K Number
K170451

Validate with FDA (Live)

Date Cleared
2017-03-16

(29 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The FDX Console is a workstation intended to associate digital (CR and/or DR) images with patient and exam information, apply image processing to facilitate diagnosis, display the image, and output the resulting image and exam data for further display, distribution, or archiving. The FDX Console is not for use in Mammography.

Device Description

The FDX Console is used by a radiographer for viewing Digital Radiography (DR) and Computed Radiography (CR) images for final quality assurance (QA) checking and image processing and optimization prior to transferring the images to external devices such as a PACS or a printer. Furthermore, when connected to Fuji Image Readers via a network, the FDX Console is used to enter patient ID information, exposure information, and in the case of CR register Image Plate (IP) barcode numbers. The FDX Console is compatible with the ACR/NEMA DICOM Version 3.0 standard. The four primary features of the FDX Console are patient identification, exam selection, image processing, and image transmission.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the FDX Console (DR-ID300CL) Software. The submission aims to demonstrate substantial equivalence to legally marketed predicate devices, not primarily to prove performance against specific acceptance criteria through clinical studies. Therefore, much of the requested information regarding acceptance criteria, specific device performance metrics, sample sizes for test sets, expert ground truth establishment, adjudication methods, MRMC studies, or standalone performance studies is explicitly stated as "not required" or not detailed in the document in the context of device performance evaluation.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not present acceptance criteria in terms of specific performance metrics (e.g., sensitivity, specificity, accuracy) for diagnosis or detection by the FDX Console software. This is because the device is a medical image processing workstation, not a diagnostic AI algorithm that directly interprets images for disease. The "acceptance criteria" discussed are related to conformance to voluntary standards and demonstrability of substantial equivalence.

Acceptance Criteria (Not in terms of performance metrics)Reported Device Performance and Compliance
Conformance to DICOM, V3.0 2007Conforms
Conformance to IEC 62304: 2006 (Medical Device Software – Software Life Cycle Processes)Conforms
Conformance to IEC 62366: 2014 (Medical devices – Application of usability engineering to medical devices)Conforms
All verification and validation activities performed as required by risk analysisResults were satisfactory
Software modifications do not alter technological characteristics of predicatesDemonstrated (by comparison table)

2. Sample Size for the Test Set and Data Provenance

Not Applicable (N/A) for a medical image processing workstation. The submission indicates that "Clinical studies were not required as there is no change in image processing." The focus was on non-clinical performance data (software V&V and conformance to standards) to demonstrate substantial equivalence to existing devices. There is no mention of a "test set" of medical images for evaluating diagnostic performance.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

N/A. As no clinical studies or test sets for diagnostic performance were conducted, no experts were used to establish ground truth for such a purpose.

4. Adjudication Method for the Test Set

N/A. No test set for diagnostic performance was utilized.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. The document explicitly states: "Clinical studies were not required as there is no change in image processing." Therefore, no MRMC study comparing human readers with and without AI assistance was performed or reported.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

No. The FDX Console is described as a workstation that "applies image processing to facilitate diagnosis, display the image, and output the resulting image and exam data," meaning it is an assistive tool within a human workflow, not a standalone diagnostic algorithm. No standalone performance study was conducted or reported.

7. Type of Ground Truth Used

N/A. Since no clinical studies were performed to evaluate diagnostic performance against a ground truth, this is not applicable. The device's function is image processing and management, not direct diagnostic interpretation.

8. Sample Size for the Training Set

N/A. The FDX Console is an image processing software with fixed algorithms. It is not an AI/ML device that requires "training data" in the typical sense of machine learning for diagnostic tasks. Its function is to apply established image processing techniques.

9. How the Ground Truth for the Training Set Was Established

N/A. As the device does not involve machine learning requiring a training set with established ground truth, this information is not applicable. The software's algorithms are based on predefined image processing techniques.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other, with flowing lines extending from the bottom profile.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 16, 2017

FUJIFILM Medical Systems USA, Inc. % Mr. Peter Altman Regulatory and Quality Consultant 419 West Avenue STAMFORD CT 06902

Re: K170451

Trade/Device Name: FDX Console (DR-ID300CL) Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 8, 2017 Received: February 15, 2017

Dear Mr. Altman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael O'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K170451

Device Name FDX Console (DR-ID300CL) Software

Indications for Use (Describe)

The FDX Console is a workstation intended to associate digital (CR and/or DR) images with patient and exam information, apply image processing to facilitate diagnosis, display the image, and output the resulting image and exam data for further display, distribution, or archiving. The FDX Console is not for use in Mammography.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE = CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Fujifilm logo. The logo consists of the word "FUJIFILM" in a bold, sans-serif font, with a red square above the "I" in "FUJIFILM". Below the logo is the tagline "Value from Innovation" in a smaller, sans-serif font. The logo is simple and modern, and the tagline conveys the company's commitment to providing value through innovation.

