K Number
K171939

Validate with FDA (Live)

Date Cleared
2018-03-07

(252 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ALVUE™ Balloon Dilation System is indicated for patients undergoing surgical procedures requiring tissue retraction including the following procedures: endoscopic; laparoscopic; general surgery; plastic and reconstructive surgery; spine surgery; orthopedic surgery; thoracoscopic surgery; and procedures in the extraperitoneal space. The system is intended to create an operative space by dissecting layers of connective tissue along natural tissue planes of separation of the extraperitoneal, subcutaneous extremity, or thoracic space.

Device Description

The ALVUE™ Balloon Dilation System is a surgical instrument that consists of an inflatable nylon balloon attached to the distal end of a dilator. ALVUE™ is used to dilate soft tissue to gain access to the surgical site and is available in a variety of sizes to accommodate the anatomical needs of the patient. The system includes accessories for inflation. ALVUE™ is a sterile single use device.

AI/ML Overview

This device is a physical medical device (ALVUE™ Balloon Dilation System) and not an AI/ML powered device, so acceptance criteria and studies are based on mechanical and biocompatibility testing rather than AI model performance metrics. Therefore, many of the requested fields are not applicable.

Here's the relevant information from the provided document:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria CategorySpecific Test/StandardReported Device Performance/Compliance
Mechanical TestingInflation Volume and Pressure (in accordance with ISO 10555-1:2013 and ISO 10555-4:2013)Performance data demonstrates substantial equivalence to the predicate. (Quantitative values not provided in this summary.)
Axial Tension (in accordance with ISO 10555-1:2013 and ISO 10555-4:2013)Performance data demonstrates substantial equivalence to the predicate. (Quantitative values not provided in this summary.)
Fatigue (in accordance with ISO 10555-1:2013 and ISO 10555-4:2013)Performance data demonstrates substantial equivalence to the predicate. (Quantitative values not provided in this summary.)
Biocompatibility TestingAccording to ISO 10993-1 on patient contacting materialsPerformed (Result is implied compliance for substantial equivalence, specific results not detailed in this summary.)
Technological CharacteristicsDesign, intended use, material composition, and function compared to predicate devices.Similar technological characteristics to predicate devices. Consists of cylindrical balloons that dilate soft tissue. Uses fluid to inflate (predicates use fluid or gas).
Intended UseFor patients undergoing surgical procedures requiring tissue retraction (endoscopic; laparoscopic; general surgery; plastic and reconstructive surgery; spine surgery; orthopedic surgery; thoracoscopic surgery; and procedures in the extraperitoneal space)The ALVUE™ Balloon Dilation System is indicated for these procedures. (Note: ALVUE™, unlike the predicates, is not indicated for vascular surgery nor laparoscopic surgery requiring a sealed port of access.)

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for Test Set: Not specified in the provided summary.
  • Data Provenance: Not specified in the provided summary (as these are physical device tests, not clinical data from patients).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

  • Not Applicable. This is not an AI/ML device relying on expert-established ground truth from clinical data. The "ground truth" for mechanical testing is compliance with established engineering standards (ISO 10555-1, ISO 10555-4) and biocompatibility standards (ISO 10993-1).

4. Adjudication Method:

  • Not Applicable. This is not an AI/ML device.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No. This is not an AI/ML device.

6. Standalone (Algorithm Only) Performance Study:

  • Not Applicable. This is a physical medical device, not an algorithm.

7. Type of Ground Truth Used:

  • For mechanical testing: Engineering standards (ISO 10555-1:2013, ISO 10555-4:2013).
  • For biocompatibility: Biocompatibility standards (ISO 10993-1).

8. Sample Size for the Training Set:

  • Not Applicable. This is a physical medical device, not an AI/ML model that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

  • Not Applicable. This is a physical medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 7, 2018

Globus Medical, Inc. Lori Burns Director, Regulatory Affairs 2560 General Armistead Ave. Valley Forge Business Center Audubon, Pennsylvania 19403

Re: K171939

Trade/Device Name: ALVUE Balloon Dilation System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: January 22, 2018 Received: January 23, 2018

Dear Lori Burns:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely. Jennifer R. Stevenson -S3 For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices

Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K171939

Device Name

ALVUE™ Balloon Dilation System

Indications for Use (Describe)

The ALVUE™ Balloon Dilation System is indicated for patients undergoing surgical procedures requiring tissue retraction including the following procedures: endoscopic; general surgery; plastic and reconstructive surgery; spine surgery; orthopedic surgery; thoracoscopic surgery; and procedures in the extraperitoneal space. The system is intended to create an operative space by dissecting layers of connective tissue planes of separation of the extraperitoneal, subcutaneous extremity, or thoracic space.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

510(k) Summary: ALVUE™ Balloon Dilation

Company:Globus Medical Inc.2560 General Armistead Ave.Audubon, PA 19403610-930-1800
Primary Contact:Lori BurnsDirector, Regulatory Affairs
Alternate Contact:Kelly J. Baker, Ph.D.Senior Vice President, Regulatory and Clinical Affairs
Date Prepared:January 22, 2018
Device Name:ALVUE™ Balloon Dilation System
Common Name:Surgical Balloon Dilator
Classification:Per 21 CFR as follows:§876.1500 Endoscope and accessoriesProduct Code(s): GCJRegulatory Class: II, Panel Code: 87
Primary Predicate:Pajunk Disposable Balloons and Balloon Systems (K090631)
Predicates:Spacemaker Surgical Balloon Dissector (K973046)
ReferenceDevices:AFFIRM™ VCF System (K110998)CRE Balloon Dilatation Catheter (K110833)Gyrus ACMI Nephro EZDilate Nephrostomy Balloon DilationCatheter (K132383)Gyrus ACMI EZDilate 3-Stage Balloon Dilatation Catheters(K143609)

Purpose:

The purpose of this submission is to request clearance for the ALVUE™ Balloon Dilation System.

Device Description:

The ALVUE™ Balloon Dilation System is a surgical instrument that consists of an inflatable nylon balloon attached to the distal end of a dilator. ALVUE™ is used to dilate soft tissue to gain access to the surgical site and is available in a variety of sizes to accommodate the anatomical needs of the patient. The system includes accessories for inflation. ALVUE™ is a sterile single use device.

{4}------------------------------------------------

Indications for Use:

The ALVUE™ Balloon Dilation System is indicated for patients undergoing surgical procedures requiring tissue retraction including the following procedures: endoscopic; laparoscopic; general surgery; plastic and reconstructive surgery; spine surgery; orthopedic surgery; thoracoscopic surgery; and procedures in the extraperitoneal space. The system is intended to create an operative space by dissecting layers of connective tissue along natural tissue planes of separation of the extraperitoneal, subcutaneous extremity, or thoracic space.

Performance Data:

Mechanical testing (inflation volume and pressure, axial tension, and fatigue) was conducted in accordance with ISO 10555-1:2013, Sterile, Single-Use Intravascular Catheters - Part 1: General Requirements and ISO 10555-4:2013, Sterile, Single-Use Intravascular Catheters – Part 4: Balloon Dilation Catheters. Performance data demonstrates substantial equivalence to the predicate. Biocompatibility testing was performed in accordance to ISO 10993-1 on patient contacting materials.

Technological Characteristics:

The ALVUE™ Balloon Dilation System has similar technological characteristics as the predicate devices including design, intended use, material composition, and function. The subject and predicate devices consist of cylindrical balloons which function to dilate soft tissue. The subject device uses fluid to inflate the balloon while the predicates use either fluid or gas. These systems are used to create an operative space by dissecting layers of connective tissue along natural planes of separation of extraperitoneal, subcutaneous extremity, or thoracic space. ALVUE™, unlike the predicates, is not indicated for vascular surgery nor laparoscopic surgery requiring a sealed port of access.

Basis of Substantial Equivalence:

ALVUE™ Balloon Dilation System have been found to be substantially equivalent to the predicate devices with respect to technical characteristics, performance, and intended use. The information provided within this premarket notification supports substantial equivalence of the subject instruments to the predicate devices.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.