K Number
K171770

Validate with FDA (Live)

Date Cleared
2017-09-12

(90 days)

Product Code
Regulation Number
866.3130
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The cobas® Cdiff Nucleic acid test for use on the cobas® Liat® System is an automated, qualitative in vitro diagnostic test, that utilizes real-time polymerase chain reaction (PCR), for the detection of the toxin B (tcdB) gene of toxigenic Clostridium difficile in unformed (liquid or soft) stool specimens suspected of having C. difficile infection (CDI). The cobas® Cdiff Nucleic acid test for use on the cobas® Liat® System is intended for use as an aid in the diagnosis of CDI in humans in conjunction with clinical and epidemiological risk factors.

Device Description

The cobas® Cdiff Nucleic acid test for use on the cobas® Liat® System (cobas® Cdiff ) is a rapid, automated in vitro diagnostic test for the qualitative detection of C. difficile DNA in human stool specimens. The cobas® Liat® System is for in vitro diagnostic use. The system is designed to identify and/or measure the presence of genetic material in a biological sample. The system automates all nucleic acid amplification test (NAAT) processes, including reagent preparation, target enrichment, inhibitor removal, nucleic acid extraction, amplification, real-time detection, and result interpretation in a rapid manner. The cobas® Cdiff test uses silica magnetic particle-based nucleic acid extraction and TaqMan probe-based real-time PCR amplification and detection. The cobas® Liat® Analyzer automates and integrates sample purification, nucleic acid amplification and detection of the target sequence in biological samples. Other than adding the sample to the cobas® Cdiff assay tube, no reagent preparation or additional steps are required. The cobas® Liat® System consists of an cobas® Liat® Analyzer with integrated software for running tests and analyzing the results, and a single-use disposable cobas® Cdiff assay tube that holds all of the sample purification and PCR reagents and hosts the sample preparation and PCR processes specific for the Cdiff analyte. The test uses the assay tube as both the sample and reaction vessel. The assay tube comprises flexible tubing containing all required unit dose reagents pre-packed in tube segments, separated by pressure-sensitive seals, in the order of reagent use. During the testing process, multiple sample processing actuators of the analyzer compress the cobas® Cdiff assay tube to selectively release reagents from tube segments, move the sample from one segment to another, and control reaction conditions such as reaction volume, temperature, pressure, and incubation time. Precise control of all these parameters provides optimal conditions for assay reactions, allowing the test to achieve high performance similar to or better than that of currently available molecular assays. The cobas® Liat® Analyzer software controls and coordinated these actions to perform all required assay processes, including sample preparation, nucleic acid extraction, target enrichment, inhibitor removal, nucleic acid elution, and real-time PCR. All assay steps are performed within the closed and self-contained cobas® Cdiff assay tube, thereby eliminating the potential for cross-contamination between samples. The collected data are automatically analyzed and the result is displayed in the assay report on the integrated LCD touch screen of the cobas® Liat® Analyzer.

AI/ML Overview

Here's an analysis of the acceptance criteria and study details based on the provided text:

Acceptance Criteria and Device Performance for cobas® Cdiff Nucleic acid test for use on the cobas® Liat® System

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state formal "acceptance criteria" in a separate section with numerical targets like a typical clinical trial report for an AI device. Instead, the performance metrics are presented as the results of the studies conducted. I will infer the acceptance criteria from the reported performance, implying that the reported values met the internal standards set by the manufacturer for substantial equivalence.

Performance MetricAcceptance Criteria (Inferred from Predicate/Study Goals)Reported Device Performance (cobas® Cdiff Test)
Analytical Sensitivity (LOD)To detect C. difficile at a concentration to yield a Hit Rate of at least 95%, where all higher concentrations also have a Hit Rate of ≥ 95%.ATCC 43255: 90 CFU/swab R12087 (CD196): 45 CFU/swab
InclusivityTo detect 100% of tested toxigenic strains at 3xLOD.All 37 toxigenic strains detected with 100% hit rate.
Analytical Specificity (Cross-reactivity)No cross-reactivity with non-toxigenic C. difficile strains, human genomic DNA, or other common stool microorganisms, and no interference with toxigenic C. difficile detection.No cross-reactivity with 149 tested non-toxigenic Cdiff strains, human epithelial cells, or other microorganisms. No interference with detection of toxigenic Cdiff at 3x LOD.
InterferenceNo interference from common medications and fecal substances at specified concentrations.Fecal Fat: No interference up to 39% (w/v) Whole blood: No interference up to 100% (v/v) Mucin: No interference up to 50% (w/v) Various other medications and substances tested and confirmed for no interference at specified concentrations (see Table 5).
Reproducibility (~1xLOD)Overall percent agreement for low positive samples (~1xLOD) to demonstrate consistency across sites, operators, and lots.98.5% (262/266) with a 95% CI of (96.2%, 99.4%)
Reproducibility (~3xLOD)Overall percent agreement for moderate positive samples (~3xLOD) to demonstrate consistency across sites, operators, and lots.99.3% (268/270) with a 95% CI of (97.3%, 99.8%)
Reproducibility (Negative)Overall percent agreement for negative samples to demonstrate consistency across sites, operators, and lots.100.0% (262/262) with a 95% CI of (98.6%, 100.0%)
Clinical Sensitivity (vs. Combined Culture)High sensitivity for detecting toxigenic C. difficile in clinical samples.87.2% (156/179; 95% CI: 81.5% to 91.3%)
Clinical Specificity (vs. Combined Culture)High specificity for detecting toxigenic C. difficile in clinical samples.98.1% (818/834; 95% CI: 96.9% to 98.8%)
Clinical PPV (vs. Combined Culture)High Positive Predictive Value.90.7% (156/172; 95% CI: 85.4% to 94.2%)
Clinical NPV (vs. Combined Culture)High Negative Predictive Value.97.3% (818/841; 95% CI: 95.9% to 98.2%)
Initial Invalid RateLow initial invalid rate.1.4%
Initial Failed RateLow initial failed rate.0.2%
Final Invalid Rate (after retest)Very low final invalid rate.0.1%
Final Failed Rate (after retest)Zero final failed rate.0%

