K Number
K172296

Validate with FDA (Live)

Device Name
Stethee Pro
Manufacturer
Date Cleared
2017-10-30

(91 days)

Product Code
Regulation Number
870.1875
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Stethee™ Pro 1 is an electronic stethoscope intended for medical diagnostic purposes only. It may be used for the detection and amplification of sounds from the heart, lungs, arteries, veins, and other internal organs. It can be used on any person undergoing a physical assessment.

Device Description

The M3DICINE Stethee™ Pro 1 is an electronic stethoscope that picks up sounds of the heart, arteries, veins, lungs and other internal organs. Stethee™ Pro 1 provides capability for auscultation in frequencies from 20-2000Hz.

After amplification and filtering, the sounds are transferred to the user's ears via any Bluetooth connected headphones. The Stethee™ Pro 1 includes features that permit it to record and store sounds on a peripheral smart device (e.g., mobile phone) by transferring the sounds as an audio buffer to the smart device via a Bluetooth® connection. The smart device makes use of a downloaded companion stand-alone Mobile Medical Application (MMA), Stethee™ App, to record, manage and replay the captured patient sounds on the smart device.

Stethee™ App is also able to display the audio file as a phonocardiograph (21 CFR 870.2390) on the user's smart device and can use the audio file to calculate and display a simple Heart Rate. Stethee™ App does not direct or influence the performance of Stethee™ Pro 1.

The Stethee™ Pro 1 device user interface includes a simple push down mechanism ("control knob") to start recording. It has a light-emitting diode (LED) display ring which indicates the device's status state (ON or OFF, charge status, and Bluetooth® synchronization status).

The Stethee™ Pro 1 operates on a rechargeable certified 3.7V lithium ion polymer battery.

The Stethee™ Pro 1 incorporates embedded software. The embedded software controls all the various features found in the Stethee™ Pro 1, such as sound capture, digital signal processing, volume control, LED display ring, and Bluetooth® transfer.

AI/ML Overview

The document describes the Stethee™ Pro 1 electronic stethoscope and its equivalence to a predicate device, the 3M™ Littmann® Model 3200. However, it does not provide acceptance criteria in terms of specific performance metrics with numerical targets or a detailed study proving the device meets those specific acceptance criteria.

Instead, the document focuses on demonstrating substantial equivalence to a predicate device through comparative performance testing.

Here's an attempt to answer your questions based on the provided text, highlighting what information is available and what is not:

1. Table of acceptance criteria and the reported device performance

The document does not explicitly state "acceptance criteria" in the format of defined thresholds for different performance metrics. It rather presents a comparison of technical characteristics between the Stethee™ Pro 1 and the predicate device. The implied "acceptance criteria" is that the Stethee™ Pro 1 performs substantially equivalent to the predicate device in these characteristics.

