K Number
K171565

Validate with FDA (Live)

Device Name
OEC Elite
Date Cleared
2017-08-10

(72 days)

Product Code
Regulation Number
892.1650
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OEC Elite mobile fluoroscopy system is designed to provide fluoroscopic and digital spot images of adult and pediatric patient populations during diagnostic, interventional, and surgical procedures. Examples of a clinical application may include orthopedic, gastrointestinal, endoscopic, urologic, vascular, critical care, and emergency procedures.

Device Description

The OEC Elite is a Mobile Fluoroscopic C-arm Imaging system used to assist trained surgeons and other qualified physicians. The system is used to provide fluoroscopic X-Ray images during diagnostic, interventional, and surgical procedures. These images help the physician visualize the patient's anatomy and interventional tools. This visualization helps to localize clinical regions of interest and pathology. The images provide real-time visualization and records of pre-procedure anatomy, in vivo-clinical activity and post-procedure outcomes. The system is composed of two primary physical components. The first is referred to as the "C - Arm" because of its "C" shaped image gantry; the second is referred to as the "Workstation", which is the primary interface for the user to interact with the system.

AI/ML Overview

The provided document describes the OEC Elite mobile fluoroscopy system, which is a medical imaging device. The document is submitted as a 510(k) premarket notification to the FDA to demonstrate substantial equivalence to a legally marketed predicate device.

However, the document specifically states: "Clinical images are not required to demonstrate the substantial equivalence to the predicate device." This means that there was no clinical study involving human patients or human readers performed to prove the device meets acceptance criteria related to clinical performance (e.g., diagnostic accuracy, reader improvement with AI assistance).

The acceptance criteria and proof of performance are based on non-clinical testing (engineering bench testing and phantom studies).

Here's a breakdown of the requested information based on the provided text, while acknowledging the absence of a clinical study:

1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical):

The acceptance criteria are implied by compliance with various industry standards and guidance documents for X-ray imaging devices. The document asserts that the device met these criteria.

Acceptance Criteria Category (Implied)Specific Metrics (Examples from SSXI guidance)Reported Device Performance
System Performance- Risk AnalysisMet design input and user needs; successful verification and validation
- Hazard MitigationSuccessfully executed
- Compliance to IEC 60601-1 Ed. 3 seriesCompliant
- Compliance to IEC 60601-2-54Compliant
- Compliance to IEC 60601-2-43Compliant
- Compliance to 21 CFR Subchapter J performance standardsAll applicable standards met
Image Quality/Performance- DOE (Detective Quantum Efficiency)Acceptable results; identical performance to predicate
- Dynamic RangeAcceptable results; identical performance to predicate
- Spatial Resolution (MTF, Limiting Resolution)Acceptable results; identical performance to predicate
- Temporal ResolutionAcceptable results; identical performance to predicate
- Contrast ResolutionAcceptable results; identical performance to predicate
- Beam AlignmentAcceptable results; identical performance to predicate
- Dose RateAcceptable results; identical performance to predicate
- Stability of device characteristics over timeAcceptable results; identical performance to predicate
- Brightness uniformityAcceptable results; identical performance to predicate
- Fluoroscopy Frame RateAcceptable results; identical performance to predicate
- Reuse RateAcceptable results; identical performance to predicate
Software Quality- Design ControlsCompliant
- Risk ManagementCompliant
- Software Development Lifecycle Processes (IEC 60324 compliant)Compliant
Functional Features- Vascular features (Subtraction, Roadmapping, Digital Cine Pulse, Peak opacification, Cine Recording/playback, etc.)Implemented and functioning as designed (implied by "built upon existing robust...software architecture")

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set Sample Size: Not applicable in the context of a clinical test set. The document refers to "engineering bench testing" and "imaging performance evaluation using anthropomorphic phantoms." The specific number of phantoms or tests is not provided.
  • Data Provenance: Not applicable for clinical data. The tests were performed internally by the manufacturer (GE OEC Medical Systems, Inc.) in Salt Lake City, Utah, USA. The data is from non-clinical sources (bench tests, phantom studies).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

  • Not applicable. As no clinical study was performed, there were no human experts establishing clinical ground truth for a test set. Ground truth for non-clinical performance (e.g., image quality metrics) is established through engineering and physics measurements against defined standards.

4. Adjudication Method for the Test Set:

  • Not applicable, as no clinical study with human readers or AI performance evaluation was conducted.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

  • No. The document explicitly states: "Clinical images are not required to demonstrate the substantial equivalence to the predicate device." Therefore, no MRMC study to compare human readers with and without AI assistance was performed.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Yes, indirectly, in a non-clinical context. The device itself is an imaging system, not explicitly an "AI algorithm" in the modern sense of a diagnostic AI. However, the performance metrics listed (DOE, spatial resolution, contrast resolution, etc., and the functioning of new software features like subtraction or roadmapping) represent the "standalone" imaging performance of the device, without human interpretation as part of the validation study. The "algorithm" here refers to the underlying image processing and control software of the fluoroscopy system.

7. The Type of Ground Truth Used:

  • Engineering specifications, physical measurements, and phantom imaging results. The ground truth for the non-clinical performance was based on:
    • Design input specifications.
    • Compliance with recognized standards (e.g., IEC 60601 series, FDA's "Information for Industry: X-ray Imaging Devices- Laboratory Image Quality and Dose Assessment, Tests and Standards", "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices (SSXI)").
    • Measurements obtained from anthropomorphic phantoms that simulate human anatomy, rather than actual patient data or clinical outcomes.

8. The Sample Size for the Training Set:

  • Not applicable. This device is a fluoroscopy system, not a machine learning model that undergoes "training" in the typical AI sense. The software development process mentioned (IEC 60324 compliant, design control, risk management) pertains to traditional software engineering, not AI model training with large datasets. The new vascular features were "built upon the existing robust and extensible software architecture."

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable, for the same reason as point 8.

{0}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 10, 2017

GE OEC Medical Systems, Inc. Rachel Schandel Regulatory Affairs Leader 384 Wright Brothers Drive Salt Lake City, Utah 84116

Re: K171565

Trade/Device Name: OEC Elite Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, JAA, OXO Dated: July 26, 2017 Received: July 27, 2017

Dear Ms. Rachel Schandel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Michael D. O'Hara For

Robert A. Ochs, Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K171565

Device Name OEC Elite

Indications for Use (Describe)

The OEC Elite mobile fluoroscopy system is designed to provide fluoroscopic and digital spot images of adult and pediatric patient populations during diagnostic, interventional, and surgical procedures. Examples of a clinical application may include orthopedic, gastrointestinal, endoscopic, urologic, vascular, critical care, and emergency procedures.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. There are three small, white, teardrop-shaped elements evenly spaced around the outside of the circle, giving the logo a distinctive and recognizable appearance.

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

This 510(k) Summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h).

Date:May 26th, 2017
Submitter:GE OEC Medical Systems, Inc (GE Healthcare)384 Wright Brothers DriveSalt Lake City, Utah 84116
Primary Contact:Rachel SchandelRegulatory Affairs LeaderT (801)5364790Rachel.schandel@ge.com
Secondary Contact:Helen PengSenior Regulatory Affairs ManagerGE Medical Systems, LLCTel: 262-424-8222e-mail: hong.peng@ge.com
Secondary Contact:John JaeckleChief Regulatory Affairs StrategistGE Medical Systems, LLCTel: 262-424-9547e-mail: john.jaeckle@ge.com

PRODUCT IDENTIFICATION

Device Trade Name:OEC Elite
Regulation Name:Image-intensified Fluoroscopic x-ray system
Regulation:21CFR 892.1650
Classification:Class II
Product Code:OWB, JAA, OXO
Manufacturer/ Design Location:GE OEC Medical Systems, Inc (GE Healthcare)384 Wright Brothers Drive.Salt Lake City, Utah 84116
Manufacturing Location(s):GE OEC Medical Systems, Inc (GE Healthcare)384 Wright Brothers Drive.

Salt Lake City, Utah 84116

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized script, enclosed within a blue circle. The circle is surrounded by decorative swirls, also in blue, which add a sense of elegance and motion to the design. The logo is simple, yet recognizable, and represents the long-standing brand identity of General Electric.

Predicate Device:

Device Name:OEC Elite
510(k) number:K170752
Manufacturer:GE OEC Medical Systems, Inc (same as proposed device)
Regulation Name:Image-intensified Fluoroscopic x-ray system
Regulation:21CFR 892.1650
Classification:Class II
Product Code:OWB, JAA, OXO

Marketed Device:

OEC Elite with Vascular option is a Mobile Fluoroscopic C-arm Imaging System with modification to the predicate device OEC Elite (K170752) to offer the users the vascular features. It is of comparable type and substantially equivalent to the predicate device OEC Elite. This 510(k) submission for the OEC Elite system with vascular option includes the same intended use and substantially equivalent indications for use as its predicate device. The existing indications for use of OEC Elite are expanded to include the vascular capability which is one of the current clinical uses of mobile c-arm systems that are well established in the medical community.

