(137 days)
The Diagnostic Ultrasound System Xario 200 Model TUS-X200 and Xario 200 Model TUS-X200S are indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, intra-operative(abdominal), laparoscopic, pediatric, small organs, neonatal cephalic, adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal (conventional), musculo-skeletal (superficial), cardiac adult, cardiac pediatric, trans-esoph(cardiac) and peripheral vessel.
The Xario200 Model TUS-X200 and Model TUS-X200S are mobile diagnostic ultrasound systems. These systems are Track 3 devices that employ a wide array of probes including flat linear array, convex linear array, and sector array with frequency ranges between approximately 2 MHz to 12 MHz.
The provided text describes modifications to an ultrasound system, the Xario 200 Diagnostic Ultrasound System V6.0, and states that these modifications do not change the intended use of the device. The modifications primarily involve the addition of new transducers and implementation of previously cleared software options. Therefore, the acceptance criteria and study information provided are primarily focused on demonstrating that the new transducers and software features perform as intended and are substantially equivalent to previously cleared devices or features.
Acceptance Criteria and Reported Device Performance
The document does not explicitly state quantitative acceptance criteria in a table format for the device's overall performance. Instead, it refers to the device performing "as intended" for the new features based on comparisons to predicate software/devices. The criteria for demonstrating safety and effectiveness likely included successfully fulfilling the intended use, maintaining image quality, and adhering to applicable medical device standards.
| Acceptance Criteria (Inferred) | Reported Device Performance |
|---|---|
| New transducers (PLU-1202BT and others for specific applications) function as intended. | The various tables for each transducer (PSU-25BT, PSU-30BT, PSU-50BT, PSU-70BT, PVU-375BT, PVU-382BT, PVU-674BT, PVU-674MV, PVU-712BT, PVU-745BTF, PVU-745BTH, PVU-745BTV, PVU-681MVL, PVU-770ST, PVU-781VT, PVU-781VTE, PVL-715RS, PLU-704BT, PLU-1005BT, PLU-1202BT, PLU-1204BT, PET-512MC, PET-512MD, PET-805LA, PC-20M, PC-50M) are listed with their supported clinical applications and modes of operation, indicating suitability for their intended uses. For new indications, they are marked as 'N'. |
| 2D Wall Motion Tracking (2D WMT) performs as intended in comparison to predicate software. | Clinical images of volunteers were obtained, and it was concluded that 2D WMT implemented on the subject device performs as intended in comparison to the predicate software. Specifically, it enables cardiac wall trace, local wall motion tracking, wall motion information analysis and display, cardiac volume measurement, and cardiac function analysis. |
| Shear Wave feature performs as expected including measurement quantification accuracy and precision. | Bench testing using an elasticity phantom demonstrated that the implementation of Shear Wave onto the subject device performed as intended, including measurement quantification accuracy and precision. |
| The device maintains conformance with applicable standards. | The device is designed and manufactured under the Quality System Regulations (21 CFR § 820 and ISO 13485 Standards). Conforms with applicable parts of IEC60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 62304, AIUM RTD2-2004 Output Display, and ISO 10993-1 standards. Software Documentation for a Moderate Level of Concern was included as per FDA guidance. |
| Software validation is successful. | Successful completion of software validation is reported. |
Study Information
-
Sample size used for the test set and the data provenance:
- 2D Wall Motion Tracking: "Representative clinical images of volunteers were obtained." The exact sample size is not specified. The data provenance is implied to be clinical, but the country of origin is not detailed. Given the company is "Toshiba Medical Systems Corporation" in Japan and "Toshiba America Medical Systems, Inc" in the USA, it could be from either or both.
- Shear Wave: "Bench testing was done using an elasticity phantom." This indicates a non-human, controlled setting. No specific sample size (e.g., number of phantoms or measurements) is provided.
- Transducers: No specific test set size or data provenance is mentioned for the general performance of the new transducers beyond their listed clinical applications.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not specified in the provided text. The document states "it was concluded" for 2D WMT and "As concluded in the study" for Shear Wave, implying expert assessment, but details about the number or qualifications of these experts are missing.
-
Adjudication method for the test set: Not specified.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not explicitly stated. The document mentions 2D WMT "compared to predicate software", but this does not describe a comparative effectiveness study involving human readers with and without AI assistance for improved performance. The device itself is a diagnostic ultrasound system, not explicitly an AI-assisted diagnostic tool in the sense of a standalone AI algorithm for interpretation.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The document describes bench testing for Shear Wave and clinical image evaluation for 2D WMT. While these tests evaluate the algorithm/feature's performance and output, it's not clear if a standalone algorithm-only performance assessment without any human review or interpretation of the output was conducted or reported. However, for an ultrasound system, the output (images, measurements) is inherently for human interpretation.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- 2D Wall Motion Tracking: The phrase "Representative clinical images of volunteers were obtained" and "performs as intended" suggests a comparison against expected normal cardiac function and/or comparisons with the predicate device's output, likely adjudicated by clinical experts.
