K Number
K171829

Validate with FDA (Live)

Device Name
EndoNaut
Manufacturer
Date Cleared
2017-09-21

(93 days)

Product Code
Regulation Number
892.1650
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EndoNaut provides image guidance by overlaying preoperative 3D vessel anatomy onto live fluoroscopic images in order to assist in the positioning of the guidewires, catheters and other endovascular devices.

EndoNaut is intended to assist endovascular procedures in the thorax, abdomen, neck, pelvis and lower limbs. Suitable procedures include (but not limited to) endovascular aortic aneurysm repair (AAA and TAA), angioplasty, stenting and embolization in iliac arteries and corresponding veins.

EndoNaut is not intended for use in the X-ray guided procedures in the liver, kidneys or pelvic organs.

Device Description

EndoNaut is a stand-alone software medical device that runs on a Windows based computer that meets the minimum requirements.

EndoNaut software provides navigation tools for image-guided endovascular surgery. The device enables to register the X-ray intra-operative images and the pre-operative data.

The device is operated by the physician or a trained operator. The client machine captures a live fluoroscopy video feed from the X-ray's machine external video port, either in digital or in analog format. Any machine that meets the hardware and software requirements can be supported.

The EndoNaut will be marketed as a software only solution.

AI/ML Overview

Below is a summary of the acceptance criteria and study information for the EndoNaut device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Performance MetricAcceptance CriteriaReported Device Performance
Registration AccuracyRegistration error less than 3mmResults meet the acceptance criteria
Panorama Creation ErrorMaximum acceptable error of 10mm for peripheral artery surgeryMean errors below the acceptance criteria
Measuring Functions (Distance)No error for distance measurement along centerline (compared to planning tool)Results meet the acceptance criteria
Measuring Functions (Length on Image)Max error 1mm for length measurements on images (compared to visible rule)Results meet the acceptance criteria
Software ComplianceCompliance with ISO 14971, IEC 62304, IEC 62366, and FDA Guidance for Software Contained in Medical DevicesSoftware verification and validation testing performed, demonstrating compliance with these standards and usability testing with clinical users.
Clinical FeasibilityFeasibility of fusion imaging during aortic endovascular procedures (Primary Endpoint)Clinical study conclusion confirms the device is safe and effective and supports the indications for use, implying feasibility was achieved. (Specific percentage not provided)
Clinical EfficiencyEvaluation of efficiency in deploying infrarenal aortic stent grafts to treat unruptured atheromatous aneurysms (Secondary Endpoint: radiation dose)Clinical study conclusion confirms the device is safe and effective and supports the indications for use. (Specific improvement in radiation dose not provided, but implies efficiency was found acceptable)

2. Sample Sizes and Data Provenance

  • Test Set (for Technical Performance - Registration Accuracy):

    • Sample Size: 5000 different registrations on 100 pre-operative CT-scan images from 50 patients.
    • Data Provenance: Not explicitly stated (e.g., country of origin). Appears to be retrospective, using existing CT-scan images.
  • Test Set (for Technical Performance - Panorama Creation):

    • Sample Size: 7 cases (6 patients and 1 phantom).
    • Data Provenance: Not explicitly stated. The use of "in-vivo data" suggests patient data, likely retrospective or collected specifically for this purpose.
  • Test Set (for Clinical Study):

    • Sample Size: Not explicitly stated, though it is described as a "single-centre, prospective feasibility pilot study."
    • Data Provenance: Prospective, single-center study. Location not specified.

3. Number of Experts and Qualifications for Ground Truth (Test Set)

  • Technical Performance (Registration Accuracy): Not applicable. The ground truth was a "gold standard transformation matrix," which suggests a numerically derived or calculated reference, not directly established by human experts in this context.
  • Technical Performance (Panorama Creation): One "perfect panorama (manually corrected)" was used as ground truth. No specific number or qualifications of the individual who manually corrected it are provided.
  • Clinical Study: Not applicable for establishing ground truth of the device's performance. The clinical endpoints (feasibility, radiation dose) are objective measures.

