(266 days)
This product is used for temporary internal drainage from the ureteropelvic junction to the bladder. They are not intended to remain indwelling more than twelve months.
The Black Silicone Filiform Double Pigtail Stent Set is a sterile, single-use device. The stents are available with outside diameters of 6.0, 7.0, or 8.5 Fr and specified lengths ranging from 20 to 30 centimeters. The stents are flexible, tubular double pigtail stents constructed of radiopaque silicone rubber. Sideports extend along the body of the stent as well as on the pigtails to provide drainage. The stents are secured in the urinary tract with pigtail loops on the proximal and distal ends. A braided tether for repositioning and removal of the device is located on the proximal pigtail (bladder end) of the stent. Along the stent are graduation marks to provide visualization during stent advancement and placement. The Black Silicone Filiform Double Pigtail Stent Set includes a stent, wire guide, a stent positioner with a lockable fitting, and a T-inserter. The lockable fitting on the proximal end of the stent positioner secures the position of the stent positioner on the wire guide and allows for one-pass placement of the stent. The Black Silicone Filiform Double Pigtail Stent Set is labeled for a 12-month indwell time; however, the tether should be removed after 14 days.
This document describes the Cook Incorporated Black Silicone Filiform Double Pigtail Stent Set (K172017), a ureteral stent. The provided text is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed study report with specific acceptance criteria and performance metrics. Therefore, several requested details, particularly quantitative data from a clinical or standalone performance study, are not available in this document.
Here's a breakdown of the available information:
1. Table of Acceptance Criteria and Reported Device Performance
The document lists various tests performed to assure "reliable design and performance." However, it does not explicitly state quantitative acceptance criteria or detailed numerical results for each test. Instead, it generally concludes that the device passed these tests and functions as intended.
| Test Performed | Acceptance Criteria (Not explicitly stated in document) | Reported Device Performance (General conclusion from document) |
|---|---|---|
| Sterility | (Expected to meet industry standards for sterile medical devices) | "Assure reliable design and performance" |
| Packaging | (Expected to ensure product integrity and sterility) | "Assure reliable design and performance" |
| Biocompatibility | (Expected to meet ISO 10993 standards) | "Assure reliable design and performance" |
| Flow Rate | (Expected to demonstrate adequate fluid drainage) | "Assure reliable design and performance" |
| Retention Strength | (Expected to demonstrate secure placement in the urinary tract) | "Assure reliable design and performance" |
| Tensile Strength | (Expected to demonstrate material integrity under tension) | "Assure reliable design and performance" |
| Curl Restoration | (Expected to demonstrate ability of pigtail to form and maintain shape) | "Assure reliable design and performance" |
| Kink Radius | (Expected to demonstrate resistance to kinking) | "Assure reliable design and performance" |
| Radiopacity | (Expected to be sufficiently visible under fluoroscopy) | "Assure reliable design and performance" |
| MRI Testing | (Expected to demonstrate compatibility with MRI environment) | "Assure reliable design and performance" |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the sample sizes used for the listed tests. It also does not provide any information about data provenance (e.g., country of origin, retrospective or prospective) as these tests appear to be bench or lab-based performance tests, not clinical studies involving patient data.
3. Number of Experts Used to Establish Ground Truth and Qualifications
This information is not applicable to the type of testing described. The tests listed are primarily engineering and material performance assessments, not clinical evaluations requiring expert interpretation of ground truth in the context of diagnostic accuracy or clinical outcomes.
4. Adjudication Method
This information is not applicable. The listed tests are objective, laboratory or bench tests, not subjective assessments requiring adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study is mentioned. This device is a passive medical implant (ureteral stent) and not an AI-powered diagnostic or assistive tool, so such a study would not be relevant.
6. Standalone Performance Study (Algorithm only without human-in-the-loop performance)
The device is a physical medical device, not an algorithm. Therefore, a "standalone algorithm only" performance study is not applicable. The document describes bench and material performance tests as listed in the table, which could be considered "standalone" in the sense that they evaluate the device itself without human intervention beyond conducting the test.
