K Number
K170462

Validate with FDA (Live)

Date Cleared
2017-05-16

(90 days)

Product Code
Regulation Number
882.1480
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The KARL STORZ ShuntScope is indicated to provide visualization and access during cranial diagnostic and therapeutic procedures such as shunt placement, tumor biopsy and resection, hydrocephalus treatment, endoscopic third ventriculostomy with choroid plexus cauterization (ETV/CPC), endoscopic third ventriculostomy, cyst fenestration, and aqueduct exploration in pediatrics and adult population.

Device Description

The components subject of this submission are: (1) the KARL STORZ ShuntScope (Model Number: 28164SSA), (2) the examination sheath (Model Number: 11577KA) and (3) operating sheath (Model Number: 11577KE)

AI/ML Overview

The KARL STORZ ShuntScope is indicated to provide visualization and access during cranial diagnostic and therapeutic procedures. This response pertains to the non-clinical performance data for the device. Clinical studies were not required or performed to demonstrate substantial equivalence to predicate devices, as non-clinical bench testing was deemed sufficient.

1. Table of Acceptance Criteria and Reported Device Performance:

Verification TestAcceptance CriterionReported Performance
Surface and EdgesThe surface of the ShuntScope shall be free of pores, cracks, and tooling agent remainders.Pass
Maximum insertion portion widthThe maximum insertion portion width shall not be larger than 1 mm, as stated in the instruction manual.Pass
Field of ViewThe deviation of field of view shall not be greater than 15%.Pass
Direction of ViewThe deviation of direction of view shall not be greater than 0° ± 10°.Pass
ResolutionCharacterization only. (No specific numerical acceptance criterion stated, but performance was assessed.)Pass

2. Sample Size Used for the Test Set and Data Provenance:

The document does not explicitly state the sample size (number of ShuntScope units) used for each bench test. The term "bench testing" refers to in-vitro testing, suggesting the tests were performed on manufactured devices. The provenance of the data is from KARL STORZ Endoscopy-America, Inc., and these tests are inherently prospective as they are performed to verify the design of the new device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

Not applicable. This is a non-clinical bench testing study assessing physical and optical characteristics of a device, not a diagnostic or clinical performance study requiring expert interpretation or ground truth establishment in a clinical context. The "ground truth" here is adherence to manufacturing specifications and design requirements.

4. Adjudication Method for the Test Set:

Not applicable. As this is non-clinical bench testing, there is no adjudication method in the context of clinical interpretation or diagnosis. The "Pass" results indicate that the device met its pre-defined engineering and performance specifications.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

No. An MRMC comparative effectiveness study was not done. This study focuses on the technical performance of the device itself (e.g., its physical dimensions, optical properties) rather than its effectiveness in a clinical setting with human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

Not applicable. This device is a neurological endoscope, which is a physical instrument for visualization and access, not an algorithm or AI system. Therefore, standalone algorithm performance is not relevant.

7. The Type of Ground Truth Used:

The ground truth used for these bench tests consisted of pre-defined technical specifications and engineering requirements. For instance, the "Maximum insertion portion width" has a ground truth of "not larger than 1 mm," and the "Direction of view" has a ground truth of "0° ± 10°." These are objective, measurable criteria established during the device's design and manufacturing process.

8. The Sample Size for the Training Set:

Not applicable. This is not a machine learning or AI-based device, so there is no concept of a "training set" in the context of this performance evaluation.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, creating a sense of depth and unity.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 16, 2017

Karl Storz Endoscopy-America, Inc. Leigh Spotten Regulatory Director 2151 E. Grand Avenue El Segundo, California 90245

Re: K170462

Trade/Device Name: KARL STORZ ShuntScope Regulation Number: 21 CFR 882.1480 Regulation Name: Neurological Endoscope Regulatory Class: Class II Product Code: GWG Dated: February 14, 2017 Received: February 15, 2017

Dear Leigh Spotten:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Michael 原田offmann -S

for

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170462

Device Name KARL STORZ ShuntScope

Indications for Use (Describe)

The KARL STORZ ShuntScope is indicated to provide visualization and access during cranial diagnostic and therapeutic procedures such as shunt placement, tumor biopsy and resection, hydrocephalus treatment, endoscopic third ventriculostomy with choroid plexus cauterization (ETV/CPC), endoscopic third ventriculostomy, cyst fenestration, and aqueduct exploration in pediatrics and adult population.

Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The word "STORZ" is written in large, bold, blue letters. Below that, the words "KARL STORZ - ENDOSKOPE" are written in smaller, blue letters. The logo is simple and professional.

510(k) Summary

This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.

Submitter:KARL STORZ Endoscopy-America, Inc.2151 E. Grand AvenueEI Segundo, CA 90245
Contact:Nozomi YagiRegulatory Affairs SpecialistPhone: (424) 218-8351Fax: (424) 218-8519
Date ofPreparation:May 15, 2017
Type of 510(k)Submission:Traditional
DeviceIdentification:Trade Name: KARL STORZ ShuntScopeClassification Name: Endoscope, Neurological (21 CFR Part 882.1480)
RegulatoryClass:II
Product Code:GWG
GuidanceDocument:Not Applicable for GWG product code
RecognizedConsensusStandards:Not Applicable for GWG product code
PredicateDevice(s):Primary Predicate Device: Medtronic PS Medical's NeuroPEN Endoscopeand Optical Accessories (K003914)Secondary Predicate Device: KARL STORZ Endoscopy-America's Neuro-Fiberscope (K002788)Reference Device: KARL STORZ Endoscopy-America's KSEA
Sialendoscopes and Accessories (K012527)
DeviceDescription:The components subject of this submission are: (1) the KARL STORZShuntScope (Model Number: 28164SSA), (2) the examination sheath(Model Number: 11577KA) and (3) operating sheath (Model Number:11577KE)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The word "STORZ" is in large, bold, blue letters. Below that, in smaller blue letters, is the text "KARL STORZ - ENDOSKOPE".

