K Number
K171365

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-11-07

(182 days)

Product Code
Regulation Number
888.3565
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Arthrex iBalance BiCompartmental Arthroplasty System is intended to be used as a multicompartmental knee arthroplasty in patients with:

  • Moderately disabling joint disease of the knee resulting from painful osteoarthritis or post traumatic arthritis;
  • Correction of functional deformities;
  • Revision of previous unsuccessful partial knee replacement or other procedure;
    The BiCompartmental Arthroplasty System is not intended to be used as a dual-condyle or tri-compartmental knee.
    These components are single use only and are intended for implantation with bone cement.

The Arthrex iBalance PFJ components are intended for use in patellofemoral knee arthroplasty in patients with:

  • Degenerative arthritis in the distal femur and patella;
  • A history of patellar dislocation or fracture;
  • Failed previous surgery (arthroscopy, tibial tubercle elevation, lateral release) where pain, deformity or dysfunction persists.
    These components are single use only and are intended for implantation with bone cement.

The Arthrex iBalance TKA System is indicated for use in individuals undergoing surgery for:

  • Painful, disabling joint disease of the knee resulting from degenerative arthritis, rheumatic arthritis:
  • Post-traumatic loss of knee joint configuration and function
  • Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability;
  • Revisions of previous unsuccessful knee replacement or other procedure.
    Additional indications for posteriorly stabilized components:
  • Ligamentous instability requiring implant bearing surfaces with increased constraint;
  • Absent or non-functioning posterior cruciate ligament.
    These devices are single use only and are intended for implantation with the exception of porous coated femoral components which can be used cemented (biological fixation).

Arthrex iBalance UKA System is intended for use in unicompartmental knee arthroplasty as a result of.

  • Moderately disabling joint disease of the knee resulting from painful osteoarthritis or post traumatic arthritis;
  • Correction of functional deformities;
  • Revision of previous unsuccessful unicompartmental knee replacement or other procedure;
  • As an alternative to tibial osteotomy in patients with unicompartmental osteoarthritis.
    These components are single use only and are intended for implantation with bone cement.
Device Description

The Arthrex Knee Systems encompass the following FDA cleared devices:

  • Arthrex iBalance TKA System
  • Arthrex iBalance UKA System
  • Arthrex iBalance PFJ
  • Arthrex iBalance BiCompartmental Arthroplasty System
    These systems are comprised of femoral, tibia, and patellar components for use in knee prosthesis.
AI/ML Overview

The provided document is a 510(k) premarket notification for Arthrex Knee Systems. It states that this particular submission (K171365) is a special 510(k) for a labeling modification to include an "MR Conditional" statement. It references general intended uses and provides some non-clinical testing for "MR Conditional" status and bacterial endotoxin but does not describe a study that proves the device meets specific performance acceptance criteria related to its clinical function, such as efficacy or longevity in patients.

Therefore, I cannot provide the requested information in full. However, I can extract the information that is present regarding the non-clinical tests mentioned.

Here's what I can provide based on the given document:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
For MR Conditional Status:Non-clinical testing and in-vivo electromagnetic simulation per the FDA Guidance, "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment", demonstrated that the Arthrex Knee Systems are "MR Conditional".
For Pyrogen Limit Specifications:Bacterial endotoxin per EP 2.6.14/USP <85> was conducted to demonstrate that the device meets pyrogen limit specifications.

Since this 510(k) is for a labeling change to include "MR Conditional" status and mentions bacterial endotoxin testing, it does not describe a clinical study for proving patient-related performance or effectiveness. Consequently, the following requested information cannot be found in the provided text:

  • Sample size used for the test set and the data provenance: Not applicable for the described tests (MR Conditional and bacterial endotoxin). These are non-clinical or laboratory tests.
  • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable for the described tests.
  • Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable for the described tests.
  • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document is not about AI or diagnostic imaging.
  • If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This document is not about an algorithm or AI.
  • The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For the "MR Conditional" status, the ground truth is established by meeting criteria specified in the FDA Guidance for MR environments. For bacterial endotoxin, the ground truth is meeting pyrogen limit specifications. This doesn't involve clinical outcomes or expert consensus in the typical sense of evaluating device effectiveness for its primary medical purpose.
  • The sample size for the training set: Not applicable. There is no mention of a training set for an algorithm.
  • How the ground truth for the training set was established: Not applicable.

