K Number
K171229

Validate with FDA (Live)

Date Cleared
2017-05-26

(29 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DELL UP3017 with QUBYX PerfectLum is intended to be used for displaying and viewing medical images, for review and analysis by trained medical practitioners.

The DELL UP3017 can be used only in conjunction with QUBYX PerfectLum.

The device can not be used in primary image diagnosis in mammography.

The device can not be used for a life-support system.

The device does not contact with the patient.

The device is intended for prescription use.

Device Description

The DELL UP3017 with QUBYX PerfectLum is a 30" color display for medical viewing. It provides 2560 x 1600 resolution with an adjustable Look Up Table and a 10 bit panel. It is combined with QUBYX PerfectLum and PerfectLum remote management, a userfriendly DICOM calibration and AAPM TG18 verification software suite. The software allows setting the display function to DICOM, displaying test pattern and performing acceptance and constancy tests.

AI/ML Overview

Here's the breakdown of the acceptance criteria and study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriterionReported Device Performance
DICOM Part 14 GSDF CompliancePassed DICOM conformance test
AAPM TG18 Standard CompliancePassed AAPM TG18 acceptance test

2. Sample Sized Used for the Test Set and Data Provenance

  • Sample Size (Test Set): Not explicitly stated. The text mentions "the display device" (singular) having passed the tests, implying testing was performed on one or more units of the DELL UP3017. However, no specific number is provided for how many individual displays were tested.
  • Data Provenance: The tests were performed by QUBYX Software Technologies Inc. The University of Arizona also verified compliance for both current and predicate devices, suggesting some level of independent verification or historical data. The data is prospective, generated specifically for this submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

  • Number of Experts: Not explicitly stated. For the visual components of the AAPM TG18 test, the text mentions "the user analyzed test patterns." This "user" could be a single technician or multiple, but details on their qualifications are not provided.
  • Qualifications of Experts: Not explicitly stated.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. The tests are objective measurements and visual assessments against a standard, not a consensus-based adjudication of subjective findings from multiple readers. The software analyzes results against defined target values.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

  • MRMC Study: No, an MRMC comparative effectiveness study was not done. This submission focuses on the display device's technical compliance with standards, not its impact on human reader performance with or without AI assistance.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study was Done

  • Standalone Study: Yes, in effect. The DICOM and AAPM compliance tests are standalone evaluations of the display's technical performance and adherence to established standards, without a human interpretation loop as the primary measure of effectiveness. The "algorithm" here refers to the display and the PerfectLum software's ability to render images and perform calibrations/verifications according to specifications.

7. Type of Ground Truth Used

  • Ground Truth: Industry standards and specifications:
    • DICOM Part 14 GSDF (Grayscale Standard Display Function) for luminance response and image fidelity.
    • AAPM TG18 (American Association of Physicists in Medicine Task Group 18) for display quality control and acceptance testing, which includes measurement and visual criteria.

8. Sample Size for the Training Set

  • Sample Size (Training Set): Not applicable. This device is a medical display and calibration software, not an AI algorithm that undergoes a "training" phase with a dataset in the typical machine learning sense. The software's functionality is based on programmed algorithms and calibration profiles, not learned from a training set.

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth for Training Set: Not applicable, as there is no "training set" in the context of an AI algorithm. The display and software are designed and verified against established industry standards (DICOM, AAPM) which serve as the reference for correct performance.

{0}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing to the right. The profiles are arranged in a way that they appear to be connected. The logo also includes the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 26, 2017

Qubyx Software Technologies Inc. Marc Leppla Director, CTO 501 Silverside Road. Suite 105 Wilmington, Delaware 19809

Re: K171229

Trade/Device Name: DELL UP3017 with OUBYX PerfectLum bundle Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: PGY Dated: April 20, 2017 Received: April 27, 2017

Dear Marc Leppla:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Robert Ochs

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K171229

Device Name Dell UP3017 with QUBYX PerfectLum bundle

Indications for Use (Describe)

The DELL UP3017 with QUBYX PerfectLum is intended to be used for displaying and viewing medical images, for review and analysis by trained medical practitioners.

The DELL UP3017 can be used only in conjunction with QUBYX PerfectLum.

The device can not be used in primary image diagnosis in mammography.

The device can not be used for a life-support system.

The device does not contact with the patient.

The device is intended for prescription use.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary

as required by 807.92

  1. Company Identification Qubyx Software Technologies Inc. 501 Silverside Road, Suite 105 Wilmington, DE 19809 USA Tel: +1 518 618 1012

  2. Official Correspondent Mr. Marc Leppla President and CTO (Chief Technical Officer) leppla@qubyx.com

  3. Date of Submission Feb 20, 2017

  4. Device Trade name DELL UP3017 with QUBYX PerfectLum bundle

  5. Common/Usual Name Image display system, Color LCD Monitor, image monitor/display

  6. Classification Number Medical displays classified in Class II per 21 CFR 892.2050

  7. Predicate device 1 Name: DELL U3014 Manufacturer: Qubyx Software Technologies Inc. 510(k) number: K131601

8. Device description

The DELL UP3017 with QUBYX PerfectLum is a 30" color display for medical viewing. It provides 2560 x 1600 resolution with an adjustable Look Up Table and a 10 bit panel. It is combined with QUBYX PerfectLum and PerfectLum remote management, a userfriendly DICOM calibration and AAPM TG18 verification software suite. The software allows setting the display function to DICOM, displaying test pattern and performing acceptance and constancy tests.

