K Number
K171410

Validate with FDA (Live)

Device Name
ePatch
Date Cleared
2018-01-04

(234 days)

Product Code
Regulation Number
870.2800
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ePatch® is indicated for use on patients who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, fatigue, chest pain and/or anxiety.

The ePatch® is intended for use by adolescents 18-21 and adults.

Device Description

The ePatch® ECG Recorder is intended for continuous recording of heart action potentials (ECG). While in use, the sensor is connected to a compatible ECG electrode and placed on the patient.

The ePatch® ECG Recorder is a small digital Holter recorder intended for use by professionals to acquire continuous ambulatory recording of heart action potentials (ECG) from a patient through a compatible ECG patch electrode. The ePatch® is intended for use in a clinical, point of care or at a patient setting. The recording time of the ECG recorder is dependent on the set configuration of the recorder (sample rate and channel selection). The patient's ECG is recorded to the ePatch® ECG Recorder and then transferred via the ePatch® USB cable to a Holter analysis system for review by physician or other qualified personnel.

The ePatch® ambulatory electrocardiograph (ECG) recorder is a small, lightweight monitor that records ECG continuously. The unit records 1, 2 or 3 channels of ECG up to 9 days. The device is composed of a large ECG electrode and an ECG recorder. The ECG recorder snaps onto the electrode and records until the recorder is removed from the electrode. At the end of the recording, the recorder can be plugged into a PC via a USB cable. The ECG recording can be transferred in an available file format at can be processed through recording processing software provided by recording processing software manufacturers and analyzed by qualified healthcare providers. The ePatch® can be worn continuously up to 5 days in homes, healthcare facilities, hospitals, and wherever the patient may go. The components that are part of the ePatch® recorder are a lead wire, USB cable and the ePatch® electrode.

During the recording the patient wears the ePatch® connected to an electrode on the upper sternum or an alternative placement on the upper torso. The ePatch® consists of a microprocessor, measuring circuit, memory, data storage, lightemitting diode (trans-illuminates the plastic casing), and contract to the electrode. In addition to that, ePatch® contains firmware to control the collection of the ECG data and software to transfer to the processing software.

AI/ML Overview

The provided text does not contain detailed acceptance criteria or a comprehensive study report with specific performance metrics for the ePatch® device. The document is a 510(k) premarket notification letter and summary, which focuses on establishing substantial equivalence to predicate devices rather than providing a standalone performance study.

However, based on the information available, here's what can be extracted and inferred:

1. A table of acceptance criteria and the reported device performance:

The document lists performance standards and categories of testing but does not provide quantitative acceptance criteria or corresponding reported performance values for the ePatch® device itself. Instead, it states that "The technological characteristics of the proposed ePatch® are very similar to the predicate devices." and broadly mentions "Design Verification" as a performance standard.

A table format would look like this, but the "Acceptance Criteria" and "Reported Performance" columns cannot be filled with specific numbers from this document:

Characteristic/Test CategoryAcceptance Criteria (from recognized standards)Reported Device Performance (ePatch®)
Frequency Response(Refer to IEC 60601-2-47 for detailed requirements)0.05-215 Hz
Input Impedance(Refer to IEC 60601-2-47 for detailed requirements)>10MOhms
Resolution(Refer to IEC 60601-2-47 for detailed requirements)12 bit or 16 bit, depending on customer preference
Biological CompatibilityConformance to ISO 10993-1Conforms to ISO 10993-1
Electrical Safety & Essential PerformanceConformance to IEC 60601-1, IEC 60601-2-47, IEC 60601-1-11Conforms to IEC 60601-1, IEC 60601-2-47, IEC 60601-1-11
Electromagnetic Compatibility (EMC)Conformance to IEC 60601-1-2Conforms to IEC 60601-1-2
Software Verification & ValidationConformance to IEC 62304Performed (stated)
UsabilityConformance to IEC 62366Performed (stated)
Patch Adhesion & Signal Quality(Implied to be comparable to predicate devices and acceptable for intended use)Validation conducted to support substantial equivalence (stated)

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

The document mentions "Validation of patch adhesion and signal quality data were conducted to support substantial equivalence" and "Non-clinical testing summary," which includes biocompatibility, electrical safety, EMC, software, and usability testing. However, it does not provide any specific sample sizes for these tests, nor does it detail the data provenance (country of origin, retrospective/prospective).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

This information is not provided in the document. The document describes a medical device for ECG recording, not an AI or diagnostic interpretation system that would typically require expert-established ground truth in the way a diagnostic algorithm would. The "ground truth" for an ECG recorder would be the accurate capture of electrical heart signals, which is validated through technical performance standards rather than expert consensus on diagnostic images or reports.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

This information is not provided and is generally not applicable to the technical validation of an ECG recording device. Adjudication methods are typically employed in studies involving human interpretation or subjective assessments, which are not the primary focus of validating an ECG recorder's hardware and software functionality.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not provided and is not relevant to the ePatch® device as described. The ePatch® is an ECG recorder, not an AI-powered diagnostic interpretation tool. Therefore, an MRMC study comparing human readers with and without AI assistance would not be performed for this type of device.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

This information is not provided and not directly applicable. The ePatch® is a physical device that records ECG data, which is then transferred to a Holter analysis system for review by qualified personnel. While it has firmware and software, it's not described as an "algorithm only" device in the context of standalone diagnostic performance. Its performance is tied to its ability to accurately record physiological signals.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

For the technical performance of the ECG recorder, the "ground truth" is established by conformance to recognized standards (e.g., IEC 60601-2-47 for ambulatory electrocardiographic systems) and the accurate capture of electrical heart activity. This is not typically based on expert consensus, pathology, or outcomes data in the same way a diagnostic algorithm's output would be. The "ground truth" for signal quality means the device faithfully reproduces the biological signal within specified technical tolerances.

8. The sample size for the training set:

This information is not provided. As the ePatch® is an ECG recorder focused on hardware and embedded software for data acquisition and storage, it typically does not involve a "training set" in the machine learning sense. The firmware and software are likely developed and validated through traditional software engineering and testing processes against specifications derived from regulatory standards, rather than being "trained" on a dataset.

9. How the ground truth for the training set was established:

This information is not provided and is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 4, 2018

Braemar Manufacturing, LLC Kent Sayler VP of RA/QA 3890 Murphy Canyon Road, Suite 100 San Diego, California 92123

Re: K171410

Trade/Device Name: ePatch® Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II Product Code: DSH Dated: December 5, 2017 Received: December 6, 2017

Dear Kent Sayler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

M.A. Willemann

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171410

Device Name ePatch®

Indications for Use (Describe)

The ePatch® is indicated for use on patients who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, fatigue, chest pain and/or anxiety.

The ePatch® is intended for use by adolescents 18-21 and adults.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

1. Sponsor

Braemar Manufacturing, LLC 3890 Murphy Canyon Road, Suite 100 San Diego, CA 92123 USA

Contact Person: Kent Sayler 619-243-7560 Telephone:

Date Prepared: May 11, 2017

DEVICE NAME 2.

Classification Name: Recorder, Magnetic Tape, Medical

Classification Regulation: 21 CFR 870.2800

Product code: DSH

Trade / Common Name: ePatch®

3. PREDICATE DEVICES

Primary Predicate:

DMS Service LLC myPatch®s- subject of K160704

Secondary Predicate:

DMS Service LLC myPatch® Ambulatory ECG Recorder- subject of K131190

{4}------------------------------------------------

4. DEVICE DESCRIPTION

The ePatch® ECG Recorder is intended for continuous recording of heart action potentials (ECG). While in use, the sensor is connected to a compatible ECG electrode and placed on the patient.

The ePatch® ECG Recorder is a small digital Holter recorder intended for use by professionals to acquire continuous ambulatory recording of heart action potentials (ECG) from a patient through a compatible ECG patch electrode. The ePatch® is intended for use in a clinical, point of care or at a patient setting. The recording time of the ECG recorder is dependent on the set configuration of the recorder (sample rate and channel selection). The patient's ECG is recorded to the ePatch® ECG Recorder and then transferred via the ePatch® USB cable to a Holter analysis system for review by physician or other qualified personnel.

The ePatch® ambulatory electrocardiograph (ECG) recorder is a small, lightweight monitor that records ECG continuously. The unit records 1, 2 or 3 channels of ECG up to 9 days. The device is composed of a large ECG electrode and an ECG recorder. The ECG recorder snaps onto the electrode and records until the recorder is removed from the electrode. At the end of the recording, the recorder can be plugged into a PC via a USB cable. The ECG recording can be transferred in an available file format at can be processed through recording processing software provided by recording processing software manufacturers and analyzed by qualified healthcare providers. The ePatch® can be worn continuously up to 5 days in homes, healthcare facilities, hospitals, and wherever the patient may go. The components that are part of the ePatch® recorder are a lead wire, USB cable and the ePatch® electrode.