This 510(k) notification focuses on demonstrating substantial equivalence to predicate devices, particularly regarding safety aspects relevant to an MRI environment and pyrogenicity for sterile devices, rather than a de novo clinical study proving the effectiveness of the knee system itself. The phrase "Performance Data" section specifically highlights "Non-clinical testing and in-vivo electromagnetic simulation" and "Bacterial endotoxin" tests.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left, and the U.S. Food & Drug Administration logo is on the right. The FDA logo is blue and white and includes the words "FDA U.S. FOOD & DRUG ADMINISTRATION".

November 7, 2017

Arthrex Inc. David L. Rogers Project Manager, Regulatory Affairs 1370 Creekside Boulevard Naples, Florida 34108-1945

Re: K171365

Trade/Device Name: Arthrex Knee Systems Regulation Number: 21 CFR 888.3565 Regulation Name: Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: MBH, HSX, KRR Dated: October 3, 2017 Received: October 11, 2017

Dear Mr. Rogers:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

Page 2 - Mr. David Rogers

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K171365

Device Name

Arthrex iBalance BiCompartmental Arthoplasty System

Indications for Use (Describe)

When used concurrently, the Arthrex iBalance UKA and PFJ systems create the Arthrex iBalance BiCompartmental Arthroplasty System. The Arthrex iBalance BiCompartmental Arthroplasty System is intended to be used as a multicompartmental knee arthroplasty in patients with:

  • Moderately disabling joint disease of the knee resulting from painful osteoarthritis or post traumatic arthritis;
  • · Correction of functional deformities;
  • · Revision of previous unsuccessful partial knee replacement or other procedure;

The BiCompartmental Arthroplasty System is not intended to be used as a dual-condyle or tri-compartmental knee.

These components are single use only and are intended for implantation with bone cement.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{3}------------------------------------------------

Indications for Use

510(k) Number (if known)

K171365

Device Name Arthrex iBalance PFJ

Indications for Use (Describe)

The Arthrex iBalance PFJ components are intended for use in patellofemoral knee arthroplasty in patients with:

  • · Degenerative arthritis in the distal femur and patella;
  • · A history of patellar dislocation or fracture;
  • Failed previous surgery (arthroscopy, tibial tubercle elevation, lateral release) where pain, deformity or dysfunction persists.

These components are single use only and are intended for implantation with bone cement.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{4}------------------------------------------------

Indications for Use

510(k) Number (if known)

K171365

Device Name Arthrex iBalance TKA System

Indications for Use (Describe)

The Arthrex iBalance TKA System is indicated for use in individuals undergoing surgery for:

· Painful, disabling joint disease of the knee resulting from degenerative arthritis, rheumatic arthritis:

  • · Post-traumatic loss of knee joint configuration and function
  • · Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability;
  • · Revisions of previous unsuccessful knee replacement or other procedure.

Additional indications for posteriorly stabilized components:

  • · Ligamentous instability requiring implant bearing surfaces with increased constraint;
  • · Absent or non-functioning posterior cruciate ligament.

These devices are single use only and are intended for implantation with the exception of porous coated femoral components which can be used cemented (biological fixation).

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CER 801 Subpart D)
Over-The-Counter Use (21 CER 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{5}------------------------------------------------

Indications for Use

510(k) Number (if known) K171365

Device Name Arthrex iBalance UKA System

Indications for Use (Describe)

Arthrex iBalance UKA System is intended for use in unicompartmental knee arthroplasty as a result of.

  • Moderately disabling joint disease of the knee resulting from painful osteoarthritis or post traumatic arthritis;
  • · Correction of functional deformities;
  • · Revision of previous unsuccessful unicompartmental knee replacement or other procedure;
  • · As an alternative to tibial osteotomy in patients with unicompartmental osteoarthritis.

These components are single use only and are intended for implantation with bone cement.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{6}------------------------------------------------