{4}------------------------------------------------

9. Intended use

The DELL UP3017 with QUBYX PerfectLum is intended to be used for displaying and viewing of digital images, for review and analysis by trained medical practitioners.

The DELL UP3017 must only be used in conjunction with QUBYX PerfectLum.

The device must not be used in primary image diagnosis in mammography. The device can not be used for a life-support system and does not contact with the patient.

DELL UP3017with PerfectLumPredicate device DELL U3014510(k) number: K131601
Panel TypeIPSIPS
Panel size30" viewable30" viewable
Native Resolution2560 x 16002560 x 1600
Aspect Ratio16:1016:10
Pixel Pitch0.25 mm0.25 mm
Brightness (typical)350 cd/m2350 cd/m2
Contrast Ratio (typical)1:10001:1000
Viewing Angle (typical)178° Vert., 176° Hor.178° Vert., 178° Hor.
Displayable Colors1.07 billion colors1.07 billion colors
BacklightWhite LEDRGB LED
DICOM calibration andAAPM verification softwarebundledbundled
Indications for UseThe DELL UP3017 withQUBYX PerfectLum isintended to be used fordisplaying and viewingmedical images, for reviewand analysis by trainedmedical practitioners.The DELL UP3017 can beused only in conjunction withQUBYX PerfectLum. Thedevice can not be used inprimary image diagnosis inmammography.The device can not be usedfor a life-support system.The device does not contactwith the patient.The DELL U3014 withQUBYX PerfectLum isintended to be used fordisplaying and viewingmedical images, for reviewand analysis by trainedmedical practitioners. TheDELL U3014 can be usedonly in conjunction withQUBYX PerfectLum.The device can not be usedin primary image diagnosis inmammography.The device can not beused for a life-supportsystem. The device doesnot contact with the patient.

10. Comparison table - predicate device 1

{5}------------------------------------------------

Conclusion

DELL UP3017 is the direct successor of the DELL U3014 model in the product line of UltraSharp displays, so the DELL UP3017 with PerfectLum has exactly the same technical characteristics that predicate device DELL U3014 with PerfectLum, except backlight type. Backlight type of the U3014 model is RGB LED, while backlight type of the UP3017 model is White LED. The next generation used a white light LED to save on power. The comparison table shows that the subject device (DELL UP3017 with PerfectLum) has the same intended use as the predicate device.

Although the devices have different backlight types, these differences do not make the subject device less safe and reliable, so the subject device fits for displaying and viewing of digital images for diagnosis by trained physicians as the predicate device does.

Both devices are compliant with DICOM Part 14 GSDF and AAPM TG18 standards, which is tested and verified by University of Arizona. To verify DICOM and AAPM compliance for the subject device, AAPM acceptance test and DICOM conformance test were also performed by QUBYX.

Details of testing:

To verify DICOM conformance, a DICOM conformance test was performed, using QUBYX PerfectLum software and an X-Rite i1 Display Pro measurement device.

The test procedure was generated by the software in accordance with the requirements of the DICOM standard. It consisted of measurement steps, where the meter measured display's characteristics and the software recorded them. Then the software analyzed the results in comparison with target values, defined by DICOM standard, and generated the report, stating that the display is DICOM-conformant.

The display device has successfully passed DICOM conformance test, so it is compliant with DICOM Part 14 GSDF standard. So is the predicate device, so the two devices are substantially equivalent in this regard.

To verify AAPM TG18 conformance, an acceptance test was performed, using QUBYX PerfectLum software and an X-Rite i1 Display Pro measurement device.

The test procedure was generated by the software in accordance with the requirements of the AAPM TG18 standard and consisted of measurement and visual parts.

During the measurement steps, the meter measured display's characteristics and the software recorded them. During the visual steps, the user analyzed test patterns, generated by the software in accordance with AAPM standard. The software recorded the user's answers. Then the software analyzed the results in comparison with target values, defined by AAPM standard. and generated the report, stating that the display passes AAPM TG18 acceptance test. The display device has successfully passed AAPM TG18 acceptance test, so it is compliant with AAPM TG18 standard and can be used as a primary category display for interpretation of medical images. The same is true for the predicate device, so the two devices are substantially equivalent in this regard.

Both devices have the same indications for use.

The new and predicate devices are substantially equivalent in the areas of technical characteristics, general function, application and indented use.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).