FeatureImplied Acceptance Criteria (Substantial Equivalence to Predicate)Stethee™ Pro 1 Reported Performance
ClassificationElectronic Stethoscope, Class II, DQDElectronic Stethoscope, Class II, DQD
Intended UseMedical diagnostic purposes, detection/amplification of internal sounds, used on any person undergoing physical assessmentSame as predicate
ContraindicationsNoneNone
Pickup SensorConvert sound waves to digital signalMicrophone (-22 dB sensitivity), Sampling Rate: 16kHz, Bit Rate: 16 bits
Frequency ResponseResponsive in audible ranges, comparable to predicate's Bell/Diaphragm/Extended RangeResponsive from 20-2000Hz
Maximum Sound LevelAmplifies up to 24XAmplifies up to 24X
Volume ControlYes, in 8 stepsYes, in 8 steps
Power SourceBattery-poweredRechargeable Lithium-ion polymer battery
Low Battery IndicatorYesYes (LED color change)
Operating Controls & IndicatorsON/OFF, status indicationON/OFF features by electronics (tapping mechanism), LEDs for status (ON/OFF, charging, Bluetooth® synchronization)
Heart Rate: Detection & DisplayYes, displayed on device or peripheralYes, using companion Stethee™ App (MMA)
Heart Rate: Minimum Audio Sample Requirements5 seconds recording0.3-2 second recording
Heart Rate: MethodologyContinuous real-time calculation, updated every few secondsContinuous real-time calculation after initial 0.3-2 second sampling, updates after each heart beat
Heart Rate: Range of Detection & Accuracy30-199 BPM with ±10% consistency or ±5 bpm30-200 BPM with allowable readout error rate of no greater than ±10 % (i.e., 10% consistency) of the input rate or ±5 bpm
Heart Rate: Handling of inconsistent soundsInconsistent sounds indicated by "-"Inconsistent sounds displayed on Stethee™ App GUI as "-"
Direct ListeningReal-time sound listening capabilityReal-time listening using a Bluetooth® enabled headset
Recording and PlaybackCapability to record and playback soundsWith Stethee™ App (not on device itself)
VisualizationSounds visualized as phonocardiograph using softwareSounds visualized as phonocardiograph using companion software Stethee™ App
Wireless Technology & Peripheral Platform CompatibilityBluetooth® at 2.4 GHz, compatible with devicesBluetooth® at 2.4 GHz, compatible with Android and iOS devices
Ambient & Frictional Noise Reduction TechnologyYesYes

2. Sample size used for the test set and the data provenance

The document states: "M3DICINE submitted performance testing information in this 510(k) demonstrating that the Stethee™ Pro 1 can perform over its intended range of operation (20-2000Hz). Comparative performance testing submitted in the 510(k) demonstrated the two devices perform in a substantially equivalent manner."

However, no specific sample sizes for test sets or data provenance (e.g., country of origin, retrospective/prospective) are mentioned for this comparative performance testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the document. The testing described is comparative performance testing against a predicate device, not necessarily a diagnostic study requiring expert ground truth for classification.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided in the document.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was described or performed. The device is an electronic stethoscope with an accompanying app that can display phonocardiographs and heart rate. It's not an AI-assisted diagnostic tool in the sense of improving human reader performance on complex image interpretation. The comparison is primarily on technical specifications and basic functionality.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

The document describes the performance of the Stethee™ Pro 1 device itself, which includes embedded software for sound capture, digital signal processing, volume control, and Bluetooth transfer. The heart rate calculation by the Stethee™ App can be considered an "algorithm only" function in terms of its calculation process.

  • Heart Rate Calculation: The Stethee™ App "calculates and display a simple Heart Rate."
    • Minimum Audio Sample Requirements: Requires initial 0.3-2 second recording (predicate requires 5 seconds).
    • Methodology: Performs continuous real-time calculation after initial 0.3-2 second sampling and updates heart rate display after each heart beat (predicate updates every 2 seconds after initial 5 seconds).
    • Range of Detection and Accuracy: 30-200 BPM with an allowable readout error rate of no greater than ±10 % or ±5 bpm (predicate: 30-199 BPM with ±10% consistency).
  • Visualization: "Sounds can be visualized as phonocardiograph using companion software Stethee™ App." This is a standalone function of the software.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the heart rate accuracy, the implied ground truth would be a validated heart rate measurement, likely from an ECG or a highly accurate reference device. However, the exact type of ground truth used for performance testing is not explicitly stated. For the broader "substantial equivalence" claim, the predicate device itself serves as the reference for comparison.

8. The sample size for the training set

The document does not mention a training set or any machine learning/AI model training in the context of diagnostic capabilities beyond basic heart rate calculation. The "embedded software" handles signal processing and device control but isn't described as a learnable AI model.

9. How the ground truth for the training set was established

As no training set is mentioned in the context of the device's diagnostic claims, this information is not applicable/provided.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

November 9, 2017

M3dicine Pty. Ltd. % Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, Minnesota 55313

Re: K172296

Trade/Device Name: Stethee™ Pro 1 Regulation Number: 21 CFR 870.1875 Regulation Name: Stethoscope Regulatory Class: Class II Product Code: DQD Dated: July 27, 2017 Received: July 31, 2017

Dear Mark Job:

This letter corrects our substantially equivalent letter of October 30, 2017.

