K Number
K172137

Validate with FDA (Live)

Device Name
PIEZOTOME CUBE
Date Cleared
2018-05-03

(290 days)

Product Code
Regulation Number
872.4120
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PIEZOTOME CUBE is an ultrasonic surgical system that supply utilities to and serve as a base for dental tips. The PIEZOTOME CUBE consist of a control unit and handpiece intended for use in intraoral surgery procedures, including osteotomy, osteoplasty, periodontics and implantology.

Device Description

SATELEC CUBE is an piezoelectric device that uses micro-vibrations of associated Tips, to perform the dental procedures defined in Indication for Use.

PIEZOTOME CUBE consists of a Console, a Multifunction footswitch and a reusable Handpiece.

The Console is the heart of the SATELEC PIEZOTOME CUBE. The Console contains the display board and the motherboard. Ultrasonic Handpiece and Footswitch are connected to the Console. The Touch screen present on the front panel is used to define the settings of the SATELEC PIEZOTOME CUBE (Ultrasonic modes, irrigation flow values).

The Ultrasonic Handpiece is held in the Practitioner's hand. The Ultrasonic Handpiece is connected to the Console via a Handpiece Cord. The Handpiece is dedicated to Dental Bone Surgery procedures. The Ultrasonic Handpiece is equipped with a Piezoelectric Transducer. The Piezoelectric transducer converts the Electrical Signal delivered by the Console into mechanical micro-vibrations. The Ultrasonic Handpiece is reusable and Sterilizable by autoclaving.

Tips are fixed by means of a Wrench at the extremity of the Ultrasonic Handpiece. Tip are in direct contact to the patient. The ultrasonic mechanical vibrations are transmitted to the Tip.

The Pump housing is designed to accommodate SATELEC Irrigation Tubing cassettes. Irrigation Tubings are Single Use and delivered under Sterile State. Irrigation is intended to cool the clinical site and rinse the fragments such as bone or teeth.

AI/ML Overview

The provided text is a 510(k) Summary for the PIEZOTOME CUBE, an ultrasonic surgical system. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a standalone study with defined acceptance criteria and performance metrics for clinical efficacy. Therefore, much of the requested information regarding an acceptance criteria table, sample sizes for test and training sets, number and qualifications of experts, adjudication methods, multi-reader multi-case studies, and specific ground truth types for a clinical performance study cannot be found or directly inferred from the provided text.

The document primarily relies on comparisons of technical characteristics and performance testing to the predicate device, PIEZOTOME SOLO.

Here's an attempt to answer the questions based on the available information:

1. A table of acceptance criteria and the reported device performance

The document does not provide a table of explicit acceptance criteria with corresponding performance results for clinical efficacy. Instead, it compares the technical characteristics and performance of the PIEZOTOME CUBE to its predicate device, PIEZOTOME SOLO, arguing that the differences have no impact on safety or effectiveness.

Below is a summary of the technical performance comparisons, which serve as the "reported device performance" in the context of demonstrating substantial equivalence, effectively meeting the "acceptance criteria" of being similar to the predicate.

