K Number
K170489

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-05-23

(95 days)

Product Code
Regulation Number
878.4630
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Clarify Medical Phototherapy System is an Ultraviolet Light Emitting Medical Device. It is intended for use in localized phototheraputic treatment of dermatologic conditions such as psoriasis, vitiligo, atopic dermatitis (eczema), seborrheic dermatitis and lenkoderma on all skintypes (I-VI).

Device Description

The Clarify Phototherapy System is a handheld medical device that produces UVB light for the purpose of phototherapy. It was developed and will be marketed by Skylit Medical, Inc. DBA Clarify Medical, Inc. and is referred to herein as "Clarify Medical" or "Clarify". The Clarify Medical Phototherapy System is an Ultraviolet Light Emitting Medical device. It is indicated for use in localized phototherapeutic treatment of dermatologic conditions such as psoriasis, vitiligo, atopic dermatitis (eczema), seborrheic dermatitis and leukoderma on all skin types (I-VI). The system consists of the following components and accessories:

  • Handheld UV light Emission device that delivers the UVB light
  • Mobile Software Application that runs on smartphone (Android and iOS) platforms. This Application is the primary user interface, and provides patient guidance when phototherapy is delivered and treatment reminders
  • Physician Portal Software is a cloud based software application that links the patient to the prescribing physician during home therapy.
AI/ML Overview

The provided text describes a 510(k) premarket notification for the "Skylit Phototherapy System" (later referred to as "Clarify Medical Phototherapy System"). This device is a handheld UV light treatment system for dermatologic conditions. The submission aims to demonstrate substantial equivalence to previously marketed devices.

Based on the provided document, the device in question is a phototherapy system, not an AI/algorithm-based device. Therefore, the questions related to AI acceptance criteria, sample size for AI test sets, number of experts for ground truth, adjudication methods, MRMC studies, standalone algorithm performance, and training set information are not applicable to this medical device submission.

The acceptance criteria provided focus on the electrical, optical, software, and biocompatibility safety and performance of the phototherapy device. The study proving these criteria are met consists of various performance tests and validations.

Here's a breakdown of the applicable information:

1. A table of acceptance criteria and the reported device performance:

Element of ComparisonMethod for Safety AssessmentTest Result Meet Design Input Specifications and Demonstrate EquivalenceNotes
UBV light Source, Max Power OutputPerformance Testing (optical emissions spectrum measured at the treatment surface, biologically weighted emissions curve computed using the International Commission on Illumination accepted erythema curve, optical power at treatment surface measured). Evaluated in accordance with IEC 60601-2-57.YesVerification testing demonstrated weighted peak of emissions within established therapeutic range (300-320nm) and optical power at treatment surface met specified requirements.
Power SourceElectrical Safety Testing, Battery Performance Testing. Evaluated in accordance with IEC 60601-1 and IEC 60601-1-4.YesNot explicitly detailed in the provided text beyond "Yes."
Prescription Stored on the deviceSystem Validation. Compliance with relevant FDA guidance documents for software (e.g., Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices) and IEC 62304.YesThis refers to prescriptions being stored on a cloud server for the subject device, a feature that distinguishes it from some predicate devices.
Treatment RemindersSoftware Validation. Compliance with relevant FDA guidance documents for software (e.g., Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices) and IEC 62304.YesThis feature is built into the mobile software application.
Auto Calculation of treatment adjustmentSystem Validation. Compliance with relevant FDA guidance documents for software (e.g., Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices) and IEC 62304.YesThis capability is built into the software in accordance with the prescribed protocol.
Skin Contact MaterialsPerformance testing. Evaluated for biocompatibility in accordance with ISO 10993-1 (cytotoxicity, sensitization, irritation) as a "surface device" with limited contact duration (< 24 hours).YesThe distal end material has been tested to ISO-10993.
Wireless communications between handheld and consoleSystem Validation. Compliance with relevant FDA guidance documents for Mobile Medical Applications and Radio Frequency Wireless Technology in Medical Devices. Also, electromagnetic compatibility (EMC) assessed according to IEC 60601-1-2.YesThe device uses Bluetooth wireless communication between the handheld unit and the mobile phone software application.
Records RetentionSystem Validation. Compliance with relevant FDA guidance documents for software (e.g., Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices) and IEC 62304.YesTreatment data is stored and uploaded to a server.
Remote Physician Review of the resultsSystem Validation. Compliance with relevant FDA guidance documents for software (e.g., Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices) and IEC 62304.YesThis feature is enabled by the cloud-based Physician Portal Software.
Electrical SafetyTesting to applicable sections of IEC 60601-1 and IEC 60601-1-4.Not explicitly stated as a row in Table 2, but mentioned in "Standards and Guidance Documents" section.Not explicitly detailed beyond "Yes."
Software and Wireless SafetyReview and compliance with FDA guidance documents: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, Mobile Medical Applications, Radio Frequency Wireless Technology in Medical Devices. Consideration of IEC 62304.Covered by System ValidationNot explicitly detailed beyond "Yes."
Electromagnetic Compatibility (EMC)Assessed according to IEC 60601-1-2.Covered by Wireless Communications validationNot explicitly detailed beyond "Yes."
Human Factors/UsabilityFormative studies and software validation. Compliance with IEC 62366 to be verified by an independent third party prior to marketing.Will be confirmedThe document states: "Clarify medical will not market the device until compliance with IEC 62366: Medical devices – Application of usability engineering to medical devices is verified by an independent third party." This is a pre-market condition.