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

M. A. Wilhelmsen

for Bram D. Zuckerman, M.D.
Director Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172296

Device Name

Stethee™ Pro 1

Indications for Use (Describe)

The Stethee™ Pro 1 is an electronic stethoscope intended for medical diagnostic purposes only. It may be used for the detection and amplification of sounds from the heart, lungs, arteries, veins, and other internal organs. It can be used on any person undergoing a physical assessment.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92.

510(k) SubmitterM3DICINE Pty Ltd.Suite 14 BTP Hub8 Clunies Ross Court,Eight Mile Plains QLD 4113AUSTRALIA
-----------------------------------------------------------------------------------------------------------------------------
Contact personGinger Cantor, RAC
Centaur Consulting, LLC
Phone: (715) 307-1850
centaurconsultingllc@gmail.com

Date Summary Prepared: October 12, 2017

Trade Name:Stethee™ Pro 1
Classification Name:Electronic Stethoscope
Regulation Number:21 CFR §870.1875(b)
Product Code:DQD
Classification Panel:Cardiovascular
Device Classification:Class II

Intended Use/Indications for Use

The Stethee™ Pro 1 is an electronic stethoscope intended for medical diagnostic purposes only. It may be used for the detection and amplification of sounds from the heart, lungs, arteries, veins, and other internal organs. It can be used on any person undergoing a physical assessment.

Predicate Device:

3MTM Littmann® Model 3200 (K083903)

{4}------------------------------------------------

Description of Device:

The M3DICINE Stethee™ Pro 1 is an electronic stethoscope that picks up sounds of the heart, arteries, veins, lungs and other internal organs. Stethee™ Pro 1 provides capability for auscultation in frequencies from 20-2000Hz.

After amplification and filtering, the sounds are transferred to the user's ears via any Bluetooth connected headphones. The Stethee™ Pro 1 includes features that permit it to record and store sounds on a peripheral smart device (e.g., mobile phone) by transferring the sounds as an audio buffer to the smart device via a Bluetooth® connection. The smart device makes use of a downloaded companion stand-alone Mobile Medical Application (MMA), Stethee™ App, to record, manage and replay the captured patient sounds on the smart device.

Stethee™ App is also able to display the audio file as a phonocardiograph (21 CFR 870.2390) on the user's smart device and can use the audio file to calculate and display a simple Heart Rate. Stethee™ App does not direct or influence the performance of Stethee™ Pro 1.

The Stethee™ Pro 1 device user interface includes a simple push down mechanism ("control knob") to start recording. It has a light-emitting diode (LED) display ring which indicates the device's status state (ON or OFF, charge status, and Bluetooth® synchronization status).

The Stethee™ Pro 1 operates on a rechargeable certified 3.7V lithium ion polymer battery.

The Stethee™ Pro 1 incorporates embedded software. The embedded software controls all the various features found in the Stethee™ Pro 1, such as sound capture, digital signal processing, volume control, LED display ring, and Bluetooth® transfer.

Comparison to Predicate:

Classification ComparisonStethee™ Pro 1Predicate3MTM Littmann® Model3200510(k) 083903Comment
ClassificationElectronicStethoscopeElectronic StethoscopeSame
Regulation21 CFR 870.1875(b)Class II device21 CFR 870.1875(b)Class II deviceSame
FDA ProceedDQDDQDSame