Feature / CharacteristicPIEZOTOME CUBE (New Device)PIEZOTOME SOLO (Predicate Device #1 - K112188)Impact of the differences on Safety / Effectiveness
Indications for UseConsistent with predicateThe intended use of the SATELEC PIEZOTOME SOLO... is to supply utilities to and serve as a base for dental tools and accessories for use by qualified dental practitioners.Similar to PD#1 - No Impact
Intended for use in intraoral surgery procedures, including osteotomy, osteoplasty, periodontics and implantology.The PIEZOTOME SOLO is a device designed to perform dental surgery including osteotomies, osteoplasties, periodontal and implant surgery.Similar to PD#1 - No Impact
Principle of OperationAn electrical signal supplied to the dental piezo-ultrasonic Handpiece, comprising a piezoelectric ceramic transducer converting electrical signal into ultrasonic vibrations.Identical (Principle is described, assumed identical as "No Impact" stated for this category)Identical to PD#1 - No Impact
Dimensions (Console)Height: 160 mm, Depth: 271 mm, Width: 251 mm, Weight: 3.5 kgHeight: 136 mm, Depth: 306 mm, Width: 378 mm, Weight: 3.7 kgNo Impact
Power Supply100-230 VAC, 50/60 Hz, 150 VA100-230 VAC, 50/60 Hz, 150 VAIdentical to PD#1 - No Impact
Electrical SafetyClass 1, BF TypeClass 1, BF TypeIdentical to PD#1 - No Impact
Piezoelectric Performances (Output Ultrasonic frequency)28 to 36 kHz28 to 36 kHzIdentical to PD#1 - No Impact
Available ModesD1/ D2 / D3 / D4D1/ D2 / D3 / D4Identical to PD#1 - No Impact
Nominal Output Current (D1, D2, D3, D4 modes)215.3 mA, 171.2 mA, 123.9 mA, 120.8 mA215.3 mA, 171.2 mA, 123.9 mA, 120.8 mAIdentical to PD#1 - No Impact
Frequency Modulation (D1-D3)60 Hz60 HzIdentical to PD#1 - No Impact
Frequency Modulation (D4)30 Hz30 HzIdentical to PD#1 - No Impact
Irrigation Performances (Pump, Off, Min, Max Flow Rate)1 pump, 0 ml/min, 10 ml/min, 120 ml/min1 pump, 0 ml/min, 10 ml/min, 120 ml/minIdentical to PD#1 - No Impact
User Interface for SettingSensitive areas on front panel (LED)Keys on front panel (Monochrome LCD)No Impact (considered a minor change with no impact on safety/effectiveness)
Environmental (Storage Temperature)0 to +50 °C-20 to +70 °CNo Impact
HandpiecesCUBE LED HandpiecePIEZOTOME SOLO HandpieceNo Impact (difference in specific handpiece model, but technology and function are similar)
Biocompatibility Material (Handpiece)Polyphenylsulfone (PPSU) used in Medical ApplicationPolyphenylsulfone (PPSU) used in Medical ApplicationIdentical to PD#1 - No Impact
Standards ComplianceIEC 60601-1, IEC 60601-1-2IEC 60601-1, IEC 60601-1-2Identical to PD#1 - No Impact
Sterilization ValidationPer ISO 17665-1, ISO 17665-2, AAMI ST55, AAMI TIR 30N/A (implied met by predicate)Validated for new device
Cleaning ValidationPer FDA Guidance "Reprocessing Medical Devices..."N/A (implied met by predicate)Validated for new device
Software V&VPer "Guidance for Industry and FDA Staff..."N/A (implied met by predicate)Validated for new device

The acceptance criterion, in essence, is that the new device's performance is either identical to the predicate or that any differences have no impact on safety or effectiveness.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document refers to "Performance Testing bench" and various validation tests (Electrical Safety, EMC, Software V&V, Sterilization, Cleaning, Biocompatibility). These are bench tests or validation processes, not clinical studies with human patient data or specific "test sets" in the context of AI/ML performance evaluation. Therefore, sample size and data provenance (country, retrospective/prospective) in the clinical sense are not applicable here.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided. The device is a physical ultrasonic surgical system, not an AI/ML diagnostic tool that requires ground truth established by medical experts for image interpretation or diagnosis.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as this is related to expert consensus for ground truth establishment in studies involving interpretation, which is not the case for this device.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No such study was mentioned or performed, as the PIEZOTOME CUBE is a surgical tool, not an AI-assisted diagnostic or interpretation system for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable, as the device is an ultrasonic surgical system, not an algorithm. Performance testing was primarily bench testing and direct comparison to a predicate device's technical specifications.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

Ground truth in the context of this submission refers to the established technical specifications and performance characteristics of the predicate device, as well as compliance with recognized standards (e.g., IEC 60601-1, ISO 10993-1). For the validation tests performed (sterilization, cleaning, biocompatibility), the "ground truth" would be the successful adherence to the requirements of the specified standards and guidance documents. There is no mention of clinical outcome data or pathology reports for ground truth.