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not applicable. This submission relies on performance testing and comparison to predicate devices, not a clinical study with a test set of patient data as might be used for an AI/algorithm-based device.
  • Data Provenance: Not applicable. The "data" here refers to engineering test results and compliance with medical device standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. Ground truth as understood in the context of AI (e.g., expert labels on medical images) is not relevant for this device's testing. Safety and performance are established through engineering and systems validation against established standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable. There is no external "test set" requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This is not an AI-assisted device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is not an AI or algorithm-only device. It's a hardware device with associated software to manage its function.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • Ground Truth: For this type of device, "ground truth" relates to compliance with engineering and safety standards. Examples include:
    • Optical Emissions: Measured values against specified therapeutic ranges (300-320nm).
    • Electrical Safety: Compliance with IEC 60601-1 standards.
    • Software Functionality: Validation against design input specifications and recognized software lifecycle processes (IEC 62304) and FDA guidance for medical device software.
    • Biocompatibility: Results of cytotoxicity, sensitization, and irritation tests against ISO 10993 standards.
    • These are objective, measurable criteria, not subjective interpretations requiring expert consensus as in AI image analysis.

8. The sample size for the training set:

  • Not applicable. This is not an AI/machine learning device that requires a training set.

9. How the ground truth for the training set was established:

  • Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it. The caduceus is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the logo.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W O66-G609 Silver Spring, MD 20993-0002

May 23, 2017

Skylit Medical Don Canal Vice President 401 North A Street Suite 775 San Diego, California 92101

Re: K170489

Trade/Device Name: Skylit Phototherapy System Regulation Number: 21 CFR 878.4630 Regulation Name: Ultraviolet Lamp For Dermatologic Disorders Regulatory Class: Class II Product Code: FTC Dated: February 22, 2017 Received: February 23, 2017

Dear Don Canal:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely, Jennifer R. Stevenson -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170489

Device Name Clarify Medical Phototherapy System

Indications for Use (Describe)

The Clarify Medical Phototherapy System is an Ultraviolet Light Emitting Medical Device. It is intended for use in localized phototheraputic treatment of dermatologic conditions such as psoriasis, vitiligo, atopic dermatitis (eczema), seborrheic dermatitis and lenkoderma on all skintypes (I-VI).

