K Number
K170634

Validate with FDA (Live)

Date Cleared
2017-11-29

(272 days)

Product Code
Regulation Number
890.3800
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The devices provide mobility to adult persons limited to a sitting position and capable to operate a few simple controls.

Device Description

There are three models in the Quingo range for this subject 510(k): Quingo Flyte Power Scooter, S551 (transportable model), Quingo Classic Power Scooter, S1301 and Quingo Toura Power Scooter, S350 (pavement and higher speed models). They all have very similar characteristics which are detailed below, with the main differences being: length, widths, motor size, controller size, and wheel size. All the Quingo models scooter frames are a welded steel construction and they all include 2 rear wheels connected by a motor/ transaxle with an electromagnetic brake, batteries, and they all have 3 front wheels and a tiller to control its turning direction. When the user activates the throttle lever, the controller receives a signal to release the brakes. With the brakes released, the scooters are allowed to move forwards or backward. When the user releases the throttle lever, the scooter slows to a stop and the brakes are automatically re-engaged. The electromechanical brakes allow the user to stop by letting to of the throttle lever. All the Quingo Models have an emergency parking brake, the lever for this brake is on the tiller and the brake itself is a carbon disc brake on to the front middle wheel.

AI/ML Overview

The provided document is a 510(k) summary for a medical device, specifically power scooters. It does not contain information about an AI/ML-driven medical device, nor does it detail acceptance criteria and a study proving an AI/ML device meets those criteria. Instead, it focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance standards and comparisons of specifications.

Therefore, I cannot fulfill the request to describe acceptance criteria and a study that proves an AI/ML device meets acceptance criteria based on the provided text. The document discusses:

  • Device Description: Power scooters (Quingo Classic, Flyte, Toura).
  • Predicate Device: Merits Transformer, Model S941, K142162.
  • Comparison Tables: Detailed comparisons of physical, electrical, and performance characteristics between the subject devices and the predicate device.
  • Non-Clinical Tests: Adherence to various ISO, ANSI, BS EN, and IEC standards for wheelchairs and motorized three-wheeled vehicles (e.g., static stability, dynamic stability, brakes, energy consumption, dimensions, speed, strength, climatic tests, electromagnetic compatibility, ignition resistance).
  • Clinical Tests: Stated as "Not applicable as there are no new indications for use which must be supported by a clinical data."
  • Software Information: Mentions "Software validation was conducted in accordance with a major level of concern designation" but provides no details on the validation study itself.

None of this information pertains to the rigorous evaluation of an AI/ML device's performance against defined acceptance criteria, including aspects like sample sizes for test/training sets, expert ground truth establishment, adjudication methods, or MRMC studies for human-AI interaction.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 29, 2017

Advanced Vehicle Concepts Ltd. % Vaibhav A. Rajal Official Correspondent for Advanced Vehicle Concepts Ltd. mdi Consultants. Inc. 55 Northern Blvd. Suite 200 Great Neck, New York 11021

Re: K170634

Trade/Device Name: Quingo Classic Power Scooter S1301, Quingo Flyte Power Scooter S551, Quingo Toura Power Scooter S350 Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized Three-Wheeled Vehicle Regulatory Class: Class II Product Code: INI Dated: October 24, 2017 Received: October 25, 2017

Dear Mr. Rajal:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Vivek J. Pinto -S

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170634

Device Name

Advanced Vehicle Concepts Ltd. Quingo Classic Power Scooter, S1301, Quingo Flyte Power Scooter, S551, Quingo Toura Power Scooter, S350.

