K Number
K171245

Validate with FDA (Live)

Device Name
Presygen/si-1
Date Cleared
2018-08-09

(468 days)

Product Code
Regulation Number
870.2700
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Presygen™/si-1 is indicated for use in hospital facilities, along with clinical assessment of signs and symptoms, by healthcare professionals for non-invasive and non-diagnostic measure of relative temporal variation in oxygen saturation (StO2) in superficial tissue. Presygen™/si-1 displays two-dimensional color-coded images of tissue oxygenation of the scanned surface and reports tissue oxygenation measurements in selected tissue regions. Presygen™/si-1 is intended for adults only (22 and over).

Device Description

Presygen™/si-1 is a standalone, non-contact tissue oxygen saturation oximeter based on multispectral optical imaging techniques. The mode of action is to transmit lights at four (4) known wavelengths through blood in superficial tissue and measure the tissue oxygen saturation (StO2), based on the amount of diffusely reflected light.

The Presygen™/si-1 software provides user control of the device and access to stored data, through a Graphical User Interface (GUI). The software controls hardware elements of the device such as illuminators, sensors, display and Input/Output, and manages communication between these elements and between processors and memory. The software also performs calculations to derive Tissue Oxygen Saturation (StO2) from optical inputs and display this to the device user in a two-dimension (2D) spatial map.

The Presygen™/si-1 device has been designed with the additional capability of compensating for melanin levels in the skin, since absorption of light by melanin affects the measurement of the oxygenated hemoglobin and deoxygenated hemoglobin through the skin.

The device consists of the following:

  • Imaging head: Contains optics, lasers, and electronics to display the %StO2 .
  • . Base compartment: Includes the wheels, brakes mechanism, SMB (Smart Multi-Media Board), power switch, certified power supply, fans, filters and cabling
  • . Touchscreen monitor display: User interface, display of image, and patient data
  • Keyboard: Used to enter user or patient information into the device .
  • Accessory: The Presygen™/si-1 Sterilized Drape is a prepackaged single-use third party . drape that is used with Presygen™/si-1 to avoid contamination during the imaging process.
AI/ML Overview

The provided text is a 510(k) summary for the Presygen™/si-1 device, an Oximeter. It details the steps taken to demonstrate the device's substantial equivalence to a predicate device (Kent Camera, K113507) and a reference device (T.Ox™, K042657).


1. Acceptance Criteria and Reported Device Performance

The general acceptance criteria for this device seem to be based on demonstrating substantial equivalence to predicate and reference devices, particularly in the measurement of tissue oxygen saturation (StO2) and its temporal variation. Performance is evaluated through phantom studies and clinical studies, focusing on accuracy, uniformity, and consistency with existing devices, especially across different Fitzpatrick Skin Types.

While direct numerical acceptance criteria for various metrics are not explicitly laid out in a table format in the provided text, the results of the studies serve as the reported device performance, indicating whether these (often implicit) criteria were met.

Here's an attempt to structure the information based on the reported results, interpreting the criteria from the outcomes:

