K Number
K171461

Validate with FDA (Live)

Date Cleared
2017-07-07

(50 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

As a part of a radiographic system, the SkyPlate Detector for Philips Radiography/Fluoroscopy Systems is intended to acquire, process, store, display and export digital radiographic images. The SkyPlate Detector for Philips Radiography/ Fluoroscopy Systems is suitable for all routine radiography exams, including specialist areas like intensive care, trauma, or pediatric work, excluding mammography.

Device Description

The SkyPlate Detector for Philips Radiography/Fluoroscopy Systems is a Solid State X-ray Imaging Device that converts x-ray patterns into electrical signals. The signals are converted into visible images for use in medical diagnosis. A cesium iodide scintillator absorbs the input x-ray photons in the detector. The cesium iodide scintillator in turn emits visible spectrum photons that illuminate an array of photodetectors that create an electrical charge representation of the x-ray input. A matrix scan of the array converts the integrated charges into a modulated electrical signal.

The SkyPlate Detector for Philips Radiography/Fluoroscopy Systems is optionally installed and is intended to be integrated into an x-ray system, where it constitutes an x-ray receptor for direct radiography x-ray imaging. It is electrically powered by and connected with the x-ray system. The SkyPlate Detector for Philips Radiography/Fluoroscopy Systems is connected to the Philips Eleva Workspot with SkyFlow (cleared via K153318) to create a complete x-ray imaging chain, and is intended to be used for radiography in Philips Radiography/Fluoroscopy systems, such as the CombiDiagnost R90 (cleared via K163210). The SkyPlate Detector for Philips Radiography/Fluoroscopy Systems, is used to acquire diagnostic radiographic images during radiographic procedures.

AI/ML Overview

This document is a 510(k) premarket notification for the SkyPlate Detector for Philips Radiography/Fluoroscopy Systems. It outlines the device's characteristics, intended use, and claims of substantial equivalence to a predicate device.

Here's an analysis of the provided text in relation to your questions:

1. A table of acceptance criteria and the reported device performance

The document does not provide a specific table of quantitative acceptance criteria for device performance. Instead, it states that the device:

  • "Complies with the aforementioned international and FDA-recognized consensus standards and device specific guidance documents."
  • "Meets the acceptance criteria and is adequate for its intended use."
  • "is as safe, as effective, and performs as well or better than the predicate device."
  • "provides images of equivalent diagnostic capability."

The full list of standards mentioned are:

  • ISO 14971: Medical Devices - Application of risk management to medical devices
  • NEMA PS 3.1-3.20 Digital Imaging and Communication in Medicine (DICOM) Set
  • AAMI ANSI IEC 62304:2006 Medical Device Software - Software lifecycle processes
  • IEC 62220-1: Medical electrical equipment, Characteristics of digital x-ray imaging devices
  • IEC 62220-1-1:2015: Medical electrical equipment, Characteristics of digital x-ray imaging devices – Part 1-1: Determination of the detective quantum efficiency – Detectors used in radiographic imaging
  • IEC 60601-1: Medical electrical equipment – Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005 + A1:2012, MOD)
  • IEC 60601-1-2: Medical electrical equipment. General requirements for basic safety and essential performance. Collateral standard: Electromagnetic compatibility requirements and tests
  • IEC 60601-1-3: Medical electrical equipment. General requirements for basic safety and essential performance. Collateral standard: Radiation in diagnostic X-ray equipment
  • IEC 60601-2-54: Medical electrical equipment. Particular requirements for basic safety and essential performance of X-ray equipment for radiography and radioscopy
  • Guidance for the Submission of 510(k)s for Solid State C X-ray Imaging Device", issued September 1, 2016
  • "Guidance for the Content of Premarket Submissions for O software contained in Medical Device", issued May 11, 2005
  • Pediatric information for X-Ray Imaging Device Premarket Notifications, Draft, issued May 10, 2012

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document explicitly states: "The SkyPlate Detector for Philips Radiography/Fluoroscopy Systems did not require a clinical study since substantial equivalence to the currently marketed and predicate device was demonstrated with the following attributes: Design features; Indication for use; Fundamental scientific technology; Non-clinical performance testing including validation; and Safety and effectiveness." Therefore, no clinical test set was used for this 510(k) submission, and consequently, no clinical sample size or data provenance is provided.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Since no clinical study was conducted and no clinical test set was used, no experts were used to establish ground truth for a test set.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

