(246 days)
WillCeph is intended for use by specialized dental practices for capturing, storing and presenting patient images and radiographs and to aid in cephalometric analysis, orthodontic treatment planning and case follow-up. Results produced by the software tools are to be interpreted by trained and licensed dental practitioners.
WillCeph is a software device for Cephalometric Analysis & Maxillo-facial Surgical Simulation, Consulting, Diagnosis, Surgical Planning, Analysis, Growth Forecast. It includes features for patient data management, various gallery formats, image fusion and comparing, cephalogram and tracing on digitized X-ray film, treatment simulation (VTO / STO), chart and analysis, superimposition, photo-film image fusion, growth forecast, and PACS integration (Option). It can be used as a client or standalone software on Microsoft Windows 7 or higher.
The provided document, a 510(k) premarket notification for the "WillCeph" device, contains a limited amount of information regarding acceptance criteria and study details. It explicitly states that "No clinical studies were considered necessary and performed." This means there is no study described in this document that proves the device meets specific acceptance criteria through clinical performance.
However, the document does refer to "Non-Clinical Test Summary" and states that "WillCeph ... was designed and developed according to a software development process and was verified and validated." While it doesn't provide specific numerical acceptance criteria or detailed performance reports, it implies that the device was tested against design specifications.
Based on the available information, here's a breakdown of what can be extracted:
1. Table of Acceptance Criteria and Reported Device Performance
As no clinical study was performed, there are no specific performance metrics reported for device clinical performance. The "acceptance criteria" here would be related to software verification and validation, but the document does not detail these criteria or their fulfillment numerically. The comparison table focuses on technological characteristics and intended use, demonstrating substantial equivalence to a predicate device, rather than explicit performance against predefined criteria.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Software designed and developed according to a software development process | "WillCeph was designed and developed according to a software development process" |
| Software verified and validated | "was verified and validated" |
| Substantial equivalence to predicate device in intended use and technological characteristics | Demonstrated through comparison table (see below for details) |
Summary of Technological Characteristics (extracted from the comparison table, serving as a basis for substantial equivalence, not performance):
| Feature | Proposed Device (WillCeph) | Predicate Device (ORTOMED - K131348) |
|---|---|---|
| Intended Use | Specialized dental practices for capturing, storing, and presenting patient images and radiographs, aiding in cephalometric analysis, orthodontic and orthognathic treatment planning, and case follow-up. Results interpreted by trained and licensed dental practitioners. | Same as Proposed Device |
| Type of Use | Prescription Use | Prescription Use |
| Modality Support | X-ray | X-ray |
| Component | Client or Standalone software | Client or Standalone software |
| Operating System | Windows 7 or higher | Windows 2000/XP, Vista and Windows 7 |
| User Interface | Mouse, Keyboard | Mouse, Keyboard |
| Image Communication Standard | DICOM | DICOM |
| Patient Data Management | Patient module for registration and file management; Image module for image management. | GESDEN module for patient registration and file management; GESIMAG module for image management. |
| Patient Database Engine | My SQL / Microsoft® SQL© | Microsoft® SQL© |
| Image Measurement | Yes (linear, distance, angle) | Yes (linear, distance, angle) |
| Matching / Adjustment of Photograph to X-ray | Yes | Yes |
| Cephalometric Analyses | User-configured analysis; Lateral, Frontal, Models Analyses; Photo studies of soft tissues. | User-configured analysis; Lateral, Frontal, Models Discrepancy Studies; Photo studies of soft tissue. |
| Treatment Planning, Simulation, and Follow-up | Orthodontic treatment (translate, tip, rotate incisors, reposition molars, rotate mandible); Growth Forecast; Growth simulation on traced x-ray; Superimpose growth tracings; VTO – Visual Treatment Objective; STO – Surgical Treatment Objective (orthognathic surgery); Warping and Morphing. | Orthodontic treatment (translate, tip, rotate incisors, reposition molars, auto-rotate mandible); Arch length discrepancy worksheet; CO/CR Conversion Growth Forecast (Ricketts algorithm); Growth simulation on traced x-ray; Superimpose growth tracings; VTO – Visual Treatment Objective; SVTO – Surgical Visual Treatment Objective (orthognathic surgery); Warping and Morphing. |
2. Sample Size Used for the Test Set and Data Provenance
Since "No clinical studies were considered necessary and performed," there is no test set in the traditional sense for clinical performance evaluation. The "test set" would implicitly refer to data used for software verification and validation, but details on sample size, country of origin, or retrospective/prospective nature are not provided.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Given the absence of clinical studies, there is no mention of experts or ground truth establishment for a clinical test set. The document focuses on regulatory compliance through substantial equivalence.
