K Number
K171689

Validate with FDA (Live)

Date Cleared
2017-10-05

(120 days)

Product Code
Regulation Number
888.3080
Age Range
All
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ARTiC-L™ 3D Ti Spinal System with TiONIC™ Technology is indicated for use as an intervertebral body fusion device in skeletally mature patients with degenerative disc disease (DDD - defined by discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine (L2-S1). Additionally, ARTiC-L™ 3D Ti Spinal System can be used in patients diagnosed with spinal deformities as an adjunct to fusion. These patients should be skeletally mature and have undergone 6 months of nonoperative treatment prior to surgery. These implants are used to facilitate fusion in the lumbar spine using autogenous bone and/or allogenic bone graft comprised of cancellous bone graft. When used as an interbody fusion device, these implants are intended for use with supplemental internal fixation systems.

The ARTiC-XL™ 3D Ti Spinal System with TiONIC™ Technology is indicated for use as an intervertebral body fusion device in skeletally mature patients with degenerative disc disease (DDD - defined by discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine (L2-S1). Additionally, ARTIC-XL™ 3D Ti Spinal System can be used in patients diagnosed with spinal deformities as an adjunct to fusion. These patients should be skeletally mature and have undergone 6 months of nonoperative treatment prior to surgery. These implants are used to facilitate fusion in the lumbar spine using autogenous bone and/or allogenic bone graft comprised of cancellous bone graft. When used as an interbody fusion device, these implants are intended for use with supplemental internal fixation systems.

Device Description

The ARTIC-L™ and ARTIC-XL™ 3D Ti Spinal Systems with TiONIC™ Technology consist of additively manufactured (AM) titanium spacers of various widths, lengths, heights and lordotic angles and reusable instruments used for implantation and extraction of the subject implant. The instruments may be transported to the customer in cases, trays and lids which can also be used for sterilization. These implants are intended to be surgically placed between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar intervertebral body fusion surgical procedures.

AI/ML Overview

This document describes the regulatory submission for the Medtronic ARTiC-L™ 3D Ti Spinal System and ARTiC-XL™ 3D Ti Spinal System with TiONIC™ Technology. The submission is a 510(k) premarket notification, which demonstrates substantial equivalence to legally marketed predicate devices.

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria for this device are based on demonstrating substantial equivalence to predicate devices, primarily through mechanical performance, biocompatibility, and MRI safety. The document doesn't present a table with explicit numerical acceptance criteria values alongside reported device performance. Instead, it states that the tests were performed "in accordance with" relevant ASTM standards and guidance documents, implying that the acceptance criteria are adherence to these established standards and comparable performance to predicates.

The reported device performance is presented as:

Test CategoryAcceptance Criteria (Implied)Reported Device Performance
BiocompatibilityAdherence to "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process" (FDA Guidance, June 16, 2016).- Evaluated for permanent, >30-day body contact for implants. - Materials: Titanium-6 Aluminum-4 Vanadium ELI. - Tests performed: Chemical Characterization, Toxicological Risk Assessment, Cytotoxicity, Acute Systemic Toxicity, Material Mediated Pyrogenicity. - Results: "Based on the results, no additional testing was required." (Implies successful completion and adherence to safety profiles). - Bacterial Endotoxin (LAL) test: Successfully met 20 EU/device pyrogen limit specification (as per ANSI/AAMI ST72 and USP <161>). - Instruments (non-sterile, external communicating for <24 hr contact): Materials are Stainless Steel (ASTM F899, A564, A693) and Titanium Alloy ELI (ASTM F136). "No new biocompatibility testing is required" due to use of same materials as predicate.
Mechanical TestingAdherence to "Guidance for Industry and FDA Staff – Spinal System 510(k)'s" and specific ASTM standards (ASTM F2077-14, F2267-04, F-04.25.02.02 Draft, F1877-16).- Bench testing completed to demonstrate substantial equivalence to predicates. - Performed tests: Static Compression, Static Compression-Shear, Compression Fatigue, Compression-Shear Fatigue (all per ASTM F2077-14). - Load-Induced Subsidence (ASTM F2267-04). - Static Push-out (ASTM Draft F-04.25.02.02). - Particle Characterization (ASTM F1877-16). - Results: "the results of these tests demonstrated the substantial equivalence of the subject devices to the predicates."
MRI SafetyAdherence to specific ASTM standards (ASTM F2052, F2213, F2119, F2182).- Performed tests: Measurement of magnetically induced displacement force (ASTM F2052), magnetically induced torque (ASTM F2213), MR image artifacts (ASTM F2119), radio frequency induced heating (ASTM F2182). - Results: Not explicitly stated as "passed" for each, but the overall conclusion states the device is "as safe and effective as the following predicates" which implicitly includes MRI safety.
Design ValidationUser needs met and intended function demonstrated.- Performed using saw bone models and/or cadaver lab (Tru-Trainer). - Results: "Design Validation demonstrated the subject implants and instruments function as intended and user needs were met."

