K Number
K172203

Validate with FDA (Live)

Date Cleared
2017-10-18

(89 days)

Product Code
Regulation Number
N/A
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Mist Spray:
CryoDose TA OTC is used like ice for the temporary relief and reduction of minor pain and swelling from sprains, strains, bruising, contusions and minor sports injuries.

Stream Spray:
CryoDose TA OTC is used like ice for muscle spasm and for the temporary relief and reduction of minor pain and swelling from sprains, strains, bruising, contusions and minor sports injuries.

Device Description

Nuance Medical's CryoDose TAOTC (subject device) is an over-the-counter (OTC) device designed to deliver a standard mixture 245fa (1,1,1,3,3-Pentafluoropropane) and 134a (1,1,2-Tetrafluoroethane), which is being offered in two delivery configurations, Mist Spray (Model No. 1815) and Stream Spray (Model No. 1821). This mixture self-propels itself from the delivery system, which is designed to account for its low vapor pressure. The device's delivery system controls the amount of the CryoDose TA TTC mixture that is dispensed, the Mist Spray configuration produces very fine droplets that create cooling at the points of contact. The Stream Spray configuration produces a pinpoint stream that contacts the skin surface at a specific single location. Both configurations contain the same standard mixture thus there is not safety or effectiveness concerns between configuration. The skin is cooled through rapid evaporation of the non-medicated propellants.

AI/ML Overview

This document is a 510(k) Summary for the CryoDose TA OTC (Mist Spray and Stream Spray) device, which is a vapocoolant for temporary pain relief. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a traditional clinical study with acceptance criteria and performance metrics for a novel AI/software device. Therefore, many of the requested categories related to AI/software performance studies (like sample sizes for test/training sets, experts for ground truth, MRMC studies, standalone performance) are not applicable or cannot be extracted from this type of regulatory submission.

However, I can extract information related to the device's characteristics and the types of tests performed to support its substantial equivalence.

1. Table of Acceptance Criteria and Reported Device Performance

Since this is a submission for substantial equivalence based on a physical medical device (vapocoolant spray) and not an AI/software product, the "acceptance criteria" are related to demonstrating similarity to a predicate device rather than numerical performance thresholds for diagnostic accuracy. The "reported device performance" is essentially that the device was found to be "substantially equivalent" and met the requirements outlined for physical and chemical characteristics.

Acceptance Criteria CategorySpecific Acceptance Criteria (Implied)Reported Device Performance
BiocompatibilityMeet acceptance requirements of ISO 10993-5:2009 (Cytotoxicity), ISO 10993-10:2010 (Sensitization), and ISO 10993-10:2010 (Irritation) for a surface-contacting/breached or compromised surface with limited contact duration (≤ 24 hours)."Biocompatibility testing met each evaluation's acceptance requirements, which demonstrated the material's safety." (Reference device's results utilized as identical to subject device.)
Chemical CompositionIdentical aerosol profile: 245fa (1,1,1,3,3-Pentafluoropropane) at 95% and 134a (1,1,2-Tetrafluoroethane) at 5%."The subject device and predicate are composed of identical aerosols profiles... The subject device's aerosol is checked and verified upon receipt from the aerosol supplier to ensure the same chemical profile as the predicate." (Identical chemical composition confirmed.)
Structural & Parts CompositionCanisters, valves, and caps equivalent to the predicate device."Engineering verification measurements were taken and visual inspections were made to determine the canisters, valves, and caps were equivalent to the predicate." (Reference device's results utilized as identical to subject device.)
Directions for Use (DFU)Key elements including indication for use, intended use, precaution, warning, and contraindication statements, and treatment regimen equivalent to the predicate."All key elements of the Directions for Use (DFU) between the predicate and the subject device are equivalent... No substantial differences exist between the predicate and subject device's Directions for Use."
Temperature & Output Bench TestingEquivalent temperature output at the application surface and volume dispensed per actuation/total compared to the predicate/reference device."Comparative performance testing was conducted as it related to temperature output at the application surface and volume dispensed per actuation/total and equivalent results were demonstrated." (Reference device's results utilized as identical to subject device.)
Stability & Shelf LifeIdentity, strength, quality, and purity maintained throughout labeled dating period under controlled and accelerated conditions."All requirements were confirmed to meet established acceptance criteria."

2. Sample size used for the test set and the data provenance
Not applicable. This is not a study involving a "test set" of medical images or patient data for an AI/software device. The testing involved laboratory and bench testing of the physical properties and chemical composition of the device, utilizing data from an identical reference device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This is not a study assessing diagnostic performance against a ground truth established by experts.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a vapocoolant spray, not an AI-powered diagnostic or assistive tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a vapocoolant spray.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the biocompatibility tests, the "ground truth" was the established standards and acceptance criteria of ISO 10993. For chemical and structural composition, the "ground truth" was the known composition and structure of the predicate device. For performance characteristics, the "ground truth" was the expected performance and temperature/volume output of the predicate/reference device.

8. The sample size for the training set
Not applicable. This is not an AI/machine learning device.

9. How the ground truth for the training set was established
Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left and features a stylized human figure. The FDA logo is on the right and includes the agency's acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 18, 2017

Nuance Medical, LLC Marc Lieberman CEO 5931 Sea Lion Place, Suite 113 Carlsbad, California 92010

Re: K172203

Trade/Device Name: CryoDose TA OTC, Mist Spray and CryoDose TA OTC, Stream Spray Regulatory Class: Unclassified Product Code: MLY Dated: July 19, 2017 Received: July 21, 2017

Dear Marc Lieberman:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Michael J. Hoffmann -S

Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172203

Device Name

CryoDose TA OTC (Mist Spray and Stream Spray)

Indications for Use (Describe)

Mist Spray:

CryoDose TA OTC is used like ice for the temporary relief and reduction of minor pain and swelling from sprains, strains, bruising, contusions and minor sports injuries.

Stream Spray:

CryoDose TA OTC is used like ice for muscle spasm and for the temporary relief and reduction of minor pain and swelling from sprains, strains, bruising, contusions and minor sports injuries.

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for "CRYODOSE" by Nuance Medical. The word "CRYO" is in blue, and the word "DOSE" is in gray. Above the word "CRYODOSE" is a series of vertical lines in blue and gray.

K172203 – 510(k) Summary

Nuance Medical, LLC – CryoDose TAoTc

This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92(c).

I. SUBMITTER

II.

III.

Owner:Nuance Medical, LLC.5931 Sea Lion Place, Suite 113Carlsbad, CA 92010(760) 585-4849
Contact Person:Marc S. LiebermanNuance Medical, LLC.5931 Sea Lion Place, Suite 113Carlsbad, CA 92010(760) 585-4849
Date Prepared:October 17, 2017
DEVICE
Trade Name:CryoDose TAOTC, Stream Spray, Model #1821CryoDose TAOTC, Mist Spray, Model #1815
Common Name:Cold Spray - 245fa (1,1,1,3,3-Pentafluoropropane) and134a (1,1,1,2-Tetrafluoroethane)
Classification Name:Refrigerant Topical, Vapocoolant
Product Code:MLY
PREDICATE DEVICE
Primary:Gebauer's Instant Ice (Mist Spray & Stream Spray)Gebauer Company.K021726Legally marketed medical device
Reference Device:PainFreeze (Mist Spray & Stream SprayNuance MedicalK162218Legally marketed medical device

{4}------------------------------------------------

DEVICE DESCRIPTION IV.

The predicate is a legally marketed device.

Nuance Medical's CryoDose TAOTC (subject device) is an over-the-counter (OTC) device designed to deliver a standard mixture 245fa (1,1,1,3,3-Pentafluoropropane) and 134a (1,1,2-Tetrafluoroethane), which is being offered in two delivery configurations, Mist Spray (Model No. 1815) and Stream Spray (Model No. 1821). This mixture self-propels itself from the delivery system, which is designed to account for its low vapor pressure. The device's delivery system controls the amount of the CryoDose TA TTC mixture that is dispensed, the Mist Spray configuration produces very fine droplets that create cooling at the points of contact. The Stream Spray configuration produces a pinpoint stream that contacts the skin surface at a specific single location. Both configurations contain the same standard mixture thus there is not safety or effectiveness concerns between configuration. The skin is cooled through rapid evaporation of the non-medicated propellants.

CryoDose TAºTC, is substantially equivalent, if not identical, to the predicate device Gebauer's Instant Ice (Mist Spray and Stream Spray). Both devices are indicated for use to provide temporary relief and reduction of minor pain and swelling from sprains, strains, bruising, contusions or minor sport injuries.

INDICATIONS FOR USE V.

The Indications for Use is similar to the predicate device.

Mist Spray:

CryoDose TA TC is used like ice for the temporary relief and reduction of minor pain and swelling from sprains, strains, bruising, contusions and minor sports injuries.

Stream Spray:

CryoDose TA OTC is used like ice for muscle spasm and for the temporary relief and reduction of minor pain and swelling from sprains, strains, bruising, contusions and minor sports injuries.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The subject device has the same technical characteristics as the predicate device including materials, design, and energy source. There are no technological differences between the predicate and the submitted device. Refer to the following table for the comparison between the subject device and the predicate:

{5}------------------------------------------------

Comparison Chart – Technological Characteristics
Trade NameNuance Medical'sCryoDose TAOTCSubject DeviceGebauer's Instant IcePredicateNuance Medical's PainFreezeReference Device
TypeOTCOTCRx
ProductDesignPressurized dispensingcontainer, which includesthe vapocoolant, canister,valve, actuator, and capPressurized dispensingcontainer, whichincludes thevapocoolant, canister,valve, actuator, and capPressurized dispensingcontainer, which includes thevapocoolant, canister, valve,actuator, and cap
Indication forUse/ IntendedUseMist Spray:CryoDose TA OTC isused like ice for thetemporary relief andreduction of minor painand swelling from sprains,strains,bruising, contusions andminor sports injuries.Stream Spray:CryoDose TA OTC isused like ice for musclespasm and for thetemporary relief andreduction of minor painandswelling from sprains,strains, bruising,contusions and minorsports injuries.Clinical conditions thatmay respond to thecooling effect's ofGebauer's Instant IceTMMist Spray are to uselike ice for temporaryrelief and reduction ofminor pain and swellingfrom sprains, strains,bruising, contusions orminor sport injuries. ForGebauer's Instant IceTMStream spray they are touse like ice for thetemporary relief andreduction of minor painand swelling fromsprains, strains, bruising,contusions or minorsports injuries andmuscle spasms.PainFreeze™ Mist Spray andMedium Stream Spray arevapocoolants (skin refrigerants)intended for topical applicationto skin, intact mucousmembranes (oral cavity, nasalpassage ways and the lips) andminor open wounds.PainFreezeTM controls painassociated with injections(venipuncture, IV starts,cosmetic procedures), minorsurgical procedures (such aslancing boils, incisions,drainage of small abscesses andsutures) and the temporaryrelief of minor sports injuries(sprains, bruising, cuts andabrasions). PainFreezeTMMedium Stream Spray is alsointended for use themanagement of myofascialpain, restricted motion andmuscle tension.
Product FillVolume3.5oz (103.5mL)3.5oz (103.5mL)3.5oz (103.5mL)
VapocoolantComposition1,1,1,3,3-Pentafluoropropane(HFC-245fa) 95% and1,1,1,2-Tetrafluoroethane(HFC-134a) 5%1,1,1,3,3-Pentafluoropropane(HFC-245fa) 95% and1,1,1,2-Tetrafluoroethane (HFC-134a) 5%1,1,1,3,3-Pentafluoropropane(HFC-245fa) 95% and1,1,1,2-Tetrafluoroethane(HFC-134a) 5%
EnergyDeliveredThermal Energy viaRefrigerant SprayThermal Energy viaRefrigerant SprayThermal Energy via RefrigerantSpray
EnergyDepositedN/AN/AN/A
VapocoolantDischargeMethodDepress the ActuatorButton to release thevapocoolantDepress the ActuatorButton to release thevapocoolantDepress the Actuator Button torelease the vapocoolant
EnvironmentalCompatibilityNon-FlammableNon-FlammableNon-Flammable
MechanicalSafetyMechanism has positiveshut off releaseMechanism has positiveshut off releaseMechanism has positive shut-off release

{6}------------------------------------------------

Substantial Equivalence:

The predicate and reference device are legally marketed devices. The subject device has similar intended use (Indications for Use) as the predicate device. The reference device's Indications for Use is similar in that it is topically applied for temporary relief of minor sports injuries, however is also associated with injections and minor surgical procedures that is conducted by healthcare professionals. 21 CFR 807.107(b)(1)

The subject device is for Over-The-Counter (OTC) like the predicate. The reference device is the Rx version of the subject device.

The subject device has the same technological characteristics as the predicate and reference device. There are no technological differences, including no changes in the materials. design, energy source, or other features of the device from those of the predicate or reference device. 21 CFR 800.107(b)(2)(i)

The subject, predicate, and reference devices use identical (substantially equivalent) chemical composition by type and percent of components: 245fa (1,1,1,3,3-Pentafluoropropane) at 95% of the total and 134a (1,1,2-Tetrafluoroethane) at 5% of the total mixture.

The subject, predicate, and reference devices use the same materials, design, energy source and other features.

The subject device is a reverse-engineering of its predicate and identical to the reference device.

Differences between the subject, predicate, and reference device, if any, would be limited to discussion and promotion of product, marketing materials, and cosmetic labeling as well as the reference device is for prescription use while the subject device and predicate are OTC.

The subject device does not raise questions of safety and effectiveness efficacy different from the predicate or the reference device.

Labeling:

The labeling of subject device has been prepared to ensure the medical professional has adequate and clear instructions for safety and usage. The canister labeling, directions for use, safety and warning statements are substantially equivalent to the predicate.

Summary: CryoDose TAOTC is substantially equivalent and no technological differences exist to the predicate device. 21 CFR 800.107(b)(2)(i)

VII. PERFORMANCE DATA

The subject, predicate, and reference devices use the same materials, design and energy source and no technological differences exist. Tests were selected and performed to ensure the subject device's output performance are the same as its predicate. The following tests were performed in support of substantial equivalence:

Biocompatibility

Since the reference device is identical to the subject device's composition and indication for use, the biocompatibility results of the reference device were utilized. The subject/reference

{7}------------------------------------------------

device is categorized as a surface-contacting/breached or compromised surface with limited contact duration (≤ 24 hours). The biocompatibility test program is based on ISO 10993-1:2009, which the following evaluations were conducted:

  • Cytotoxicity (ISO 10993-5: 2009) ●
  • Sensitization (ISO 10993-10:2010) ●
  • Irritation (ISO 10993-10:2010)

Biocompatibility testing met each evaluation's acceptance requirements, which demonstrated the material's safety.

Chemical Composition Confirmation

The subject device and predicate are composed of identical aerosols profiles, which is 245fa (1,1,1,3,3-Pentafluoropropane) at 95% of the total and 134a (1,1,2-Tetrafluoroethane) at 5% of the total mixture.

The subject device's aerosol is checked and verified upon receipt from the aerosol supplier to ensure the same chemical profile as the predicate.

Structural and Parts Composition

Since the reference device is identical to the subject device's construction, engineering verification results for the reference device were utilized. Engineering verification measurements were taken and visual inspections were made to determine the canisters, valves, and caps were equivalent to the predicate.

Directions for Use (Clinical Use) Application and Methodology:

All key elements of the Directions for Use (DFU) between the predicate and the subject device are equivalent including indication for use, intended use, precaution statements, warning statements, and contraindication statements, and treatment regiment. No substantial differences exist between the predicate and subject device's Directions for Use.

Side-by-Side Temperature & Output Bench Testing

Since the reference device is identical to the subject device's construction and composition, the performance results for the reference device were utilized. Comparative performance testing was conducted as it related to temperature output at the application surface and volume dispensed per actuation/total and equivalent results were demonstrated.

Stability Protocol and Shelf Life Testing

A stability protocol was developed to ensure that the identity, strength, quality, and purity of the product is maintained throughout its labeled dating period. Testing assessments were conducted under controlled conditions at room temperature and under accelerated conditions. All requirements were confirmed to meet established acceptance criteria.

Summarv

Based on the performance evaluations conducted as it relates to the subject device, as compared to the predicate, it was found to be safety and effectiveness and similar, if not identical, to the predicate. The following was demonstrated for substantial equivalence:

  • Material safety through biocompatibility

{8}------------------------------------------------

  • Identical chemical composition ●
  • Identical canister volume and construction ●
  • Identical clinical use, application and methodology
  • Equivalent temperature at application surface and volume dispersed from canister ●

VIII. CONCLUSION

Based on the information provided in the submission, it is concluded that the CryoDose TAOTC is safe and effective for its intended use and has demonstrated substantially equivalent to the predicated device.

N/A