K Number
K171729

Validate with FDA (Live)

Device Name
Fixtemp C&B
Date Cleared
2017-12-13

(184 days)

Product Code
Regulation Number
872.3770
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Fixtemp® C&B is a resin based material used to fabricate temporary crowns and bridges.

Device Description

Fixtemp® C&B is a two-component automatically mixable composite on the basis of multifunctional (meth)acrylates provided in standard double cartridges. The material is dispensed and mixed by cartridge/mixing tip combination and it is used for the manufacturing of temporary crowns and bridges for use until the permanent restoration is fabricated. It can be also used for inlays, onlays and veneers and is available in tooth-colors A1, A2, A3, A3.5, B1, Bleach X and D2. This is a prescription only material. The labeling and working instructions are designed for health care professionals.

AI/ML Overview

Acceptance Criteria and Device Performance for Fixtemp® C&B

The Fixtemp® C&B device is a resin-based material used to fabricate temporary crowns and bridges. Its performance was evaluated against a set of non-clinical criteria to establish substantial equivalence to a predicate device.

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicAcceptance Criteria (or Reference Standard)Reported Device Performance
Working / Processing time≥ 45 sec≥ 45 sec
Curing time in mouth (flexible phase)≥ 2 - 3 min≥ 2 - 3 min
Setting time≥ 6 min≥ 6 min
Hardness (1h)≥ 75 Shore D≥ 75 Shore D
Hardness (24h)≥ 80 Shore D≥ 80 Shore D
Compressive strength≥ 200 MPa (according to DIN EN ISO 4049:2010)≥ 200 MPa (according to DIN EN ISO 4049:2010)
Flexural strength≥ 60 MPa (according to DIN EN ISO 4049:2010)≥ 60 MPa (according to DIN EN ISO 4049:2010)
E-modulus> 1500 (according to DIN EN ISO 4049:2010)> 1500 (according to DIN EN ISO 4049:2010)
Water absorptionPass (according to DIN EN ISO 4049:2010)Pass (according to DIN EN ISO 4049:2010)
Radiopacity≥ 1.00 mm aluminium/mm material (according to DIN EN ISO 4049:2010)≥ 1.00 mm aluminium/mm material (according to DIN EN ISO 4049:2010)
CytotoxicityNo biological reactivity (Grade 0) based on ISO 10993-5No biological reactivity (grade 0) of the cells exposed to the test article. (Test report: "Toxikon Europe Final GLP Report: 17-01389-G1")
Carcinogenic potentialNo evidence for causing cancer within the time period of use (up to 29 days)Different evaluations and assessments have been made for cured Fixtemp® C&B and they have shown that there is no evidence for causing cancer within the time period of use in the patients mouth for up to 29 days.
Mutagenic potentialNo evidence for mutagenic reactions within the time period of use (up to 29 days)Different evaluations and assessments have been made for cured Fixtemp® C&B and they have shown that there is no evidence for mutagenic reactions within the time period of use in the patients mouth for up to 29 days.
Irritating potentialNo evidence for irritation within the time period of use (up to 29 days)Different evaluations and assessments have been made for cured Fixtemp® C&B and they have shown that there is no evidence for irritation within the time period of use in the patients mouth for up to 29 days.
Sensitizing potentialNo evidence for sensitization within the time period of use (up to 29 days)Different evaluations and assessments have been made for cured Fixtemp® C&B and they have shown that there is no evidence for sensitization within the time period of use in the patients mouth for up to 29 days.
Shelf life2 yearsThe results of stability tests at 23 °C/37 °C justify to assume a shelf life of 2 years for Fixtemp® C&B.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific sample sizes used for each individual test (e.g., number of specimens for compressive strength). It generally refers to "a series of in-house tests" and adherence to specific ISO standards (e.g., DIN EN ISO 4049:2010).

For the cytotoxicity testing, the test was conducted by the independent laboratory Toxikon Europe. The provenance of the data is indicated as "Toxikon Europe Final GLP Report: 17-01389-G1". The country of origin for the manufacturing company (Dreve Dentamid GmbH) is Germany. It is implied that these are prospective tests performed on the Fixtemp® C&B material.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not provided in the document. The performance evaluation relies on standardized tests (e.g., ISO standards) and laboratory findings rather than expert consensus on a test set in the typical sense of diagnostic imaging or clinical studies. The cytotoxicity test was conducted by an "independent laboratory," but the qualifications of the individuals establishing the ground truth (i.e., interpreting the cytotoxicity results) are not detailed.

