K Number
K172110

Validate with FDA (Live)

Date Cleared
2018-01-11

(182 days)

Product Code
Regulation Number
892.1000
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ECHELON Oval MRI System is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.

Device Description

The ECHELON OVAL is a Magnetic Resonance Imaging System that utilizes a 1.5 Tesla superconducting magnet in a gantry design. The design was based on the ECHELON MRI system. The ECHELON OVAL has been designed to enhance clinical utility as compared to the ECHELON by taking advantage of open architecture.

AI/ML Overview

The provided text is a 510(k) summary for the ECHELON OVAL V6.0A MRI system, which is a software update to an existing MRI system. The document states that "The ECHELON OVAL V6.0A MRI System is only a software update with new and improved features." This implies that the core hardware and fundamental MRI functionalities (such as image acquisition, magnet strength, etc.) remain as per the predicate device (ECHELON OVAL V5.1 MRI System, K153547).

The performance evaluation primarily focuses on the new and improved software features rather than establishing a new acceptance criteria for the entire MRI system from scratch. The acceptance criteria essentially revolve around demonstrating that these new and improved features perform as intended and produce diagnostically acceptable images.

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria CategorySpecific Acceptance Criteria (Inferred)Reported Device Performance
New Features PerformanceEach new feature (Computed DWI HiMAR, QSM) performs as intended for diagnostic use."Test data confirmed that each new and improved feature perform as intended for diagnostic use."
Improved Features PerformanceEach improved feature (Soft Sound Package, R2* Map Analysis, Advanced contrast enhanced abdominal imaging, ASL-Perfusion (GRASE), Cine-PC Analysis) performs as intended for diagnostic use."Test data confirmed that each new and improved feature perform as intended for diagnostic use."
Clinical Image Quality for New/Improved FeaturesClinical images from applicable new and improved features have acceptable image quality for clinical use."Clinical image examples are provided for each applicable new and improved feature and that we judged to be sufficient to evaluate clinical usability. In addition, a radiologist validated that the clinical images have acceptable image quality for clinical use."
Overall Substantial EquivalenceThe ECHELON OVAL V6.0A MRI system is substantially equivalent to the predicate device (ECHELON Oval V5.1 MRI System (K153547)) in terms of hardware, base elements of software, safety, effectiveness, and functionality.This is the overall conclusion of the 510(k) submission, based on the rationale that hardware and coil differences are not significant, and functionality enhancements do not constitute a new intended use or raise different questions of safety and effectiveness.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not explicitly state the numerical sample size for the test set. It mentions "Clinical image examples are provided for each applicable new and improved feature." The provenance of the data (country of origin, retrospective or prospective) is not specified.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Number of experts: Singular ("a radiologist").
Qualifications of experts: Only "a radiologist" is mentioned. No specific experience level (e.g., 10 years of experience) is provided.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

No formal adjudication method (like 2+1 or 3+1) is described. The text states: "a radiologist validated that the clinical images have acceptable image quality for clinical use." This implies a single expert review for validation.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No such study (MRMC comparative effectiveness study with AI assistance) is mentioned or implied. The device is purely an MRI system software update, not an AI-powered diagnostic aid that assists human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable in the context of this device. The device is a Magnetic Resonance Imaging System, designed to produce images for interpretation by a trained physician. It's not an an algorithm performing a standalone diagnostic task without human intervention.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth for the test set, specifically for the clinical image quality, appears to be based on expert opinion/validation by a radiologist. The document doesn't mention pathology, outcomes data, or a formal expert consensus process.

8. The sample size for the training set

The document does not mention a "training set" or its sample size. This is consistent with a software update for an MRI system, where the focus is on verifying the performance of new/improved features rather than an AI/ML algorithm that undergoes a distinct training phase.

9. How the ground truth for the training set was established

Not applicable, as no training set is mentioned in the context of this device's evaluation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 11, 2018

Hitachi Healthcare Americas Corporation Doug Thistlethwaite Manager of Regulatory Affairs 1959 Summit Commerce Park Twinsburg, Ohio 44087

Re: K172110

Trade/Device Name: ECHELON OVAL V6.0A MRI System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH Dated: November 30, 2017 Received: December 01 2017

Dear Doug Thistlethwaite:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

Page 2 - Doug Thistlethwaite

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael D. O'Hara For

Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172110

Device Name ECHELON OVAL V6.0A MRI system

Indications for Use (Describe)

The ECHELON Oval MRI System is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.

