(28 days)
The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.
The Varian Eclipse™ Treatment Planning System (Eclipse TPS) provides software tools for planning the treatment of malignant or benign diseases with radiation. Eclipse TPS is a computer-based software device used by trained medical professionals to design and simulate radiation therapy treatments. Eclipse TPS is capable of planning treatments for external beam irradiation with photon, electron, and proton beams, as well as for internal irradiation (brachytherapy) treatments.
This document is a 510(k) summary for the Eclipse Treatment Planning System (Eclipse TPS), specifically version 15.5, from Varian Medical Systems. It describes changes from a previous version (v15.1.1, K170969) and argues for substantial equivalence.
Here's an analysis of the provided information concerning acceptance criteria and supporting studies, particularly regarding the newly included "Smart Segmentation Knowledge Based Contouring (SS KBC)" module, which appears to be the most relevant AI/software feature:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not provide a formal table of explicit acceptance criteria with quantitative performance metrics for the entire Eclipse Treatment Planning System, nor for the specific new features. Instead, it states a general conclusion from non-clinical testing:
- Acceptance Criteria/Goal (Inferred): The product conforms to defined user needs and intended uses, and there are no discrepancy reports (DRs) remaining with a priority of Safety Intolerable or Customer Intolerable. The device is considered safe and performs at least as well as the predicate device.
- Reported Device Performance: "The outcome was that the product conformed to the defined user needs and intended uses and that there were no DRs (discrepancy reports) remaining which had a priority of Safety Intolerable or Customer Intolerable."
For the Smart Segmentation Knowledge Based Contouring (SS KBC) module, the document notes that it was previously cleared under K141248 (v2.2). This implies that the specific performance metrics and acceptance criteria for this module would have been detailed in that earlier submission. In this current submission (K172163), the SS KBC module is being integrated into the broader Eclipse TPS, and the claim is that its previously established safety and effectiveness hold.
2. Sample Size Used for the Test Set and Data Provenance
The document provides very limited detail on specific test sets for the current submission (K172163). It broadly refers to "Verification and Validation" being performed for all new features and "regression testing" for existing features.
- Sample Size: Not explicitly stated for the overall V&V or for specific features.
- Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). The document refers to "test outcomes" but doesn't specify data sources.
For the Smart Segmentation Knowledge Based Contouring (SS KBC) module, since it was previously cleared under K141248, the relevant test set details would be found in that earlier submission. This document only states that the feature "is included as a part of Eclipse Treatment Planning System (v15.5)" and references its prior clearance.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This information is not provided in the current document. The document primarily focuses on software engineering V&V processes (tracing requirements to test outcomes, DRs). For the SS KBC module, this information would have been part of the K141248 submission where its performance as a standalone tool was assessed and ground truth established. Given it's a "Knowledge Based Contouring" system, human expert input for ground truth would be expected.
4. Adjudication Method for the Test Set
This information is not provided in the current document.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No MRMC study is mentioned in this 510(k) summary. The summary focuses on comparing technical features with a predicate device and confirming software functionality, not on human-in-the-loop performance improvement with AI assistance.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- The document implies that standalone performance testing was done as part of the overall "Verification and Validation" process for the software features, as it addresses "product conformance to the defined user needs and intended uses." However, specific details or quantitative results of such standalone performance are not provided in this summary directly.
- For the Smart Segmentation Knowledge Based Contouring (SS KBC) module, as it was previously cleared under K141248, it is highly probable that its standalone algorithmic performance (e.g., accuracy of auto-segmentation compared to ground truth) was evaluated in that prior submission. This document lists it as a feature of the Eclipse TPS v15.5, meaning its existing, cleared functionality is being integrated.
7. Type of Ground Truth Used
- This information is not explicitly stated for the general V&V.
- For the Smart Segmentation Knowledge Based Contouring (SS KBC) module, given its function ("Knowledge Based Contouring"), the ground truth for its original clearance (K141248) would typically involve expert delineation consensus (e.g., radiation oncologists, radiologists) as the gold standard for anatomical structures.
