K Number
K172600

Validate with FDA (Live)

Date Cleared
2017-10-25

(56 days)

Product Code
Regulation Number
886.5925
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AIR OPTIX® COLORS (lotrafilcon B) spherical soft contact lenses with refractive power are indicated for daily wear for the optical correction of refractive ametropia (myopia) in phakic or aphakic persons with non-diseased eyes and up to approximately 1.50 diopters of astigmatism that does not interfere with visual acuity.

AIR OPTIX® COLORS (lotrafilcon B) toric soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes with 6.00 diopters (D) or less of astigmatism.

AIR OPTIX® COLORS (lotrafilcon B) multifocal soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic persons with nondiseased eyes who may require a reading addition of +3.00 diopters (D) or less and who may have up to approximately 1.50 diopters of astigmatism.

AIR OPTIX® COLORS (lotrafilcon B) lenses with or without refractive power act to enhance or alter the apparent color of the eye.

The lenses may be prescribed for frequent/planned replacement wear with daily removal for cleaning and disinfection (chemical, not heat) prior to reinsertion, as recommended by the eye care professional.

Device Description

The lens material is 33% water and 67% lotrafilcon B. a fluoro-silicone containing hydrogel which is surface treated. Lotrafilcon B is classified as a Group V (silicone hydrogel) hydrogel contact lens material according to ISO 18369-1:2006/Amd.1:2009. A cosmetic pattern is embedded into the back surface of the lens, containing a combination of the following color additives: carbazole violet, iron oxides, [phthalocyaninato (2-)] copper, phthalocyanine green and titanium dioxide.

Lens designs for AIR OPTIX® COLORS (lotrafilcon B) lenses include spherical, toric, and multifocal lenses in the following parameter range:

Diameter Range: 13.0 to 15.0 mm
Base Curve Range: 8.0 to 9.2 mm
Power Range: -20.00D to +20.00D
Center Thickness: varies with design and power (0.08 mm for -3.00D spherical)

Lenses have the following properties:

• Refractive index: 1.42 (hydrated)
• Luminous transmittance: 95 ± 5%
• Water content: 33% by weight in normal saline
• Oxygen permeability 110 x 10 -11 (cm²/sec)(ml O2 /ml x mm Hg), measured at 35 °C (intrinsic Dk - Coulometric method)

Lenses are supplied sterile in sealed blister packs containing isotonic phosphate buffered saline solution (PBS), or PBS with 1% Copolymer 845 (labeled as buffered saline containing 0.2 % VP/DMAEMA Copolymer). The compatibility and package integrity of the blister pack packaging system has been demonstrated and successfully used for other marketed lens products, and packaged lenses are effectively steam sterilized in a validated autoclave. Blister pack containers are labeled with the lens parameters, lens color, lot number and product expiration date. The expiration date has been established through stability studies that have assessed the chemical stability of the lens and package integrity (ability to maintain sterility). Shelf-life studies are ongoing to further extend the labeled expiration date.

AI/ML Overview

I'm sorry, but this document does not contain the detailed information necessary to complete your request. This document is an FDA 510(k) clearance letter and a 510(k) summary for AIR OPTIX® COLORS contact lenses.

Here's why I cannot fulfill your request:

  • No Acceptance Criteria or Device Performance Table: The document states that "new color options are being added" and that "Key product characteristics remain unchanged," indicating that the performance would be substantially equivalent to the predicate device. However, it does not provide a table of specific acceptance criteria (e.g., minimum visual acuity improvement, maximum complication rate) or reported device performance against those criteria.
  • No Detailed Study Description: While it mentions "Performance testing was conducted in consideration of the May 1994 FDA guideline titled Premarket Notification 510(k) Guidance Document for Class II Contact Lenses" and "Non-clinical biocompatibility testing was conducted in accordance with the GLP regulation (21 CFR Part 58)," it does not describe the studies in detail. It explicitly states: "The scope of the device modification did not require clinical testing to establish safety and effectiveness of the modified device." This means there aren't the types of clinical studies (with test sets, ground truth, experts, MRMC studies, standalone performance) you're asking about.
  • Focus on Substantial Equivalence: The document is primarily focused on demonstrating substantial equivalence to a predicate device, meaning the new device is as safe and effective as a previously cleared device. This often involves showing that fundamental characteristics and performance remain the same, rather than conducting new, large-scale clinical trials with new acceptance criteria.

Therefore, I cannot extract information regarding:

  1. A table of acceptance criteria and reported device performance.
  2. Sample sizes for test sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, type of ground truth for test sets, sample size for training sets, or how ground truth for training sets was established.