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

"510(k) Summary" as required by section 807.92(c)

Date Prepared:May 22, 2017
Manufacturer NameClarify Medical, Inc.401 N. A Street, Suite 775San Diego, CA 92101, USATelephone: (972) 955-7644
Official ContactDon CanalRegulatory Affairs401 N. A StreetSan Diego, CA 92101, USATelephone:(972) 955-7644Email: don@clarifymed.com

DEVICE NAME AND CLASSIFICATION

Trade/Proprietary NameClarify Phototherapy System
Common NameLight, Ultraviolet, Dermatological
Classification NameClassification RegulationsUltraviolet lamp for dermatologic disorders21 CFR878.4630Class II
Product CodeFTC
Submission Type510(k)
Classification PanelPremarket ReviewGeneral & Plastic SurgeryODE/DSD/General Surgery Devices Branch One(GSDB1)

Predicate Devices

The Clarify System is substantially equivalent to the other legally marketed phototherapy, UVB treatment devices. Specifically, the Clarify Phototherapy System is substantially equivalent to the Lerner Medical Inc. Levia Phototherapy System (K040062/K090097), Daavlin Distributing Co. DermaPal (K073587) and Allux, Inc. Resolve UVB Phototherapy System (K072035).

Device Description

The Clarify Phototherapy System is a handheld medical device that produces UVB light for the purpose of phototherapy. It was developed and will be marketed by Skylit Medical, Inc. DBA Clarify Medical, Inc. and is referred to herein as "Clarify Medical" or "Clarify". The Clarify Medical Phototherapy System is an Ultraviolet Light Emitting Medical device.

It is indicated for use in localized phototherapeutic treatment of dermatologic conditions

{4}------------------------------------------------

such as psoriasis, vitiligo, atopic dermatitis (eczema), seborrheic dermatitis and leukoderma on all skin types (I-VI). The system consists of the following components and accessories:

  • Handheld UV light Emission device that delivers the UVB light 1.
    1. Mobile Software Application that runs on smartphone (Android and iOS) platforms. This Application is the primary user interface, and provides patient guidance when phototherapy is delivered and treatment reminders
    1. Physician Portal Software is a cloud based software application that links the patient to the prescribing physician during home therapy.

Federal (U.S.) law restricts this device to sale by or on the order of a physician.

The Clarify System delivers UV light between the wavelengths of 300 and 320 nm using Light Emitting Diodes (LEDs). The handheld device is powered by a rechargeable battery and is connected to a smartphone software application that enables the device and provides feedback during treatment. The smartphone also provides reminders on treatment days and provides a means for patient feedback and dose adjustment. The smartphone application also provides the means for the patient to take photographs for comparison purposes.

Indications for Use Statement

The system is indicated for the treatment of psoriasis, vitiligo, atopic dermatitis (eczema). seborrheic dermatitis and leucoderma on all skin types (I-VI). The device can be used in the comfort of a patient's home or in a physician's office.

Comparison to Predicate Devices

The Clarify System is substantially equivalent to the other legally marketed phototherapy, UVB treatment devices. Specifically, the Clarify Phototherapy System is substantially equivalent to the Lerner Medical Inc. Levia Phototherapy System (K040062/K090097). Daavlin Distributing Co. DermaPal (K073587) and Allux, Inc. Resolve UVB Phototherapy System (K072035). The detailed comparison between the Subject Device and the Predicate devices is provided in Table 1 below.

