K Number
K172450

Validate with FDA (Live)

Date Cleared
2018-09-13

(395 days)

Product Code
Regulation Number
882.5891
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TENS device-HeadaTerm is indicated for the prophylactic treatment of episodic migraine in patients 18 years of age or older.

Device Description

TENS device-HeadaTerm offers a drug-free scheme, which introduces electric impulses to act on the supraorbital nerve and supratrochlear nerve, the very nerves that transmit migraine pain, to treat and prevent migraine headaches. TENS device-HeadaTerm enables an electronic feedback mechanism to adjust the electrical impulses to the specific requirements of the individual user for personalized efficacy and comfort.

The device could adhere to patients forehead with conductive paste, the current passing through the electrodes and release current stimulate the nerve underneath patients' forehead skin; the stimulation would alleviate migraine headaches, and during ramping up, patients could choose comfortable intensity level by pressing the host key to lock up the current voltage, and the patient contact material is biocompatibility safe. The conductive paste could be replaced once the paste lost its adhesiveness and patients are recommended to replace the conductive paste after 7 times of treatment uses.

AI/ML Overview

The provided document is a 510(k) premarket notification for the TENS device-HeadaTerm, indicating it is substantially equivalent to a legally marketed predicate device (Cefaly). This submission focuses on demonstrating substantial equivalence through non-clinical testing and comparison of technological characteristics, rather than presenting a performance study with acceptance criteria often seen for AI/ML-driven medical devices that require a clinical effectiveness study.

Therefore, many of the requested details, such as acceptance criteria based on clinical performance metrics (e.g., sensitivity, specificity, AUC), sample sizes for test sets in clinical studies, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, and details on training sets for an AI/ML model, are not applicable or not provided in this type of 510(k) submission.

This device, a Transcutaneous Electrical Nerve Stimulator (TENS), is cleared based on its equivalence to an existing TENS device for prophylactic treatment of episodic migraine, demonstrated through non-clinical testing (electrical safety, electromagnetic safety, performance of nerve and muscle stimulators, performance for home use, biocompatibility).

Here's an attempt to address the points based on the provided document, highlighting what is available and what is not:


1. A table of acceptance criteria and the reported device performance

The document does not present "acceptance criteria" in terms of clinical performance metrics like sensitivity or specificity, nor does it report device performance against such metrics. Instead, it focuses on demonstrating equivalence to a predicate device through non-clinical testing and comparison of physical and electrical specifications.

The "acceptance criteria" can be inferred as successful completion of the listed non-clinical tests and demonstrating comparable technical specifications to the predicate device.

Acceptance Criteria (Implied)Reported Device Performance (WAT Medical - HeadaTerm)Predicate Device (Cefaly) Performance (for comparison)
Non-Clinical Testing:Results(N/A - Predicate's test results not detailed here)
Electrical Safety (IEC60601-1)Pass (per Section 1.9 & 1.10)Yes (per Section 1.10)
Electromagnetic Safety (IEC60601-1-2)Pass (per Section 1.9 & 1.10)Yes (per Section 1.10)
Performance of nerve and muscle stimulators (IEC60601-2-10)Pass (per Section 1.9 & 1.10)Yes (per Section 1.10)
Performance of Device for Home Use (IEC 60601-1-11)Pass (per Section 1.9) / Yes (per Section 1.10)No (per Section 1.10 - This is a difference noted)
Biocompatibility: CytotoxicityPass (per Section 1.9)(N/A - Predicate's test results not detailed here)
Biocompatibility: IrritationPass (per Section 1.9)(N/A - Predicate's test results not detailed here)
Biocompatibility: SensitizationPass (per Section 1.9)(N/A - Predicate's test results not detailed here)
Key Technical Specifications (Comparison for Equivalence):
Intended UseProphylactic treatment of episodic migraine in patients ≥ 18 yearsProphylactic treatment of episodic migraine in patients ≥ 18 years
Power SourceOne 3V lithium coin cellsOne lithium battery
Follow CurrentYesYes
Voltage Overload DetectionYesYes
Adjustable IntensityON/OFF button on front of deviceON/OFF button on front of device
Channel11
Software-controlledNoYes
Time SetNo - Manually AdjustNo - Manually Adjust
Constant CurrentNoNo
Automatic overload trip voltage levelYesYes
Patient override control methodOn/Off button on front of deviceOn/Off button on front of device
Indicator displaysUnit functioning, Electrical connectionUnit functioning, Low battery, Electrical connection
Timer SettingYesYes
Weight11g30g
Dimensions127x35x12mm160 x 170 x 40 mm
WaveformAC Symmetric Square WaveAC Symmetric Square Wave
Phase Duration (µsec)250us(±20)250us(±0.5%)
Phase Interval10us(±0.5%)10us(±0.5%)
Pulse Period500us(±0.5%)505us(±0.5%)
Frequency (Hz)60Hz(±1)60Hz(±0.5%)
Net Charge (µC) per pulse00
Max output voltage (V) @500 ohms8.00V(±0.5%)8.00V(±0.5%)
Max output voltage (V) @2000 ohms32V(±0.5%)32V(±0.5%)
Max output voltage (V) @10000 ohms60V(±3)60V(±0.5%)
Max output current (mA) @500 ohms16mA(±0.5%)16mA(±0.5%)
Max output current (mA) @2000 ohms16mA(±0.5%)16mA(±0.5%)
Max output current (mA) @10000 ohms6mA(±0.5%)6mA(±0.5%)
Max phase charge (μC) @500Ω3.954
Max Current Density (mA/cm2, r.m.s.) @500Ω2.352.37
Max Average Power Density (W/cm2) @500Ω0.0000170.000017
Max Average Current (mA) @500Ω0.480.48