The document indicates that the modification (adding new color options) did not necessitate new clinical studies to establish safety and effectiveness, as the key product characteristics and material remained the same as the predicate device. The performance data mentioned are non-clinical (physicochemical characteristics and biocompatibility) to show equivalence.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

October 25, 2017

Alcon Laboratories, Inc. Martina Heim, Ph.D., RAC Director, Global Regulatory Affairs 6201 South Freeway Fort Worth, TX 76134-2099

Re: K172600

Trade/Device Name: AIR OPTIX COLORS® Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lens Regulatory Class: Class II Product Code: LPL Dated: August 25, 2017 Received: August 30, 2017

Dear Martina Heim, Ph.D., RAC:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling

(21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events)

(21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions

(Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Denise L. Hampton -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K172600

Device Name AIR OPTIX® COLORS soft contact lenses

Indications for Use (Describe)

AIR OPTIX® COLORS (lotrafilcon B) spherical soft contact lenses with refractive power are indicated for daily wear for the optical correction of refractive ametropia (myopia) in phakic or aphakic persons with non-diseased eyes and up to approximately 1.50 diopters of astigmatism that does not interfere with visual acuity.

AIR OPTIX® COLORS (lotrafilcon B) toric soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes with 6.00 diopters (D) or less of astigmatism.

AIR OPTIX® COLORS (lotrafilcon B) multifocal soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic persons with nondiseased eyes who may require a reading addition of +3.00 diopters (D) or less and who may have up to approximately 1.50 diopters of astigmatism.

AIR OPTIX® COLORS (lotrafilcon B) lenses with or without refractive power act to enhance or alter the apparent color of the eye.

The lenses may be prescribed for frequent/planned replacement wear with daily removal for cleaning and disinfection (chemical, not heat) prior to reinsertion, as recommended by the eye care professional.

Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D)[] Over-The-Counter Use (21 CFR 801 Subpart C)[X] Prescription Use (Part 21 CFR 801 Subpart D)[] Over-The-Counter Use (21 CFR 801 Subpart C)
[X] Prescription Use (Part 21 CFR 801 Subpart D)
[] Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information, Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This 510(k) summary document has been prepared in accordance with section 21 CFR 807.92.

I. Submitter of the 510(k)

Company:Alcon Laboratories, Inc.
6201 South FreewayFort Worth, TX 76134-2099, USA
Primary Contact Person:Martina Heim, Ph.D., RACDirector, Global Regulatory Affairs
Phone:678-415-3565
Fax:678-415-3454
Email:martina.heim@alcon.com
Date Prepared:21 August 2017

II. Devices Subject to this 510(k)

Trade Names:AIR OPTIX® COLORS
Common Name:Soft Contact Lens
Classification Name:Soft (Hydrophilic) Contact Lenses
Device Classification:Class II [21 CFR 886.5925 (b) (1)]
Product Code:LPL

III. Predicate Device

The subject device, AIR OPTIX® COLORS (lotrafilcon B) soft contact lenses, is a modification of the currently commercialized AIR OPTIX® COLORS contact lens. The predicate AIR OPTIX® COLORS soft contact lenses have received FDA clearance per Premarket Notification 510(k) K133176.

{4}------------------------------------------------

IV. Device Description

The lens material is 33% water and 67% lotrafilcon B. a fluoro-silicone containing hydrogel which is surface treated. Lotrafilcon B is classified as a Group V (silicone hydrogel) hydrogel contact lens material according to ISO 18369-1:2006/Amd.1:2009. A cosmetic pattern is embedded into the back surface of the lens, containing a combination of the following color additives: carbazole violet, iron oxides, [phthalocyaninato (2-)] copper, phthalocyanine green and titanium dioxide.

Lens designs for AIR OPTIX® COLORS (lotrafilcon B) lenses include spherical, toric, and multifocal lenses in the following parameter range:

Diameter Range:13.0 to 15.0 mm
Base Curve Range:8.0 to 9.2 mm
Power Range:-20.00D to +20.00D
Center Thickness:varies with design and power (0.08 mm for -3.00D spherical)

Lenses have the following properties:

• Refractive index:1.42 (hydrated)
• Luminous transmittance:95 ± 5%
• Water content:33% by weight in normal saline
• Oxygen permeability110 x 10 -11 (cm²/sec)(ml O2 /ml x mm Hg),measured at 35 °C (intrinsic Dk - Coulometricmethod)

Lenses are supplied sterile in sealed blister packs containing isotonic phosphate buffered saline solution (PBS), or PBS with 1% Copolymer 845 (labeled as buffered saline containing 0.2 % VP/DMAEMA Copolymer). The compatibility and package integrity of the blister pack packaging system has been demonstrated and successfully used for other marketed lens products, and packaged lenses are effectively steam sterilized in a validated autoclave. Blister pack containers are labeled with the lens parameters, lens color, lot number and product expiration date. The expiration date has been established through stability studies that have assessed the chemical stability of the lens and package integrity (ability to maintain sterility). Shelf-life studies are ongoing to further extend the labeled expiration date.

