K Number
K170481
Manufacturer
Date Cleared
2017-05-16

(89 days)

Product Code
Regulation Number
866.2360
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Strep B Carrot Broth™ Kit is a selective and differential medium which is intended for the detection of Group B Streptococus (GBS) from anovaginal specimens collected from pregnant women. The medium is used as an aid in the qualitative determination of GBS colonization in pregnant women. The color change reaction from white to orange is representative of a positive result for presence of GBS. The medium requires 24 hours of incubation but positive results can be interpreted and reported as early as 16 hours. Due to the properties of Strep B Carrot Broth™ Kit, non-hemolytic GBS cannot be detected by the medium's color change and require subculture for identification. Any presumptive negative indicated by lack of color change at the end of the incubation period must be subcultured to a non-selective medium (e.g., Tryptic Soy Agar with 5% Sheep Blood) to confirm absence of GBS. Subculture must also be performed to recover isolates for conducting susceptibility testing as recommended for penicillin-allergic women.

Device Description

Strep B Carrot Broth™ is a selective and differential medium with selective components designed to enrich for Group B Streptococci. The production of a light orange, or red-orange pigment is a unique characteristic of β-hemolytic GBS due to reaction with substrates such as starch, and folate pathway inhibitors. GBS detection by color with Strep B Carrot Broth™ is only possible with β-hemolytic Group B Streptococci colonies, which provides evidence of a direct genetic linkage between pigment production in this media and hemolysin production. Non-hemolytic GBS can be recovered Strep B Carrot Broth™ upon subculture to 5% sheep blood agar plates.

AI/ML Overview

The device in question is the Strep B Carrot Broth™ Kit, a selective and differential medium intended for the detection of Group B Streptococcus (GBS) from anovaginal specimens collected from pregnant women. It aids in the qualitative determination of GBS colonization.

Here's an analysis of the acceptance criteria and the study proving the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria in terms of specific sensitivity or specificity thresholds. However, the performance data presented implies a comparison against a reference method (LIM Broth with subculture and biochemical testing). The "Substantially Equivalent?" column in the comparison table (page 4) suggests that similar performance for "Intended Use," "Methodology," "Inoculation," and "Sample Type" were considered for equivalence. For quantitative performance, the comparison against the reference method served as the implicit acceptance benchmark.

Based on the provided performance tables, here's a summary of the device's performance:

Criterion / Performance MetricReported Device Performance (Overall) - Color Reaction Only (Table 1)Reported Device Performance (Overall) - Color Reaction + Subculture of Negatives (Table 2)
Sensitivity for GBS detection (color reaction)87.73% (95% CI: 81.8 - 91.9)N/A (This metric is less relevant when incorporating subculture for negatives)
Specificity for GBS detection (color reaction)98.83% (95% CI: 97.6 - 99.4)N/A
Sensitivity for GBS detection (color reaction + subculture of negatives)N/A98.8% (95% CI: 95.6 - 99.7)
Specificity for GBS detection (color reaction + subculture of negatives)N/A97.9% (95% CI: 96.4 - 98.7)
Recovery of β-hemolytic GBS (Color reaction only, considering non-hemolytic as negative)90.4% (95% CI: 84.7-94.1) * (based on footnote 3, page 5)N/A
Specificity for β-hemolytic GBS (Color reaction only, considering non-hemolytic as negative)98.5% (95% CI: 97.2-99.2) * (based on footnote 3, page 5)N/A
Limit of Detection (LoD)10^0 CFU/mL (for S. agalactiae ATCC 12386 and ATCC 12403)N/A
Analytical Reactivity (GBS strains)54/54 (100%) ATCC and clinical GBS strains produced expected color or recovered upon subcultureN/A
Analytical Specificity (Non-target organisms)78/78 (100%) non-target organisms produced negative color reactionN/A
Microbial InterferenceTarget organisms recovered/color produced in presence of high conc. (1.5 x 10^8 CFU/mL) of all but one non-target. E. faecalis required lowering its concentration for recovery/color reaction.N/A
Incubation Time Range16-24 hoursN/A
Specimen Stability for Transport MediaVaried by transport system; up to 96 hours at 2-8°C for Healthlink, and up to 120 hours at 2-8°C for TransPRO™.N/A
Reproducibility>95% agreement with known test resultsN/A

