K Number
K172213

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-10-23

(91 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MECRON™ Cannulated Screw System is indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, ligament fixation, fracture repair and fracture fixation, appropriate for the device.

Device Description

The MECRON™ Cannulated Screw System consists of headed and headless bone screws and corresponding washers. The screws are offered in diameters from 2.0mm thru 7.0mm (in 0.5 mm increments) and overall lengths from 8 mm (for smaller diameters) thru 120 mm (for larger diameters). The cannulated screws have self-tapping and self-drilling features and are manufactured from medical grade titanium alloy (per DIN EN ISO 5832-3 and ASTM F136).

AI/ML Overview

This document is a 510(k) premarket notification for a medical device (MECRON™ Cannulated Screw System). It does not describe acceptance criteria and a study proving a device meets those criteria in the context of an AI/ML or diagnostic device. Instead, it focuses on demonstrating substantial equivalence to a predicate device through mechanical testing for a physical orthopedic implant.

Therefore, I cannot extract the information required by your request about acceptance criteria, device performance, sample sizes, expert involvement, ground truth, or MRMC studies, as these concepts are not applicable or detailed in this specific type of regulatory submission for this device.

The document states:

  • "Clinical Data were not needed for these devices to show substantial equivalence."
  • "Merete has conducted mechanical analysis and functional worst case tests to demonstrate that the MECRON™ Cannulated Screw System has the mechanical properties necessary to perform as well or better than the predicate device... These tests have been performed in accordance with ASTM F543-13."

This indicates that the "study" proving the device met its criteria was a series of mechanical and functional worst-case tests, not a clinical study involving human data, AI models, or expert human readers. The "acceptance criteria" were likely performance benchmarks derived from the predicate device and established standards like ASTM F543-13.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 23, 2017

Merete GmbH Paul Munch Head of Regulatory Affairs Alt-Lankwitz 102 Berlin, 12247 Germany

Re: K172213

Trade/Device Name: MECRON Cannulated Screw System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC. HTN Dated: September 27, 2017 Received: October 2, 2017

Dear Mr. Munch:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.qov

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K172213

Device Name

MECRON™ Cannulated Screw System

Type of Use (Select one or both, as applicable)

Indications for Use (Describe)

The MECRON™ Cannulated Screw System is indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, ligament fixation, fracture repair and fracture fixation, appropriate for the device.

☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Merete. The logo consists of two interlocking circles inside of a larger circle on the left, and the word "merete" in a sans-serif font on the right. There is a registered trademark symbol to the upper right of the word "merete".

510(k) Summary of Safety and Effectiveness Information as required by 21 CFR 807.92

Date Prepared:27th July 2017
Submitted by:Merete GmbHAlt-Lankwitz 10212247 Berlin, Germany
Contact Person:Merete GmbHPaul Münch (Head of Regulatory Affairs)Alt-Lankwitz 10212247 Berlin, GermanyPhone: +49-30 779980-181Fax: +49-30 779980-177
Device Name:MECRON™ Cannulated Screw System
Regulation/Classification Names:888.3040 Smooth or threaded metallic bone fixation fastener888.3030 Single/ multiple metallic bone fixation appliancesand accessories
Device Product Code:HWC, HTN
Proposed Regulatory Class:Class II

Reference devices:

VILEX/DUVAL/ORTHEX Cannulated Bone Screw, Model F20-xx-00 K014154

Device Description:

The MECRON™ Cannulated Screw System consists of headed and headless bone screws and corresponding washers. The screws are offered in diameters from 2.0mm thru 7.0mm (in 0.5 mm increments) and overall lengths from 8 mm (for smaller diameters) thru 120 mm (for larger diameters). The cannulated screws have self-tapping and self-drilling features and are manufactured from medical grade titanium alloy (per DIN EN ISO 5832-3 and ASTM F136).

Indications for Use

The MECRON™ Cannulated Screw System is indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, ligament fixation, fracture repair and fracture fixation, appropriate for the size of the device.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Merete. The logo consists of two interlocking circles inside of a larger circle on the left. To the right of the circle is the name "merete" in a stylized font, with a registered trademark symbol above the "e" at the end of the name.

Comparison of technological characteristics with the predicate devices:

Merete has conducted mechanical analysis and functional worst case tests to demonstrate that the MECRON™ Cannulated Screw System has the mechanical properties necessary to perform as well or better than the predicate device and the reference device. These tests have been performed in accordance with ASTM F543-13. The MECRON Cannulated Screws have successfully passed all tests within the mechanical verification and have been proven to be mechanically as good as or better than the predicate device.

There are no differences in sterilization method or material in comparison to the predicate device. Sterilization has to be performed according to common valid standards as well as the material is specified according to common valid standards.

Clinical Data were not needed for these devices to show substantial equivalence.

Substantial Equivalence:

The MECRON™ Cannulated Screw System has passed all defined criteria, has performed as well or better than the predicate device and is therefore considered substantially equivalent to the cleared predicate device.

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.