Metric / Acceptance CriteriaReported Device Performance (Summary)
Phantom Study (Accuracy & Spatial Uniformity)
Measurement accuracy (StO2 % range 50%-90%) and spatial uniformity.Results: Significant linear trend of %StO2 in actual vs. theoretical values (target confidence 95%). Results acceptably uniform over the device field of view.
Clinical Study 1 (Presygen™/si-1 vs. Kent Camera (Predicate), & T.Ox™ (Reference))
Primary Equivalence (Presygen™/si-1 vs. Kent Camera in Fitzpatrick I & II):
- Agreement: >95% of observations within Limits of Agreement (Bland-Altman).Results: Presygen™/si-1 compared to Kent Camera in Fitzpatrick Scale I with >95% of observations within the Limits of Agreement on the Bland-Altman plot.
- Slope (Deming Regression): 95% Confidence Limits including 1.Results: Deming Regression with bootstrap resampling results showing a slope with 95% confidence limits including 1.
- Intercept (Deming Regression): 95% Confidence Limits including 0 (indicating no bias).Results: The intercept for the Deming regression equation did not contain 0, demonstrating a small positive bias in the measurement of StO2. (This indicates this specific criterion was not fully met, but the overall conclusion was still substantial equivalence).
- Precision (Coefficient of Variation): Within 2 standard deviations of the mean for >95% of observations.Results: Precision found coefficient of variation was within 2 standard deviations of the mean for >95% of all observations in StO2 measurement in Fitzpatrick I and II.
Vascular Occlusion Test (VOT) Consistency: Similar reactions to ischemia and reperfusion.Results: Presygen™/si-1, Kent Camera, and T.Ox™ all demonstrate similar reactions to ischemia and reperfusion in the measurement of StO2. VOT graphs show similar patterns of baseline, response to ischemia and reperfusion.
Re-analysis of VOT (Presygen™/si-1 vs. T.Ox™ for temporal ΔStO2): Similar temporal StO2 change.Results: Bland-Altman plot showed 91% of observations within Limits of Agreement for ΔStO2. Mean bias close to 0 (0.110 %StO2). Deming regression for ΔStO2 shows equivalence to T.Ox™ with intercepts close to 0 and slopes close to 1.
Clinical Study 2 (Presygen™/si-1 vs. Kent Camera & T.Ox™):
Equivalence across Fitzpatrick Scales (II-V) for Presygen™/si-1 vs. T.Ox™: All observations within Limits of Agreement.Results: StO2 values with Presygen™/si-1 and T.Ox™ show equivalence across all Fitzpatrick Scales including II-V, with all observations within the Limits of Agreement on the Bland-Altman plot.
Consistency in VOT response for all devices.Results: Presygen™/si-1, Kent Camera and T.Ox™ all demonstrate similar fluctuations in response to ischemia and reperfusion. Data shows stable baseline, decreasing through ischemia, and rebounding during reperfusion.
Re-analysis of VOT (Presygen™/si-1 vs. T.Ox™ for temporal ΔStO2): Similar temporal StO2 change.Results: Bland-Altman plot showed 90% of observations within Limits of Agreement for ΔStO2. Mean bias close to 0 (0.016 %StO2). Deming regression for ΔStO2 shows equivalence with intercepts close to 0 and slopes close to 1.
Safety: No Serious Adverse Events (SAE) or Adverse Events (AE).Results: Throughout both studies, there were no SAE or AE observed.

2. Sample Sizes and Data Provenance

  • Clinical Study 1:
    • Sample Size: 55 subjects (199 data points). 52 subjects included in data analysis, 3 excluded due to incomplete data.
    • Data Provenance: Prospective, unblinded, convenience sample population volunteers from the general community. No specific country of origin is mentioned, but typically for FDA submissions, studies conducted in the US are preferred, or international studies must meet FDA standards. Given the context of a US FDA submission, it's highly likely to be US-based or globally recognized clinical research. Retrospective/Prospective: Prospective.
  • Clinical Study 2:
    • Sample Size: 20 subjects (6 male, 14 female).
    • Data Provenance: Prospective, unblinded, convenience sample population who self-reported as being healthy. Laboratory setting. Retrospective/Prospective: Prospective.
  • Phantom Study for Bench Validation:
    • Sample Size: 3 phantom sets.
    • Data Provenance: Bench study, independent validation.

3. Number of Experts and Qualifications for Ground Truth

The document does not mention the use of experts to establish a "ground truth" in the traditional sense for the clinical studies. Instead, the ground truth for performance comparison is established by comparing the device's measurements (StO2) against measurements from a legally marketed predicate device (Kent Camera) and a reference device (T.Ox™ Tissue Oximeter). The evaluation is based on statistical equivalence of these measurements rather than expert consensus on images or findings.

For the phantom study, the "ground truth" was established by spectrophotometer measurements of the controlled phantom samples, which were fabricated with known, targeted StO2 values. This is an objective, technical ground truth rather than human expert opinion.


4. Adjudication Method for the Test Set

No adjudication method is described. As the studies compare quantitative measurements of StO2 from different devices rather than subjective interpretations or classifications that would require expert review and adjudication (e.g., in an imaging study), an adjudication method is generally not applicable or described in this context.