4. Adjudication Method for the Test Set

  • Technical Performance: Not applicable as the ground truth was either a numerical matrix or a manually corrected panorama, without a listed multi-reader adjudication process.
  • Clinical Study: No explicit adjudication method for clinical endpoints (e.g., disagreement resolution for radiation dose measurements) is described.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC comparative effectiveness study is mentioned, nor is there any information about the effect size of human readers improving with AI vs. without AI assistance. The clinical study focused on feasibility and efficiency of the device itself.

6. Standalone Performance

  • Yes, performance evaluations were conducted for the algorithm in a standalone manner. The technical performance tests (registration accuracy, panorama creation, measuring functions) demonstrate the algorithm's capabilities without direct human interaction beyond setting up the test scenarios. The device itself is described as a "stand-alone software medical device."

7. Type of Ground Truth Used

  • Technical Performance (Registration Accuracy): "Gold standard transformation matrix."
  • Technical Performance (Panorama Creation): "Perfect panorama (manually corrected)."
  • Technical Performance (Measuring Functions): Comparison with measurements from "EndoSize (planning tool)" and "a visible rule on the fluoroscopic image."
  • Clinical Study: Clinical endpoints such as "feasibility rate of fusion" and "radiation dose as measured by fluoroscopy time, dose-area product and air kerma" were used as measures of outcome.

8. Sample Size for the Training Set

  • The document does not explicitly state the sample size for any training set. The performance data section describes testing on CT-scan images and patient cases but does not distinguish between training and test data or provide details on the training methodology.

9. How the Ground Truth for the Training Set Was Established

  • Since no information about a distinctive training set or its sample size is provided, there is no description of how ground truth for a training set was established.

{0}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 21, 2017

Therenva SAS % Mr. Matthis Hamy QA & RA Manager 4 rue Jean Jaures Rennes. 35000 FRANCE

Re: K171829

Trade/Device Name: EndoNaut Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB Dated: August 25, 2017 Received: August 31, 2017

Dear Mr. Hamy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-

Image /page/0/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Michael D. O'Hara For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K171829

Device Name EndoNaut

Indications for Use (Describe)

EndoNaut provides image guidance by overlaying preoperative 3D vessel anatomy onto live fluoroscopic images in order to assist in the positioning of the guidewires, catheters and other endovascular devices.

EndoNaut is intended to assist endovascular procedures in the thorax, abdomen, neck, pelvis and lower limbs. Suitable procedures include (but not limited to) endovascular aortic aneurysm repair (AAA and TAA), angioplasty, stenting and embolization in iliac arteries and corresponding veins.

EndoNaut is not intended for use in the X-ray guided procedures in the liver, kidneys or pelvic organs.

Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5 - EndoNaut 510(K) Summary

This 510(k) summary is submitted in accordance with the requirements of 21 CFR Part 807.92(c)

Therenva SAS hereby submits this traditional 510(k) to provide a notification submission for a new device

1. Submitter information

Manufacturer Name:Therenva SAS4, rue Jean Jaurès35000 RennesFrance
Contact Person:Matthis Hamy, QA & RA ManagerPhone: +33 6 06 86 16 86E-mail: matthis.hamy@therenva.com
Establishment Registration Number:3011240766
Date prepared:05/31/2017
Device Identification
Device Trade Name:EndoNaut
Device Common Name:Interventional Fluoroscopic X-ray System
Regulation Class:Class II (21 CFR 892.1650, OWB)
Classification Name:Interventional Fluoroscopic X-ray System
Classification Panel:Radiology

3. Predicate Device Identification

Device Classification Name:Interventional Fluoroscopic X-ray System
Regulation Number:892.1650
510(k) Number:K160088
Device Name:Cydar EV
Product Code:OWB, Interventional Fluoroscopic X-ray System

{4}------------------------------------------------

Decision: SUBSTANTIALLY EQUIVALENT (SE)

4. Device Description

EndoNaut is a stand-alone software medical device that runs on a Windows based computer that meets the minimum requirements.

EndoNaut software provides navigation tools for image-guided endovascular surgery. The device enables to register the X-ray intra-operative images and the pre-operative data.

The device is operated by the physician or a trained operator. The client machine captures a live fluoroscopy video feed from the X-ray's machine external video port, either in digital or in analog format. Any machine that meets the hardware and software requirements can be supported.

The EndoNaut will be marketed as a software only solution.

