K Number
K172554

Validate with FDA (Live)

Date Cleared
2018-05-01

(250 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Prescription: LUOFUCON® Extra Silver Alginate Dressing (Rx only) is indicated for the management of moderate to heavily exuding partial to full thickness wounds, including, postoperative wounds, trauma wounds, leg ulcers, pressure ulcers, diabetic ulcers, graft and donor sites.
OTC: LUOFUCON® Antibacterial Alginate Wound Dressing (OTC) is first aid to help in minor abrasions, minor cuts, lacerations, scrapes, minor scalds and burns.

Device Description

LUOFUCON® Extra Silver Alginate Dressing (Rx only)/LUOFUCON® Antibacterial Alginate Wound Dressing (OTC) is a sterile, single-use dressing composed of calcium alginate and silver antibacterial agent, which absorbs wound exudate and release silver ions in the presence of wound fluid in the dressing. As wound exudate is absorbed, the alginate forms a gel, which assists in maintaining a moist environment for optimal wound healing, and allows intact removal. The alginate material contains silver which inhibits bacterial growth within the dressing. Based on in vitro laboratory testing, the silver has been shown to protect the dressing against both Gram positive and Gram negative bacteria, such as Staphylococcus aureus, Enterococcus faecalis, Streptococcus pyogenes, Pseudomonas aeruginosa, Escherichia coli, Klebsiella pneumoniae, Methicillin-resistant Staphylococcus aureus (MRSA), and Vancomycin-resistant Enterococcus (VRE). And the silver protects the dressing from a broad spectrum of bacteria up to seven days. All dressings are sterilized and sold after sterilization by radiation using conditions validated following ISO 11137-2:2013.

AI/ML Overview

This document describes the premarket notification for the LUOFUCON® Extra Silver Alginate Dressing (Rx Only) and LUOFUCON® Antibacterial Alginate Wound Dressing (OTC). The acceptance criteria and supporting studies are presented below.

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance
Free Swell Absorption CapacityNot explicitly stated; implied to be comparable to predicate device.Performance comparable to the predicate device.
Loss on DryingNot explicitly stated; implied to be comparable to predicate device.Performance comparable to the predicate device.
pH ValueNot explicitly stated; implied to be comparable to predicate device.Performance comparable to the predicate device.
Silver contentNot explicitly stated; implied to be controlled and effective for indicated antibacterial properties.Contained silver which inhibits bacterial growth within the dressing. Shown to protect against Staphylococcus aureus, Enterococcus faecalis, Streptococcus pyogenes, Pseudomonas aeruginosa, Escherichia coli, Klebsiella pneumoniae, Methicillin-resistant Staphylococcus aureus (MRSA), and Vancomycin-resistant Enterococcus (VRE).
Antibacterial ActivityThe subject device achieves antibacterial duration for seven days.Passed (Achieved antibacterial duration for seven days). In vitro laboratory testing demonstrated protection against both Gram positive and Gram negative bacteria, including MRSA and VRE, for up to seven days.
Biocompatibility (Surface device; contacting breached or compromised skin for prolonged duration)No new safety concerns raised as per ISO 10993-1:2009. Must be safe for intended use. Must pass tests for Cytotoxicity, Sensitization, Irritation, Systemic toxicity, Implantation, and Material-mediated pyrogenicity.No new safety concerns were raised. Device demonstrated to be safe for its intended use, having been evaluated for: Cytotoxicity, Sensitization, Irritation, Systemic toxicity, Implantation, Material-mediated pyrogenicity. The porcine wound healing study showed silver did not cause cytotoxicity or delay normal wound healing.
SterilizationValidated following ISO 11137-2:2013 for radiation sterilization.All dressings were sterilized by radiation using conditions validated following ISO 11137-2:2013.

2. Sample Size and Data Provenance for Test Set (Performance Tests)

The document primarily describes in vitro and animal studies, not human clinical trials with a "test set" in the typical sense for AI/software devices.

  • Antibacterial Activity Test: The sample size for the AATCC 100-2012 testing is not specified. The data provenance is from in vitro laboratory testing performed by the manufacturer, Huizhou Foryou Medical Devices Co., Ltd., likely in China (country of origin for the submitter). This would be considered retrospective data in the context of the submission.
  • Biocompatibility Tests: The sample sizes for the various ISO 10993-1:2009 biocompatibility tests (Cytotoxicity, Sensitization, Irritation, Systemic toxicity, Implantation, Material-mediated pyrogenicity) are not specified in this summary. These are laboratory tests.
  • Porcine Wound Healing Study: The sample size for the porcine wound healing study is not specified. This is an animal model study.

3. Number of Experts and Qualifications for Ground Truth (Test Set)

Not applicable. The reported studies are laboratory performance tests and an animal study, not studies involving human experts establishing ground truth for diagnostic or AI performance.

4. Adjudication Method (Test Set)

Not applicable. This is not a study requiring adjudication of human interpretations.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. This document describes the performance of a wound dressing, not an AI or diagnostic device that would involve human readers or MRMC studies.

6. Standalone Performance

Yes, the performance tests (Free Swell Absorption Capacity, Loss on Drying, pH Value, Silver content, Antibacterial Activity, and Biocompatibility tests, including the porcine wound healing study) represent the standalone performance of the device itself. These tests assess the inherent physical, chemical, and biological properties of the wound dressing without human-in-the-loop intervention for diagnostic or interpretative purposes.

7. Type of Ground Truth Used

The ground truth for the performance tests is based on:

  • Quantitative measurements against established standards and methodologies (e.g., AATCC 100-2012 for antibacterial activity, ISO 10993-1:2009 for biocompatibility).
  • Direct observation and assessment in the porcine wound healing study (e.g., absence of cytotoxicity, normal wound healing).
  • Laboratory analysis for material properties (e.g., silver content, pH).

8. Sample Size for Training Set

Not applicable. This is a medical device (wound dressing), not an AI or machine learning algorithm that requires a "training set."

9. How Ground Truth for Training Set was Established

Not applicable, as there is no training set for this device type.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 1, 2018

Huizhou Foryou Medical Devices Co., Ltd. Huida Huang Development Engineer North Shangxia Rd., Dongjiang Hi-tech Industry Park Huizhou, 516005 Cn

Re: K172554

Trade/Device Name: Luofucon Extra Silver Alginate Dressing (Rx Only), Luofucon Antibacterial Alginate Wound Dressing (OTC) Regulatory Class: Unclassified

Product Code: FRO Dated: March 29, 2018 Received: March 29, 2018

Dear Huida Huang:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may; therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

David Krause -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172554

Device Name

Luofucon Extra Silver Alginate Dressing (Rx Only) Luofucon Antibacterial Alginate Wound Dressing (OTC)

Indications for Use (Describe)

Prescription:

Luofucon Extra Silver Alginate Dressing is indicated for the management of moderate to heavily exuding partial to full thickness wounds, including: postoperative wounds, leg ulcers, pressure ulcers, diabetic ulcers, graft and donor sites.

OTC:

Luofucon Antibacterial Alginate Wound Dressing is first aid to help in minor cuts, lacerations, scrapes, minor scalds and burns.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary or 510(k) Statement Section 5

This 510(k) Summary information is being submitted in accordance with the requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number: K172554

1. Date of Submission:

April 25, 2018

2. Submitter Identification:

Huizhou Foryou Medical Devices Co., Ltd. North Shangxia Rd., Dongjiang Hi-tech Industry Park, 516005, Huizhou, P. R. China. Establishment Registration Number: 3007735241 Contact Person: Huida Huang Position: Development Engineer Tel: +86-0752-5302185 Fax: +86-0752-5302020 Email: hdhuang@foryougroup.com

3. Subject Device Identification:

Device Name: LUOFUCON® Extra Silver Alginate Dressing (Prescription use) LUOFUCON® Antibacterial Alginate Wound Dressing (OTC use) Common Name: Silver Alginate Dressing Classification Name: Dressing, Wound, Drug; Product Code: FRO; Regulation Number: Unclassified; Review Panel: General & Plastic Surgery;

4. Predicate Device Identification:

510(k) Number: K120181 Product Name: LUOFUCON® Silver Alginate Dressing (Prescription Use)/ LUOFUCON® Antibacterial Alginate Dressing (OTC Use) Manufacturer: Huizhou Foryou Medical Devices Co., Ltd.

5. Device Description:

LUOFUCON® Extra Silver Alginate Dressing (Rx only)/LUOFUCON® Antibacterial Alginate Wound Dressing (OTC) is a sterile, single-use dressing composed of calcium alginate and silver antibacterial agent, which absorbs wound exudate and release silver

{4}------------------------------------------------

ions in the presence of wound fluid in the dressing. As wound exudate is absorbed, the alginate forms a gel, which assists in maintaining a moist environment for optimal wound healing, and allows intact removal.

The alginate material contains silver which inhibits bacterial growth within the dressing. Based on in vitro laboratory testing, the silver has been shown to protect the dressing against both Gram positive and Gram negative bacteria, such as Staphylococcus aureus, Enterococcus faecalis, Streptococcus pyogenes, Pseudomonas aeruginosa, Escherichia coli, Klebsiella pneumoniae, Methicillin-resistant Staphylococcus aureus (MRSA), and Vancomycin-resistant Enterococcus (VRE). And the silver protects the dressing from a broad spectrum of bacteria up to seven days.

All dressings are sterilized and sold after sterilization by radiation using conditions validated following ISO 11137-2:2013.

6. Intended Use Statement:

Prescription:

LUOFUCON® Extra Silver Alginate Dressing (Rx only) is indicated for the management of moderate to heavily exuding partial to full thickness wounds, including, postoperative wounds, trauma wounds, leg ulcers, pressure ulcers, diabetic ulcers, graft and donor sites.

