K Number
K172438

Validate with FDA (Live)

Date Cleared
2018-01-12

(154 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ARBOR™ External Fixation System is indicated for use in construction of an external fixation frame for the treatment of pediatric and adult fractures and/or reconstruction of long bones (including metacarpal and metatarsal), and the pelvis.

The ARBOR™ External Fixation System is intended for:

  • Stabilization of open or closed fractures with soft tissue injuries;
  • Polytrauma;
  • Vertically stable pelvic fractures or as a treatment adjunct for vertically unstable pelvic fractures;
  • Arthrodesis and osteotomies with soft tissue problems;
  • Revision procedures where other devices have been unsuccessful including failures of total joints;
  • Neutralization of fractures stabilized with limited internal fixation;
  • Non-unions/septic non-unions;
  • Intraoperative reduction/stabilization tool to assist with indirect reduction;
  • Correction of deformity; and
  • Unilateral rectilinear bone segment transport or leg lengthening.
Device Description

The ARBOR™ External Fixation System is comprised of Schanz pins, external fixation clamps, and bars. The pins and bars are available in various sizes, and the fixation clamps are available in several designs all capable of use with any size pins and bars.

AI/ML Overview

This document describes the regulatory clearance of a medical device, the ARBOR™ External Fixation System, based on its substantial equivalence to previously cleared predicate devices. It does not present a study proving the device meets specific acceptance criteria through clinical or AI-driven performance metrics.

Therefore, I cannot provide a detailed answer to your request in the format you've outlined because the provided text describes a 510(k) premarket notification which relies on demonstrating similarity to existing devices, rather than presenting novel performance data from an independent study with an AI component or a clinical trial.

Here's why I cannot fulfill your request based on this document:

  • No AI Component: The document describes a traditional mechanical external fixation system. There is no mention of any artificial intelligence (AI) or machine learning (ML) components, nor any software algorithms that would require a study with acceptance criteria related to AI performance (e.g., sensitivity, specificity, AUC).
  • No Clinical Study Data with Acceptance Criteria: The performance data section refers to mechanical testing (ASTM F543, F1541 for pins, bars, clamps, including insertion torque, pullout, static torsion, static axial compression), bacterial endotoxin testing, biocompatibility, and MRI testing. These are engineering and material safety tests, not clinical performance studies with patient outcomes or diagnostic accuracy metrics. The key takeaway is substantial equivalence to predicates, not novel performance claims.
  • No Human Reader Study (MRMC): Since there's no diagnostic or AI component, there are no human readers involved in interpreting images, and thus no MRMC study.
  • No Ground Truth Establishment: Ground truth in the context of diagnostic AI refers to adjudicated labels for medical images or patient outcomes. For a mechanical device, performance is evaluated against engineering standards and material properties.

To directly address your points based on the absence of the requested information in this specific document:

  1. A table of acceptance criteria and the reported device performance:

    • Acceptance Criteria (Implied for Mechanical Device): Substantial equivalence in:
      • Mechanical properties (e.g., static torsion, static axial compression within ranges similar to predicates).
      • Biocompatibility.
      • Material composition.
      • Functional characteristics.
    • Reported Device Performance: "Performance data demonstrates substantial equivalence to the predicate devices" for mechanical tests. Specific numerical values are not provided as this is a summary for regulatory clearance, not a research paper.
  2. Sample size used for the test set and the data provenance: Not applicable in the context of an AI-driven or clinical diagnostic study. The "test set" here refers to the actual physical devices subjected to mechanical and materials testing. The provenance would be the manufacturer's testing labs.

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a diagnostic AI is established by clinical experts. For a mechanical device, performance is measured against engineering specifications.

  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.

  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable as there is no AI component or human reader interpretation involved.

  6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable as there is no algorithm or AI component.

  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): For this mechanical device, the "ground truth" relates to meeting established engineering and material standards (e.g., ASTM F543, F1541 for mechanical strength; ANSI/AAMI ST-72:2011 for bacterial endotoxin; ASTM F2052, F2182, F2213 for MRI compatibility).