Anatomical Region:Head, Body, Spine, Extremities
Nucleus excited:Proton
Diagnostic uses:T1, T2, proton density weighted imaging
Diffusion weighted imaging
MR Angiography
Image processing
Spectroscopy
Whole Body
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5

510(k) Summary

{4}------------------------------------------------

Submitter:Hitachi Healthcare Americas1959 Summit Commerce ParkTwinsburg, Ohio 44087-2371
Contact:Douglas J. Thistlethwaite
Telephone number:330-425-1313
Telephone number:330-963-0749
E-mail:thistlethwaited@HitachiHealthcare.com
Date:June 14, 2017

Subject Device Name

Trade/Proprietary Name:ECHELON OVAL V6.0A MRI system
Regulation Number:21 CFR 892.1000
Regulation Name:System, Nuclear Magnetic Resonance Imaging
Product Code:LNH
Class:2
Panel:Radiology

Predicate Device Name

Predicate Device(s):ECHELON OVAL V5.1 MRI System (K153547)
Regulation Number:21 CFR 892.1000
Regulation Name:System, Nuclear Magnetic Resonance Imaging
Product CodeLNH
Class2
PanelRadiology

Device Intended Use

The ECHELON Oval System is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved crosssectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.

Anatomical Region: Head, Body, Spine, Extremities Nucleus excited: Proton Diagnostic uses: T1, T2, proton density weighted imaging Diffusion weighted imaging MR Angiography Image processing Spectroscopy

Whole Body

{5}------------------------------------------------

Device Description

Function

The ECHELON OVAL is a Magnetic Resonance Imaging System that utilizes a 1.5 Tesla superconducting magnet in a gantry design. The design was based on the ECHELON MRI system. The ECHELON OVAL has been designed to enhance clinical utility as compared to the ECHELON by taking advantage of open architecture.

Scientific Concepts

Magnetic Resonance imaging (MRI) is based on the fact that certain atomic nuclei have electromagnetic properties that cause them to act as small spinning bar magnets. The most ubiquitous of these nuclei is hydrogen, which makes it the primary nuclei currently used in magnetic resonance imaging. When placed in a static magnetic field, these nuclei assume a net orientation or alignment with the magnetic field, referred to as a net magnetization vector. The introduction of a short burst of radiofrequency (RF) excitation of a wavelength specific to the magnetic field strength and to the atomic nuclei under consideration can cause a re-orientation of the net magnetization vector. When the RF excitation is removed, the protons relax and return to their original vector. The rate of relaxation is exponential and varies with the character of the proton and its adiacent molecular environment. This re-orientation process is characterized by two exponential relaxation times, called T1 and T2. A RF emission or echo that can be measured accompanies these relaxation events.

The emissions are used to develop a representation of the relaxation events in a three dimensional matrix. Spatial localization is encoded into the echoes by varying the RF excitation, applying appropriate magnetic field gradients in the x, y, and z directions, and changing the direction and strength of these gradients. Images depicting the spatial distribution of the NMR characteristics can be reconstructed by using image processing techniques similar to those used in computed tomography.

Physical and Performance Characteristics

MRI is capable of producing high quality anatomical images without the associated risks of ionizing radiation. The biological properties that contribute to MR image contrast are different from those responsible for x-ray image contrast. In MR imaging, difference in proton density, blood flow, and T1 and T2 relaxation times can all contribute to image contrast. By varying the pulse sequence characteristics, the resulting images can emphasize T1, T2, proton density, or the molecular diffusion of water or other proton containing molecules. And MR system has the Function of measuring spectroscopy.

Performance Evaluation

The ECHELON OVAL V6.0A MRI System is only a software update with new and improved features.

Therefore, a clinical and performance evaluation was conducted on the new features which include:

  • . Computed DWI HiMAR . QSM .
    HiMAR can be used for reduction of metal artifacts around passive MR Conditional implants.

In addition, a clinical and performance evaluation was conducted on the improved features which include:

  • Soft Sound Package

  • R2* Map Analysis o

  • Advanced contrast enhanced abdominal imaging ●

  • ASL-Perfusion (GRASE) ●

  • Cine-PC Analysis

{6}------------------------------------------------

A rationale analysis was then conducted and the results are contained in Table 1.

Table 1 Performance Analysis

Testing TypeRationale Analysis
Performance Testing - BenchPerformance bench testing was conducted on the applicable new and improved features.Test data confirmed that each new and improved feature perform as intended for diagnostic use.
Performance Testing - ClinicalClinical image examples are provided for each applicable new and improved feature and that we judged to be sufficient toevaluate clinical usability. In addition, a radiologist validated that the clinical images have acceptable image quality for clinicaluse.

