(206 days)
EZFluence is intended to assist radiation treatment planning professionals in generating optimal fluences for producing a homogeneous dose distribution in external beam radiation therapy treatment plans consisting of photon treatment fields.
The EZFluence device (model RADEZ) is software is intended to assist radiation treatment planning professionals in generating optimal fluences for producing a homogeneous dose distribution in external beam radiation therapy treatment plans consisting of photon treatment fields. Inputs are obtained from plan and patient data obtained from the Eclipse Treatment Planning System (also referred to as Eclipse TPS) of Varian Medical Systems. EZFluence runs as a dynamic link library (DLL) plugin to Varian Eclipse. It is designed to run on the Windows Operating System. EZFluence performs calculations on the plan obtained from Eclipse TPS (Version 13.5 (K141283) and Version 13.7 (K152393)) which is a software device used by trained medical professionals to design and simulate radiation therapy treatment plans for malignant or benign diseases.
The document provided discusses the EZFluence device, a software intended to assist radiation treatment planning professionals. However, it does not contain specific details about acceptance criteria, reported device performance figures (like sensitivity, specificity, or accuracy), sample sizes for test or training sets, data provenance, the number or qualifications of experts, ground truth establishment, or any MRMC studies.
The document primarily focuses on establishing substantial equivalence to a predicate device (Eclipse Treatment Planning System) based on similar indications for use and technological characteristics. It mentions that "Verification tests were performed to ensure that the software works as intended and pass/fail criteria were used to verify requirements," but it does not elaborate on these criteria or the results.
Therefore, based solely on the provided text, I cannot generate a table of acceptance criteria and reported performance, nor can I answer many of the specific questions about the study design and results, as this information is not present in the excerpt.
Here's what I can extract and what's missing:
Inability to Fulfill Request Due to Lack of Information
The provided document (K171352) is a 510(k) summary for the EZFluence device. While it describes the device's intended use and compares it to a predicate device, it does not contain the detailed performance data, acceptance criteria, sample sizes, expert qualifications, or ground truth methodology that would be required to answer the questions in the prompt.
The document explicitly states under "5.7 Performance Data": "As with the Predicate Device, no clinical trials were performed for EZFluence. Verification tests were performed to ensure that the software works as intended and pass/fail criteria were used to verify requirements." However, it does not provide the specific "pass/fail criteria" or the results of these "verification tests."
Therefore, I cannot construct the requested table or provide answers to most of the specific questions.
What can be inferred or directly stated from the document:
- Device Name: EZFluence
- Intended Use: To assist radiation treatment planning professionals in generating optimal fluences for producing a homogeneous dose distribution in external beam radiation therapy treatment plans consisting of photon treatment fields. (Section 5.5)
- Regulatory Class: Class II (Section 5.2)
- Predicate Device: Eclipse Treatment Planning System (K152393) (Section 5.3)
- Study Type: Verification tests were performed; no clinical trials were performed. (Section 5.7)
- Performance Metrics Reported: None explicitly stated (e.g., no sensitivity, specificity, accuracy, or any quantitative metric of "optimal fluence" or "homogeneous dose distribution" quality).
Missing Information (Cannot be answered from the provided text):
- A table of acceptance criteria and the reported device performance: This information is not present. The document mentions "pass/fail criteria were used to verify requirements" but does not detail them or the results.
- Sample sizes used for the test set and the data provenance: Not mentioned.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not mentioned.
- Adjudication method for the test set: Not mentioned.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: The document explicitly states "no clinical trials were performed for EZFluence," suggesting no such MRMC study was conducted. The device is also described as assisting professionals, not as an AI-driven diagnostic tool for human readers in the typical MRMC context.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not explicitly detailed. The "verification tests" mentioned are likely standalone software tests, but no performance metrics are given.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not mentioned.
- The sample size for the training set: Not mentioned.
- How the ground truth for the training set was established: Not mentioned.
Conclusion based on provided text: The document serves as a regulatory submission demonstrating substantial equivalence rather than a detailed scientific study report detailing performance metrics and validation methodologies.
{0}------------------------------------------------
December 1, 2017
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Radformation, Inc. % Mr. Kurt Sysock Co-founder/CEO 335 Madison Avenue, 16th Floor NEW YORK NY 10017
Re: K171352
Trade/Device Name: EZFluence Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE, MUJ Dated: October 4, 2017 Received: October 11, 2017
Dear Mr. Sysock:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael D. O'Hara For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
Device Name EZFluence
Indications for Use (Describe)
EZFluence is intended to assist radiation treatment planning professionals in generating optimal fluences for producing a homogeneous dose distribution in external beam radiation therapy treatment plans consisting of photon treatment fields.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Section 5. 510(k) Summary
. . . . . . . .
:
Radformation, Inc.
:
510(k) Submission
{4}------------------------------------------------
This 510(k) Summary has been created per the requirements of the Safe Medical Device Act (SMDA) of 1990, and the content is provided in conformance with 21 CFR Part 807.92.
| Table 1 : Submitter's Information | |
|---|---|
| Submitter's Name: | Kurt Sysock |
| Company: | Radformation, Inc. |
| Address: | 335 Madison Avenue, 16th FloorNew York, NY 10017 |
| Contact Person: | Alan NelsonChief Science Officer, Radformation |
| Phone: | 518-888-5727 |
| Fax: | ---------- |
| Email: | anelson@radformation.com |
| Date of Summary Preparation | 05/01/2017 |
5.1. Submitter's Information
5.2. Device Information
| Table 2 : Device Information | |||
|---|---|---|---|
| Trade Name: | EZFluence | ||
| Common Name: | Oncology Information System | ||
| Classification Name: | Class II | ||
| Classification: | Medical charged-particle radiation therapy system,dosimetric quality control system | ||
| Regulation Number: | 892.5050 | ||
| Product Code:: | IYE | ||
| Classification Panel: | Radiology |
5.3. Predicate Device Information
Eclipse Treatment Planning System (K152393)
{5}------------------------------------------------
5.4. Device Description
The EZFluence device (model RADEZ) is software is intended to assist radiation treatment planning professionals in generating optimal fluences for producing a homogeneous dose distribution in external beam radiation therapy treatment plans consisting of photon treatment fields. Inputs are obtained from plan and patient data obtained from the Eclipse Treatment Planning System (also referred to as Eclipse TPS) of Varian Medical Systems. EZFluence runs as a dynamic link library (DLL) plugin to Varian Eclipse.