2. Sample size used for the test set and the data provenance

  • Analytical Sensitivity (LOD): Not specified as a per-sample-size, but replicates of each 5-member test panel were tested with 2 different lots of assay tubes. Additional replicates were tested for confirmation.
  • Inclusivity: 3 replicates per strain (37 toxigenic strains) were tested.
  • Analytical Specificity (Cross-reactivity): Not specified as a per-sample-size, but 149 non-toxigenic organisms/cells were tested.
  • Interference: Not specified as a per-sample-size, but 38 commonly used medications, fecal fat, whole blood, and mucin were tested.
  • Reproducibility: 818 total tests were performed, with 798 valid for analysis across 3 panel members (Negative, ~1xLOD, ~3xLOD). 3 replicates per panel member, run on 5 nonconsecutive days by 2 operators at 3 sites for 3 reagent lots.
  • Clinical Performance: 1,013 evaluable fresh remnant stool specimens were prospectively collected.
    • Data Provenance:
      • Country of Origin: Not explicitly stated, but "Nine (9) study sites from geographically diverse locations participated in this study." This suggests a multi-center study, likely within a single country (such as the US, given the FDA submission) or potentially across a few countries, but a specific list is not provided.
      • Retrospective/Prospective: The clinical performance study was prospective. "Evaluable fresh remnant specimens were prospectively collected from 1,013 patients."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Analytical (LOD, Inclusivity, Specificity, Interference): Not applicable for these studies as the "ground truth" is based on engineered samples (known concentrations of specific strains, or presence/absence of interfering substances).
  • Reproducibility: Not applicable, as this assessed the consistency of the device's own results, not against an external ground truth.
  • Clinical Performance:
    • Number of experts: Not explicitly stated.
    • Qualifications of experts: Not explicitly stated. The "ground truth" was established by "combined results of direct and enriched toxigenic culture using leftover, deidentified, unformed stool samples" performed at a "reference laboratory." The document mentions "Nineteen (19) operators distributed across these study sites performed testing with the cobas® Cdiff test," but these are operators, not necessarily the adjudicating experts for the ground truth. For microbiological culture, laboratory personnel with specific training in microbiology would establish the ground truth.

4. Adjudication method for the test set

  • Analytical, Reproducibility: Not applicable.
  • Clinical Performance:
    • Initial Ground Truth: "Combined results of direct and enriched toxigenic culture" served as the primary comparative method.
    • Discrepant Analysis: "All evaluable discordant specimens and an equal number of randomly selected and representative concordant samples were used for discrepant testing by an FDA-cleared comparator NAAT. Testing was performed at an external lab that was preselected and qualified by RMS." This implies a form of 2-method consensus with a third method for discrepancies. If direct and enrichment culture disagreed, or if the cobas test disagreed with the combined culture, a third (comparator NAAT) test was used for "discrepant analysis." However, the final clinical performance tables still directly compare the cobas test to the combined culture results, with the discrepant analysis results discussed qualitatively ("Of the 23 specimens with false-negative cobas® Cdiff test results relative to combined direct culture and enrichment culture, 19 were negative. 3 were positive and 1 was invalid by the second NAAT method.").

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

  • No, an MRMC comparative effectiveness study was not done. This device is an in vitro diagnostic (IVD) test, not an imaging AI device that relies on human-in-the-loop interpretation. The clinical study compares the device's output to laboratory culture results, not to human readers' interpretations with or without AI assistance.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Yes, the performance presented is purely standalone (algorithm/device only). The cobas® Cdiff Nucleic acid test is an automated, qualitative in vitro diagnostic test. The device provides "result interpretation" directly on its integrated LCD touch screen (page 5). The clinical performance data reflects the device's ability to detect the target gene independently, compared against culture methods.

7. The type of ground truth used

  • Analytical Studies (LOD, Inclusivity, Specificity, Interference): Lab-prepared spiked samples with known concentrations/compositions.
  • Clinical Performance: Microbiological culture results. Specifically, "combined results of direct and enriched toxigenic culture." A "second NAAT method" was used for discrepant analysis.

8. The sample size for the training set

  • Not applicable / Not explicitly stated. This is a molecular diagnostic test for Clostridium difficile, not a machine learning or AI algorithm in the typical sense that requires explicit "training data" for a model to learn from. The assay uses predefined primers and probes for real-time PCR detection (page 4). While the assay design and optimization would involve extensive internal development and testing, this is not characterized as a "training set" in the context of typical AI algorithm development.

9. How the ground truth for the training set was established

  • Not applicable / Not explicitly stated. As mentioned above, this is not an AI/ML device that uses a "training set" in the conventional sense. The "ground truth" for developing the test was based on well-characterized isolates of C. difficile and other microorganisms to design and validate the specificity and sensitivity of the primers and probes.

{0}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 12, 2017

Roche Molecular Systems, Inc. Clare Santulli Senior Manager 4300 Hacienda Drive Pleasanton, California 94588-2722

Re: K171770

Trade/Device Name: cobas Cdiff Nucleic acid test for use on the cobas Liat System Regulation Number: 21 CFR 866.3130 Regulation Name: Clostridium difficile toxin gene amplification assay Regulatory Class: Class II Product Code: OZN, OOI Dated: June 13, 2017 Received: June 14, 2017

Dear Clare Santulli:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements

{1}------------------------------------------------

as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Ribhi Shawar -S For

Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171770

Device Name

cobas® Cdiff Nucleic acid test for use on the cobas® Liat® System

Indications for Use (Describe)

The cobas® Cdiff Nucleic acid test for use on the cobas® Liat® System is an automated, qualitative in vitro diagnostic test, that utilizes real-time polymerase chain reaction (PCR), for the detection of the toxin B (tcdB) gene of toxigenic Clostridium difficile in unformed (liquid or soft) stool specimens suspected of having C. difficile infection (CDI). The cobas® Cdiff Nucleic acid test for use on the cobas® Liat® System is intended for use as an aid in the diagnosis of CDI in humans in conjunction with clinical and epidemiological risk factors.

Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Submitter NameRoche Molecular Systems, Inc.
Address4300 Hacienda DrivePleasanton, CA 94588-2722
ContactClare SantulliPhone: (925) 730-8886FAX: (925) 225-0207Email: clare.santulli@roche.com
Date PreparedSeptember 11, 2017
Proprietary Namecobas® Cdiff Nucleic acid test for use on the cobas® Liat® System
Common NameClostridium difficile Test
Classification Name21 CFR 866.3130 - Clostridium difficile toxin gene amplification assay21 CFR 862.2570 - Real Time Nucleic Acid Amplification System
Product CodesOZN, OOI
Predicate Devicescobas® Cdiff Test on the cobas® 4800 system
Establishment RegistrationBranchburg: 2243471Pleasanton: 3004141078Indianapolis: 1823260

{4}------------------------------------------------

1. DEVICE DESCRIPTION

The cobas® Cdiff Nucleic acid test for use on the cobas® Liat® System (cobas® Cdiff ) is a rapid, automated in vitro diagnostic test for the qualitative detection of C. difficile DNA in human stool specimens.

The cobas® Liat® System is for in vitro diagnostic use. The system is designed to identify and/or measure the presence of genetic material in a biological sample. The system automates all nucleic acid amplification test (NAAT) processes, including reagent preparation, target enrichment, inhibitor removal, nucleic acid extraction, amplification, real-time detection, and result interpretation in a rapid manner.

1.1. Target Selection

The cobas® Cdiff test detects the tcdB target-specific oligonucleotide sequences. Toxin B (or tcdB) is a major toxin that is implicated in C. difficile pathogenesis and allows the differentiation between toxigenic and non-toxigenic C. difficile strains. Primers and probe oligonucleotide sequences were designed to detect C. difficile conserved sequences without cross-reacting with other Clostridium genus organisms, as well as with organisms commonly found in normal gut flora.

1.2. Test Principle

The cobas® Cdiff test uses silica magnetic particle-based nucleic acid extraction and TaqMan probe-based real-time PCR amplification and detection. The cobas® Liat® Analyzer automates and integrates sample purification, nucleic acid amplification and detection of the target sequence in biological samples. Other than adding the sample to the cobas® Cdiff assay tube, no reagent preparation or additional steps are required. The cobas® Liat® System consists of an cobas® Liat® Analyzer with integrated software for running tests and analyzing the results, and a single-use disposable cobas® Cdiff assay tube that holds all of the sample purification and PCR reagents and hosts the sample preparation and PCR processes specific for the Cdiff analyte. The test uses the assay tube as both the sample and reaction vessel. The assay tube comprises flexible tubing containing all required unit dose reagents pre-packed in tube segments, separated by pressure-sensitive seals, in the order of reagent use.

{5}------------------------------------------------

During the testing process, multiple sample processing actuators of the analyzer compress the cobas® Cdiff assay tube to selectively release reagents from tube segments, move the sample from one segment to another, and control reaction conditions such as reaction volume, temperature, pressure, and incubation time. Precise control of all these parameters provides optimal conditions for assay reactions, allowing the test to achieve high performance similar to or better than that of currently available molecular assays. The cobas® Liat® Analyzer software controls and coordinates these actions to perform all required assay processes, including sample preparation, nucleic acid extraction, target enrichment, inhibitor removal, nucleic acid elution, and real-time PCR. All assay steps are performed within the closed and self-contained cobas® Cdiff assay tube, thereby eliminating the potential for cross-contamination between samples. The collected data are automatically analyzed and the result is displayed in the assay report on the integrated LCD touch screen of the cobas® Liat® Analyzer.

2. INDICATIONS FOR USE

The cobas® Cdiff Nucleic acid test for use on the cobas® Liat® System is an automated, qualitative in vitro diagnostic test, that utilizes real-time polymerase chain reaction (PCR), for the detection of the toxin B (tcdB) gene of toxigenic Clostridium difficile in unformed (liquid or soft) stool specimens obtained from patients suspected of having C. difficile infection (CDI). The cobas® Cdiff Nucleic acid test for use on the cobas® Liat® System is intended for use as an aid in the diagnosis of CDI in humans in conjunction with clinical and epidemiological risk factors.

TECHNOLOGICAL CHARACTERISTICS 3.

The primary technological characteristics and intended use of the RMS cobas® Cdiff Nucleic acid test for use on the cobas Liate System are substantially equivalent to other legally marketed nucleic acid amplification tests intended for the qualitative detection of Clostridium difficile. As indicated in Table 1, the RMS cobas® Cdiff Nucleic acid test for use on the cobas® Liat® System is substantially equivalent to significant characteristics of the identified predicate device, the currently cleared cobas® Cdiff Test for use on the cobas® 4800 System, 510(k) 142422. The predicate device and the submitted device both utilize the cobas® PCR Media Uni Swab Sample Kit for specimen transfer into cobas PCR Media.