Device Technological Characteristics

The control and image processing hardware and the base elements of the system software are identical to the predicate device. The ECHELON OVAL V6.0A MRI system software is substantially equivalent to the ECHELON OVAL V5.1 (K153547). See tables below.

The technological characteristics in regards to hardware of the ECHELON OVAL V6.0A MRI system and the predicate are listed in Table 2.

ITEMPREDICATE DEVICEECHELON OVAL V5.1 (K153547)SUBJECT DEVICEECHELON OVAL V6.0ADIFFERENCE
SystemStandards MetNEMA: MS 1, MS 2, MS 3, MS 4, MS 5, MS 8,IEC: 60601-1, 60601-1-2, 60601-2-33, 62304NEMA: MS 1, MS 2, MS 3, MS 4, MS 5, MS 8,IEC: 60601-1, 60601-1-2, 60601-2-33, 62304No
Magnet and GantryType and Field StrengthSuper-conducting magnet, horizontal bore, 1.5 TeslaSuper-conducting magnet, horizontal bore, 1.5 TeslaNo
Gradient SystemResonant Frequency63.86 MHz63.86MHzNo
Gradient Strength34mT/m34mT/mNo
Slew Rate150 T/m/sec150 T/m/secNo
Rise Time227µsec to 34mT/m227µsec to 34mT/mNo
Audible Noise (MCAN)Ambient58 dBA58 dBANo
Lpeak125 dBA125 dBANo
Leq117 dB117 dBANo
RF SystemTransmitter channels22No
Peak Envelop Power40 kW40 kWNo
Duty Cycle100% (Gating max), 12.5% at full power100% (Gating max), 12.5% at full powerNo
RF receiver channel16, 3216, 32No

Table 2 Comparison: Hardware

The hardware differences from the ECHELON OVAL V6.0A MRI system to the predicate device are analyzed in Table 3.

Table 3 Hardware Comparison Analysis

FDARequirementsAnalyze why any differences between the subject device and predicate(s) do not render the device NSE (e.g., does notconstitute a new intended use; and any differences in technological characteristics are accompanied by information thatdemonstrates the device is as safe and effective as the predicate and do not raise different questions of safety andeffectiveness than the predicate ), affect safety or effectiveness, or raise different questions of safety and effectiveness (seesection 513(i)(1)(A) of the FD&C Act and 21 CFR 807.87(f)).
DeviceModificationSummaryThere are no significant hardware changes that affect technological characteristics and safety.
SignificantChanges□ Manufacturing Process□ Labeling□ Technology□ Performance
□ Engineering□ Materials□ Others☑ None (See rationale statement)
HITACHIRationaleStatementModified specification doesn't constitute a new intended use. There are no significant changes in technological characteristics.For safety, gradient system and RF system is controlled according to same regulation as ECHELON Oval V5.1 (K153547). So,safety and effectively of the device are same as ECHELON Oval V5.1 (K153547)

{7}------------------------------------------------

The technological characteristics in regards to coils of the ECHELON OVAL V6.0A MRI system and the predicate are listed in Table 4.

ITEMPREDICATE DEVICESUBJECT DEVICEDIFFERENCE
ECHELON OVAL V5.1 (K153547)ECHELON OVAL V6.0A
RF CoilsTransmit CoilT/R BodyT/R BodyNo
Receiver CoilsWIT Posterior Head/Neck coil, WIT AnteriorHead attachmentWIT Posterior Head/Neck coil, WITAnterior Head attachmentNo
WIT Posterior Head/Neck coil BSee Table 5
WIT Torso coilWIT Torso coilNo
WIT Torso coil 12See Table 5
WIT Torso coil 8See Table 5
Extremity coil (Knee)Extremity coil (Knee)No
WIT Anterior Neck attachmentWIT Anterior Neck attachmentNo
WIT Anterior Neck attachment BNo
Hand/Wrist coilHand/Wrist coilNo
WIT Anterior NV attachmentWIT Anterior NV attachmentNo
BreastBreastNo
MP coil 140A, BMP coil 140A, BNo
Micro coil (S) A, BSee Table 5
ShoulderShoulderNo
Shoulder coil 8See Table 5
WIT Spine coil 12WIT Spine coil 12No
WIT Spine coil ASee Table 5
WIT Spine coil 8WIT Spine coil 8No
WIT Spine coil BSee Table 5
Foot/AnkleFoot/AnkleNo
Flexible Extremity (Long Bone)Flexible Extremity (Long Bone)No
WIT CardiacWIT CardiacNo
PVPVNo

Table 4 Comparison: RF Coils

The coil differences from the ECHELON OVAL V6.0A MRI system to the predicate device are analyzed in Table 5.