It is designed to run on the Windows Operating System. EZFluence performs calculations on the plan obtained from Eclipse TPS (Version 13.5 (K141283) and Version 13.7 (K152393)) which is a software device used by trained medical professionals to design and simulate radiation therapy treatment plans for malignant or benign diseases.
5.5. Indications for Use
EZFluence is intended to assist radiation treatment planning professionals in generating optimal fluences for producing a homogeneous dose distribution in external beam radiation therapy treatment plans consisting of photon treatment fields.
5.6. Technological Characteristics
EZFluence (Subject Device) makes use of a Predicate Device, Eclipse Treatment Planning System (K152393) for substantial equivalence comparison.
5.6.1. EZFluence vs. Eclipse Treatment Planning System (K152393)
The Eclipse Treatment Planning System (TPS) External Beam Planning workspace has two features that are considered substantially equivalent with the functionality provided by EZFluence:
-
- Irregular Surface Compensator
-
- Optimal Fluence Editing
| Table 3: Substantial Equivalence Table for EZFluence | |||
|---|---|---|---|
| Parameters | Subject Device: EZFluence Radformation | Predicate Device: Eclipse Treatment Planning System, Varian Medical Systems (K152393) | Equivalence |
| Indications for use | Used to assist radiation treatment planners in generating optimal fluences for external beam radiation therapy treatment plans with photon treatment fields. | Used to plan radiotherapy treatments for patients with malignant or benign diseases. | Subject Device is a subset of the Predicate Device |
{6}------------------------------------------------
| Pure software | Yes | Yes | Equivalent |
|---|---|---|---|
| Intended users | Trained medicalprofessionals to design andsimulate radiation therapytreatments | Trained medical professionalsto design and simulateradiation therapy treatments | Equivalent |
| OTC/Rx | Rx | Rx | Equivalent |
| EZFluence vs. Eclipse Irregular Surface Compensator | |||
| Functionality | Optimizes optimal fluencesfor treatment fields to obtaina homogeneous dose in themiddle of the patient and tocontrol the maximum doseallowed in the patient. | Optimizes optimal fluences fora specific field to obtain ahomogeneous dose at a planewith depth specified by theuser. | Minordifferences |
| EZFluence vs. Eclipse Optimal Fluence Editing | |||
| Functionality | Allows user to manuallychange fluence intensities ofan optimal fluence. | Allows user to manuallychange fluence intensities ofan optimal fluence. | Equivalent |
As shown in Table 3, EZFluence's indications for use are a subset of Eclipse TPS's indication for use, and the core functionality provided by EZFluence is substantially equivalent with Eclipse TPS's Irregular Surface Compensator and Optimal Fluence Editing features.
5.6.2. Differences
The minor differences between EZFluence and the Irregular Surface Compensator are as follows:
-
- EZFluence automatically preserves static multi-leaf collimator blocking by assigning 0 fluence intensity wherever the field was blocked while Eclipse Irregular Surface Compensator does not. By doing this, EZFluence increases safety and efficiency as radiation treatment planners generally have to manually set fluence intensities to 0 wherever there were blocks in the original treatment plan (a step that can easily be forgotten, inadvertently increasing dose, in breast radiotherapy for example, to the heart).
-
- EZFluence automatically adds "flash" wherever the field does not intersect the body by extending the fluence nearest to the edge of the patient to the edge of the field. High quality breast radiation therapy, for example, always requires flash in order to ensure the breast is adequately dosed with radiation even when there are day-to-day setup errors. The Eclipse Irregular Surface Compensator does not account for flash and so
{7}------------------------------------------------
the user must manually add flash in order to ensure high quality treatment when they use it for breast treatments. In this way, EZFluence's differences from the Irregular Surface Compensator only increase safety relative to the predicate device.
-
- Because EZFluence simultaneously generates optimal fluences for multiple beams instead of for single beams as the Eclipse Irregular Surface Compensator does, EZFluence produces optimal fluences that result in a generally more acceptable and high quality homogeneous dose distribution. Thus the differences between EZFluence and Eclipse Irregular Surface Compensator represent an increase in safety and quality by EZFluence.
- EZFluence allows the user to limit the maximum dose to the patient while 4. the Eclipse Irregular Surface Compensator does not have that feature.
5.7. Performance Data
As with the Predicate Device, no clinical trials were performed for EZFluence. Verification tests were performed to ensure that the software works as intended and pass/fail criteria were used to verify requirements.
5.8 Conclusion
EZFluence is deemed substantially equivalent to the Predicate Device, Eclipse Treatment Planning System (K152393) due to the similarities with the Irregular Surface Compensator and Optimal Fluence Editing features of Eclipse. Verification and Validation testing and Hazard Analysis demonstrate that EZFluence is as safe and effective as the Predicate Device. The minor technological differences between EZFluence and the Predicate Device do not raise any guestions on the safety and effectiveness of the Subject Device.
N/A