- Shear Wave: "Bench testing using an elasticity phantom" implies a known, controlled physical property as ground truth for measurement accuracy and precision.
-
The sample size for the training set: Not applicable and not mentioned. This document pertains to a diagnostic ultrasound system and its software features, not an AI/ML model that typically requires a training set.
-
How the ground truth for the training set was established: Not applicable, as there is no mention of a training set for an AI/ML model.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. The word "Administration" is on the second line.
December 12, 2017
Toshiba Medical Systems Corporation % Orlando Tadeo Manager, Regulatory Affairs Toshiba America Medical Systems, Inc 2441 Michelle Drive TUSTIN, CA 92780
Re: K172276
Trade/Device Name: Xario 200 Diagnostic Ultrasound System V6.0 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: July 26, 2017 Received: July 28, 2017
Dear Orlando Tadeo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
Page 2 - Orlando Tadeo
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert A. Ochs. Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
Device Name
Xario 200 Diaqnostic Ultrasound System V6.0
Indications for Use (Describe)
The Diaqnostic Ultrasound System Xario 200 Model TUS-X200 and Xario 200 Model TUS-X200S are indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diaqnosis in the following clinical applications: fetal, abdominal, intra-operative(abdominal), laparoscopic, pediatric, small orqans, neonatal cephalic, adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal (conventional), musculo-skeletal (superficial), cardiac adult, cardiac pediatric, trans-esoph(cardiac) and peripheral vessel.
Type of Use (Select one or both, as applicable)
ك Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
System: Xario 200 TUS-X200,TUS-X200S V6.0
Transducer:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI> | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | P | P | P | P | 2 | P | P | P | P | 4,5,6,7,8,9,11,12 | ||
| Abdominal | P | P | P | P | P | 2,3 | P | P | P | P | 4,5,6,7,8,9,11,12,14 | |
| Intra-operative (Abdominal) | P | P | P | P | 2 | P | P | P | 5,6,7,8,11 | |||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | P | P | P | P | 2 | P | P | P | 5,7 | |||
| Pediatric | P | P | P | P | P | 2,3 | P | P | P | P | 4,5,6,7,8,9,11,12 | |
| Small Organ (Note 1) | P | P | P | P | 2 | P | P | P | 4,5,6,7,8,9,10,11,12 | |||
| Neonatal Cephalic | P | P | P | P | P | 2,3 | P | P | P | 5,6,7,8,11 | ||
| Adult Cephalic | P | P | P | P | P | 3 | P | P | P | 7 | ||
| Trans-rectal | P | P | P | P | 2 | P | P | P | P | 4,5,6,7,8,9,12 | ||
| Trans-vaginal | P | P | P | P | 2 | P | P | P | P | 4,5,6,7,8,9,12 | ||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | P | P | P | P | 2 | P | P | P | 4,5,6,7,8,9,10,11,12 | |||
| Musculo-skeletal (Superficial) | P | P | P | P | 2 | P | P | P | 4,5,6,7,8,9,10,11,12 | |||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | P | P | P | P | P | 3 | P | P | P | 4, 7, 13(N) | ||
| Cardiac Pediatric | P | P | P | P | P | 3 | P | P | P | 4, 7, 13(N) | ||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | P | P | P | P | P | 3 | P | 4, 7, 13(N) | ||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | P | P | P | P | P | 2 | P | P | P | 4,5,6,7,8,9,10,11,12 | ||
| Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM
- Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking
- Note 14 Shear Wave
{4}------------------------------------------------
Xario 200 TUS-X200, TUS-X200S V6.0 System: _ Transducer: PSU-25BT
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | P | P | P | P | P | 3 | P | P | P | 7 | ||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | P | P | P | P | P | 3 | P | P | P | 7 | ||
| Small Organ (Note 1) | ||||||||||||
| Neonatal Cephalic | P | P | P | P | P | 3 | P | P | P | 7 | ||
| Adult Cephalic | P | P | P | P | P | 3 | P | P | P | 7 | ||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | P | P | P | P | P | 3 | P | P | P | 4, 7, 13(N) | ||
| Cardiac Pediatric | P | P | P | P | P | 3 | P | P | P | 4, 7, 13(N) | ||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking
Note 14 Shear Wave
{5}------------------------------------------------
System: Xario 200 TUS-X200, TUS-X200S V6.0 Transducer:_ PSU-30BT
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | P | P | P | P | P | 3 | P | P | P | 7 | ||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | P | P | P | P | P | 3 | P | P | P | 7 | ||
