K Number
K172311

Validate with FDA (Live)

Device Name
BioFlux Device
Manufacturer
Date Cleared
2017-12-15

(136 days)

Product Code
Regulation Number
870.2920
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The bioflux Device is intended for use by patients who either have or are at risk of having cardiac disease and those that demonstrate intermittent symptoms indicative of cardiac disease and require cardiac monitoring on a continuing basis for up to 30 days. The device continuously records ECG data and upon detection by an ECG analysis algorithm or manually initiated by the patient, automatically delivers the recorded cardiac activity to the server where it is presented and can be reviewed by a medical professional.

The data received from the bioflux device can be used by another device for arrhythmia analysis, reporting and signal measurements. The bioflux device is not intended to sound any alarms. The device does not deliver any therapy, administer any drugs, provide interpretive or diagnostic statements or provide for any life support.

bioflux is for prescription use only.

Device Description

The bioflux system consists of the bioflux device and the server. The bioflux device is a portable, battery-powered, wireless cardiac monitor which may be worn by a patient to record ECG and activity level data for up to 30 consecutive days. The device can capture patient activated and auto-triggered events such as Bradycardia, pause and Atrial Fibrillation as identified by an embedded arrhythmia detection algorithm. The device is designed to automatically deliver the data to the server. The data is delivered to the server wirelessly via mobile cellular dedicated network connection. A medical professional, using the server, can adjust and program the device configuration and autotriggering parameters.

bioflux device is not a life-supporting or life-sustaining system. Clinical judgment and experience are used to check and interpret the data.

AI/ML Overview

This document describes the BioFlux Device, a cardiac monitor. Since this device is a hardware device for monitoring and not an AI/ML algorithm for analysis, the typical acceptance criteria and study data provided in an AI/ML submission are not present.

However, I can extract information regarding overall performance testing and the product's intended function.

Here's a summary of the device's functional performance:

1. Acceptance Criteria and Reported Device Performance:

Acceptance Criteria CategoryReported Device Performance
ECG Signal RecordingContinuously records ECG signal.
ECG Data StorageStores ECG data in the device memory.
Event TransmissionTransmits manual or auto-activated event recordings to the server via mobile network connection.
Medical Professional Review CapabilityTransmitted data can be evaluated by a medical professional.
Arrhythmia Detection (Algorithm)Captures patient-activated and auto-triggered events such as Bradycardia, pause, and Atrial Fibrillation as identified by an embedded arrhythmia detection algorithm.
Server FunctionalityFacilitates data communication, provides data storage, and presents data for evaluation.
Device ProgrammingMedical professional can adjust and program device configuration and auto-triggering parameters via the server.

Study that proves the device meets the acceptance criteria:

The document mentions "Bench test results verify that Bioflux Monitor system can continuously record ECG signal, store ECG data in the device memory, and transmit manual or auto activated event recordings to the server via mobile network connection for evaluation by a medical professional. Test results verify that all requirements were met and that the Bioflux Monitor performs as designed."

2. Sample Size for Test Set and Data Provenance: Not applicable. The document describes bench testing for a hardware device, not a performance study involving a test dataset for an algorithm.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications: Not applicable. Ground truth for an algorithm's performance is not relevant for this type of hardware device submission.

4. Adjudication Method: Not applicable.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: Not applicable. This is for assessing human reader performance with and without AI assistance, which is not relevant for this device.

6. Standalone Performance Study (Algorithm Only):
While there's an "embedded arrhythmia detection algorithm," the document describes the device as the subject of the performance testing, not the algorithm in isolation. The core function is recording and transmitting, with the algorithm being a part of how events are "auto-triggered." The performance verification focuses on the system's ability to record, store, and transmit, effectively implying that the algorithm's output (triggered events) are then transmitted. There isn't a dedicated standalone performance study for just the algorithm discussed in this document.

7. Type of Ground Truth Used: Not applicable in the context of an algorithm's performance. The "ground truth" for the device's functionality would be its physical capability to record, store, and transmit ECG data, which was verified via bench testing against design specifications.

8. Sample Size for the Training Set: Not applicable. This document does not describe the development or training of a machine learning algorithm.