{5}------------------------------------------------

Indications for Use V.

AIR OPTIX® COLORS (lotrafilcon B) spherical soft contact lenses with refractive power are indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes and up to approximately 1.50 diopters of astigmatism that does not interfere with visual acuity.

AIR OPTIX® COLORS (lotrafilcon B) toric soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes with 6.00 diopters (D) or less of astigmatism.

AIR OPTIX® COLORS (lotrafilcon B) multifocal soft contact lenses are indicated for daily wear for the optical correction of presbyopia with or without refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who may require a reading addition of +3.00 diopters (D) or less and who may have up to approximately 1.50 diopters of astigmatism.

AIR OPTIX® COLORS (lotrafilcon B) lenses with or without refractive power act to enhance or alter the apparent color of the eye.

The lenses may be prescribed for frequent/planned replacement wear with daily removal for cleaning and disinfection (chemical, not heat) prior to reinsertion, as recommended by the eye care professional.

VI. Comparison to Technological Characteristics with the Predicate Device

New color options are being added to the line of AIR OPTIX® COLORS (lotrafilcon B) soft contact lenses. Key product characteristics remain unchanged, as listed in Table 1. Carbazole violet is added to the list of color additives that are used to achieve the desired shade variations in the cosmetic print pattern.

{6}------------------------------------------------

Predicate DeviceModified Device
Administrative / Regulatory Information
Trade NameAIR OPTIX® COLORSSame
Submission numberK133176510(k) # to be assigned
Device ClassificationNameDaily Wear Soft Contact Lens21 CFR 886.5925 (b) (1)Same
Indications For Use Information
With refractive power:Vision CorrectionSame
Intended UseWith or without refractive power:Enhance or alter the apparentcolor of the eyeSame
Wearing ScheduleDaily WearSame
Replacement ScheduleUp to Monthly ReplacementSame
Material and Technology Information
Lens Materiallotrafilcon BSame
Surface TreatmentPlasma treatedSame
Manufacturing MethodDouble-side molding;integrated print stepSame
Print TechnologyIn-mold pad print technologySame
Color Additives for Print:PCN greenPCN blueTitanium dioxideYellow iron oxideRed iron oxideBlack iron oxidePCN greenPCN blueTitanium dioxideYellow iron oxideRed iron oxideBlack iron oxideCarbazole violet
Water Content33%Same
Refractive Index1.42Same
Oxygen Permeability~110*Same
SterilizationSteam sterilization,validated autoclaveSame
Predicate DeviceModified Device
PackagingBlister packSame
Package Storage SalineSolutionPhosphate buffered saline with(or without) 1% Copolymer 845Same
Lens Design and Parameters
Lens DesignsSpherical, toric, multifocalSame
Power Range+20.00 to -20.00DSame
Base Curve Range8.0 to 9.2 mmSame
Diameter Range13.0 to 15.0 mmSame

Table 1. Substantial Equivalence Comparison

{7}------------------------------------------------

Table 1. Substantial Equivalence Comparison

  • intrinsic Dk - Coulometric method; barrer units

VII. Performance Data

Performance testing was conducted in consideration of the May 1994 FDA guideline titled Premarket Notification 510(k) Guidance Document for Class II Contact Lenses. The following performance data are provided in support of the substantial equivalence determination:

Non-clinical Testing

A series of non-clinical testing was performed to characterize the lens material properties of AIR OPTIX® COLORS lenses and demonstrate the substantial equivalence of the modified device to the predicate device. Non-clinical biocompatibility testing was conducted in accordance with the GLP regulation (21 CFR Part 58).

The results of all non-clinical testing demonstrate:

  • Physicochemical characteristics of the modified device are substantially equivalent to the predicate lens, AIR OPTIX® COLORS.
  • The modification does not impact the established safety profile for AIR OPTIX® ● COLORS soft contact lenses.

Successful stability testing supports the labeled expiration date.

Clinical Testing

The scope of the device modification did not require clinical testing to establish safety and effectiveness of the modified device.

{8}------------------------------------------------

VIII. Conclusions

AIR OPTIX® COLORS (lotrafilcon B) soft contact lenses as modified are equivalent to the predicate device lenses and similar to other daily wear soft contact lenses in terms of technological characteristics and intended use.

Non-clinical data demonstrates that the modified device is as safe, as effective, and performs as well as or better than the legally marketed, predicate device, AIR OPTIX® COLORS soft contact lenses.

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.