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Clinical Performance Study: 771 valid specimens. An initial 884 specimens were collected, but 113 were excluded.
  • Data Provenance:
    • Country of Origin: Not explicitly stated, but the study was conducted at "four geographically diverse hospitals with routine GBS specimen." Given the FDA submission from a US company (Hardy Diagnostics, Santa Maria, CA), it is highly probable the data is from the United States.
    • Retrospective or Prospective: The study describes evaluating the performance of the Strep B Carrot Broth™ "against routine culture," using "routine GBS specimens." This phrasing, along with the collection of specimens and subsequent testing, suggests a prospective collection and testing of new specimens, though the document does not explicitly state "prospective study."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not explicitly specify the number of experts used or their qualifications for establishing the "ground truth" for the clinical test set. The ground truth was established by:

  • "routine culture, defined as the selective enrichment of specimen in LIM Broth, followed by subculture to Blood Agar and confirmed by biochemical testing."
  • Organisms that grew on Blood Agar were confirmed using "gram-stain, catalase test, and latex agglutination."
  • For discrepant analysis, isolates were sent back to Hardy Diagnostics for confirmation, where their identity was confirmed (β GBS, NH GBS, or non-GBS).

This suggests standard microbiology laboratory procedures were followed, implying trained microbiologists or laboratory technicians were involved in establishing the ground truth.

4. Adjudication Method for the Test Set

The data indicates an adjudication process for discrepant isolates.

  • "All discrepant isolates were frozen in CryoSavers™ with Brucella Broth and returned to Hardy Diagnostics for testing."
  • "The identity of each isolate was confirmed (β Group B Streptococi, NH Group B Streptococi, or non-Group B Streptococi)."
  • "Once the identity was confirmed, positive organisms (β Group B Streptococi or NH Group B Streptococi) were tested at LoD (10^0 CFU/mL) in donated negative-vaginal rectal matrix for their recovery from the LIM reference method, color development in Carrot Broth™ Kit, and recovery from the Carrot Broth™ Kit to Blood Agar System."
  • Footnotes 1 and 2 on page 5 further detail the re-testing and confirmation process for false positives and false negatives based on the initial comparison.

This represents a form of independent adjudication performed by Hardy Diagnostics on all discrepant samples to determine the true status of the sample.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not conducted. This device is a culture medium, not an AI-assisted diagnostic tool interpreted by human readers. The clinical study compares the performance of the new culture medium (Strep B Carrot Broth™ Kit) against a predicate culture medium (LIM Broth) and standard microbiological confirmatory tests.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

The device is a standalone diagnostic tool in the sense that its primary output (color change and/or subculture) is directly interpreted. There is no "algorithm only" performance as it's a biochemical reaction in a broth medium. The interpretation of the color change is visual, and subsequent subculture and biochemical testing for negatives are laboratory procedures. Its performance as a standalone culture medium for GBS detection is what was evaluated.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

The ground truth for the clinical performance study was established by standard microbiological culture methods and confirmatory tests. Specifically:

  • Selective enrichment in LIM Broth.
  • Subculture to Blood Agar.
  • Confirmation by biochemical testing (gram-stain, catalase test, and latex agglutination).
  • For discrepant analysis, full re-testing and confirmation of isolates at Hardy Diagnostics.

This represents a robust, laboratory-based "reference standard" or "gold standard" for GBS detection.

8. The Sample Size for the Training Set

This type of product (a culture medium) does not typically involve a "training set" in the machine learning sense. Its performance is based on its inherent biochemical and selective properties. Therefore, there is no training set for this device.

9. How the Ground Truth for the Training Set Was Established

As there is no training set for a culture medium, this question is not applicable. The "ground truth" principles applied to its development would stem from established scientific principles of bacteriology, media formulation, and GBS characteristics rather than a data-driven training process.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a caduceus, a symbol often associated with healthcare, featuring a staff with a serpent entwined around it.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 16, 2017

HARDY DIAGNOSTICS RIANNA MALHERBE LEAD PERFORMANCE STUDIES MICROBIOLOGIST 1430 WEST MCCOY LANE SANTA MARIA CA 93455

Re: K170481

Trade/Device Name: Strep B Carrot Broth Kit Regulation Number: 21 CFR 866.2360 Regulation Name: Selective culture medium Regulatory Class: I Product Code: POZ Dated: January 20, 2017 Received: February 16, 2017