8. The sample size for the training set

Not applicable. This is not an AI/ML device that requires a training set.

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

Satelec - Acteon Group Argie Zoubroulis Quality Manager 124 Gaither Drive Suite # 140 Mt. Laurel, New Jersey 08054 April 3, 2018

Re: K172137

Trade/Device Name: PIEZOTOME CUBE Regulation Number: 21 CFR 872.4120 Regulation Name: Bone Cutting Instrument and Accessories Regulatory Class: Class II Product Code: DZI Dated: March 23, 2018 Received: April 3, 2018

Dear Argie Zoubroulis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Mary S. Runner -S

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172137

Device Name PIEZOTOME CUBE

Indications for Use (Describe)

The PIEZOTOME CUBE is an ultrasonic surgical system that supply utilities to and serve as a base for dental tips. The PIEZOTOME CUBE consist of a control unit and handpiece intended for use in intraoral surgery procedures, including osteotomy, osteoplasty, periodontics and implantology.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary [As Required by 21 CFR 807.92.c] K172137

I - SUBMITTER

01.a. 510(k) Submitter:

SATELEC 17, Avenue Gustave Eiffel BP 30216 33708 Merignac Cedex France 011-33-556-340-607 Telephone: 011-33-556-349-292 Fax: philippe.girard@acteongroup.com email:

01.b. Contact Person:

Argie ZOUBROULIS SATELEC c/o ACTEON, Inc. 124 Gaither Drive, Suite 140 Mt. Laurel, NJ 08054 800-289-6367 Ext. 390 Telephone: 856-222-4726 Fax: argie.zoubroulis@acteongroup.com Email:

8044015 1.c. Establishment Registration Number: 1.d. Date Prepared: May 2, 2018

  • 1.e. Type of 510(k) submission:
  • II DEVICE

02.a. Trade Name of Device:

02.b. Common Name of Device:

02.c. Classification Regulation:

02.d. Regulation Identification: Traditional 510(k) Submission.

PIEZOTOME CUBE

Piezoelectric Handpiece and Console

Classification Name : Bone Cutting Instrument and Accessories (21 CFR 872.4120)

A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting Handpiece, and ACpowered bone saw.

02.e. Medical Device Class: ll

Image /page/3/Picture/20 description: The image shows the logo for Acteon. The logo consists of the word "ACTEON" in a bold, sans-serif font. The letters are dark gray. A blue curved line is below the word, starting from the left and extending upwards, partially encircling the word.

{4}------------------------------------------------

02.f.Panel:Dental
02.g.Product Code:DZI
02-h.Subsequent Product Codes:Not Applicable

III - PREDICATE DEVICES

The Substantial Equivalence (SE) of PIEZOTOME CUBE is based on the Predicate Device identified in the Table 01.

Trade NameManufacturerProduct Code510(k) numberDate Cleared
PIEZOTOME SOLOPIEZOTOME SOLO LEDSATELECDZIK112188Feb 03, 2012

Table 01 Identification of I egally Marketed Predicate Devices

IV -DEVICE DESCRIPTION

SATELEC CUBE is an piezoelectric device that uses micro-vibrations of associated Tips, to perform the dental procedures defined in Indication for Use.

PIEZOTOME CUBE consists of a Console, a Multifunction footswitch and a reusable Handpiece.

The Console is the heart of the SATELEC PIEZOTOME CUBE. The Console contains the display board and the motherboard. Ultrasonic Handpiece and Footswitch are connected to the Console. The Touch screen present on the front panel is used to define the settings of the SATELEC PIEZOTOME CUBE (Ultrasonic modes, irrigation flow values).

The Ultrasonic Handpiece is held in the Practitioner's hand. The Ultrasonic Handpiece is connected to the Console via a Handpiece Cord. The Handpiece is dedicated to Dental Bone Surgery procedures. The Ultrasonic Handpiece is equipped with a Piezoelectric Transducer. The Piezoelectric transducer converts the Electrical Signal delivered by the Console into mechanical micro-vibrations. The Ultrasonic Handpiece is reusable and Sterilizable by autoclaving.