2. Sample Size and Data Provenance

The document does not specify exact sample sizes for the mechanical tests, biocompatibility tests, or design validation beyond referring to the standards by which they were conducted. For mechanical testing, the standards (e.g., ASTM F2077-14) would typically define minimum sample sizes.

The data provenance is not explicitly stated in terms of country of origin. The submission is made by Medtronic Sofamor Danek, USA Inc., and the FDA is a US regulatory body, suggesting the studies were likely conducted in the US or in adherence to US regulatory expectations. The studies described are pre-market design validation and verification activities, inherently prospective in their nature as they involve testing the medical device itself. They are not clinical studies on patient populations.

3. Number of Experts and Qualifications for Ground Truth

This submission does not involve AI or algorithms that require human expert assessment for "ground truth" in the typical sense of image interpretation or diagnostic accuracy. The "ground truth" here is established through bench testing against engineering standards and material properties, and validation through physical models (saw bones/cadavers). Therefore, the concept of "number of experts," their qualifications, or adjudication methods for establishing "ground truth" (as it relates to diagnostic accuracy) is not applicable to this type of device submission. Expertise would lie within the engineering and scientific teams conducting the tests and interpreting the physical and chemical properties.

4. Adjudication Method

Not applicable, as described above.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. This is a spinal implant device, not an AI-assisted diagnostic tool. Therefore, an MRMC study comparing human reader performance with and without AI assistance is not relevant to this submission.

6. Standalone (Algorithm only) Performance

Not applicable, as there is no standalone algorithm in this context. The device is a physical implant.

7. Type of Ground Truth Used

The ground truth used for this device is based on:

  • Engineering standards and material specifications: For biocompatibility, mechanical strength, and MRI safety, adherence to established ASTM and ISO standards defines acceptable performance.
  • Physical models/Cadaveric studies: For design validation, including proper fit, function, and surgical feasibility, using saw bones and cadaver models serves as functional ground truth in a pre-clinical setting.

8. Sample Size for Training Set

Not applicable. This is a physical medical device, not a machine learning model, so there is no "training set."

9. How the Ground Truth for the Training Set was Established

Not applicable, as there is no training set.

{0}------------------------------------------------

Medtronic Sofamor Danek, USA Inc. Mr. Tejas Patel Senior Regulatory Affairs Specialist 1800 Pvramid Place Memphis, Tennessee 38132

October 5, 2017

Re: K171689

Trade/Device Name: ARTiC-L™ 3D Ti Spinal System with TiONIC Technology and ARTiC-XL™ 3D Ti Spinal System with TiONIC Technology Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: September 1, 2017 Received: September 5, 2017

Dear Mr. Patel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.qov

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Katherine D. Kavlock -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171689

Device Name

ARTiC-LTM 3D Ti Spinal System with TiONICTM Technology

Indications for Use (Describe)

