K Number
K171828

Validate with FDA (Live)

Device Name
ETCO2 Sensor
Manufacturer
Date Cleared
2018-05-31

(345 days)

Product Code
Regulation Number
868.1400
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use of the ETCO2 sensor is to provide carbon dioxide monitoring to a host monitoring system during anesthesia / recovery, in the intensive care unit (ICU), and in Emergency Medicine/Transport or Respiratory care.

Device Description

The ETCO2 Sensor is comprised of three main components, a sensor with photo detector and light emitter, cable and connector, The CO2 Sensor incorporates an infrared light source, of specified wavelength, and an infrared detector. The photo detector and light emitter end of the CO2 sensor is connected to an airway adapter. The airway adapter is connected between the patient airway and the respirator. As the patient completes an expiratory breath the sensor measures the CO2 levels in the expiratory breath and sends that data to the compatible CO2 monitor. The monitor then reads the data and converts the data so it can be displayed on screen.

AI/ML Overview

This document is a 510(k) premarket notification for an ETCO2 Sensor (Model CTM-RP01), not a study report detailing specific acceptance criteria and performance of an AI/ML device. Therefore, the requested information (acceptance criteria, study details, sample sizes, ground truth establishment, expert adjudication, MRMC studies, standalone performance, training set details) is not present in this document.

The document focuses on demonstrating substantial equivalence to a predicate device (Capnostat 5 CO2 sensor, K042601) based on similar technological characteristics and performance specifications for a carbon dioxide gas analyzer.

Here's what can be extracted and what is not available based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document provides the performance specifications of the subject device (ETCO2 Sensor, Model CTM-RP01) and compares them to the predicate device. These specifications serve as de facto "acceptance criteria" through the lens of substantial equivalence.

Acceptance Criteria (from Predicate Device Specifications)Reported Device Performance (ETCO2 Sensor, Model CTM-RP01)Comments
CO2 Measurement Range: 0 to 150 mmHg, 0 to 19.7%0 to 150 mmHg, 0 to 19.7%Same
CO2 Resolution: 0.1 mmHg 0 to 69 mmHg; 0.25 mmHg 70 to 150 mmHg0.1 mmHg 0 to 69 mmHg; 0.25 mmHg 70 to 150 mmHgSame
CO2 Accuracy:Same
0 - 40 mmHg ± 2 mmHg0 - 40 mmHg ± 2 mmHg
41 - 70 mmHg ± 5% of reading41 - 70 mmHg ± 5% of reading
71 - 100 mmHg ± 8% of reading71 - 100 mmHg ± 8% of reading
101 - 150 mmHg ± 10% of reading101 - 150 mmHg ± 10% of reading
Above 80 breath per minute ± 12% of readingAbove 80 breath per minute ± 12% of reading
*NOTE: Gas temperature at 25°C*NOTE: Gas temperature at 25°C
Respiration Rate Range: 0 to 150 breaths per minute (bpm)0 to 150 breaths per minute (bpm)Same
Respiration Rate Accuracy: ± 1 breaths per minute (bpm)± 1 breaths per minute (bpm)Same
Voltage Requirements: 5.0 VDC ±5%5.0 VDC ±5%Same
Interconnection: Lemo Redel 8-pin plasticLemo Redel 8-pin plasticSame
Operating Temperature/Humidity: 0-45°C, 10-90% RH, non-condensing0-40°C, 15-85% RH, non-condensingSimilar
Storage Temperature/Humidity: -40-70°C, <90% RH, non-condensing-10-40°C, <90% RH, non-condensingSimilar
Water Resistance: IPX4 – Splash-proof (sensor head only)IPX2 – Splash-proof (sensor head only)Similar
Compliance: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ISO 80601-2-55 (Performance)IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ISO 80601-2-55 (Performance)Same

The following information is NOT available in the provided text:

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • The document states "Non-clinical test data" were used, including compliance with industry standards. It does not specify a "test set" in the context of data for an AI/ML device, nor does it provide sample sizes or data provenance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable, as this is a medical device (sensor) and not an AI/ML diagnostic or predictive algorithm based on expert-labeled data. The performance criteria are met through physical measurements and engineering validation.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable for this type of device and testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, an MRMC study was not done. This device is a sensor performing direct measurements, not an AI assisting human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • The device's performance is inherently standalone in its measurement function, but it's a sensor, not an "algorithm only" device in the AI sense. Its output is displayed by a compatible monitor.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • The "ground truth" for this device would be established by reference gas standards and engineering measurement techniques for CO2 concentration, flow rates, and timing. The document references compliance with ISO 80601-2-55 (Respiratory gas monitors) which would outline such testing methods.

