K Number
K172159

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-01-30

(196 days)

Product Code
Regulation Number
880.6880
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Getinge GSS67N Series Steam Sterilizer is intended for use by health care facilities to sterilize wrapped and unwrapped, porous and nonporous heat and moisture stable items such as surgical instruments and linens by means of pressurized steam. The GSS67N Series Steam Sterilizer is available in 3 models differentiated by chamber length; GSS67N Model 6710 (39 inch chamber), GSS67N Model 6713 (51 inch chamber) and GSS67N Model 6717 (67 inch chamber).

Device Description

The Getinge GSS67N Series Steam Sterilizer is designed for sterilization of heat and moisture stable materials used in healthcare facilities. There are three model designations to identify three different chamber lengths. The model 6710 is 1000 mm (39") long, model 6713 is 1300 mm (51") long and model 6717 is 1700 mm (67") lona.

The Getinge GSS67N Series Steam Sterilizer employs both gravity/downward displacement with positive pulse conditioning and pressure/vacuum pulsing for dynamic air removal. Up to 11 standard cycles can be easily accessed and custom cycle names can be designated by the user. All cycle phases are sequenced and monitored by the control system, providing both audible and visual notification of deviation from certain operating parameters.

AI/ML Overview

This document pertains to the Maquet GmbH Getinge GSS67N Series Steam Sterilizer (K172159).

1. Table of Acceptance Criteria and Reported Device Performance:

The acceptance criteria for the GSS67N Series Steam Sterilizer are derived from the ANSI/AAMI ST8:2013 Hospital Steam Sterilizers standard. The reported device performance indicates that the sterilizer meets these requirements.

Performance MetricAcceptance Criteria (ANSI/AAMI ST8:2013)Reported Device Performance (Summary of Performance Testing Section)
Sterility Assurance Level (SAL)At least 10^-6 (complete BI kill and sufficient Fo value)Achieved SAL of at least 10^-6 through complete BI kill and Fo value of at least 12 for all PREVAC, GRAVITY, and IUSS cycles.
Moisture Retention (Fabric Packs)Less than 3% increase in pre-sterilization test pack weight, no visible wet spots.Demonstrated less than 3% increase in pre-sterilization test pack weight, including no visible wet spots for PREVAC (excluding PREVAC 2 and PREVAC 5) and GRAVITY cycles.
Moisture Retention (Wrapped Instrument Trays)Less than 20% increase in pre-sterilization weight of towel, no visible wet spots on outer wrapper.Demonstrated less than 20% increase in pre-sterilization weight of the towel, including no visible wet spots on the outer wrapper for PREVAC and GRAVITY cycles using wrapped instruments.
Empty Chamber Thermal ConditionsCapable of providing steady-state thermal conditions consistent with predicated SAL.Demonstrated capability to provide steady-state thermal conditions within the chamber consistent with the predicated SAL.
Bowie-Dick TestSatisfactory performance as described in ANSI/AAMI ST8:2013 section 5.6.1.1.Bowie-Dick cycle validated using the Bowie-Dick test pack as described in ANSI/AAMI ST8:2013 section 5.6.1.1. (Implicitly passed, as performance testing summary states validation).
Liquid Load Cycles (Water Loss)Water loss not exceeding 50 mL for 1-liter flasks.Demonstrated water loss not exceeding 50 mL for 1-liter flasks.
Software ValidationPer FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."Software validation for cycle operation performed according to the specified FDA guidance document.
Electrical Safety & EMCCompliance with IEC 61010-1, IEC 6101-2-040, and IEC 61326:2013.Electrical safety and EMC testing completed per IEC 61010-1, IEC 6101-2-040, and IEC 61326:2013.

2. Sample Size Used for the Test Set and Data Provenance:

The document does not explicitly state a specific "sample size" in terms of number of sterilizers or distinct test runs for the validation of each cycle. However, the performance testing was conducted according to ANSI/AAMI ST8:2013 Hospital Steam Sterilizers, which outlines rigorous testing protocols. For each cycle type (PREVAC, GRAVITY, IUSS, Liquids, Bowie-Dick), validation was performed using specific process challenge devices and configurations as detailed in the standard.