Indications for Use (Describe)

The devices provide mobility to adult persons limited to a sitting position and capable to operate a few simple controls.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)X

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

The assigned 510(k) number is: K170634

    1. Date Prepared: November 28, 2017
      Submitter's Identification:

Contact Person: Mr. Mark Nicholls Advanced Vehicle Concepts Ltd. 7 Stadium Way Cradock Road Luton Bedfordshire LU4 OJF United Kingdom Tel: 44 1582 430889 Fax: 44 1582 430898 mark@quingoscooters.com

    1. Device Trade/Proprietary Name:
      Advanced Vehicle Concepts Ltd. Quingo Classic Power Scooter, S1301, Quingo Flyte Power Scooter, S551, Quingo Toura Power Scooter, S350.

Common Name: Motorized 3 Wheeled Vehicles (5 Wheeled Scooters)

Classification Name:

Regulation Number: 21 CFR Part 890.3800 Regulation Name: Motorized 3 Wheeled Vehicles (5 Wheeled Scooters) Regulatory Class: II Product Code: INI Classification Panel: Physical Medicine

    1. Predicate Device:
      Merits Health Products Co., Ltd., Merits Transformer, Model S941, K142162

{4}------------------------------------------------

4. Device Description

There are three models in the Quingo range for this subject 510(k)

Quingo Flyte Power Scooter, S551 (transportable model), Quingo Classic Power Scooter, S1301 and Quingo Toura Power Scooter, S350 (pavement and higher speed models)

They all have very similar characteristics which are detailed below, with the main differences being: length, widths, motor size, controller size, and wheel size.

All the Quingo models scooter frames are a welded steel construction and they all include 2 rear wheels connected by a motor/ transaxle with an electromagnetic brake, batteries, and they all have 3 front wheels and a tiller to control its turning direction.

When the user activates the throttle lever, the controller receives a signal to release the brakes. With the brakes released, the scooters are allowed to move forwards or backward. When the user releases the throttle lever, the scooter slows to a stop and the brakes are automatically re-engaged. The electromechanical brakes allow the user to stop by letting to of the throttle lever. All the Quingo Models have an emergency parking brake, the lever for this brake is on the tiller and the brake itself is a carbon disc brake on to the front middle wheel.

{5}------------------------------------------------

This is best illustrated by the picture below:

Image /page/5/Picture/1 description: The image shows a close-up of a five-wheel mobility scooter. The text above the scooter reads, "Agility: 5 Wheel Stabilizing System. Extra tight and dynamically safe turning." There is a green check mark with the word "YES!" next to it. There are also green arrows pointing in opposite directions below the scooter.

The Quingo Classic Power Scooter, S1301 is battery powered, rear transaxle driven and is controlled by the Dynamics Rhino 2 DS70C amp controller. The user interface is a wig-wag control lever and a speed knob on the control panel. The S1301 is powered by two 12 VDC 35ah batteries. The batteries are charged by a 5A off-board charger connect with 3-pin microphone connector to charging socket on the tiller. The approximate driving range on fully charged batteries is up to 37 kilometers (23 miles). The scooter frame is a welded steel construction and includes 2 rear wheels connected by a motor/transaxle with an electromagnetic brake, batteries, 3 front wheels and a tiller to control its turning direction.

When the user activates the throttle lever, the controller receives a signal to release the brakes. With the brakes released, the scooter is allowed to move forward or backwards. When the user releases the throttle lever, the scooter slows to a stop and the brakes are automatically re-engaged. The electromechanical brakes allow the user to stop by letting go of the throttle lever.

The S1301 also has fitted to the tiller, an emergency/parking brake which works via a disc brake on to one of the front wheels.

The upholstery of the device complies with ISO 7176-16:2012, resistance to ignition of postural support devices.

{6}------------------------------------------------

The device can be operated on dry, level surfaces composed of concrete, blacktop, or asphalt under normal driving conditions.

The Quingo Toura Power Scooter, S350 is battery powered, rear transaxle driven and is controlled by the Dynamics Rhino 2 DS160 amp controller. The user interface is a wig-wag control lever and a speed knob on the control panel. The S350 is powered by two 12 VDC 100ah batteries. The batteries are charged by a 8A off-board charger connect with 3-pin microphone connector to charging socket on the tiller. The approximate driving range on fully charged batteries is up to 88 kilometers (55 miles). The scooter frame is a welded steel construction and includes 2 rear wheels connected by a motor/transaxle with an electromagnetic brake, batteries, 3 front wheels and a tiller to control its turning direction.