Tips are fixed by means of a Wrench at the extremity of the Ultrasonic Handpiece. Tip are in direct contact to the patient. The ultrasonic mechanical vibrations are transmitted to the Tip.

The Pump housing is designed to accommodate SATELEC Irrigation Tubing cassettes. Irrigation Tubings are Single Use and delivered under Sterile State. Irrigation is intended to cool the clinical site and rinse the fragments such as bone or teeth.

Image /page/4/Picture/12 description: The image shows the logo for ACTEON. The logo is composed of the company name in a bold, sans-serif font, with the letters in a dark gray color. A curved, light blue line extends from the bottom left of the image, arching upwards and towards the left of the word ACTEON. The overall design is simple and modern.

{5}------------------------------------------------

V - INDICATION FOR USE

The PIEZOTOME CUBE is an ultrasonic surgical system that supply utilities to and serve as a base for dental tips. The PIEZOTOME CUBE consist of a control unit and handpiece intended for use in intraoral surgery procedures, including osteotomy, osteoplasty, periodontics and implantology.

VI - COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE (SUBSTANTIAL EQUIVALENCE)

A comparison Between SATELEC PIEZOTOME CUBE and the Predicate Device is present in Table 02.

New DevicePredicate Device #1(PD#1)Impact of the differenceson Safety /Effectiveness
Trade /Device NamePIEZOTOME CUBEPIEZOTOME SOLO
KnumberK172137K112188
Indication for UseThe PIEZOTOME CUBEis an ultrasonic surgicalsystem that supplyutilities to and serve as abase for dental tips. ThePIEZOTOME CUBEconsist of a control unitand handpiece intendedfor use in intraoralsurgery procedures,including osteotomy,osteoplasty, periodonticsand implantology.The intended use of theSATELEC PIEZOTOMESOLO (or PIEZOTOMESOLO LED available inoption) is to supplyutilities to and serve as abase for dental tools andaccessories for use byqualified dentalpractitioners.Similar to PD#1 - No Impact
Intended UseThe intended use of theSATELEC PIEZOTOMECUBE is to supply utilitiesand to serve as a base fordental tools andaccessories for use byqualified dentalpractitioners.The intended use of thePIEZOTOME SOLO is tosupply utilities to andserve as a base fordental tools andaccessories for use byqualified dentalpractitioners.ThePIEZOTOME SOLO is adevice designed toperform dental surgeryincluding osteotomies,osteoplasties,periodontaland implantsurgery.Similar to PD#1 - No Impact
Code ProductDZIDZIIdentical to PD#1 - No Impact
Subsequent product//Identical to PD#1 - No Impact
Part872 - Dental devices872 - Dental devicesIdentical to PD#1 - No Impact
Regulation Number21 CFR 872.412021 CFR 872.4120Identical to PD#1 - No Impact

Table 02 – Comparison Between SATELEC PIEZOTOME CUBE and the Predicate Device

Image /page/5/Picture/6 description: The image shows the word "ACTEON" in a bold, sans-serif font. The letters are dark gray, almost black. A curved, light blue line extends from the bottom left of the image, arching upwards and to the right, partially surrounding the word. The line adds a sense of motion or energy to the logo.

{6}------------------------------------------------

006 – 510(k) Summary (SMDA Requirements) S001 Additional Inform natio n

Regulation IdentificationA bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. TheA bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert aIdentical to PD#1 - No Impact
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{7}------------------------------------------------