The ARTiC-L™ 3D Ti Spinal System with TiONIC™ Technology is indicated for use as an intervertebral body fusion device in skeletally mature patients with degenerative disc disease (DDD - defined by discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine (L2-S1). Additionally, ARTiC-L™ 3D Ti Spinal System can be used in patients diagnosed with spinal deformities as an adjunct to fusion. These patients should be skeletally mature and have undergone 6 months of nonoperative treatment prior to surgery. These implants are used to facilitate fusion in the lumbar spine using autogenous bone and/or allogenic bone graft comprised of cancellous bone graft. When used as an interbody fusion device, these implants are intended for use with supplemental internal fixation systems.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K171689

Device Name

ARTiC-XL™ 3D Ti Spinal System with TiONIC™ Technology

Indications for Use (Describe)

The ARTiC-XL™ 3D Ti Spinal System with TiONIC™ Technology is indicated for use as an intervertebral body fusion device in skeletally mature patients with degenerative disc disease (DDD - defined by discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine (L2-S1). Additionally, ARTIC-XL™ 3D Ti Spinal System can be used in patients diagnosed with spinal deformities as an adjunct to fusion. These patients should be skeletally mature and have undergone 6 months of nonoperative treatment prior to surgery. These implants are used to facilitate fusion in the lumbar spine using autogenous bone and/or allogenic bone graft comprised of cancellous bone graft. When used as an interbody fusion device, these implants are intended for use with supplemental internal fixation systems.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY

MEDTRONIC Sofamor Danek ARTiC-L™ 3D Ti Spinal System ARTiC-XL™ 3D Ti Spinal System with TiONIC ™ Technology

September 28, 2017

Submitter:Medtronic Sofamor Danek, USA Inc.
1800 Pyramid Place
Memphis, Tennessee 38132
Telephone: (901)396-3133
Fax: (901) 346-9738
Tejas Patel
Contact PersonSr. Regulatory Affairs Specialist
Telephone: (901) 344-1443 (Direct)
Alternate Contact:Raphael McInnis
Regulatory Affairs Manager
Telephone: (901) 399-2057 (Direct)
Date PreparedSeptember 28, 2017
Name of DeviceARTiC-L™ 3D Ti Spinal System and ARTiC-XL™ 3D Ti Spinal System with
TiONIC ™ Technology
Common NameInterbody fusion device
ClassificationIntervertebral Body Fusion Device with bone graft
Name(21 CFR 888.3080)
Regulatory ClassClass II
Product CodeMAX
Predicate 1 (Primary) - DePuy Synthes T-PAL Spacer System
K162358 SE 11/01/2016
Predicate DevicesPredicate 2 CRESCENT® Ti
K122037 SE 03/22/2013
Predicate 3 K2M CASCADIA™ TL Interbody System
K160547 SE 03/24/2016
Predicate 4 NuVasive® Brigade® Hyperlordotic System
K123045 SE 04/16/2013
Predicate 5 CLYDESDALE® Spinal System
K151128 SE 08/06/2015
Predicate 6 WAVE-D Spinal System
K121333 SE 06/29/2012
Predicate 7 CAPSTONE® PEEK
K073291 SE 04/24/2008, K123027 SE 07/25/2013
Predicate 8 CAPSTONE® PTC
K133205 SE 03/13/2014
Predicate 9 CRESCENT® Spinal System
K094025 SE 04/26/2010
Predicate 10 LOOP® Spinal SystemK121332 SE 06/29/2012
Description ofDeviceThe ARTIC-L™ and ARTIC-XL™ 3D Ti Spinal Systems with TiONIC™
Technology consist of additively manufactured (AM) titanium spacers of
various widths, lengths, heights and lordotic angles and reusable instruments
used for implantation and extraction of the subject implant. The instruments
may be transported to the customer in cases, trays and lids which can also be
used for sterilization. These implants are intended to be surgically placed
between two lumbar or lumbosacral vertebral bodies to give support and
correction during lumbar intervertebral body fusion surgical procedures.
ARTIC-L™ 3D Ti Spinal System with TiONIC™ Technology
The ARTIC-L™ 3D Ti Spinal System with TiONIC™ Technology is indicated
for use as an intervertebral body fusion device in skeletally mature patients
with degenerative disc disease (DDD - defined by discogenic back pain with
degeneration of the disc confirmed by patient history and radiographic studies)
at one or two contiguous levels of the lumbar spine (L2-S1). Additionally,
ARTIC-L™ 3D Ti Spinal System can be used in patients diagnosed with spinal
deformities as an adjunct to fusion. These patients should be skeletally mature
and have undergone six months of non-operative treatment prior to surgery.
These implants are used to facilitate fusion in the lumbar spine using
autogenous bone and/or allogenic bone graft comprised of cancellous and/or
cortico-cancellous bone graft. When used as an interbody fusion device these
Indications for Useimplants are intended for use with supplemental internal fixation systems.
ARTIC-XL™ 3D Ti Spinal System with TiONIC™ Technology
The ARTIC-XL™ 3D Ti Spinal System with TiONIC™ Technology is indicated
for use as an intervertebral body fusion device in skeletally mature patients
with degenerative disc disease (DDD - defined by discogenic back pain with
degeneration of the disc confirmed by patient history and radiographic studies)
at one or two contiguous levels of the lumbar spine (L2-S1). Additionally,
ARTIC-XL™ 3D Ti Spinal System with TiONIC ™ Technology can be used in
patients diagnosed with spinal deformities as an adjunct to fusion. These
patients should be skeletally mature and have undergone six months of non-
operative treatment prior to surgery. These implants are used to facilitate fusion
in the lumbar spine using autogenous bone and/or allogenic bone graft
comprised of cancellous and/or cortico-cancellous bone graft. When used as an
interbody fusion device these implants are intended for use with supplemental
internal fixation systems.
Comparison ofTechnologicalCharacteristics withthe PredicateDevices:ARTiC-L™ 3D Ti Spinal System and ARTiC-XL™ 3D Ti Spinal System with
TiONIC TM Technology has the same fundamental scientific technology;
indications for use, intended use, design, material levels of attachment as the
predicate devices. The predicate and subject devices are intended for
stabilization use and to promote bone fusion during the normal healing process
following surgical correction of disorders of the spine.
The following performance data were provided in support of substantial
equivalence.
Biocompatibility
The biocompatibility evaluation for the ARTiC-L™ 3D Ti Spinal System and
ARTiC-XL™ 3D Ti Spinal System implants was conducted in accordance with
FDA's Guidance for Industry and FDA Staff "Use of International Standard
ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and
Testing within a Risk Management Process" issued June 16, 2016.
The subject implants are temporary implants and will be classified as
Performance Data:permanent, >30-day body contact according to with FDA's Guidance for
Industry and FDA Staff "Use of International Standard ISO-10993, Biological
Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk
Management Process" issued June 16, 2016.
The subject implants are manufactured from Titanium-6 Aluminum-4
Vanadium ELI (Extra Low Interstitial) Alloy. The following tests have been
conducted to ensure biocompatibility:
Chemical Characterization●
Toxicological Risk Assessment●
Cytotoxicity●
Acute Systemic Toxicity●

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

Material Mediated Pyrogenicity ● Based on the results, no additional testing was required.

The bacterial endotoxin test, also known as Limulus amebocyte lysate (LAL) test, was performed utilizing worst case ARTiC-L™ 3D Ti Spinal System and ARTiC-XL" 3D Ti Spinal System implants to verify that the subject implants meet the 20 endotoxin units (EU)/device pyrogen limit specification. Testing was successfully performed and it was confirmed that the subject implants meet the 20 EU/device testing limit for general medical devices that are implanted as outlined in ANSI/AAMI ST72: Bacterial endotoxins - Test methods, routine monitoring, and alternatives to batch testing and USP <161>, Medical Devices – Bacterial Endotoxin and Pyrogen Tests.

The subject instruments are external communicating devices and are classified as limited, up to 24 hours of body contact according to with FDA's Guidance for Industry and FDA Staff "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process" issued June 16, 2016. The instruments are manufactured from Stainless Steel, in accordance with ASTM F899, ASTM A564 and ASTM A693 and Titanium Alloy ELI, in accordance with F136. The subject instruments are provided non-sterile and are intended for steam sterilization by the end user prior to use. The subject instruments can be packaged and shipped in case, trays and lids, the subject instruments can be steam sterilized in these case, trays, and lids.