7. Type of Ground Truth Used
The concept of "ground truth" as typically used in AI/diagnostic studies (e.g., pathology, clinical outcomes, expert consensus) is not directly applicable here. For the listed tests, the "ground truth" would be established by the physical and chemical properties of the materials, engineering specifications, and established testing methodologies (e.g., ISO standards for biocompatibility or sterility).
8. Sample Size for the Training Set
This information is not applicable. This device is not an AI-driven system that requires a "training set."
9. How the Ground Truth for the Training Set Was Established
This information is not applicable for the same reason as point 8.
In summary, the provided document is a 510(k) summary for a traditional medical device (a ureteral stent). It focuses on demonstrating substantial equivalence through a comparison to a predicate device and presenting a list of engineering and material performance tests. It does not contain the detailed quantitative data, clinical study results, or AI-specific information that many of the questions pertain to.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
March 26, 2018
Cook Incorporated Karthik Pillai Regulatory Science Specialist 750 Daniels Wav P.O. Box 489 Bloomington, IN 47402
Re: K172017
Trade/Device Name: Black Silicone Filiform Double Pigtail Stent Set Regulation Number: 21 CFR 876.4620 Regulation Name: Ureteral stent Regulatory Class: II Product Code: FAD Dated: February 16, 2018 Received: February 20, 2018
Dear Karthik Pillai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Joyce M. Whang -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
510(k) Number (if known) K172017
Device Name
The Black Silicone Filiform Double Pigtail Ureteral Stent Set
Indications for Use (Describe)
This product is used for temporary internal drainage from the ureteropelvic junction to the bladder. They are not intended to remain indwelling more than twelve months.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is in white, uppercase letters, with a registered trademark symbol next to it. Below "COOK" is the word "MEDICAL" in white, uppercase letters, set against a slightly darker red background.
510(k) Summary
COOK INCORPORATED 750 DANIELS WAY, P.O. BOX 489 BLOOMINGTON, IN 47402-0489 V.COOKM
Black Silicone Filiform Double Pigtail Stent Set 21 CFR §876.4620 Date Prepared: June 30, 2017
Submitted By:
| Submission: | Traditional 510(k) Premarket Notification |
|---|---|
| Applicant: | Cook Incorporated |
| Applicant Address: | Cook Incorporated750 Daniels WayBloomington, IN 47404 |
| Contact: | Karthik Pillai |
| Email: | RegSubmissions@CookMedical.com |
| Contact Phone Number: | 812-335-3575 x104929 |
| Contact Fax Number: | 812-332-0281 |
Device Information:
| Trade Name: | Black Silicone Filiform Double Pigtail Ureteral Stent Set |
|---|---|
| Common Name: | Ureteral Stent |
| Classification Name: | Ureteral Stent |
| Regulation: | 21 CFR §876.4620 |
| Product Code: | FAD |
| Device Class: | II |
| Classification Panel: | Gastroenterology/Urology |
Predicate Device:
The predicate device is Universa™ Firm Ureteral Stents and Stent Sets, cleared under 510(k) K161236 on January 27, 2017.
Device Description:
The Black Silicone Filiform Double Pigtail Stent Set is a sterile, single-use device. The stents are available with outside diameters of 6.0, 7.0, or 8.5 Fr and specified lengths ranging from 20 to 30 centimeters. The stents are flexible, tubular double pigtail stents constructed of radiopaque silicone rubber. Sideports extend along the body of the stent as
{4}------------------------------------------------
well as on the pigtails to provide drainage. The stents are secured in the urinary tract with pigtail loops on the proximal and distal ends. A braided tether for repositioning and removal of the device is located on the proximal pigtail (bladder end) of the stent. Along the stent are graduation marks to provide visualization during stent advancement and placement. The Black Silicone Filiform Double Pigtail Stent Set includes a stent, wire guide, a stent positioner with a lockable fitting, and a T-inserter. The lockable fitting on the proximal end of the stent positioner secures the position of the stent positioner on the wire guide and allows for one-pass placement of the stent. The Black Silicone Filiform Double Pigtail Stent Set is labeled for a 12-month indwell time; however, the tether should be removed after 14 days.