{6}------------------------------------------------

Submitted Device:cobas® Cdiff Nucleic acid test for useon the cobas® Liat® SystemPredicate Device:cobas® Cdiff Test for use on the cobas®4800 System , K142422
Intended UseThe cobas® Cdiff Nucleic acid test for useon the cobas® Liat® System is anautomated, qualitative in vitro diagnostictest, that utilizes real-time polymerasechain reaction (PCR), for the detection ofthe toxin B (tcdB) gene of toxigenicClostridium difficile in unformed (liquid orsoft) stool specimens obtained frompatients suspected of having C. difficileinfection (CDI). The cobas® Cdiff Nucleicacid test for use on the cobas® Liat®System is intended for use as an aid in thediagnosis of CDI in humans in conjunctionwith clinical and epidemiological riskfactors.The cobas® Cdiff Test on thecobas® 4800 system is an automated,qualitative in vitro diagnostic test, thatutilizes real-time polymerase chainreaction (PCR), for the direct detection ofthe toxin B (tcdB) gene of toxigenicClostridium difficile in unformed (liquid orsoft) stool specimens obtained frompatients suspected of having C. difficileinfection (CDI). The cobas® Cdiff Test isintended for use as an aid in the diagnosisof CDI in humans in conjunction withclinical and epidemiological risk factors.
Conditions for useFor prescription useSame
Regulation Number,Classification andSubsequent ProductCodes21 CFR 866.2660: Microorganismdifferentiation and identification device.OMN: C. difficile Nucleic AcidAmplification Test Assay.OOI: Real Time Nucleic AcidAmplification System.Same
Sample TypesUnformed soft stool specimensSame
Amplification TechnologyReal-time PCRSame
Detection TechnologyTaqMan probes with fluorescent dyesSame
Internal ControlA gram-positive Bacillus thuringiensis israelensis bacterial organism to monitorthe full process on the cobas® Liat®Analyzer. Native sequence in the bacteriais used as the Internal Control target.Lambda phage with encapsulated internalcontrol sequence
Positive ControlPlasmid in bufferSame
Negative ControlBuffer onlySame
Analyte TargetsToxin B ( tcdB ) geneSame
Sample Transfer Devicescobas® PCR Media Uni Swab Sample KitSame
Sample PreparationMagnetic bead-based nucleic acidextraction automated bycobas® Liat® AnalyzerMagnetic bead-based nucleic acidextraction automated bycobas® 4800 System
Result AnalysisBased on PCR cycle threshold analysisSame
Subject StatusSymptomaticSame

Similarities and Differences between the cobas® Cdiff Nucleic acid test for use Table 1: on the cobas® Liat® System and the Predicate Device

{7}------------------------------------------------

NON-CLINICAL PERFORMANCE EVALUATION 4.

4.1. Analytical Sensitivity

The analytical sensitivity (Limit of Detection or LOD) for the cobas® Cdiff test was determined by analyzing 2 toxigenic C. difficile strains ATCC 43255(VPI 10463) and R12087 (CDI 196). Five-member test panels were prepared with the two different strains of Cdiff at concentrations that bracketed the expected Limit of Detection (LOD). Replicates of each test panel member were then tested with two different lots of cobas® Liat® Cdiff assay tubes to determine the Limit of Detection, or the lowest Cdiff concentration to yield a Hit Rate of at least 95%, where all higher concentrations also have a Hit Rate of ≥ 95%. Additional replicates were tested to confirm the LOD.

Negative stool background samples were also run to verify the Limit of Blank (LoB).

The LOD results among 2 reagent lots are shown in Table 2.

LOD cobas® Cdiff Nucleic acid test for use on the cobas® Liat® System Table 2:

Strain IDToxinotypeREA*TypePFG†TypeRibotypePhenotypeLOD(CFU/swab)
ATCC 43255(VPI 10463)0N/AN/A87A+B+CDT-90
R12087 (CD196)IIIBINAP127A+B+CDT+45

*Restriction endonuclease analysis; *Pulse Field Gel

4.2. Inclusivity

The limit of detection of cobas Cdiff test on 37 toxigenic strains representing additional toxinotypes was verified by testing three replicates per strain at three times the LOD level (~60 CFU/ml equivalent to 270 CFU/swab) of ATCC 43255.

All 37 toxigenic strains were detected with a 100% hit rate for each toxigenic strain in this study, confirming that the cobas Cdiff test can detect these C. difficile toxinotypes. Results are shown in Table 3.

{8}------------------------------------------------

Sample IDCdiff StrainToxinotypeRibotype
1RMSCC 11251(ATCC#BAA-1382; 630)0012
2EX 623I102
3AC 008II103
4RMSCC 12827 [2004118; CDC-204118 (NAP-1)]III027
5SE 844IIIa080
6CH6230IIIcN/A
7RMSCC 11298 (P43)IVN/A
855767IV023
9RMSCC 11300 (2748-06)V078
10SE 881V045
11RMSCC 11302 (SE 1203)VI033
1257267VII063
13RMSCC 12472(ATCC# 43598; 1470)VIII017
14RMSCC 11299 (51680)IX019
15RMSCC 11304 (CCUG 8864/STCC20309)X036
16RMSCC 11308 (F15)XIIN/A
17IS 25XII056
18R 9367XIII070
19R 10870XIV(New-XIVa)111
20R 9385XV(New XIVb)122
21SUC36XVI078
22RMSCC 11309 (No 1313)XVII232
23K095XVIII014
24TR13XIXN/A
25TR14XXN/A
26CH6223XXIN/A
27CD07-468XXIIN/A
288785XXIII(New-IXc)N/A
29597BXXIV131
307325XXV027
317459XXVIN/A
32KK2443/2006XXVIIN/A
33CD08-070XXVIII126
34CD07-140XXIX056
35ES 130XXXN/A
36WA 151XXXIN/A
37173070XXXIIN/A

Table 3: Results of Testing Three Replicates of 40 Cdiff Strains at 60CFU/mL (~3x LOD)

{9}------------------------------------------------

Analytical Specificity (Cross-reactivity) 4.3.

The cobas Liat Cdiff test was examined for analytical specificity by testing 146 non-toxigenic strains (toxinotype XI) of Clostridium difficile, human genomic DNA, and other microorganisms which could be present in clinical stool specimens at ~3xLOD concentration.

