K Number
K170580

Validate with FDA (Live)

Date Cleared
2017-04-11

(42 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CARESTREAM Vue PACS is an image management system whose is to provide completely scalable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, retrieve and display images and data from all hospital modalities and information systems. The system is to be used by trained professionals including, but not limited to, physicians and medical technicians.

The system contains interactive tools in order to ease the process of analyzing and comparing three dimensional (3D) images. It is a single system that integrates review, dictation and reporting tools to create a productive work environment for the radiologists and physicians.

The system contains a Perfusion module with interactive tools to analyze and compare Computed Tomography Perfusion (CTP) and MR Perfusion (MRP) images of adult patients. Blood perfusion parameters are automatically calculated and displayed as a set of perfusion maps and perfusion tables include the calculation of parameters related to tissue flow (perfusion) and tissue blood volume.

The system contains a Diffusion Module with interactive tools to ease the process of analyzing and comparing MR Diffusion Weighted images (DWI) and MR Diffusion Tensor Imaging (DTI) of adult patients. This module is used to visualize local water diffusion properties from the analysis of diffusion- weighted MRI data.

The system supports Subtractive tools to aid with the analysis of Digital Subtraction Angiography (DSA) images in both interventional radiology and cardiology. Subtraction automatically subtracts a mask from contrast frames of an X-Ray Angiography study for visualization of vascular anatomy and pathology of adult patients.

Device Description

CARESTREAM Vue PACS is an image management system. It provides functionality to allow remote site access to image and patient data enabling diagnostic reading through industry standard interfaces. The system contains interactive tools in order to ease the process of analyzing and comparing three dimensional (3D) images.

The modification to the Vue PACS device described in this submission includes the addition of the MR Perfusion and Diffusion Module.

AI/ML Overview

The provided document describes the Carestream Vue PACS system, specifically focusing on the addition of MR Perfusion and Diffusion Modules. It outlines the performance testing conducted to demonstrate substantial equivalence to a predicate device (Olea Sphere v3.0). However, the document primarily discusses high-level details of the study rather than providing specific, granular acceptance criteria and detailed performance results with numerical thresholds.

Below is a structured attempt to answer your questions based on the provided text, supplemented with interpretations where specific details are missing (and noted as such).


Acceptance Criteria and Device Performance

The document states, "Predefined acceptance criteria were met and demonstrated that the device is as safe, as effective, and performs as well as or better than the predicate device." However, the exact, quantifiable acceptance criteria are not explicitly listed. The study objective was to "compare the similarity of the parametric maps," implying that the acceptance criterion was a qualitative or quantitative measure of this similarity, but no specific threshold (e.g., a certain correlation coefficient, a maximum allowed difference) is provided.

Given the available information, the table would look like this, with the understanding that specific numerical criteria are not provided in the text:

Table 1: Acceptance Criteria and Reported Device Performance

Feature/Metric Tested (Implied)Acceptance Criteria (Not Explicitly Stated, But Implied)Reported Device Performance (Summary)
Similarity of Perfusion Parametric Maps (rCBV, rCBF, TTP, MTT, TMAX, TMIP)Parametric maps generated by the investigational device should be "similar" to those generated by the predicate device (Olea Sphere v3.0)."Results of this study demonstrated substantial equivalence in the similarity of the parametric maps of the Carestream Vue PACS MR Perfusion and Diffusion module as compared with the Olea Sphere V3.0."
Similarity of Diffusion Parametric Maps (ADC, FA, RA, VR, AD, RD)Parametric maps generated by the investigational device should be "similar" to those generated by the predicate device."Results of this study demonstrated substantial equivalence in the similarity of the parametric maps of the Carestream Vue PACS MR Perfusion and Diffusion module as compared with the Olea Sphere V3.0."
Overall Function and UsabilityIntended workflow, related performance, and overall function established."These studies have demonstrated the intended workflow, related performance, overall function, shipping performance, verification and validation of requirements for intended use, and reliability of the system including both software and hardware requirements."
Reliability (Software and Hardware)System reliability confirmed."Reliability of the system including both software and hardware requirements" was demonstrated.
Conformance to SpecificationsDevice conforms to its specifications."Non-clinical test results have demonstrated that the device conforms to its specifications."