9. How the Ground Truth for the Training Set Was Established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health and Human Services logo. On the right is the FDA logo, which is a blue square with the letters FDA in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 15, 2017

Biotricity Inc. % Mark Job Official Correspondent Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, Minnesota 55313

Re: K172311

Trade/Device Name: BioFlux Device Regulation Number: 21 CFR 870.2920 Regulation Name: Telephone Electrocardiograph Transmitter and Receiver Regulatory Class: Class II Product Code: DXH Dated: November 14, 2017 Received: November 21, 2017

Dear Mark Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

MS. Hillebrand

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Food and Drug Administration

Indications for Use

Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K172311

Device Name BioFlux Device

Indications for Use (Describe)

The bioflux Device is intended for use by patients who either have or are at risk of having cardiac disease and those that demonstrate intermittent symptoms indicative of cardiac disease and require cardiac monitoring on a continuing basis for up to 30 days. The device continuously records ECG data and upon detection by an ECG analysis algorithm or manually initiated by the patient, automatically delivers the recorded cardiac activity to the server where it is presented and can be reviewed by a medical professional.

The data received from the bioflux device can be used by another device for arrhythmia analysis, reporting and signal measurements. The bioflux device is not intended to sound any alarms. The device does not deliver any therapy, administer any drugs, provide interpretive or diagnostic statements or provide for any life support.

bioflux is for prescription use only.

Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

(as required by 21CFR 807.92)

SUBMITTER l.

Biotricity - Measuring Vitals 275 Shoreline Drive, Suite 150 Redwood City CA 94065 Contact Person: Spencer LaDow sladow@biotricity.com Phone: 585-414-7407 Date Prepared: November 14, 2017

DEVICE II.

Name of Device: BioFlux Device Classification Name: 870.2920 Telephone electrocardiograph transmitter and receiver. Common or Usual Name: Mobile Cardiac Monitor Device Panel: Cardiovascular Regulatory Class: Class 2 Product Code: DXH

III. PREDICATE DEVICE

The bioflux System is substantially equivalent in intended use and similar technological characteristics the following device, Rhythmedix, LLC. RhythmStar System which was cleared under K141813.

{4}------------------------------------------------

DEVICE DESCRIPTION IV.

The bioflux system consists of the bioflux device and the server. The bioflux device is a portable, battery-powered, wireless cardiac monitor which may be worn by a patient to record ECG and activity level data for up to 30 consecutive days. The device can capture patient activated and auto-triggered events such as Bradycardia, pause and Atrial Fibrillation as identified by an embedded arrhythmia detection algorithm. The device is designed to automatically deliver the data to the server. The data is delivered to the server wirelessly via mobile cellular dedicated network connection. A medical professional, using the server, can adjust and program the device configuration and autotriggering parameters.

bioflux device is not a life-supporting or life-sustaining system. Clinical judgment and experience are used to check and interpret the data.

V. INDICATIONS FOR USE

The bioflux Device is intended for use by patients who either have or are at risk of having cardiac disease and those that demonstrate intermittent symptoms indicative of cardiac disease and require cardiac monitoring on a continuing basis for up to 30 days. The device continuously records ECG data and upon detection by an ECG analysis algorithm or manually initiated by the patient, automatically delivers the recorded cardiac activity to the server where it is presented and can be reviewed by a medical professional.

The data received from the bioflux device can be used by another device for arrhythmia analysis, reporting and signal measurements. The Bioflux device is not intended to sound any alarms. The device does not deliver any therapy, administer

any drugs, provide interpretive or diagnostic statements or provide for any life support. bioflux is for prescription use only.

{5}------------------------------------------------

SUMMARY OF TECHNOLIGICAL CHARACTERISTICS WITH THE PREDICATE VI. DEVICE

The bioflux system is substantially equivalent in intended use and similar technological characteristics of RhythmStar System cleared under K141813.