Table 5 Coil Comparison Analysis

FDARequirementsAnalyze why any differences between the subject device and predicate(s) do not render the device NSE (e.g., does notconstitute a new intended use; and any differences in technological characteristics are accompanied by information thatdemonstrates the device is as safe and effective as the predicate and do not raise different questions of safety andeffectiveness than the predicate ), affect safety or effectiveness, or raise different questions of safety and effectiveness (seesection 513(i)(1)(A) of the FD&C Act and21 CFR 807.87(f)).
DeviceModificationSummaryWIT Posterior Head/Neck coil B, WIT Anterior Neck attachment B, WIT Spine coil A, WIT Spine coil B, Micro coil (S) A, B,Shoulder coil 8, WIT Torso coil 12 and WIT Torso coil 8 are added. The performance or safety and effectiveness are thesame as WIT Posterior Head/Neck coil, WIT Anterior Neck attachment, WIT Spine coil 12, MP coil 140A,B, Shoulder and WIT Torso coil for ECHELON Oval V5.1 (K153547).
SignificantChanges□ Manufacturing Process□ Labeling□ Technology□ Performance
□ Engineering□ Materials□ Others☑ None (See rationale statement)
HITACHIRationaleStatementModified functions don't constitute a new intended use. There are no significant changes in technological characteristics.During transmitter coil operation, RF Coils are de-resonated by same scheme as ECHELON Oval V5.1 (K153547).

{8}------------------------------------------------

The technological characteristics in regards to changes in functionality of the ECHELON Oval MRI System as compared to the predicate are listed in Table 6.

Table 6 Comparison: Functionality

ITEMDIFFERENCESANALYSIS
Operating SystemNoneNo
CPU PlatformNoneNo
Application SoftwareGoing from V5.1A to V6.0ASee Table 7
Scan TasksRAPID capability is changed to 4.0 per a direction (16.0 per 2 dir.)See Table 7
2D Processing TasksImage Division, and R2/R2* calculation are available.See Table 7
3D Processing TasksNoneNo
Analysis TasksComputed DWI and volume flow analysis are available.See Table 7
Maintenance TasksNoneNo
Viewport ToolsNoneNo
Film, Archive ToolsNoneNo
Network ToolsNoneNo
Protocol EnhancementsRADAR (2D Soft GE, 2D Soft FSE, 2D Soft FIR, 2D Soft SE, and 3D Soft TOF), Soft Sound (2D SoftFSE (Rephase: off), 2D Soft FSE (Rephase:Slice), 2D Soft FSE (RADAR), 2D Soft FIR (RADAR), 2DSoft SE (RADAR), 3D Soft TOF (RADAR), 2D Soft GE, 2D Soft GE (RADAR), and 2D Soft SARGE),ASL-Perfusion (GRASE), Advanced contrast enhanced abdominal imaging, HiMAR, and QSM areavailable.See Table 7
Pulse Sequences2D Soft GE, 2D Soft SARGE, and 3D GRASE are available.See Table 7

The functionality differences from the ECHELON Oval MRI System to the predicate device are analyzed in

Table 7 Functionality Comparison Analysis

FDARequirementsAnalyze why any differences between the subject device and predicate(s) do not render the device NSE (e.g., does notconstitute a new intended use; and any differences in technological characteristics are accompanied by information thatdemonstrates the device is as safe and effective as the predicate and do not raise different questions of safety andeffectiveness than the predicate ), affect safety or effectiveness, or raise different questions of safety and effectiveness (seesection 513(i)(1)(A) of the FD&C Act and 21 CFR 807.87(f)).
DeviceModificationSummaryImage Division, and R2/R2* calculation of Parameter Analysis in the 2D Processing Tasks are added to productspecification.Computed DWI of DWI Analysis and volume flow of Velocity Analysis in the Analysis Tasks category are added toproduct specification.2D Soft GE, 2D Soft FSE, 2D Soft FIR, 2D Soft SE, and 3D Soft TOF of RADAR in the Protocol Enhancementscategory are added to product specification.2D Soft FSE (Rephase: off), 2D Soft FSE (Rephase:Slice), 2D Soft FSE (RADAR), 2D Soft FIR (RADAR), 2D Soft SE(RADAR), 3D Soft TOF (RADAR), 2D Soft GE, 2D Soft GE (RADAR), and 2D Soft SARGE of Soft Sound in theProtocol Enhancements category are added to product specification.ASL-Perfusion (GRASE), Advanced contrast enhanced abdominal imaging, HiMAR, and QSM in the ProtocolEnhancements category are added to product specification.Soft GE and Soft SARGE, which are used for the Protocol Enhancements of Soft Sound, and GRASE, which are usedfor the Protocol Enhancements of ASL-Perfusion (GRASE) in pulse sequences category are added to functionalspecification.
SignificantChanges□ Manufacturing Process□ Labeling□ Technology□ Performance
□ Engineering□ Materials□ Others☑ None (See rationale statement)
HITACHIRationaleStatementModified functions do not constitute a new intended use. There are no significant changes in technological characteristics. Forsafety, pulse sequences are controlled according to same safety limits as ECHELON Oval V5.1 (K153547). So safety andeffectivity of the device are equivalent to the ECHELON Oval V5.1 (K153547).

{9}------------------------------------------------

Substantial Equivalence

A summary decision was based on analysis of Table 8.

Table 8 Rationale Analysis: ECHELON Oval MRI vs. Predicate

ITEMOverall Rationale Analysis
HardwareDifferent specifications do not constitute a new intended use. There are no significant changes in technological characteristics, safety and effectiveness.
CoilsModified functions don't constitute a new intended use. There are no significant changes in technological characteristics, safety and effectiveness.
FunctionalityEnhanced features do not constitute a new intended use. There are no significant changes in technological characteristics, safety and effectiveness. The feature set of the device is generally equivalent to the Predicate.

Therefore, based on a thorough analysis and comparison of the functions, scientific concepts, physical and performance characteristics, performance comparison and technological characteristics, the proposed ECHELON Oval MRI is considered substantially equivalent to the currently marketed predicate device (ECHELON Oval V5.1 MRI System (K153547)) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.

Summary of Non-Clinical Testing

The revisions to the ECHELON Oval MRI System software will have no effect on the standards tests which were conducted on the ECHELON Oval V5.1 MRI System (K153547) and included in the original submission.

Therefore, ECHELON Oval MRI System is in conformance with the applicable parts of the following standards:

  • NEMA MS 1-2008, Determination of Signal-to-noise Ratio (SNR) in Diagnostic Magnetic . Resonance Images
  • . NEMA MS 2-2008, Determination of Two-Dimensional Geometric Distortion in Diagnostic Magnetic Resonance Images
  • NEMA MS 3-2008, Determination of Image Uniformity in Diagnostic Magnetic Resonance Images ●
  • NEMA MS 4-2010, Acoustic Noise Measurement Procedure for Diagnostic Magnetic Resonance Imaging Devices
  • . NEMA MS 5-2010, Determination of Slice Thickness in Diagnostic Resonance Imaging
  • NEMA MS 8-2008, Characterization of the Specific Absorption Rate for Magnetic Resonance Imaging Systems
  • AAMI / ANSI ES60601-1:2005/(R) 2012 and A1:2012, c1:2009/(r) 2012 and a2:2010/(r ) 2012 (consolidated text) medical electrical equipment - part 1: general requirements for basic safety and essential performance (iec 60601-1:2005, mod).
  • IEC 60601-1-2 Edition 3:2007-03, medical electrical equipment part 1-2: general requirements ● for basic safety and essential performance - collateral standard: electromagnetic compatibility requirements and tests.
  • IEC 60601-2-33 Edition 3.1 2013-04. medical electrical equipment part 2-33; particular ● requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnostic.
  • IEC 62304 First edition 2006-05, medical device software software life cycle processes. ●

{10}------------------------------------------------

In addition, tests were conducted on the new and improved features of the ECHELON OVAL V6.0A MRI system, they include:

  • Test Report: ASL-Perfusion (GRASE) ●
  • Test Report: Advanced contrast enhanced abdominal imaging
  • Test Report: Cine-PC Analysis
  • Test Report: Computed DWI ●
  • Test Report: HiMAR .
  • Test Report: QSM .
  • Test Report: R2* Map Analysis .
  • Test Report: Soft Sound Package o

Summary of Clinical Testing

Clinical images were collected and analyzed, to ensure that images from the new features meet user needs.

As a result of the analysis:

Testing TypeRationale Analysis
Performance Testing - ClinicalClinical image examples are provided for each applicable new and improved feature and that we judged to be sufficient toevaluate clinical usability. In addition, a radiologist validated that the clinical images have acceptable image quality forclinical use.

Conclusions

It is the opinion of Hitachi, the ECHELON OVAL V6.0A MRI system is substantially equivalent with respect to hardware, base elements of the software, safety, effectiveness, and functionality to the ECHELON Oval V5.1 MRI System (K153547).

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.