K Number
K171602

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-07-28

(57 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Williams Cystoscopic Injection Needle is used for cystoscopic injection into the urethra, bladder neck, and bladder wall.

Device Description

The Williams Cystoscopic Injection Needle is a semi-rigid injection needle. It is provided in 3.7 or 5.0 French sizes with 35 or 45 centimeters long polyurethane catheter, which ends with a 7 or 8 millimeter length stainless steel beveled tip cannula that is available in 20 to 25 needle gages. The subject device is sterilized with EtO and is intended for one-time use.

AI/ML Overview

The provided text describes a medical device, the Williams Cystoscopic Injection Needle, and its substantial equivalence application (K171602) to the FDA. The submission focuses on demonstrating that the device meets design and performance requirements through a series of tests.

Here's an analysis of the acceptance criteria and the study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

TestAcceptance CriteriaReported Device Performance
BiocompatibilityAcceptable for intended uses per ISO 10993-1 and FDA Guidance.Biological risks evaluated and determined to be acceptable.
Shelf LifeMaintain device performance for a three-year shelf life.All predetermined acceptance criteria were met after accelerated aging to the equivalence of three years.
Functional VerificationDevice functions as intended.Acceptance criteria were met.
Dimension VerificationDevice functions as intended (implies meeting dimensional specifications).Acceptance criteria were met.

2. Sample Size for Test Set and Data Provenance:

The document does not specify the sample sizes used for the Biocompatibility, Shelf Life, Functional Verification, or Dimension Verification tests. It also does not mention the country of origin of the data or whether the studies were retrospective or prospective.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

The document does not mention the involvement of experts to establish a "ground truth" for the performance tests. The tests described (biocompatibility, shelf life, functional, dimensional) typically rely on standardized laboratory procedures and measurements against predetermined specifications, rather than expert interpretation of a 'truth' in a clinical context.

4. Adjudication Method for the Test Set:

Given the nature of the performance tests described (laboratory-based measurements and compliance with standards), an "adjudication method" like 2+1 or 3+1 is not applicable or mentioned. Acceptance is typically based on whether measured values fall within predefined ranges or meet specific pass/fail criteria.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

An MRMC comparative effectiveness study was not performed or described. The device is an injection needle, and the testing focuses on its physical and functional properties, not on diagnostic or interpretive accuracy that would typically involve human readers.

6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance:

This concept is not applicable to the Williams Cystoscopic Injection Needle. This device is a manual, single-use medical instrument and does not involve algorithms or AI. Therefore, no standalone algorithm performance was assessed.

7. Type of Ground Truth Used:

For the performance tests mentioned, the "ground truth" implicitly refers to:

  • Biocompatibility: Adherence to established biological safety standards (ISO 10993-1).
  • Shelf Life: Stability of physical and functional properties over time, measured against initial design specifications.
  • Functional Verification: The device performing its mechanical and operational actions as designed.
  • Dimension Verification: Adherence to the specified dimensional blueprints and tolerances.

These are not "expert consensus, pathology, or outcomes data" in the traditional sense, but rather objective compliance against engineering specifications and international standards.

8. Sample Size for the Training Set:

The document does not mention a training set. This is because the device is a physical instrument, not a learning algorithm or AI, therefore, a training set is not relevant.

9. How the Ground Truth for the Training Set Was Established:

This question is not applicable as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, creating a sense of depth and unity.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 28, 2017

Cook Incorporated Rebecca Odulio (Li-Chun Liu) Regulatory Affairs Specialist 750 Daniels Way, P.O. Box 489 Bloomington, IN 47402

Re: K171602

Trade/Device Name: Williams Cystoscopic Injection Needle Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FBK Dated: May 31, 2017 Received: June 1, 2017

Dear Rebecca Odulio (Li-Chun Liu):

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Charles Viviano -S

For

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K171602

Device Name Williams Cystoscopic Injection Needle

Indications for Use (Describe)

The Williams Cystoscopic Injection Needle is used for cystoscopic into the urethra, bladder wall.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

2.0 510(k) Summary

Williams Cystoscopic Injection Needle 21 CFR §807.92 Date Prepared: May 31, 2017

Submitted By:

Submission:Traditional 510(k) Premarket Notification
Applicant:Cook Incorporated
Contact:Rebecca Odulio (Li-chun Liu)
Applicant Address:Cook Incorporated750 Daniels WayBloomington, IN 47404
Contact Phone:(812) 339-2235 x104673
Contact Fax:(812) 332-0281

Device Information:

Trade Name:Williams Cystoscopic Injection Needle
Common Name:Endoscopic Injection Needle, Gastroenterology-Urology
Classification Name:Endoscope and Accessories
Classification Regulation:21 CFR §876.1500, Product Code FBK
Device Class/Classification Panel:Class II, Gastroenterology/Urology

Predicate Device:

Cook Injection Needles (Cook Urological Inc., K022484) is the predicate device for the Williams Cystoscopic Injection Needle.

Device Description:

The Williams Cystoscopic Injection Needle is a semi-rigid injection needle. It is provided in 3.7 or 5.0 French sizes with 35 or 45 centimeters long polyurethane catheter, which ends with a 7 or 8 millimeter length stainless steel beveled tip cannula that is available in 20 to 25 needle gages. The subject device is sterilized with EtO and is intended for one-time use.

{4}------------------------------------------------

Indications for Use:

The Williams Cystoscopic Injection Needle is used for cystoscopic injection into the urethra, bladder neck, and bladder wall.