Study Details

  1. Sample Size used for the test set and the data provenance:

    • Sample Size: Not specified. The document only mentions "A Clinical study was performed."
    • Data Provenance: Not specified (e.g., country of origin, retrospective or prospective).
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This was a comparative study between an investigational device and a predicate device, focusing on the "similarity of the parametric maps." It doesn't appear to establish a ground truth through expert consensus in the traditional diagnostic sense (e.g., identifying disease vs. no disease). Instead, the "ground truth" seems to be conceptually based on the output of the predicate device.
    • Therefore, the number of experts and their qualifications for establishing ground truth are not applicable/not mentioned in the context of this study as described.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Adjudication methods are typically used when multiple human readers interpret cases and their agreement or disagreement needs to be resolved to establish a "ground truth" or consensus. Since this study focused on comparing the algorithmic output (parametric maps) of two systems, and not on human interpretation needing resolution, adjudication methods are not applicable/not mentioned.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC study assessing human reader improvement with AI assistance was not done based on this document. The study explicitly states its objective was to "compare the similarity of the parametric maps of MR DSC-Perfusion (Perfusion Module), MR DWI (Diffusion Module) and MR DTI (Diffusion Module) in Carestream Vue PACS ("investigational device") to the Olea Sphere PACS with Perfusion and DWI Modules ("predicate device")." This describes a comparison of the software's output, not a human reader performance study.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Yes, the primary clinical study described was effectively a standalone comparison between the investigational device's algorithmic output and the predicate device's algorithmic output. It focused on the "similarity of the parametric maps" generated by the respective modules, which implies direct algorithmic comparison rather than assessing human-in-the-loop performance.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • As interpreted above, the "ground truth" in this comparative study was essentially the outputs (parametric maps) generated by the predicate device (Olea Sphere v3.0). The study aimed to show that the investigational device's outputs were "similar" or "substantially equivalent" to those of the legally marketed predicate. This is not a ground truth derived from pathology, expert consensus on disease, or outcomes data.
  7. The sample size for the training set:

    • The document describes a "Clinical study" for validation and "non-clinical (bench) testing" for verification. It does not mention a training set or details about the development of the algorithms. The focus is purely on the validation of the integrated modules. Therefore, the sample size for the training set is not provided.
  8. How the ground truth for the training set was established:

    • Since the document does not mention details about the training set or algorithm development, how ground truth for the training set was established is not provided.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of three human profiles facing right, stacked on top of each other. The profiles are black and have a flowing, abstract design. The image is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in black, uppercase letters.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 11, 2017

Carestream Health, Inc. % Ms. Diane Koetter Sr. Manager, Regulatory Affairs and Quality Systems 150 Verona Street ROCHESTER NY 14608

Re: K170580

Trade/Device Name: Carestream Vue PACS Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 17, 2017 Received: February 28, 2017

Dear Ms. Koetter:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael D.'Hara
For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170580

Device Name Carestream Vue PACS

Indications for Use (Describe)

"The CARESTREAM Vue PACS is an image management system whose is to provide completely scalable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, retrieve and display images and data from all hospital modalities and information systems. The system is to be used by trained professionals including, but not limited to, physicians and medical technicians.

The system contains interactive tools in order to ease the process of analyzing and comparing three dimensional (3D) images. It is a single system that integrates review, dictation and reporting tools to create a productive work environment for the radiologists and physicians.

The system contains a Perfusion module with interactive tools to analyze and compare Computed Tomography Perfusion (CTP) and MR Perfusion (MRP) images of adult patients. Blood perfusion parameters are automatically calculated and displayed as a set of perfusion maps and perfusion tables include the calculation of parameters related to tissue flow (perfusion) and tissue blood volume.

The system contains a Diffusion Module with interactive tools to ease the process of analyzing and comparing MR Diffusion Weighted images (DWI) and MR Diffusion Tensor Imaging (DTI) of adult patients. This module is used to visualize local water diffusion properties from the analysis of diffusion- weighted MRI data.

The system supports Subtractive tools to aid with the analysis of Digital Subtraction Angiography (DSA) images in both interventional radiology and cardiology. Subtraction automatically subtracts a mask from contrast frames of an X-Ray Angiography study for visualization of vascular anatomy and pathology of adult patients."

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Carestream

"510(k) Summary"

510(k) Owner Name:510(k) Owner Address:Carestream Health, Inc.150 Verona StreetRochester, New York 14608
510(k) Owner Phone:510(k) Owner Fax:585-627-6505585-627-8802
Contact Person & Info:Diane KoetterSr. Manager, Regulatory Affairs and Quality Systemsdiane.koetter@carestream.com585-627-6505
Date Summary Prepared:February 17, 2017
Device Trade Name:Device Common Name:Classification Name:Carestream Vue PACSSystem, image processing, radiologicalPicture archiving and communications system
Device Class:Device Code:Regulation Number:2LLZ21 CFR 892.2050
Predicate Device:Olea Sphere v3.0Manufactured by: Olea Medical510(k) No .: K152602 (March 3, 2016)

Device Description:

CARESTREAM Vue PACS is an image management system. It provides functionality to allow remote site access to image and patient data enabling diagnostic reading through industry standard interfaces. The system contains interactive tools in order to ease the process of analyzing and comparing three dimensional (3D) images.

