K Number
K171844

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-07-18

(28 days)

Product Code
Regulation Number
872.5470
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Inspire ICE, a ceramic orthodontic bracket, is intended for use during orthodontic treatment of malocclusions. A malocclusion is imperfect positioning of the teeth when the jaws are closed.

Device Description

Inspire ICE, is a single-use orthodontic bracket made from monocrystalline alumina. The brackets are intended for use during orthodontic treatment of malocclusions. A malocclusion is imperfect positioning of the teeth when the jaws are closed. After following standard office protocols for tooth preparation, the brackets are bonded to the teeth with an orthodontic adhesive or cement. An archwire is then threaded through the bracket's archwire slot and is held in place with a ligature tie. These ligatures are tightened, or ligated, around the 'wings' of the bracket and over the archwire. To aid in optimal teeth movement, the individual brackets can also be ligated to other brackets or other orthodontic devices. After the teeth have moved into their proper positions, the orthodontic / dental professional will remove the brackets with the Inspire ICE Debonding Pliers. This step is also known as debonding.

AI/ML Overview

This document is a 510(k) premarket notification for the Inspire ICE™ aka Inspire ICE™ Clear Brackets, an orthodontic plastic bracket. The document details the device's characteristics and its substantial equivalence to predicate devices.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied by the non-clinical performance data and the comparison to predicate devices, demonstrating the device performs "as safe, as effective, and performs as well as the predicate devices". Specific quantitative acceptance criteria or pass/fail thresholds for each test are not explicitly stated in this summary. The table below summarizes the reported non-clinical performance data categories.

Test ElementAcceptance Criteria (Implied)Reported Device Performance (Inspire ICE)
Bond StrengthDemonstrates comparable bond strength to predicate devices and meets relevant standards.Testing results performed. Implied to be acceptable based on the conclusion of substantial equivalence.
Tie Wing StrengthDemonstrates comparable tie wing strength to predicate devices and meets relevant standards.Testing results performed. Implied to be acceptable based on the conclusion of substantial equivalence.
Torque StrengthDemonstrates comparable torque strength to predicate devices and meets relevant standards.Testing results performed. Implied to be acceptable based on the conclusion of substantial equivalence.
BiocompatibilityMeets the requirements of ISO 10993 series for biological evaluation of medical devices.Testing results performed, utilizing relevant ISO 10993 standards. The finished device is considered to have met biological evaluation and risk assessment requirements.
Stability TestingDemonstrates stability over time under specified conditions.Testing results performed. Implied to be acceptable based on the conclusion of substantial equivalence.
Material (Alumina)Material composition similar to predicate, ensuring equivalent performance and safety.Made from monocrystalline alumina, identical to predicate and predicate reference materials.
Physical DimensionsConformity to established ranges/values for orthodontic brackets (In/Out, Torque, Angulation, Rotation).In/Out: .035" - .050" Torque: +17° thru -22° Angulation: Up to +13° Rotation: Up to 4° These are within or comparable to the ranges of predicate devices.
Manufacturing Method (Grinding)Manufacturing method consistent with predicate, ensuring similar structural integrity.Grinding, identical to predicate and predicate reference manufacturing methods.
Ligation (Non Self-Ligating)Ligation mechanism consistent with predicate, ensuring proper orthodontic function.Non Self-Ligating, identical to predicate and predicate reference ligation methods.
Debonding PerformanceCompatible with a specific debonding instrument, ensuring safe removal.Debonding Pliers (REF: 866-4020) are used, similar to the predicate reference and comparable/compatible with predicate device debonding.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample sizes used for the non-clinical performance testing (Bond Strength, Tie Wing Strength, Torque Strength, Biocompatibility, Stability testing). It mentions that "internal test methods" were utilized alongside applicable ISO standards.

The data provenance is from non-clinical performance data (laboratory testing) conducted by the manufacturer, Ormco Corporation, or accredited research service companies. The country of origin of the data is not explicitly stated, but the submitter (Ormco Corporation) is based in Orange, California, USA, suggesting the testing was likely conducted in the USA or by labs adhering to international standards (ISO). The testing is prospective for the Inspire ICE device, comparing its performance to already marketed predicate devices.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This document describes non-clinical performance testing and a comparison to predicate devices for substantial equivalence. It does not involve a "test set" in the context of human interpretation or diagnostic accuracy where expert ground truth would be established by panels of human experts. Instead, the "ground truth" for material properties and mechanical performance would be derived from:

  • Established international standards (e.g., ISO 27020, ISO 10993 series).
  • Validated laboratory testing methodologies.