When the user activates the throttle lever, the controller receives a signal to release the brakes. With the brakes released, the scooter is allowed to move forward or backwards. When the user releases the throttle lever, the scooter slows to a stop and the brakes are automatically re-engaged. The electromechanical brakes allow the user to stop by letting go of the throttle lever.

The S350 also has fitted to the tiller, an emergency/parking brake which works via a disc brake on to one of the front wheels.

The upholstery of the device complies with ISO 7176-16:2012, resistance to iqnition of postural support devices.

The device can be operated on dry, level surfaces composed of concrete, blacktop, or asphalt under normal driving conditions.

The Quinqo Flyte Power Scooter, S551 is battery powered, rear transaxle driven and is controlled by the Dynamics Rhino 2 DS90C amp controller. The user interface is a wig-wag control lever and a speed knob on the control panel. The S551 is powered by four 12 VDC 22ah batteries. The batteries are charged by a 5A off-board charger connect with 3-pin microphone connector to charging socket on the tiller. The approximate driving range on fully charged batteries is up to 37 kilometers (23 miles). The scooter frame is a welded steel construction and includes 2 rear wheels connected by a motor/transaxle with an electromagnetic brake, batteries, 3 front wheels and a tiller to control its turning direction.

When the user activates the throttle lever, the controller receives a signal to release the brakes. With the brakes released, the scooter is allowed to move forward or backwards. When the user releases the throttle lever, the scooter slows to a stop and the brakes are automatically re-engaged. The electromechanical brakes allow the user to stop by letting go of the throttle lever.

The S551 also has fitted to the tiller, an emergency/parking brake which works via a disc brake on to one of the front wheels.

{7}------------------------------------------------

The upholstery of the device complies with ISO 7176-16:2012, resistance to ignition of postural support devices.

The device can be operated on dry, level surfaces composed of concrete, blacktop, or asphalt under normal driving conditions.

5. Indications for Use:

The devices provide mobility to adult persons limited to a sitting position and capable to operate a few simple controls.

6. Comparison to the Predicate Device:

The Quingo Classic Power Scooter, S1301 is substantially equivalent to the Merits Transformer, Model S941, (K142162).

The Quingo Toura Power Scooter, S350 is substantially equivalent to the Merits Transformer, Model S941 (K142162).

The Quingo Flyte Power Scooter, S551 is substantially equivalent to the Merits Transformer, Model S941 (K142162).

Below are our "Comparison Tables" comparing the five subject device models to the predicate device:

COMPARISON OF SUBJECT DEVICE (QUINGO CLASSIC) AND PREDICATE DEVICE (Merits Transformer K142162)

ITEMSSUBJECTDEVICEPREDICATEDEVICESafety &effectiveness ofsubject devicecompared to thepredicate device.
Brand NameClassicMerits TransformerDifferent Make
ManufacturerAdvanced VehicleConcepts LtdMerits HealthProducts Co., Ltd.SameManufacturer
ModelNumberS1301S941Different Model
510K NoK170634K142162DifferentSubmission
SeriesPower MobilityPower MobilitySame