CategoryIdentical/DifferentbiofluxRhythmStar
510(k) NumberK172311K141813
ClassificationIdenticalMedicalMedical
NameIdenticalMobile Cardiac MonitorMobile Cardiac Monitor
Product CodeSimilarDXHDXH
Intended UseSimilarThe bioflux Device is intended foruse by patients who either have orare at risk of having cardiac diseaseand those that demonstrateintermittent symptoms indicative ofcardiac disease and require cardiacmonitoring on a continuing basis forup to 30 days. The devicecontinuously records ECG data andupon detection by an ECG analysisalgorithm or manually initiated by thepatient, automatically delivers therecorded cardiac activity to the serverwhere it is presented and can bereviewed by a medical professional.The data received from the biofluxdevice can be used by another devicefor arrhythmia analysis, reporting andsignal measurements. The biofluxdevice is not intended to sound anyalarms. The device does not deliverany therapy, administer any drugs,provide interpretive or diagnosticstatements or provide for any lifesupport. Bioflux is for prescription useonly.The RhythmStar system is intendedfor use by patients who either haveor are at risk of having cardiacdisease and those that demonstrateintermittent symptoms indicative ofcardiac disease and require cardiacmonitoring on a continuing basis. Thedevice continuously records ECG dataand upon detection by an ECGanalysis algorithm or manuallyinitiated by the patient, automaticallydelivers the recorded cardiac activityto the server where it is presentedand can be reviewed by a medicalprofessional.The data received from theRhythmStar device can be used byanother device for arrhythmiaanalysis, reporting and signalmeasurements. The RhythmStarsystem is not intended to sound anyalarms. The device does not deliverany therapy, administer any drugs,provide interpretive or diagnosticstatements or provide for any lifesupport. RhythmStar is forprescription use only.
Delivered deviceincludesSimilarbioflux device-patient 3 lead ECG cableRhythmStar monitor- Patient 3 lead ECG lead cable
CategoryIdentical/DifferentbiofluxRhythmStar
-internal rechargeable battery-Wall Battery charger-Carry pouch- 2 externally re-chargeable batteries- Wall battery charger
Monitor functional blocksSimilaranalog ECG front end,MCU,flash data storage,RF modem fordata transmission,LCD screen, andRecord buttonanalog ECG front end,accelerometer,MCU,flash data storage,RF modem fordata transmission,LCD screen, andRecord button
The server:identicalfacilitate data communication with the bioflux device, provide data storage, and present the data for evaluation by a medical professionalfacilitate data communication with the RhythmStar device, provide data storage, and present the data for evaluation by a medical professional
Device form factorSimilarsmall, lightweightambulatory cardiac monitors.small, lightweightambulatory cardiac monitors.
Wireless technology used to transmit data to serverIdenticalYesYes
Device is battery powered by a rechargeable Li-Ion batteryIdenticalYesYes
using a server, can adjust device programming parameters such as pre-post recording times and auto-triggering configuration.IdenticalYesYes
CategoryIdentical/DifferentbiofluxRhythmStar
Devices incorporate anaccelerometer tocapture activity leveldata related to patientmotion and deviceorientation.DifferentNoYes
devices have Recordbutton for manualevent recordings and auser interface toindicate device statusand mode of operation.similarYesYes
Device incorporateembedded ECG analysisalgorithm to auto-capture Bradycardia,Tachycardia and AtrialFibrillation andarrhythmia eventsbetween the signalacquisition point andthe server.SimilarYesYes
device has at least 2ECG channels and 3-lead electrodesSimilarYesYes
Functional,Environmental andElectricalcharacteristicsSimilarYesYes
USB connectiondifferentYes, used to charge the battery,marks ECG data during recording incase noise is introduced, not used fordata download.Yes, used only for data download andsecondary communication, cannot beconnected during ECG recording

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

PERFORMANCE TESTING VII.

The following performance and safety tests have been passed successfully:

  • . IEC 60601-2-47:2012 Medical electrical equipment - Part 2-47: Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems.
  • IEC 60601-1:2012 3rd Edition with amendment 1, Medical electrical equipment -Part 1: General requirements for basic safety and essential performance.
  • IEC 60601-1-2:2014 4th Edition, Medical Electrical Equipment - Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility - Requirements and tests.
  • IEC 60601-1-11:2015 Edition 1.1, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
  • ANSI/AAMI-EC 57:2012, Testing and Reporting Performance Results of Cardiac Rhythm and ST Segment Measurement Algorithms.
  • IEC 62366-1:2015 Edition 1.0, Medical devices Part 1: Application of usability ● engineering to medical devices
  • Biocompatibility testing of patient contacting materials according to ISO 10993-1.
  • . Bench test results verify that Bioflux Monitor system can continuously record ECG signal, store ECG data in the device memory, and transmit manual or auto activated event recordings to the server via mobile network connection for evaluation by a medical professional. Test results verify that all requirements were met and that the Bioflux Monitor performs as designed.

SUBSTANTIAL EQUIVALENCE RATIONALE VIII.

The intended use, performance and technological characteristics of the bioflux Monitor system compared to the named predicate device demonstrates that the bioflux Monitor is substantially equivalent to the predicate.

CONCLUSIONS IX.

The analysis of the differences between bioflux Monitor and the predicate deice does not raise new questions of safety and effectiveness. Based on device performance test results, Biotricity determines that the bioflux Monitor system performs within its design specifications, and is substantially equivalent to the predicate devices.

The information in this 510(k) submission demonstrates that the bioflux Monitor system is substantially equivalent to the predicate device.

§ 870.2920 Telephone electrocardiograph transmitter and receiver.

(a)
Identification. A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.(b)
Classification. Class II (performance standards).