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, an MRMC comparative effectiveness study was not conducted. This type of study is typically performed for diagnostic imaging devices where human readers interpret medical images, and the AI's assistance to these readers is being evaluated. The Presygen™/si-1 is an oximeter measuring tissue oxygen saturation, not an imaging device requiring human interpretation in the sense that would necessitate an MRMC study. The studies focused on the device's measurement equivalence to existing oximetry devices.


6. Standalone (Algorithm Only) Performance

Yes, a standalone performance evaluation was done. The entire study focuses on the Presygen™/si-1 device itself, which includes the software's ability to calculate and display StO2 from optical inputs. The phantom study, in particular, assesses the standalone measurement accuracy and spatial uniformity of the device. The clinical studies compare its direct output to other devices, without integrating a human "reader" into the core function of StO2 measurement generation. The device is intended for use by healthcare professionals as a tool, but its fundamental output (StO2 values) is algorithm-generated.


7. Type of Ground Truth Used

  • Phantom Study: The ground truth was objective, calibrated measurements from a spectrophotometer of phantoms specifically designed to have known, targeted StO2 values.
  • Clinical Studies: The "ground truth" was effectively the measurements from the predicate device (Kent Camera) and a reference device (T.Ox™ Tissue Oximeter). The goal was to prove substantial equivalence, meaning the new device's measurements were statistically similar to those of the established devices. The Vascular Occlusion Test (VOT) provided a physiological process (ischemia and reperfusion) to evaluate dynamic changes, with the expectation that all devices would respond similarly.

8. Sample Size for the Training Set

The document does not specify any training set size or methodology. The Presygen™/si-1 is described as a non-contact oximeter based on multispectral optical imaging techniques, and the software performs calculations to derive StO2. While such a device would inherently have a model (algorithm) that was developed and potentially "trained" or optimized during its development, this 510(k) summary focuses on the validation data (performance testing and clinical studies) rather than the internal development or training of the core StO2 calculation algorithm. This is typical for a 510(k) where the focus is on post-development verification and validation for regulatory approval.


9. How the Ground Truth for the Training Set Was Established

Since no training set details are provided in the document, there is no information on how its ground truth was established.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym and name on the right. The FDA part of the logo is in blue, with the acronym "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION" next to it.

August 9, 2018

CHRISTIE MEDICAL HOLDINGS, Inc. % Diane Horwitz Regulatory Consultant Mandell Horwitz Consultants LLC 2995 Steven Martin Drive Fairfax, Virginia 20031

Re: K171245

Trade/Device Name: Presygen™/si-1 Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: MUD Dated: July 11, 2018 Received: July 12, 2018

Dear Diane Horwitz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice

(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

M & Tillelrem

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K171245

Device Name PresygenTM/si-1

Indications for Use (Describe)

Presygen™/si-1 is indicated for use in hospital facilities, along with clinical assessment of signs and symptoms, by healthcare professionals for non-invasive and non-diagnostic measure of relative temporal variation in oxygen saturation (StO2) in superficial tissue. Presygen™/si-1 displays two-dimensional color-coded images of tissue oxygenation of the scanned surface and reports tissue oxygenation measurements in selected tissue regions. Presygen™/si-1 is intended for adults only (22 and over).

Type of Use ( Select one or both, as applicable )
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the word "CHRISTIE" in large, bold, blue letters. To the right of the word is a plus sign made up of small squares in shades of green. The plus sign is slightly tilted.

510(k) SUMMARY

1. GENERAL INFORMATION

Submitter and 510(k) Owner 1.1

CHRISTIE MEDICAL HOLDINGS, INC. 3175 Lenox Park Boulevard, Suite 200 Memphis, TN 38115

1.2 Official Correspondent

Salwa Rezk, Director, Quality & Regulatory

Telephone: 519-741-3794 Fax: 901-721-0350 Email: salwa.rezk(@christiedigital.com

1.3 Date of Preparation

August 9, 2018

NAME OF THE DEVICE 2.

2.1.1 Trade/Proprietary Name, 510(k) Number

Presygen™/si-1, K171245

2.1.2 Common/Usual Name

Tissue Saturation Oximeter

2.1.3 Classification Information

Classification Name:Oximeter
Classification Regulation:21 CFR 870.2700
Class:2
Product Code:MUD
Panel:Cardiovascular

3. PREDICATE DEVICE AND REFERENCE DEVICE

The predicate device is Kent Camera (Kent Imaging, Inc., K113507). The reference device is T.OxTM (Vioptix, Inc., K042657).