8. Sample Size for the Training Set
- This information is not provided in the current document. Training set details would be most relevant for the knowledge-based/AI component (SS KBC).
9. How the Ground Truth for the Training Set Was Established
- This information is not provided in the current document. For the SS KBC module, again, this would have been part of the K141248 submission. Typically, for knowledge-based contouring systems, ground truth for training data involves expert delineation and potentially review processes.
Summary of AI-Specific Information from the Provided Document:
The most AI-related component mentioned is the "Smart Segmentation Knowledge Based Contouring (SS KBC) as an Eclipse module." Crucially, the document states: "Smart Segmentation - Knowledge Based Contouring (SS KBC) v2.2 was cleared under K141248. The corresponding features above were cleared previously as a part of that submission."
This means that the current submission (K172163) does NOT provide the detailed acceptance criteria or study results for the SS KBC module itself. Instead, it incorporates a previously cleared module and relies on its prior clearance for safety and effectiveness. To find the specific details regarding the SS KBC's performance, a review of the K141248 submission would be necessary. This current document validates that the integration of this module into Eclipse TPS v15.5 maintains its intended functionality and does not introduce new safety or effectiveness concerns.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three human profiles facing to the right, with flowing lines beneath them that resemble waves or fabric.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 15, 2017
Varian Medical Systems, Inc. % Mr. Peter Coronado Director, Regulatory Affairs 911 Hansen Way PALO ALTO CA 94304
Re: K172163
Trade/Device Name: Eclipse Treatment Planning System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: MUJ Dated: July 17, 2017 Received: July 19, 2017
Dear Mr. Coronado:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael D'Hara For
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
K172163
Device Name Eclipse Treatment Planning System
Indications for Use (Describe)
The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.
Type of Use (Select one or both, as applicable):
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
PREMARKET NOTIFICATION
510(k) Summary
Eclipse Treatment Planning System
As required by 21 CFR 807.92
| Submitter's Name: | Varian Medical Systems3100 Hansen Way, m/s E110Palo Alto, CA 94304 |
|---|---|
| Contact Name: Peter J. Coronado-Director Regulatory AffairsPhone: 650/424.6230Fax: 650/646.9200E-mail: submissions.support@varian.com | |
| Proprietary Name: | Eclipse Treatment Planning System |
| Classification Name: | system, planning, radiation therapy treatment21CFR892.5050, MUJ, Class II |
| Common/Usual Name: | Eclipse TPS, Eclipse, Treatment Planning System |
| Predicate Devices: | Eclipse Treatment Planning System 15.1.1 (K170969). |
| Device Description: | The Varian Eclipse™ Treatment Planning System (Eclipse TPS)provides software tools for planning the treatment of malignantor benign diseases with radiation. Eclipse TPS is a computer-basedsoftware device used by trained medical professionals todesign and simulate radiation therapy treatments. Eclipse TPS iscapable of planning treatments for external beam irradiation withphoton, electron, and proton beams, as well as for internalirradiation (brachytherapy) treatments. |
| Indications for Use: | The Eclipse Treatment Planning System (Eclipse TPS) is used toplan radiotherapy treatments for patients with malignant orbenign diseases. Eclipse TPS is used to plan external beamirradiation with photon, electron and proton beams, as well as forinternal irradiation (brachytherapy) treatments. In addition, theEclipse Proton Eye algorithm is specifically indicated forplanning proton treatment of neoplasms of the eye. |
{4}------------------------------------------------
Changes in Technological Characteristics:
The significant changes compared with the predicate are associated with:
-
- Support for HyperArc Planning
- Support for Encompass Fixation Device (Encompass Structures for HyperArc) a.