510(k) Summary or 510(k) Statement

Date PreparedMay 5, 2017
SubmitterArthrex Inc.
1370 Creekside Boulevard
Naples, FL 34108-1945
Contact PersonDavid L Rogers
Project Manager, Regulatory Affairs
1-239-643-5553, ext. 71924
david.rogers@arthrex.com
Name of DeviceArthrex Knee Systems
Common NameKnee Prosthesis
Product CodeMBH, HSX, KRR
Classification Name21 CFR 888.3565: Knee joint patellofemorotibial metal/polymer porous-coated
uncemented prosthesis
21 CFR 888.3560: Prosthesis, Knee, Patellofermorotibial, Semi-Constrained,
Cemented, Polymer/Metal/Polymer
21 CFR 888.3520: Prosthesis, Knee, Femorotibial, Non-Constrained,
Metal/Polymer
Regulatory ClassII
Predicate DeviceK161060: Arthrex iBalance UKA System Vitamin E Tibial Bearing
K160461: Arthrex iBalance BiCompartmental Arthroplasty System
K153586: Arthrex iBalance TKA System
K152382: Arthrex iBalance TKA System
K152252: Arthrex iBalance TKA System
K143047: Arthrex iBalance Patella Implant, Dome
K133854: Arthrex iBalance TKA System
K073120: ACCIN Patellafemoral System
K063782: ACCIN UNI-Knee System
K060670: ACCIN UNI-Knee System
Purpose ofSubmissionThis Special 510(k) premarket notification is submitted to obtain clearance for the
modification to the device labeling for the Arthrex Knee Systems to include the
"MR Conditional" statement in accordance with the FDA Guidance, "Establishing
Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR)
Environment".
Device DescriptionThe Arthrex Knee Systems encompass the following FDA cleared devices:
● Arthrex iBalance TKA System
● Arthrex iBalance UKA System
● Arthrex iBalance PFJ
● Arthrex iBalance BiCompartmental Arthroplasty System
These systems are comprised of femoral, tibia, and patellar components for use
in knee prosthesis.
Indications for UseThe Arthrex iBalance TKA System is indicated for use in individuals undergoing
surgery for:
● Painful, disabling joint disease of the knee resulting from degenerative
arthritis, rheumatoid arthritis or post-traumatic arthritis;
● Post-traumatic loss of knee joint configuration and function
● Moderate varus, valgus, or flexion deformity in which the ligamentous
structures can be returned to adequate function and stability;

{7}------------------------------------------------

  • Revisions of previous unsuccessful knee replacement or other procedure.
    Additional indications for posteriorly stabilized components:

  • Ligamentous instability requiring implant bearing surfaces with ● increased constraint;

  • Absent or non-functioning posterior cruciate ligament.

These devices are single use only and are intended for implantation with bone cement, with the exception of porous coated femoral components which can be used cemented or uncemented (biological fixation).

Arthrex iBalance UKA System is intended for use in unicompartmental knee arthroplasty as a result of:

  • Moderately disabling joint disease of the knee resulting from painful osteoarthritis or post traumatic arthritis;
  • Correction of functional deformities;
  • Revision of previous unsuccessful unicompartmental knee replacement ● or other procedure;
  • . As an alternative to tibial osteotomy in patients with unicompartmental osteoarthritis.

These components are single use only and are intended for implantation with bone cement.

The Arthrex iBalance PFJ components are intended for use in patellofemoral knee arthroplasty in patients with:

  • Degenerative arthritis in the distal femur and patella;
  • A history of patellar dislocation or fracture;
  • Failed previous surgery (arthroscopy, tibial tubercle elevation, lateral release) where pain, deformity or dysfunction persists.

These components are single use only and are intended for implantation with bone cement.

When used concurrently, the Arthrex iBalance UKA and PFJ systems create the Arthrex iBalance BiCompartmental Arthroplasty System. The Arthrex iBalance BiCompartmental Arthroplasty System is intended to be used as a multicompartmental knee arthroplasty in patients with:

  • Moderately disabling joint disease of the knee resulting from painful ● osteoarthritis or post traumatic arthritis;
  • Correction of functional deformities;
  • Revision of previous unsuccessful partial knee replacement or other procedure;

The BiCompartmental Arthroplasty System is not intended to be used as a dualcondyle or tri-compartmental knee.

These components are single use only and are intended for implantation with bone cement.

Performance Data Non-clinical testing and in-vivo electromagnetic simulation per the FDA Guidance, "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment", demonstrated that the Arthrex Knee Systems are "MR Conditional".

{8}------------------------------------------------

Bacterial endotoxin per EP 2.6.14/USP <85> was conducted to demonstrate thatthe device meets pyrogen limit specifications.
ConclusionThe Arthrex Knee Systems are substantially equivalent to the predicate device inwhich the basic design features and intended uses are the same. Any differencesbetween the proposed device and the predicate device are considered minor anddo not raise questions concerning safety or effectiveness.
Based on the indications for use, technological characteristics, and the summaryof data submitted, Arthrex Inc. has determined that the proposed device issubstantially equivalent to the currently marketed predicate device.

N/A