As no clinical study was conducted and no clinical test set was used, no adjudication method was applied.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

The document does not describe any MRMC comparative effectiveness study. This device is an X-ray detector, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This refers to a standalone performance of an algorithm. The SkyPlate Detector is an X-ray imaging device, not an algorithm. Therefore, this question is not applicable to the device described. The submission relies on "non-clinical performance testing including validation" for the physical device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the non-clinical performance testing, the ground truth would be established by engineering and laboratory measurements demonstrating compliance with the specified international and FDA-recognized consensus standards. This would involve physical measurements of detector characteristics (e.g., image resolution, detective quantum efficiency, electrical signals, safety parameters) against defined benchmarks within those standards, rather than clinical ground truth from patient data.

8. The sample size for the training set

This question typically applies to machine learning or AI models. Since the device is a physical X-ray detector and not an AI algorithm, there is no concept of a "training set" in the context of this submission. The validation was based on non-clinical performance and engineering tests.

9. How the ground truth for the training set was established

As explained above, this question is not applicable as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 7, 2017

PHILIPS MEDICAL SYSTEMS DMC GMBH MING XIAO REGULATORY AFFAIRS MANAGER, NORTH AMERICA ROENTGENSTRASSE 24-26 HAMBURG 22335 DE

Re: K171461

Trade/Device Name: SkyPlate Detector for Philips Radiography/Fluoroscopy Systems Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: May 18, 2017 Received: May 18, 2017

Dear Mr. Xiao:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Robert Oolo

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out.

5. Statement of Indication for Use

DEPARTMENT OF HEALTH AND HUMAN SERVICESFood and Drug AdministrationForm Approved: OMB No. 0910-0120Expiration Date: January 31, 2017See PRA Statement below.
510(k) Number (if known)K171461
Device NameSkyPlate Detector for Philips Radiography/Fluoroscopy Systems
Indications for Use (Describe)As a part of a radiographic system, the SkyPlate Detector for Philips Radiography/Fluoroscopy Systems is intended to acquire, process, store, display and export digital radiographic images. The SkyPlate Detector for Philips Radiography/ Fluoroscopy Systems is suitable for all routine radiography exams, including specialist areas like intensive care, trauma, or pediatric work, excluding mammography.
Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

SkyPlate Detector for Philips Radiography/Fluoroscopy Systems Premarket Notification – Special 510(k)

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is plain white.

6. 510(k) Summary

510(k) Summary of Safety and Effectiveness

This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.

Date Prepared:May 16, 2017 (original Submission Date)July 4, 2017 (Revision)
Manufacturer:Philips Medical Systems DMC GmbHRoentgenstrasse 24-2622335 HamburgGERMANYEstablishment registration number: 3003768251
Contact Person:Ming XiaoRegulatory Affairs Manager North AmericaPhone: +49 40 5078-2306Fax: +49 40 5078-2425E-mail: ming.xiao@philips.com
Trade / Device Name:SkyPlate Detector for Philips Radiography/Fluoroscopy Systems
Classification:(Primary) Regulation Name: Stationary x-ray system
(Primary) Regulation Number: 21CFR 892.1680
Classification Panel: 90 -- Radiology
Device Regulatory Class: Class II
(Primary) Product Code: MQB
(Secondary) Product Code: LLZ
(Secondary) Regulation Name: Picture archiving andcommunication system
(Secondary) Regulation Number: 21 CFR 2050
Predicate Device:Trade / Device Name: Philips Eleva Workspotwith SkyPlate Detectors
Manufacturer: Philips Medical SystemsDMC GmbH
510(k) Clearance: K141736 – July 25, 2014
Regulation Name: Stationary x-ray system,Picture archiving andcommunications system
Regulation Number: 21CFR 892.1680, 21 CFR 2050
Classification Panel:90 -- Radiology
Device Regulatory Class:Class II
Product code:MOB, LLZ
Reference Device I:Trade / Device Name:Philips CombiDiagnost R90
Manufacturer:Philips Medical Systems
DMC GmbH
510(k) Clearance:K163210 - January31.2017
Regulation Name:Image-intensified
Radiography/Fluoroscopy
x-ray system
Regulation Number:21 CFR 892.1650
Product Code:JAA, KPR and MQB
Device Regulatory Class:Class II
Reference Device II:Trade / Device Name:Philips Eleva Workspotwith SkyFlow
Manufacturer:Philips MedicalSystemsDMC GmbH
510(k) Clearance:K153318 – December 22,2015
Regulation Name:Stationary X-ray System
Regulation Number:21 CFR 892.1680
Product Code:MQB, LLZ
Device Regulatory Class:Class II
Device Description:SkyPlate Detector for for Philips Radiography/FluoroscopyThe
Systems is a Solid State X-ray Imaging Device that converts x-ray