Test panels were prepared that contained each of these microorganisms, and each test panel was tested with the cobas Liat® Cdiff assay tubes. The microorganism concentrations that were tested correspond to approximately 1E+06 units (CFU, IFU, cells) per mL of stool specimen for bacteria and human epithelial cells (as a source for human genomic DNA) and 1E+05 TCID50and/or PFU per mL of stool specimen for viruses.

Three non-toxigenic Cdiff strains (toxinotype XI) tested during inclusivity study [RMSCC # 11305 (ES 1103), 11306 (6035/06) and 12414 (F14)] were not detected by the cobas® Cdiff test are included in this section.

The cobas® Liat Cdiff Test did not cross react with Cdiff non-toxigenic (toxinotype XI) strains, human epithelial cells (tested as a source of human genomic DNA) or other microorganisms which could be present in clinical stool specimens and that were tested in the study. Further, the presence of any of these potential cross reactants tested did not interfere with detection of toxigenic Cdiff strains tested at approximately 3x LOD level.

All 149 non-toxigenic strains (toxinotype XI) of Clostridium difficile, human genomic DNA, and other microorganisms are listed in Table 4.

{10}------------------------------------------------

Organism / CellNumberOrganism / Cell Name:Source#
1Abiotrophia defectivaATCC# 49176
2Acinetobacter baumanniiATCC# 19606
3Acinetobacter IwoffiiATCC# 15309
4Aeromonas hydrophilaATCC# 7966
5Alcaligenes faecalisATCC# 35655
6Alcaligenes faecalis subsp. FaecalisATCC# 8750
7Alcaligenes faecalis subsp. FaecalisATCC# 15554
8Anaerococcus tetradiusATCC# 35098
9Bacillus cereusATCC# 13472
10Bacillus cereusATCC# 11778 (also known as HER 1414)
11Bacteroides caccaeATCC# 43185
12Bacteroides fragilisATCC# 25285
13Bacteroides merdaeATCC# 43184
14Bacteroides stercorisATCC# 43183
15Bifidobacterium adolescentisATCC# 15703
16Bifidobacterium longumATCC# 15707
17Campylobacter coliATCC# 33559
18Campylobacter jejuni (f.k.a Campylobacter coli)ATCC# 43479
19Campylobacter jejuni Subsp. jejuniATCC# 33292
20Candida albicansATCC# 10231
21Candida catenulataATCC# 10565
22Cedecea davisaeATCC# 33431
23Chlamydia Trachomatis Serovar L2Zeptometrix # 0801776
24Citrobacter amalonaticusATCC# 25405
25Citrobacter freundiiATCC# 8090
26Citrobacter koseriATCC# 27028
27Citrobacter sedlakiiATCC# 51115
28Clostridium beijerinckiiATCC# 8260
29Clostridium bifermentansATCC# 638
30Clostridium bolteaeATCC# BAA-613
31Clostridium butyricumATCC# 19398
32Clostridium chauvoeiATCC# 11957
33Clostridium difficile (Non-toxigenic, Serogroup B)ATCC# 43593
34Clostridium difficile (Non-toxigenic, Serogroup I)ATCC# 43601
35Clostridium fallaxATCC# 19400
36Clostridium haemolyticumATCC# 9650
37Clostridium histolyticumATCC# 19401
38Clostridium innocuumATCC# 14501
39Clostridium methylpentosumATCC# 43829
40Clostridium nexileATCC# 27757
41Clostridium novyiATCC# 19402
42Clostridium orbiscindens (re-namedFlavonifractor plautii)ATCC# 49531
Organism / CellNumberOrganism / Cell Name:Source#
43Clostridium paraputrificumATCC# 17796
44Clostridium perfringensATCC# 13124
45Clostridium ramosumATCC# 25582
46Clostridium scindensATCC# 35704
47Clostridium septicumATCC# 12464
48Clostridium sordelliiATCC# 9714
49Clostridium sphenoidesATCC# 19403
50Clostridium spiroformeATCC# 29899
51Clostridium sporogenesCCRI# 11128
52Clostridium sporogenesATCC# 15579
53Clostridium symbiosumATCC# 14940
54Clostridium tertiumDSMZ# 662
55Clostridium tetaniATCC# 19406
56Collinsella aerofaciensATCC# 25986
57Corynebacterium genitaliumATCC# 33030
58Cytomegalovirus (AD-169)ZeptoMetrix # 0810003CF
59Desulfovibrio pigerATCC# 29098
60Edwardsiella tardaATCC# 15947
61Eggerthella lentaATCC# 25559
62Enterobacter aerogenesATCC# 13048
63Enterobacter cloacaeATCC# 13047
64Enterococcus casseliflavusATCC# 25788
65Enterococcus cecorumATCC# 43198
66Enterococcus disparATCC# 51266
67Enterococcus faecalis Van BATCC# 51299
68Enterococcus faecium Van AATCC# 35667
69Enterococcus gallinarum Van CATCC# 49573
70Enterococcus hiraeATCC# 8043
71Enterococcus raffinosusATCC# 49427
72Escherichia coliATCC# 11775
73Escherichia coliATCC# 25922
74Escherichia coli O157:H7ATCC# 700927
75Escherichia fergusoniiATCC# 35469
76Escherichia hermanniiATCC# 33650
77Fusobacterium variumATCC# 8501
78Gardnerella vaginalisATCC# 14018
79Gemella morbillorumATCC# 27824
80Hafnia alveiCMCC# 147
81HCT-15 Human CellsATCC# CCL-225
82Helicobacter fennelliaeATCC# 35683
83Helicobacter pyloriATCC# 43504
84Human Adenovirus 41ZeptoMetrix # 0810085CF
85Human Coxsackievirus A4ZeptoMetrix # 0810142CF
86Human Coxsackievirus B4ZeptoMetrix # 0810075CF
87Human Echovirus 11ZeptoMetrix # 0810023CF
Organism / CellNumberOrganism / Cell Name:Source#
88Human Enterovirus 71ZeptoMetrix # 0810236CF
89Human RotavirusZeptoMetrix # 0810041CF
90Klebsiella oxytocaATCC# 33496
91Klebsiella pneumoniae subsp. PneumoniaeATCC# 13883
92Lactobacillus acidophilusATCC# 4356
93Lactobacillus reuteriATCC# 23272
94Lactococcus lactisATCC# 19435
95Leminorella grimontiiATCC# 33999
96Listeria grayiATCC# 19120
97Listeria innocuaATCC# 33090
98Listeria monocytogenesATCC# 15313
99Listeria monocytogenesATCC# BAA-839
100Mitsuokella multacidaATCC# 27723
101Mobiluncus curtisiiATCC# 35241
102Molellerella wisonsensisATCC# 35017
103Morganella morganiiATCC# 25830
104Neiserria gonorrhoeaeATCC# 35201
105Norovirus GIIZeptoMetrix # 0810087CF
106Peptoniphilus asaccharolyticusATCC# 14963
107Peptostreptococcus anaerobiusATCC# 27337
108Plesiomonas shigelloidesATCC# 14029
109Porphyromonas asaccharolyticaATCC# 25260
110Prevotella melaninogenicaATCC# 25845
111Proteus mirabilisATCC# 29906
112Proteus mirabilisATCC# 25933
113Proteus penneriATCC# 35198
114Providencia alcalifaciensATCC# 9886
115Providencia rettgeriATCC# 9250
116Providencia stuartiiATCC# 29914
117Pseudomonas aeroginosaATCC# 35554
118Pseudomonas aeruginosaATCC# 33584
119Pseudomonas putidaATCC# 12633
120Ruminococcus bromiiATCC# 27255
121Salmonella enterica subsp. arizonae (f.k.a.Salmonella choleraesuis ssp. arizonae)ATCC# 13314
122Salmonella enterica subsp. entericaCMCC# 1975
123Salmonella enterica subsp. enterica serovarCholeraesuisATCC# 7001
124(^)Salmonella enterica subsp. enterica serovarTyphiATCC# 19430
125Salmonella enterica subsp.enterica serovarTyphimuriumATCC# 14028
126Serratia liquefaciensCMCC# 169
127Serratia marcescensATCC# 8100
128Serratia marcescensATCC# 13880
129Serratia liquefaciensATCC# 27592
Organism / CellNumberOrganism / Cell Name:Source#
130Shigella boydiiATCC# 9207
131Shigella dysenteriaeATCC# 11835
132Shigella sonneiATCC# 29930
133Staphylococcus aureusATCC# 43300
134Staphylococcus epidermidisATCC# 14990
135Stenotrophomonas maltophiliaATCC# 13637
136Streptococcus agalactiaeATCC# 13813
137Streptococcus dysgalactiaeATCC# 43078
138Streptococcus intermediusATCC# 27335
139Streptococcus sp.; strain V8ATCC# 12973
140Streptococcus uberisATCC# 19436
141Trabulsiella guamensisATCC# 49490
142Veillonella parvulaATCC# 10790
143Vibrio choleraeATCC# 25870
144Vibrio parahaemolyticusATCC# 17802
145Yersinia bercovieriATCC# 43970
146Yersinia rohdeiATCC# 43380
147Clostridium difficile (Non-toxigenic- Xla)(ES 1103)*BMTU# 13799 (RMSCC 11305)
148Clostridium difficile (Non-toxigenic- Xia)(6035/06)*BMTU#9961 (RMSCC 11306)
149Clostridium difficile (Non-toxigenic-Xlb) (F14)*BMTU# 9962 (RMSCC 12414 and 11307)

Table 4: All Cross Reactivity Panel Members

{11}------------------------------------------------

{12}------------------------------------------------

{13}------------------------------------------------

  • Non-toxigenic Clostridium difficile (#147,148,149) were tested during Inclusivity study.

{14}------------------------------------------------

Interference 4.4.

Thirty eight commonly used medications, as well as fecal fat, whole blood, and mucin, were tested for potential interference effects with the cobas Cdiff test. All substances were tested at levels above what could be reasonably expected to be collected by a swab in a stool specimen. The amount of interference substance is expressed as concentration in primary stool specimen. Two toxigenic C. difficile isolates were spiked to 3 x Limit of Detection (LOD) of the cobas Cdiff test and used as targets in the tests. Exogenous substances at the highest tolerable concentration with no interference on cobas Cdiff test are shown in Table 5. Exogenous substances concentrations higher than listed in Table 5 may generate false negative or invalid results.

For fecal fat, no interference was observed up to 39% (w/v), and for mucin, no interference was observed up to 50% (w/v). For whole blood, no interference was observed up to 100% (v/v), which is equivalent to 100% of the capacity of the transfer swab. These results are summarized in Table 5.

SubstancePrimary StoolSpecimenConcentration
Fecal Fat0.22% - 39% (w/v)
Whole blood100% (v/v)
Mucin50% (w/v)
Aleve100% (w/v)
Mylanta100% (w/v)
Anusol100% (w/v)
Dulcolax23% (w/v)*
Equate Laxative50% (w/v)*
Equate Hydrocortisone100% (w/v)
E-Z-HD Barium Sulfate100% (w/v)
Fleet100% (w/v)
Glycerin Suppositories100% (w/v)
Gravol Suppositories100% (w/v)
Gynol II Contraceptive10% (w/v)*
Imodium100% (w/v)
Kaopectate100% (w/v)
K-Y Jelly100% (w/v)
Metronidazole100% (w/v)

Results from Interference Substances Testing Table 5:

{15}------------------------------------------------

SubstancePrimary StoolSpecimenConcentration
Miconazole100% (w/v)
Mineral Oil100% (w/v)
Monistat Cream100% (w/v)
Monistat Complete Care100% (w/v)
Nystatin Ointment100% (w/v)
Palmitic Acid100% (w/v)
Pedia Lax100% (w/v)
Pepto Bismol25% (w/v)*
Witch Hazel50% (w/v)*
Preparation H Hemorrhoidal Cream100% (w/v)
Preparation H Hemorrhoidal ointment100% (w/v)
Dramamine12.5% (w/v)*
Steric Acid100% (w/v)
Docusate Sodium100% (w/v)
Tums50% (w/v)*
Mesalamine Rectal Suspension100% (w/v)
Vagisil Anti-itch Cream12.5% (w/v)*
Vancomycin100% (w/v)
Vaseline100% (w/v)
Sun Screen100% (w/v)
Monistat Vaginal Insert100% (w/v)
Vaginal Contraceptive Film1 film vortexed with20 ml of primary stoolsample
Spermicidal Condoms1 Condom vortexedwith 20 mL of primarystool sample

5. CLINICAL PERFORMANCE EVALUATION

5.1. Reproducibility

The reproducibility of the cobas Cdiff test was established in a multi-site investigation (2 external sites, 1 internal site) using simulated clinical samples evaluated across reagent lot, site, operator, and testing day. Overall, 818 tests were performed in this study, out of which 798 were valid for the study and included in the final percent agreement analysis.

Panels consisted of 3 members: 1 negative specimen and 2 specimens with different concentrations of 1 strain of toxigenic C. difficile – the first a low positive concentration at ~ 1 x limit of detection (LOD) and the second a moderate positive at ~3 x LOD. There were

{16}------------------------------------------------

3 replicates per panel member; testing of 1 replicate sample was 1 test. A run was defined as testing of 3 replicates of a panel member. For each of 3 lots, panels were run on 5 different nonconsecutive days by 2 different operators at each of the 3 sites. Table 6 shows the percent agreement results by panel member concentration, negative percent agreement for negative panel members and positive percent agreement for positive panel members. For the ~1 x LOD panel members (low positive), the overall percent agreement was 98.5% with a lower bound of the two-sided 95% Score CI of 96.2%. Overall percent agreement was 99.3% for the ~3 x LOD panel member (moderate positive) and 100% for the negative panel member.

Percent Agreement by Panel Member Compared to Acceptance Criteria Table 6:

Percent Agreement
Panel MemberNumber of ValidTest ResultsEstimate(95% CI)*MetAcceptanceCriteria
Negative262100.0% (262/262)(98.6%, 100.0%)n/a
~1xLOD26698.5% (262/266)(96.2%, 99.4%)Yes
~3xLOD27099.3% (268/270)(97.3%, 99.8%)n/a
  • 95% CI = two-sided 95% Score binomial confidence interval.

Results were in agreement when a positive panel member had a valid result of Positive (Cdiff Detected) for the analyte orwhen the negative panel member had a valid result of Negative (Cdiff Not Detected) for the analyte. n/a = not applicable

Table 7 presents the mean, total standard deviation (SD), and total percent coefficient of variation (CV %) of Ct values by panel member concentration.

Table 7 also presents percent agreement separately by lot, site, operator, and testing day.

{17}------------------------------------------------

Panel MemberNegative~1x LOD~3x LOD
Number of Valid Test Results262266270
Overall Percent Agreement, % (n/N)*
PercentAgreement100.0(262/262)98.5(262/266)99.3(268/270)
Overall95% Score CI(98.6, 100.0)(96.2, 99.4)(97.3, 99.8)
Percent Agreement for Component Variables, % (n/N)*
Reagent Lot1100.0 (85/85)100.0 (87/87)100.0 (90/90)
2100.0 (88/88)100.0 (90/90)100.0 (90/90)
3100.0 (89/89)95.5 (85/89)97.8 (88/90)
Site1100.0 (85/85)100.0 (88/88)100.0 (90/90)
2100.0 (89/89)97.8 (87/89)98.9 (89/90)
3100.0 (88/88)97.8 (87/89)98.9 (89/90)
1100.0 (43/43)100.0 (44/44)100.0 (45/45)
2100.0 (42/42)100.0 (44/44)100.0 (45/45)
3100.0 (45/45)97.7 (43/44)100.0 (45/45)
Operator4100.0 (44/44)97.8 (44/45)97.8 (44/45)
5100.0 (44/44)100.0 (44/44)97.8 (44/45)
6100.0 (44/44)95.6 (43/45)100.0 (45/45)
Testing Day1100.0 (53/53)100.0 (54/54)100.0 (54/54)
2100.0 (54/54)96.3 (52/54)100.0 (54/54)
3100.0 (52/52)96.0 (48/50)98.1 (53/54)
4100.0 (52/52)100.0 (54/54)98.1 (53/54)
5100.0 (51/51)100.0 (54/54)100.0 (54/54)

Table 7: Percent Agreement by Panel Member for Lot, Site, Operator and Testing Day

Note: CI = confidence interval.

  • For the negative panel member: Percent agreement = (number of not detected results) x 100. For the positive panel members: Percent agreement = (number of detected results/total valid results) x 100.

Table 8 below presents the overall SD and percent CV (%) of Ct values for positive panel members at ~1 x LOD and ~3 x LOD concentrations, as well as the variance attributed to individual components (lot, site, operator, testing day, and within-run). Within-run variation refers to the variation within a 'study run' that consists of the 3 replicates for a given panel member processed by the same operator on the same analyzer on the same day.

Across all components, the total CV (%) was ≤ 1.9% with respect to the cycle threshold (Ct) value for all positive panel members. Within each component, the CV (%) was ≤ 1.6% across positive panel members.

{18}------------------------------------------------

Standard Deviation and Percent Coefficient of Variation
LotSiteOperatorDayWithin-RunTotal
Panel MemberNMean CtSDCVSDCVSDCVSDCVSDCVSDCV
~1xLOD26231.40.130.4%0.260.8%0.000.0%0.130.4%0.511.6%0.601.9%
~3xLOD26830.00.230.8%0.250.8%0.080.3%0.000.0%0.381.3%0.511.7%

Table 8: Overall Mean, Standard Deviation (SD) and Percent Coefficients of Variation (CV) for Ct Values from Valid Results for Positive Panel Members

5.2. Clinical Performance

The clinical performance of the cobas® Cdiff test was established in a prospective, multi-site investigation designed to evaluate the sensitivity and specificity of the cobas Cdiff test compared to the combined results of direct and enriched toxigenic culture using leftover, deidentified, unformed stool samples from patients suspected of having C. difficile infection.

