K Number
K172490

Validate with FDA (Live)

Device Name
eUnity
Date Cleared
2018-02-06

(173 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

eUnity is a software application that displays medical image data and associated clinical reports to aid in diagnosis for healthcare professionals. It performs operations relating to the transfer, storage, display, and measurement of image data.

eUnity allows users to perform image manipulations, including window/level, rotation, measurement and markup. eUnity provides 2D display, Multi-Planar Reformating and 3D visualization of medical image data, and mobile access to images.

eUnity displays both lossless and lossy compressed images. For lossy images, the medical professional user must determine if the level of loss is acceptable for their purposes. Display monitors or mobile devices used for reading medical images for diagnostic purposes must comply with applicable regulatory approvals and with quality control requirements for their use and maintenance. For mobile diagnostic usage when a full workstation is not available.

Mobile usage for mammography is for reference and referral only.

Device Description

Client Outlook has developed eUnity to load, display and manipulate medical (DICOM) images within a web-browser without installing client software. eUnity is a server-based software solution that extends common web-browsers on the most popular operating systems into medical review stations; removing a technical barrier that had long been a key contributor to poor medical image access.

eUnity is an enterprise medical image viewer that provides access to full quality images from anywhere using nothing more than a standard web browser. Combined with a calibrated monitor, it can be used to make diagnostic decisions. Secure, fast, immediate access to information means less time spent searching for specialized workstations and supports greater efficiency for care, greater collaboration, and faster turnaround times.

This device is the successor to eUnity predicate (K161515) and adds the following functionality: MIP/MPR/3D and Mobile Diagnostic Use.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the eUnity device based on the provided text, structured to answer your specific questions.

1. Table of Acceptance Criteria and Reported Device Performance

The provided document (K172490) is a 510(k) summary for a medical image processing software (eUnity). It's primarily a comparison document to demonstrate substantial equivalence to predicate devices rather than a detailed report of a performance study with numerical acceptance criteria. The "acceptance criteria" here are implied by the comparison to already cleared predicate devices and the statement that the device "meets its design requirements and intended use, and that it is safe and effective."

The capabilities added to the new eUnity version (K172490) are:

  • MIP/MPR/3D functionality
  • Mobile Diagnostic Use

The reported performance is qualitative and based on clinical validation:

Acceptance Criteria (Implied)Reported Device Performance
Functional Equivalence to Predicate for Existing Features: Demonstrates the same or similar essential features for visualization of radiological data.The eUnity device (K172490) offers the same functionality as the K161515 eUnity device for existing features (e.g., Window Level, Rotate/Pan/Zoom, Multi-Study viewing, Measurement tools, Metadata display, etc.) (See Comparison Chart in the document).
Successful Implementation of New Features (MIP/MPR/3D): New 3D viewing capabilities are functionally similar to those in reference predicate devices.The device adds MIP/MPR/3D functionality, which is compared to the Resolution MD (K133508) and Vue Motion (K151774) reference predicate devices, both of which have these features. The documentation does not provide a specific performance metric for these new features but implies their functionality is equivalent to the predicates.
Successful Implementation of Mobile Diagnostic Use: Enables diagnostic quality viewing on mobile devices."Additional Clinical Validation testing based on typical clinical workflows was performed by trained radiologists in comparison with an existing device and on several different hardware devices. ... There was consensus among all the Radiologists that the same diagnosis would be made on the mobile device with eUnity as on the predicate device in various lighting conditions." This suggests the mobile diagnostic capability was deemed diagnostically equivalent. The validated mobile devices include: iPhone 6 and higher, iPad Mini and higher, iPad Pro and higher, Samsung Galaxy Note 5 and higher, and Samsung Galaxy Tab E and higher. Additionally, a "Mobile Luminance Check" is implemented, requiring the user to determine lighting conditions prior to diagnosis.
Meets Design Requirements and Intended Use, Safe and Effective: Overall regulatory compliance and performance."designated individuals performed all verification and validation activities and results demonstrated that the device meets its design requirements and intended use, and that it is safe and effective." This is a general statement affirming the successful completion of internal V&V activities. The substantial equivalence argument itself serves as the primary "proof" of safety and effectiveness for a 510(k) device.

