K Number
K171375

Validate with FDA (Live)

Date Cleared
2018-02-06

(272 days)

Product Code
Regulation Number
878.4460
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Gammex PI Breach Detect Powder Free Surgical Gloves are intended to be worn by operating room personnel to protect a surgical wound from contamination. When used as an underglove, the glove is intended to promote visibility of a fluid that breaches/enters the glove via an increased spreading rate.

Device Description

Gammex Pl Breach Detect Powder Free Surgical Glove is a sterile, green in color, single use disposable powder free surgical glove. The glove is made of synthetic polyisoprene rubber. A polymer is applied to the inner surface of the glove to make donning easy. The glove's outer surface is coated with a surface-active agent and when used as an under-glove, this coating and the glove's green color can improve visibility of a glove breach by promoting wetting of the glove surface and spreading of bodily fluid penetrating the breach.

AI/ML Overview

This document describes a 510(k) premarket notification for a medical device called "Gammex PI Breach Detect Powder Free Surgical Glove" (K171375). However, the document does not contain the kind of detailed information about acceptance criteria and performance study results that would typically be found for an AI/ML-driven medical device, such as a diagnostic algorithm.

The device in question is a surgical glove, which is a Class I device. The FDA's review for Class I devices generally focuses on demonstrating substantial equivalence to a predicate device based on defined physical, material, and biocompatibility characteristics, rather than extensive clinical performance studies or AI model validation.

Therefore, many of the specific questions you've asked (e.g., sample sizes for test sets, number of experts for ground truth, MRMC studies, standalone performance of an algorithm, ground truth for training sets) are not applicable to this device and the information provided in this 510(k) summary.

Here's a breakdown based on the provided text, addressing what is available and explaining why other aspects are absent:

1. A table of acceptance criteria and the reported device performance

The document provides a table outlining "Technological Characteristics" compared to ASTM or equivalent standards. These function as the acceptance criteria for a surgical glove.

CharacteristicsStandard/Test/FDA GuidanceAcceptance Criteria (Implied by Standard)Reported Device Performance (Result Summary)
Physical Characteristics:
Dimensions: LengthASTM D3577-09Minimum 265mmAverage value: Minimum 297 mm (Meets)
Dimensions: Palm Width (mm)ASTM D3577-09Size 5.5: $70 \pm 6$ Size 6.0: $76 \pm 6$ Size 6.5: $83 \pm 6$ Size 7.0: $89 \pm 6$ Size 7.5: $95 \pm 6$ Size 8.0: $102 \pm 6$ Size 8.5: $108 \pm 6$ Size 9.0: $114 \pm 6$Size 5.5: 76 Size 6.0: 82 Size 6.5: 86 Size 7.0: 91 Size 7.5: 98 Size 8.0: 100 Size 8.5: 112 Size 9.0: 120 (Meets)
Dimensions: Thickness (mm) – single wall (Finger, Palm, Cuff)ASTM D3577-09Minimum 0.10 for Finger, Palm, CuffFinger: 0.25 Palm: 0.21 Cuff: 0.17 (Meets)
Physical Properties:ASTM D3577-09Requirements for tensile strength, ultimate elongation, and stress at 500% elongation before and after accelerated aging for synthetic surgical gloves.Meets ASTM D3577-09 requirements (Meets)
Freedom from holesASTM D3577-09, ASTM D5151-06AQL 1.5Meets ASTM D3577-09 and ASTM D5151-06 requirements of AQL 1.5 (Meets)
Powder ResidualASTM D3577-09, ASTM D6124-06≤ 2 mg per gloveMeets applicable requirement for powder free; ≤ 2 mg per glove (Meets)
SterilityANSI/AAMI/ISO 11137-1:2006$10^{-6}$ SAL (Sterility Assurance Level)Meets ANSI/AAMI/ISO 11137-1:2006 requirement of $10^{-6}$ SAL. (Meets)
Biocompatibility:
ISO In Vitro CytotoxicityISO 10993-5:2009Undiluted and 1:2 dilution not cytotoxic.Under the conditions of the study, undiluted and 1:2 dilution were cytotoxic. 1:4, 1:8, 1:16, 1:32 and 1:64 are not cytotoxic. (Note: This result is reported rather than a pass/fail against typical acceptance of "not cytotoxic at all concentrations". However, the overall conclusion for substantial equivalence indicates this was acceptable.)
ISO Skin Irritation StudyISO10993-10:2010Not an irritantUnder the conditions of the study, not an irritant (Meets)
ISO Maximization Sensitization StudyISO 10993-10:2010Not a sensitizerUnder the conditions of the study, not a sensitizer (Meets)
ISO acute systemic toxicityISO 10993-11: 2006No evidence of systemic toxicityUnder the conditions of the study, no evidence of systemic toxicity (Meets)