During the recording the patient wears the ePatch® connected to an electrode on the upper sternum or an alternative placement on the upper torso. The ePatch® consists of a microprocessor, measuring circuit, memory, data storage, lightemitting diode (trans-illuminates the plastic casing), and contract to the electrode. In addition to that, ePatch® contains firmware to control the collection of the ECG data and software to transfer to the processing software.

The ePatch® components are described below:

Sensor: The sensor has interfaces to a compatible patch and the ePatch® Micro USB cable. The sensor is rechargeable and reusable. The sensor contains internal non-volatile storage that stores the ECG data until it is emptied or, optionally, until the start of a new recording. The sensor also contains embedded software for recording ECG, storing ECG, and charging. When the sensor is connected to a PC through the ePatch® Micro USB cable, the recorded ECG files are accessible as files on an external drive.

{5}------------------------------------------------

The battery is a rechargeable lithium polymer battery with a built-in protection circuit. The charge of the ePatch® is by use of a USB cable which can only be connected to the ePatch® when it is not connected to the electrode. The ePatch® device must be snapped onto the electrode to turn on and acquire data. The ePatch® has a serial number and product number in a bar code format and numbers to secure the traceability of the equipment and the data during and after the investigation. The ePatch® is a reusable device.

Electrode:

The electrode (accessory to ePatch® ) consists of one layer of one-sided adhesive tape that runs the full length of the electrode. This adhesive tape allows the electrode to be placed on the upper sternum or torso. The electrode gel is placed on the underneath side of the adhesive tape. Through this electrode gel the ECG of the patient is measured. The top side of the electrode has the connector to the ePatch® and one snap for an optional lead wire. The conductor paths are flex strips that run between the connector and the electrode gel. A release liner is placed on the bottom of the electrode. The release liner covers the adhesive tape and the electrode gel. This liner is torn off when placing the ePatch® on the upper sternum. The electrode has to be connected to the ePatch® to have any practical use. The electrode is a single use device and is disposed of after the one single use.

When recording a 3 channel ECG, a standard lead wire is attached to the snap on the front, bottom of the electrode. This lead wire attaches to any standard ECG electrode. The normal placement of this electrode is the V5 position.

ട. INTENDED USE / INDICATIONS FOR USE

The ePatch® is indicated for use on patients who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, fatigue, chest pain and/or anxiety.

The ePatch® is intended for use by adolescents 18-21 and adults.

{6}------------------------------------------------

Technical EquivalenceePatch®ProposedmyPatch®sK160704PrimaryPredicatemyPatchK131190SecondaryPredicate
Indications for UseePatch® is indicated for use on patientswho may be asymptomatic or who maysuffer from transient symptoms such aspalpitations, shortness of breath, dizziness,light headedness, pre-syncope,syncope, fatigue, chest pain and/oranxiety.The ePatch® is intended for use byadolescents 18-21 and adults.myPatch®s is indicatedfor use on patients whomay be asymptomatic orwho may suffer fromtransient symptoms suchas palpitations, shortnessof breath, dizziness, lightheadedness, pre-syncope,syncope, fatigue, chestpain and/or anxiety. ThemyPatch®s is intendedfor use by Adults and allPediatric subgroups.The myPatch is indicatedfor use on patients whomay be asymptomatic orwho may suffer fromtransient symptoms such apalpitations, shortness ofbreath, dizziness, lightheadedness, presyncope,syncope, fatigue, chestpain and/or anxiety.
Type of ECG RecorderPatchPatchPatch
Number ofECGChannels1,2 or 31, 2 or 31 or 2
Wear TimeUp to 5 daysUp to 7 daysUp to 3 days
Recording FormatContinuousContinuousContinuous
Power RequirementLithium PolymerNominal voltage: 3.7VCharging voltage: 4.2VCapacity: Typical 350 mAhLithium PolymerCharging voltage: 3.7vCapacity350 mAhLithium Polymer Chargingvoltage: 3.7v Capacity: 280mAh
Sensor Dimensions10 x 40 x 49mm10 x 40 x 49 mm8.8 x 39.1 x 48mm
Sensor Weight16 grams7 grams15 grams
Frequency Response0.05-215 Hz0.05Hz - 110Hz0.67Hz - 40 Hz
Input Impedance>10MOhms>10MOhms>10MOhms
Resolution12 bit or 16 bit, depending on customerpreference12 bits12 bits
Performance StandardDesign VerificationIEC 60601-2-47Design VerificationIEC 60601-2-47Design VerificationIEC 60601-2-47
SafetyIEC 60601-1 BasicSafety & Essential PerformanceIEC 60601-2-47IEC 60601-1-11IEC 60601-1 BasicSafety & EssentialPerformanceIEC 60601-2-47IEC 60601-1 BasicSafety & EssentialPerformanceIEC 60601-2-47
BiologicalCompatibilityISO 10993-1ISO 10993-1ISO 10993-1
Single Use PatchReusable/Rechargeable Monitor, SingleUse ElectrodeReusable/RechargeableMonitor, Single UseElectrodeReusable/RechargeableMonitor, Single UseElectrode
Recording StandardHolterHolterHolter
Technical EquivalenceePatch®ProposedmyPatch®sK160704PrimaryPredicatemyPatchK131190SecondaryPredicate
Recording FormatContinuousContinuousContinuous
Patch PlacementSternum or appropriate position on torsoLeft SternumSternum
Recording PeriodUp to 9 daysUp to 7 DaysUp to 3 Days