2. Sample Size Used for the Test Set and the Data Provenance

  • Test Set Sample Size: The document does not specify a numerical sample size (e.g., number of images, number of cases) for the "Additional Clinical Validation testing." It mentions "several different hardware devices" and testing "on several different hardware devices."
  • Data Provenance: Not explicitly stated whether the data used for clinical validation was retrospective or prospective, nor its country of origin.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

  • Number of Experts: The document states "multiple team members" performed verification and validation activities, and "trained radiologists" performed the clinical validation. It also notes "consensus among all the Radiologists." The exact number of radiologists involved is not specified, but it implies more than one.
  • Qualifications of Experts: The experts are identified as "trained radiologists." No further details on their years of experience or specific subspecialties are provided.

4. Adjudication Method for the Test Set

  • The document states, "There was consensus among all the Radiologists that the same diagnosis would be made on the mobile device with eUnity as on the predicate device in various lighting conditions." This indicates a consensus-based adjudication method, but the specific process (e.g., initial independent reads followed by a consensus meeting, voting, etc.) is not detailed. It implies that agreement was reached among the radiologists.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

  • No, an MRMC comparative effectiveness study, particularly one measuring the improvement of human readers with AI assistance, was not conducted or reported. This device is a medical image viewing and processing software, not an AI-powered diagnostic aid. The clinical validation was a comparison of diagnostic capability on different platforms (mobile with eUnity vs. predicate device).

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

  • No, a standalone (algorithm-only) performance study was not conducted as this device is a viewing platform and requires human interpretation for diagnosis. Its intended use states "a licensed medical practitioner reviews the output, providing ample opportunity for competent human intervention for the interpretation of images and information being displayed."

7. The Type of Ground Truth Used

  • The ground truth for the clinical validation appears to be the diagnosis made using an existing predicate device/workstation. The test aimed to determine if "the same diagnosis would be made on the mobile device with eUnity as on the predicate device." This suggests the predicate's output served as the reference.

8. The Sample Size for the Training Set

  • This document describes a 510(k) submission for a software that displays and manipulates medical images. It is not an AI/ML device in the sense of requiring a "training set" to learn diagnostic patterns. Therefore, there is no mention of a training set or its sample size. The software's functionality is based on established imaging principles and display technologies, not learned algorithms from a dataset.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as there is no "training set" for this type of medical device software.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Client Outlook Inc. Christie Eby Director of Operations & Quality 103 Bauer Place. Suite #3 Waterloo, ON N2T 2V2 CANADA

February 6, 2018

Re: K172490

Trade/Device Name: eUnity Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: LLZ Dated: January 2, 2018 Received: January 4, 2018

Dear Christie Eby:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Sincerely,

Michael D. O'Hara For

Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K172490

Device Name eUnity

Indications for Use (Describe)

eUnity is a software application that displays medical image data and associated clinical reports to aid in diagnosis for healthcare professionals. It performs operations relating to the transfer, storage, display, and measurement of image data.

eUnity allows users to perform image manipulations, including window/level, rotation, measurement and markup. eUnity provides 2D display, Multi-Planar Reformating and 3D visualization of medical image data, and mobile access to images.

eUnity displays both lossless and lossy compressed images. For lossy images, the medical professional user must determine if the level of loss is acceptable for their purposes. Display monitors or mobile devices used for reading medical images for diagnostic purposes must comply with applicable regulatory approvals and with quality control requirements for their use and maintenance. For mobile diagnostic usage when a full workstation is not available.