  • Data Provenance: The testing was carried out by Maquet GmbH, the manufacturer, to demonstrate compliance with the ANSI/AAMI ST8:2013 standard. The data is prospective as it involves validation testing specifically for this device series, rather than retrospective analysis of existing data. The country of origin of the data is implicitly Germany (manufacturer location).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This document does not provide information on the "number of experts" or their specific "qualifications" for establishing ground truth for the test set. For steam sterilizers, the "ground truth" is typically established by physical and biological indicators showing successful sterilization, rather than expert judgment in interpreting images or clinical outcomes. The adherence to the ANSI/AAMI ST8:2013 standard itself implies that the methodologies for determining effective sterilization are standardized and technically validated.

4. Adjudication Method for the Test Set:

Not applicable in the context of sterilizer performance validation. The outcomes of sterilization cycles are determined by objective measurements (temperature, time, pressure) and biological indicator analysis. There is no subjective interpretation or "adjudication" in the sense of expert review of discrepancies.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done:

No. This is not a diagnostic device that involves human readers interpreting data.

6. If a Standalone (i.e., algorithm only without human-in-the loop performance) Was Done:

Yes, in essence. The device (steam sterilizer) operates autonomously based on its programmed cycles. The performance validation assessed the sterilizer's ability to achieve sterilization without human intervention in the sterilization process itself. Human involvement is in operating the device and loading/unloading items, but the sterilization process is automated ("algorithm only" in the context of its defined cycles).

7. The Type of Ground Truth Used:

The ground truth used for performance validation is based on:

  • Biological Indicators (BI kill): Complete kill of biological indicators is a primary ground truth for sterility.
  • Physical/Chemical Indicators: Achievement of sufficient Fo value (a measure of sterilization lethality related to time and temperature).
  • Physical Measurements: Temperature, pressure, and time parameters are monitored.
  • Moisture Retention: Quantitative measurement of weight increase and qualitative assessment of wet spots.
  • Bowie-Dick Test Pack: Changes indicating proper air removal and steam penetration.
  • Water Loss: Quantitative measurement for liquid cycles.

These are objective, outcome-based measurements directly demonstrating the physical and biological effectiveness of the sterilization process according to established industry standards.

8. The Sample Size for the Training Set:

Not applicable. This device is a hardware sterilizer, not an AI/ML algorithm that requires a "training set" in the conventional sense. The "training" for the device would be its engineering design and validation processes against established standards.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no training set in the context of AI/ML. The device's "ground truth" for its operation is established by adherence to the physical and biological principles of steam sterilization, codified in standards like ANSI/AAMI ST8:2013, and confirmed through extensive engineering design and testing.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 30, 2018

Maquet GmbH Holger Ullrich Director Product Compliance Kehler Strasse 31 Rastatt, 76437 De

Re: K172159

Trade/Device Name: Getinge GSS67N Series Steam Sterilizer Regulation Number: 21 CFR 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: Class II Product Code: FLE Dated: December 27, 2017 Received: December 28, 2017

Dear Holger Ullrich:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172159

Device Name Getinge GSS67N Series Steam Sterilizer

Indications for Use (Describe)

The Getinge GSS67N Series Steam Sterilizer is intended for use by health care facilities to sterilize wrapped and unwrapped, porous and nonporous heat and moisture stable items such as surgical instruments and linens by means of pressurized steam. The GSS67N Series Steam Sterilizer is available in 3 models differentiated by chamber length; GSS67N Model 6710 (39 inch chamber), GSS67N Model 6713 (51 inch chamber) and GSS67N Model 6717 (67 inch chamber).

List of available cycles: See table below

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Getinge GSS67N Series Steam Sterilizer Cycles and Load Chart