Comparison to Predicate Device:

The Williams Cystoscopic Injection Needle has similar methods of operation, materials, and fundamental technological characteristics as the predicate device. Differences between the subject device and the predicate device include minor dimensional variations that fall within the range of the predicate device and an additional feature of a grip wing hub. Characteristics of the subject device that differ from the predicate device are supported by testing to demonstrate that the subject device met the design input requirements for the instruction for use. The change for which this device is submitted is the more specific indications for use.

Indications for Use
To deliver a variety of injectable material into tissues during laparoscopic,
hysteroscopy, cystoscopic, endoscopic, transurethral procedures and open surgical
Predicate deviceprocedures. The type of material to be injected will be dependent on the nature of the
procedure, but such materials may include for example, saline, contrast media,
collagen, silicone, Teflon, local anesthetics, etc.
Subject deviceUsed for cystoscopic injection into the urethra, bladder neck, and bladder wall.

Performance Data:

The following testing were performed to demonstrate that the subject Williams Cystoscopic Injection Needle met applicable design and performance requirements and support a determination of substantial equivalence. Where applicable, testing was done per applicable ISO international standards.

  • Biocompatibility Biological risks have been evaluated in accordance with ISO . 10993-1 and FDA's Guidance on ISO 10993-1 and determined to be acceptable for the intended uses of the subject devices.
  • Shelf Life A three-year shelf life is supported by device performance testing after . accelerated aging to the equivalence of three years. All predetermined acceptance criteria were met.
  • . Functional Verification - Testing demonstrated that the subject device functions as intended. The results showed that acceptance criteria were met.
  • . Dimension Verification - Testing demonstrated that the subject device functions as intended. The results showed that acceptance criteria were met.

{5}------------------------------------------------

Conclusion:

The results of these tests provide reasonable assurance that the subject Williams Cystoscopic Injection Needle functions as intended and does not raise new questions of safety or effectiveness as compared to the predicate device.

{6}------------------------------------------------

2.0 510(k) Summary

Williams Cystoscopic Injection Needle 21 CFR §807.92 Date Prepared: May 31, 2017

Submitted By:

Submission:Traditional 510(k) Premarket Notification
Applicant:Cook Incorporated
Contact:Rebecca Odulio (Li-chun Liu)
Applicant Address:Cook Incorporated750 Daniels WayBloomington, IN 47404
Contact Phone:(812) 339-2235 x104673
Contact Fax:(812) 332-0281

Device Information:

Trade Name:Williams Cystoscopic Injection Needle
Common Name:Endoscopic Injection Needle, Gastroenterology-Urology
Classification Name:Endoscope and Accessories
Classification Regulation:21 CFR §876.1500, Product Code FBK
Device Class/Classification Panel:Class II, Gastroenterology/Urology

Predicate Device:

Cook Injection Needles (Cook Urological Inc., K022484) is the predicate device for the Williams Cystoscopic Injection Needle.

Device Description:

The Williams Cystoscopic Injection Needle is a semi-rigid injection needle. It is provided in 3.7 or 5.0 French sizes with 35 or 45 centimeters long polyurethane catheter, which ends with a 7 or 8 millimeter length stainless steel beveled tip cannula that is available in 20 to 25 needle gages. The subject device is sterilized with EtO and is intended for one-time use.

{7}------------------------------------------------

Indications for Use:

The Williams Cystoscopic Injection Needle is used for cystoscopic injection into the urethra, bladder neck, and bladder wall.

Comparison to Predicate Device:

The Williams Cystoscopic Injection Needle has similar methods of operation, materials, and fundamental technological characteristics as the predicate device. Differences between the subject device and the predicate device include minor dimensional variations that fall within the range of the predicate device and an additional feature of a grip wing hub. Characteristics of the subject device that differ from the predicate device are supported by testing to demonstrate that the subject device met the design input requirements for the instruction for use. The change for which this device is submitted is the more specific indications for use.

Indications for Use
Predicate deviceTo deliver a variety of injectable material into tissues during laparoscopic, hysteroscopy, cystoscopic, endoscopic, transurethral procedures and open surgical procedures. The type of material to be injected will be dependent on the nature of the procedure, but such materials may include for example, saline, contrast media, collagen, silicone, Teflon, local anesthetics, etc.
Subject deviceUsed for cystoscopic injection into the urethra, bladder neck, and bladder wall.

Performance Data:

The following testing were performed to demonstrate that the subject Williams Cystoscopic Injection Needle met applicable design and performance requirements and support a determination of substantial equivalence. Where applicable, testing was done per applicable ISO international standards.

  • . Biocompatibility - Biological risks have been evaluated in accordance with ISO 10993-1 and FDA's Guidance on ISO 10993-1 and determined to be acceptable for the intended uses of the subject devices.
  • . Shelf Life - A three-year shelf life is supported by device performance testing after accelerated aging to the equivalence of three years. All predetermined acceptance criteria were met.
  • . Functional Verification - Testing demonstrated that the subject device functions as intended. The results showed that acceptance criteria were met.
  • . Dimension Verification - Testing demonstrated that the subject device functions as intended. The results showed that acceptance criteria were met.

{8}------------------------------------------------

Conclusion:

The results of these tests provide reasonable assurance that the subject Williams Cystoscopic Injection Needle functions as intended and does not raise new questions of safety or effectiveness as compared to the predicate device.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.