8. The sample size for the training set

  • Not applicable. This is a hardware sensor, not an AI/ML model that requires a training set.

9. How the ground truth for the training set was established

  • Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

May 31, 2018

Orantech Inc. % Mei Tan RA Consultant Chonconn Medical Device Consulting Co. LTD. 22A, Hai Jing Square, No.18 Taizi Road, Nanshan District, Shenzhen, 51800 China

Re: K171828

Trade/Device Name: ETCO2 Sensor, Model CTM-RP01 Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: Class II Product Code: CCK Dated: April 30, 2018 Received: April 30, 2018

Dear Mei Tan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Geeta K. Pamidimukkala -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K171828

Device Name EtCO2 Sensor, Model CTM-RP01

Indications for Use (Describe)

The intended use of the ETCO2 sensor is to provide carbon dioxide monitoring to a host monitoring system during anesthesia / recovery, in the intensive care unit (ICU), and in Emergency Medicine/Transport or Respiratory care.

Type of Use (Select one or both, as applicable)
☑Prescription Use (Part 21 CFR 801 Subpart D) ☐Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

510(K) Summary K171828

1. Prepared Date: 2018/5/30

2. Submitter Information

NameOrantech Inc.
AddressZone#A,4F,1st Bld,7th Industrial Zone,Yulv Community,GongMing, Guangming New District, Shenzhen,China 518106
Contact personMr. Hsin Xiong
Tel0755-23699939
Fax0755-23243594
EmailYunxi.xiong@orantech.cn

3. Submission Correspondent

NameChonconn Medical Device Consulting Co.LTD.
Address22A,HaiJing Square,No.18 Taizi Road,Nanshan District,Shenzhen,Guangdong,P.R.China,
Contact personMei(RA consultant)
E-mailFDA@chonconn.com/Xinmeitan@chonconn.com
Tel+86-755-33941160

4. Proposed Device Information

Trade NameETCO2 Sensor
ModelCTM-RP01
Common nameCarbon Dioxide (CO2) Analyzer
Regulatory classII
Production regulation21 CFR §868.1400
Product codeCCK
PanelAnesthesiology

5. Predicate Device Information

510(K)No.Trade Name/modelSubmitter
K042601Capnostat 5 CO2 sensorRespironics Novametrix, LLC

6. Device description

The ETCO2 Sensor is comprised of three main components, a sensor with photo detector and light emitter, cable and connector, The CO2 Sensor incorporates an infrared light source, of specified wavelength, and an infrared detector. The photo ETCO2 Sensor-510(k) Submission Page 1 of 4

{4}------------------------------------------------

detector and light emitter end of the CO2 sensor is connected to an airway adapter. The airway adapter is connected between the patient airway and the respirator. As the patient completes an expiratory breath the sensor measures the CO2 levels in the expiratory breath and sends that data to the compatible CO2 monitor. The monitor then reads the data and converts the data so it can be displayed on screen.

The ETCO2 sensor are described as follows:

ModelDescriptionCompatible Monitor
CTM-RP01Mainstream/ 3.0MGeneral Meditech INC. Patient Monitor model G3C

7. Intended use & Indications for Use

The intended use of the ETCO2 sensor is to provide carbon dioxide monitoring to a host monitoring system during anesthesia / recovery, in the intensive care unit (ICU), and in Emergency Medicine/Transport or Respiratory care.