Cycle TypeFactory SettingsLoad Configuration(Note 1)Maximum Items perChamber Length
Exp.Temp.Exp.TimeDryingTime67101000mm(39 in.)67131300mm(51 in.)67171700mm(67 in.)
PREVAC 1135.0°C(275.0°F)3 min16 minDouble-wrapped instrumenttrays, up to 11.3 kg (25.0 lb)(per tray)81216
PREVAC 2135.0°C(275.0°F)3 min3 min(Note 4)Fabric Packs182430
PREVAC 4132.2°C(270.0°F)4 min30 minDouble-wrapped instrumenttrays, up to 11.3 kg (25.0 lb)(per tray)81216
PREVAC 5132.2°C(270.0°F)4 min5 min(Note 4)Fabric packs182430
B & D TEST134°C(273.0°F)3 min,30 sec0 min1 B&D Test Pack in anEMPTY chamber (other thanloading accessories)1 TestPack1 TestPack1 TestPack
GRAVITY 1121.1°C(250.0°F)30 min45 minDouble-wrapped instrumenttrays, up to 11.3 kg (25.0 lb)(per tray)81216
GRAVITY 1Fabric packs182430
GRAVITY 2135.0°C(275.0°F)10 min45 minDouble-wrapped instrumenttrays, up to 11.3 kg (25.0 lb)(per tray)81216
GRAVITY 2Fabric packs182430
GRAVITY 3132.2°C(270.0°F)15 min45 minDouble-wrapped instrumenttrays, up to 11.3 kg (25.0 lb)(per tray)81216
GRAVITY 3Fabric packs182430
IUSS 1Vac(Note 7)135.0°C(275.0°F)3 min1 min(Note 4)Instrument Tray (25lb)111
(Note 7)Single Instrument Tray111
IUSS 2Grav(Note 7)135.0°C(275.0°F)10 min0.5 min(Note 4)Instrument Tray (25lb)111
Single Instrument/Tray111
IUSS 3Grav(Note 7)132.2°C(270.0°F)4 min1 min(Note 4)Instrument Tray (25lb)111
Single Instrument/Tray111
LIQUIDS121.1°C(250.0°F)45 min3 kPa/min(0.44psi/min)(Note 3)Each container1000 mL (34 fl oz) or smaller(Notes 5, 6)333
LEAK TEST(Note 2)131.1°C(268.0°F)N/AN/AEmpty Chamber(other than loadingaccessories)

{4}------------------------------------------------

NOTE: Liquid Cycles are not intended for the sterilization of liquids used for direct patient contact.

TABLE NOTES:

    1. The load configurations listed in these tables are those used during testing validations of the sterilizer. These configurations follow AAM Standard ST8: Hospital steam sterilizers where applicable (fabric packs are process challenge devices as described in ANSI/AAMI ST8 and were made to be consistent with the packs described in ANSI/AAMI ST8). For guidance on processing loads in the sterilizer, refer to AAM Standard ST79: Comprehensive guide to steam sterility assurance in health care facilities.
    1. Vacuum leak test parameters are not adjustable.
    1. Cooldown rate
    1. Items may NOT be dry at the end of the following cycles:
    • · IUSS 1, 2, 3
    • PREVAC 2
    • PREVAC 5

Drying time may be added if required.

    1. Facility must validate the cycle if the load includes containers larger than 1000 mL (34 fl oz).
    1. Use vented or open containers only.
    1. The recommended minimum exposure time and temperature for unwrapped, nonporous loads (e.g., metal instruments) that are sterilized for immediate use is 3 minutes at 132°C (270°F) or 135°C (275°F).

{5}------------------------------------------------

K172159 510(k) SUMMARY

Getinge GSS67N Series Steam Sterilizer

Submitted by:Maquet GmbHKehler Strasse 31Rastatt DE-BWGermany 76437
Contact Person:Barb Smith, RACSr. Regulatory Affairs SpecialistPhone: (585) 214-6049Email: barb.smith@getinge.com
Date prepared:Jan 30, 2018
Proprietary Name:GSS67N Series Steam Sterilizer
Common Name:Steam Sterilizer
Device Classification:Steam Sterilizer (80 FLE)Class II, as listed per 21 CFR 880.6880
Predicate Device:Getinge 633HC Series Steam Sterilizer [K070657]

Description of Device:

The Getinge GSS67N Series Steam Sterilizer is designed for sterilization of heat and moisture stable materials used in healthcare facilities. There are three model designations to identify three different chamber lengths. The model 6710 is 1000 mm (39") long, model 6713 is 1300 mm (51") long and model 6717 is 1700 mm (67") lona.

The Getinge GSS67N Series Steam Sterilizer employs both gravity/downward displacement with positive pulse conditioning and pressure/vacuum pulsing for dynamic air removal. Up to 11 standard cycles can be easily accessed and custom cycle names can be designated by the user. All cycle phases are sequenced and monitored by the control system, providing both audible and visual notification of deviation from certain operating parameters.

Intended Use:

The Getinge GSS67N Series Steam Sterilizer is intended for use by health care facilities to sterilize wrapped and unwrapped, porous and nonporous heat and moisture stable items such as surgical instruments and linens by means of pressurized steam. The GSS67N Series Steam Sterilizer is available in 3 models differentiated by chamber length; GSS67N Model 6710 (39 inch chamber), GSS67N Model 6713 (51 inch chamber) and GSS67N Model 6717 (67 inch chamber).