| Small Organ (Note 1) | ||||||||||||
| Neonatal Cephalic | P | P | P | P | P | 3 | P | P | P | 7 | ||
| Adult Cephalic | P | P | P | P | P | 3 | P | P | P | 7 | ||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | P | P | P | P | P | 3 | P | P | P | 4, 7, 13(N | ||
| Cardiac Pediatric | P | P | P | P | P | 3 | P | P | P | 4, 7, 13(N | ||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM
- Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking
- Note 14 Shear Wave
{6}------------------------------------------------
System: Xario 200 TUS-X200, TUS-X200S V6.0 Transducer:_ PSU-50BT
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI> | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | P | P | P | P | P | 3 | P | P | P | 7 | ||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | P | P | P | P | P | 3 | P | P | P | 7 | ||
| Small Organ (Note 1) | ||||||||||||
| Neonatal Cephalic | P | P | P | P | P | 3 | P | P | P | 7 | ||
| Adult Cephalic | P | P | P | P | P | 3 | P | P | P | 7 | ||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | P | P | P | P | P | 3 | P | P | P | 4, 7, 13(N) | ||
| Cardiac Pediatric | P | P | P | P | P | 3 | P | P | P | 4, 7, 13(N) | ||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking
- Note 14 Shear Wave
{7}------------------------------------------------
System: Xario 200 TUS-X200, TUS-X200S V6.0 Transducer:_ PSU-70BT
| Clinical Application | Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI | 4D | Other[Note] | |
| Ophthalmic | |||||||||||||
| Fetal | |||||||||||||
| Abdominal | P | P | P | P | P | 3 | P | P | P | 7 | |||
| Intra-operative (Abdominal) | |||||||||||||
| Intra-operative (Neuro) | |||||||||||||
| Laparoscopic | |||||||||||||
| Pediatric | P | P | P | P | P | 3 | P | P | P | 7 | |||
| Small Organ (Note 1) | |||||||||||||
| Neonatal Cephalic | P | P | P | P | P | 3 | P | P | P | 7 | |||
| Adult Cephalic | P | P | P | P | P | 3 | P | P | P | 7 | |||
| Trans-rectal | |||||||||||||
| Trans-vaginal | |||||||||||||
| Trans-urethral | |||||||||||||
| Trans-esoph. (non-Card.) | |||||||||||||
| Musculo-skeletal (Conventional) | |||||||||||||
| Musculo-skeletal (Superficial) | |||||||||||||
| Intravascular | |||||||||||||
| Other (Specify) | |||||||||||||
| Cardiac Adult | P | P | P | P | P | 3 | P | P | P | 4, 7, 13(N) | |||
| Cardiac Pediatric | P | P | P | P | P | 3 | P | P | P | 4, 7, 13(N) | |||
| Intravascular (Cardiac) | |||||||||||||
| Trans-esoph. (Cardiac) | |||||||||||||
| Intra-cardiac | |||||||||||||
| Other (Specify) | |||||||||||||
| Peripheral vessel | |||||||||||||
| Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM
- Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking
- Note 14 Shear Wave
{8}------------------------------------------------
System: Xario 200 TUS-X200, TUS-X200S V6.0 Transducer: __ PVU-375BT
| Clinical ApplicationSpecific(Tracks 3) | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI | 4D | Other[Note] | |
| Ophthalmic | ||||||||||||
| Fetal | P | P | P | P | P | 2 | P | P | P | 4,5,6,7,8,9,11,12 | ||
| Abdominal | P | P | P | P | P | 2 | P | P | P | 4,5,6,7,8,9,11,12,1 | ||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | P | P | P | P | P | 2 | P | P | P | 4,5,6,7,8,9,11,12 | ||
| Small Organ (Note 1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | ||||||||||||
| Musculo-skeletal(Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
Note 4 TDI
Note 5 ApliPure
Note 6 ApliPure Plus
Note 7 Precision Imaging
Note 8 Differential THI Note 9 Elastography
Note 9 Elastography
No. 10 BEAM
Note 10 BEAM
Note 11 Smart 3D Note 12 SMI
Note 13 2D Wall Motion Tracking
Note 14 Shear Wave
{9}------------------------------------------------
System: Xario 200 TUS-X200, TUS-X200S V6.0 Transducer: __ PVU-382BT
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI> | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | P | P | P | P | P | 2 | P | P | P | 5,6,7,8,11 | ||
| Abdominal | P | P | P | P | P | 2 | P | P | P | 5,6,7,8,11 | ||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | P | P | P | P | P | 2 | P | P | P | 5,6,7,8,11 | ||
| Small Organ (Note 1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking
- Note 14 Shear Wave
{10}------------------------------------------------
System: Xario 200 TUS-X200, TUS-X200S V6.0 Transducer: PVU-674BT
| Clinical ApplicationSpecific(Tracks 3) | Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI | 4D | Other[Note] | ||
