K Number
K172036

Validate with FDA (Live)

Date Cleared
2017-08-03

(29 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Medrobotics Flex Robotic System is intended to provide robot-assisted control of the Flex Colorectal Drive during visualization of and surgical site access to the anus, recturn and distal colon. The Flex Robotic System is intended for use in adults (≥22 years of age).

The Flex Colorectal Drive is intended for robot-assisted visualization of and surgical site access to the anus, rectum, and distal colon in adults (≥22 years of age). The Flex Colorectal Drive also provides accessory channels for compatible flexible instruments used in surgery.

Device Description

The Medrobotics Flex Robotic System is an operator controlled flexible scope that include the benefits of both a rigid scope and a computer assisted controller. This allows for the Flex Colorectal Drive to be introduced via an operator controlled user interface, easily providing transanal access to the anus, rectum and distal colon. Visualization is provided by a user selectable 2D or 3D HD camera incorporated in distal end of the scope. The Flex Robotic System's scope also provides accessory channels for the use of varied flexible surgical instruments.

AI/ML Overview

This document is a 510(k) summary for the Medrobotics Flex Robotic System. It details the device, its intended use, and substantial equivalence to a predicate device, but does not contain specific acceptance criteria or a study proving the device meets those criteria in the context of clinical performance or diagnostic accuracy. Instead, it focuses on general performance, safety, and regulatory compliance.

Therefore, many of the requested details about acceptance criteria, study design, and ground truth establishment cannot be extracted from this document as they are not present.

Based solely on the provided text, here is what can be extracted:

1. A table of acceptance criteria and the reported device performance

The document mentions that the device "has been successfully tested for function, performance, and safety as per FDA recognized Standards" and "met acceptance criteria." However, it does not provide a table of specific acceptance criteria or the numerical performance results against those criteria. It lists only the categories of testing performed.

Acceptance Criteria CategoryReported Device Performance
FunctionSuccessfully tested
PerformanceSuccessfully tested
SafetySuccessfully tested
Biocompatibility and ToxicityMet acceptance criteria to ISO 10993-1
Labeled Shelf LifeMet acceptance criteria per FDA recognized standards
ShippingMet acceptance criteria per FDA recognized standards
Sterility (ETO and Steam)Validated to a SAL of 10-6

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document. The document refers to engineering and biocompatibility testing, not clinical performance studies with patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided in the document. This type of information would be relevant for studies evaluating diagnostic accuracy or clinical outcomes, which are not detailed here.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided in the document.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not provided in the document. The device is a robot-assisted surgical system, not an AI-powered diagnostic tool, so an MRMC study comparing human readers with and without AI assistance is not applicable in this context and is not mentioned.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not provided in the document. As a robot-assisted system, it inherently involves human operators, so a standalone algorithm performance without human-in-the-loop would not be applicable in the sense of a diagnostic algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the mechanical, electrical, and materials testing mentioned, the "ground truth" would be established by reference to engineering specifications, validated test methods, and regulatory standards (e.g., IEC 60601-1, ISO 10993-1). The document doesn't detail what specific performance metrics were used as "ground truth" but implies compliance with these standards. For sterility, the ground truth is a Sterility Assurance Level (SAL) of 10-6.

8. The sample size for the training set

This information is not provided in the document. "Training set" is typically relevant for machine learning or AI models, which is not the focus of the testing described here.

9. How the ground truth for the training set was established

This information is not provided in the document.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, arranged in a stacked formation.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 3, 2017

Medrobotics Corporation John D. Bonasera Vice President of Clinical. Regulatory and Ouality Affairs 475 Paramount Drive Raynham, MA 02767

Re: K172036

Trade/Device Name: Medrobotics Flex Robotic System Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FDF Dated: July 5, 2017 Received: July 5, 2017

Dear John D. Bonasera:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Charles Viviano -S

For

Beniamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172036

Device Name Medrobotics Flex Robotic System

Indications for Use (Describe)

The Medrobotics Flex Robotic System is intended to provide robot-assisted control of the Flex Colorectal Drive during visualization of and surgical site access to the anus, recturn and distal colon. The Flex Robotic System is intended for use in adults (≥22 years of age).

The Flex Colorectal Drive is intended for robot-assisted visualization of and surgical site access to the anus, rectum, and distal colon in adults (≥22 years of age). The Flex Colorectal Drive also provides accessory channels for compatible flexible instruments used in surgery.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary Medrobotics Flex Robotic System

Submitter's Name:Medrobotics Corporation
Submitter's Address:475 Paramount DriveRaynham, MA 02767
Contact Person:John D. Bonasera
Phone Number:508-692-6460
Fax Number:508-823-1703
Date Prepared:July 5, 2017
Device Trade Name:Medrobotics Flex® Robotic System
Device Common Name:Endoscope/Endoscope and Accessories
Product Code:FDF
Classification:The Medrobotics Flex Robotic System is classified as Class II per 21 CFR §876.1500
Predicate Device:Medrobotics Flex Robotic System (K162330)
Device Description:The Medrobotics Flex Robotic System is an operator controlled flexible scopethat include the benefits of both a rigid scope and a computer assistedcontroller. This allows for the Flex Colorectal Drive to be introduced via anoperator controlled user interface, easily providing transanal access to theanus, rectum and distal colon. Visualization is provided by a user selectable2D or 3D HD camera incorporated in distal end of the scope. The FlexRobotic System's scope also provides accessory channels for the use of variedflexible surgical instruments.
Intended Use:The Medrobotics Flex Robotic System is intended to provide robot-assistedcontrol of the Flex Colorectal Drive during visualization of and surgical siteaccess to the anus, rectum and distal colon. The Flex Robotic System isintended for use in adults (≥22 years of age).The Flex Colorectal Drive is intended for robot-assisted visualization of andsurgical site access to the anus, rectum, and distal colon in adults (≥22 yearsof age). The Flex Colorectal Drive also provides accessory channels forcompatible flexible instruments used in surgery.
Performance Data:The Flex Robotic system has been successfully tested for function,performance, and safety as per FDA recognized Standards: IEC 60601-1,IEC 60601-1-2, IEC 60601-2-18, and Biocompatibility and Toxicity testing ofthe patient contacting materials to ISO 10993-1. It has been tested and metacceptance criteria per FDA recognized standards for the establishment oflabeled shelf life and shipping, and validated for sterility by ETO and Steamto a SAL of 10-6.
Substantial Equivalence:The Medrobotics Flex Robotic System is substantially equivalent to and hasthe intended use as the predicate device.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.