The modification to the Vue PACS device described in this submission includes the addition of the MR Perfusion and Diffusion Module.

Indications for Use / Intended Use:

{4}------------------------------------------------

The Indications for Use for the device, as described in its labeling, are:

"The Carestream Vue PACS is an image management system whose intended use is to provide completely scalable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute, retrieve and display images and data from all hospital modalities and information systems. The system is to be used by trained professionals including, but not limited to, physicians and medical technicians.

The system contains interactive tools in order to ease the process of analyzing and comparing three dimensional (3D) images. It is a single system that integrates review, dictation and reporting tools to create a productive work environment for the radiologists and physicians.

The system contains a Perfusion module with interactive tools to analyze and compare Computed Tomography Perfusion (CTP) and MR Perfusion (MRP) images of adult patients. Blood perfusion parameters are automatically calculated and displayed as a set of perfusion maps and perfusion tables. The perfusion tables include the calculation of parameters related to tissue flow (perfusion) and tissue blood volume.

The system contains a Diffusion Module with interactive tools to ease the process of analyzing and comparing MR Diffusion Weighted images (DWI) and MR Diffusion Tensor Imaging (DTI) of adult patients. This module is used to visualize local water diffusion properties from the analysis of diffusion- weighted MRI data.

The system supports Subtraction with interactive tools to aid with the analysis of Digital Subtraction Angiography (DSA) images in both interventional radiology and cardiology. Subtraction automatically subtracts a mask from contrast frames of an X-Ray Angiography study for visualization of vascular anatomy and pathology of adult patients."

The intended use for this device, as determined by descriptions and the proposed labeling contained in this submission, is similar to the Indications for Use statement provided above.

The Indications for Use for the subject device is similar to the predicate device and the intended use remains unchanged. Any variation in features or technical specifications have been identified and addressed through testing (described below) to support a substantial equivalence determination.

Comparison of Technological Characteristics:

Based upon information provided within this submission, we believe that the Carestream Vue PACS is substantially equivalent to the legally marketed Olea Sphere v3.0 Perfusion and Diffusion Modules (predicate device).

Comparison of the investigational and predicate devices:

{5}------------------------------------------------

  • . Both devices provide a Perfusion plug-in to compute optimized rCBV, rCBF, TTP, MTT, TMAX, and TMIP parametric maps from raw perfusion series.
  • Both devices provide automatic or manual Arterial Input Function (AIF). ●
  • Both devices provide automatic or manual Venous Output Function (VOF). ●
  • Both devices provide background segmentation, where the hypoperfused and lesion ● tissues are calculated based upon AIF and VOF.
  • Both devices support sSVD (standard singular value decomposition), cSVD . (blockcirculant singular value decomposition), and oSVD (oscillation singular value decomposition) deconvolution methods/algorithms for generation of the perfusion maps.
  • Both devices support motion correction to minimize motion artifacts when . generating the perfusion maps.
  • Both devices calculate lesion and hypoperfused volumes based upon parameters and ● values supplied by the user.
  • The predicate device differs from the investigational device in that it supports a ● Bayesian deconvolution methods/algorithms for generation of the perfusion maps, while the investigational device does not.
  • Both devices support the ADC map for Diffusion Weighted Imaging. ●
  • Both devices support map display for Diffusion Tensor Imaging of the following ● types Fractional Anisotropy (FA), Relative Anisotropy (RA), Volume Ratio (VR), ADC/Mean Diffusivity (ADC), Axial Diffusivity (AD), and Relative Diffusivity (RD).

Discussion of Testing

The performance characteristics and operation / usability of the Carestream Vue PACS MR Perfusion and Diffusion Module were evaluated in non-clinical (bench) testing. These studies have demonstrated the intended workflow, related performance, overall function, shipping performance, verification and validation of requirements for intended use, and reliability of the system including both software and hardware requirements. Non-clinical test results have demonstrated that the device conforms to its specifications. Predefined acceptance criteria were met and demonstrated that the device is as safe, as effective, and performs as well as or better than the predicate device.

A Clinical study was performed to establish substantial equivalence for the Carestream Vue PACS MR Perfusion and Diffusion Module to its predicate, the Olea Sphere V3.0. The objective of this study was to compare the similarity of the parametric maps of MR DSC-Perfusion (Perfusion Module), MR DWI (Diffusion Module) and MR DTI (Diffusion Module) in Carestream Vue PACS ("investigational device") to the Olea Sphere PACS with Perfusion and DWI Modules ("predicate device"). The results of this study demonstrated substantial equivalence in the similarity of the parametric maps of the Carestream Vue PACS MR Perfusion and Diffusion module as compared with the Olea Sphere V3.0.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).