{5}------------------------------------------------

Intended Use/Indications for Use/Contraindications Comparison
Stethee™ Pro 13M™ Littmann®Model 3200FDA 510(k) K083903Comment
Intended Use/Indications forUseThe Stethee™ Pro 1 is anelectronic stethoscopeintended for medicaldiagnostic purposes only.It may be used for thedetection and amplificationof sounds from the heart,lungs, arteries, veins, andother internal organs. Itcan be used on any personundergoing a physicalassessment.The 3M™ Littmann®Electronic Stethoscope, Model3200 is intended for medicaldiagnostic purposes only. Itmay be used for the detectionand amplification of soundsfrom the heart, lungs, arteries,veins, and other internalorgans with the use of aselective frequency. It can beused on any personundergoing a physicalassessment.Substantiallyequivalent
ContraindicationsNoneNoneSame
Technical Characteristics Comparison
FeatureStethee™ Pro 13MTM Littmann®Model 3200 (K083903)Comments
Pickup sensorand processingMicrophone (-22 dB sensitivity)Sampling Rate: 16kHzBit Rate: 16 bitsPiezoelectric sensorSampling Rate: 4 kHzBit Rate: 16 bitsBoth mechanismsconvert sound wavesinto a digital signal viadigital signalprocessing. They arefunctionally equivalent.
FrequencyResponseResponsive from 20-2000HzBell (20-1000 Hz)Diaphragm (20-2000 Hz)"Extended Range" (20-2000 Hz)Note that 3MLittmann® hasqualified these responseclaims in theirInstruction for Use asfollows:Littmann®'s Bell modeemphasizes 20-200Hz).Littmann®'sDiaphragm modeemphasizes 100-500Hz).Littmann®'s "ExtendedRange" provides morelow frequency responsebetween 50-500Hz.
MaximumSound LevelAmplifies up to 24XAmplifies up to 24XSame
Volume ControlYes, in 8 stepsYes, in 8 stepsSame
FeatureStethee™ Pro 13MTM Littmann®Model 3200 (K083903)Comments
Power SourceRechargeable Lithium-ionpolymer batterySingle AA, NiMH(rechargeable) or Lithiumbatteries may be usedThe Littmann® Model3200 uses a single AAalkaline battery as thedefault setting butstates rechargeableNiMH or Lithiumbatteries may be used.
Low BatteryIndicatorYesYesThe color of Stethee™Pro 1's LEDs on thedevice indicate lowbattery condition;Littmann® Model 3200uses a low battery iconin the LCD display onthe stethoscope.
OperatingControls andIndicatorsON/OFF features by electronics.LEDs on device indicate devicesstatus (ON/OFF, charging,Bluetooth® synchronization).ON/OFF features by electronics.LCD display on device indicatesdevice status.Stethee™ Pro 1 uses atapping mechanism toturn it on and off.Littmann® Model 3200uses a physical on/offbutton. Both arecontrolled throughelectronics andsoftware.
Heart Rate:Detection andDisplayYes -using companion Stethee™App (a MMA).Yes - displayed on theLittmann® Model 3200integrated LCD.Substantiallyequivalent. Heart Rateis not displayed ondevice, but is displayedon a peripheral devicerunning the companionStethee™ App.
Heart Rate:Minimum AudioSampleRequirementsRequires initial 0.3-2 secondrecording.Requires initial 5 secondrecording.Substantiallyequivalent
Heart Rate:MethodologyPerforms continuous real-timecalculation after initial 0.3-2second sampling and updatesheart rate display after eachheart beatThe heart rate is updated every 2seconds after the initial 5 secondsampling.Substantiallyequivalent
Heart Rate:Range ofDetection andAccuracy30-200 BPM with an allowablereadout error rate of no greaterthan ±10 % (i.e., 10%consistency) of the input rate or±5 bpm.30-199 BPM with an allowablereadout error rate of no greaterthan ±10 % (i.e., 10%consistency) of the input rate.Substantiallyequivalent
Heart Rate-Handling ofinconsistentsoundsInconsistent sounds displayed onStethee™ App GUI as "-".Inconsistent sounds displayed onLittmann® Model 3200 LCDscreen as "-".Substantiallyequivalent
Technical Characteristics Comparison
FeatureStethee™ Pro 13MTM Littmann®Comments
Model 3200 (K083903)
Direct ListeningStethee™ Pro 1 sounds can belistened to in real time using aBluetooth® enabled headset.Littmann® Model 3200 allowsdirect listening to sounds in realtime through the device'sattached binaurals.Substantiallyequivalent. Thehealthcare professionalcan use the device tolisten in real-time.
Recording andPlaybackNot on the device itself, but withStethee™ App.Yes - stores twelve (12) 30second tracks on deviceStethee™ App can beused to record andplayback the sounds onthe User's Smartdevice. Number ofsounds is only limitedby the phone'smemory.
VisualizationSounds can be visualized asphonocardiograph usingcompanion software Stethee™AppSounds can be visualized asphonocardiograph usingsoftware StethAssist™.Substantiallyequivalent. Bothdevices rely on externalsoftware to visualizethe sound, under thesame softwareclassification ofphonocardiograph(21 CFR 870.2390).
WirelessTechnology andPeripheralPlatformCompatibilityYes - uses Bluetooth® at 2.4GHzCompatible with Android andiOS devicesYes - uses Bluetooth® at 2.4GHz,Compatible with Androiddevices, not compatible withiOS devices**Littmann® web pagestates it is notcompatible with Appledevices.Littmann® Model 3200manual states anadditional USBBluetooth® adaptor isrequired for thisfunction.
Ambient &Frictional NoiseReductionTechnologyYesYesSubstantiallyequivalent.