Dear Ms. Malherbe:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Ribhi Shawar -S

For Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170481

Device Name Strep B Carrot Broth™ Kit

Indications for Use (Describe)

Strep B Carrot Broth™ Kit is a selective and differential medium which is intended for the detection of Group B Streptococus (GBS) from anovaginal specimens collected from pregnant women. The medium is used as an aid in the qualitative determination of GBS colonization in pregnant women. The color change reaction from white to orange is representative of a positive result for presence of GBS. The medium requires 24 hours of incubation but positive results can be interpreted and reported as early as 16 hours. Due to the properties of Strep B Carrot Broth™ Kit, non-hemolytic GBS cannot be detected by the medium's color change and require subculture for identification. Any presumptive negative indicated by lack of color change at the end of the incubation period must be subcultured to a non-selective medium (e.g., Tryptic Sov Agar with 5% Sheep Blood) to confirm absence of GBS. Subculture must also be performed to recover isolates for conducting susceptibility testing as recommended for penicillin-allergic women.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Appendix D - 510(k) Summary

I. SUBMITTER

Rianna Malherbe Lead Performance Studies Microbiologist Hardy Diagnostics 1430 W. McCoy Lane Santa Maria, CA 93455 Phone: 805-346-2766 x 5714 E-mail: MalherbeR@hardydiagnostics.com

II. DEVICE

Name of Device: Strep B Carrot Broth™ Kit Classification Name: Selective Culture Medium Regulatory Class: I Product Code: PQZ

III. PREDICATE DEVICE

LIM Broth, K871447

IV. DEVICE DESCRIPTION

Approximately 10-35% of women are asymptomatic carriers of group B streptococci (GBS) in the genital and gastrointestinal tracts. Group B Streptococi (GBS) remains a leading cause of serious illness and death in newborn populations and, therefore, the detection of Group B Streptococci in the vaginal-anorectal area is critical to the prevention of neonatal GBS disease. The Centers for Disease Control and Prevention (CDC) recommends the screening of all pregnant women for vaginal and rectal Group B Streptococci colonization between 35 and 37 weeks of gestation using an enrichment broth followed by subculture.

Strep B Carrot Broth™ is a selective and differential medium with selective components designed to enrich for Group B Streptococci. The production of a light orange, or red-orange pigment is a unique characteristic of β-hemolytic GBS due to reaction with substrates such as starch, and folate pathway inhibitors. GBS detection by color with Strep B Carrot Broth™ is only possible with ß-hemolytic Group B Streptococci colonies, which provides evidence of a direct genetic linkage between pigment production in this media and hemolysin production. Non-hemolytic GBS can be recovered Strep B Carrot Broth™ upon subculture to 5% sheep blood agar plates.

V. INDICATIONS FOR USE

Strep B Carrot Broth™ Kit is a selective and differential medium which is intended for the detection of Group B Streptoccus (GBS) from anovaginal specimens collected from pregnant women. The medium is used as an aid in the qualitative determination of GBS colonization in pregnant women. The color change reaction from white to orange is representative of a positive result for presence of GBS. The medium requires 24 hours of incubation but positive results can be interpreted and reported as early as 16 hours. Due to the properties of Strep B Carrot Broth™ Kit, non-hemolytic GBS cannot be detected by the medium's color change and require subculture for identification. Any presumptive negative indicated by lack of color change at the incubation period must be subcultured to a non-selective medium (e.g., Tryptic Soy Agar with 5% Sheep Blood) to confirm absence of GBS. Subculture must also be performed to recover isolates for conducting susceptibility testing as recommended for penicillin-allergic women.