New DevicePredicate Device #1(PD#1)Impact of the differences onSafety / Effectiveness
Trade /Device NamePIEZOTOME CUBEPIEZOTOME SOLO
KnumberK172137K112188
device includes themanual bone drill andwire driver, poweredbone drill, rotary bonecutting handpiece, andAC-powered bone sawwire, pin, or screw. Thedevice includes themanual bone drill andwire driver, poweredbone drill, rotary bonecutting handpiece, andAC-powered bone saw
Regulation NameBone Cutting Instrumentand AccessoriesBone Cutting Instrumentand AccessoriesIdentical to PD#1 - No Impact
Regulatory ClassIIIIIdentical to PD#1 - No Impact
Principle of operation
PrincipleAn electrical signal emitted by the medical device issupplied to the dental piezo-ultrasonic Handpiece. Itcomprises a piezoelectric ceramic transducer, whichconverts the electrical signal into ultrasonicvibrations. Mechanical vibrations are transmitted to atip attached to the end of the ultrasonic HandpieceIdentical to PD#1 - No Impact
Dimensions of the Console
Height (mm)160136No Impact
Depth (mm)271306No Impact
Width (mm)251378No Impact
Weight (kg)3.53.7No Impact
Mounting UnitTable TopTable TopIdentical to PD#1 - No Impact
Power Supply
Supply Voltage (VAC)100 VAC to 230 VAC100 VAC to 230 VACIdentical to PD#1 - No Impact
Frequency (Hz)50 Hz / 60 Hz50 Hz / 60 HzIdentical to PD#1 - No Impact
Power Consumption (VA)150150Identical to PD#1 - No Impact
Electrical Safety Classification
Equipment ClassificationClass 1Class 1Identical to PD#1 - No Impact
Electrical TypeBF TypeBF TypeIdentical to PD#1 - No Impact
Applied Parts on the Console
Quantity of connectors forApplied Parts11Identical to PD#1 - No Impact
LocalizationFront Panel of theDeviceFront Panel of theDeviceIdentical to PD#1 - No Impact
Internal constitution of the Console
Power supplySwitch Power SupplySwitch Power SupplyIdentical to PD#1 - No Impact
Mother Board Fire aspectsUL94-V0 (self-extinguishing material)UL94-V0 (self-extinguishing material)Identical to PD#1 - No Impact
Display Board FireaspectsUL94-V0 (self-extinguishing material)UL94-V0 (self-extinguishing material)Identical to PD#1 - No Impact
External constitution of the Console
Fire aspects (for Casing)UL94-V0 (self-extinguishing material)UL94-V0 (self-extinguishing material)Identical to PD#1 - No Impact
Pictures
Picture of the ConsoleImage: Picture of the ConsoleImage: Picture of the ConsoleNo impact
New DevicePredicate Device #1(PD#1)Impact of the differences onSafety / Effectiveness
Trade /Device NamePIEZOTOME CUBEPIEZOTOME SOLO
KnumberK172137K112188
Picture of the FootswitchImage: footswitchImage: footswitchIdentical to PD#1 - No Impact
Piezoelectric Performances for Dental Bone Surgery
TechnologySATELEC PIEZOTOMEtopologySATELEC PIEZOTOMEtopologyIdentical to PD#1 - No Impact
OutputUltrasonicfrequency (kHz)28 to 3628 to 36Identical to PD#1 - No Impact
Available modesD1/ D2 / D3 / D4D1/ D2 / D3 / D4Identical to PD#1 - No Impact
Nominal specification forOutput Current Value inD1 mode (mA)215.3215.3Identical to PD#1 - No Impact
Nominal specification forOutput Current Value inD2 mode (mA)171.2171.2Identical to PD#1 - No Impact
Nominal specification forOutput Current Value inD3 mode (mA)123.9123.9Identical to PD#1 - No Impact
Nominal specification forOutput Current Value inD4 mode (mA)120.8120.8Identical to PD#1 - No Impact
Frequency Modulation forD1 mode (Hz)6060Identical to PD#1 - No Impact
Frequency Modulation forD2 mode (Hz)6060Identical to PD#1 - No Impact
Frequency Modulation forD3 mode (Hz)6060Identical to PD#1 - No Impact
Frequency Modulation forD4 mode (Hz)3030Identical to PD#1 - No Impact
Irrigation Performances
Irrigation Pump(s)11Identical to PD#1 - No Impact
Off Irrigation (ml/min)00Identical to PD#1 - No Impact