5. Indications for Use

EndoNaut provides image guidance by overlaying preoperative 3D vessel anatomy onto live fluoroscopic images in order to assist in the positioning of the guidewires, catheters and other endovascular devices.

EndoNaut is intended to assist endovascular procedures in the thorax, abdomen, neck, pelvis and lower limbs. Suitable procedures include (but not limited to) endovascular aortic aneurysm repair (AAA and TAA), angioplasty, stenting and embolization in iliac arteries and corresponding veins.

EndoNaut is not intended for use in the X-ray guided procedures in the liver, kidneys or pelvic organs.

6. Predicate Device Comparison

EndoNaut has similar indications for use, principle of operation and technical characteristics than the predicate device listed above. The differences between the devices do not raise any question with respect to the safety and effectiveness of the subject device.

ManufacturerTherenva SASCydar LtdDifferences
Device NameEndoNaut(Subject Device)Cydar EV(Predicate Device)K160088
Product CodeOWBOWBIdentical
RegulationNumber892.1650892.1650Identical
Regulation NameInterventional FluoroscopicX-Ray SystemInterventional FluoroscopicX-Ray SystemIdentical
Indications forUseEndoNaut provides imageguidance by overlayingpreoperative 3D vesselanatomy onto liveCydar EV provides imageguidance by overlayingpreoperative 3D vesselanatomy, from a previouslySimilar. Bothdevices are imageguidance systemsfor fluoroscopy-
ManufacturerTherenva SASCydar LtdDifferences
Device NameEndoNaut(Subject Device)Cydar EV(Predicate Device)K160088
fluoroscopic images in order to assist in the positioning of the guidewires, catheters and other endovascular devices.EndoNaut is intended to assist the treatment of endovascular diseases during procedures such as (but not limited to) AAA, TAA, carotid stenting, iliac interventions.EndoNaut is not intended for use in the X-ray guided procedures in the liver, kidneys or pelvic organs.acquired contrast-enhanced, diagnostic CT scan, onto live fluoroscopic images in order to assist in the positioning of guidewires, catheters and other endovascular devices.Cydar EV is intended to assist fluoroscopy-guided endovascular procedures in the lower thorax, abdomen and pelvis. Suitable procedures include (but are not limited to) endovascular aortic aneurysm repair (AAA and mid-distal TAA), angioplasty, stenting and embolization in the common iliac, proximal external iliac and proximal internal iliac arteries and corresponding veins.Cydar EV is not intended for use in the X-ray guided procedures in the liver, kidneys or pelvic organs.contrast-guided endovascular procedures.
HardwareSoftware only product; runs on a separate interventional tools workstation.Software only product; runs on a separate interventional tools workstation.Identical
RegistrationOverview2D-3D registration is achieved by manual initialization and machine vision tracking of vertebral anatomy2D-3D registration is achieved by machine vision tracking of vertebral anatomySimilar. The manual initialization in EndoNaut allows securing and improving the registration algorithm. 2D-3D registration is performed the same with no questions raised
ManufacturerTherenva SASCydar LtdDifferences
Device NameEndoNaut(Subject Device)Cydar EV(Predicate Device)K160088for safety andeffectiveness.
Registration targetVertebral anatomyVertebral anatomyIdentical
Patient contactingNoNoIdentical
Energy emitted or absorbedNoNoIdentical
Dynamic update on C-arm / table / patient motionAutomatic motion detection; registration is updated manually.AutomaticDifferent. In the subject device, the registration is updated through a semi-automatic workflow with no additional questions raised for safety and effectiveness.
Anatomical LocationVascular anatomy of the chest, abdomen, pelvis and lower limbs.Vascular anatomy of the chest, abdomen and pelvis.Similar. The subject device includes guidance tools for lower limbs similar than for aortic region, with no additional questions raised for safety and effectiveness.
Ability to store roadmapsYes, in Lower Limbs module onlyYesSimilar. Ability to store roadmaps does not impact the intended use and does not raise additional questions for safety and effectiveness
Ability to store snapshotsYesYesIdentical
ManufacturerTherenva SASCydar LtdDifferences
Device NameEndoNaut(Subject Device)Cydar EV(Predicate Device)K160088
IEC 62304AppliedAppliedIdentical
IEC 62366AppliedAppliedIdentical
ISO 14971AppliedAppliedIdentical
NEMA PS 3.1-3.20DICOMNoAppliedDifferent. Thesubject devicedoes not connectto the PACS.Input patient dataare providedthrough EndoSizesoftware(K160376),DICOMcompliant, withno additionalquestion raisedfor safety andeffectiveness