OTC:

LUOFUCON® Antibacterial Alginate Wound Dressing (OTC) is first aid to help in minor abrasions, minor cuts, lacerations, scrapes, minor scalds and burns.

7. Comparison to the Predicate Device

LUOFUCON® Extra Silver Alginate Dressing (Rx only)/LUOFUCON® Antibacterial Alginate Wound Dressing (OTC) is compared with the following Predicate Device in terms of intended use, design, material, specifications, and performance.

K120181, LUOFUCON® Silver Alginate Dressing (Prescription Use)/ ● LUOFUCON® Antibacterial Alginate Dressing (OTC Use), Manufactured by Huizhou Foryou Medical Devices Co., Ltd.

The following table shows similarities and differences of use, design, material, and processing methods between subject device and its predicate device.

{5}------------------------------------------------

ItemSubject Device (K172554)Predicate Device (K120181)
Intended UsePrescription:LUOFUCON® Extra Silver AlginateDressing (Rx only) is indicated for themanagement of moderate to heavilyexuding partial to full thicknesswounds, including, postoperativewounds, trauma wounds, leg ulcers,pressure ulcers, diabetic ulcers, graftand donor sites.OTC:LUOFUCON® Antibacterial AlginateWound Dressing (OTC) is first aid tohelp in minor abrasions, minor cuts,lacerations, scrapes, minor scalds andburns.Prescription:LUOFUCON® Silver AlginateDressing is indicated for themanagement of moderate to heavilyexuding partial to full thicknesswounds, including, postoperativewounds, trauma wounds, leg ulcers,pressure ulcers, diabetic ulcers, graftand donor sites.OTC:LUOFUCON® Antibacterial AlginateDressing is first aid to help in minorabrasions, minor cuts, lacerations,scrapes, minor scalds and burns.
MechanismAlginate for absorbing liquid, silverpresent in the alginate for reducingbacteria colonization in the dressing.Alginate for absorbing liquid, silverpresent in the alginate for reducingbacteria colonization in the dressing.
MaterialAlginate and silverAlginate and silver
AntibacterialDurationSeven daysOne day
Single UseYesYes
SterilizationRadiationRadiation

Table 5-1 Comparison of Intended Use, Design and Material

LUOFUCON® Extra Silver Alginate Dressing (Rx only)/LUOFUCON® Antibacterial Alginate Wound Dressing (OTC) and its predicate device (K120181) utilize silver as the antibacterial agent and utilize alginate pad for the exudates absorption and wound So, LUOFUCON® Extra management. Silver Dressing (Rx only)/LUOFUCON® Antibacterial Alginate Wound Dressing (OTC) and its predicate device are made from similar materials, utilize the same antibacterial mechanism, and have the same intended use.

All dressings are sterilized and sold after sterilization by radiation using conditions validated following ISO 11137-2:2013.

The following performance tests were conducted on LUOFUCON® Extra Silver Alginate Dressing (Rx only)/LUOFUCON® Antibacterial Alginate Wound Dressing (OTC) in comparison to the predicate:

  • Free Swell Absorption Capacity ●

{6}------------------------------------------------

  • Loss on Drying ●
  • pH Value
  • Silver content
  • Antibacterial Activity

LUOFUCON® Extra Silver Alginate Dressing (Rx only)/LUOFUCON® Antibacterial Alginate Wound Dressing (OTC) raised no new safety concerns for biocompatibility. Assessment was carried out on the final device as per ISO 10993-1:2009 for a surface device; contacting breached or compromised skin for prolonged duration. The device has been demonstrated to safe for its intended use.

LUOFUCON® Extra Silver Alginate Dressing (Rx only)/LUOFUCON® Antibacterial Alginate Wound Dressing (OTC) was evaluated for:

  • Cytotoxicity
  • Sensitization
  • Irritation
  • Systemic toxicity
  • Implantation
  • Material-mediated pyrogenicity

Modified AATCC 100-2012 was used to evaluate the antibacterial activity of the subject device. The predicate device also chooses the AATCC 100 as the test method. The subject device achieves antibacterial duration for seven days while the predicate device achieves one day.

A porcine wound healing study was carried out to evaluate the silver cytotoxicity of subject device. The test results had shown that the silver does not cause cytotoxicity in the healing period or delay normal wound healing.

Therefore, the performance results are comparable to the predicate device when the dressings are used for antibacterial purpose for seven days. The product is safe and effective for its intended use.

Substantial Equivalent Statement

Based on the comparison of intended use, design, materials, performance, biocompatibility, and animal study, the subject device, LUOFUCON® Extra Silver Alginate Dressing (Rx only)/LUOFUCON® Antibacterial Alginate Wound Dressing (OTC), is determined to be Substantially Equivalent (SE) to the predicate device, LUOFUCON® Silver Alginate Dressing (Prescription Use)/ LUOFUCON® Antibacterial Alginate Dressing (OTC Use) (K120181), in respect of safety and effectiveness.

N/A