Intended Use:
This product is used for temporary internal drainage from the ureteropelvic junction to the bladder. They are not intended to remain indwelling more than twelve months.
Comparison to Predicate Device:
The Black Silicone Filiform Double Pigtail Ureteral Stent Set is substantially equivalent to the predicate device, the Universa™ Firm Ureteral Stents and Stent Sets, as described in K161236. The table below presents the similarities and differences between the devices for substantial equivalence purposes. The differences between the subject device and the predicate device do not raise any new issues of safety and effectiveness. Performance data are available to support substantial equivalence and were developed using acceptable scientific methods for evaluation.
| Universa™ Firm Ureteral Stent andStent Set (K161236) | Black Silicone Filiform Double PigtailUreteral Stent Set (Subject of thissubmission) | |
|---|---|---|
| Indications for Use | Used for temporary internal drainagefrom the ureteropelvic junction to thebladder. Ureteral stents have been used torelieve obstruction in a variety of benign,malignant, and post-traumatic conditions.The stents may be placed usingendoscopic, percutaneous, or opensurgical techniques. | Used for temporary internal drainagefrom the ureteropelvic junction to thebladder. They are not intended to remainindwelling more than twelve months. |
| Maximum IndwellTime | 12 months | Identical |
| Sterilization | Ethylene oxide | |
| Maximum ShelfLife | 3 years | Identical |
{5}------------------------------------------------
| UniversaTM Firm Ureteral Stent andStent Set (K161236) | Black Silicone Filiform Double PigtailUreteral Stent Set (Subject of thissubmission) | |
|---|---|---|
| Principle OperationComponents | Double pigtail stent Stent positioner Wire guide | Double pigtail stent Stent positioner with lockable fitting Wire guide T-inserter |
| Dimensions | ||
| Loop Diameter | 13 mm - 20 mm | 15 mm - 18 mm |
| Stent OuterDiameter | 5 Fr - 8 Fr | 6 Fr - 8.5 Fr |
| Stent Length | 18 cm - 30 cm | 20 cm - 30 cm |
| Pusher/Positioner | 5 Fr and 7 Fr | 7 Fr, 8 Fr, and 9 Fr |
| Materials | ||
| Stent | Radiopaque Polycarbonate-basedPolyurethane | Radiopaque Silicone Rubber |
| Stent Markings | Black Ink | White Ink |
| Tether | Tevdek® (Polyethylene Terephthalate) | Identical |
| Pusher/Positioner | Polyethylene Non-radiopaque or Vinyl | Identical |
| Pusher/PositionerMarker | Stainless Steel | Stainless Steel or Tungsten |
| Wire Guide | PTFE-coated Stainless Steel | Identical |
Technological Characteristics:
The Black Silicone Filiform Double Pigtail Ureteral Stent Set was subjected to the following tests to assure reliable design and performance under the specified testing parameters:
-
- Sterility
-
- Packaging
-
- Biocompatibility
-
- Flow Rate
-
- Retention Strength
-
- Tensile Strength
-
- Curl Restoration
-
- Kink Radius
-
- Radiopacity
-
- MRI Testing
{6}------------------------------------------------
Conclusion:
The results of the testing provide reasonable assurance that the Black Silicone Filiform Double Pigtail Ureteral Stent Set has been designed such that it will function as intended. The subject device does not raise new questions of safety or effectiveness as compared to the predicate.
§ 876.4620 Ureteral stent.
(a)
Identification. A ureteral stent is a tube-like implanted device that is inserted into the ureter to provide ureteral rigidity and allow the passage of urine. The device may have finger-like protrusions or hooked ends to keep the tube in place. It is used in the treatment of ureteral injuries and ureteral obstruction.(b)
Classification. Class II (performance standards).