| Ophthalmic | |||||||||||||
| Fetal | P | P | P | P | 2 | P | P | P | 5,6,7,8,11,1 | ||||
| Abdominal | P | P | P | P | 2 | P | P | P | 5,6,7,8,11,1 | ||||
| Intra-operative (Abdominal) | |||||||||||||
| Intra-operative (Neuro) | |||||||||||||
| Laparoscopic | |||||||||||||
| Pediatric | P | P | P | P | 2 | P | P | P | 5,6,7,8,11,1 | ||||
| Small Organ (Note 1) | |||||||||||||
| Neonatal Cephalic | |||||||||||||
| Adult Cephalic | |||||||||||||
| Trans-rectal | |||||||||||||
| Trans-vaginal | |||||||||||||
| Trans-urethral | |||||||||||||
| Trans-esoph. (non-Card.) | |||||||||||||
| Musculo-skeletal (Conventional) | |||||||||||||
| Musculo-skeletal (Superficial) | |||||||||||||
| Intravascular | |||||||||||||
| Other (Specify) | |||||||||||||
| Cardiac Adult | |||||||||||||
| Cardiac Pediatric | |||||||||||||
| Intravascular (Cardiac) | |||||||||||||
| Trans-esoph. (Cardiac) | |||||||||||||
| Intra-cardiac | |||||||||||||
| Other (Specify) | |||||||||||||
| Peripheral vessel | |||||||||||||
| Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM
- Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking
- Note 14 Shear Wave
{11}------------------------------------------------
System: Xario 200 TUS-X200, TUS-X200S V6.0 Transducer:_ PVU-674MV
| Clinical ApplicationSpecific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI | 4D | Other[Note] |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Ophthalmic | ||||||||||||
| Fetal | P | P | P | P | 2 | P | P | P | P | 5,6,7,8 | ||
| Abdominal | P | P | P | P | 2 | P | P | P | P | 5,6,7,8 | ||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | P | P | P | P | 2 | P | P | P | P | 5,6,7,8 | ||
| Small Organ (Note 1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking
- Note 14 Shear Wave
{12}------------------------------------------------
System: Xario 200 TUS-X200, TUS-X200S V6.0 Transducer: PVU-712BT
| Clinical Application | Mode of Operation | Other[Note] | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | |
| Ophthalmic | ||||||||||
| Fetal | ||||||||||
| Abdominal | P | P | P | P | P | 2 | P | P | P | 5,6,7,8,1 |
| Intra-operative (Abdominal) | ||||||||||
| Intra-operative (Neuro) | ||||||||||
| Laparoscopic | ||||||||||
| Pediatric | P | P | P | P | P | 2 | P | P | P | 5,6,7,8,1 |
| Small Organ (Note 1) | ||||||||||
| Neonatal Cephalic | P | P | P | P | P | 2 | P | P | P | 5,6,7,8,1 |
| Adult Cephalic | ||||||||||
| Trans-rectal | ||||||||||
| Trans-vaginal | ||||||||||
| Trans-urethral | ||||||||||
| Trans-esoph. (non-Card.) | ||||||||||
| Musculo-skeletal (Conventional) | ||||||||||
| Musculo-skeletal (Superficial) | ||||||||||
| Intravascular | ||||||||||
| Other (Specify) | ||||||||||
| Cardiac Adult | ||||||||||
| Cardiac Pediatric | ||||||||||
| Intravascular (Cardiac) | ||||||||||
| Trans-esoph. (Cardiac) | ||||||||||
| Intra-cardiac | ||||||||||
| Other (Specify) | ||||||||||
| Peripheral vessel | ||||||||||
| Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM
- Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking
- Note 14 Shear Wave
{13}------------------------------------------------
System: Xario 200 TUS-X200, TUS-X200S V6.0 Transducer: PVU-745BTF
| Clinical Application | Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI | 4D | Other[Note] | |
| Ophthalmic | |||||||||||||
| Fetal | |||||||||||||
| Abdominal | P | P | P | P | 2 | P | P | P | 5,6,7,8,1 | ||||
| Intra-operative (Abdominal) | P | P | P | P | 2 | P | P | P | 5,6,7,8,1 | ||||
| Intra-operative (Neuro) | |||||||||||||
| Laparoscopic | |||||||||||||
| Pediatric | |||||||||||||
| Small Organ (Note 1) | P | P | P | P | 2 | P | P | P | 5,6,7,8,1 | ||||
| Neonatal Cephalic | |||||||||||||
| Adult Cephalic | |||||||||||||
| Trans-rectal | |||||||||||||
| Trans-vaginal | |||||||||||||
| Trans-urethral | |||||||||||||
| Trans-esoph. (non-Card.) | |||||||||||||
| Musculo-skeletal (Conventional) | |||||||||||||
| Musculo-skeletal (Superficial) | |||||||||||||
| Intravascular | |||||||||||||
| Other (Specify) | |||||||||||||
| Cardiac Adult | |||||||||||||
| Cardiac Pediatric | |||||||||||||
| Intravascular (Cardiac) | |||||||||||||
| Trans-esoph. (Cardiac) | |||||||||||||
| Intra-cardiac | |||||||||||||
| Other (Specify) | |||||||||||||
| Peripheral vessel | |||||||||||||
| Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking
- Note 14 Shear Wave
{14}------------------------------------------------