DESCRIPTION OF THE DEVICE 4.

Presygen™/si-1 is a standalone, non-contact tissue oxygen saturation oximeter based on multispectral optical imaging techniques. The mode of action is to transmit lights at four (4) known

1

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the Christie logo. The word "CHRISTIE" is written in large, bold, blue letters. To the right of the word "CHRISTIE" is a plus sign made up of small green squares. The squares are arranged in a grid pattern, with the color of the squares gradually changing from light green to dark green.

wavelengths through blood in superficial tissue and measure the tissue oxygen saturation (StO2), based on the amount of diffusely reflected light.

The Presygen™/si-1 software provides user control of the device and access to stored data, through a Graphical User Interface (GUI). The software controls hardware elements of the device such as illuminators, sensors, display and Input/Output, and manages communication between these elements and between processors and memory. The software also performs calculations to derive Tissue Oxygen Saturation (StO2) from optical inputs and display this to the device user in a two-dimension (2D) spatial map.

The Presygen™/si-1 device has been designed with the additional capability of compensating for melanin levels in the skin, since absorption of light by melanin affects the measurement of the oxygenated hemoglobin and deoxygenated hemoglobin through the skin.

The device consists of the following:

  • Imaging head: Contains optics, lasers, and electronics to display the %StO2 .
  • . Base compartment: Includes the wheels, brakes mechanism, SMB (Smart Multi-Media Board), power switch, certified power supply, fans, filters and cabling
  • . Touchscreen monitor display: User interface, display of image, and patient data
  • Keyboard: Used to enter user or patient information into the device .
  • Accessory: The Presygen™/si-1 Sterilized Drape is a prepackaged single-use third party . drape that is used with Presygen™/si-1 to avoid contamination during the imaging process.

INDICATIONS FOR USE 5.

The indication for use statement for Presygen™/si-1 is shown below.

Presygen™/si-1 is indicated for use in hospital facilities, along with clinical assessment of signs and symptoms, by healthcare professionals for non-invasive and non-diagnostic measure of relative temporal variation in oxygen saturation (StO2) in superficial tissue. Presygen™/si-1 displays twodimensional color-coded images of tissue oxygenation of the scanned surface and reports tissue oxygenation measurements in selected tissue regions. Presygen™/si-1 is intended for adults only (22 and over).

INDICATIONS FOR USE COMPARED TO THE PREDICATE 6.

The indications for use statement for Presygen™/si-1 system is similar to the predicate device. The devices share the same target patient population, the same users and conditions of use. Both devices are Rx-only.

Subject DevicePredicate: K113507
PresygenTM/si-1(Christie Medical Holdings, Inc.)Kent Camera(Kent Imaging Inc.)
PresygenTM/si-1 is indicated for use in hospitalfacilities, along with clinical assessment of signsand symptoms, by healthcare professionals for non-invasive and non-diagnostic measure of relativetemporal variation in oxygen saturation (StO2) inThe Kent Camera is intended for use by healthcareprofessionals as a non-invasive tissue oxygenationmeasurement system that reports an approximatevalue of:• oxygen saturation (StO2)j,

Table 1. Indications for Use Comparison for Presygen™/si-1 Versus the Predicate

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Christie. The logo is blue and green. The word "CHRISTIE" is in blue, and there is a green plus sign to the right of the word.

Subject DevicePredicate: K113507
Presygen™/si-1Kent Camera
(Christie Medical Holdings, Inc.)(Kent Imaging Inc.)
superficial tissue. Presygen™/si-1 displays two-dimensional color-coded images of tissueoxygenation of the scanned surface and reportstissue oxygenation measurements in selected tissueregions. Presygen™/si-1 is intended for adults only(22 and over).• oxyhemoglobin level (HbO2), and• deoxyhemoglobin (Hb) levelin superficial tissue. The Kent Camera displays two-dimensional color-coded images of tissue oxygenationof the scanned surface and reports multispectral tissueoxygenation measurements for selected tissue regions.The Kent Camera is indicated for use to determineoxygenation levels in superficial tissues for patients

7. TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE

A comparison of the technological features between Presygen™/si-1 device and the predicate is shown in Table 1 below.