2. Sample sizes used for the test set and the data provenance

  • Sample Size for Test Set: Not applicable. This submission relies on non-clinical testing (bench testing for electrical safety, biocompatibility, etc.) and a comparison of technical specifications to the predicate device, not on a clinical performance test set with patient data.
  • Data Provenance: Not applicable for a clinical test set. The non-clinical testing results would originate from the testing laboratories.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable. There is no clinical test set requiring expert ground truth for diagnosis/classification in this 510(k) submission as it is not an AI/ML diagnostic device. The ground truth for electrical and biocompatibility testing is based on established engineering and safety standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. There is no clinical test set requiring adjudication in this submission.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device is a TENS device, not an AI-assisted diagnostic or treatment planning tool. Therefore, no MRMC study or AI assistance evaluation was conducted or required for this type of submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This device is a TENS device, not an AI algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

  • The "ground truth" for this type of device is based on engineering standards (e.g., IEC standards for electrical safety, performance, and home use) and biocompatibility standards. It's about meeting specified physical, electrical, and material safety parameters, not diagnostic accuracy.

8. The sample size for the training set

  • Not applicable. This submission is for a physical medical device (TENS) that does not employ machine learning or AI models, and thus does not have a "training set" in the context of AI/ML development.

9. How the ground truth for the training set was established

  • Not applicable for the same reason as above (no AI/ML training set). The device is tested against established performance and safety standards, not trained on data with established ground truth.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

September 13, 2018

WAT Medical Technology Inc. % Rodney Zhang CTO WAT Medical Technology Inc. Room703-711, No.2 North Taoyuan Road Ningbo, 315600 China

Re: K172450

Trade/Device Name: TENS device-HeadaTerm, eEspresso Regulation Number: 21 CFR 882.5891 Regulation Name: Transcutaneous electrical nerve stimulator to treat headache Regulatory Class: Class II Product Code: PCC Dated: August 1, 2017 Received: August 14, 2017

Dear Rodney Zhang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice

(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Timothy A. Marjenin -S

For Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172450

Device Name

TENS device-HeadaTerm, eEspresso

Indications for Use (Describe)

The TENS device-HeadaTerm is indicated for the prophylactic treatment of episodic migraine in patients 18 years of age or older.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for WAT Med. The logo consists of a stylized "W" in green and pink, followed by the text "WAT Med" in black. The text is in a bold, sans-serif font.

SECTION 1

SUMMARY OF SAFETY AND EFFECTIVENESS

SECTION 1 - 1 of 7

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for WAT Med. The logo features a stylized checkmark symbol in blue, green, and pink. To the right of the checkmark is the text "WAT Med" in a bold, sans-serif font.