{4}------------------------------------------------

AttributeDeviceComparatorSubstantiallyEquivalent?
NameStrep B Carrot Broth™ KitLIM Broth
510(k) Details510(k) number K170481Product Code PQZ21 CFR 866.2360"Culture media, selective and differential"Class IPanel 83 Microbiology510(k) number K871447Product Code JSD21 CFR 866.2360"Culture Media, Selective Broth"Class IPanel 83 MicrobiologyYes
Intended UseStrep B Carrot Broth™ Kit is a selective anddifferential medium which is intended for thedetection of Group B Streptococcus (GBS) fromanovaginal specimens collected from pregnantwomen. The medium is used as an aid in thequalitative determination of GBS colonizationin pregnant women. The color change reactionfrom white to orange is representative of apositive result for presence of GBS. Themedium requires 24 hours of incubation butpositive results can be interpreted andreported as early as 16 hours. Due to theproperties of Strep B Carrot Broth™ Kit, non-hemolytic GBS cannot be detected by themedium's color change and require subculturefor identification. Any presumptive negativeindicated by lack of color change at the end ofthe incubation period must be subcultured toa non-selective medium (e.g., Tryptic Soy Agarwith 5% Sheep Blood) to confirm absence ofGBS. Subculture must also be performed torecover isolates for conducting susceptibilitytesting as recommended for penicillin-allergicwomen.Remel Todd Hewitt Broth w/ CNA (LimBroth) is a liquid medium recommended foruse in qualitative procedures for theisolation of Group B Streptococcus (GBS)from clinical specimens containing mixedbacterial flora.Yes
MethodologyEnrichment Broth, ChromogenicEnrichment BrothYes
InoculationDirectDirectYes
Sample TypeVaginal/rectal swabVaginal/rectal swabYes
InterpretationManual/visual and subculture negativesManual/visual and subcultureYes

vi. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

{5}------------------------------------------------

VII. PERFORMANCE DATA

Performance of Strep B Carrot Broth™ was evaluated at four geographically diverse hospitals with routine GBS specimen in the form of an anovaginal swab. The detection of Group B Streptocci by orange color development in Strep B Carrot Broth™ Kit was compared to routine culture, defined as the selective enrichment of specimen in LIM Broth, followed by subculture to Blood Agar and confirmed by biochemical testing. Additionally, the recovery of Group B Streptococi from Strep B Carrot Broth™ that been subsequently subcultured to Blood Agar was also compared to LIM broth routine culture. Organisms that grew on Blood Agar were confirmed to be Group B Streptococci using gram-stain, catalase test, and latex agglutination.

A total of 884 specimens were tested against routine culture, 113 specimens did not meet enrollment criteria, and were therefore excluded from the analysis. Of the remaining 771 valid samples tested, a total of 143 specimens were positive for Group B Streptococi by orange color development in Strep B Carrot Broth™ after 24 hours of incubation at 35-37ºC with concordant results obtained from the LIM reference method. Those results are shown in Table 1.

All discrepant isolates were frozen in CryoSavers™ with Brucella Broth and returned to Hardy Diagnostics for testing. The identity of each isolate was confirmed (β Group B Streptococci, NH Group B Streptococci, or non-Group B Streptococci). Once the identity was confirmed, positive organisms (β Group B Streptococci or NH Group B Streptococi) were tested at LoD (10° CFU/mL) in donated negative-vaginal rectal matrix for their recovery from the LIM reference method, color development in Carrot Broth™ Kit, and recovery from the Carrot Broth™ Kit to Blood Agar System.

SiteTPFP1FN2TNSensitivity95% CISpecificity95% CI
CCP471414192.281.596.999.396.199.9
NY343913779.164.888.697.993.999.3
CC26028392.977.498.0100.095.6100.0
MCW363524087.874.594.798.896.499.6
Overall14372060187.7381.891.998.8397.699.4

Table 1. LIM Reference Method vs. Strep B Carrot Broth™ Kit Color reaction

There were 7 False Positives observed. All isolates were re-tested and confirmed at Hardy Diagnostics using the discrepant analysis protocol described above. Three isolates were originally negative by LIM reference method, showed a positive color reaction in Carrot Broth, but no ß-hemolytic colonies were recovered upon subculture to Blood Agar. Of these three, both Blood Agar plates from one sample (from LIM and CB subcultured plates) were swarmed with Proteus and no β-hemolytic colonies could be isolated for confirmation. No ß-hemolytic colonies were recovered from the frozen samples of the other two isolates, and therefore could not be confirmed. The remaining four discrepant isolates were confirmed to be ß Group B Streptococci.

² There were 20 False Negatives observed. All isolates were re-tested and confirmed at Hardy Diagnostics using the discrepant analysis protocol described above. Fourteen of the β Group B Streptococi isolates recovered from LIM, but originally gave a negative Carrot Broth color reaction, were confirmed as β Group B Streptococi and subsequently confirmed to have a positive Strep B Carrot Broth™ Kit color reaction at LoD. Two isolates were identified as a very weak β Group B Streptococi and did not produce the expected color reaction in Strep B Carrot Broth™ Kit. Four isolates were confirmed as non-hemolytic Group B Streptococci with a negative color reaction in Strep B Carrot Broth™.