{6}------------------------------------------------

List of available cycles:

Cycle TypeFactory SettingsLoad Configuration(Note 1)Maximum Items perChamber Length
Exp.TempExp.TimeDryingTime67101000mm(39 in.)67131300mm(51 in.)67171700mm(67 in.)
PREVAC 1135.0°C(275.0°F)3 min16 minDouble-wrapped instrumenttrays, up to 11.3 kg (25.0 lb)(pertray)81216
PREVAC 2135.0°C(275.0°F)3 min3 min(Note 4)Fabric Packs182430
PREVAC 4132.2°C(270.0°F)4 min30 minDouble-wrapped instrumenttrays, up to 11.3 kg (25.0 lb)(pertray)81216
PREVAC 5132.2°C(270.0°F)4 min5 min(Note 4)Fabric packs182430
B & D TEST134°C(273.0°F)3 min,30 sec0 min1 B&D Test Pack in anEMPTY chamber (other thanloading accessories)1 TestPack1 TestPack1 TestPack
GRAVITY 1121.1°C(250.0°F)30 min45 minDouble-wrapped instrumenttrays, up to 11.3 kg (25.0 lb)(pertray)81216
Fabric packs182430
GRAVITY 2135.0°C(275.0°F)10 min45 minDouble-wrapped instrumenttrays, up to 11.3 kg (25.0 lb)(pertray)81216
Fabric packs182430
GRAVITY 3132.2°C(270.0°F)15 min45 minDouble-wrapped instrumenttrays, up to 11.3 kg (25.0 lb)(pertray)81216
Fabric packs182430
IUSS 1Vac(Note 7)135.0°C(275.0°F)3 min1 min(Note 4)Instrument Tray (25lb)111
Single Instrument Tray111
IUSS 2Grav(Note 7)135.0°C(275.0°F)10 min0.5 min(Note 4)Instrument Tray (25lb)111
Single Instrument/Tray111
IUSS 3Grav(Note 7)132.2°C(270.0°F)4 min1 min(Note 4)Instrument Tray (25lb)111
Single Instrument/Tray111
LIQUIDS121.1°C(250.0°F)45 min3 kPa/min(0.44psi/min)(Note 3)Each container1000 mL (34 fl oz) or smaller(Notes 5, 6)333
LEAK TEST(Note 2)131.1°C(268.0°F)N/AN/AEmpty Chamber(other than loadingaccessories)

Getinge GSS67N Series Steam Sterilizer Cycles and Load Chart

{7}------------------------------------------------

Maquet GmbH Getinge Group FDA 510(k) Summary Device: GSS67N Series Steam Sterilizer

NOTE: Liquid Cycles are not intended for the sterilization of liquids used for direct patient contact.

TABLE NOTES:

    1. The load configurations listed in these tables are those used during testing validations of the sterilizer. These configurations follow AAMI Standard ST8: Hospital steam sterilizers where applicable (fabric packs are process challenge devices as described in ANSI/AAMI ST8 and were made to be consistent with the packs described in ANSI/AAMI ST8). For guidance on processing loads in the sterilizer, refer to AAMI Standard ST79: Comprehensive guide to steam sterility assurance in health care facilities.
    1. Vacuum leak test parameters are not adjustable.
    1. Cooldown rate
    1. Items may NOT be dry at the end of the following cycles:
    • · IUSS 1, 2, 3
    • PREVAC 2
    • PREVAC 5

Drying time may be added if required.

  1. Facility must validate the cycle if the load includes containers larger than 1000 mL (34 fl oz).
    1. Use vented or open containers only.
  1. The recommended minimum exposure time and temperature for unwrapped, nonporous loads (e.g., metal instruments) that are sterilized for immediate use is 3 minutes at 132°C (270°F) or 135°C (275°F).

{8}------------------------------------------------

Comparisons to Predicate Device:

Similarities between the Getinge GSS67N Series Steam Sterilizer and the identified predicate device (Getinge 633HC Series Steam Sterilizer) are:

  • · Intended use is the same: Intended for use by health care facilities to sterilize wrapped and unwrapped, porous and nonporous heat and moisture stable items such as surgical instruments and linens by means of pressurized steam.
  • · Operating Principle is the same: Saturated steam is the sterilizing agent.
  • · Materials of construction are the same: Vessel material is Stainless Steel SA240. There is no direct patient contact associated with this device.
  • · Cycle Types: The cycle types offered are the same; Prevacuum, Gravity, Immediate Use (Flash) and Liquids (not for sterilization of liquids used directly for patient contact).
  • · Performance Testing: Factory recommended cycles were tested per industry standards and guidelines and effectiveness of sterilizer function was demonstrated by complete kill of biological indicators. Getinge Sterilizers have been validated to meet the requirements of ANSI/AAMI ST8 Hospital Steam Sterilizers.