{8}------------------------------------------------

Scooter SeriesScooter Series
Indications for UseThe devices provide mobility to adult persons limited to a sitting position and capable to operate a few simple controls.The devices provide mobility to adult persons limited to a sitting position and capable to operate a few simple controls.Same Intended use and target population
Warranty1 Year1 YearSame
Product CodeINIINISame
Weight Limit350 lbs. / 159 Kg350 lbs. / 159 KgSame Carrying Capacity
Electronics ControllerDynamics Rhino 2 DS70CDynamics Rhino 2 DS120Same type with different volume controllers
MotorDC 24V, 250 WDC 24V, 400 WSmaller Motor (top speed is less on subject device)
FootplatesAluminum Adjustable FootplatesABSIndividual Adjustable footplates ideal to stretch out legs
SuspensionFront Lever Arm Coil SpringIntegral Lever Arm Coil SpringsSame
Rear Independent Coil SpringIndependent Coil SpringsSame
Back UpholsteryFabricFabricSame Material
BatteriesQuantity 22Same Quantity
Type 12V, 35Ah12V, 50AhSmaller Capacity
Patient Contacting MaterialsSeat Vinyl FabricVinyl FabricSame
Hand Grip RubberRubberSame
Seat Belt Steel/ ABS/ Polyester FabricN/A (No Seat Belt)Additional Safety Feature
Overall DimensionsOverall Length 1220mm / 48"1300mm / 51.2"Smaller Length
Overall Width 640mm / 25"686mm / 27.2"Smaller width
Overall Height 900mm / 36"1245mm / 49"Lower in overall height
Armrest TypesRemovable & Flip-backwardRemovable & Flip-backwardSame
FrameSteel FrameSteel FrameSame
Seat Dimensions
Seat Width457mm / 18"470mm / 18"Same
Seat Height (From Ground)710mm / 28"635-685mm / 25-27"Higher Seating position
Seat Depth (Actual)400mm / 15.7"400mm / 16"Marginally Smaller Depth
Seat Height (From Deck)450-550m / 18-22"430-480mm / 17-19"Higher Seating position
Wheels
Front10" Pneumatic Tires12" Pneumatic TiresSmaller size
Front (2 x Outer Wheels)7" Pu TiresN/AExtra (5 Wheel design)
Rear (Drive Wheels)10" Pneumatic Tires12" Pneumatic TiresSmaller size
Turning Radius1070mm / 42"1670mm / 65.7"Smaller Radius making it easier to maneuver
Ground Clearance127mm / 5"90mm / 3.5"Larger clearance
Kerb Climbing Ability76mm / 3"100mm / 4"Slightly less kerb climbing ability
Scooter WeightTotal Unlaiden220 lbs. / 100 Kg118kg / 260 lbs.Lower Weight
Without Batteries170.2 lbs. / 77.2 Kg84kg / 185 lbs.Lower Weight
Maximum Speed4 Mph / 6.4 Kph7.5 mph / 12 KphLess Speed
Incline / Gradient12 Degrees23 Miles / 37 Kilometers10 Degrees25 Miles / 40 KilometersSafer SlopeSlightly smaller range
Battery ChargerTypeExternal (off-board) ChargerExternal (off-board) ChargerSame manufacture
Voltage Output24 VDC, 5Ah24 VDC, 5AhSame type, higher ampage
ModelHP8204-5AHP8204B-5ASame Manufacturer

{9}------------------------------------------------

{10}------------------------------------------------

COMPARISON OF SUBJECT DEVICE (QUINGO Toura) AND PREDICATE DEVICE (Merits Transformer K142162)