Therefore, the concept of "number of experts" or their qualifications for establishing ground truth as typically understood in studies involving diagnostic accuracy is not directly applicable here. The experts involved would be engineers, material scientists, and toxicologists familiar with the respective ISO standards and testing procedures.

4. Adjudication Method for the Test Set

As this study focuses on non-clinical performance data and comparison to predicate devices, there is no "adjudication method" in the context of resolving discrepancies in expert interpretations (e.g., 2+1, 3+1). The performance is assessed against established technical specifications and international standards.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was done. This device is an orthodontic bracket, a physical medical device, not a software algorithm that assists human readers.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a physical orthodontic bracket, not an algorithm.

7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

The "ground truth" for this submission is based on:

  • Established performance specifications and standards: Primarily ISO 27020 for orthodontics brackets and tubes, and the ISO 10993 series for biological evaluation of medical devices.
  • Validated laboratory testing methodologies: Used to measure Bond Strength, Tie Wing Strength, Torque Strength, Biocompatibility, and Stability.
  • Comparison to legally marketed predicate devices: The "ground truth" for demonstrating safety and effectiveness relies on showing that the new device performs equivalently to previously cleared devices which have an established history of safe and effective use.

8. The Sample Size for the Training Set

Not applicable. This device is a physical orthodontic bracket and does not involve AI or machine learning models requiring a "training set."

9. How the Ground Truth for the Training Set was Established

Not applicable, as no training set is involved.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human profile, composed of three overlapping shapes.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 18, 2017

Ormco Corporation c/o Mr. Dave Yungvirt Third Party Review Group, LLC The Old Station House, 24 Lackawanna Place Millburn, New Jersey 07041

Re: K171844

Trade/Device Name: Inspire ICETM aka Inspire ICETM Clear Brackets Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic plastic bracket Regulatory Class: Class II Product Code: NJM Dated: June 18, 2017 Received: June 20, 2017

Dear Mr. Dave Yungvirt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Mary S. Runner -A

Lori A. Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Change Control Table, Change History

Change Control Table

VersionDocument AuthorDocument ApproverDate Approved
1.00Name, Title, OfficeName, Title, OfficeMM/DD/YYYY

Complete Change Control Table (all versions) retained in SWIFT Docs.

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Food and Drug Administration
Indications for Use
Expiration Date: January 31, 2017See PRA Statement below.
510(k) Number (if known)TBD
Device NameInspire ICE™, aka Inspire ICE™ Clear Braces
Indications for Use (Describe)Inspire ICE, a ceramic orthodontic bracket, is intended foruse during orthodontic treatment of malocclusions. Amalocclusion is imperfect positioning of the teeth whenthe jaws are closed.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human ServicesFood and Drug AdministrationOffice of Chief Information OfficerPaperwork Reduction Act (PRA) StaffPRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

SECTION 5. 510(k) SUMMARY K171844

For Inspire ICE (Ceramic Orthodontic Brackets)

This 510(k) summary is submitted in accordance with the requirements of 21 CFR 807.92.

1.510(k) SubmitterOrmco Corporation1717 W. Collins AvenueOrange, California 92867
2.ContactTara BonnyRegulatory AffairsPhone: 909-962-5856Fax: 909-962-5694Email: Tara.Bonny@kavokerr.com
3.Additional ContactWendy GarmanVice President, Regulatory AffairsPhone: 909-962-5666Fax: 909-962-5694Email: Wendy.Garman@kavokerr.com
4.Date PreparedMay 15, 2017
5.Device
Proprietary NameInspire ICE™, aka Inspire ICE™ Clear Braces
Classification NameBracket, Ceramic Dental
Regulation Number21 CFR 872.5470
Product CodeNJM
Medical Specialty Panel76 Dental
ClassificationMedical Device, Class II

6. Predicate Device

Inspire ICE is substantially equivalent to American Orthodontics', Radiance Plus® brackets, K080749, product code NJM, cleared on August 4, 2008.

7. Predicate Reference Device

The Inspire ICE orthodontic ceramic bracket is the next generation of the original Vanish Orthodontic Bracket, K844067, cleared on November 23, 1984. The Vanish bracket is included in this 510(k) submission as a 'predicate reference'. Although the Inspire ICE bracket is substantially equivalent to the Vanish bracket, Vanish could not be declared as the predicate because it was cleared under Product Code EJF, Metal Orthodontic Brackets.