-
- Trade-Off Exploration based on Multi-Criteria Optimization (MCO)
-
- GPU-calculation capability for select optimization and dose calculations
-
- Updated handling for Numerical Input Values (NIV)
-
- Inclusion of Smart Segmentation Knowledge Based Contouring (SS KBC) as an Eclipse module
- a. Updated Eclipse Image Registration and Segmentation IFU & Reference Guides
The complete list of Eclipse changes and their related requirements can be found in the document "Tracing Changed and New Features to System Requirements" in Section 18 of this submission; the indications for use and the intended use are the same in the subject device as for the predicate.
| PREDICATE DEVICEFEATURE/ SPECIFICATION510(к) K170969ECLIPSE TPS v15.1.1 | MODIFIED DEVICEFEATURE/ SPECIFICATIONECLIPSE TPS v15.5 | |
|---|---|---|
| 1. Intended Use / Indications for Use | ||
| Statement of Intended Use and Indications for Use | Eclipse is intended to be used to plan radiotherapytreatments for patients with malignant or benigndiseases. Eclipse is used to plan external beamirradiation with photon, electron and proton beams,as well as for internal irradiation (brachytherapy)treatments. In addition, the Eclipse Proton Eyealgorithm is specifically indicated for planning protontreatment of neoplasms of the eye. | Eclipse is intended to be used to plan radiotherapytreatments for patients with malignant or benigndiseases. Eclipse is used to plan external beamirradiation with photon, electron and proton beams,as well as for internal irradiation (brachytherapy)treatments. In addition, the Eclipse Proton Eyealgorithm is specifically indicated for planning protontreatment of neoplasms of the eye. |
| 2. General Usage | ||
| External beam PHOTON planning | Yes | Yes |
| External beam PHOTON inverse planning | Yes | Yes |
| External beam ELECTRON planning | Yes | Yes |
| External beam PROTON planning | Yes | Yes |
| External beam OCULAR PROTON planning (EOPP) | No *) | No *) |
| Internal BRACHYTHERAPY planning | Yes | Yes |
| Stereotactic Frame Localization | Yes | Yes |
| Treatment Planning for Varian RDS (Halcyon) | Yes | Yes |
| Automated Structure Delineation based on ExpertCase Library**) | No, feature was part ofSmart SegmentationKnowledge BasedContouring (v3.6) | Yes, feature is includedas a part of EclipseTreatment PlanningSystem (v15.5) |
| 3. Supported External Beams & Accessories | ||
| Photon beams | Yes | Yes |
| Electron beams | Yes | Yes |
| Proton beams | Yes | Yes |
| Coplanar fields | Yes | Yes |
| Non-coplanar fields | Yes | Yes |
| Multi-leaf Collimators | Yes | Yes |
| Asymmetric collimators | Yes | Yes |
| Stereotactic Cone collimators | Yes | Yes |
| Arc fields | Yes | Yes |
| Poured Blocks | Yes | Yes |
| Compensators | Yes | Yes |
| Physical wedges | Yes | Yes |
| Dynamic wedges | Yes | Yes |
| Flattening filter free support (FFF) | Yes | Yes |
| Rotating treatment couch | Yes | Yes |
| Non-rotating RDS couch | Yes | Yes |
| Elekta 160 MLC | Yes | Yes |
| Varian SX1 MLC | Yes | Yes |
| 4. Supported Brachytherapy Sources &Accessories | ||
| Plan for high dose rate afterloader | Yes | Yes |
| Manual low dose rate brachytherapy: seeds, linesources, wire | Yes | Yes |
| Applicator library | Yes | Yes |
| Needle templates | Yes | Yes |
| Seed templates | Yes | Yes |
| 5. Graphical User Interface | ||
| Multiple-instance application | Yes | Yes |
| Multiple-workspace layout | Yes | Yes |
| Graphical display/editing of field parameters | Yes | Yes |
| Beam's-Eye-View display | Yes | Yes |
| 3D patient image display | Yes | Yes |
| Model for human Eye | No *) | No *) |
| SRS Localization application | Yes | Yes |
| SRS Planning application | Yes | Yes |
| Biological Optimization application | Yes | Yes |
| Biological Evaluation application | Yes | Yes |