TechnologicalCharacteristicSubject DevicePresygen™/si-1(Christie Medical Holdings, Inc.)Predicate: K113507Kent Camera(Kent Imaging Inc.)
Measured Parameters• StO2• StO2• HbO2• Hb
Operating principleSpectral Analysis at specificwavelengths of light returned from targettissue.Same
Patient ContactSensors in imaging head do not touchpatient skin.Same
Output DisplayTwo-dimensional color-coded map(image) of relative % oxygen saturation(%StO2).Numeric data output available.Same
Light EnergyDeliveredMultiple wavelengths of Near-infraredlight (NIR)Source: Light emitting diodes (LEDs)Same
Energy deliveredWavelengths: 740, 780, 850, 945 nmVisible lasers for properly focusingdevice (wavelength 515 nm)Wavelengths: 670, 735, 890, 940 nmVisible lasers for properly focusingdevice (wavelength details notavailable)
Power sourceMains poweredRating: 100-240 Vac, 50-60Hz, 198 VAMains poweredRating: 90-240 Vac, 50-60Hz, 150Watts
Output to userColor-coded %StO2 map. User captureimages for future reference.Same

Table 2. Technology Comparison for Presygen™/si-1 Versus the Predicate

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the word "CHRISTIE" in blue, with a stylized "K" that looks like two triangles. To the right of the word is a plus sign made up of small squares in shades of green. The logo is simple and modern, with a focus on the company name and a visual element that suggests growth or addition.

TechnologicalCharacteristicSubject DevicePresygenTM/si-1(Christie Medical Holdings, Inc.)Predicate: K113507Kent Camera(Kent Imaging Inc.)
Monitor viewImmediate live view of the image duringa surgical procedure using the touch-screen monitorNo immediate view of the imageduring an operation in the touch-screen monitor
SMC (Small MotionCorrection) module inthe softwarePresygen™/si-1 system uses SMC(Small Motion Correction) to track ROI(regions of interest) using spatialregistration to compensate for up to 23 x23 mm translation (X and Y axis) forsmall motions of the imaging head orsubject (e.g., during breathing)Feature not available. Kent cameraacts as a still-image camera.
Melanin CalibrationPrior to each session, a melanincalibration is performed for plannedimaging areas in the subject tocompensate for skin melanin levels intissueNo calibration for melanin isperformed.Note that the ViOptix ODISsey TissueOximeter (K042657) is a 510(k)-cleared device with measurements thatare independent of melaninconcentration. This device isconsidered a Reference Device.
ROI Plot"Region of Interest" Plot: The user hasthe option of both saving and displayingStO2 levels from user-selected regions ofinterest over a maximum of 5 minutes inthe form of a line graphFeature not available
Presygen™/si-1Sterilized DrapeAccessoryYes. A sterile single-use drape isobtained sterilized and pre-packagedfrom a third-party vendor and is usedduring imaging.Kent Camera does not have a steriledrape

Similarities and Differences in Technology Comparison 7.1

The fundamental technology of the Presygen™/si-1 device is the same as the predicate device: The measured parameters, operating principle, patient contact, output display, materials, light energy delivered, power source and output to user are all the same. The energy delivered (NIR and Visible wavelengths) are similar.

It is acknowledged that there are also some differences: the subject device has features such as live view of images on the monitor, a calibration method to aid in the imaging of different skin tones, a correction for small motion changes of the test location. The subject device includes a sterile single-use drape accessory to cover the optical head and arm. Despite the differences between Presygen™/si-1 and the predicate, extensive bench testing, software verification and validation, usability testing and two comparative clinical performance studies were conducted under design controls to demonstrate that the Presygen™/si-1 device meets its design specifications and that the differences from the predicate do not adversely impact safety or effectiveness.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the Christie logo. The word "CHRISTIE" is written in large, bold, blue letters. To the right of the word, there is a plus sign made up of small squares in varying shades of green. The logo is simple and modern.

8. PERFORMANCE TESTING

The 510(k) submission provided performance data to establish the substantial equivalence of the Presygen™/si-1 to the predicate device. A summary of these performance tests is provided below.