{5}------------------------------------------------

8. Device Description

Inspire ICE, is a single-use orthodontic bracket made from monocrystalline alumina. The brackets are intended for use during orthodontic treatment of malocclusions. A malocclusion is imperfect positioning of the teeth when the jaws are closed. After following standard office protocols for tooth preparation, the brackets are bonded to the teeth with an orthodontic adhesive or cement. An archwire is then threaded through the bracket's archwire slot and is held in place with a ligature tie. These ligatures are tightened, or ligated, around the 'wings' of the bracket and over the archwire. To aid in optimal teeth movement, the individual brackets can also be ligated to other brackets or other orthodontic devices. After the teeth have moved into their proper positions, the orthodontic / dental professional will remove the brackets with the Inspire ICE Debonding Pliers. This step is also known as debonding.

Accessory Used withInspire ICEManufacturer of AccessoryPremarketNotification
Inspire ICE Debonding PliersOrmco1332 S. Lone Hill AvenueGlendora, CA 91741 USAExempt from premarketnotification requirements

9. Indications for Use

Inspire ICE, an orthodontic ceramic bracket, is intended for use during orthodontic treatment of malocclusions. A malocclusion is imperfect positioning of the teeth when the jaws are closed.

10. Description of Safety and Substantial Equivalence

Technological Characteristics

The designs of Inspire ICE are similar to the predicate American Orthodontics', Radiance Plus® (K080749), Product Code NJM, cleared on August 4, 2008 and to the reference predicate Vanish Orthodontic Bracket, K844067, Product Code EJF, cleared on November 23, 1984. All three are clear ceramic twin brackets, created from a single crystal of pure grown sapphire.

Non-Clinical Performance Data

Non-clinical performance data included testing results for Bond Strength. Tie Wing Strength and Torque Strength, as well as Biocompatibility and Stability testing. Along with internal test methods, the following applicable standards were utilized for the nonclinical performance testing:

{6}------------------------------------------------

  • . ISO 27020 Dentistry - Brackets and Tubes for use in Orthodontics
  • . ISO 10993-1:2009 Biological evaluation of medical devices Part 1: Evaluation and testing in the risk management process
  • ISO 10993-2:2006 Biological evaluation of medical devices Part 2: Animal . welfare requirements
  • . ISO 10993-3:2014 Biological evaluation of medical devices Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
  • . ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity
  • . ISO 10993-6:2007 Biological evaluation of medical devices Part 6: Tests for local effects after implantation
  • ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for . irritation and skin sensitization
  • . ISO 10993-11:2006 Biological evaluation of medical devices Part 11: Tests for systemic toxicity
  • . ISO 10993-12:2012 Biological evaluation of medical devices Part 12: Sample preparation and reference materials (available in English only)
  • ISO 10993-17:2002 Biological evaluation of medical devices Part 17: . Establishment of allowable limits for leachable substances
  • . ISO 10993-18:2005 Biological evaluation of medical devices Part 18: Chemical characterization of materials
  • . ISO 14971:2007 Medical Devices – Application of risk management to medical devices, Annex 1: Guidance on Risk Analysis Procedures for Biological Hazards
ElementRadiance Plus(Predicate Device)Vanish(Predicate Reference)Inspire ICE(Proposed Device)
510(k)K080749K844067To be assigned
Trade NameRadiance PlusVanish OrthodonticBracketInspire ICE,aka Inspire ICE ClearBraces
Target UsersLicensed Orthodontic /Dental ProfessionalsLicensed Orthodontic /Dental ProfessionalsLicensed Orthodontic /Dental Professionals
DeviceDescriptionRadiance Plus is asingle-usemonocrystallineVanish is a single-usemonocrystallineInspire ICE is a single-use monocrystalline
ElementRadiance Plus(Predicate Device)Vanish(Predicate Reference)Inspire ICE(Proposed Device)
sapphire orthodontictwin bracket.sapphire orthodontictwin bracket.sapphire orthodontictwin bracket.
Indications forUseOrthodontic brackets areprescribed for patientswith teeth that are notnormally positioned inthe mouth. Orthodontictreatment is used tocorrect dentaldeficiencies and toimprove the appearanceof the patient.Orthodontic bracket isintended for use byorthodontists to assist inrestoring the naturalocclusion of teeth.Intended for use duringorthodontic treatment ofmalocclusions. Amalocclusion is theimperfect positioning ofthe teeth when the jawsare closed.
Common NameOrthodontic CeramicBracketOrthodontic MetalBracketOrthodontic CeramicBracket
ClassificationNameBracket, Ceramic,DentalBracket, Metal,OrthodonticBracket, Ceramic,Dental
ClassIIIII
Product CodeNJMEJFNJM
MaterialAlumina (Al2O3)Alumina (Al2O3)Alumina (Al2O3)
ClarityTransparentTransparentTransparent
Bracket Design1-Piece Base andBracket Construction1-Piece Base andBracket Construction1-Piece Base andBracket Construction
LigationNon Self-LigatingNon Self-LigatingNon Self-Ligating
ManufacturingMethodGrindingGrindingGrinding
HooksMachined HooksMachined HooksMachined Hooks
In/OutUnknown.035" - . 059".035" - .050"
Torque+17° thru -22°+17° thru -22°+17° thru -22°
AngulationUp to +10°Up to +13°Up to +13°
RotationUp to 4°Up to 4°Up to 4°
ElementRadiance Plus(Predicate Device)Vanish(Predicate Reference)Inspire ICE(Proposed Device)
BracketIdentificationand PlacementColor Coded VisualPlacement Aids (VPAs)Colored dot to indicateUpper/Lower, Left/RightWater-Soluble FacePaint™ IdentificationSystemColored dot to indicateUpper/Lower, Left/RightWater-Soluble FacePaint™ IdentificationSystemColored dot to indicateUpper/Lower, Left/Right
MaterialComposition forBracketIdentificationand PlacementUnknownSupplier's proprietaryRed, Purple, Green,Yellow, Blue, Orange orBlack water-solubletemporary ink/colorant*.*Per an independentaccredited medicalresearch service company,the finished device isconsidered to have metbiological evaluation andrisk assessmentrequirements.Supplier's proprietaryRed, Purple, Green,Yellow, Blue, Orange orBlack water-solubletemporary ink/colorant*.*Per an independentaccredited medicalresearch service company,the finished device isconsidered to have metbiological evaluation andrisk assessmentrequirements.
Bracket BaseQuad Matte™ base thatis intended to providevariable bond strengthacross different locationsalong the baseBall-base technologythat is evenly distributedand fused onto thebottom of the bracketwhich provides optimalbond strength betweenthe bracket and toothBall-base technologythat is evenly distributedand fused onto thebottom of the bracketwhich provides optimalbond strength betweenthe bracket and tooth
Single UseYesYesYes
Non-SterilePackagingYesYesYes
Debonding HandInstrumentAccessoryRadiance PlusDebonding Pliers(REF: 001-343E) orOmega CeramicBracket DebondingPliers (Sushi pliersREF: 001-301E)ICE Debonding Plier(REF: 866-4020)ICE Debonding Plier(REF: 866-4020)

Predicate and Proposed Device Comparison

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

Clinical Performance Data

Clinical performance testing has not been performed for Inspire ICE.

Conclusion as to Substantial Equivalence

Although there are semantic differences between the indications for use of Radiance Plus brackets (predicate device), Vanish brackets (predicate reference) and Inspire ICE brackets (proposed device), these differences do not change the indications. Therefore, the indication of use for Inspire ICE brackets is equivalent to the predicate device Radiance Plus (K080749) and predicate reference Vanish (K844067).

The Radiance Plus (predicate device), Vanish (predicate reference) and Inspire ICE bracket systems (proposed device) all offer a hand instrument accessory designed for bracket debonding.

The technological characteristics of Inspire ICE brackets are very similar to the predicate device Radiance Plus brackets (K080749) and the predicate reference Vanish brackets (K844067). Results of the nonclinical testing, demonstrate that Inspire ICE is as safe, as effective, and performs as well as the predicate devices in terms of the design, intended use, indications for use, manufacturing method, mechanical properties and biocompatibility. The information provided herein, demonstrates that Inspire ICE brackets are substantially equivalent to the legally marketed predicate device Radiance Plus brackets (K080749) and predicate reference Vanish brackets (K844067).

§ 872.5470 Orthodontic plastic bracket.

(a)
Identification. An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.(b)
Classification. Class II.