SkyPlate Detector for Philips Radiography/Fluoroscopy Systems Premarket Notification – Special 510(k) 510(k) Summary, revised July 4, 2017

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, providing a strong contrast that makes the word stand out.

Systems is a Solid State X-ray Imaging Device that converts x-ray patterns into electrical signals. The signals are converted into visible images for use in medical diagnosis. A cesium iodide scintillator absorbs the input x-ray photons in the detector. The cesium iodide scintillator in turn emits visible spectrum photons that illuminate an array of photodetectors that create an electrical charge representation of the x-ray input. A matrix scan of the array converts the integrated charges into a modulated electrical signal.

The SkyPlate Detector for Philips Radiography/Fluoroscopy Systems is optionally installed and is intended to be integrated into an x-ray system, where it constitutes an x-ray receptor for direct radiography x-ray imaging. It is electrically powered by and connected with the x-ray system. The SkyPlate Detector for Philips

SkyPlate Detector for Philips Radiography/Fluoroscopy Systems Premarket Notification - Special 510(k) 510(k) Summary, revised July 4, 2017

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out. The word is horizontally oriented and centered.

Radiography/Fluoroscopy Systems is connected to the Philips Eleva Workspot with SkyFlow (cleared via K153318) to create a complete x-ray imaging chain, and is intended to be used for radiography in Philips Radiography/Fluoroscopy systems, such as the CombiDiagnost R90 (cleared via K163210). The SkyPlate Detector for Philips Radiography/Fluoroscopy Systems, is used to acquire diagnostic radiographic images during radiographic procedures. Key features for the SkyPlate Detector for Philips Radiography/Fluoroscopy Systems are provided below:

  • Detector Size: ● SkyPlate Large (also called Pixium 3543EZ): 384.5mm x 460.5mm x 16mm SkyPlate Small (also called Pixium 2430EZ): 268.5mm x 328.5mm x 16mm
  • Image Size: . SkyPlate Large (also called Pixium 3543EZ): 2400 x 2880 pixels SkyPlate Small (also called Pixium 2430EZ): 1560 x 1920 pixels
  • Pixel Size: 148um ●
  • Image Resolution up to 3.38 LP/mm ●

The identical Skyplate detector was previously cleared under K141736 (July 25, 2014) for use with currently marketed Philips radiography systems, such as the Philips DigitalDiagnost and the Philips ProGrade.

The SkyPlate Detector for Philips Radiography/Fluoroscopy Systems is integrated for use with Philips radiography/fluoroscopy systems, such as the CombiDiagnost R90 (K163210). The modifications included mechanical integration of the SkyPlate detector and the radiography/fluoroscopy system and integration of the SkyPlate Detector for Philips Radiography/Fluoroscopy Systems software interface with the Philips Eleva Workspot with SkyFlow software (K153318) to facilitate communication.

Indications for Use: The Indication for Use for the SkyPlate Detector for Philips Radiography/Fluoroscopy Systems is as follows:

As a part of a radiographic system, the SkyPlate Detector for Philips Radiography/Fluoroscopy Systems is intended to acquire, process, store, display and export digital radiographic images. The SkyPlate Detector for Philips Radiography/Fluoroscopy Systems is suitable for all routine radiography exams, including

SkyPlate Detector for Philips Radiography/Fluoroscopy Systems Premarket Notification - Special 510(k)

510(k) Summary, revised July 4, 2017

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, providing a strong contrast that makes the word stand out.

specialist areas like intensive care, trauma, or pediatric work, excluding mammography.

