K Number
K172240

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-01-31

(189 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SPI Dental Implant System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function.

It is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

Device Description

SPI Dental Implant System is an internal hex implant system which comes in 3.3, 3.7, 4.2, 5.0, and 6.0 diameter. The implants come in lengths of 8, 10, 11.5, 13 and 16. Both straight and angled abutments are available.

The implant is a conical shape one with a grit blasted and acid etched surface. Standard abutments, standard narrow abutments, standard shoulder abutments, standard wide shoulder abutments, multiunit abutments, ball attachments, healing caps and standard 15° and 25° abutments are included in the system.

AI/ML Overview

This is an FDA 510(k) summary for a medical device (SPI Dental Implant System), not a study report for an AI/ML powered device. As such, it does not contain the information requested in your prompt regarding acceptance criteria and a study that proves the device meets those criteria, specifically concerning AI/ML performance.

The document describes standard engineering and biocompatibility testing for a dental implant, confirming its substantial equivalence to a predicate device. It includes:

  • Testing Summary:
    • Dynamic fatigue testing according to ISO 14801
    • Surface analysis (SEM and EDS)
    • Sterilization according to ISO 11137-1 and 11137-2 (implants), ISO 17665-1 and ISO 17665-2 (abutments)
    • Package integrity testing according to ASTM F1929-12
    • Accelerated aging according to ASTM F1980-07
    • Material compliance with ASTM F136
    • Biocompatibility (cytotoxicity) according to ISO 10993-5
    • Endotoxin testing according to USP 161

These tests are designed to demonstrate the physical, chemical, and biological safety and performance of the dental implant itself, not the performance of an AI/ML algorithm.

Therefore, I cannot provide the requested information about acceptance criteria, device performance, sample sizes, ground truth establishment, or multi-reader multi-case studies because the provided document does not pertain to an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.

January 31. 2018

MSI France Gerard Bouana 5 Rue des Halles Paris 75001 France

Re: K172240

Trade/Device Name: SPI Dental Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: October 24, 2017 Received: November 2, 2017

Dear Gerard Bouana:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mary S. Runner -S

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K172240

Device Name SPI Dental Implant System

Indications for Use (Describe)

The SPI Dental Implant System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function.

It is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

510k Summary January 29, 2018 SPI Dental Implant System K172240

Name and address: MSI Implants 5 Rue des Halles Paris 75001 France Contact Person: Gerard Bouana Phone Number: +6-52129536 Name of device: SPI Dental Implant System Classification Name: Endosseous dental implants CFR: 21 CFR 872.3640 Product Code: DZE, NHA

Device Description: SPI Dental Implant System is an internal hex implant system which comes in 3.3, 3.7, 4.2, 5.0, and 6.0 diameter. The implants come in lengths of 8, 10, 11.5, 13 and 16. Both straight and angled abutments are available.

The implant is a conical shape one with a grit blasted and acid etched surface. Standard abutments, standard narrow abutments, standard shoulder abutments, standard wide shoulder abutments, multiunit abutments, ball attachments, healing caps and standard 15° and 25° abutments are included in the system.

Indications for Use: The SPI Dental Implant System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function.

It is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

Testing Summary: Dynamic fatigue testing according to ISO 14801 was conducted to determine the abutments are strong enough for their intended use. Surface analysis according to the FDA guidance document was done including SEM and EDS. Sterilization according to ISO 11137-1 and 11137-2 was conducted on the implants. Abutment steam sterilization was done according to ISO 17665-1 and ISO 17665-2. Package integrity testing according to ASTM F1929-12 and accelerated aging according to ASTM F1980-07 was conducted. Materials used in the product meet ASTM F136 and the biocompatibility was demonstrated by testing the cytotoxicity according to ISO 10993-5. Endotoxin testing according to USP 161 was conducted.

Predicate Device: GP Implants Spiral Shape Dental Implant System K162299 Reference Device: PALTOP Dental Implant System K112795

{4}------------------------------------------------

Substantial Equivalence:

SPI Dental Implant System is substantially equivalent to Spiral Shape Dental Implant System in indications for use, materials, design, and fatigue performance.

CompanySPI Dental Implant SystemSpiral Shape Dental Implant System
Indications for UseThe Spiral Shape Dental implant System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function.It is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.The Spiral Shape Dental implant System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function.It is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Implant Diameters3.3, 3.7, 4.2, 5, 63.3, 3.7, 4.2, 5, 6
Implant Lengths8,10,11.5,13,16 (no 6 x 16)8,10,11.5,13,16 (no 6 x 16)
Material of devices included in the submissionTi-6AL-4V ELITi-6AL-4V ELI
Type of abutment and maximum angulationPre-manufactured of no more than 25°Pre-manufactured of no more than 25°
Interface type/shapeInternal hexInternal hex
MSI DeviceGP Implants Device
Spiral Shape InternalHex ImplantsDiameters of 3.3, 3.7,4.2, 5, and 6mmDiameters of 3.3, 3.7,4.2, 5, and 6mm
Cover screwCover screwCover screw

{5}------------------------------------------------

Multi-Unit AbutmentsMulti-unit abutments in heights of 1,2,3 and 4 mmMulti-unit abutments in heights of 1,2,3 and 4 mm
Ball attachmentsBall attachments in heights of 1,2,3,4,5, and 6mmBall attachments in heights of 1,2,3,4,5, and 6mm
Healing Caps 4.5 diameterHealing cap in 3,4,5,6 and 7mm heightHealing cap in 2 and 7mm height
Healing Caps 5.5 diameterHealing cap in 2,3,4,and 5mm heightHealing cap in 2,3,4,5,6,and 7mm height
Standard Titanium AbutmentStandard Titanium Abutment with height of 7 and 9 mmStandard Titanium Abutment with height of 7mm
Standard Narrow AbutmentStandard narrow abutment with heights of 7 and 9 mmStandard narrow abutment with heights of 7, 9, and 11mm
Standard Shoulder AbutmentStandard shoulder abutment in heights of 1,3 and 4mmStandard shoulder abutment in heights of 1,2,3 and 4mm
Standard Wide Shoulder AbutmentStandard Wide Shoulder Abutment with heights of 1,2,3,and 4mmStandard Wide Shoulder Abutment with heights of 1,2,3,and 4mm
Standard 15° AbutmentStandard 15° Abutment with heights of 8, 12 and 13mmStandard 15° Abutment with heights of 8, 12, and 13mm
Standard 25° AbutmentStandard 25° Abutment with heights of 9 and 12mmStandard 25° Abutment with heights of 9 and 12mm

Conclusion:

SPI Dental Implant System is substantially equivalent to GP Implants Spiral Shape Dental Implant System. They both have the same indications for use, are of the same material, and have internal hex connections. The abutments, healing caps, and angled abutments are offered in similar designs and heights. Performance testing demonstrates substantial equivalence to the identified predicate devices.

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.