ITEMSSUBJECT DEVICEREFERENCE DEVICESafety & effectiveness of subject device compared to the Reference device.
Brand NameToura 2Merits TransformerDifferent Make
ManufacturerAdvanced VehicleConcepts LtdMerits HealthProducts Co., Ltd.SameManufacturer
ModelNumberS350S941Different Model
510K NoK170634K142162DifferentSubmission
SeriesPower MobilityScooter SeriesPower MobilityScooter SeriesSame
Indications forUseThe devicesprovide mobility toadult personslimited to a sittingposition andcapable tooperate a fewsimple controls.The devicesprovide mobility toadult personslimited to a sittingposition andcapable to operatea fewsimple controls.Same Intendeduse and targetpopulation
Warranty1 Year1 YearSame
Product CodeINIINISame
Weight Limit485 lbs. / 220 Kg350 lbs. / 159 KgHeavier Carrying/User WeightCapacity
ElectronicsControllerDynamics Rhino 2DS160Dynamics Rhino 2DS120Same type withdifferent volumecontrollers
MotorDC 24V, 650 WDC 24V, 400 WLarger Power
FootplatesAluminumAdjustableFootplatesABSIndividualAdjustablefootplates
SuspensionFrontIntegral Lever ArmCoil SpringsRearIndependent CoilSpringsIntegral Lever ArmCoil SpringsIndependent CoilSpringsSame
BackUpholsteryFabricFabricSame Material
BatteriesQuantity22Same Quantity
Type12V, 100Ah12V, 50AhLarger Capacity
PatientContactingMaterialsSeatVinyl FabricVinyl FabricSame
HandGripRubberRubberSame
Seat BeltSteel / ABS /Polyester FabricN/A (No Seat Belt)Additional SafetyFeature
OverallDimensionsOverallLength1390mm / 54"1300mm / 51.2"Larger Length
OverallWidth670mm / 26.4"686mm / 27.2"Smaller Width
OverallHeight1430 / 56.3"1245mm / 49"Higher Seatingposition
ArmrestTypesRemovable & Flip-backwardRemovable & Flip-backwardSame
FrameSteel FrameSteel FrameSame
SeatDimensionsSeatWidth540mm / 21.3"470mm / 18"Larger Width
SeatHeight(FromGround)720-790mm /28.3-33.1"635-685mm / 25-27"Higher Seatingposition
SeatDepth(Actual)470mm / 18.5"400mm / 16"Larger Depth
SeatHeight(FromDeck)520-590mm / 20.5-23.2"430-480mm / 17-19"Higher Seatingposition
WheelsFront10" PneumaticTires12" PneumaticTiresSmaller size,same type
Front (2 xOuterWheels)9" Pu TiresN/AExtra (5 Wheeldesign)
Rear(DriveWheels)12" PneumaticTires12" PneumaticTiresSmaller size,same type
TurningRadius1320mm / 52"1670mm / 65.7"Smaller turningcircle
GroundClearance155mm / 6.1"90mm / 3.5"Larger clearance
Kerb ClimbingAbility100mm / 4"100mm / 4"Same
ScooterWeightTotalUnlaiden361.3 lbs. / 163.9Kg118kg / 260 lbs.Larger Weight
WithoutBatteries216.7 lbs. / 98.3Kg84kg / 185 lbs.Larger Weight
Maximum Speed6 Mph / 12.9 Kph7.5 mph / 12 KphLarger speed
Incline / Gradient12 Degrees10 DegreesSafer Slope
Range (upto)55 Miles / 88 Kilometers25 Miles / 40 KilometersLarger range
Battery ChargerTypeExternal (off-board) ChargerExternal (off-board) ChargerSame
Voltage Output24 VDC, 8Ah24 VDC, 5AhSame type, higher ampage
Model4C24080AHP8204B-5ADifferent type

{11}------------------------------------------------

{12}------------------------------------------------

COMPARISON OF SUBJECT DEVICE (QUINGO FLYTE) AND PREDICATE DEVICE (Merits Transformer K142162)