Minimum Irrigation FlowRate (nominal) (ml/min)1010Identical to PD#1 - No Impact
Maximum Irrigation FlowRate (nominal) (ml/min)120120Identical to PD#1 - No Impact
Localization of the Mains Power Switch
Power-On / Power-OffMains switch localized atthe rear of the DeviceMains switch localizedat the rear of the DeviceIdentical to PD#1 - No Impact
User Interface for setting
User interface forinformation on settingsLED (Light EmittedDiode)Monochrome LCDNo Impact
Ultrasonic Power SettingkeysSensitive areas on frontpanelKeys on front panelNo Impact
Irrigation Flow SettingkeysSensitive areas on frontpanelKeys on front panelNo Impact
Footswitch
Type of FootswitchMultifunctionMultifunctionIdentical to PD#1 - No Impact
Number of actuators onFootswitch2 Functions2 FunctionsIdentical to PD#1 - No Impact
Environmental
New DevicePredicate Device #1(PD#1)Impact of the differences onSafety / Effectiveness
Trade /Device NamePIEZOTOME CUBEPIEZOTOME SOLO
KnumberK172137K112188
Where usedUsable in all medicalpremisesExcept Operatingtheater / outdoorUsable in all medicalpremisesExcept Operatingtheater / outdoorIdentical to PD#1 - No Impact
OperatingTemperature(°C)+10 to +30+10 to +30Identical to PD#1 - No Impact
Storage Temperature (°C)0 to +50-20 to +70No Impact
MaximumOperatingAltitude (m)20002000Identical to PD#1 - No Impact
Anatomical siteBones in the oral sphereBones in the oralsphereIdentical to PD#1 - No Impact
Handpieces
DentalBoneSurgeryHandpieceCUBE LED HandpiecePIEZOTOME SOLOHandpieceNo Impact
Type of Applied PartUltrasonic HandpieceUltrasonic HandpieceIdentical to PD#1 - No Impact
TechnologyusedforHandpiecesPiezoelectric technologyPiezoelectric technologyIdentical to PD#1 - No Impact
Dimensions of the Dental Bone Surgery Handpieces
Length (mm)130130Identical to PD#1 - No Impact
Maximum Diameter (mm)2323Identical to PD#1 - No Impact
Irrigation Tubing
Length (mm)25502550Identical to PD#1 - No Impact
Principle of irrigationPeristaltic tapePeristaltic tapeIdentical to PD#1 - No Impact
Dimensions of the Footswitch
Width (mm)173173Identical to PD#1 - No Impact
Height (mm)140140Identical to PD#1 - No Impact
Depth (mm)176176Identical to PD#1 - No Impact
Ingress Protection Rating(IPxx)IPX1IPX1Identical to PD#1 - No Impact
Cleaning
DentalBoneUltrasonicHandpieceRe-usable, CleanableRe-usable, CleanableIdentical to PD#1 - No Impact
Unit CasingRe-usable, CleanableRe-usable, CleanableIdentical to PD#1 - No Impact
Sterilization / Sterile State
DentalBoneUltrasonicHandpieceSterilizable132°C / 4 minSterilizable132°C / 4 minIdentical to PD#1 - No Impact
Irrigation TubingProvided under SterileState - Single UseProvided under SterileState - Single UseIdentical to PD#1 - No Impact
Biocompatibility
UltrasonicHandpiecematerialincontacttoPatientPolyphenylsulfone(PPSU) used in MedicalApplicationPolyphenylsulfone(PPSU) used in MedicalApplicationIdentical to PD#1 - No Impact
Standards
Safety StandardIEC 60601-1IEC 60601-1Identical to PD#1 - No Impact
EMC StandardIEC 60601-1-2IEC 60601-1-2Identical to PD#1 - No Impact
Summary of Major Changes
Major Changes - User Interface for setting
User interface for SettingSensitive areas withLight Emitted Diode(LED)Monochrome LCDNo Impact
Ultrasonic Power SettingAccessible on Sensitivearea on front panelAccessible on Keyboardon front panelNo Impact
New DevicePredicate Device #1(PD#1)
Trade /Device NamePIEZOTOME CUBEPIEZOTOME SOLOImpact of the differences onSafety / Effectiveness
KnumberK172137K112188
Irrigation Flow SettingAccessible on Sensitivearea on front panelAccessible on Keyboardon front panelNo Impact
Handpieces
Dental Bone SurgeryHandpieceCUBE LED HandpiecePIEZOTOME SOLOHandpiece
Pictures of Dental BoneSurgery HandpiecesImage: CUBE LED HandpieceImage: PIEZOTOME SOLO HandpieceNo Impact