4. Adjudication Method for the Test Set
As no clinical test set is described, no adjudication method is mentioned.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study was performed or described. The document does not discuss human reader performance with or without AI assistance.
6. Standalone Performance
The product, WillCeph, is a standalone software that aids in cephalometric analysis. However, its effectiveness is assessed through substantial equivalence to a predicate device, not through a standalone clinical performance study. The statement "Results produced by the software tools are to be interpreted by trained and licensed dental practitioners" indicates human-in-the-loop, but the core functionality is algorithm-only. The document does not provide standalone performance metrics for the algorithm.
7. Type of Ground Truth Used
As no clinical study was performed, there is no explicit mention of ground truth types (e.g., expert consensus, pathology, outcomes data) for clinical performance evaluation. The "ground truth" for software verification and validation would likely be defined by internal software requirements and specifications.
8. Sample Size for the Training Set
The document does not mention a training set or its sample size. This suggests the device's functionality is based on established algorithms and calculations for cephalometric analysis, rather than a machine learning model trained on a large dataset for image interpretation.
9. How the Ground Truth for the Training Set Was Established
Since no training set is mentioned, there is no information on how ground truth for a training set was established.
In summary: The WillCeph device received 510(k) clearance based on demonstrating substantial equivalence to a legally marketed predicate device (ORTOMED - K131348). This means the FDA determined that the device is as safe and effective as the predicate device, primarily through a comparison of their intended use and technological characteristics, and through non-clinical software verification and validation activities (which are not detailed in terms of their criteria or results). No clinical studies or performance data against specific numerical acceptance criteria were deemed necessary or provided in this submission for device clearance.
{0}------------------------------------------------
March 21, 2018
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
HDX WILL CORP. % Myoung-Joon Lee General Manager #105, 106, 201, 202, 203, 204, 38, Osongsaengmyeong 4-ro, Osong-eup. Heungdeok-gu Cheongju-si, Chungcheongbuk-do 28161 KOREAN
Re: K172168
Trade/Device Name: WillCeph Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: January 22, 2018 Received: January 30, 2018
Dear Myoung-Joon Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K172168
Device Name WillCeph
Indications for Use (Describe)
WillCeph is intended for use by specialized dental practices for capturing, storing and presenting patient images and radiographs and to aid in cephalometric analysis, orthodontic treatment planning and case follow-up. Results produced by the software tools are to be interpreted by trained and licensed dental practitioners.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for HDX Will. The logo consists of a green and blue square with a white "W" inside, followed by the text "HDX Will" in blue. The text is in a bold, sans-serif font.
510(k) Summary
[As Required by 21 CFR 807.92]
1. Date Prepared [21 CFR 807.92(a)(a)]
July 12, 2017
2. Submitter's Information [21 CFR 807.92(a)(1)]
| • | Name of Sponsor: | HDX WILL CORP. |
|---|---|---|
| - | Address: | #105, 106, 201, 202, 203, 204, 38,Osongsaengmyeong 4-ro, Osong-eup, Heungdeok-gu,Cheongju-si, Chungcheongbuk-do, 28161, Korea |
| • | Contact Name: | Myoung-Joon Lee / General Manager |
| - | Telephone No.: | +82-43-710-7318 |
| - | Fax No.: | +82-43-710-7312 |
| - | Email Address: | mjlee@iwillmed.com |
| • | Name of Manufacturer: | Same as sponsor |
- -Address: Same as sponsor
3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
- Trade Name: WillCeph
- Common Name: ● PACS Software
- Classification:
| Classification Name | System, Image Processing, Radiological |
|---|---|
| Classification Regulation | 21 CFR 892.2050 |
| Product Code | LLZ |
| Device Class | II |
| Classification Panel | Radiology |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for HDX Will. The logo consists of a green square with a white "W" inside, followed by the text "HDX Will" in blue. The text is in a bold, sans-serif font and is slightly slanted to the right.
4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]
The identified predicate devices within this submission are shown as follow:
Predicate Device
- 510(k) Number: K131348 Applicant: ● INFOMED SERVICIOS INFORMATICOS Classification Name: Imaging Processing System ● ● Trade Name: ORTOMED
5. Description of the Device [21 CFR 807.92(a)(4)]
■ Overview
- Description
- Cephalometric Analysis & Maxillo-facial Surgical Simulation
- Consulting, Diagnosis, Surgical Planning, Analysis, Growth Forecast
- Specification
- Patient data management : Patient Database
- Various Gallery format & User-defined gallery
- Image Fusion & Comparing : Pre / Post Treatment
- Cephalogram & Tracing on digitized X-ray film
- Treatment Simulation (VTO / STO)
- Chart & Analysis / Export into excel file
- Superimposition
- Photo-Film Image Fusion
- Growth Forecast
- PACS Integration (Option)
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for HDX WILL. The logo consists of a green and blue square with the letter 'W' inside, followed by the text 'HDX WILL' in blue. The text is bold and slightly slanted to the right.
■ System Requirements
- Client or Standalone
- OS : Microsoft Windows 7 or higher
- CPU : Intel Core2Duo or higher
- Memory : DDR2 2GB or higher
- Storage Installation : 100MB free-space
Data storage: depends on size of a data
- Monitor : 1024 x 768 or higher
- Input device : General keyboard & wheel mouse
- Hardware Lock key
- OS : Microsoft Windows 7 or higher
Program configuration
- Connection method : USB 2.0
Image /page/5/Picture/14 description: The image shows a window with a program's menu. The menu has options such as 'Print' and 'Close'. There is also an 'Exit' button at the bottom of the menu. The number '1' is circled in red twice in the image.
Image /page/5/Picture/15 description: The image shows a software interface, likely for dental or orthodontic analysis, with various panels and data displays. The top section includes menu options such as "Analysis," "Measure," and "Excel," along with methods like "Steiner" and "McNamara Kim." The central area displays a grid of images, including X-rays, photos, and traces, labeled as "Lateral," "Frontal," and "SMV." The left side features patient and study lists, with details like chart number, name, and study date.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for HDX Will. The logo consists of a green and blue square with a white "W" inside, followed by the text "HDX Will" in a stylized blue font. The logo is simple and modern, and the colors are bright and eye-catching.
- (1) Menu: To print the currently selected screen, or you can exit the program.
- (2) Top toolbar: Diagnostic items, measurements, and environment setting is available
- (3) Patient editing window:
I) Provides functions such as register new patient, modify, and delete.
-
II) Selected patient's information of the Study, and the image files can be exported as one output file.
III) Copy the image data in connection with the Will-Master. -
(4) Study editing window: Provides functions such as Study registration, modification, deletion and importing/exporting of images.
-
(5) Output window:
Output the images of the selected Study in the Consultation screen, and provides functions such as Trace, VTO/STO, Growth Forecast, Super Impose, Chart, Gallery, etc.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for HDX Will. The logo consists of a green and blue square with a white "W" inside, followed by the text "HDX Will" in a bold, blue font. The logo is simple and modern, and the colors are eye-catching.
6. Intended Use/Indications for Use [21 CFR 807.92(a)(5)]
WillCeph is intended for use by specialized dental practices for capturing, storing and presenting patient images and radiographs and to aid in cephalometric analysis, orthognathic treatment planning and case follow-up. Results produced by the software tools are to be interpreted by trained and licensed dental practitioners.
7. Technological Characteristics [21 CFR 807.92(a)(6)]
There are no significant differences in the technological characteristics of this device compared to the predicate device which adversely affect safety or effectiveness. Provided below is a table summarizing and comparing the technological characteristics of the WillCeph and the predicate device:
| Proposed Device | Predicate Device | |
|---|---|---|
| K Number | Not known | K131348 |
| Model Name | WillCeph | ORTOMED |
| Manufacturer | HDX WILL CORP. | INFOMED SERVICIOS INFORMATICOS |
| Classification Name | System, Image Processing, Radiological | System, Image Processing, Radiological |
| Regulatory Number | 21 CFR 892.2050 | 21 CFR 892.2050 |
| Product Code | LLZ | LLZ |
| Classification | II | II |
| Intended Use | WillCeph is intended for use by specializeddental practices for capturing, storing andpresenting patient images and radiographsand to aid in cephalometric analysis,orthodontic and orthognathic treatmentplanning and case follow-up. Resultsproduced by the software tools are to beinterpreted by trained and licensed dentalpractitioners. | ORTOMED is intended for use byspecialized dental practices for capturing,storing and presenting patient images andradiographs and to aid in cephalometricanalysis, orthodontic and orthognathictreatment planning and case follow-up.Results produced by the software tools areto be interpreted by trained and licenseddental practitioners. |
| Type of Use | Prescription Use | Prescription Use |
| Modality Support | X-ray | X-ray |
| Component | Client or Standalone software | Client or Standalone software |
| Operating System | Windows 7 or higher | Windows 2000/XP, Vista and Windows 7 |
| User Interface | Mouse, Keyboard | Mouse, Keyboard |
| ImageCommunicationStandard | DICOM | DICOM |
| Patient DataManagement | Patient module for patient registration andfile management.Image module for image management. | GESDEN module for patient registrationand file management.GESIMAG module for image management. |
| Patient DatabaseEngine | My SQL / Microsoft® SQL© | Microsoft® SQL© |
| Proposed Device | Predicate Device | |
| ImageMeasurement | Yes (linear, distance, angle) | Yes (linear, distance, angle) |
| Matching /Adjustment ofPhotograph to X-ray | Yes | Yes |
| CephalometricAnalyses | In addition to user-configured analysis,standard orthodontic tracing analysesinclude:- Lateral Analyses- Frontal Analyses- Models Analyses- Photo studies of soft tissues | In addition to user-configured analysis,standard orthodontic tracing analysesinclude:- Lateral Analyses- Frontal Analyses- Models Discrepancy Studies- Photo studies of soft tissue |
| TreatmentPlanning,Simulation andFollow-up | Orthodontic treatment- Translate, tip and rotate incisors,reposition molars, rotate mandible- Growth Forecast- Growth simulation on a traced x-ray ortracing overlaid on photograph- Superimpose one or more growthtracings over original tracingVTO – Visual Treatment ObjectiveSTO – Surgical Treatment Objective(orthognathic surgery)Warping and Morphing | Orthodontic treatment- Translate, tip and rotate incisors,reposition molars, auto-rotate mandible- Arch length discrepancy worksheet- CO/CR Conversion Growth Forecast(Ricketts algorithm)- Growth simulation on a traced x-ray ortracing overlaid on photograph- Superimpose one or more growthtracings over original tracingVTO – Visual Treatment ObjectiveSVTO – Surgical Visual Treatment Objective(orthognathic surgery)Warping and Morphing |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Hoxwill. The logo consists of a square with a green top half and a blue bottom half, with a white "W" in the center. To the right of the square is the text "Hoxwill" in blue, with the letters connected to each other.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for HDX Will. The logo consists of a green square with a white "W" inside, followed by the text "HDX Will" in blue. The text is in a bold, sans-serif font and is slightly slanted to the right.
Non-Clinical Test Summary
WillCeph contains MODERATE level of concern software was designed and developed according to a software development process and was verified and validated.
Software information is provided in accordance with FDA guidance: the content of premarket submissions for software contained in medical devices, issued on May 11, 2005.
Clinical Test Summary:
No clinical studies were considered necessary and performed.
8. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]
WillCeph has the similar technical characteristics with the predicate devices are intended for use by specialized dental practices for capturing and presenting patient images and radiographs and to aid in cephalometric analysis, orthodontic and orthognathic treatment planning and case follow-up. Results produced by the software tools are to be interpreted by trained and licensed dental practitioners.
The subject and primary predicate device share the same intended use, type of use, modality support and image communication standard. They also share equivalent operating systems. Therefore, WillCeph is substantially equivalent to legally marketed predicate device with respect to indications for use and technology characteristics.
9. Conclusion [21 CFR 807.92(b)(3)]
In according with the Federal Food & Drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification HDX WILL CORP, concludes that the WillCeph is substantially equivalent in safety and effectiveness to the predicate devices as described herein.
N/A