Sterilization, Storage, and Packaging: Presygen™/si-1 is a non-sterile imaging device and it is used with the Presygen™/si-1 Sterilized Drape as a pre-packaged single-use accessory from a Third-Party. Verification and validation testing for usage with Presygen™/si-1 together with the Presygen™/si-1 sterile drape met all acceptance criteria. In addition, the sterile drape was used successfully in the usability studies and in the clinical study that support this application.

Presygen™/si-1 device storage has been validated successfully under the proposed storage conditions of -9° to +140° F (-23° to 60°C), and operating conditions of 60° to 86°F (16° to 30°C), which is standard for many electronic medical device products.

Transportation, handling, and packaging strength were tested according to the applicable sections of ASTM D4169-16 by a third party and met the acceptance criteria.

Software Verification and Validation Testing: Software verification and validation testing was conducted according applicable sections of IEC 62304-06 and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern. Unit- and system-level software verification and validation testing demonstrated that the system performs as intended. The software successfully passed all requirements.

Electrical and Electromagnetic Compatibility: The Presygen™/si-1 was tested to international electrical and electromagnetic safety testing in accordance with AAMI ANSI ES 60601-1-12 and IEC 60601-1-2-07., and the device met the standards.

Bench (Performance) Testing: Christie Medical Holdings, Inc. performed a series of bench tests to demonstrate that Presygen™/si-1 meets its performance specifications; the device passed all test requirements and met the acceptance criteria for performance. The Bench Performance Tests were as follows:

  • . Mechanical component tests
  • . Vibration, temperature, and humidity specific conditions tests
  • Presygen™/si-1 sterile drape usage functionality device test •
  • . Cleaning agents and cleaning process in health care setting tests
  • Electrical sensors functionality tests ●
  • . Electrical circuit boards functionality tests
  • Optical functionality tests .
  • . Phantom study

Optical Safety: The system belongs in the Low Risk Group according to IEC 62471-06, Photobiological safety of lamps and lamp systems.

Laser Safety: The system is a laser product of Class 1 according to 21 CFR $1040.10 and complies with IEC 60825-07.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the word "CHRISTIE" in large, bold, blue letters. To the right of the word is a plus sign made up of nine squares in varying shades of green. The plus sign is smaller than the word "CHRISTIE" and is positioned so that the top of the plus sign is aligned with the top of the word.

Usability Testing: Presygen™/si-1 performed as intended according to AAMI ANSI IEC 62366-1-15. Application of usability engineering to medical devices.

Phantom Testing:

Study Objectives:

An independent bench validation phantom study was conducted to evaluate measurement accuracy and spatial uniformity of StO2 values as a secondary validation for (StO2% measurement range of 50%-90%) in which 3 phantom sets were generated with varying %StO2 at high, medium and low levels (target levels of 90%, 70% and 50% StO2, respectively). These phantoms were fabricated using a mixture of human hemoglobin tuned to the target StO2 value and polystyrene microspheres, which act as optical scatterers similar to human tissue. The spectrophotometer measurements for the controlled samples, vs. the measurements of Presygen™/si-1 %StO2 were statistically evaluated to demonstrate the Presygen™/si-1 measurement accuracy in %StO2 and the spatial variations across the imaging field.

Results:

The results of this testing demonstrated a significant linear trend of %StO2 in a plot of actual vs. theoretical values using a target confidence of 95%. Phantom results indicate that the results are acceptably uniform over the device field of view.

Clinical Studies: A series of two studies were conducted comparing tissue oxygen saturation (StO2) measurements taken with Presygen™/si-1 system versus the Kent Camera [K113507] and with Presygen™/si-1 system and the T.Ox™ Tissue Oximeter [K042657]. The study objectives were:

  • To provide clinical evidence to support the claim that Presygen™/si-1 is substantially equivalent to Kent Camera, with a particular focus on subjects in Fitzpatrick Skin Type Scale I and II.
  • . To compare the measurement of StO2 between Presygen™/si-1 and T.Ox™ Tissue Oximeter in subjects across Fitzpatrick Skin Type Scales I through V since the device output is not significantly affected by melanin.

