K Number
K172325

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-10-26

(86 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Rhapsody H-30 Holmium Laser System is intended for incision, resection, ablation, vaporization, coagulation, hemostasis of soft tissue, or stone fragmentation during open surgical, laparoscopic, percutaneous or endoscopic procedures in urologic, gynecologic, gastroenterological, or pulmonary applications.

Device Description

The Rhapsody H-30 Holmium Laser System is an enclosed, transportable solid-state (i.e., Ho:YAG) laser source unit. The laser system unit produces a beam of coherent nearinfrared light (2100 nm wavelength) upon activation by depressing a footswitch pedal. The beam is directed to the treatment zone by means of a compatible optical delivery system, such as a holmium laser fiber. Components of the unit include a footswitch pedal and rotatable control panel. The system is a reusable, non-sterile device and is supplied in a wooden shipping crate.

AI/ML Overview

Here's an analysis of the provided text regarding the Rhapsody H-30 Holmium Laser System's acceptance criteria and studies:

Based on the provided FDA 510(k) summary, the Rhapsody H-30 Holmium Laser System is a medical device, not an AI/ML-driven device. Therefore, many of the requested categories related to AI/ML software performance (e.g., sample sizes for test/training sets, expert consensus, MRMC studies, standalone algorithm performance) are not applicable or explicitly mentioned in this type of submission.

The acceptance criteria and study information provided primarily relate to the safety and essential performance of the physical laser system and its firmware, demonstrating substantial equivalence to a predicate device.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Firmware Performance: Device firmware (LCK 1.6) performs as intended.Test results met predetermined criteria.
Electrical Safety and Essential Performance: Device complies with IEC 60601-1 (Medical electrical equipment Part 1: General requirements for basic safety and essential performance).Test results met predetermined criteria.
Output Energy Accuracy and Endurance: Acceptable output energy within ± 10% of factory setting. System fires for a minimum of 30 minutes at maximum output energy of 2.5 J @ 12 Hz on short pulse width.Test results met predetermined criteria.

2. Sample size used for the test set and the data provenance

  • Sample Size: Not explicitly stated for each test. These are likely engineering and performance tests on the physical device, not statistical tests on patient data.
  • Data Provenance: Not applicable in the context of patient data for a physical device. The tests were likely conducted by the manufacturer (Cook Incorporated) as part of their design verification and validation process.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable as this is a physical medical device, not an AI/ML diagnostic system requiring expert consensus on image interpretation or other data. The "ground truth" for these tests is defined by engineering standards and specifications (e.g., a specific power output, adherence to IEC 60601-1).

4. Adjudication method for the test set

  • Not applicable. The tests involve objective measurements against predefined engineering specifications and regulatory standards.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. This is not an AI/ML device, so MRMC studies are not relevant. This device is a surgical laser system.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • No. This is a physical medical device. The "firmware" mentioned (LCK 1.6) is embedded control software for the laser, not a standalone diagnostic or assistive algorithm. Its performance is integrated with the device's function.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • The "ground truth" for these performance tests is based on engineering specifications, manufacturing tolerances, and compliance with recognized international standards (e.g., IEC 60601-1 for electrical safety). For example, the laser's output energy is compared against its design specification (±10% of factory setting).

8. The sample size for the training set

  • Not applicable. This is a physical device, and the "firmware" mentioned does not involve machine learning requiring training data.

9. How the ground truth for the training set was established

  • Not applicable. (See #8).

Summary of the Study:

The studies performed for the Rhapsody H-30 Holmium Laser System were primarily engineering verification and validation tests. These tests were designed to ensure that the device met its design specifications and complied with relevant safety standards, demonstrating its reliability and performance. Key tests included:

  • Firmware Verification and Validation: To confirm that the internal LCK 1.6 firmware operates as intended.
  • Electrical Safety and Essential Performance Testing: To ensure compliance with IEC 60601-1, a critical standard for medical electrical equipment.
  • Laser Output Performance and Endurance Testing: To verify the laser's output energy accuracy (within ±10% of factory settings) and its ability to operate for extended periods at maximum output (30 minutes at 2.5 J @ 12 Hz).

The conclusion was that all predetermined acceptance criteria were met, supporting that the device meets its design requirements and is substantially equivalent to its predicate device (H-30 Holmium Laser System, K130444) without raising new safety or effectiveness concerns.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 26, 2017

Cook Incorporated Colin Jacob Capital Equipment Specialist, Regulatory Affairs 750 Daniels Way P.O. Box 489 Bloomington, Indiana 47402

Re: K172325

Trade/Device Name: Rhapsody H-30 Holmium Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: July 31, 2017 Received: August 1, 2017

Dear Colin Jacob:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820):

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely, Jennifer R. Stevenson -53 For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172325

Device Name Rhapsody H-30 Holmium Laser System

Indications for Use (Describe)

The Rhapsody H-30 Holmium Laser System is intended for incision, resection, ablation, vaporization, coagulation, hemostasis of soft tissue, or stone fragmentation during open surgical, laparoscopic, percutaneous or endoscopic procedures in urologic, gynecologic, gastroenterological, or pulmonary applications.