2. Sample sizes used for the test set and the data provenance:

  • Sample Size: Not explicitly stated as a "test set" in the context of an AI/ML algorithm. The performance data refers to standard compliance tests (e.g., ASTM standards for glove characteristics). These tests involve sampling plans as defined within the respective standards, but the specific number of gloves tested for each characteristic is not provided in this summary.
  • Data Provenance: Not applicable. These are laboratory test results based on manufactured glove samples. The data is not patient-derived from specific countries or retrospective/prospective studies.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not Applicable. This device is a physical product (surgical glove), not an AI/ML diagnostic device requiring expert interpretation of medical images or data for ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not Applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This is not an AI-assisted diagnostic device. The document explicitly states: "A clinical study was not conducted on the subject or predicate devices."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not Applicable. There is no algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • Ground truth for this device is based on established engineering standards and material science properties (e.g., ASTM standards for dimensions, tensile strength, freedom from holes, powder residue; ISO standards for biocompatibility and sterility). The "truth" is whether the glove meets the specified physical and biological performance characteristics.

8. The sample size for the training set:

  • Not Applicable. There is no AI/ML model for which to define a training set.

9. How the ground truth for the training set was established:

  • Not Applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. The word "Administration" is on the second line.

February 6, 2018

Ansell Healthcare Products LLC Don Cronk Sr Manager, QARA / Technical Services 111 Wood Avenue South, Suite 210 Iselin, New Jersey 08830

Re: K171375

Trade/Device Name: Gammex PI Breach Detect Powder Free Surgical Glove Regulation Number: 21 CFR 878.4460 Regulation Name: Surgeon's Glove Regulatory Class: Class I Product Code: KGO Dated: December 20, 2017 Received: December 27, 2017

Dear Don Cronk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171375

Device Name:

Gammex PI Breach Detect Powder Free Surgical Glove

Indications for Use (Describe)

Gammex PI Breach Detect Powder Free Surgical Gloves are intended to be worn by operating room personnel to protect a surgical wound from contamination.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

x | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510k Summary

510(k): K171375

Submitter

Ansell Healthcare Products LLC. 111 Wood Avenue South, Suite 210 Iselin, NJ 08830 USA

Contact Person:

Donald J. Cronk Senior Manager QA/RA/Technical Services Phone: (775) 470-7106 Email: don.cronk@ansell.com

Date Prepared

January 22, 2018

Name of Device

Trade Names:Gammex PI Breach Detect Powder Free Surgical Glove
Common Name:Surgeon's Gloves
Classification Name:Surgeon's Gloves
Classification Regulation:21 CFR 878.4460
Device Class:I
Product Code:KGO
Classification Panel:General and Plastic Surgery

Legally Marketed Predicate Device

K071746 – DermaPrene® Pl or IsoTouch Green Sterile Powder-Free Polyisoprene Surgical Gloves Surgeon's Gloves Common Name: Classification Name: Surgeon's Gloves Classification Regulation: 21 CFR 878.4460 Device Class: -Product Code: KGO Classification Panel: General and Plastic Surgery

Device Description

Gammex Pl Breach Detect Powder Free Surgical Glove is a sterile, green in color, single use disposable powder free surgical glove. The glove is made of synthetic polyisoprene rubber. A polymer is applied to the inner surface of the glove to make donning easy. The glove's outer surface is coated with a surface-active agent and when used as an under-glove, this coating and the glove's green color can improve visibility of a glove breach by promoting wetting of the glove surface and spreading of bodily fluid penetrating the breach.

Indication for Use Statement

Gammex PI Breach Detect Powder Free Surgical Gloves are intended to be worn by operating room personnel to protect a surgical wound from contamination.

{4}------------------------------------------------

Technological Characteristics

Gammex Pl Breach Detect Powder Free Surgical Gloves have the following technological characteristics as compared to ASTM or equivalent standards:

CharacteristicsStandard/Test/FDA GuidanceResult Summary
Physical Characteristics:
DimensionsASTM D3577-09Meets ASTM D3577-09 requirements for length, width and thickness
LengthMinimum 265mmAverage value:Minimum 297 mm
Palm Width (mm)Average value in mm:
Size 5.5$70 \pm 6$76
Size 6.0$76 \pm 6$82
Size 6.5$83 \pm 6$86
Size 7.0$89 \pm 6$91
Size 7.5$95 \pm 6$98
Size 8.0$102 \pm 6$100
Size 8.5$108 \pm 6$112
Size 9.0$114 \pm 6$120
Thickness (mm) – single wallAverage value in mm
FingerMinimum 0.100.25
PalmMinimum 0.100.21
CuffMinimum 0.100.17
Physical Properties:ASTM D3577-09Meets ASTM D3577-09 requirementsfor tensile strength, ultimateelongation and stress at 500%elongation before and afteraccelerated aging for syntheticsurgical gloves
Freedom from holesASTM D3577-09ASTM D5151-06Meets ASTM D3577-09 and ASTMD5151-06 requirements of AQL 1.5
Powder ResidualASTM D3577-09ASTM D6124-06Meets applicable requirement forpowder free; ≤ 2 mg per glove
SterilityANSI/AAMI/ISO 11137-1:2006Meets ANSI/AAMI/ISO 11137-1:2006requirement of $10^{-6}$ SAL.
Biocompatibility
ISO In Vitro CytotoxicityISO 10993-5:2009Under the conditions of the study,undiluted and 1:2 dilution werecytotoxic. 1:4, 1:8, 1:16, 1:32 and 1:64are not cytotoxic
ISO Skin Irritation StudyISO10993-10:2010Under the conditions of the study,not an irritant
ISO MaximizationSensitization StudyISO 10993-10:2010Under the conditions of the study,not a sensitizer
ISO acute systemictoxicityISO 10993-11: 2006Under the conditions of the study, noevidence of systemic toxicity