3 considering that non-hemolytic GBS cannot be detected by the medium's color change and require subculture for identification, there were seven specimens that were found to be non-hemolytic upon subculture and identification by the reference method. If these specimens are included as negatives, then the overall sensitivity and specificity values observed when comparing the recovery of β-hemolytic GBS by the LIM reference method to the Strep B Carrot Broth™ Kit color reaction were 90.4% (141/156) [95% Cl: 84.7-94.1] and 98.5% (606/615) [95% Cl: 97.2-99.2], respectively.

{6}------------------------------------------------

When comparing the number of Group B Streptococi positive specimens recovered by the LIM reference method to the number identified by Strep B Carrot Broth™ color change in conjunction with the subculture of presumptive negatives to the Blood Agar, an additional 18 specimens showed concordant positive results for a total of 161 true positive results. The Lim reference method included the identification of both ß-hemolytic GBS from samples by culture. Those results are shown in Table 2.

SiteTPFP1FN2TNSensitivity95% CISpecificity95% CI
CCP501114198.089.799.799.396.199.9
NY4370133100.091.8100.095.090.097.6
CC280083100.087.9100.0100.095.6100.0
MCW405123897.687.499.697.995.399.1
Overall16113259598.895.699.797.996.498.7

Table 2. LIM Reference Method vs. Strep B Carrot Broth™ Kit (color), plus Subculture of Presumptive Negatives to Blood Agar with Biochemical Testing

' There were thirteen False Positives observed. All isolates were re-tested and confirmed at Hardy Diagnostics using the discrepant analysis protocol described above. All isolates recovered from the Strep B Carrot Broth to Blood Agar system were confirmed to be ß Group B Streptococci.

2 There were two False Negatives observed. Both isolates were re-tested and confirmed at Hardy Diagnostics using the discrepant analysis protocol described above. Both of the ß Group B Streptococi isolates recovered from LIM were confirmed as β Group B Streptococci.

RECOVERY RATE

To determine the recovery (Limit of Detection (LoD)) of Strep B Carrot Broth™, the media was challenged with two beta-hemolytic ATCC strains of Group B Streptococi at 10-fold decreasing concentrations and evaluated for color change. The lowest concentration at which a positive reaction was seen, indicated by an orange color, was determined to be the LoD. The determined by testing Strep B Carrot Broth™ with 20 replicate dilutions of the determined LoD concentrations. Strep B Carrot Broth™ was able to recover both S. agalactiae ATCC 12386 and S. agalactiae ATCC 12403 at a LoD of 10' CFU/mL of the fluid from a flocked anovaginal swab specimen (30 CFU/tube when using a 30μL inoculum). Variable recovery was seen at lower concentrations. Blood Agar plates were used to determine the concentration of organisms present in each dilution.

ANALYTICAL REACTIVITY

Fifty-four ATCC reference and clinical Group B Streptococci strains representing seven of the nine known serotypes were recovered in Strep B Carrot Broth™ when inoculated at the limit of detection concentration of 10° CFU/mL. The GBS serotypes included in this study were serotypes la, Ib, III, IV, V, and VI. Four strains that were nontypable against the nine known serotypes were also included. All the beta-hemolytic strains of Group B Streptococi produced the expected orange color in Strep B Carrot Broth™. The non-hemolytic strains showed no orange color in Strep B Carrot Broth™ but were successfully recovered upon subculture to Blood Agar.

ANALYTICAL SPECIFICITY

Seventy-eight non-target organisms that are phylogenetically-related to Group B Streptococci or potentially encountered in an anovaginal swab were tested in Strep B Carrot Broth™ at a concentration of 10° CFU/mL. All

{7}------------------------------------------------

organisms tested are listed in Table 3 below. After 24 hours of incubation, all Carrot Broth tubes were evaluated for color reaction. In order to determine if Strep B Carrot Broth™ supported the growth of non-target organisms in the absence of a color reaction, all negative tubes were subcultured to an appropriate medium for the non-target organism. All organisms tested produced a negative color reaction in Strep B Carrot Broth™ and 48/78 (61.5%) of the organisms were recoverable when subcultured after enrichment.