| 3D Conformal Optimization application | Yes | Yes |
| 6. Image Processing | ||
| Orthogonal image displays (3) | Yes | Yes |
| Oblique image display | Yes | Yes |
| Edge enhancement filters | Yes | Yes |
| Image smoothing filters | Yes | Yes |
| CT/MR/PET Image Registration | Yes | Yes |
| Image blending utility | Yes | Yes |
| 4D image display (registration of time series of 3Dimages) | Yes | Yes |
| Digitally reconstructed radiographs | Yes | Yes |
| Enclosed Volume measurement | Yes | Yes |
| Stereotactic Frame Coordinate transformation | Yes | Yes |
| 7. Image Segmentation | ||
| Geometrical shapes | Yes | Yes |
| Manual editing and manipulation tools | Yes | Yes |
| Automatic /semi-automatic tools | Yes | Yes |
| Automatic/semi-automatic on-demand and post-processing tools for individual organs/structures | Yes | Yes |
| Automatic on-demand and pre-processing tools formultiple organs/structures | Yes | Yes |
| 3D Automargin | Yes | Yes |
| Logical operators | Yes | Yes |
| Enhanced 2D and 3D contouring tools | Yes | Yes |
| Enhanced 4D functionality, including structurepropagation and display of respiratory amplitudedistribution | Yes | Yes |
| Expert Case Library**) | No, feature was part ofSmart SegmentationKnowledge BasedContouring (v3.6) | Yes, feature is includedas a part of EclipseTreatment PlanningSystem (v15.5) |
| Multi-atlas Segmentation | ||
| Expert cases: Nasopharynx, Tonsil, Base ofTongue, Hypopharynx, Larynx**) | ||
| Structure templates and dictionary with standardizelabels**) | ||
| Calypso segmentation **) | ||
| Segmentation of mandible **) | ||
| Semi-Automatic Segmentation of Lung Tumors**)(Lung Tumor Segmentation Tool) | ||
| Encompass Couch Model (Fixation Device) | No | Yes |
| 8. Dose Calculation | ||
| Distributed Calculation Framework | Yes | Yes |
| Photon calculation | Yes | Yes |
| - Energy Range | 1 MV – 50 MV | 1 MV – 50 MV |
| - CT-based volumetric calculation | Yes | Yes |
| - Non-CT based IRREG calculation | Yes | Yes |
| - Convolution method | No | No |
| - Combined electron/photon scatter | Yes | Yes |
| - Directional heterogeneity correction | Yes | Yes |
| - Treatment head modelling | Yes | Yes |
| - Photon Monitor Unit calculation | Yes | Yes |
| - Compensator monitor unit calculation | Yes | Yes |
| - Beam Angle Optimization (GEOS) | Yes | Yes |
| - Leaf Motion Sequencing | Yes | Yes |
| - Dose Dynamic Arc planning | Yes | Yes |
| - Cone Dose Calculation | Yes | Yes |
| - Biological optimization | Yes | Yes |
| - 3D Conformal Optimization | Yes | Yes |
| - IMRT optimization | Yes | Yes (also GPU) |
| - AcurosXB dose calculation algorithm | Yes | Yes (also GPU) |
| - AAA | Yes | Yes |
| - Range Uncertainty feature for photons | Yes | Yes |
| - Siemens mArc | Yes | Yes |
| - Siemens FFF (unflattened beam) | Yes | Yes |
| - Elekta Agility MLC 160 | Yes | Yes |
| - RaySearch PlanConverter | Yes | Yes |
| - Varian SX1 MLC | Yes | Yes |
| - FTDC Imaging dose calculation for RDS machine | Yes | Yes (also GPU) |
| Electron calculation | Yes | Yes |
| - Energy Range | 1 MeV – 50 MeV | 1 MeV – 50 MeV |
| - Gaussian Pencil Beam Model | Yes | No, obsoleted |
| - Electron Monte Carlo algorithm | No | No |
| - Electron Monte Carlo algorithm v2 | Yes(incl. Siemens, Elekta) | Yes(incl. Siemens, Elekta) |
| - Electron Monitor Unit calculation | Yes | Yes |
| - Portal Dose Image Prediction | Yes | Yes |
| Proton calculation | Yes | Yes |
| - Energy Range | 50 MeV - 300 MeV | 50 MeV - 300 MeV |
| - Single scattering technique | Yes | Yes |
| - Double scattering technique | Yes | Yes |
| - Uniform scanning technique | Yes | Yes |
| - Modulated scanning technique | Yes | Yes |
| - Monitor unit calculation for modulated scanning | Yes | Yes |
| - Range uncertainty feature for Protons | Yes | Yes |
| - Robust Proton Optimization | Yes | Yes |
