K Number
K172610

Validate with FDA (Live)

Date Cleared
2018-04-17

(229 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The THUNDERBEAT Type S hand instruments are intended to be used with the Ultrasonic Generator (USC-400), the Electrosurgical Generator (ESG-400), and the THUNDERBEAT Transducer (TD-TB400).

Seal & Cut mode:
The THUNDERBEAT Type S hand instruments when used in combination with the Seal & Cut mode are indicated for open, laparoscopic (including single-site surgery and gynecological surgery (including urologic, thoracic, plastic and reconstructive, bowel resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectomy, hysterectomies (both vaginal assisted and abdominal) etc.), and endoscopic surgery or in any procedure in which cutting, vessel ligation (sealing and cutting), coagulation, grasping, and dissection is performed. These devices have been designed to seal and cut vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics.

Seal mode:
The THUNDERBEAT Type S hand instruments when used in combination with the Seal mode are indicated for open, laparoscopic (including single-site surgery) general surgery (including urologic, thoracic, plastic and reconstructive, bowel resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectomy, hysterectomies (both vaginal assisted and abdominal) etc.), and endoscopic surgery or in any procedure in which vessel sealing, coagulation, grasping is performed. These devices have been designed to seal vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics.

The THUNDERBEAT Type S hand instruments have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.

Device Description

The THUNDERBEAT Type S Hand Instruments is intended to be used with the Ultrasonic Generator (USG-400), the Electrosurgical Generator (ESG-400), and the THUNDERBEAT Transducer, (TD-TB400).

The THUNDERBEAT Type S Hand Instruments is provided as a sterile, single use device. It is a functional device capable of vessel sealing & cutting, tissue coagulating & cutting, grasping, and dissecting. This device has been designed to seal and cut vessels up to and including 7 mm in diameter, tissue bundles, and lymphatics.

AI/ML Overview

This document is a 510(k) Summary for the THUNDERBEAT Type S Hand Instruments, which are electrosurgical cutting and coagulation devices. It compares the new device to a predicate device and provides performance testing summaries to demonstrate substantial equivalence.

Here's an analysis of the provided information concerning acceptance criteria and the study that proves the device meets them:

1. Acceptance Criteria and Reported Device Performance

The document doesn't present specific numerical acceptance criteria in a table format with corresponding reported performance values explicitly labeled as "acceptance criteria met". Instead, it describes performance tests designed to compare the new device (THUNDERBEAT Type S) to a predicate device (THUNDERBEAT K132703) to demonstrate substantial equivalence. The implication is that the new device's performance results were "similar" or "demonstrated substantial equivalence" to the predicate, which serves as the de facto acceptance benchmark.

Here's a table based on the performance testing summarized, with the understanding that "acceptance criteria met" implies performance comparable/substantially equivalent to the predicate device. The document does not provide quantitative results for the performance tests, only that they were conducted.

Acceptance Criteria (Implied: Performance Comparable/Substantially Equivalent to Predicate)Reported Device Performance (Qualitative Summary)
Bench Testing:
Ex-vivo Vessel Burst Pressure (vessel sealing performance)Conducted on both subject and predicate devices to demonstrate vessel sealing performance. (Implied: Comparable performance to predicate).
Ex-vivo Cutting Performance (Cutting time)Conducted on both subject and predicate devices to demonstrate cutting performance. (Implied: Comparable performance to predicate).
Max Temperature (inside/outside temperature at grasping section)Conducted on both subject and predicate devices to demonstrate for the inside/outside temperature at the grasping section. (Implied: Comparable performance to predicate).
Grasping force (homogeneousness of grasping force)Conducted on both subject and predicate devices to demonstrate homogeneousness of grasping force across the grasping section. (Implied: Comparable homogeneousness to predicate).
Durability of the coating on underside of probe tipConducted to confirm durability of coating on underside of probe tip after device output. (Implied: Acceptable durability).
Durability against twisting (torque when spark occurs)Conducted on both subject and predicate devices to demonstrate torque when spark occurs between probe and jaw during activation. (Implied: Comparable durability to predicate).
Animal Testing:
Chronic Animal Study (seal performance, thermal spread, healing degree for 7.0mm vessels, lymphatics, tissue bundles)Conducted on both subject and predicate devices to confirm seal performance, thermal spread, and healing degree. (Implied: Acceptable and comparable performance to predicate).
Acute Animal Study (seal performance, thermal spread, degeneration degree for 7.0mm vessels, lymphatics, tissue bundles)Conducted on both subject and predicate devices to confirm seal performance, thermal spread, and degeneration degree. (Implied: Acceptable and comparable performance to predicate).
Biocompatibility Testing:Compliance with ISO-10993 series.
Sterilization:Compliance with ISO11135:2014.
Shelf Life:Three-year shelf life.