FundamentalScientific Technology:The fundamental scientific technology of the SkyPlate Detector forPhilips Radiography/Fluoroscopy Systems is that it converts x-raypatterns into electrical signals, which are then converted intovisible images for use in medical diagnosis.The fundamental scientific technology of the SkyPlate Detectorfor Philips Radiography/Fluoroscopy Systems remains unchangedfrom the currently marketed and predicate Philips Eleva Workspotwith SkyPlate Detectors (K141736, July 25, 2014) thusdemonstrating substantial equivalence.
Summary ofNon-clinicalPerformance Data:The SkyPlate Detector for Philips Radiography/FluoroscopySystems complies with the following international and FDA-recognized consensus standards:International and FDA-recognized consensus standards: ISO 14971: Medical Devices - Application of riskmanagement to medical devices NEMA PS 3.1-3.20 Digital Imaging and Communicationin Medicine (DICOM) Set AAMI ANSI IEC 62304:2006 Medical Device Software- Software lifecycle processes IEC 62220-1: Medical electrical equipment, Characteristicsof digital x-ray imaging devices IEC 62220-1-1:2015: Medical electrical equipment,Characteristics of digital x-ray imaging devices – Part 1-1: Determination of the detective quantum efficiency –Detectors used in radiographic imaging IEC 60601-1: Medical electrical equipment – Part 1:General requirements for basic safety and essentialperformance (IEC 60601-1:2005 + A1:2012, MOD) IEC 60601-1-2: Medical electrical equipment. Generalrequirements for basic safety and essential performance.Collateral standard: Electromagnetic compatibilityrequirements and tests IEC 60601-1-3: Medical electrical equipment. Generalrequirements for basic safety and essential performance.Collateral standard: Radiation in diagnostic X-rayequipment IEC 60601-2-54: Medical electrical equipment. Particularrequirements for basic safety and essential performance of

SkyPlate Detector for Philips Radiography/Fluoroscopy Systems Premarket Notification – Special 510(k) 510(k) Summary, revised July 4, 2017

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is white.

X-ray equipment for radiography and radioscopy

  • Device specific guidance document: ●
    • "Guidance for the Submission of 510(k)s for Solid State C X-ray Imaging Device", issued September 1, 2016
    • "Guidance for the Content of Premarket Submissions for O software contained in Medical Device", issued May 11, 2005
    • Pediatric information for X-Ray Imaging Device o Premarket Notifications, Draft, issued May 10, 2012

Non-clinical verification and validation tests have been performed with regards to the intended use, technical claims, requirement specifications, and risk management results. The software information was provided according to a moderate level of concern.

Non-clinical verification and validation test results demonstrate that the SkyPlate Detector for Philips Radiography/Fluoroscopy Systems:

  • Complies with the aforementioned international and FDArecognized consensus standards and device specific guidance documents.
  • Meets the acceptance criteria and is adequate for its ● intended use.

Therefore, the SkyPlate Detector for Philips Radiography/Fluoroscopy Systems is substantially equivalent to the currently marketed and predicate Philips Eleva Workspot with SkyPlate Detectors (K141736. July 25, 2014) thus demonstrating substantial equivalence.

The SkyPlate Detector for Philips Radiography/Fluoroscopy Summary of Clinical Systems did not require a clinical study since substantial Data: equivalence to the currently marketed and predicate device was demonstrated with the following attributes:

  • Design features; ●
  • Indication for use;
  • Fundamental scientific technology; ●
  • Non-clinical performance testing including validation; and
  • . Safety and effectiveness.

Substantial The comparison of technological characteristics, non-clinical Equivalence performance data, safety testing, software validation, and clinical image concurrence data demonstrates that the device is as safe, as

SkyPlate Detector for Philips Radiography/Fluoroscopy Systems Premarket Notification - Special 510(k)

510(k) Summary, revised July 4, 2017

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is plain white.

effective, and performs as well or better than the predicate device. Conclusion:

The SkyPlate Detector for Philips Radiography/Fluoroscopy Systems has the same indications for use and intended use, has identical or equivalent technological characteristics, provides images of equivalent diagnostic capability, and does not introduce new potential hazards or safety risks.

The SkyPlate Detector for Philips Radiography/Fluoroscopy Systems is substantially equivalent to the currently legally marketed predicate device (K141736, July 25, 2014) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness. Additionally, substantial equivalence was demonstrated with non-clinical performance (verification and validation) tests, which complied with the requirements specified in the international and FDA-recognized consensus standards: IEC 62304 and ISO 14971. The results of these tests demonstrate that the SkyPlate Detector for Philips Radiography/Fluoroscopy Systems met the acceptance criteria and is adequate for its intended use.

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.