510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS

(As Required by 21 CFR 807.92)

1. Date Prepared [21 CFR 807.92(a)(1)]

09/11/2018

2. Submitter's Information [21 CFR 807.92(a)(1)]

Company Name:WAT Medical Technology Inc.
Company Address:Room703-711, No.2 North Taoyuan Road, 315600,Ningbo, Zhejiang Province, P.R.C
Contact Person:Dr. Rodney Zhang
Phone:86-574-6506-0811
Email:13500799711@139.com

3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]

Trade Name:TENS device-HeadaTerm, eEspresso
Common Name:Transcutaneous electrical nerve stimulator to treat headache
Model No.YF-HT-W1.
Product Code:PCC
Regulation Number:21 CFR 882.5891
Device Class:II

4. Identification of Predicate Devices(s) [21 CFR 807.92(a)(3)[

The identification of predicates within this submission is as follow:

Predicate I
Manufacturer:STX-Med SPRL
Trade Name:Cefaly
Product Code:PCC
Classification Name:Transcutaneous electrical nerve stimulator to treatheadache
Regulation Number:21 CFR 882.5891
Classification:Class II
FDA 510 (k) #:K122566 (Primary), K160237

5. Description of the Device [21 CFR 807.92(a)(4)]

TENS device-HeadaTerm offers a drug-free scheme, which introduces electric impulses to act on the supraorbital nerve and supratrochlear nerve, the very nerves that transmit migraine pain, to treat and prevent migraine headaches. TENS device-HeadaTerm enables an electronic feedback mechanism to adjust the electrical

SECTION 1 - 2 of 7

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for WAT Med. The logo consists of a stylized "W" made up of blue, green, and pink elements, followed by the text "WAT Med" in bold, black letters. The logo is simple and modern, with a focus on the company name.

impulses to the specific requirements of the individual user for personalized efficacy and comfort.

The device could adhere to patients forehead with conductive paste, the current passing through the electrodes and release current stimulate the nerve underneath patients' forehead skin; the stimulation would alleviate migraine headaches, and during ramping up, patients could choose comfortable intensity level by pressing the host key to lock up the current voltage, and the patient contact material is biocompatibility safe. The conductive paste could be replaced once the paste lost its adhesiveness and patients are recommended to replace the conductive paste after 7 times of treatment uses.

6. Previous Device Names and Company:

Previous Device Name:

Other names appears in various test reports including but not limit to Biocompatibility reports, electrical safety, EMC are recorded as HeadaTerm-wet, HeadaTerm-wet(shell), HeadaTerm-wet device for preventing and treating primary headache, these names are the previous names being used for proposed products. Previous Company:

The company appears in various test reports is WAT Medical Technology (Ningbo) Co., Ltd, which is the same company as the current WAT Medical Technology Inc.

7. Indications for Use [21 CFR 807.92(a)(5)]

The TENS device-HeadaTerm is indicated for the prophylactic treatment of episodic migraine in patients 18 years of age or older.

8. Technological Characteristic [21 CFR 807.92(a)(6)]

By releasing the low frequency pulse with a particular frequency and reaching the advanced nerve center of cerebral cortex via nervus supraorbitalis, they can adjust all signals causing headache which come from the biological and physical channels, so as to stop or postpone the transmission of headache signal to cerebral center.

Clinical TestingNot Applicable
Non-Clinical TestingElectrical Safety
Electromagnetic Safety
Performance of nerve and muscle stimulators
Performance of Device for Home Use

9. Clinical Testing and Non-Clinical Testing

SECTION 1 - 3 of 7

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for WAT Med. The logo features a stylized "W" in blue, green, and pink, followed by the text "WAT Med" in bold, black letters. The logo is simple and modern, and the colors are bright and eye-catching.

Biocompatibility: Cytotoxicity
Biocompatibility: Irritation
Biocompatibility: Sensitization

The non-clinical testing conducted are listed above, and all tests come back with "Pass" results which demonstrating the same safety and effectiveness with the predicate device.

10. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92(b)(2)]