2. Sample Sizes Used for the Test Set and Data Provenance

The document mentions ex-vivo and animal studies using porcine models. However, it does not specify the sample sizes (e.g., number of vessels, number of animals) used for these tests. The data provenance is described as:

  • Country of origin: Not explicitly mentioned for the data itself, but the manufacturer and associated entities are in Japan and Massachusetts, USA. The animal studies were likely conducted in a research facility.
  • Retrospective or prospective: The studies described (bench and animal testing) are inherently prospective in nature, as they involve actively conducting tests with the devices.

3. Number of Experts and Qualifications for Ground Truth

This type of submission for an electrosurgical device does not typically involve human expert interpretation of images or data for ground truth establishment in the same way an AI-powered diagnostic device would. The "ground truth" for the performance tests (e.g., burst pressure, cutting time, temperature, seal performance, thermal spread, healing) would be established by objective measurements and observations conducted by trained personnel using standardized protocols, rather than expert consensus on a subjective measure. Therefore, the concept of "number of experts" and their "qualifications" for establishing ground truth as seen in AI/imaging studies is not directly applicable here.

4. Adjudication Method for the Test Set

As the ground truth is based on objective measurements and observations, an adjudication method like 2+1 or 3+1 (common in image interpretation studies) is not applicable. The data would likely be collected and analyzed according to predefined statistical methods and quality control procedures.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI devices assisting human readers in diagnostic tasks (e.g., radiologists interpreting images). The THUNDERBEAT device is a surgical instrument, not a diagnostic AI system, so an MRMC study would be irrelevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

N/A. This question is highly relevant for AI/software devices. The THUNDERBEAT is a physical electrosurgical device; it does not have a standalone "algorithm only" performance that would be evaluated in this context. Its performance is always in the context of its physical operation.

7. The Type of Ground Truth Used

The ground truth for the performance testing was based on objective measurements and observations from ex-vivo (e.g., porcine blood vessels, mesentery) and in-vivo (porcine animal studies) experiments. This includes quantifiable measures like burst pressure, cutting time, temperatures, grasping force, torque, and qualitative assessments of seal performance, thermal spread, and healing observed in animal models. It is not based on expert consensus, pathology, or outcomes data in the clinical sense, but rather on the direct results of the physical and biological tests.

8. The Sample Size for the Training Set

This device does not involve machine learning or AI algorithms that require a "training set" in the computational sense. Therefore, the concept of a "training set" and its sample size is not applicable. The device's design and manufacturing are based on conventional engineering principles and testing, not AI model training.

9. How the Ground Truth for the Training Set was Established

As there is no "training set" for an AI algorithm, this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 17, 2018

Olympus Medical Systems Corp. % Christina Flores Senior Specialist, Regulatory Affairs Gyrus ACMI Inc. 136 Turnpike Road Southborough, Massachusetts 01772

Re: K172610

Trade/Device Name: THUNDERBEAT 5 mm, 20 cm, Front-actuated Grip Type S, THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip Type S, THUNDERBEAT 5 mm, 45 cm, Front-actuated Grip Type S Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI, LFL Dated: February 23, 2018 Received: February 26, 2018

Dear Christina Flores:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Jennifer R. Stevenson -S For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172610

Device Name

THUNDERBEAT 5 mm, 20 cm, Front-actuated Grip Type S THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip Type S THUNDERBEAT 5 mm, 45 cm, Front-actuated Grip Type S

Indications for Use (Describe)

The THUNDERBEAT Type S hand instruments are intended to be used with the Ultrasonic Generator (USC-400), the Electrosurgical Generator (ESG-400), and the THUNDERBEAT Transducer (TD-TB400).

Seal & Cut mode:

The THUNDERBEAT Type S hand instruments when used in combination with the Seal & Cut mode are indicated for open, laparoscopic (including single-site surgery and gynecological surgery (including urologic, thoracic, plastic and reconstructive, bowel resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectomy, hysterectomies (both vaginal assisted and abdominal) etc.), and endoscopic surgery or in any procedure in which cutting, vessel ligation (sealing and cutting), coagulation, grasping, and dissection is performed. These devices have been designed to seal and cut vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics.