ITEMSSUBJECTDEVICEPREDICATEDEVICESafety &effectiveness ofsubject devicecompared to thepredicate device.
Brand NameFlyteMeritsTransformerDifferent Make
ManufacturerAdvancedVehicleConcepts LtdMerits HealthProducts Co., Ltd.Same Manufacturer
ModelNumberS551S941Different Model
510K NoK170634K142162Different Submission
SeriesPower MobilityScooter SeriesPower MobilityScooter SeriesSame
Indications forUseThe devicesprovide mobilityto adult personslimited to a sittingposition andcapable tooperate a fewsimple controls.The devicesprovide mobility toadult personslimited to a sittingposition andcapable to operatea fewsimple controls.Same Intended useand targetpopulation
Warranty1 Year1 YearSame
Product CodeINIINISame
Weight Limit350 lbs. / 159 Kg350 lbs. / 159 KgSame Carrying/ User Weight Capacity
ElectronicsControllerDynamics Rhino 2 DS90Dynamics Rhino 2 DS120Same type with different volume controllers
MotorDC 24V, 250 WAluminumAdjustableFootplatesDC 24V, 400 WLess Power due to lower max speed on the Subject device
FootplatesFootplatesABSIndividual Adjustable footplates
SuspensionFrontN/AIntegral Lever ArmCoil SpringsNo suspension on subject device
RearN/AIndependent CoilSpringsNo suspension on subject device
BackUpholsteryFabricFabricSame Material
BatteriesQuantity42Greater QuantityLower amp ofbatteries in subject.Total amps issubject is 88ampagainst 100amps inpredicate
Type12V, 22Ah12V, 50Ah
PatientContactingMaterialsSeatVinyl FabricVinyl FabricSame
HandGripRubberRubberSame
Seat BeltSteel/ ABS/Polyester FabricN/A (No Seat Belt)Additional SafetyFeature
OverallDimensionsOverallLength1100mm / 43.3"1300mm / 51.2"Smaller Length -more maneuverable
OverallWidth590mm / 23.2"686mm / 27.2"Smaller Width -more maneuverable
OverallHeight1010mm / 39.8"1245mm / 49"Smaller in Length toassistmaneuverability
ArmrestTypesRemovable &Flip-backwardRemovable & Flip-backwardSame
FrameSteel FrameSteel FrameSame
SeatDimensionsSeatWidth410mm / 16"470mm / 18"Smaller Width
SeatHeight(FromGround)710mm / 28"635-685mm / 25-27"Higher Seatingposition
SeatDepth(Actual)380mm / 14.9"400mm / 16"Smaller Depth
SeatHeight(FromDeck)500m / 19.7"430-480mm / 17-19"Higher Seatingposition
WheelsFront10" PneumaticTires12" PneumaticTires
Front (2 xOuterWheels)7" Pu TiresN/AExtra (5 Wheeldesign)
Rear(DriveWheels)10" PneumaticTires12" PneumaticTiresSmaller size
TurningRadius1020mm / 40.2"1670mm / 65.7"Smaller turningRadius
GroundClearance127mm / 5"90mm / 3.5"Larger ability
Kerb ClimbingAbility76mm / 3"100mm / 4"Slightly lessclearance
ScooterWeightTotalUnlaiden257 lbs / 116.5Kg118kg / 260 lbsLower Weight
WithoutBatteriesN/A84kg / 185 lbsStill Lower on weight
MaximumSpeed4 Mph / 6.4 Kph7.5 mph / 12 KphLower max speed
Incline /Gradient6 Degrees10 DegreesSmaller angle
Range (upto)23 Miles / 37Kilometers25 Miles / 40KilometersSlightly less range
BatteryChargerTypeExternal (off-board) ChargerExternal (off-board) ChargerSame
VoltageOutput24 VDC, 5Ah24 VDC, 5AhSame type
ModelHP8204-5AHP8204B-5ASame Manufacturer

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:

Testing information demonstrating safety and effectiveness of the subject device in the intended environment of use is supported by testing that was conducted in accordance with the FDA July 26, 1995 Draft, Reformatted 12/18/97, "Guidance Document for the Preparation of Premarket Notification [510(k)] Applications for Mechanical and Powered Wheelchairs, and Motorize Three-Wheeled Vehicles".

None of the testing demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazards. It was Advanced Vehicle Concepts Ltd.'s conclusion that testing met all relevant requirements of "Guidance Document for the Preparation of Premarket Notification [510(k)] Applications for Mechanical and Powered Wheelchairs, and Motorize Three-Wheeled Vehicles".