  8. The sample size for the training set: Not applicable as there is no AI model that requires a training set.

  9. How the ground truth for the training set was established: Not applicable as there is no AI model or training set.

In summary, the provided FDA 510(k) clearance document is for a traditional mechanical medical device, not an AI-enabled one. The "study" mentioned refers to engineering and biocompatibility testing, designed to demonstrate the device's equivalence to existing products, rather than a clinical trial or AI performance study with human readers and ground truth establishment.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 12, 2018

Globus Medical Inc. Kelly Baker Senior Vice President/Regulatory and Clinical Affairs 2560 General Armistead Ave. Audubon, Pennsylvania 19403

Re: K172438

Trade/Device Name: ARBOR External Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: KTT, LXT Dated: December 12, 2017 Received: December 13, 2017

Dear Kelly Baker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172438

Device Name ARBORTM External Fixation System

Indications for Use (Describe)

The ARBOR™ External Fixation System is indicated for use in construction of an external fixation frame for the treatment of pediatric and adult fractures and/or reconstruction of long bones (including metacarpal and metatarsal), and the pelvis.

The ARBORTM External Fixation System is intended for:

  • · Stabilization of open or closed fractures with soft tissue injuries;
  • · Polytrauma;
  • · Vertically stable pelvic fractures or as a treatment adjunct for vertically unstable pelvic fractures;
  • · Arthrodesis and osteotomies with soft tissue problems;
  • · Revision procedures where other devices have been unsuccessful including failures of total joints;
  • · Neutralization of fractures stabilized with limited internal fixation;
  • · Non-unions/septic non-unions;
  • · Intraoperative reduction/stabilization tool to assist with indirect reduction;
  • · Correction of deformity; and
  • · Unilateral rectilinear bone segment transport or leg lengthening.
Type of Use (Select one or both, as applicable)
-----------------------------------------------------------------------------------------------------------------------------------------------

X | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary: ARBOR™ External Fixation System

Company:Globus Medical Inc.2560 General Armistead Ave.Audubon, PA 19403610-930-1800
Contact:Kelly J. Baker, Ph.D.Senior Vice President, Regulatory and Clinical Affairs
Date Prepared:January 9, 2018
Device Name:ARBOR™ External Fixation System
Common Name:External Fixation System
Classification:Per 21 CFR as follows:§888.3030 Single/multiple component metallic bonefixation appliances and accessoriesProduct Codes: KTT, LXTRegulatory Class: II, Panel Code: 87

Primary Predicate: Synthes Large External Fixation System (K031428)

AdditionalPredicates:Gexfix External Fixation System (K052605 & K160972)SBi Mini Rail External Fixation System (K093550)Stryker Hoffmann 3 (K122284) & Apex Pins (K061493)Synthes Medium External Fixation System (K040258)
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Purpose:

The purpose of this submission is to request clearance for the ARBOR™ External Fixation System.

Device Description:

The ARBOR™ External Fixation System is comprised of Schanz pins, external fixation clamps, and bars. The pins and bars are available in various sizes, and the fixation clamps are available in several designs all capable of use with any size pins and bars.

Indications for Use:

The ARBOR™ External Fixation System is indicated for use in construction of an external fixation frame for the treatment of pediatric and adult fractures and/or reconstruction of long bones, small bones (including metacarpal and metatarsal), and the pelvis.

{4}------------------------------------------------

The ARBOR™ External Fixation System is intended for:

  • . Stabilization of open or closed fractures with soft tissue injuries;
  • Polytrauma: ●
  • Vertically stable pelvic fractures or as a treatment adjunct for vertically . unstable pelvic fractures;
  • Arthrodesis and osteotomies with soft tissue problems; ●
  • . Revision procedures where other devices have been unsuccessful including failures of total joints;
  • . Neutralization of fractures stabilized with limited internal fixation;
  • . Non-unions/septic non-unions;
  • Intraoperative reduction/stabilization tool to assist with indirect reduction: ●
  • Correction of deformity; and ●
  • Unilateral rectilinear bone segment transport or leg lengthening. ●

Performance Data:

Performance of the ARBOR™ External Fixation System was evaluated in accordance with ASTM F543 and F1541. Engineering analysis and mechanical testing of pins, bars, and clamps was performed, including insertion torque, pullout, static torsion, and static axial compression. Performance data demonstrates substantial equivalence to the predicate devices.

Bacterial endotoxin testing (BET) was conducted in accordance with ANSI/AAMI ST-72:2011. Biocompatibility of patient-contacting materials was demonstrated by using materials that meet applicable standards or are used in 510(k) cleared devices. MRI testing was performed on the subject devices per ASTM F2052, F2182, and F2213.

Technological Characteristics:

The ARBOR™ External Fixation System implants have the same technological characteristics as the predicate devices including design, intended use, material composition, function, and range of sizes.

Conclusions:

The ARBOR™ External Fixation System has been found to be substantially equivalent to the predicate devices with respect to technical characteristics, performance, and intended use. The information provided within this premarket notification supports substantial equivalence to the predicate devices.

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.