Clinical Study 1

Clinical Study 1 Objectives:

Study number 1 was a prospective, unblinded, convenience sample population volunteers from the general community (no health information collected) for 55 subjects (199 data points) from ages 22 years and older, male and female. Fifty-two subjects (10 male and 42 female) were included in data analysis and three subjects were excluded due to incomplete data associated with device issues. To evaluate the effect of patient skin type, subjects were stratified across five (5) Fitzpatrick Skin Type Scales. Subjects had static images taken of target areas [on the thenar eminence and ventral forearm] with all three devices.

Clinical Study 1 Results:

A Vascular Occlusion Test (VOT) was conducted with all devices to evaluate performance during ischemia and reperfusion over a clinically meaningful dynamic range of StO2. The %StO2 results measurement range was 40% to 90%.

Presvgen™/si-1 compared to the Kent camera in Fitzpatrick Scale I with >95% of the observations within the Limits of Agreement on the Bland-Altman with repeated measures and Deming Regression with bootstrap resampling results showing a slope with 95% confidence limits

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the Christie logo. The word "CHRISTIE" is written in blue, block letters. To the right of the word is a plus sign made up of small green squares. The logo is simple and modern.

including 1. The intercept for the Deming regression equation did not contain 0. however, demonstrating a small positive bias in the measurement of StO2. The assessment of precision found the coefficient of variation was within 2 standard deviations of the mean for >95% of all observations in the measurement of StO2 in Fitzpatrick I and II. Presygen™/si-1, Kent Camera, and T.Ox™ all demonstrate similar reactions to ischemia and reperfusion in the measurement of StO2.

Clinical Study 1 VOT Results:

VOT data for Presygen™/si-1. Kent Camera, and T.Ox™ Tissue Oximeter all demonstrate similar reactions to ischemia and reperfusion in the measurement of %StO2. The VOT graphs for all three devices show similar patterns of baseline, response to ischemia and reperfusion. Data from the vascular occlusion test shows %StO2 stable at baseline, decreasing through ischemia and rebounding during reperfusion for both Presygen™/si-1 and Kent Camera, and between Presygen™/si-1 and T.Ox™ Tissue Oximeter.

Clinical Study 1 VOT Re-analysis:

Reanalysis of Clinical Study 1 VOT data was conducted to evaluate the measured temporal StO2 change between Presygen™/si-1 and the reference device (T.Ox™). The measured temporal StO2 change (ΔStO2) was calculated as the measured StO2 value subtracted by the mean StO2 value of the entire time series for a specific subject. The Bland-Altman plot showed most of the data is centered close to a Mean Bias that is close to 0; 91% of the observations are within the Limits of Agreement. It should be noted that the paired VOT measurements were collected on two different arms and at separate time intervals which generates an additional source of variation. Bland-Altman analysis showed a mean difference of 0.110 %StO2 with Limits of Agreement (-9.052, 9.273). The analysis of the correlation between the AStO2 by Deming regression shows equivalence to T.Ox™ with intercepts that are close to 0 and slopes that are close to 1. Deming Regression Analysis showed an intercept of 0.00 %StO2 with a 95% CI (-0.282, 0.278) and a slope of 0.773 with a 95% CI (0.708, 0.845). The histogram plot demonstrates that the differences in the △StO2 are distributed evenly around 0.

Throughout the study there was no SAE (Serious Adverse Events) or AE (Adverse Events) observed.

Clinical Study 2

Clinical Study 2 Objectives:

Study number 2 was also a prospective, unblinded, convenience sample population who selfreported as being healthy for 20 subjects (6 male, 14 female) from ages 22 years and older, male and female in a laboratory setting. The aim of this study was to provide data to compare StQz values in the left and right forearm using Presvgen™/si-1 (subject device). Kent Camera (predicate device; Fitzpatrick Scales I and II) and T.Ox™ (reference device; Fitzpatrick Skin Type Scales I through V). Test conditions included static measurements and VOT.

Primary effectiveness included evaluation with the predicate device (Kent Camera) and reference device (T.OxTM Tissue Oximeter) using Bland-Altman with repeated measures and Deming Regression with bootstrap resampling. Secondary effectiveness included evaluation across all Fitzpatrick Scales with all three devices (Presygen™/si-1, Kent Camera and T.Ox™ Tissue Oximeter). Presygen™/si-1 and Kent Camera results from this study should be interpreted with caution due to the small sample size and the influence of unusual values from the Kent camera with several observations.