OrganismOrganismOrganism
Acinetobacter baumanniiEnterococcus duransProteus mirabilis
Aeromonas hydrophilaEnterococcus faecalisProvidencia alcalifaciens
Aspergillus brasiliensisEnterococcus faeciumPseudomonas aeruginosa
Bacillus cereusEnterococcus flavescensPseudomonas fluorescens
Bacillus subtilisEnterococcus hiraeSaccharomyces cerevisiae
Bacteroides fragilisEnterococcus raffinosusSalmonella enterica (typhii)
Bifidobacterium breveEnterococcus saccharolyticusSalmonella enterica arizonae
Campylobacter coliEscherichia coliSerratia marcescens
Campylobacter jejuniGardnerella vaginalisShigella boydii
Candida albicansGeotrichum candidumShigella flexneri
Candida glabrataHafnia alveiShigella sonnei
Candida parapsilosisKlebsiella oxytocaStaphylococcus aureus
Candida tropicalisKlebsiella pneumoniaeStaphylococcus epidermidis
Citrobacter brakkiiLactobacillus acidophilusStaphylococcus saprophyticus
Citrobacter freundiiLactobacillus gasseriStenotroph. maltophilia
Citrobacter koseriLactobacillus leichmanniiStreptococcus mutans
Clostridium difficileLactococcus lactisStreptococcus anginosus
Clostridium novyiLegionella pneumophilaStreptococcus bovis
Clostridium perfringensListeria monocytogenesStreptococcus dysgalactiae
Clostridium sporogenesMicrococcus luteusStreptococcus mitis
Corynebacterium jekeiumMoraxella catarrhalisStreptococcus pneumoniae
Enterobacter aerogenesMorganella morganiiStreptococcus pyogenes
Enterobacter cloacaeNeisseria gonorrhoeaeStreptococcus salivarius
Enterococcus casseliflavusPediococcus acidilactiStreptococcus uberis
Enterococcus cecorumPeptostreptococcus anaerobiusVibrio parahaemolyticus
Enterococcus disparPlesiomonas shigelloidesYersinia enterocolitica

Table 3. List of non-target organisms tested in Analytical Specificity

MICROBIAL INTERFERENCE

Strep B Carrot Broth™ was challenged to determine if target organisms at low concentration could be recovered in the presence of non-target organisms at a high concentration. All organisms that were recovered upon subculture from Strep B Carrot Broth™ in the Analytical Specificity study were used in this Microbial Interference study. Nontarget organisms at a high concentration (1.5 x 10° CFU/mL) were mixed with each target organism at the LoD and

{8}------------------------------------------------

inoculated into Strep B Carrot Broth™. If the target organism was not recovered, the concentration of the nontarget organism was lowered 10-fold until the target organism was recovered.

Strep B Carrot Broth™ was able to produce the expected color reaction with target organisms and allowed the recovery of both GBS strains in the presence of high concentrations (1.5 x 10° CFU/mL) of all but one of the nontarget organisms used in this study. The only organism found to affect recovery of Group B Streptococci was E. faecalis ATCC 29212. S. agalactiae ATCC 12386 produced the expected color reaction when the concentration of E. faecalis ATCC 29212 was 10 CFU/mL or lower. S. agalactiae ATCC 13813, a non-hemolytic strain, was recovered upon subculture when the concentration of E. faecalis ATCC 29212 was 10° CFU/mL or lower.

INTERFERENCE

Commonly used or encountered endogenous and exogenous substances that may be present in anovaginal swabs were evaluated for potential interference of growth or chromogenic reaction in Strep B Carrot Broth™. The substances tested are listed in the table below. No interference was observed with any substance at the highest clinically relevant concentration in the GBS-negative specimen matrix.