{5}------------------------------------------------

Substantial Equivalence

Predicate DeviceSubject DeviceSubstantialEquivalence
Trade NameDermaPrene® PI or IsoTouchGreen Sterile Powder-FreePolyisoprene Surgical GlovesGammex® PI Breach DetectPowder Free Surgical GloveNot applicable
510(k) NumberK071746K171375Not applicable
SubmitterAnsell Healthcare Products LLCAnsell Healthcare Products LLCSame
Product CodeKGOKGOSame
RegulationNumber21 CFR 878.446021 CFR 878.4460Same
RegulationNameSurgeon's GloveSurgeon's GloveSame
Regulatory ClassClass IClass ISame
Device TypeType 2Gloves compounded primarilyfrom rubber cement or syntheticrubber latexType 2Gloves compounded primarilyfrom rubber cement orsynthetic rubber latexSame
Indications foruseThese gloves are intended to beworn by operating roompersonnel to protect a surgicalwound from contaminationThese gloves are intended tobe worn by operating roompersonnel to protect a surgicalwound from contamination.When used as an underglove,the glove is intended topromote visibility of a fluid thatbreaches/enters the glove viaan increased spreading rate.Different
Prescription orOver-TheCounter-UseOver-The-Counter-UseOver-The-Counter-UseSame
MaterialSynthetic Polyisoprene RubberSynthetic Polyisoprene RubberSame
Coating (inner)Polyurethane polymer innercoating to aid donningPolyurethane polymer innercoating to aid donningSame
Coating (outer)NoneProprietary solution to aid inwetting/spreading of fluidwhen used as an undergloveDifferent
and the outer surgeon's gloveis breached.
DesignSingle useSingle useSame
SterileSterileSame
Powder-FreePowder-FreeSame
Hand-specificHand-specificSame
Beaded cuffBeaded cuffSame
ColorGreenGreenSame
Dimension &physicalpropertiesMeets ASTM D3577-09requirementsMeets ASTM D3577-09requirementsSame
Freedom fromholesMeets ASTM D3577-09requirements of AQL 1.5Meets ASTM D3577-09requirements of AQL 1.5Same
Powder-FreeMeets applicable definition forpowder free; ≤ 2mg per gloveMeets applicable definition forpowder free; ≤ 2mg per gloveSame
SterilizationmethodRadiationRadiationSame
SterilityAssurance Level(SAL)10-6 SAL10-6 SALSame
Shelf life3 years3 yearsSame
Biocompatibility"Under the conditions of thestudy, not an irritant" and"Under the conditions of thestudy, not a sensitizer""Under the conditions of thestudy, not an irritant" and"Under the conditions of thestudy, not a sensitizer"Same

Substantial Equivalence Comparison Table

{6}------------------------------------------------

The subject device is manufactured from polyisoprene with polyurethane polymer inner coating to aid donning. The glove's outer surface is coated with a surface-active agent and when used as an under-glove, this coating and the glove's green color can improve visibility of a glove breach by promoting wetting of the glove surface and spreading of bodily fluid penetrating the breach. Aside from the outer surface treatment, the subject device's materials are functionally equivalent to those of the cited predicate.

The subject device meets the applicable requirements for surgeon's gloves regarding dimensions and sizes, physical properties, freedom from holes, and powder residues as found in the following standards: ASTM D3577, ASTM D5151 and ASTM D6124. The subject device passes biological reactivity testing for dermal sensitization and irritation, in accord with the ISO 10993 series of standards.

Performance Data

A clinical study was not conducted on the subject or predicate devices.

Substantial Equivalence Statement

{7}------------------------------------------------

The Gammex PI Breach Detect Powder Free Surgeon's Glove are as safe and effective as the predicate device with respect to design, technological characteristics, intended use and conformance to standard requirements.

Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed device, K071746

§ 878.4460 Non-powdered surgeon's glove.

(a)
Identification. A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls).