K Number
K172077

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-10-27

(109 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The enlighten III laser system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery.

1064 nm:
The 1064 nm wavelength of the enlighten III laser system is indicated for:

  • treatment of benign pigmented lesions on patients with all skin types (Fitzpatrick I-VI)
  • tattoo removal for dark colored tattoo inks and for multicolored tattoos containing dark colored tattoo inks on patients with all skin types (Fitzpatrick I-VI)

532 nm:

  • The 532 nm wavelength of the enlighten III laser system is indicated for:
  • treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-IV
  • tattoo removal for lighter colored tattoo inks, including red and yellow inks, on patients with Fitzpatrick skin types I-III

670 nm:

  • The 670 nm wavelength of the enlighten III laser system is indicated for:
  • treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-III
  • tattoo removal for lighter colored tattoo inks, including green and blue inks, on patients with Fitzpatrick skin types II-IV
Device Description

The enlighten III laser system is a multi-wavelength, pulsed laser system. A key feature of the device is its ability to produce multiple laser wavelengths and pulse widths (1064 and 532 nm: 750 ps and 2 ns; 670 nm: 660 ps and 2 ns).

The laser, power supplies, and control electronics are housed inside a console equipped with a touchscreen control panel. The laser treatment parameters are selected using the control panel. Laser emission is activated by depressing a footswitch. An articulated arm with rotational mirror knuckles delivers the laser beam from a laser head inside the console to a handpiece. The handpiece is equipped with an optical zoom assembly that determines the spot size of the laser beam on the treatment surface. The beam of a low-power diode laser is also transmitted through the arm to provide an aiming beam.

AI/ML Overview

The Cutera enlighten III Laser System was studied to assess the safety and efficacy of its 670 nm wavelength for clearing tattoos containing green and/or blue ink.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Primary Efficacy Endpoint)Reported Device Performance (670 nm wavelength for green/blue ink)
≥ 50% clearing of green and blue inks in ≥ 80% of patientsGreater than 50% clearing was seen in green and blue ink in the tattoos for 84% of subjects.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 27 subjects were initially enrolled. Data analysis was conducted on 25 subjects (2 subjects voluntarily withdrew).
  • Data Provenance: The study was an IRB-approved, prospective, two-center clinical study. The country of origin is not explicitly stated in the provided text.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Number of Experts: Three blinded dermatologists.
  • Qualifications: "Blinded dermatologists" is the only qualification mentioned. Specific experience levels (e.g., years of experience) are not provided.

4. Adjudication Method for the Test Set

  • The text states that "Before and after images of the twenty-five subject tattoos were randomized and sent to three blinded dermatologists for correct identification of the before from after image, and grading of the clearing of the green and/or blue ink."
  • It then mentions that "All blinded dermatologists correctly identified the before image for all subjects."
  • Regarding grading, the aggregated result is reported (84% of subjects achieved >50% clearing). The specific adjudication method for combining the dermatologists' individual grading scores (e.g., simple majority, average, or if discrepancies were resolved) is not explicitly stated. However, the fact that an overall performance metric is given implies a consensus or combined outcome.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study comparing AI-assisted vs. non-AI-assisted human readers was not conducted. This study focused on the device's standalone performance in a clinical setting.

6. Standalone Performance Study

  • Yes, a standalone study was conducted. The clinical trial assessed the safety and efficacy of the enlighten III laser system (the algorithm/device) for tattoo removal when used by medical professionals. This is a direct measure of the device's performance without specific "human-in-the-loop" AI assistance considerations.

7. Type of Ground Truth Used

  • Expert Consensus (Blinded Dermatologist Assessment): The ground truth for tattoo clearing was established through the photographic assessment and grading by three blinded dermatologists.
  • Investigator Assessment: There was also an investigator assessment of tattoo clearing at six-week post-treatment.

8. Sample Size for the Training Set

  • The document does not provide information about a separate training set or its sample size. The clinical study described appears to be a validation/test set for the device's performance.

9. How the Ground Truth for the Training Set Was Established

  • As no information about a separate training set is provided, the method for establishing its ground truth is not applicable from this document. The device's performance in the clinical study was evaluated against the ground truth established by the blinded dermatologists and investigators using the actual subjects' tattoos.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.

October 27, 2017

Cutera, Inc. Bradley Renton Vice President, Regulatory and Medical Affairs, & Compliance Officer 3240 Bayshore Blvd. Brisbane, California 94005

Re: K172077

Trade/Device Name: Cutera enlighten III Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: October 5, 2017 Received: October 6, 2017

Dear Bradley Renton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Jennifer R. Stevenson -S3

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172077

Device Name Cutera enlighten III Laser System

Indications for Use (Describe)

The enlighten III laser system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery.

1064 nm:

The 1064 nm wavelength of the enlighten III laser system is indicated for:

  • · treatment of benign pigmented lesions on patients with all skin types (Fitzpatrick I-VI)
  • tattoo removal for dark colored tattoo inks and for multicolored tattoos containing dark colored tattoo inks on patients with all skin types (Fitzpatrick I-VI)

532 nm:

  • The 532 nm wavelength of the enlighten III laser system is indicated for:
  • · treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-IV
  • · tattoo removal for lighter colored tattoo inks, including red and yellow inks, on patients with Fitzpatrick skin types I-III

670 nm:

  • The 670 nm wavelength of the enlighten III laser system is indicated for:
  • · treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-III
  • · tattoo removal for lighter colored tattoo inks, including green and blue inks, on patients with Fitzpatrick skin types II-IV

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)□ Over-The-Counter Use (21 CFR 801 Subpart C)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5 510(K) Summary

This 510(K) Summary of safety and effectiveness for the enlighten III laser system is submitted in accordance with the requirements of the SMDA 1990 and following guidance concerning the organization and content of a 510(K) summary.

Applicant:Cutera, Inc.
Address:3240 Bayshore Blvd., Brisbane, CA 94005
Contact Person:Bradley Renton
415-657-5568 - phone415-715-3568 - faxbrenton@cutera.com
Preparation Date:October 5, 2017
Device Trade Name:enlighten III Laser System
Common Name:Dermatology Laser
Classification Name:Instrument, Surgical, Powered, laser79-GEX, 21 CFR 878.4810
Legally MarketedPredicate Devices:Cutera enlighten III Laser System (K170936)Cynosure PicoSure (K160480)Syneron Candela PicoWay (K170597)Cynosure RevLite (K133254)
Device Description:The enlighten III laser system is a multi-wavelength, pulsed lasersystem. A key feature of the device is its ability to produce multiplelaser wavelengths and pulse widths (1064 and 532 nm: 750 ps and2 ns; 670 nm: 660 ps and 2 ns).The laser, power supplies, and control electronics are housedinside a console equipped with a touchscreen control panel. Thelaser treatment parameters are selected using the control panel.Laser emission is activated by depressing a footswitch. Anarticulated arm with rotational mirror knuckles delivers the laserbeam from a laser head inside the console to a handpiece. Thehandpiece is equipped with an optical zoom assembly thatdetermines the spot size of the laser beam on the treatmentsurface. The beam of a low-power diode laser is also transmittedthrough the arm to provide an aiming beam.

{4}------------------------------------------------

Section 5 510(K) Summary

Indications for Use:

1064 nm

The 1064 nm wavelength of the enlighten III laser system is indicated for:

  • . treatment of benign pigmented lesions on patients with all skin types (Fitzpatrick I-VI)
  • tattoo removal for dark colored tattoo inks and for . multicolored tattoos containing dark colored tattoo inks on patients with all skin types (Fitzpatrick I-VI)

532 nm

The 532 nm wavelength of the enlighten III laser system is indicated for:

  • . treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-IV
  • . tattoo removal for lighter colored tattoo inks, including red and yellow inks, on patients with Fitzpatrick skin types I-III

670 nm

The 670 nm wavelength of the enlighten III laser system is indicated for:

  • . treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-III
  • tattoo removal for lighter colored tattoo inks, including green . and blue inks, on patients with Fitzpatrick skin types II-IV

Performance Data: IEC 60601-1 Medical Electrical Equipment - Part 1: General Requirements for Safety (Edition 3.1, 2012), including:

  • . IEC 60601-1-6 Medical Electrical Equipment - Part 1-6: General Requirements for Safety - Collateral Standard: Usability (Edition 3.1, 2013) - Test Report Attachment 4
  • IEC 62366 Medical Devices Application of Usability . Engineering to Medical Devices (Edition 1.1, 2014) - Test Report Attachment 5
  • IEC 60601-2-22 Medical Electrical Equipment Part 2: . Particular Requirements for Basic Safety and Essential Performance of Surgical, Cosmetic, Therapeutic and Diagnostic Laser Equipment (Edition 3.1, 2012) – Test Report Attachment 6
  • IEC 60825-1 Safety of Laser Products Part 1: Equipment ● Classification and Requirements (Edition 2.0, 2007) - Test Report Attachment 7

The product also fulfills the requirements of AAMI/ANSI ES60601-1:2005+A2 (R2012) + A1.

IEC 60601-1-2 Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility (Edition 3, 2007)

{5}------------------------------------------------

Section 5 510(K) Summary

Results of Clinical An IRB-approved, prospective, two-center clinical study was Study: conducted to assess the safety and efficacy of the 670 nm wavelength of the enlighten III laser system for the clearing of tattoos containing green and/or blue ink.

Twenty-seven subjects with Fitzpatrick skin types II-V with unwanted tattoos containing green and/or blue ink, either alone or in combination with other ink colors were enrolled. The subjects received nanosecond and/or picosecond treatments with the Cutera enlighten III laser system with the 670 nm wavelength used to target green and blue inks; the 1064 nm wavelength used to target black and gray inks; and the 532 nm wavelength used to target lighter red and yellow inks. All wavelengths of the system could be used as required by the ink colors contained within the tattoo; however, only the clearing of green and blue inks were graded by blinded dermatologists. Standardized photographs were taken prior to and immediately after all study visits. Pain levels during treatment and adverse events were recorded during all visits.

An analysis of data was conducted on twenty-five of the twentyseven subjects originally enrolled (two subjects had voluntarily withdrawn from the study after receiving one and two treatments respectively, but prior to attending the required follow-up visits for photography). Before and after images of the twenty-five subject tattoos were randomized and sent to three blinded dermatologists for correct identification of the before from after image, and grading of the clearing of the green and/or blue ink. All blinded dermatologists correctly identified the before image for all subjects. Greater than 50% clearing was seen in green and blue ink in the tattoos for 84% of subjects indicating that the primary efficacy endpoint for the study (≥ 50% clearing of green and blue inks in ≥ 80% of patients) had been met.

Blinded photographic assessment of the 25 subject tattoos resulted in a clinically and statistically significant improvement score of 3.24 on a 0-4 point scale. The mean and standard deviation were 3.24 ±0.71. Investigator assessment of tattoo clearing at six-week post treatment resulted in a mean improvement score of 3.43 ± 0.7.

Based on subject questionnaires, sixty-nine percent of subjects reported that they were "extremely satisfied" or "satisfied" with the level of clearing and thirty-one percent of subjects reported they were neutral. No patients reported dissatisfaction. Eighty-five percent of subjects would have the investigational treatment aqain.

All subjects tolerated treatments well with few requests for topical anesthetic and without any requests for local infiltration of anesthetic even though allowed by protocol (mean pain score 4.2 ± 2.9 [0-9] on a 10-point scale). As expected, all subjects experienced erythema, edema and frosting immediately following laser treatment, and most experienced crusting in the treated area. All adverse effects resolved with no required intervention. No

{6}------------------------------------------------

Section 5 510(K) Summary

unexpected device related serious adverse events were noted.

The 670 nm wavelength of the enlighten III laser system was found to be safe and effective, with acceptable adverse effects, for tattoo removal for lighter colored tattoo inks including green and blue inks, in patients with skin types II-IV, allowing the conclusion that the enlighten laser system is substantially equivalent to the predicate devices for the requested indications.

Summary of The Cutera enlighten III laser systems used at each site in the Technological clinical trial had identical device design, components, and Characteristics: performance specifications as the currently cleared enlighten III laser system (K170936). The enlighten III device design and components are also very similar to those of the Cynosure PicoSure, Syneron Candela PicoWay, and Cynosure RevLite predicates. For each of these systems:

  • the user interface is located at the front/top of the console;
  • . the treatment handpiece is attached to an articulating arm that is connected to the main system console;
  • . laser energy is generated within the console and is transmitted through the articulated arm to the handpiece; and
  • . the handpiece allows the spot size on the skin to be adjusted according to device specifications.

Each system thus consists of an articulating arm and attached handpiece, as well as an electrically powered system console that produces the laser energy and houses the software and user interface.

The enlighten III laser system provides similar key design aspects, including the same or similar spot sizes, laser wavelengths, pulse widths, and laser types as its predicate devices. The repetition rate of the enlighten III laser system is the same as or within the repetition rate range of the predicate devices. Further, each of the devices presents a range of spot sizes to allow the user to choose the most appropriate spot size for each patient. Therefore, the minor differences do not raise any new safety or effectiveness questions, because the enlighten III parameters are the same as or within the range of the predicates.

Conclusion: Clinical testing of the enlighten III laser system demonstrated that the device performs as intended with a favorable safety profile. Results in the study were similar to those reported for the predicate devices, in support of substantial equivalence. The non-clinical data further support the safety of the device, and software verification and validation testing demonstrates that the enlighten III is expected to perform as intended in the specified use conditions. The Cutera enlighten III laser system is substantially equivalent to the predicate devices.

Summary of
Technological

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.