Seal mode:

The THUNDERBEAT Type S hand instruments when used in combination with the Seal mode are indicated for open, laparoscopic (including single-site surgery) general surgery (including urologic, thoracic, plastic and reconstructive, bowel resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectomy, hysterectomies (both vaginal assisted and abdominal) etc.), and endoscopic surgery or in any procedure in which vessel sealing, coagulation, grasping is performed. These devices have been designed to seal vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics.

The THUNDERBEAT Type S hand instruments have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.#### This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary Gyrus ACMI, Inc. THUNDERBEAT Type S Hand Instruments

General Information

Regulation Number:

Regulatory Class:

Product Codes:

Review Panel:

Applicant:OLYMPUS MEDICAL SYSTEMSCORP.2951 Ishikawa-cho, Hachioji-shi, Tokyo,Japan 192-8507Phone: (+81) 42-642-2694Fax: (+81) 42-642-2307
Establishment Registration Number:8010047
Manufacturer:Aomori Olympus2-248-1 okkonokikuroishi-shi aomori, Japan 036-0357Phone: (+81) 172-52-8543Fax: (+81) 172-52-8515
Establishment Registration Number:9614641
510(k) Submitter:Gyrus ACMI, Inc.136 Turnpike Rd.Southborough, MA 01772-2104
Establishment Registration Number:3003790304
Contact Person:Christina Flores, RACSenior Specialist, Regulatory Affairs
Date Prepared:April 16, 2018
Device Description
Classification Name:Electrosurgical cutting and coagulationdevice and accessories

21 CFR 878.4400

General & Plastic Surgery

Class II,

GEI, LFL

{4}------------------------------------------------

Trade Name:

THUNDERBEAT Type S Hand Instruments

Model NameDevice Name
TB-0520FCSTHUNDERBEAT 5 mm, 20 cm, Front-actuated Grip Type S
TB-0535FCSTHUNDERBEAT 5 mm, 35 cm, Front-actuated Grip Type S
TB-0545FCSTHUNDERBEAT 5 mm, 45 cm, Front-actuated Grip Type S

Generic/Common Name:

Ultrasonic and electrosurgical devices

Predicate Devices

Olympus Medical Systems Corp. THUNDERBEAT TB-0520IC/0535PC/0545PC/0535IC/0545IC Hand Instruments K132703

Product Description

The THUNDERBEAT Type S Hand Instruments is intended to be used with the Ultrasonic Generator (USG-400), the Electrosurgical Generator (ESG-400), and the THUNDERBEAT Transducer, (TD-TB400).

The THUNDERBEAT Type S Hand Instruments is provided as a sterile, single use device. It is a functional device capable of vessel sealing & cutting, tissue coagulating & cutting, grasping, and dissecting. This device has been designed to seal and cut vessels up to and including 7 mm in diameter, tissue bundles, and lymphatics.

Comparison of Technological Characteristics

The THUNDERBEAT Type S Hand Instruments have the same intended use and fundamental scientific technological characteristics as the predicate THUNDERBEAT Hand Instruments cleared under K132703. The subject and predicate devices activate combined HF Bipolar (FineCoag) output and Ultrasonic output [Seal & Cut mode] simultaneously while grasping vessels, tissue bundles and lymphatics between the Probe and the Grasping section.

The THUNDERBEAT Type S Hand Instruments, like the predicate THUNDERBEAT, also activates the HF Bipolar (HardCoag) output [Seal mode]. Those outputs can be used for sealing and cutting/sealing of vessels, tissue bundles and lymphatics and or coagulating and cutting/coagulating tissues. The hand switches provided on the grip handle enable those output operations.

{5}------------------------------------------------

The differences between the THUNDERBEAT Type S and the predicate THUNDERBEAT include the following:

  • Grasper shape difference in upper jaw .
  • Coating on underside of probe tip
  • . Front actuation grip on handle

Material

Full biocompatibility testing on all patient contacting surfaces has been performed in compliance to the relevant requirements of ISO-10993 series.

Indications for use

The THUNDERBEAT Type S hand instruments are intended to be used with the Ultrasonic Generator (USG-400), the Electrosurgical Generator (ESG-400), and the THUNDERBEAT Transducer (TD-TB400).