System: Xario 200 TUS-X200, TUS-X200S V6.0 Transducer: PVU-745BTH
| Clinical ApplicationSpecific(Tracks 3) | Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI | 4D | Other[Note] | ||
| Ophthalmic | |||||||||||||
| Fetal | |||||||||||||
| Abdominal | P | P | P | P | P | 2 | P | P | P | 5,6,7,8,11 | |||
| Intra-operative (Abdominal) | P | P | P | P | P | 2 | P | P | P | 5,6,7,8,11 | |||
| Intra-operative (Neuro) | |||||||||||||
| Laparoscopic | |||||||||||||
| Pediatric | |||||||||||||
| Small Organ (Note 1) | P | P | P | P | P | 2 | P | P | P | 5,6,7,8,11 | |||
| Neonatal Cephalic | |||||||||||||
| Adult Cephalic | |||||||||||||
| Trans-rectal | |||||||||||||
| Trans-vaginal | |||||||||||||
| Trans-urethral | |||||||||||||
| Trans-esoph. (non-Card.) | |||||||||||||
| Musculo-skeletal (Conventional) | |||||||||||||
| Musculo-skeletal (Superficial) | |||||||||||||
| Intravascular | |||||||||||||
| Other (Specify) | |||||||||||||
| Cardiac Adult | |||||||||||||
| Cardiac Pediatric | |||||||||||||
| Intravascular (Cardiac) | |||||||||||||
| Trans-esoph. (Cardiac) | |||||||||||||
| Intra-cardiac | |||||||||||||
| Other (Specify) | |||||||||||||
| Peripheral vessel | |||||||||||||
| Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking
- Note 14 Shear Wave
{15}------------------------------------------------
System: Xario 200 TUS-X200, TUS-X200S V6.0 Transducer: PVU-745BTV
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI> | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | P | P | P | P | P | 2 | P | P | P | 5,6,7,8,1 | ||
| Intra-operative (Abdominal) | P | P | P | P | P | 2 | P | P | P | 5,6,7,8,1 | ||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Note 1) | P | P | P | P | P | 2 | P | P | P | 5,6,7,8,1 | ||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking
- Note 14 Shear Wave
{16}------------------------------------------------
System: _Xario 200 TUS-X200, TUS-X200S V6.0 Transducer: PVU-681MVL
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Note 1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | P | P | P | P | P | 2 | P | P | P | P | 5,6,7,8,12 | |
| Trans-vaginal | P | P | P | P | P | 2 | P | P | P | P | 5,6,7,8,12 | |
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM
- Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking
- Note 14 Shear Wave
{17}------------------------------------------------
System: Xario 200 TUS-X200, TUS-X200S V6.0 Transducer: PVU-770ST
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Note 1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | P | P | P | P | 2 | P | P | P | 4,5,6,7,8, | |||
| Trans-vaginal | P | P | P | P | 2 | P | P | P | 4,5,6,7,8, | |||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM
- Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking
- Note 14 Shear Wave
{18}------------------------------------------------
System: _Xario 200 TUS-X200, TUS-X200S V6.0 Transducer: PVU-781VT
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI> | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Note 1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | P | P | P | P | P | 2 | P | P | P | 4,5,6,7,8,9,1 | ||
| Trans-vaginal | P | P | P | P | P | 2 | P | P | P | 4,5,6,7,8,9,1 | ||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM
- Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking
- Note 14 Shear Wave
{19}------------------------------------------------
System: Xario 200 TUS-X200, TUS-X200S V6.0 Transducer: PVU-781VTE
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Note 1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | P | P | P | P | P | 2 | P | P | P | 4,5,6,7,8,9,10 | ||
| Trans-vaginal | P | P | P | P | P | 2 | P | P | P | 4,5,6,7,8,9,10 | ||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM
- Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking
- Note 14 Shear Wave
{20}------------------------------------------------
System: Xario 200 TUS-X200, TUS-X200S V6.0 Transducer: PVL-715RS
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Note 1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | P | P | P | P | P | 2 | P | P | P | 5,6,7,8,12 | ||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM
- Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking
- Note 14 Shear Wave
{21}------------------------------------------------
System: Xario 200 TUS-X200, TUS-X200S V6.0 Transducer:_ PLU-704BT