{5}------------------------------------------------

Subject DevicePredicate DevicePredicate DevicePredicate Device
Clarify MedicalPhototherapy System(Subject Device)Lerner MedicalDevices, Inc.Levia PhototherapySystem / LH-75TPhototherapySystemK040062 / K090097Daavlin DistributingCo.DermaPal K073587AlluxResolveUVBPhototherapy SystemK072035
Image: Clarify Medical Phototherapy SystemImage: Levia Phototherapy SystemImage: DermaPal K073587Image unavailable
Indicationsfor UseThe Clarify MedicalPhototherapy System isan Ultraviolet LightEmitting Medicaldevice. It is intendedfor use in localizedphototherapeutictreatment ofdermatologicconditions such aspsoriasis, vitiligo,atopic dermatitis(eczema), seborrheicdermatitis andleukoderma on all skintypes (I-VI).K040062: The LeviaPhototherapy System isintended for use inUVB phototherapy forthe treatment ofpsoriasis, vitiligo,atopic dermatitis(eczema), seborrheicdermatitis andleucoderma. The LeviaPhototherapy System isintended for use on allskin types (I-VI).K090097: The LH-75TPhototherapy System isintended for use, by orunder the direction of aphysician, in UVBphototherapy for thetreatment of psoriasis,vitiligo, atopicdermatitis (eczema),seborrheic dermatitisand leucoderma. TheLH-75T PhototherapySystem is intended foruse on all skin types (I-VI).The DermaPal® is ahandheld ultravioletlight emittingfluorescent lamp. It isintended for use, by orunder the direction of aphysician, for thetreatment of psoriasis,vitiligo, and atopicdermatitis (eczema) onall skin types (I - VI).The Resolve™ UVBPhototherapy System isan ultraviolet lightemitting medical devicefor localizedphototherapeutictreatment ofdermatologic conditionssuch as psoriasis,vitiligo, atopicdermatitis (eczema),and seborrheicdermatitis. TheResolve™ UVBPhototherapy System isintended to be usedwith all Skin Types (I -VI).
Product Codes/ RegulationNumberFTC / 21 CFR878.4630FTC / 21 CFR878.4630FTC / 21 CFR878.4630FTC / 21 CFR878.4630
Treatmentarea$4.5 cm^2$$3 cm^2$11.4 x 2.54 cm12.8 x 9.0 cm
Subject DevicePredicate DevicePredicate DevicePredicate Device
Clarify MedicalPhototherapy System(Subject Device)Lerner MedicalDevices, Inc.Levia PhototherapySystem / LH-75TPhototherapySystemK040062 / K090097Daavlin DistributingCo.DermaPal K073587AlluxResolveUVBPhototherapy SystemK072035
UVB lightsourceLED UV lampsArc LampFluorescent UV lampsLED UV lamps
UserInterfaceTouch screen onMobile Device andStart button onControl Box withcables to handhelddeviceControl Box withcables to handhelddeviceControl Box withcables to handhelddevice
Max PowerOutput3-15 mW/cm290-100 mW/ cm23.9-6.8 mW/ cm21 mW/cm2
UV lightWavelength300-320 nm300-320 nm300-320 nm300-320nm
Mode ofoperationHand piece used withwireless connectionto mobile phoneHand piececonnected to consolewith touch screencontrol panel tocontrol treatmentHand piece used instandalone treatmentmodePatient attachedunit connectedto a powersupply used instandalonetreatment mode
Physicianprovides dosinginstructions andphysician canadjust time/dosebased onPhysician providesdosing instructionsand patient canadjust time/dosebased on treatmentoutcomePhysician providesdosing instructionsand patient canadjust time/dosebased on treatmentoutcomeUnknown
Increase UVBtreatment Doseafter eachtreatmentYes, thiscapability is builtinto the SW inaccordance withthe prescribedprotocol.Yes, this capabilityis built into the SWin accordance withthe prescribedprotocol.Instructions areprovided in theDFU. Not built intothe the SW, thepatient mustmanually calculateand enter the dose.Unknown
DeliverymethodSpot treatment windowSpot treatment attachment(LiteSpot™), Quartz fiber-optic brush attachment(LiteBrush™)Spot treatment wand withintegrated combSpot treatment window
PowerSourceRechargeable batteryAC outletAC outletAC outlet
Prescriptionstored onDeviceYes - Cloud serverYes - USBNoNo
ScalpAttachmentNoYesYesNo
UseEnvironmeHome / Physician'sofficeHome / Physician'sofficeHome / Physician'sofficeHome / Physician'soffice
Subject DevicePredicate DevicePredicate DevicePredicate Device
Clarify MedicalPhototherapy System(Subject Device)Lerner MedicalDevices, Inc.Levia PhototherapySystem / LH-75TPhototherapySystemK040062 / K090097Daavlin DistributingCo.DermaPal K073587AlluxResolveUVBPhototherapy SystemK072035
IPX -Rating/waterResistanceNot water resistantNot water resistantNot water resistantNot water resistant
CommunicationsbetweenBluetooth wirelessWired connectionWired connectionWired connection
Recordsretentionand remotephysicianreview ofresults.Yes. Treatment data isstored and uploaded toServer for remotereviewYes. Treatment data isstored locally on thedevice. no remotereview of data ispossibleNo record retentioncapabilityNo record retentioncapability
Treatmentday/Timereminderset by UserYesNoNoNo
Timer tocontroltreatmentdurationYes - stored in theSoftware, based onPrescriptionYes - entered by theuser for each treatmentYes - entered by theuser for each treatmentYes - entered by theuser for each treatment
TargetPopulationPatients and cliniciansPatients andcliniciansPatients andcliniciansPatients andclinicians