Seal & Cut mode:

The THUNDERBEAT Type S hand instruments when used in combination with the Seal & Cut mode are indicated for open, laparoscopic (including single-site surgery) general surgery and gynecological surgery (including urologic, thoracic, plastic and reconstructive, bowel resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectomy, hysterectomies (both vaginal assisted and abdominal) etc.), and endoscopic surgery or in any procedure in which cutting, vessel ligation (sealing and cutting), coagulation, grasping, and dissection is performed. These devices have been designed to seal and cut vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics.

Seal mode:

The THUNDERBEAT Type S hand instruments when used in combination with the Seal mode are indicated for open, laparoscopic (including single-site surgery) general surgery and gynecological surgery (including urologic, thoracic, plastic and reconstructive, bowel resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectomy, hysterectomies (both vaginal assisted and abdominal) etc.), and endoscopic surgery or in any procedure in which vessel sealing, coagulation, grasping is performed. These devices have been designed to seal vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics.

The THUNDERBEAT Type S hand instruments have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.

{6}------------------------------------------------

Compliance to Voluntary Standards

The design of the THUNDERBEAT Type S Hand Instruments comply with the following standards:

IEC60601-1: 2005+A1:2012 IEC60601-1-2: 2001+A1, 2007 IEC60601-2-2:2009 IEC60601-2-18:2009 ISO10993-1:2009 ISO10993-5: 2009 ISO10993-7: 2008 ISO10993-10: 2010 ISO10993-11: 2006 ISO11135:2014 ISO14971:2007 ASTM F1980-16

Device-specific guidance

  • Premarket Notification (510(k)) Submissions for Bipolar Electrosurgical Vessel Sealers for General Surgery - Guidance for Industry and Food and Drug Administration Staff, 08/15/2016
  • Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery - Guidance for Industry and Food and Drug Administration Staff, 08/15/2016

Summary of Sterilization and Shelf Life Discussion

The sterilization has been tested to comply with ISO11135:2014. Accelerated Aging testing confirmed that the subject device has a three-year shelf life.

Summary of Performance Testing

The following performance testing was conducted in support of the substantial equivalence determination.

ItemContents
Ex-vivo Vessel BurstPressureEx-vivo burst pressure testing of porcine blood vessels (arteries, veins, and tissue bundles) was conducted on both the subject and predicate devices to demonstrate vessel sealing performance.
Ex-vivo CuttingPerformanceEx-vivo Cutting performance (Cutting time) testing of porcine mesentery was conducted on both the subject and predicate devices to demonstrate cutting performance.

1. Bench Testing

{7}------------------------------------------------

ItemContents
Max TemperatureMax Temperature testing was conducted on both the subject and predicate devices to demonstrate for the inside/outside temperature at the grasping section.
Grasping forceGrasping force Testing was conducted on both the subject and predicate devices to demonstrate homogeneousness of grasping force across grasping section.
Durability of the coating on underside of probe tipDurability testing was conducted to confirm durability of coating on underside of probe tip after device output.
Durability against twistingDurability testing against twisting was conducted on both the subject and predicate devices to demonstrate torque when spark occurs between probe and jaw during activation.

2. Animal Testing

ItemContents
Chronic Animal StudyChronic animal study of porcine was conducted on both the subject and predicate devices to confirm seal performance, thermal spread and healing degree including vessels up to 7.0mm in diameter, lymphatics and tissue bundles.
Acute Animal StudyAcute animal study of porcine was conducted on both the subject and predicate devices to confirm seal performance, thermal spread and degeneration degree including vessels (artery and artery with vein) up to 7.0mm in diameter, lymphatics and tissue bundles.
  1. Clinical Testing

Clinical testing was not performed.

{8}------------------------------------------------

Substantial Equivalence

The subject and predicate devices have the same fundamental technology and indications for use. The performance, dimensions and materials of the subject device are similar to those of the predicate. Compared to the predicate device (K132703), the subject device has a different shape in the Grasping section, a coating of insulating material has been added to the underside of the probe, and the handle is now a frontactuated grip. To support the proposed design modifications, the performance tests summarized above were conducted.

Conclusion:

In summary, the THUNDERBEAT Type S Hand Instruments are substantially equivalent to the predicate devices and present no new questions of safety or efficacy.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.