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamic Flow | Power | CHI> | 4D | Other [Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Note 1) | P | P | P | P | P | 2 | P | P | P | 5,6,7,8,10,11,1 | ||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | P | P | P | P | P | 2 | P | P | P | 5,6,7,8,10,11,1 | ||
| Musculo-skeletal (Superficial) | P | P | P | P | P | 2 | P | P | P | 5,6,7,8,10,11,1 | ||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | P | P | P | P | P | 2 | P | P | P | 5,6,7,8,10,11,1 | ||
| Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM
- Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking
- Note 14 Shear Wave
{22}------------------------------------------------
System: Xario 200 TUS-X200, TUS-X200S V6.0 Transducer: PLU-1005BT
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Note 1) | P | P | P | P | P | 2 | P | P | P | 4,5,6,7,8,9,10,11, | ||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | P | P | P | P | P | 2 | P | P | P | 4,5,6,7,8,9,10,11, | ||
| Musculo-skeletal (Superficial) | P | P | P | P | P | 2 | P | P | P | 4,5,6,7,8,9,10,11, | ||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | P | P | P | P | P | 2 | P | P | P | 4,5,6,7,8,9,10,11, | ||
| Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
Note 6 ApliPure Plus
Note 7 Precision Imaging
- Note 8 Differential THI
Note 9 Elastography
Note 10 BEAM
Note 11 Smart 3D
Note 12 SMI
Note 13 2D Wall Motion Tracking
Note 14 Shear Wave
{23}------------------------------------------------
System: Xario 200 TUS-X200, TUS-X200S V6.0 Transducer:_ PLU-1202BT
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI | 4D | Other [Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | N | N | N | N | N | 2 | N | N | N | 5,6,7,8,11 | ||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Note 1) | N | N | N | N | N | 2 | N | N | N | 4,5,6,7,8,9,11,12 | ||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | N | N | N | N | N | 2 | N | N | N | 4,5,6,7,8,9,11,12 | ||
| Musculo-skeletal (Superficial) | N | N | N | N | N | 2 | N | N | N | 4,5,6,7,8,9,11,12 | ||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | N | N | N | N | N | 2 | N | N | N | 4,5,6,7,8,9,11,12 | ||
| Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking
- Note 14 Shear Wave
{24}------------------------------------------------
System: Xario 200 TUS-X200, TUS-X200S V6.0 Transducer:_ PLU-1204BT
| Clinical ApplicationSpecific(Tracks 3) | Mode of Operation | Other[Note] | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI> | 4D | ||
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Note 1) | P | P | P | P | P | 2 | P | P | P | 4,5,6,7,8,9,10,11,1 | ||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | P | P | P | P | P | 2 | P | P | P | 4,5,6,7,8,9,10,11,1 | ||
| Musculo-skeletal (Superficial) | P | P | P | P | P | 2 | P | P | P | 4,5,6,7,8,9,10,11,1 | ||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | P | P | P | P | P | 2 | P | P | P | 4,5,6,7,8,9,10,11,1 | ||
| Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking
Note 14 Shear Wave
{25}------------------------------------------------
System: Xario 200 TUS-X200, TUS-X200S V6.0 Transducer:_ PET-512MC
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI | 4D | Other[Note] | |
| Ophthalmic | |||||||||||||
| Fetal | |||||||||||||
| Abdominal | |||||||||||||
| Intra-operative (Abdominal) | |||||||||||||
| Intra-operative (Neuro) | |||||||||||||
| Laparoscopic | |||||||||||||
| Pediatric | |||||||||||||
| Small Organ (Note 1) | |||||||||||||
| Neonatal Cephalic | |||||||||||||
| Adult Cephalic | |||||||||||||
| Trans-rectal | |||||||||||||
| Trans-vaginal | |||||||||||||
| Trans-urethral | |||||||||||||
| Trans-esoph. (non-Card.) | |||||||||||||
| Musculo-skeletal (Conventional) | |||||||||||||
| Musculo-skeletal (Superficial) | |||||||||||||
| Intravascular | |||||||||||||
| Other (Specify) | |||||||||||||
| Cardiac Adult | |||||||||||||
| Cardiac Pediatric | |||||||||||||
| Intravascular (Cardiac) | |||||||||||||
| Trans-esoph. (Cardiac) | P | P | P | P | P | 3 | P | 4, 7, 13(N) | |||||
| Intra-cardiac | |||||||||||||
| Other (Specify) | |||||||||||||
| Peripheral vessel | |||||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking
- Note 14 Shear Wave
{26}------------------------------------------------
System: Xario 200 TUS-X200, TUS-X200S V6.0 Transducer:_ PET-512MD
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI | 4D | Other[Note] | |
| Ophthalmic | |||||||||||||
| Fetal | |||||||||||||
| Abdominal | |||||||||||||
| Intra-operative (Abdominal) | |||||||||||||
| Intra-operative (Neuro) | |||||||||||||
| Laparoscopic | |||||||||||||
| Pediatric | |||||||||||||
| Small Organ (Note 1) | |||||||||||||
| Neonatal Cephalic | |||||||||||||
| Adult Cephalic | |||||||||||||
| Trans-rectal | |||||||||||||
| Trans-vaginal | |||||||||||||
| Trans-urethral | |||||||||||||
| Trans-esoph. (non-Card.) | |||||||||||||
| Musculo-skeletal (Conventional) | |||||||||||||
| Musculo-skeletal (Superficial) | |||||||||||||
| Intravascular | |||||||||||||
| Other (Specify) | |||||||||||||
| Cardiac Adult | |||||||||||||
| Cardiac Pediatric | |||||||||||||
| Intravascular (Cardiac) | |||||||||||||
| Trans-esoph. (Cardiac) | P | P | P | P | P | 3 | P | 4, 7, 13(Note) | |||||
| Intra-cardiac | |||||||||||||
| Other (Specify) | |||||||||||||
| Peripheral vessel | |||||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: N/A
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking
- Note 14 Shear Wave
{27}------------------------------------------------
System: Xario 200 TUS-X200, TUS-X200S V6.0 Transducer: PET-805LA
| Clinical Application | Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | DynamicFlow | Power | CHI | 4D | Other | [Note] |
| Ophthalmic | |||||||||||||
| Fetal | |||||||||||||
| Abdominal | |||||||||||||
| Intra-operative (Abdominal) | |||||||||||||
| Intra-operative (Neuro) | |||||||||||||
| Laparoscopic | P | P | P | P | 2 | P | P | P | 5,7 | ||||
| Pediatric | |||||||||||||
| Small Organ (Note 1) | |||||||||||||
| Neonatal Cephalic | |||||||||||||
| Adult Cephalic | |||||||||||||
| Trans-rectal | |||||||||||||
| Trans-vaginal | |||||||||||||
| Trans-urethral | |||||||||||||
| Trans-esoph. (non-Card.) | |||||||||||||
| Musculo-skeletal (Conventional) | |||||||||||||
| Musculo-skeletal (Superficial) | |||||||||||||
| Intravascular | |||||||||||||
| Other (Specify) | |||||||||||||
| Cardiac Adult | |||||||||||||
| Cardiac Pediatric | |||||||||||||
| Intravascular (Cardiac) | |||||||||||||
| Trans-esoph. (Cardiac) | |||||||||||||
| Intra-cardiac | |||||||||||||
| Other (Specify) | |||||||||||||
| Peripheral vessel | |||||||||||||
| Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM
- Note 11 Smart 3D Note 12 SMI
- Note 13 2D Wall Motion Tracking
- Note 14 Shear Wave
{28}------------------------------------------------
System: Xario 200 TUS-X200, TUS-X200S V6.0 Transducer: PC-20M
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Note 1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | P | |||||||||||
| Cardiac Pediatric | P | |||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | P | |||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking Note 14 Shear Wave
{29}------------------------------------------------
System: Xario 200 TUS-X200, TUS-X200S V6.0 Transducer:_ PC-50M
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) * | THI | AdvancedDynamicFlow | Power | CHI> | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Note 1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | P | |||||||||||
| Cardiac Pediatric | P | |||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | P | |||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K143027
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
- Note 10 BEAM Note 11 Smart 3D
- Note 12 SMI
- Note 13 2D Wall Motion Tracking
- Note 14 Shear Wave
{30}------------------------------------------------
510(k) SUMMARY
- SUBMITTER'S NAME: 1. Toshiba Medical Systems Corporation 1385 Shimoishigami Otawara-shi, Tochigi-ken, Japan 324-8550
-
- OFFICIAL CORRESPONDENT Naofumi Watanabe
- ESTABLISHMENT REGISTRATION: 3. 9614698
CONTACT PERSON: 4.
Orlando Tadeo, Jr. Manager, Regulatory Affairs Toshiba America Medical Systems, Inc 2441 Michelle Drive Tustin, CA 92780 (714) 669-7459
5. Date Prepared:
July 25, 2017
6. TRADE NAME(S):
Xario 200 Diagnostic Ultrasound System V6.0
7. COMMON NAME:
System, Diagnostic Ultrasound
8. DEVICE CLASSIFICATION:
Class II
Ultrasonic Pulsed Doppler Imaging System - Product Code: 90-IYN [per 21 CFR 892.1550] Ultrasonic Pulsed Echo Imaging System – Product Code: 90-IYO [per 21 CFR 892.1560] Diagnostic Ultrasonic Transducer - Product Code: 90-ITX [per 21 CFR 892.1570]
TOSHIBA AMERICA MEDICAL SYSTEMS
Image /page/30/Picture/18 description: The image shows the logo for Canon Group. The word "Canon" is written in red, with a stylized "o" that has a white diagonal line through it. Below the word "Canon" is the text "CANON GROUP" in black, block letters. The logo is simple and recognizable.