6. TECHNOLOGICAL COMPARISON OF EPATCH® WITH PREDICATE DEVICES

{7}------------------------------------------------

{8}------------------------------------------------

7. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE

The technological characteristics of the proposed ePatch® are very similar to the predicate devices. All of the devices are comprised of essentially identical materials and configurations in that they all consist of a sensor, patch and a USB cable. The design and operational characteristics are identical for the proposed and predicate devices in that they all record ECG up to 9 days (9 days for ePatch® , 7 and 3 days for myPatch/s configurations) on the torso of a patient through a compatible ECG electrode. The proposed and predicate devices provide up to 3 channels with similar weights, frequencies and duration of use. All of these technological characteristics are identical in fit, form and function resulting in the ePatch® being substantially equivalent to the predicate myPatch/s devices.

The operational characteristic of the ePatch® and myPatch/s recorders are identical in that they are all small digital Holter recorder operated by professionals to acquire ECG data from a patient in a clinical, point of care or at a patient setting. The ePatch® ECG Recorder can record ECG up to 9 days on the torso of a patient through the ePatch® compatible ECG electrode (length or recording time is based on the sample rate and channel selection) whereas the predicate myPatch/s devices record from 3-7 days. The patient's ECG results are managed identically for both the ePatch® and the predicate myPatch/s devices in they are recorded to the ECG Recorder and then transferred via USB cable to a Holter analysis system for review by physician or other qualified personnel.

{9}------------------------------------------------

8. PERFORMANCE STANDARDS

The FDA Recognized Standards for which ePatch® ®s conforms to are listed below:

  • . IEC 60601-1-2:2007 (3rd Edition) Clause 6.1 Emission, Clause 6.2 Immunity Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests FDA Recognition #19-1
  • ISO 15223-1:2012 Symbols to be used with medical devices labels, . labeling and information supplied FDA Recognition #5-91
  • IEC 62304:2006 Software Life Cycle Process FDA Recognition #13-32 .
  • . IEC 60601-1-11:2010 Medical Electrical Equipment used in Home Healthcare Environment FDA Recognition #19-6
  • . IEC 60601-1:2012 Medical Electrical Equipment Part 1-2 General requirements for basic safety and essential performance – Collateral standard: Electromagnetic disturbances - Requirements and tests FDA Recognition #19-4
  • . IEC 60601-2-47 Edition 2.0 2012 Ambulatory Electrocardiographic Systems FDA Recognition #3-128
  • . ANSI/AAMI EC12 "Disposable ECG Electrodes" FDA Recognition #3-52
  • . ISO 10993-1:2009 Part1, Biological Evaluation of Medical Devices, Evaluation & Testing within a risk management process; FDA Recognition #2-156
  • ISO 14971:2007 Applications of Risk Management to Medical Devices ● FDA Recognition # 5-70
  • IEC 62366:2008 Application of Usability Engineering to Medical Devices ●

9. NON-CLINICAL TESTING SUMMARY

Biocompatibility, electrical safety, electromagnetic compatibility, software verification and validation and usability testing were performed to support the safety and effectiveness of the ePatch® device.

10. CLINICAL TESTING SUMMARY

{10}------------------------------------------------

Validation of patch adhesion and signal quality data were conducted to support substantial equivalence.

11. SUBSTANTIAL EQUIVALENCE CONCLUSION

The above testing supports that the proposed device is as safe, effective and performs as well as the legally marketed predicate in its intended use. Therefore, the proposed device is substantially equivalent to the predicate.

§ 870.2800 Medical magnetic tape recorder.

(a)
Identification. A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.(b)
Classification. Class II (performance standards).