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

Performance Testing

M3DICINE submitted performance testing information in this 510(k) demonstrating that the Stethee™ Pro 1 can perform over its intended range of operation (20-2000Hz). Comparative performance testing submitted in the 510(k) demonstrated the two devices perform in a substantially equivalent manner.

Safety, Electromagnetic Compatibility (EMC) and Coexistence Compliance

M3DICINE submitted information demonstrating compliance with the requirements of IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance (2012 Reprint; Ed. 3.1) and with the US translation of the IEC 60601-1 international standard (AAMI ES60601-1:2005 +A1).

M3DICINE also submitted information demonstrating full compliance with the international standard IEC 60601-1-2:2007, Medical Electrical Equipment -Part 1: General Requirements for Safety: Collateral Standard: Electromagnetic Compatibility Requirements and Tests (Ed. 3.0).

In addition to general safety and EMC testing, M3DICINE has successfully addressed FDA's requirements for wireless coexistence testing of the Stethee™ Pro 1.

Software Testing and Standards Compliance

M3DICINE submitted software verification and validation information and documentation required under FDA's Guidance Content of Premarket Submission for Software Contained in Medical Devices (May 11, 2005). M3DICINE is compliant with the requirements of IEC 62304:2015, Medical Device Software - Software Life Cycle Processes.

Animal and Clinical Testing

No animal or clinical testing was submitted in this 510(k).

Biocompatibility

M3DICINE's biocompatibility assessment and testing used standard risk assessment techniques of ISO14971:2007 and consideration of FDA & internationally recognized guidelines and standards (including ISO 10993-1 series). M3DICINE has concluded that the Stethee™ Pro 1 is safe for its intended use.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the word "MEDICINE" in a simple, sans-serif font. The letters are all in uppercase and are a light gray color. The "E" in the word is stylized with three horizontal lines instead of the traditional vertical line.

Conclusion

M3DICINE's evaluation of the substantial equivalence of the Stethee™ Pro 1 (with companion Stethee™ App) to the 3M™ Littmann® Model 3200 was based on a comparison of device classification, intended use, indications for use and contraindications, and technical characteristics. M3DICINE also compared the two device's risks and labeling. Based on this comparison, M3DICINE concludes that Stethee™ Pro 1 is substantially equivalent to the 3M™ Littmann® Model 3200 predicate device (K083903) without raising any new safety or effectiveness issues.

§ 870.1875 Stethoscope.

(a)
Manual stethoscope —(1)Identification. A manual stethoscope is a mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs.(2)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.(b)
Electronic stethoscope —(1)Identification. An electronic stethoscope is an electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.(2)
Classification. Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.