Nine (9) study sites from geographically diverse locations participated in this study. Nineteen (19) operators distributed across these study sites performed testing with the cobas® Cdiff test. Reference culture (combined direct and enriched toxigenic culture) was performed at a reference laboratory. The remnant sample was stored frozen for discrepant analyses or possible additional testing, as determined by Roche Molecular Solution (RMS).

All evaluable discordant specimens and an equal number of randomly selected and representative concordant samples were used for discrepant testing by an FDA-cleared comparator NAAT. Testing was performed at an external lab that was preselected and qualified by RMS.

Evaluable fresh remnant specimens were prospectively collected from 1,013 patients; 483 males (47.7%) and 530 females (52.3%) with a median age of 59 years (range 5 to 98). All 1,013 specimens had valid results for both combined direct and enrichment culture and the cobas Cdiff test. Of the 1,013 specimens, 179 were positive for toxigenic C. difficile using the combined results from direct and enrichment toxigenic culture, for a prevalence rate of 17.7% for the study.

{19}------------------------------------------------

Comparison with Combined Direct and Broth Enrichment Culture 5.3.

The clinical performance of the cobas® Cdiff test compared with the combined results of direct and broth enrichment toxigenic culture are shown in Table 9. The sensitivity and specificity of the cobas® Cdiff test were 87.2% (156/179; 95% CI: 81.5% to 91.3%) and 98.1% (818/834; 95% CI: 96.9% to 98.8%), respectively; and the PPV and NPV were 90.7% (95% CI: 85.4% to 94.2%) and 97.3% (95% CI: 95.9% to 98.2%), respectively. Of the 23 specimens with falsenegative cobas® Cdiff test results relative to combined direct culture and enrichment culture, 19 were negative. 3 were positive and 1 was invalid by the second NAAT method. Of the 16 specimens with false-positive cobas Cdiff test results relative to combined direct and enrichment culture, 14 were positive and 2 were negative by the second NAAT method.

Table 9: Comparison of cobas Ciat® Cdiff Test with Combine Direct and Enrichment
Culture Results
Combined Direct and Enrichment Culture Result
cobas® Liat Cdiff TestResultPositiveNegativeTotal
Detected15616172
Not Detected23818841
Total1798341013
Sensitivity (95% CI)87.2% (156/179; 95% CI = 81.5% to 91.3%)
Specificity (95% CI)98.1% (818/834; 95% CI = 96.9% to 98.8%)
PPV (95% CI)90.7% (156/172; 95% CI = 85.4% to 94.2%)
NPV (95% CI)97.3% (818/841; 95% CI = 95.9% to 98.2%)

Note: Specimens with both combined direct and enrichment culture and valid cobas Liat Cdiff test results are considered evaluable and included in this summary table.

Note: Cl = (score) confidence interval, PPV = positive predictive value, NPV = negative predictive value, OPA = overall percent agreement.

Of the 1,016 specimens tested with the cobas® Cdiff test, 1.4% were initially invalid and 0.2% initially had failed results. Following 1 retest per invalid or failed result, the final invalid rate was 0.1% and the final failed rate was 0%.

{20}------------------------------------------------

Comparison with Direct Culture 5.4.

The performance of the cobas® Cdiff test compared to direct culture is shown in Table 10. The positive percent agreement (PPA) and negative percent agreement (NPA) of the cobas® Cdiff test compared to the direct culture for all 1.013 specimens were 94.6% (139/147) and 96.2% (833/866). respectively. Of the 8 specimens with false-negative cobas® Cdiff test results relative to direct culture, 7 were negative and 1 was positive by a second NAAT method. Of the 33 specimens with false-positive cobas 9 Cdiff test results relative to direct culture, only 17 were tested with the second NAAT method: 14 were positive and 3 were negative by that second NAAT method. The remaining 16/33 specimens were positive by enrichment culture and hence not tested with the second NAAT method per the discrepant analysis protocol.

Table 10: Comparison of cobas Liat® Cdiff Test with Direct Culture Results
Direct Culture Result
cobas Liat Cdiff Test ResultPositiveNegativeTotal
Detected13933172
Not Detected8833841
Total1478661013
PPA94.6% (139/147; 95% CI = 89.6% to97.2%)
NPA96.2% (833/866; 95% CI = 94.7% to97.3%)

Note: Specimens with both direct culture and valid cobas Liat Cdiff test results are considered evaluable and included in this summary table.

Note: CI = (score) confidence interval.

CONCLUSIONS 6.

A comparison of the intended use, technological characteristics, and the results of non-clinical analytical and clinical performance studies demonstrate that the cobas® Cdiff Nucleic acid test for use on the cobas "Liat" System is substantially equivalent to the predicate device.

§ 866.3130 Clostridium difficile toxin gene amplification assay.

(a)
Identification. AClostridium difficile toxin gene amplification assay is a device that consists of reagents for the amplification and detection of target sequences inClostridium difficile toxin genes in fecal specimens from patients suspected of havingClostridium difficile infection (CDI). The detection of clostridial toxin genes, in conjunction with other laboratory tests, aids in the clinical laboratory diagnosis of CDI caused byClostridium difficile. (b)
Classification. Class II (special controls). The special controls are set forth in FDA's guideline document entitled: “Class II Special Controls Guideline: Toxin Gene Amplification Assays for the Detection ofClostridium difficile; Guideline for Industry and Food and Drug Administration Staff.” See § 866.1(e) for information on obtaining this document.