Image /page/7/Picture/1 description: The image shows the logo for ACTEON. The logo is in a dark gray color and is in all capital letters. There is a light blue swoosh that starts below the A and goes up and over the top of the letters.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the word "ACTEON" in a bold, sans-serif font. The letters are dark gray, almost black. A curved, light blue line appears below and to the left of the word, adding a visual element to the logo.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for ACTEON. The text "ACTEON" is written in a bold, sans-serif font and is dark gray. A blue swoosh design is located to the left and below the text, adding a visual element to the logo.

{10}------------------------------------------------

A lot of features and characteristics of PIEZOTOME CUBE are identical to the Predicate Device #1. The ldentified differences have no impact on the Substantial Equivalence. Moreover, the identified Major Changes have no impact on the Substantial Equivalence.

VII - PERFORMANCE DATA

Electromagnetic Compatibility Test: The Electromagnetic Compatibility Tests have been performed according to IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances -- requirements and tests (Recognition Number 19-12).

Electrical Safety Tests: The Electrical Safety Tests have been performed according to IEC 60601-1:2005, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (General). (Recognition Number 5-71) included USA and Canadian Deviations.

Software Verification and Validation: Software Verification activities were conducted and documented according to the document named "Guidance for Industry and FDA Staff - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" dated May 11, 2005.

Performance Testing bench: The comparison of performances between PIEZOTOME CUBE and Predicate Device has been performed. The Technical Characteristics for PIEZOTOME CUBE (Piezoelectric Performances, Frequency Modulations and Irrigation Performances) are similar as the Predicate Device PIEZOTOME SOLO (K112188, cleared February 03, 2012).

Sterilization Validation: The Sterilizability Tests of SATELEC CUBE LED Handpiece have been performed according to applicable Standards for Re-Usable accessories (i.e., ISO 17665-1, ISO 17665-2, AAMI ST55, and AAMI TIR 30).

Cleaning Validation: Cleaning validation of the PIEZOTOME CUBE was conducted per the FDA Guidance Document entitled, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" dated March 17, 2015.

Biocompatibility Validation: The Biocompatibility Tests have been performed according to applicable Standards for Accessories in contact to the Patient. Tests were conducted according to the ISO 10993-1 Standard and 510(k) Memorandum - #G95-1 Table 1 "Initial Evaluation Tests for Consideration

Image /page/10/Picture/10 description: The image shows the logo for Acteon. The logo consists of the word "ACTEON" in a bold, sans-serif font. A blue curved line is on the left side of the word. The word is in a dark gray color.

{11}------------------------------------------------

requirements (Body Contact: External communicating Device – Type of Tissues: Tissue / Bone Dentin Communicating - Contact Duration: A limited less than 24 hours), SATELEC CUBE LED Handpiece is the same as SATELEC PIEZOTOME SOLO Handpiece used with the Predicate Device PIEZOTOME SOLO (K112188, cleared February 03, 2012).

VIII -CONCLUSION

Following all information contained in this Pre-Market Notification Submission File, we can declare that the SATELEC PIEZOTOME CUBE is Substantially Equivalent (SE) to the SATELEC Predicate Device PIEZOTOME SOLO (K112188, cleared February 03, 2012).

End of Section (No other information after this line expect Appendix if applicable)

Image /page/11/Picture/4 description: The image shows the logo for Acteon. The logo consists of the word "ACTEON" in a bold, sans-serif font. A blue curved line is placed below and to the left of the word. The color of the text is a dark gray.

§ 872.4120 Bone cutting instrument and accessories.

(a)
Identification. A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.(b)
Classification. Class II.