The differences between the Getinge GSS67N Series Steam Sterilizer and the predicate device (Getinge 633HC Series Steam Sterilizer) are:

  • · The Getinge GSS67N series provides standard cycles validated to the current AAMI ST8:2013 Hospital Steam Sterilizers standard and thus provides testing of instrument trays up to 25 lbs. In addition the GSS67N has included 132°C cycles that were not validated on the predicate device.
  • · The Getinge GSS67N Series Steam Sterilizer offers one larger vessel size than the 633HC Series.
  • · The Getinge GSS67N Series Steam Sterilizer removes air from the chamber by means of a vacuum pump as opposed to a water injector (used in 633HC predicate) thus allowing for less water usage.
  • · The Getinge GSS67N Series Steam Sterilizer has an updated touchscreen display.

{9}------------------------------------------------

Summary of Performance Testing:

Factory recommended cycles were tested per industry standards and quidelines and effectiveness of sterilizer function was demonstrated by complete kill of biological indicators. Getinge GSS67N Series Steam Sterilizers have been validated to meet the requirements of ANSI/AAMI ST8:2013 Hospital Steam Sterilizers.

The results of Getinge GSS67N Series Steam Sterilizer validation testing demonstrate that the sterilizer performs as intended. Summary of testing:

  • Empty chamber testing performed for all cvcles as described in ANSI/AAMI . ST8:2013 Hospital Steam Sterilizers section 5.4.2.5. The results demonstrated that the sterilizer is capable of providing steady-state thermal conditions within the chamber that are consistent with the predicated sterility assurance level (SAL) in the load.
  • . All PREVAC and GRAVITY cycles were validated using fabric process challenge packs as described in ANSI/AAMI ST8:2013 section 5.5.2. The results from this testing demonstrated a sterility assurance level of at least 10° through achievement of time at temperature sufficient to produce an Fo value of at least 12. complete Bl kill and moisture retention of less than 3% increase in pre-sterilization test pack weight including no visible wet spots.
  • All PREVAC (excluding PREVAC 2 and PREVAC 5 that have shortened drying . times) and GRAVITY cycles were validated using wrapped instrument process challenge devices as described in ANSI/AAMI ST8:2013 section 5.5.4. The results from this testing demonstrated a sterility assurance level of at least 10° through achievement of time at temperature sufficient to produce an Fo value of at least 12, complete Bl kill and moisture retention of less than 20% increase in pre-sterilization weight of the towel including no visible wet spots on the outer wrapper.
  • . All Immediate Use (IUSS) cycles were validated using an unwrapped nonporous process challenge device as described in ANSI/AAMI ST8:2013 section 5.5.5. The results from this testing demonstrated a sterility assurance level of at least 10° through achievement of time at temperature sufficient to produce an Fo value of at least 12 and complete BI kill.
  • Liquid loads cycles were validated using 3 one liter flasks as described in . ANSI/AAMI ST8:2013 section 5.5.3. The results from this testing demonstrated a sterility assurance level of at least 10° through achievement of time at temperature sufficient to produce an Fo value of at least 12, complete Bl kill and water loss not exceeding 50ml.
  • Bowie Dick cycle was validated using the Bowie-Dick test pack as described in . ANSI/AAMI ST8:2013 section 5.6.1.1.

{10}------------------------------------------------

  • . The software validation for the cycle operation was performed according to FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
  • . The electrical safety and EMC testing was completed per IEC 61010-1 "Safety requirements for electrical equipment for measurement, control, and laboratory use, Part 1: General requirements" (IEC 61010-1:2010 (Third Edition) +A1:2016; IEC 6101-2-040 "Safety requirements for electrical equipment for measurement, control, and laboratory use Part 2-040, Particular requirements for sterilizers and washer-disinfectors used to treat medical materials; and IEC 61326:2013 "Electrical Equipment for Meassurement, Control and Laboratory Use - EMC requirements", Part 1: General Requirements".

Clinical Data:

No clinical data is required for this device classification submission.

Conclusion:

The GSS67N Series Steam Sterilizer meets the applicable requirements of AAMI ST8:2013 performance standards. The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device K070657.

§ 880.6880 Steam sterilizer.

(a)
Identification. A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.(b)
Classification. Class II (performance standards).