Mobile usage for mammography is for reference and referral only.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

This 510(k) summary is submitted in accordance with the requirements of 21 CFR Part 807.92(c)

Basis for the submission:Client Outlook hereby submits this traditional 510(k) submission for eUnity software which is substantially equivalent to the FDA cleared previous version of eUnity Software (K161515)
Submitter:Client Outlook Inc.103 Bauer Place, Suite #3Waterloo, OntarioCanada N2L 6B5
Date:August 14, 2017
Establishment Registration:3009601121
Contact:Christie Eby, Director of Operations & QualityTel: 519-342-3049 x206Fax: 519-725-2351
Trade Name:eUnity
Common Name:Medical Image Processing Software
Classification:Picture Archiving and Communications Software (PACS)
Product Code:LLZ
Device Class:Class II
Regulation #:21 CFR 892.2050
Device Panel:Radiology

Predicate Devices:

Trade Name510 (k) Submitter/ Manufacturer510 (k) number
eUnityClient Outlook Inc.K161515

Reference Predicate Devices:

Trade Name510 (k) Submitter/ Manufacturer510 (k) number
Resolution MDCalgary ScientificK161130
Vue MotionCarestreamK151774

Device Description:

Client Outlook has developed eUnity to load, display and manipulate medical (DICOM) images within a web-browser without installing client software. eUnity is a server-based software solution that extends common web-browsers on the most popular operating systems into medical review stations; removing a technical barrier that had long been a key contributor to poor medical image access.

eUnity is an enterprise medical image viewer that provides access to full quality images from anywhere using nothing more than a standard web browser. Combined with a calibrated monitor, it can be used to make diagnostic decisions. Secure, fast, immediate access to information means less time spent

{4}------------------------------------------------

searching for specialized workstations and supports greater efficiency for care, greater collaboration, and faster turnaround times.

This device is the successor to eUnity predicate (K161515) and adds the following functionality: MIP/MPR/3D and Mobile Diagnostic Use.

The following devices are validated for use with eUnity for Mobile:

iPhone version 6 and higher versions, iPad Mini and higher versions, iPad pro and higher versions, Samsung Galaxy Note 5 and higher versions, and Samsung Galaxy Tab E and higher versions.

Intended Use:

Client Outlook's eUnity enables health professionals to access, manipulate and collaborate real-time over full quality medical images using any web-browser without installing client software. eUnity is a server-based solution that connects to any PACS and displays DICOM images within the hospital, securely from remote locations, or as an integrated part of an EHR or portal. eUnity offers diagnostic quality images with the performance of installed PACS viewing software.

Indications for Use:

eUnity is a software application that displays medical image data and associated clinical reports to aid in diagnosis for healthcare professionals. It performs operations relating to the transfer, storage, display, and measurement of image data.

eUnity allows users to perform image manipulations, including window/level, rotation, measurement and markup.

eUnity provides 2D display, Multi-Planar Reformatting and 3D visualization of medical image data, and mobile access to images.

eUnity displays both lossless and lossy compressed images. For lossy images, the medical professional user must determine if the level of loss is acceptable for their purposes. Display monitors or mobile devices used for reading medical images for diagnostic purposes must comply with applicable regulatory approvals and with quality control requirements for their use and maintenance. For Mobile diagnostic usage only when a full workstation is not available.

Mobile usage for mammography is for reference and referral only.

Patient Interaction:

eUnity is a software device that handles medical images. It does not contact the patient, nor control any life sustaining devices. Prior to any medical decisions, a licensed medical practitioner reviews the output, providing ample opportunity for competent human intervention for the interpretation of images and information being displayed.

Client Outlook Inc. - eUnity 510(k) Summary - Section 5

{5}------------------------------------------------

Comparison to predicate devices CHART:

System, Image Processing, Radiological
CategoryClient Outlook'eUnity'A)PredicateClient Outlook'eUnity'B)CalgaryScientificResolutionMDMobileC)CarestreamVue Motion
510KK161515K133508K151774
ClassClass IIClass IIClass IIClass II
Intended Use - for full descriptions - see appendices
Intended UseDiagnosticQualityDiagnosticQualityDiagnosticQualityDiagnostic Quality
User InstallRequirementsThin Client - noinstall, runswithin browserThin Client - noinstall, runswithin browserThin Client - noinstall, runs withinbrowserThin Client - noinstall, runs withinbrowser
CommunicationsDICOM, IHE,Non-DICOMDICOM, IHEDICOM, Non-DICOMDICOM, Non-DICOM
ModalitiesCR, CT, DX,ECG, MR, MG,NM, OP, PR,PT, RF, SC,SR, US, XA, VLCR, CT, DX,ECG, MR, MG,NM, OP, PR,PT, RF, SC,SR, US, XA, VLCT, MR, CR, DX,ES, KO, NM, OP,OT, PT, SC, US,XA, IO, XC,RTIMAGE, OPT,SR, RFCR, DR, CT, MR,NM, ECG, US
Tools
Window Level,Rotate/Pan/Zoom,Reset, Presets,Invertyesyesyesyes
Multi-Studyviewing, Imageprinting, ReportPrinting, ImageExportyesyesyesyes
Metadatadisplay/hideyesyesyesyes
Orientationlabels,Keyboardshortcuts,yesyesyesyes
Measurementtools, Annotationtools (Line,Arrow, Polygon,Freehand, Text)yesyesyesyes
Full-screenmode,Collaboration,Multimonitor,Linking Series,Revert toOriginal, ImageSharing,Triangulation,Image scrolling,Layouts, Linkedscrolling,Reference lines,Image flip androtate, Imagemeasurements,yesyesyesyes
Grayscalesoftcopypresentationstates (GSPS)yesyesyesyes
KINyesyesyesyes
Mag lensyesyesyesyes
MIP/MPR/3D Features:
Multi-Planarreformat (MPR)yesnoyesyes
MaximumIntensityProjection (MPR)yesnoyesyes
Oblique anddouble-obliquereformat,Triangulate,Rotateyesnoyesyes
3D Volume,orientationwidget,Opacity preset,Scalpel Tool,Bone Removalyesnoyesyes
Mobile Specific Features:
MobileLuminanceCheck(test to beperformed byuser to determinelightingconditions priorto diagnosis)yesnoyesyes
Security
Data EncryptionHTTPSHTTPSHTTPSHTTPS
Data SecurityStored onServerStored onServerStored on ServerStored on Server
Access ControlCan either usebuilt in accesscontrol or whenlaunched fromparentapplication canCan either usebuilt in accesscontrol or whenlaunched fromparentapplication canCan either usebuilt in accesscontrol or whenlaunched fromparent applicationCan either usebuilt in accesscontrol or whenlaunched fromparent application
Client Outlook™Inc.510(k) Summary Document – eUnity TMApproved
utilize itsaccess controlutilize itsaccess controlcan utilize itsaccess controlcan utilize itsaccess control

{6}------------------------------------------------

Approved

{7}------------------------------------------------

{8}------------------------------------------------

Summary of Testing and conclusion

As required by the Risk Analysis, designated individuals performed all verification and validation activities and results demonstrated that the device meets its design requirements and intended use, and that it is safe and effective. Verification testing executed by multiple team members included functional, smoke and regression tests and was complimented by beta tests performed by Client Outlook partners and Validation in test systems at customer sites.

Additional Clinical Validation testing based on typical clinical workflows was performed by trained radiologists in comparison with an existing device and on several different hardware devices. Refer to the Validation Summaries in Section 12 for additional information. There was consensus among all the Radiologists that the same diagnosis would be made on the mobile device with eUnity as on the predicate device in various lighting conditions.

Conclusion:

The Intended use and technological characteristics of the Client Outlook eUnity software are substantially equivalent, in our opinion, to those of the predicate device and reference devices and do not pose any new issues of safety and effectiveness. The device and the predicates are post-processing and provide the same or similar essential features of visualization of radiological data on mobile devices.

The modified eUnity device offers the same functionality as the K161515 eUnity device with the addition of the Mobile enhanced capabilities and 3D viewing capabilities that are found in the Reference Predicate device.

N/A