Intended Use:The KARL STORZ ShuntScope is intended for visualization purposes during cranial diagnostic and therapeutic procedures.
Indications For Use:The KARL STORZ ShuntScope is indicated to provide visualization and access during cranial diagnostic and therapeutic procedures such as shunt placement, tumor biopsy and resection, hydrocephalus treatment, endoscopic third ventriculostomy with choroid plexus cauterization (ETV/CPC), endoscopic third ventriculostomy, cyst fenestration, and aqueduct exploration in pediatrics and adult population.
Technological Characteristics:Comparison Table: Subject vs. Primary and Secondary Predicate Devices
Subject DevicePrimary Predicate Device, K003914Secondary Predicate Device, K002788Reference Device, K012527
KARL STORZ ShuntscopeMedtronic NeuroPENKSEA Neuro-FiberscopeMarchal Mini Sialendoscope 0°
Physical Characteristics
Type of ScopeSame as the subject deviceFlexibleSame as the subject device
Insertion Shaft Diameter1 mm (wo sheath)1.3 mm (w 11577KA)N/A1.14 mmN/ASame as the subject device
2.6 mm (w 11577KE)3.7 mmSame as the subject device
Insertion Shaft Length161.8 mm155 or 190 mm340 mmSame as the subject device
Working Channel Diameter1.3 mm (w 11577KE)N/A, No working channel1.2 mmSame as the subject device
Optical Characteristics
Type of ImagerFiberopticSame as the subject deviceSame as the subject device
Direction of ViewInformation not publicly availableSame as the subject device
Pixel Fibers10,000Same as the subject device7,000Same as the subject device
Light SourceExternalSame as the subject deviceSame as the subject device
Cleaning and Sterilization Methods
CleaningManualSingle-useSame as the subject device
SterilizationSteamSterileSame as the subject device
Non-Clinical Performance Data:Bench Testing Summary
Verification TestTest Method Summary
Surface and EdgesThe surface of the ShuntScope shall be free of pores, cracks, and remainders of tooling agents.
Maximum insertion portion widthThe maximum insertion portion width shall not be larger than 1 mm, stated in the instruction manual.
Field of viewThe deviation of field of view shall not be greater than 15%.
Direction of viewThe deviation of direction of view shall not be greater than 0° ± 10°.
ResolutionCharacterization only.
Result
Pass
Pass
Pass
Pass
Pass

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for STORZ. The word "STORZ" is written in large, bold, blue letters. Below that, the words "KARL STORZ - ENDOSKOPE" are written in smaller, blue letters.

Biocompatibility SummaryThe biocompatibility evaluation for the patient contacting components of the neuroscopewas performed according to ISO 10993-1 and FDA Guidance. The following tests wereconducted based contact type and duration:Cytotoxicity (ISO 10993-5) Sensitization (ISO 10993-11) Irritation (ISO 10993-11) Systemic Toxicity (ISO 10993-10) The biological-toxicological safety of the patient is not affected by the materials used in thetested items.
Sterilization Validation Summary
TestTest Method SummaryResult
SteamSAL=10-6, Half cycle methodPass
BibliographyThe following published literature was provided to support the pediatric useof the subject device:1. Kulkarni, Abhaya V., et al. "International Infant Hydrocephalus Study: initial resultsof a prospective, multicenter comparison of endoscopic third ventriculostomy (ETV)and shunt for infant hydrocephalus." Child's Nervous System 32.6 (2016): 1039-1048.2. Depreitere, Bart, et al. "Endoscopic biopsy for intraventricular tumors in children."Journal of Neurosurgery: Pediatrics 106.5 (2007): 340-346.3. Stone, Scellig SD, and Benjamin C. Warf. "Combined endoscopic thirdventriculostomy and choroid plexus cauterization as primary treatment for infanthydrocephalus: a prospective North American series: Clinical article." Journal ofNeurosurgery: Pediatrics 14.5 (2014): 439-446.4. Zada, Gabriel, et al. "Pathogenesis and treatment of intracranial arachnoid cysts inpediatric patients younger than 2 years of age." Neurosurgical focus 22.2 (2007): 1-5.
ClinicalPerformanceData:Clinical testing was not required to demonstrate the substantial equivalenceto the predicate devices. Non-clinical bench testing was sufficient toestablish the substantial equivalence of the modifications.
Conclusion:The conclusions drawn from the nonclinical tests demonstrate that thesubject device, KARL STORZ ShuntScope performs as well as or better thanthe legally marketed predicate devices.

§ 882.1480 Neurological endoscope.

(a)
Identification. A neurological endoscope is an instrument with a light source used to view the inside of the ventricles of the brain.(b)
Classification. Class II (performance standards).