Interfering Substances
CategorySubstance/SupplierConcentration inSample Matrix*
Anti-diarrhealPepto-Bismol® (Bismuth subsalicylate solution)1% v/v
MedicationImodium A-D® (Loperamide HCl)2% w/v
Body OilNeutrogena Body Oil2% v/v
Body PowderGold Bond Body Powder1% w/v
Contraceptive GelOptions Gynol II® (Nonoxynol-9)0.59% w/v
Enema SolutionPhysiological saline0.25% v/v
Lubricating GelK-Y® Jelly0.57% w/v
Milk of Magnesia1.78% v/v
Oral LaxativeDulcolax® (Sodium picosulfate solution)1% w/v
Polysorbate 80Tween®8010% v/v
Rectal LaxativeFleet® Glycerin Suppositories10% v/v
Topical HemorrhoidOintmentPreparation-H®0.26% w/v
Vaginal Anti-ItchMedicationVagisil® Cream0.41% w/v
Vaginal Anti-FungalMonistat® (Miconazole nitrate)0.29% w/v
MedicationLotrimin® (Clotrimazole)0.29% w/v
Endogenous Substances
Human Amniotic FluidMedfusion2% v/v
Human FecesCentral Coast Pathology2% v/v
Human MeconiumLEE Biosolutions2% v/v
Human UrineCentral Coast Pathology2% v/v
Human Whole BloodIn-house2% v/v
MucinSigma, M23780.05% w/v

'Specific amounts of substance added to anovaginal specimen matrix calculated using C.V-=C.V2 with the assumption that 1g=1mL.

INCUBATION STUDY

In order to determine a recommended incubation time range, the performance of Strep B Carrot Broth™ was evaluated using nine beta-hemolytic strains of GBS and one non-hemolytic strain of GBS at the LoD from 12 to 24

{9}------------------------------------------------

hours at 35°C. The enrichment broth was subcultured to a Tryptic Soy Agar plate with 5% sheep blood every two hours to confirm the presence of GBS. All hemolytic organisms tested produced some kind of orange color reaction by 12 hours. All organisms, including the non-hemolytic strain tested, were recovered upon subculture of Strep B Carrot Broth™ at 12 hours. The incubation range for Strep B Carrot Broth™ was set from 16-24 hours.

SPECIMEN STABILITY

Various types of specimen transport swabs were evaluated to determine the acceptable storage conditions required to recover GBS from Strep B Carrot Broth™. Swabs were spiked with Group B Streptococci and a contrived matrix consisting of organisms commonly found in vaginal flora, kept at both room temperature and refrigerated conditions, and were inoculated to Carrot Broth at 0, 24, 48, 72, 96, and 120 hours. Eight different GBS strains were used in this study and were spiked into the contrived matrix near LoD. The contrived matrix containing nontarget organisms consisted of E. faecalis, E. coli, C. albicans, and L. acidophilus. TransPRO™ swabs with Liquid Amies (flocked swab liquid-based transport system) and four types of Healthlink transport systems: Sponge-based Liquid Amies and Liquid Stuart's, and Gel-based: Amies Gel and Stuart's Gel were used in this study.

Strep B Carrot Broth™ was able to recover 8/8 (100%) of GBS strains and produce orange coloration from Healthlink swabs in Liquid Amies Stuart's liquid, Amies Gel, and Stuart's Gel when stored at 2-8°C for up to 96 hours. 100% of GBS strains also produced the expected orange color reaction from the flocked swab TransPRO™ Liquid Amies stored at 2-8°C for up to 120 hours. All transport systems tested saw a decline in recovery of GBS after 24 hours of room temperature storage.

REPRODUCIBILITY

Prior to initiating the study, a panel of 12 blinded isolates provided by Hardy Diagnostics was tested at three distinct study sites in triplicate on five work days to demonstrate reproducibility and to document proficiency in the performance of the test. Agreement of >95% with known test results was required before proceeding with the study. The testing was done with at least one operator and two readers, blinded to each other's results, per site. Strains in the reproducibility panel produced the expected color results with Strep B Carrot Broth™ > 95% of the time at 24 hours. All non-hemolytic GBS isolates tested (100%) were recovered upon subculture to TSA with 5% Sheep's Blood.

CONCLUSIONS

The clinical and performance studies summarized above demonstrate the substantial equivalence of Strep B Carrot Broth™ Kit to the predicate device.

§ 866.2360 Selective culture medium.

(a)
Identification. A selective culture medium is a device that consists primarily of liquid or solid biological materials intended for medical purposes to cultivate and identify certain pathogenic microorganisms. The device contains one or more components that suppress the growth of certain microorganisms while either promoting or not affecting the growth of other microorganisms. The device aids in the diagnosis of disease caused by pathogenic microorganisms and also provides epidemiological information on these diseases.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 866.9.