8. Comparison to predicate device

The following is provided as a summary of how the technological characteristics of the device compare to the predicate device

ComparisonitemSubject DevicePredicate DeviceComments
510(K) numberPending applicationK042601--
DeviceCTM-RP01Capnostat 5 CO2 sensor--
Intended use&Indications forUseThe intended use of the ETCO2sensor is to provide carbondioxide monitoring to a hostmonitoring system duringanesthesia / recovery, in theintensive care unit (ICU), andin EmergencyMedicine/Transport orRespiratory care.The intended use of theCapnostat 5 CO2 sensor is toprovide carbon dioxidemonitoring to a host monitoringsystem during anesthesia /recovery, in the intensive careunit (ICU), and in EmergencyMedicine/Transport orRespiratory careSame
Environment ofUseHospitals, clinics, and otherhealthcare environmentsHospitals, clinics, and otherhealthcare environmentsSame
MeasurementTechniqueNon-dispersive infrared (NDIR)Infrared spectroscopyNon-dispersive infrared (NDIR)Infrared spectroscopySame
CO2measurementRange0 to 150 mmHg, 0 to 19.7%0 to 150 mmHg, 0 to 19.7%Same
CO2 Resolution0.1 mmHg 0 to 69 mmHg0.25 mmHg 70 to 150 mmHg0.1 mmHg 0 to 69 mmHg0.25 mmHg 70 to 150 mmHgSame

Table1 Comparison between CTM-RP01 sensor and Capnostat 5 CO2 sensor

{5}------------------------------------------------

CO2 Accuracy0 - 40 mmHg ± 2 mmHg0 - 40 mmHg ± 2 mmHgSame
41 - 70 mmHg ± 5% ofreading41 - 70 mmHg ± 5% of reading
71 - 100 mmHg ± 8% ofreading71 - 100 mmHg ± 8% of reading
101 - 150 mmHg ± 10% ofreading101 - 150 mmHg ± 10% of reading
Above 80 breath per minute ±12% of readingAbove 80 breath per minute ±12% of reading
* NOTE: Gas temperature at25°C* NOTE: Gas temperature at25°C
RespirationRate Range0 to 150 breaths per minute(bpm)0 to 150 breaths per minute(bpm)Same
RespirationRate Accuracy± 1 breaths per minute (bpm)± 1 breaths per minute (bpm)Same
VoltageRequirements5.0 VDC ±5%5.0 VDC ±5%Same
InterconnectionLemo Redel 8-pin plasticLemo Redel 8-pin plasticSame
Temperatureand HumidityOperating: 0 to 40°C, 15 to85% RH, non-condensingStorage: -10 to 40°C, <90%RH, non-condensingOperating: 0 to 45°C, 10 to 90%RH, non-condensingStorage: -40 to 70°C, <90% RH,non-condensingSimilar
WaterResistanceIPX2 – Splash-proof(sensor head only)IPX4 – Splash-proof(sensor head only)Similar
ComplianceIEC 60601-1(safety)IEC 60601-1-2(EMC)ISO 80601-2-55(Performance)IEC 60601-1(safety)IEC 60601-1-2(EMC)ISO80601-2-55(Performance)Same

From the comparison form above, the subject and predicate devices are based on the following same technological elements:

  • Same measurement techniques
  • Same intended use & indications for Use
  • Same CO2 Resolution & Accuracy
  • Same respiration rate range & Accuracy
  • Same voltage Requirements
  • Both comply to the same FDA recognized standards

The following slightly technological differences exist between the subject and predicate devices:

  • Similar temperature and humidity range
  • Water Resistance of predicate device is higher than Orantech's ●

These differences do not raise different questions of safety and effectiveness.

{6}------------------------------------------------

9. Clinical test data

Not applicable.

10. Non-clinical test data

1)Industry Standards for Electrical Safety, EMC and Essential Performance Testing of the ECO2 Sensor has been completed to verify compliance with FDA recognized national and international standards for electrical safety and performance for medical devices, and particular requirements applicable to this device including:

  • IEC 60601-1:2012 Basic safety and essential performance
  • IEC 60601-1-2:2007 and 2014 EMC
  • ISO 80601-2-55:2011 Respiratory gas monitors

2) Biocompatibility

The ETCO₂ Sensor does not contain any direct or indirect patient contacting components. Patient contacting accessories are previously cleared by FDA. Biocompatibility test is not applicable for subject device.

3) Software Verification

Software verification and validation testing were conducted and documentation was provided as recommended by FDA Software Guidance.

11. Conclusion

The results for testing demonstrate that the ETCO2 Sensor is substantially equivalent to the above listed predicate device.

§ 868.1400 Carbon dioxide gas analyzer.

(a)
Identification. A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.(b)
Classification. Class II (performance standards).