Type of Use (Select one or both, as applicable)

✖ Prescription Use (Part 21 CFR 801 Subpart D)
❍ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Cook Medical logo, which is white text on a red background. The word "COOK" is in large, bold letters at the top. Below that, the word "MEDICAL" is in smaller letters. Below the logo is the text "510(k) Summary" in a large, bold font.

COOK
INCORPORATED
750 DANIELS WAY, P.O. BOX 489
BLOOMINGTON, IN 47402-0489 U.S.A.
PHONE: 812.339.2235 TOLL FREE: 800.457.4500
WWW.COOKMEDICAL.COM

Rhapsody H-30 Holmium Laser System (21 CFR § 878.4810) Date Prepared: October 26, 2017

Submitted By:

Applicant:Cook Incorporated
Contact:Colin S. Jacob
Applicant Address:Cook IncorporatedP.O. Box 489750 Daniels WayBloomington, IN 47402
Contact Phone Number:(812) 335-3575 x 104965
Contact Fax Number:(812) 332-0281

Device Information:

Trade Name:Rhapsody H-30 Holmium Laser System
Common Name:Powered Laser Surgical Instrument
Classification Name/Panel:General & Plastic Surgery
Regulation:21 CFR § 878.4810
Product Code:GEX

Predicate Devices:

  • . H-30 Holmium Laser System (K130444)

Device Description:

The Rhapsody H-30 Holmium Laser System is an enclosed, transportable solid-state (i.e., Ho:YAG) laser source unit. The laser system unit produces a beam of coherent nearinfrared light (2100 nm wavelength) upon activation by depressing a footswitch pedal. The beam is directed to the treatment zone by means of a compatible optical delivery system, such as a holmium laser fiber. Components of the unit include a footswitch pedal and rotatable control panel. The system is a reusable, non-sterile device and is supplied in a wooden shipping crate.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Cook Medical. The logo is a red rectangle with the word "COOK" in white, sans-serif font. Below the word "COOK" is the word "MEDICAL" in a smaller, white, sans-serif font. The bottom left corner of the rectangle is cut off at an angle.

COOK INCORPORATED

750 DANIELS WAY, P.O. BOX 489

BLOOMINGTON, IN 47402-0489 U.S.A.

PHONE: 812.339.2235 TOLL FREE: 800.457.4500

WWW.COOKMEDICAL.COM

Indications for Use:

The Rhapsody H-30 Holmium Laser System is intended for incision, excision, resection, ablation, vaporization, coagulation, hemostasis of soft tissue, or stone fragmentation during open surgical, laparoscopic, percutaneous or endoscopic procedures in urologic, gynecologic, gastroenterological, or pulmonary applications.

Comparison to Predicate Devices:

The Rhapsody H-30 Holmium Laser System and the predicate device, H-30 Holmium Laser System (K130444), are substantially equivalent in that these devices have the same intended use and principles of operation. The differences in technological characteristics (i.e., software and minor internal components) do not raise different questions of safety and/or effectiveness when comparing to the predicate device.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Cook Medical logo. The word "COOK" is in white, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in white, sans-serif font on a red background.

COOK INCORPORATED 750 DANIELS WAY, P.O. BOX 489 BLOOMINGTON, IN 47402-0489 U.S.A. PHONE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM

Comparison to Predicate Device

PRIMARY PREDICATESUBJECT DEVICE
H-30 Holmium Laser SystemRhapsody H-30 Holmium Laser System
510(k)K130444Subject of submission
ManufacturerCook IncorporatedCook Incorporated
Regulation21 CFR § 878.4810General & Plastic SurgeryIDENTICAL TO PREDICATE
Product CodeGEXIDENTICAL TO PREDICATE
ClassificationPowered Laser Surgical InstrumentIDENTICAL TO PREDICATE
Indications for UseThe H-30 Holmium Laser System is indicated for use in fragmentation of urinary calculi and soft tissue applications. The H-30 Holmium Laser System is indicated for use in urinary procedures where fragmentation of stones and soft tissue incision, hemostasis, vaporization, and ablation are indicated.The Rhapsody H-30 Holmium Laser System is intended for incision, excision, resection, ablation, vaporization, coagulation, hemostasis of soft tissue, or stone fragmentation during open surgical, laparoscopic, percutaneous or endoscopic procedures in urologic, gynecologic, gastroenterological, or pulmonary applications.
Principles of OperationGenerates pulsed laser energy for delivery to the surgical site via silica core laser fibersIDENTICAL TO PREDICATE
Fundamental Scientific TechnologySolid-state, holmium yttrium aluminum garnet (Ho:YAG), pulsed laser operating at 2100 nanometersIDENTICAL TO PREDICATE
PRIMARY PREDICATESUBJECT DEVICE
H-30 Holmium Laser SystemRhapsody H-30 Holmium Laser System
Technological Characteristics
Laser Wavelength2100 nmIDENTICAL TO PREDICATE
Maximum Output30 WIDENTICAL TO PREDICATE
Repetition Rate5 to 20 HzIDENTICAL TO PREDICATE
Visible Aiming BeamVariable Intensity <3 mW,532 nmIDENTICAL TO PREDICATE
Laser ActivationFoot switchIDENTICAL TO PREDICATE
Minimum Energy0.5 J at 5 HzIDENTICAL TO PREDICATE
Maximum Energy3.5 J at 5 HzIDENTICAL TO PREDICATE
Electrical Pulse WidthShort: 350 to 800 μsLong: 800 to 1200 μsIDENTICAL TO PREDICATE
Input Voltage115 & 230 V (full power)IDENTICAL TO PREDICATE
Fiber Communication1) Detects a resistor embedded within the fiber to identifythe size of the attached fiber and limit the laser energy tothe fiber's maximum allowable power output.2) Communicates with a printed circuit board (two-waycommunication) embedded within the laser fiber todisplay extra information onscreen, or recording useinformation for the laser system and the fiber such aspower setting, number of pulses, and error codes.IDENTICAL TO PREDICATE
Max Flash-lamp Voltage700 VIDENTICAL TO PREDICATE
Blast ShieldUser accessibleIDENTICAL TO PREDICATE
Water ReservoirPlasticIDENTICAL TO PREDICATE
FilterDe-ionized cartridge & strainerIDENTICAL TO PREDICATE
EnclosureMetal with cabinet & flat topIDENTICAL TO PREDICATE
Control PanelRotatableIDENTICAL TO PREDICATE
Water PumpFixedFixed with slightly different orientation
Firmware1.5 version1.6 version
Fan ShroudsNonePresent
FerriteNonePresent

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is in white, uppercase letters, and above the word "MEDICAL" which is also in white, uppercase letters.

COOK INCORPORATED 750 DANIELS WAY, P.O. BOX 489 BLOOMINGTON, IN 47402-0489 U.S.A. PHONE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM

Comparison to Predicate Devices (continued)

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image features the logo for Cook Medical. The logo consists of the word "COOK" in large, white, sans-serif font, set against a red square. Below the word "COOK", there is a smaller red rectangle containing the word "MEDICAL" in white, sans-serif font. The overall design is simple and clean, with a focus on the company name and its association with the medical field.

COOK INCORPORATED WW.COOKMEDICA

Performance Data:

The subject device, Rhapsody H-30 Holmium Laser System, was subjected to applicable testing to assure reliable design and performance under the testing parameters. The tests are listed below:

Table 2.0-2: Rhapsody H-30 Holmium Laser System Testing Summary
Test ReportResult
Test # RT20110616Testing must demonstrate that LCK 1.6 firmware performs asintended. Test results met predetermined criteria.
H-30 Litho family: Software Verification andValidation
Test # 28110456-001Test article must demonstrate to comply with IEC 60601-1
IEC 60601-1 Medical electrical equipmentPart 1: General requirements for basic safetyand essential performanceMedical electrical equipment Part 1: General requirements forbasic safety and essential performance. Test results metpredetermined criteria.
Test # 031714_502_PR #95252Test articles must demonstrate acceptable output energy within ±10% of the factory setting and each system must fire for aminimum of 30 minutes at maximum output energy of 2.5J@12Hz on short pulse width. Test results met predeterminedcriteria.
Evaluation of Rhapsody H-30 - Addition ofFan Shroud, Optimization of Cooling Linesand Alignment Procedure

Conclusion:

For these tests, all pre-determined acceptance criteria were met. The results of these tests show that the Rhapsody H-30 Holmium Laser System meets the design input requirements based on the intended use. Furthermore, these results support the conclusion that the differences between the subject device, Rhapsody H-30 Holmium Laser System, and the predicate device, H-30 Holmium Laser System, do not raise new questions of safety or effectiveness and support a determination of substantial equivalence.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.