| - AcurosPT Dose Calculation Algorithm | Yes | Yes |
| Brachytherapy calculation | Yes | Yes |
| - AAPM TG 43 compliant | Yes | Yes |
| - Point Dose calculation | Yes | Yes |
| - Optimization to point dose constraints | Yes | Yes |
| - Geometric optimization | Yes | Yes |
| - AcurosBV dose calculation algorithm | Yes(w/ support for report thedose to medium) | Yes(w/ support for report thedose to medium) |
| - Intermediate dose capability for AcurosBV algorithm | Yes | Yes |
| Eclipse Algorithm Application ProgrammingInterface (EAAPI) | Yes | Yes |
| RapidPlan - Dose Volume Histogram (DVH)Estimation | Yes(extended Dose VolumeHistogram (DVH)Estimation) | Yes(extended Dose VolumeHistogram (DVH)Estimation) |
| 9. Dose evaluation | ||
| Dose color wash | 2D, 3D | 2D, 3D |
| lsodose levels | 2D, 3D | 2D, 3D |
| Isodose Surface | 3D | 3D |
| Reference point dose summary | Yes | Yes |
| Dose Volume Histogram plot | Yes | Yes |
| Plan summing tool | Yes | Yes |
| Plan comparison tools | Yes | Yes |
| Evaluation using biological models | Yes | Yes |
| Evaluation using MCO (Multi-Criteria Optimization) | No | Yes |
| 10. Plan Output - Hardcopy | ||
| Graphics window screen dump | Yes | Yes |
| Patient administration data | Yes | Yes |
| Plan parameters | Yes | Yes |
| Geometrical displays of plan data | Yes | Yes |
| Dose distribution | Yes | Yes |
| DVH plot | Yes | Yes |
| BEV display | Yes | Yes |
| Patient orientation | Yes | Yes |
| Patient orientation | Yes | Yes |
| 11. Import/Export Interfaces | ||
| ARIA RadOnc integration | Yes | Yes |
| DICOM RT | Yes | Yes |
| Other image formats | Yes | Yes |
| Electromagnetic Digitizer | Import | Import |
| Film Scanner | No | No |
| Eclipse Scripting API (ESAPI) read only access | Yes(includes BrachyVisionand Proton) | Yes(includes BrachyVisionand Proton) |
| Eclipse Scripting API (ESAPI) write access | Yes(in research database,only, with additionalproton planning support) | Yes(in research database,only, with additionalproton planning support) |
| Eclipse Automation | Yes | Yes |
| Export field coordinates to Laser System | Export | Export |
| Basic RT Prescription information available | Yes | Yes |
Device Comparison Table
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
*) A compatible feature has not been released and using the feature is blocked either by not including it in the distribution media for the indicated version(s) or by licensing.
**) Smart Segmentation - Knowledge Based Contouring (SS KBC) v2.2 was cleared under K141248. The corresponding features above were cleared previously as a part of that submission.
Non-clinical Testing
Verification and Validation were performed for all the new features and regression testing was performed against the existing features of Eclipse. System requirements created or affected by the changes can be traced to the test outcomes.
Conclusion of Non-Clinical testing
The outcome was that the product conformed to the defined user needs and intended uses and that there were no DRs (discrepancy reports) remaining which had a priority of Safety Intolerable or Customer Intolerable. Varian therefore considers Eclipse 15.5 to be safe and to perform at least as well as the predicate device.
Argument for Substantial Equivalence to the Predicate Device
A subset of features of the current device is different to the predicate. These changes are all considered by Varian to be enhancements of the predicate. The Indications for Use and the Intended Use remain unchanged. There are no changes in the principle of operation of the software. The Verification and Validation demonstrates that the device is as safe and effective as the predicate. Varian therefore believes that Eclipse Treatment Planning System is substantially equivalent to the predicate.
N/A