The following National and International Standards were utilized for testing the subject devices, which established substantial equivalence between the subject devices and the predicate devices:

Performance Standards:

16-195 ISO 7176-1 Third edition 2014-10-01:- Wheelchairs - Part 1: Determination of static stability

16-159 ISO 7176-2 Second edition 2001-06-15:- Wheelchairs - Part 2: Determination of dynamic stability of electric wheelchairs

16-192 ISO 7176-3 Third edition 2012-12-15: - Wheelchairs - Part 3: Determination of effectiveness of brakes

16-162 ISO 7176-4 Third edition 2008-10-01:- Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range

16-163 ISO 7176-5 Second edition 2008-06-01:- Wheelchairs - Part 5: Determination of overall dimensions, mass and manoeuvring space

16-29 ISO 7176-6 Second edition 2001-10-01:- Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairs

16-196 ISO 7176-7 First Edition 1998-05-15:- Wheelchairs - Part 7: Measurement of seating and wheel dimensions

16-197 ISO 7176-8 Second edition 2014-12-15:- Wheelchairs - Part 8: Requirements and test methods for static, impact and fatique strengths

{16}------------------------------------------------

16-167 ISO 7176-9: Third edition 2009-11-15:- Wheelchairs - Part 9: Climatic tests for electric wheelchairs

16-164 ISO 7176-10 Second edition 2008-11-01:- Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs

16-190 ISO 7176-11 Second edition 2012-12-01:- Wheelchairs - Part 11: Test dummies

16-25 ISO 7176-13 First edition 1989-08-01:- Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces

16-165 ISO 7176-14 Second edition 2008-02-15:- Wheelchairs - Part 14: Power and control systems for electrically powered wheelchairs and scooters - Requirements and test methods

16-191 ISO 7176-16 Second edition 2012-12-01:- Wheelchairs - Part 16: Resistance to ignition of postural support devices

16-166 ISO 7176-21 Second edition 2009-04-01:- Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters, and battery chargers

16-198 ISO 7176-22 Second edition 2014-09-01:- Wheelchairs - Part 22: Set-up procedures

16-194 ISO 7176-25 First edition 2013-07-15:- Wheelchairs - Part 25: Batteries and chargers for powered wheelchairs

16-187 ANSI RESNA WC-1:2009- American National Standard for Wheelchairs - Volume 1 :- Requirements and Test Methods for Wheelchairs (including Scooters) Section 26: Vocabulary

ISO 9001:2008:- Requirements for a quality management system

ISO 13485:2003 + AC: 2007:- Medical devices -- Quality management systems --Requirements for regulatory purposes

5-40: ISO 14971 Second Edition 2007-03-01:- Medical Devices - Application of Risk Management to Medical Devices. (General I (QS/RM))

BS EN 1041:2008:- Information supplied by the manufacturer of medical devices

BS EN 12184:2014:- Electrically powered wheelchairs, scooters and their chargers. Requirements and test methods

BS EN 12182:1999 Technical aids for disabled persons. General requirements and test methods

{17}------------------------------------------------

IEC 60601-1-2 Edition 3: 2007-03, Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests. (General II (ES/EMC))

EN 55011:2007 Industrial, scientific and medical equipment - Radio-frequency disturbance characteristics - Limits and methods of measurement

BS EN 1021-1/2: 2006 Fire Test to Upholstered Furniture

8. Discussion of Clinical Tests Performed:

Not applicable as there are no new indications for use which must be supported by a clinical data.

9. Software Information:

Software validation was conducted in accordance with a major level of concern designation in accordance with the FDA November 2005 document "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices".

10. Conclusions:

Conclusions drawn from the performance, bench and non-clinical tests demonstrate that the subject devices are as safe, effective and perform as well as the predicate devices.

§ 890.3800 Motorized three-wheeled vehicle.

(a)
Identification. A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.(b)
Classification. Class II (performance standards).