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the word "CHRISTIE" in large, bold, blue letters. To the right of the word is a plus sign made up of small squares in shades of green. The logo appears to be for a company or organization named Christie.

Clinical Study 2 Results:

Results of this study were as follows: StO2 values with Presygen™/si-1 and T.OxTM Tissue Oximeter shows equivalence across all Fitzpatrick Scales including within the higher melanin scales of II-V with all of the observations within the Limits of Agreement on the Bland-Altman plot (Table 1). Results of Presygen™/si-1 versus Kent Camera were limited to Fitzpatrick I and II. The %StO2 results measurement range was 67% to 90%.

Clinical Study 2 VOT Results:

For the VOT study, Presygen™/si-1, Kent Camera and T.Ox™ Tissue Oximeter all demonstrate similar fluctuations in response to ischemia and reperfusion in the measurement of %StO2. Data from the vascular occlusion test shows %StO2 stable at baseline, decreasing through ischemia and rebounding during reperfusion for both Presygen™/si-1 and Kent Camera, and between Presygen™/si-1 and T.Ox™ Tissue Oximeter.

Clinical Study 2 VOT Re-analysis:

Reanalysis of Clinical Study 2 VOT data was conducted as described above for Clinical Study 1 VOT Re-analysis. The Bland-Altman plot shows most of the data is centered close to a Mean Bias that is close to 0; 90% of the observations are within the Limits of Agreement. It should be noted that the paired VOT measurements were collected on two different arms and at separate time intervals which generates an additional source of variation. Bland-Altman analysis showed a mean difference of 0.016 %StO2 with Limits of Agreement (-8.563, 8.596). The analysis of the correlation between the AStO2 by Deming regression shows equivalence with intercepts that are close to 0 and slopes that are close to 1. Deming Regression Analysis showed an intercept of 0.00 %StO2 with a 95% CI (-0.423, 0.401) and a slope of 0.815 with a 95% CI (0.716, 0.937). The histogram plot demonstrates that the differences in the AStO2 are distributed evenly around 0.

Throughout the study there was no SAE or AE observed.

In conclusion, the weight of the evidence from both clinical study 1 and study 2 shows that Presygen™/si-1 and Kent Camera have consistent directional temporal %StO2 changes, particularly during the VOT. Presygen™/si-1 compared to the Kent camera in Fitzpatrick Scale I with >95% of the observations within the Limits of Agreement on the Bland-Altman plot and the Deming regression results showing a slope with 95% confidence limits including 1 (Study 1). The intercept for the Deming regression equation did not contain 0 however, demonstrating a small positive bias in the measurement of %StO2. Presygen™/si-1 showed equivalence to Kent Camera in subjects with minimal tissue melanin. These studies shows equivalence between Presygen™/si-1 and T.Ox™ Tissue Oximeter %StO2 measurements overall, and within each Fitzpatrick Scale, demonstrating that Presygen™/si-1 shows equivalence with the reference device measurement of %StO2 across Fitzpatrick Scales I through V and is not significantly affected by tissue melanin.

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the Christie logo. The word "CHRISTIE" is written in large, bold, blue letters. To the right of the word is a green plus sign made up of smaller squares. The logo is simple and modern.

Clinical Study 1Clinical Study 2
Y-Intercept(95% agreement)119.803(98.08, 151.617)-0.590(-21.244, 16.253)
Slope(95% agreement)-0.734(-1.204, -0.399)1.007(0.788, 1.278)

Table 3. Y-Intercept and Slope from Deming Regression for Presygen vs. Reference Device in Clinical Study 1 vs. Study 2

Table 4. Summary Table of Bland-Altman Data for Presygen vs. Reference Device in Clinical Study 1 vs. Study 2

Clinical Study 1Clinical Study 2
Bias (%StO2),(95% agreement)6.4595(4.25, 8.67)0.1(-1.64, 1.84)
LOA (%StO2)-31.45, 44.37-10.65, 10.85
RMSE (%StO2)20.355.38

CONCLUSIONS 9.

The information presented in this 510(k) submission demonstrates that the Presygen™/si-1 is substantially equivalent to the predicate device.

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).