The system is labeled as the OEC Elite.

Device Description:

The OEC Elite is a Mobile Fluoroscopic C-arm Imaging system used to assist trained surgeons and other qualified physicians. The system is used to provide fluoroscopic X-Ray images during diagnostic, interventional, and surgical procedures. These images help the physician visualize the patient's anatomy and interventional tools. This visualization helps to localize clinical regions of interest and pathology. The images provide real-time visualization and records of pre-procedure anatomy, in vivo-clinical activity and post-procedure outcomes. The system is composed of two primary physical components. The first is referred to as the "C - Arm" because of its "C" shaped image gantry; the second is referred to as the "Workstation", which is the primary interface for the user to interact with the system.

The C-arm is a stable mobile platform capable of performing linear motions (vertical, horizontal) and rotational motions (orbital, lateral, wig-wag) that allow the user to position the X-Ray image chain at various and distances with respect to the patient anatomy to be imaged. The C - arm is mechanically balanced allowing for ease of movement and capable of being "locked" in place using a manually activated lock. The C-Arm is comprised of the high voltage generator, software, X-ray control, and a "C" shaped image gantry, which supports an X-ray tube and a Flat Panel Detector or Image Intensifier, depending on the choice of detector configuration desired.

The workstation is a stable mobile platform with an articulating arm supporting a color image, high resolution, LCD display monitor. It also includes image processing equipment/software, recording devices, data input/output devices and power control systems.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. The circle is surrounded by a series of curved lines that resemble water droplets or waves. The logo is simple and recognizable, and it is associated with a well-known and established company.

The primary purpose of the mobile fluoroscopy system is to provide fluoroscopic images of the patient during diagnostic, interventional, and surgical procedures such as orthopedic, gastrointestinal, endoscopic, urologic, neurologic, vascular, critical care and emergency procedures.

The primary technology change for the subject device compared to the unmodified predicate, OEC Elite, is the introduction of the vascular capabilities.

Intended Use

The OEC Elite Mobile C-arm is intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional tools/devices, and contrast agents during diagnostic, interventional, and surgical procedures.

Indications for Use:

The OEC Elite mobile fluoroscopy system is designed to provide fluoroscopic and digital spot images of adult and pediatric patient populations during diagnostic, interventional, and surgical procedures. Examples of a clinical application may include orthopedic, gastrointestinal, endoscopic, urologic, neurologic, vascular, critical care, and emergency procedures.

Technology:

The OEC Elite with Vascular option employs the same fundamental scientific technology as that of the predicate device OEC Elite (K170752).

The image chain including the X-Ray source and generator remain unchanged from that of the predicate OEC Elite. The primary change was in the software to implement the vascular features such as Subtraction, Roadmapping, and Digital cine pulse to help perform vascular procedures with the easiest workflow and least intervention by the user. A suite of supporting features such as Peak opacification, Cine Recording/playback, Cine Automatic image playback, Cine Frame-by-frame review, Reregistration, Variable landmarking, and Mask save/recall, were added to further enhance the vascular workflows. Additionally a new vascular profile was developed to optimize the visualization of positive contrast agent-filled vessels, catheters, stents, and other tools in vascular sperformed in thick anatomy, while a Bolus Chase profile was created to optimize the visualization of the run-off procedures and interventional procedures using small guidewires.

The modifications however, were built upon the existing robust and extensible software architecture, following the same design control process and software development lifecycle process that is compliant to IEC60324. The changes described above do not change the control mechanism, operating principle, energy type, or intended use from the predicate device.

Determination of Substantial Equivalence:

The main change in the proposed device is the introduction of the vascular feature. The table below summarizes the substantive feature/technological differences between the predicate device and the proposed device:

{6}------------------------------------------------

Predicate DeviceOEC Elite (K170752)Proposed DeviceOEC Elite with Vascular Option
Imaging Modes> Continuous – Fluoroscopy> Continuous – Fluoroscopy
○ Normal dose○ Normal dose
○ High Level dose○ High Level dose
○ Low dose○ Low dose
> Pulsed Fluoroscopy> Pulsed Fluoroscopy
○ Normal dose○ Normal dose
○ High Level dose○ High Level dose
○ Low dose○ Low dose
> Digital Spot> Digital Spot
> Digital Cine Pulse
○ Normal dose
○ Low dose
> Roadmap
○ Normal dose
○ Low dose
> Subtraction
○ Normal dose
○ Low dose
Imaging Features> Auto X-Ray technique control> Auto X-Ray technique control
> Noise and motion reduction> Noise and motion reduction
> Auto/Manual Brightness and Contrast Control> Auto/Manual Brightness and Contrast Control
> Negate> Negate
> Swap and auto-swap> Swap and auto-swap
> Save and auto-save> Save and auto-save
> Last image hold> Last image hold
Predicate DeviceOEC Elite (K170752)Proposed DeviceOEC Elite with Vascular Option
Edge enhancementEdge enhancement
Zoom & RoamZoom & Roam
Image rotationImage rotation
Image flip/ invertImage flip/ invert
Manual/Auto Smart MetalManual/Auto Smart Metal
AutoTrakAutoTrak
Window/LevelWindow/Level
Patient AnnotationPatient Annotation
MarkersMarkers
Measurement FunctionsMeasurement Functions
Peak opacification
Cine Recording/playback
Cine Automatic image playback
Cine Frame-by-frame review
Re-registration
Variable landmarking
Mask save/recall

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a blue circle. There are decorative swirls around the outer edge of the circle, giving it a classic and recognizable appearance.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a blue circle. There are decorative white swirls around the perimeter of the circle, adding a touch of elegance to the design.

GE Healthcare believes the OEC Elite with the Vascular option is of comparable type and substantially equivalent to our currently marketed system OEC Elite.

The substantial equivalence was also based on software documentation for a "Moderate" level of concern device.

Summary of Additional Testing

In addition to the verification and validation testing successfully completed as required by GE OEC Medical System's Quality Management System, additional engineering (non-Clinical) testing was performed to provide the requisite data to substantiate performance claims, the revised indications, and ultimately substantial equivalence.

Non-Clinical Testing

Verification and validation, including hazard mitigation has been executed with results demonstrating the OEC Elite mobile fluoroscopy system met design input and user needs.

The system has been tested and is compliant with the IEC 60601-1 Ed. 3 series, including IEC60601-2-54 and IEC 60601-2-43. All applicable 21CFR Subchapter J performance standards are met. The OEC Elite system was developed under the GE OEC Medical Systems Quality Management System, including design controls, risk management and software development life cycle processes. The following quality assurance measures were applied to the development of the system:

  • Risk Analysis
  • · Required Reviews
  • · Design Reviews
  • · Testing on unit level (Sub System verification)
  • · Integration testing (System verification)
  • · Performance testing (Verification)
  • · Safety testing (Verification)
  • · Simulated use testing (Validation)

Additional engineering bench testing was performed including demonstration of system performance; and an imaging performance evaluation using anthropomorphic phantoms. All the image quality/performance testing identified for fluoroscopy found in FDA's "Information for Industry: X-ray Imaging Devices- Laboratory Image Quality and Dose Assessment, Tests and Standards" was performed with acceptable results.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized script, enclosed within a blue circle. The circle has a white border with decorative swirls, giving it a classic and recognizable appearance.

The OEC Elite with Vascular option was also evaluated for its performance per the FDA "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices'' (SSXI). It was determined the modified system continues to comply to this guidance. The evaluation includes but not limited to the following performance metrics identified in the SSXI guidance, showing identical performance of the subject device to the predicate device OEC Elite:

  • DOE ●
  • Dynamic Range
  • Spatial Resolution (MTF, Limiting Resolution) ●
  • Temporal Resolution
  • Contrast Resolution
  • Beam Alignment
  • Dose Rate ●
  • Stability of the device characteristics over time
  • brightness uniformity
  • . Fluoroscopy Frame Rate
  • Reuse Rate .

Clinical Testing

Because OEC Elite with the vascular option does not change the system's intended use and represents equivalent technological characteristics, this type of change supports using scientific, established/standardized, engineering/physics-based performance testing, without inclusion of clinical images. Therefore, for OEC Elite with Vascular option configurations, clinical images are not required to demonstrate the substantial equivalence to the predicate device.

Substantial Equivalence Conclusion:

The differences discussed in this submission do not introduce any adverse effects nor raise new questions of safety and effectiveness. Based on the successful verification and validation testing, additional engineering bench testing, conformance to standards, and development under GE OEC Medical System's Quality Management System, we believe that the OEC Elite is of comparable type and substantially equivalent to the predicate device OEC Elite (K170752), and therefore is safe and effective for its intended use.

N/A