareXmaru View 1 / Xmaru PACS(Version 2.0.0)Xmaru View 1 / XmaruPACS (Version 2.0.0)Same
Dimensions460 × 460 × 15.5 mm460 × 460 × 15.5 mmSame
Weight4 kg4 kgSame
ApplicationGeneral Radiology system orPortable systemAvailable with upright stand,table, universal stand.General Radiology system orPortable systemAvailable with upright stand,table, universal stand.Same

The Edge Digital Flat Panel Detector is identical to the wireless hardware and functionality ● of the predicate device, 1717SCC 140 µm.

Maximum wireless signal rate derived from IEEE standard specifications. Actual data throughput will vary. Network conditions and environmental factors, including volume of

{8}------------------------------------------------

network traffic, building materials and construction, and network overhead, lower actual data throughput rate.

6. Summary of Performance Testing:

The Edge Digital Flat Panel X-Ray Detector has the same indications for use, material, form factor, performance, and safety characteristics compared to the predicate devices, 1717SCC 140 um (K171420) respectively.

The non-clinical test report and clinical consideration report for each subject device were prepared and submitted to FDA separately to demonstrate the substantial equivalency of the subject devices compared to each respective predicate device. The non-clinical test report contains the MTF, DQE and NPS test results of Edge by using the identical test equipment and same analysis method described by IEC 62220-1.

The comparative result of the MTF and DQE test for Edge detector with respect to each respective predicate demonstrated that the MTF and DQE of the subject device performed similarly compared with the predicate device. The MTF and DQE represents the ability to visualize object details of a certain size and contrast. Edge has similar MTF and DQE performance at all spatial frequencies. The comparison of the MTF and DOE for Edge detector demonstrated that it performed almost same 1717SCC 140 um (K171420).

To further demonstrate the substantial equivalency of two devices, clinical images are taken from both subject devices and reviewed by a licensed US radiologist to render an expert opinion. Both the test subject (Edge) and the predicate device (1717SCC 140 um (K171420)) have been evaluated and compared by taking sample radiographs of similar age group and anatomical structures in accordance with the test protocol of diagnostic radiography evaluation procedure.

After a broad review of plain radiographic images taken with Edge, the images obtained with Edge are superior to the same view obtained from a similar patient with the predicate devices, 1717SCC 140 µm (K171420). In general, both the spatial and soft tissue contrast resolution are superior using Edge. Specifically, the soft tissues on extremity films were seen with better clarity.

Based on the non-clinical and clinical consideration test and the outcome of a comparative

{9}------------------------------------------------

review by an expert for both devices, the sponsor can claim the substantial equivalency between the subject devices and their predicate devices in terms of diagnostic image quality. The manufacturing facility is in conformance with the design control procedure requirements and the relevant EPRC standards as specified in 21 CFR 802.30 and the records are available for review.

Note: Clinical images were provided; even though these images are not necessary to establish substantial equivalence based on the modifications to the device, they provide additional evidence in addition to bench testing data to show that the complete system works as intended.

7. Summary for any testing in the submission:

Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1:2005 (3rd Edition) + CORR. 1:2006 + CORR. 2:2007 + A1:2012 (Medical electrical equipment Part 1: General requirements for basic safety and essential performance) was performed, and EMC testing were conducted in accordance with standard IEC 60601-1-2: 2007.

The Non-clinical consideration & performance test have been conducted according to the following FDA Guidance.

  • Performance Standards for Ionizing Radiation Emitting Products, 21 CFR 1020

  • Diagnostic x-ray systems and their major components, 21 CFR 1020.30

  • Radiographic Equipment, 21 CFR 1020.31

  • FDA Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices

  • FDA Guidance titled "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices"

All test results were satisfactory.

8. Conclusions :

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification OSKO, INC. concludes that Edge is substantially equivalent in comparison with 1717SCC_140μm (K171420) the predicate device as described herein.

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.