  • . ASTM F899 Standard Specification for Wrought Stainless Steels for Surgical Instruments
  • ASTM A564 Standard Specification for Hot Rolled and Cold Finished Age Hardening Stainless Steel Bars and Shapes
  • . ASTM A693 Standard Specification for Precipitation-Hardening Stainless and Heat Resisting Steel Plate, Sheet and Strip
  • ASTM F136: Standard Specification for Wrought Titanium-● 6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications.

The system specific tray used for shipment and sterilization of instruments is

{8}------------------------------------------------

manufactured from stainless steel, aluminum, nylon coated stainless steel, radel
polypropylene with the brackets securing the instruments into the case/tray
made of silicone and/or nylon coated stainless steel and/or polypropylene.
These materials are non-patient contacting and indirect patient contacting and
are the same materials as the predicate. Therefore, no new biocompatibility
testing is required.

Mechanical Testing

In accordance with, Guidance for Industry and FDA Staff – Spinal System 510(k)'s", Medtronic has evaluated the subject devices to demonstrate substantial equivalence to the predicate devices. A combination of engineering rationales and testing were used to establish substantial equivalence. The following bench testing was completed and the results of these tests demonstrated the substantial equivalence of the subject devices to the predicates.

Performance testing was completed in accordance with:

  • ASTM F2077-14 Test Methods for Intervertebral Body Fusion Devices
    Following mechanical tests were performed per ASTM F2077-14:

  • Static Compression Testing

  • A Static Compression-Shear Testing

  • Compression Fatigue Testing

  • Compression-Shear Fatigue Testing

  • ASTM F2267-04 (2011) Standard Test Method for Measuring Load ● Induced Subsidence of Intervertebral Body Fusion Device Under Static Axial Compression

  • ASTM Draft Standard F-04.25.02.02: Static Push-out Test Method for Intervertebral Body Fusion Devices

  • ASTM F1877-16: Standard Practice for Characterization of Particle ●

MRI Safety Evaluation was completed in accordance with:

  • . ASTM F2052:2015: Standard test method for measurement of magnetically induced displacement force on passive implants in the magnetic resonance environment
  • ASTM F2213:2006 (R11): Standard test method for measurement of magnetically induced torque on medical devices in the magnetic resonance environment

{9}------------------------------------------------

Conclusion:ASTM F2119:2007(R13) : Standard test method for evaluation of MR image artifacts from passive implants ASTM F2182:2002a, 2011, 2011a : Standard test method for measurement of radio frequency induced heating on or near passive implant during magnetic resonance imaging
Design Validation
Design Validation was performed using saw bone models and/or cadaver lab (Tru-Trainer) for the subject ARTiC-LTM 3D Ti Spinal System and ARTiC-XLTM 3D Ti Spinal System implants and instruments. Design Validation demonstrated the subject implants and instruments function as intended and user needs were met.
Based on the test results and additional supporting documentation provided in the pre-market notification, the subject ARTiC-LTM 3D Ti Spinal System and ARTiC-XLTM 3D Ti Spinal System are as safe and effective as the following predicates:
Predicate 1 (Primary) – DePuy Synthes T-PAL Spacer SystemK162358 SE 11/01/2016
Predicate 2 CRESCENT® TiK122037 SE 03/22/2013
Predicate 3 K2M CASCADIATM TL Interbody SystemK160547 SE 03/24/2016
Predicate 4 NuVasive® Brigade® Hyperlordotic SystemK123045 SE 04/16/2013
Predicate 5 CLYDESDALE® Spinal SystemK151128 SE 08/06/2015
Predicate 6 WAVE-D Spinal SystemK121333 SE 06/29/2012
Predicate 7 CAPSTONE® PEEKK073291 SE 04/24/2008, K123027 SE 07/25/2013
Predicate 8 CAPSTONE® PTCK133205 SE 03/13/2014
Predicate 9 CRESCENT® Spinal SystemK094025 SE 04/26/2010
Predicate 10 LOOP® Spinal SystemK121332 SE 06/29/2012

T

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.