Proposed DevicePredicate Device
Product NameTENS device-HeadaTermCEFALY
Product CodePCCPCC
Regulation No.21 CFR 882.589121 CFR 882.5891
ClassificationClass IIClass II
Intended UseThe TENS device-HeadaTerm isindicated for the prophylactictreatment of episodic migraine inpatients 18 years of age or older.The Cefaly® device is indicated forthe prophylactic treatment ofepisodic migraine in patients 18years of age or older.
510 (k) NumberN/AK160237, K122566
Device Feature
Device NameTENS device-HeadaTermCEFALY
Power SourceOne 3V lithium coin cellsOne lithium battery
Follow CurrentYesYes
Voltage OverloadDetectionYesYes
Adjustable IntensityON/OFF button on frontof deviceON/OFF button on frontof device
Channel11
StartYesYes
Operation TipsLowBatteryNoYes
WorkYesYes
Software-controlledNoYes
Time SetNo - Manually AdjustNo - Manually Adjust
Constant CurrentNoNo
Automatic overloadtrip voltage levelYesYes
Patient overridecontrol methodOn/Off button on front of deviceOn/Off button on front of device
Indicator displaysUnit functioningElectrical connectionUnit functioningLow batteryElectrical connection
Timer SettingYesYes
Weight11g30g
Dimensions127x35x12mm160 x 170 x 40 mm
WaveformAC Symmetric SquareWaveAC Symmetric SquareWave
Phase Duration(µsec)250us(±20)250us(±0.5%)
Phase Interval10us(±0.5%)10us(±0.5%)
Pulse Period500us(±0.5%)505us(±0.5%)
Frequency (Hz)60Hz(±1)60Hz(±0.5%)
Net Charge (µC) perpulse00
Maximum outputvoltage (V):8.00V(±0.5%)8.00V(±0.5%)
@500 ohms
Maximum outputvoltage (V):@2000 ohms32V(±0.5%)32V(±0.5%)
Maximum outputvoltage (V):@10000 ohms60V(±3)60V(±0.5%)
Maximum outputcurrent (mA):@500 ohms16mA(±0.5%)16mA(±0.5%)
Maximum outputcurrent (mA):@2000 ohms16mA(±0.5%)16mA(±0.5%)
Maximum outputcurrent (mA):@10000 ohms6mA(±0.5%)6mA(±0.5%)
Maximum phasecharge (μC) @500Ω3.954
Maximum CurrentDensity, (mA/cm2,r.m.s.) @500Ω2.352.37
Maximum AveragePower Density,(W/cm2) @500Ω0.0000170.000017
Maximum AverageCurrent (averageabsolute value, mA)@500Ω0.480.48
IEC60601-1YesYes
IEC60601-1-2YesYes
IEC60601-2-10YesYes
IEC 60601-1-11YesNo

SECTION 1 - 4 of 7

{7}------------------------------------------------

W. WAT Med

SECTION 1 - 5 of 7

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for WAT Med. The logo consists of a stylized "W" in blue, green, and pink, followed by the text "WAT Med" in black. The text is in a bold, sans-serif font.

SECTION 1 - 6 of 7

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for WAT Med. The logo consists of a stylized "W" symbol in blue, green, and pink, followed by the text "WAT Med" in bold, black letters. The "W" symbol appears to be made up of three curved lines, with the left line in blue, the middle line in green, and the right line in pink.

11. Conclusion [21 CFR 807.92(b)(3)]

TENS device-HeadaTerm are substantially equivalent to predicate device because both devices use identical technology and same intended use, also testing standards are identical with the predicates. The differences between both devices are insignificant in terms of safety and effectiveness.

SECTION 1 - 7 of 7

§ 882.5891 Transcutaneous electrical nerve stimulator to treat headache.

(a)
Identification. A transcutaneous electrical nerve stimulator to treat headache is a device used to apply an electrical current to a patient's cranium through electrodes placed on the skin.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Appropriate analysis/testing must validate electromagnetic compatibility and electrical, mechanical, and thermal safety.
(3) The technical parameters of the device, including waveform, output modes, maximum output voltage and current (with 500, 2,000, and 10,000 ohm loads), pulse duration, frequency, net charge (µC) per pulse, maximum phase charge at 500 ohms, maximum current density (mA/cm
2 , r.m.s.), maximum average current (mA), maximum average power density (W/cm2 ), and the type of impedance monitoring system must be fully characterized.(4) Electrical performance, adhesive integrity, shelf life, reusability, and current distribution testing of the electrodes must be conducted.
(5) Appropriate software verification, validation, and hazard analysis must be performed.
(6) Clinical performance data must demonstrate that the device is safe and effective as a treatment for headache in the indicated patient population.
(7) Labeling must include the following:
(i) Appropriate contraindications such as not for use in subjects with an implanted metallic or electronic device in the head, a cardiac pacemaker, or an implanted or wearable defibrillator.
(ii) Appropriate warnings such as not to apply the device on the neck or chest, not to use the device in the presence of electronic monitoring equipment, not to use in the bath or shower, not to use while sleeping, not to use while driving, not to use while operating machinery.
(iii) Appropriate precautions such as the long-term effects of chronic use of the device are unknown.
(iv) A summary of the expected risks and benefits of using the device.
(v) A summary of the clinical performance data, including information on the patient population for which the device has and has not been demonstrated to be effective, and any adverse events and complications.
(vi) Information on how the device operates and the typical sensations experienced during treatment.
(vii) A detailed summary of the device technical parameters.
(viii) An expiration date/shelf life for the electrodes and the number of times they can be reused.
(ix) Disposal instructions.