4. Adjudication Method for the Test Set

Not applicable. The reported tests are objective measurements based on standardized procedures and laboratory analysis, not subjective interpretations requiring adjudication by multiple experts.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done. This type of study is typically performed for diagnostic devices where human readers interpret medical images or data. Fixtemp® C&B is a material for fabricating dental prosthetics, and its performance is evaluated through material properties and biocompatibility, not reader interpretation.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Not applicable. Fixtemp® C&B is a dental material, not an algorithm or software. Its performance is inherent to its physical and chemical properties.

7. The Type of Ground Truth Used

The ground truth for the performance tests consists of:

  • Standardized measurement directly from testing: This applies to properties like working time, curing time, setting time, hardness, compressive strength, flexural strength, E-modulus, water absorption, and radiopacity, where the "ground truth" is a measured physical or chemical property that meets or exceeds a predefined numerical threshold from an established standard (e.g., DIN EN ISO 4049:2010).
  • Laboratory findings: For cytotoxicity, the ground truth is established by the results of the ISO 10993-5 test conducted by an independent laboratory, indicating the absence of biological reactivity.
  • Evaluations and assessments: For carcinogenic, mutagenic, and irritating/sensitizing potentials, the ground truth is based on "different evaluations and assessments" indicating no evidence of these effects within the specified time frame. While the specifics of these evaluations are not detailed, they likely rely on established toxicological principles and potentially other in-vitro or in-vivo studies not explicitly listed.

8. The Sample Size for the Training Set

Not applicable. Fixtemp® C&B is a physical material, not an AI model or software that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

4/10/2018

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Dreve Dentamid GmbH Reiner Altmann Head of Quality Management & Regulatory Affairs Max-Planck-Strasse 31 Unna, DE 59423 NRW

Re: K171729

Trade/Device Name: Fixtemp C&B Regulation Number: 21 CFR 872.3770 Regulation Name: Temporary crown and bridge resin Regulatory Class: Class II Product Code: EBG Dated: November 17, 2017 Received: November 20, 2017

Dear Reiner Altmann:

This letter corrects our substantially equivalent letter of December 13, 2017.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

Page 2 - Reiner Altmann

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mary S. Runner -S

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171729

Device Name Fixtemp® C&B

Indications for Use (Describe)

Fixtemp® C&B is a resin based material used to fabricate temporary crowns and bridges.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image features a red circle with the word "Dreve" in white letters at the center. Below the circle, the text "Dreve Dentamid" is displayed in a larger, bold font. The red circle has a white border, providing contrast against the background.

• Submitter:Dreve Dentamid GmbHMax-Planck-Straße 31Unna, Nordrhein-Westfalen, 59423GermanyEstablishment Reg. No. 1000486347
• Contact:Dr. Reiner Altmann(Head of Quality Management andRegulatory Affairs)Phone: +49-2303-88070E-mail: reiner.altmann@dreve.de
• Official Correspondent:Mr. Michael Breer(Quality Management Representative)
• Date PreparedDecember 12th, 2017
• Common Name:Temporary C&B resin
• Trade/Device Name:Fixtemp® C&B
• Model No.:D60391, D60392, D603925, D60393, D603935, D60394, D60395, D60396, D60399
• Classification Name:Crown and bridge, Temporary, Resin
• Device Classification:Class 2 per 21 CFR 872.3770
• Product Code:EBG

Predicate Device:

K102917 FixTemp C&B™

Device Description

Fixtemp® C&B is a two-component automatically mixable composite on the basis of multifunctional (meth)acrylates provided in standard double cartridges. The material is dispensed and mixed by cartridge/mixing tip combination and it is used for the manufacturing of temporary crowns and bridges for use until the permanent restoration is fabricated. It can be also used for inlays, onlays and veneers and is available in tooth-colors A1, A2, A3, A3.5, B1, Bleach X and D2. This is a prescription only material. The labeling and working instructions are designed for health care professionals.

Image /page/3/Picture/7 description: The image contains two logos stacked on top of each other. The top logo is a circular seal with the text "TUV NORD" at the top and "ISO 9001 / ISO 13485" at the bottom. The bottom logo is a circular design with a square shape attached to the right side. Below the logos, the text "Deutsche-Dental-Industrie" is written.

Dreve Dentamid GmbH Max-Planck-Straße 31 59423 Unna/Germany Tel.: +49 2303 8807-0 Fax: +49 2303 8807-55 E-Mail: info@dreve.de

Internet: www.dreve.com Lieferadresse: Einsteinstr. 36 Geschäftsführer: Dr. Volker Dreve Sitz: 59423 Unna Reg.-Gericht: Hamm HRB 3712

Sparkasse Unna Kto.-Nr. 43 000 BLZ 443 500 60 BIC: WELADED1UNN IBAN: DE 26 44350060 0000043000

Commerzbank AG Dortmund Kto.-Nr. 03 222 354 00 BLZ 440 800 50 BIC: DRESDEFF440 IBAN: DE 35 44080050 0322235400

Page 1 of 5

Postbank Dortmund Kto .- Nr. 6 493 469 BLZ 440 100 46 BIC: PBNKDEFF IBAN: DE 42 44010046 0006493469

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a red circle with a white border. Inside the circle, the word "Dreve" is written in white, sans-serif font. The background is white.

Dreve Dentamid

The associated accessories include mixing tips as follows:

  • Trade/Device Name: Mixing tips .
  • Product code: . EID
  • Classification name: . Syringe, restorative and impression material
    • Device classification: Class 1 as per 21 CFR 872,4565
  • . Listing No.: D244388

Indications for Use:

e

Fixtemp® C&B is a resin based material used to fabricate temporary crowns and bridges.

Substantial Equivalence

Information provided in this application shows that the product is substantially equivalent to the predicate device in indications for use, performance, materials and application. The main difference between these devices is the mixing ratio but the Fixtemp® C&B is substantial equivalent as shown in the substantial equivalence comparison. The difference in the mixing ratio does not alter the intended use of the device.

Technological Characteristics

Design

The Fixtemp® C&B of Dreve is similar in design to the predicate listed above. Same as the predicate the Fixtemp® C&B is intended to manufacture temporary crowns and bridges, as well as inlays, onlays and veneers. They use similar technological characteristics and principles. Both materials are two component materials being mixed to start the curing process.

Material

Same as the predicate material, the Dreve Fixtemp® C&B is based on multifunctional (meth)acrylates, glass powder, aerosil, polyester and catalyst. An assessment of the biocompatibility according to FDA Recognized Consensus Standard DIN EN ISO 7405 / ISO 10993-1 is included in this application. As a result of this assessment/testing we conclude that the device is substantial equivalent.

Some chemicals are part of Fixtemp® C&B (Dreve) but not available in the predicate`s composition, but also used in other devices of product code EBG for crown and bridge, temporary, resin like K122039, K033022, K042820 and K013674.

Image /page/4/Picture/17 description: The image is a circular logo with the text "TUV NORD" at the top. Below that, it says "TUV NORD CERT GmbH". The logo also contains the text "ISO 9001 / ISO 13485" around the bottom of the circle. The top left quadrant of the circle is black, while the rest is white.

B V D

Kooperatives Mitglied im
Bundesverband Dentalhandel e.V.

F-Mail: dentamid@dreve.de

nternet: www.dreve.com/dentami ieferadresse: Einsteinstr. 36 Geschäftsführer: Dr. Volker Dreve, Dino Marchetti Sitz: 59423 Unna Reg.-Gericht: Hamm HRB 3712

to .- Nr. 43 000 0000043000

Commerzbank AG Dortmun
Kto.-Nr. 03 222 354 00
BLZ 440 800 50
BIC: DRESDEFF440
IBAN: DE 35 44080050
0322235400

Page 2 of 5

DANI. DE 47 44010046

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a red circle with a white border. Inside the circle, the word "Drewe" is written in white, bold letters. The background is white.

Testing

Dreve Dentamid

The Fixtemp® C&B has benefited from design, development, testing and production procedures that are being certified according to ISO 13485; CAN/CSA ISO 13485 and European Medical Device Directive 93/42/EEC by a Notified Body.

Testing has confirmed that this device meets its product specification. A series of in-house tests have been conducted to verify the intended signals are accurate and can maintain performance over its useful life. The testing also demonstrated that the specification is substantial equivalent as the predicate and for details please refer to the substantial equivalence comparison.

Furthermore the material is:

  • patient-contacting o
  • non-sterile .

Summary non-clinical performance data

CHARACTERISTICSComments
Materials1Mixture of multifunctional (meth)acrylates, glass powder, aerosil, polyester and catalyst.The used raw materials are state-of-the-art and are notknown to have unacceptable risks or dangers when used as ingredients.
Chemical DescriptionSelf-curing resin (radical polymerization)
Method ofmanipulationImpression or thermoforming blank
Flow propertiesPaste-like liquid
Working /Processing time≥ 45 sec
Curing time in themouth-flexiblephase≥ 2 - 3 min
Setting time≥ 6 min
Hardness1h: ≥ 75 Shore D24h: ≥ 80 Shore D
Compressivestrength≥ 200 MPa according to DIN EN ISO 4049:2010.
Flexural strength≥ 60 MPa according to DIN EN ISO 4049:2010.
E-modulus> 1500 according to DIN EN ISO 4049:2010
Water absorptionPass according to DIN EN ISO 4049:2010.
Radiopacity≥ 1.00 mm aluminium/mm material according to DIN ENISO 4049:2010.
Safety - toxicThe cytotoxicity of cured Fixtemp® C&B was tested basedon ISO 10993-5 by the independent laboratory Toxikon.The potential biological reactivity of a mammalian cellculture (mouse fibroblasts L929) in response to the testarticle Fixtemp® C&B was determined. There was nobiological reactivity (grade 0) of the cells exposed to thetest article.
The test item Fixtemp® C&B (Lot# FE140814-27C -601606x0) was considered to have no cytotoxic potentialand to meet the requirements of ISO 10993-5. For detailsplease refer to the test report "Toxikon Europe Final GLPReport: 17-01389-G1" in section B11.
Safety - carcinogenicDuring the application uncured Fixtemp® C&B has directcontact with the patients' dentin and mucosal membrane.The biological risk during the application is negligiblebecause of the short intraoral time < 3min.
Different evaluations and assessments have been made forcured Fixtemp® C&B and they have shown that there is noevidence for causing cancer within the time period of usein the patients mouth for up to 29 days.
Safety - mutagenicDuring the application uncured Fixtemp® C&B has directcontact with the patients' dentin and mucosal membrane.The biological risk during the application is negligiblebecause of the short intraoral time < 3min.
Different evaluations and assessments have been made forcured Fixtemp® C&B and they have shown that there is noevidence for mutagenic reactions within the time period ofuse in the patients mouth for up to 29 days.
Safety - irritatingDuring the application uncured Fixtemp® C&B has directcontact with the patients' dentin and mucosal membrane.The biological risk during the application is negligiblebecause of the short intraoral time < 3min.
Different evaluations and assessments have been made forcured Fixtemp® C&B and they have shown that there is noevidence for irritation within the time period of use in thepatients mouth for up to 29 days.

Image /page/5/Picture/10 description: The image contains two logos. The first logo is a circular seal with the text "TUV NORD" at the top and "ISO 9001 / ISO 13485" at the bottom. The second logo is a stylized design featuring a circle with a geometric shape inside, resembling a stylized letter or symbol.

Image /page/5/Picture/11 description: The image shows a logo for "Deutsche-Dental-Industrie". Below the text are three vertical lines that are curved at the top and bottom. The letters "B", "V", and "D" are placed between the lines. Below the lines is the text "Kooperatives Mitglied im".

Dreve Dentamid GmbH Max-Planck-Straße 31 59423 Unna/Germany Tel .: +49 2303 8807-0 Fax: +49 2303 8807-55 E-Mail: info@dreve.de

B6.

Internet: www.dreve.com Lieferadresse: Einsteinstr. 36 Geschäftsführer: Dr. Volker Dreve Sitz: 59423 Unna Reg - Gericht: Hamm HRB 3712

Sparkasse Unna Kto.-Nr. 43 000 BI 7 443 500 60 BIC: WELADED 1UNN 1BAN: DE 26 44350060 0000043000

1 For a detailed formulation for Fixtemp® C&B of Dreve Dentamid please refer to table 1 Section

Commerzbank AG Dortmund Kto -Nr. 03 222 354 00 BLZ 440 800 50 BIC: DRESDEFF440 IBAN: DE 35 44080050 0322235400

Page 3 of 6

Postbank Dortmund Kto .- Nr. 6 493 469 BLZ 440 100 46 BIC: PBNKDEFF IBAN: DE 42 44010046 0006493469

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image is a red circle with a white border. Inside the circle, the word "Dreve" is written in white, sans-serif font. The word is centered horizontally and vertically within the circle. The background of the image is plain white.

Dreve Dentamid

Image /page/6/Picture/3 description: The image shows two logos. The top logo is the "TUV NORD" certification mark, indicating compliance with ISO 9001 and ISO 13485 standards. The bottom logo is for "Deutsche Dental-Industrie", featuring a stylized circle and a square with a triangle cut out, representing the German dental industry association.

Dreve Dentamid GmbH Max-Planck-Straße 31 59423 Unna/Germany Tel.: +49 2303 8807-0 Fax: +49 2303 8807-55 E-Mail: info@dreve de

Internet: www.dreve.com Lieferadresse: Einsteinstr. 36 Geschäftsführer: Dr. Volker Dreve Sitz: 59423 Unna Reg.-Gericht: Hamm HRB 3712 Sparkasse Unna Kto.-Nr. 43 000 BLZ 443 500 60 BIC: WELADED1UNN IBAN: DE 26 44350060 0000043000

Commerzbank AG Dortmund Kto.-Nr. 03 222 354 00 BLZ 440 800 50 BIC: DRESDEFF440
IBAN: DE 35 44080050 0322235400

Postbank Dortmund
Kto.-Nr. 6 493 469
BLZ 440 100 46 BIC: PBNKDEFF IBAN: DE 42 44010046 0006493469

Page 4 of 5

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image is a red circle with the word "Dreve" written in white letters in the center. The circle has a white border around it. The background of the image is white.

Dreve Dentamid
--------------------
Safety - sensitizingDuring the application uncured Fixtemp® C&B has directcontact with the patients' dentin and mucosal membrane.The biological risk during the application is negligiblebecause of the short intraoral time < 3min.
Different evaluations and assessments have been made forcured Fixtemp® C&B and they have shown that there is noevidence for sensitization within the time period of use inthe patients mouth for up to 29 days.
Keeping (storage)quantitiesThe results of stability tests at 23 °C/37 °C justify to assumea shelf life of 2 years for Fixtemp® C&B.
DurationShort term contact < 30 days
Type of contactoral cavity as far as the pharynx
Conformance toconsensus standardEN ISO 13485:2012
ISO 7405:2008
ISO 10993-1:2009
ISO 14971:2007

Conclusion

There are no substantial differences between the Fixtemp® C&B defined in this 510(k) submission and other legally marketed devices in the United States. The device is substantially equivalent to predicate device FixTemp C&B™ listed by company Exacta Dental Direct. The materials function similar and have the same indications for use.

Same as the predicate the Fixtemp® C&B is state-of-the-art resin based material used to fabricate temporary crowns and bridges by intra-oral application and extra-oral final polymerization.

Dreve Dentamid GmbH

Signature

i.A. Dr. Reiner Altmann (Head of Quality Management / Quality Control / Regulatory Affairs & Safety Representative for Medical Devices)

Image /page/7/Picture/9 description: The image shows two logos stacked on top of each other. The top logo is a circular seal with the text "TUV NORD" and "ISO 9001 / ISO 13485". The bottom logo is a circular design with a square shape attached to the right side. Below the logos, the text "Deutsche-Dental-Industrie" is visible.

Dreve Dentamid GmbH Max-Planck-Straße 31 59423 Unna/Germany Tel.: +49 2303 8807-0 Fax: +49 2303 8807-55 E-Mail: info@dreve.de

Internet: www.dreve.com Lieferadresse: Einsteinstr. 36 Geschäftsführer: Dr. Volker Dreve Sitz: 59423 Unna Reg.-Gericht: Hamm HRB 3712

Sparkasse Unna Kto .- Nr. 43 000 BLZ 443 500 60 BIC: WELADED1UNN IBAN: DE 26 44350060 0000043000

Commerzbank AG Dortmund Kto.-Nr. 03 222 354 00 BLZ 440 800 50 BIC: DRESDEFF440 IBAN: DE 35 44080050 0322235400

Page 5 of 5

Postbank Dortmund Kto .- Nr. 6 493 469 BLZ 440 100 46 BIC: PBNKDEFF IBAN: DE 42 44010046 0006493469

§ 872.3770 Temporary crown and bridge resin.

(a)
Identification. A temporary crown and bridge resin is a device composed of a material, such as polymethylmethacrylate, intended to make a temporary prosthesis, such as a crown or bridge, for use until a permanent restoration is fabricated.(b)
Classification. Class II.