Table 1 – Predicate Comparison Table

{6}------------------------------------------------

{7}------------------------------------------------

Summary of Performance Data – Bench

The determination of substantial equivalence is based on an assessment of non-clinical performance data. Table 2 below contains a summary of the performance testing that was submitted, referenced, or relied on in the premarket notification submission to determine substantial equivalence and how their results support a determination of substantial equivalence.

Element of ComparisonMethod for SafetyAssessmentTest Result MeetDesign InputSpecifications andDemonstrateEquivalence
UBV light Source, Max PowerOutputPerformance TestingYes

Table 2 – Performance Test Summary

{8}------------------------------------------------

Element of ComparisonMethod for SafetyAssessmentTest Result MeetDesign InputSpecifications andDemonstrateEquivalence
Power SourceElectrical Safety Testing,Battery PerformanceTestingYes
Prescription Stored on the deviceSystem ValidationYes
Treatment RemindersSoftware ValidationYes
Auto Calculation of treatmentadjustmentSystem ValidationYes
Skin Contact MaterialsPerformance testingYes
Wireless communications betweenhandheld and consoleSystem ValidationYes
Records RetentionSystem ValidationYes
Remote Physician Review of theresultsSystem ValidationYes

Performance Test Conclusion

Based on technical comparison, the Subject device has been shown to be equivalent to the predicate devices.

Standards and Guidance Documents

Optical emissions testing:

Clarify has verified that optical emissions within the 300-320nm range is equivalent to that of the predicate devices. With the LEDs switched on, the optical emissions spectrum is measured at the treatment surface. The biologically weighted emissions curve is then computed using the International Commission on Illumination accepted erythema curve. Clarify has completed verification testing to demonstrate that the weighted peak of emissions is within the established therapeutic range (300-320nm). The optical power at the treatment surface is also measured to ensure it meets the specified requirements. The device has been evaluated for safety and performance in accordance with IEC 60601-2-57 Medical electrical equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use.

Electrical safety:

Clarify has completed testing to the applicable sections of IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance, and IEC 60601-1-4 Medical electrical equipment – Part 1-4: General requirements for safety - Collateral Standard: Programmable electrical medical systems.

{9}------------------------------------------------

Software and wireless safety:

Due to the software component and wireless functionality of the device. Clarify will review and comply with the following FDA guidance documents: Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (Issued 5/11/2005), Guidance for Industry and FDA Staff: Mobile Medical Applications (Issued 9/25/2013), and Guidance for Industry and FDA Staff: Radio Frequency Wireless Technology in Medical Devices (Issued 8/13/2013). The following standard has been considered when evaluating the software: IEC 62304: Medical device software - Software life cycle processes.

Electromagnetic compatibility (EMC):

Since the system includes electronics, EMC has been assessed according to IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances.

Biocompatibility:

The distal end of the Clarify Phototherapy System will have limited contact duration (< 24 hours) with skin and is therefore considered a "surface device" according to Annex A of ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process. Based on the device type and contact duration, FDA Blue Book Memorandum #G95-1 Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," recommends the following biocompatibility tests be performed: cytotoxicity, sensitization, and irritation. The Clarify Device distal end material has been tested to ISO-10993.

Human factors:

Clarify medical has completed numerous formative studies and software validation of the Clarify System. Clarify medical will not market the device until compliance with IEC 62366: Medical devices – Application of usability engineering to medical devices is verified by an independent third party.


§ 878.4630 Ultraviolet lamp for dermatologic disorders.

(a)
Identification. An ultraviolet lamp for dermatologic disorders is a device (including a fixture) intended to provide ultraviolet radiation of the body to photoactivate a drug in the treatment of a dermatologic disorder if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug.(b)
Classification. Class II.