K Number
K171225

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-08-21

(117 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Sorbact® Ribbon Gauze is intended for use in the management of exuding partial to full thickness wounds (including clean, colonized, contaminated or infected wounds). Sorbact® Ribbon Gauze is indicated for shallow cavity wounds and fistulas.

Device Description

Sorbact® Ribbon Gauze is a sterile (gamma irradiation), single use only, hydrophobic microbe binding wound dressing. It consists of a Sorbact® wound contact layer, which allows passage of wound exudate into a secondary dressing.

AI/ML Overview

Based on the provided text, the device in question is the "Sorbact® Wound Dressing – Ribbon Gauze". However, this document is a 510(k) premarket notification for a modification to an existing device, specifically a change to the color additive. The provided information does not describe a study that uses a test set to prove the device meets performance criteria related to its clinical efficacy or diagnostic performance as an AI/ML device.

The performance data section explicitly states: "The modification addressed by this 510(k) is a change to the color additive in the Sorbact® Ribbon Gauze. As the intended use, device description, wound contact material, instructions for use, mechanism of action, storage conditions, and shelf life of the modified device are the same as that of the predicate device, both devices have the same fundamental scientific technology."

Therefore, the "performance data" presented is entirely focused on non-clinical testing related to the safety and material properties of the new color additive, rather than a clinical performance study involving a test set, ground truth, or human readers, as would be expected for an AI/ML device.

Given this, I cannot fill in the requested table and answer the specific questions (2-9) because the information is not present in the provided document. The document describes a traditional medical device (wound dressing) and a 510(k) submission for a minor change (color additive), not an AI/ML device or its performance evaluation in the context you've outlined.

Here's what I can extract regarding acceptance criteria and performance, as related to the specific modification:

1. Table of Acceptance Criteria and Reported Device Performance (for the color additive change):

Acceptance Criteria CategoryDescription of Criteria (Implied)Reported Device Performance
Modifications/EquivalencyIntended use, device description, wound contact material, instructions for use, mechanism of action, storage conditions, and shelf life must remain the same as the predicate device despite the color additive change.Stated that all these characteristics are "the same as that of the predicate device."
Functional PerformanceDevice must continue to meet functional performance requirements after the modification."The results of nonclinical testing demonstrate that the device met all performance requirements."
Safety (Biocompatibility)Device must be biocompatible with the new color additive."Testing was performed on representative samples of the devices and included the following tests: Cytotoxicity, Intracutaneous reactivity, Sensitization, Systemic toxicity (acute)." Results are stated to have met requirements.
Safety (Extractables)Evaluation of extractable colorants to ensure no harmful substances are released."Evaluation of extractable colorants" was performed. Results are implied to be acceptable based on overall conclusion.

Missing Information (as per your prompt's format, because it's not an AI/ML device):

  1. Sample sizes used for the test set and data provenance: Not applicable. This was non-clinical material and biocompatibility testing, not a clinical test set.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No clinical ground truth established.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. Safety and material testing, not clinical diagnosis.
  7. The sample size for the training set: Not applicable. No AI/ML training set.
  8. How the ground truth for the training set was established: Not applicable.

In summary, the provided document pertains to a regulatory submission for a minor modification to a non-AI/ML medical device. It does not contain the information required to address your specific questions about AI/ML device performance studies.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is written in a simple, sans-serif font.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 21, 2017

ABIGO Medical AB % Deborah Graveski Sr. Project Manager M Squared Associates. Inc. 575 8th Avenue, Suite 1212 New York, New York 10018

Re: K171225

Trade/Device Name: Sorbact Wound Dressing-Ribbon Gauze Regulatory Class: Unclassified Product Code: FRO Dated: May 23, 2017 Received: May 24, 2017

Dear Deborah Grayeski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

David Krause -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171225

Device Name Sorbact Ribbon Gauze

Indications for Use (Describe)

Sorbact Ribbon Gauze is intended for use in the management of exuding partial to full thickness wounds (including clean, colonized, contaminated or infected wounds). Sorbact Ribbon Gauze is indicated for shallow cavity wounds and fistulas.

Type of Use (Select one or both, as applicable)

X | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 6: 510(k) Summary

Submitter:ABIGO Medical ABEkonomivägen 5SE-436 33 AskimSwedenPhone: +46 31 748 49 50Fax: +46 31 68 39 51Registration Number: 8043991
Official Correspondent:Deborah Lavoie GrayeskiM Squared Associates, Inc.575 8th Avenue, Suite 1212New York, NY 10018 USAOffice: 703-562-9800x250Mobile: 202-550-7395Fax: 703-562-9797Email: dgrayeski@msquaredassociates.com
Date Prepared:April 25, 2017
Trade Name:Sorbact® Wound Dressing – Ribbon Gauze
Common Name:Wound Dressing
Regulatory Class:Unclassified (Pre-amendment)
Review Panel:General & Plastic Surgery
Product Code:FRO
Predicate Device:K063059, Sorbact® Wound Dressings – Ribbon GauzeABIGO Medical AB

Device Description:

Sorbact® Ribbon Gauze is a sterile (gamma irradiation), single use only, hydrophobic microbe binding wound dressing. It consists of a Sorbact® wound contact layer, which allows passage of wound exudate into a secondary dressing.

Indication for Use:

Rx: Sorbact® Ribbon Gauze is intended for use in the management of exuding partial to full thickness wounds (including clean, colonized, contaminated or infected wounds). Sorbact® Ribbon Gauze is indicated for shallow cavity wounds and fistulas.

{4}------------------------------------------------

Comparison of Technological Characteristics with the Predicate Device:

The modification addressed by this 510(k) is a change to the color additive in the Sorbact® Ribbon Gauze. As the intended use, device description, wound contact material, instructions for use, mechanism of action, storage conditions, and shelf life of the modified device are the same as that of the predicate device, both devices have the same fundamental scientific technology.

Performance Data:

To verify that the device design met its functional performance and safety requirements, representative samples of the device underwent the following testing. The results of nonclinical testing demonstrate that the device met all performance requirements and that the subject device is substantially equivalent to the predicate device.

● Extraction testing

  • o Evaluation of extractable colorants

● Biocompatibility testing

The biocompatibility evaluation of the devices was conducted in accordance with the FDA Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing". Testing was performed on representative samples of the devices and included the following tests:

  • o Cytotoxicity
  • Intracutaneous reactivity O
  • Sensitization O
  • Systemic toxicity (acute) O

Conclusion:

ABIGO Medical AB considers the Sorbact® Ribbon Gauze to be substantially equivalent to the predicate device.

N/A