Comparison table:

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

7. Performance Data

Non-clinical performance testing has been performed and demonstrates compliance with following standards:

  • ISO 14971 – Medical Devices – Application of risk management to medical devices
  • IEC 62304 – Medical Devices software – Software life-cycle processes
  • IEC 62366 Medical Devices Application of usability engineering to medical devices ●
  • . Guidance for Industry and FDA Staff - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (issued May 11, 2005)

Software verification and validation testing includes:

  • . Verification and validation of system, user requirements and hazard mitigations through use case scenarios,
  • Usability testing with representative clinical users considering ease of use, intuitiveness and . visibility of data,
  • . Verification and Validation of registration accuracy:
    • o Acceptance criteria: registration error less than 3mm

{8}------------------------------------------------

  • O Methodology: 5000 different registrations have been performed on 100 peroperative CT-scan images from 50 patients undergoing EVAR procedure. Results have been compared to a gold standard transformation matrix
  • O Results: meet the acceptance criteria
  • Verification and Validation of panorama creation: quantification of parallax and X-ray ● panorama errors on in-vivo data
    • Acceptance criteria: max acceptable error of 10mm for peripheral artery surgery o
    • o Methodology: on 7 cases (6 patients and 1 phantom), comparison between panoramas generated by EndoNaut and a perfect panorama (manually corrected)
    • Results: mean errors below the acceptance criteria O
  • Verification and Validation of measuring functions of EndoNaut: .
    • Acceptance criteria: No error for distance measurement along centreline. Max error o 1mm for length measurements on images.
    • Methodology: comparison between centerline measurements given by EndoSize o (planning tool) and EndoNaut. Comparison between length measurements given by EndoNaut and a visible rule on the fluoroscopic image.
    • Results: meet the acceptance criteria o

The test results in this 510(k) demonstrate that EndoNaut complies with the aforementioned international and FDA-recognized consensus standards, and meets the acceptance criteria and is adequate for its intended use.

8. Clinical Data

A clinical investigation was conducted to evaluate safety and efficacy of the EndoNaut to provide image guidance by overlaying preoperative 3D vessel anatomy onto live fluoroscopic images in order to assist in the positioning of guidewires, catheters and other vascular devices.

The study was a single-centre, prospective feasibility pilot study. The objectives were to evaluate the feasibility of fusion imaging during aortic endovascular procedures using EndoNaut software, and to evaluate the efficiency of the device when deploying infrarenal aortic stent grafts to treat unruptured atheromatous aneurysms (EVAR).

  • . Clinical Endpoints: Primary endpoint was the feasibility rate of fusion. Secondary endpoints were radiation dose as measured by fluoroscopy time, dose-area product and air kerma.
  • Inclusion criteria:
    • o Patients eligible for endovascular treatment of aneurysm disease of the aorta.
    • Patients who received written and verbal information about the protocol and did not o object to participating in the trial.
  • . Exclusion criteria:
    • o Patients who also required a conventional surgical revascularisation procedure or who required an endovascular revascularisation procedure in another site.
    • o Patients who underwent MR angiography during preoperative evaluation.
    • Non-analysable CT angiogram (no or poor injection). O
    • Procedure performed in a hybrid room or in an operating room not equipped with a o mobile flat-panel detector.

{9}------------------------------------------------

Results of the clinical study support the indications for use of the EndoNaut to provide image guidance by overlaying preoperative 3D vessel anatomy onto live fluoroscopic images. Clinical study conclusion confirms that the device is safe and effective as used according to the instructions for use.

9. Substantial Equivalence Conclusion

The EndoNaut software is substantially equivalent to the predicate device in terms of intended use, indications for use and technical characteristics. The EndoNaut software has successfully undergone every bench testing.

Based on the information supplied in this 510(k), Therenva SAS concludes that EndoNaut is substantially equivalent to the predicate device and is safe and effective.

N/A