{31}------------------------------------------------
9. PREDICATE DEVICE:
| Product | Marketed by | 510(k) Number | Clearance Date |
|---|---|---|---|
| Xario200, V5.0 (Models:TUS-X200 and TUS-X200S) | Toshiba AmericaMedical Systems | K162155 | October 14, 2016 |
10. REASON FOR SUBMISSION:
Modification of a cleared device
11. DEVICE DESCRIPTION:
The Xario200 Model TUS-X200 and Model TUS-X200S are mobile diagnostic ultrasound systems. These systems are Track 3 devices that employ a wide array of probes including flat linear array, convex linear array, and sector array with frequency ranges between approximately 2 MHz to 12 MHz.
12. INDICATIONS FOR USE:
The Diagnostic Ultrasound System Xario 200 Model TUS-X200 and Xario 200 Model TUS-X200S are indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, intra-operative(abdominal), laparoscopic, pediatric, small organs, neonatal cephalic, adult cephalic, trans-vaginal, musculo-skeletal (conventional), musculo-skeletal(superficial), cardiac adult, cardiac pediatric, transesoph(cardiac) and peripheral vessel.
13. SUBSTANTIAL EQUIVALENCE:
This device is substantially equivalent to the Xario200 V5.0 Diagnostic Ultrasound System, 510(k) cleared under K162155, marketed by Toshiba America Medical Systems. The Xario 200 Model TUS-X200 Version 6.0 and Xario200 Model TUS-X200S Version 6.0 functions in a manner similar to and is intended for the same use as the predicate devices referenced within this submission. The subject device includes modifications to the cleared device which improves upon existing features. Also, new features determined to be substantially equivalent to features cleared under the predicate devices referenced within this submission. A comparison table is included in this submission which details the similarities and differences between the predicate devices and the subject device.
| Xario 200Model TUS-X200,TUS-X200S v5.0 | Xario 200Model TUS-X200 andTUS-X200S v6.0 | Comment | |
|---|---|---|---|
| 510(K) ControlNumber | K162155(Predicate Device) | N/A(Subject Device) | |
| Uninterruptible PowerSupply (UPS) | Not Available | Available | New UPS |
| PLU-1202BT | Not Available | Available | New transducer |
{32}------------------------------------------------
Previously cleared software options being implemented to the subject device:
| Application | 510(k) Clearance | Comments |
|---|---|---|
| Tissue Specific Optimization | Previously cleared under K151451 | No change from previous clearance |
| 2D Wall Motion Tracking | Previously cleared under K151451 | No change from previous clearance |
| Shear wave | Previously cleared under K151451 | No change from previous clearance. Available on TUS-X200 only |
| SMIColor | c-SMI previously cleared under K162155 | SMI available on TUS-X200 only |
| SMIMonochrome | m-SMI previously cleared under K151451 | Addition of monochrome mapping |
14. SAFETY:
The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the IEC60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 62304, AIUM RTD2-2004 Output Display and ISO 10993-1 standards.
15. TESTING
Design Control Activities including risk management following ISO14971, verification/validation testing and Acoustic Output testing (UD3, 2004) were conducted and included in this submission.
Performance Testing - Clinical Images
Representative clinical images of volunteers were obtained and it was concluded that the 2D WMT implemented on the subject device performs as intended in comparison to the predicate software. More specifically, 2D WMT enables the cardiac wall trace (initial contour settinq), the local wall motion tracking and provides wall motion information analysis and display, cardiac volume measurement, and cardiac function analysis.
Performance Testing - Bench
Bench testing was done using an elasticity phantom to demonstrate that the implementation of Shear Wave onto the subject device performed as expected including measurement quantification accuracy and precision. As concluded in the study, Shear Wave performed as intended.
Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in
{33}------------------------------------------------
Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.
Additionally, testing of the modified system was conducted in accordance with the applicable standards published by the International Electrotechnical Commission (IEC) for Medical Devices.
16. CONCLUSION
The modifications incorporated into the Xario 200 Model TUS-X200 Version 6.0 and Xario200 Model TUS-X200S Version 6.0 do not change the intended use of the device. Based upon bench testing, acquisition of representative clinical images, successful completion of software validation, application of risk management and design controls, it is concluded that this device is safe and effective for its intended use.
N/A