510(k) Summary

FDX Console (DR-ID300CL) Software

Date: February 8, 2017

Submitter's Information:

FUJIFILM Medical Systems U.S.A., Inc. 419 West Avenue Stamford, CT, 06902, USA

Contact Person:

Name:Peter Altman
Title:Regulatory and Quality Consultant
Telephone:(203) 602-3576
Facsimile:(203) 602-3785

Identification of the Device:

Proprietary/Trade Name:FDX Console (DR-ID300CL) Software
Classification Name:Picture Archiving and Communications System
Regulations Number:21 CFR 892.2050
Product Codes:LLZ
Device Class:Class II
Review Panel:Radiology
Common Name:MedicalImageProcessingWorkstation or Acquisition Workstation

Identification of the Legally Marketed Predicate Devices:

    1. FDR D-EVO Flat Panel Detector System (DR-ID600) with Improved Virtual Grid Software, K153464 cleared 4/8/2016
    1. Fuji Medical CR Console, K041990, cleared 8/6/2004

DEVICE DESCRIPTION l.

The FDX Console is used by a radiographer for viewing Digital Radiography (DR) and Computed Radiography (CR) images for final quality assurance (QA) checking and image processing and optimization prior to transferring the images to external devices such as a PACS or a printer. Furthermore, when connected to Fuji Image Readers via a network, the FDX Console is used to enter patient ID information, exposure information, and in the case of CR register Image Plate (IP) barcode numbers. The FDX Console is compatible with the ACR/NEMA DICOM Version 3.0 standard.

The four primary features of the FDX Console are patient identification, exam selection, image processing, and image transmission.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Fujifilm logo. The logo consists of the word "FUJIFILM" in a bold, sans-serif font, with a red square above the "I" in "FUJIFILM". Below the logo is the tagline "Value from Innovation" in a smaller, sans-serif font. The logo is simple and modern, and the tagline conveys the company's commitment to innovation.

  • Patient Identification feature: Patient demographics can be manually . entered using a keyboard or captured from a RIS/HIS using DICOM Worklist Management. Also, a Touch screen monitor and barcode reader streamlines this data entry.
  • Exam Selection feature: Select the patient, select the applicable exam, for CR barcode the Imaging Plate (IP), and touch Start to select the exam or use the DICOM Worklist Management feature with Auto Exam Select.
  • . Image Processing feature: DR and CR Images are processed, displayed and reviewed prior to image transfer.
  • Image Transmission feature: Images are transferred to other medical imaging devices for softcopy viewing or storage devices for archiving or to a printer for hardcopy film output.

INDICATIONS FOR USE ..............................................................................................................................................................................

The FDX Console is a workstation intended to associate digital (CR and/or DR) images with patient and exam information, apply image processing to facilitate diagnosis, display the image, and output the resulting image and exam data for further display, distribution, or archiving. The FDX Console is not for use in Mammography.

SUMMARY OF TECHNOLOGICAL CHARACTERISTICS .

The FDX Console is a software device based upon the predicate devices. The modifications implemented in Version 10.0 do not alter the technological characteristics of the predicates, i.e., software is installed on off-the-shelf computer hardware running Microsoft Windows operating systems.

SUBSTANTIAL EQUIVALENCE IV.

The FDX Console Version 10.0 Software is substantially equivalent to the following legally marketed device:

Legally Marketed Device510(k) #Clearance Date
Fuji Medical CR Console Model Flash IIPK0419908/6/2004
FDR D-EVO FPD System (DR-ID600) (includes theFDX Console Version 9.0 software)K1534644/8/2016

The CR Console has been chosen as the predicate device because it is the FDX Console's predecessor and therefore the general design including software, GUI, features, and capabilities are extremely similar. One of the major differences is that the FDX Console has the capability to acquire and process images from all Fujifilm digital radiography system whereas the CR Console was limited to Computed Radiography only.

The FDX Console version 9.0 software was cleared with the FDR D-EVO FPD System (DR-ID600) in K153464. It has exactly the same image processing software as the subject Version 10.0 device. The only differences are that FDX Console V10.0 will be able to use the Microsoft Windows 10 operating

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Fujifilm logo. The logo is black, except for a small red square above the "I" in Fujifilm. Below the logo, the words "Value from Innovation" are written in a smaller font.

system and that this submission introduces the SmartController and RemoteDesktop software capabilities allowing remote operation of the FDX Console. The following comparison applies:

Subject DevicePredicate DevicePredicate Device
FDX Console (DR-ID300CL) SoftwareFDX Console (DR-ID300CL) With Virtual GridSoftware Optionfor the FDR D-EVO FPDsystem (DR-ID600)(K153464 cleared4/8/2016)Fuji CR Console (FlashIIP)(K041990 cleared8/6/2004)
Indicationsfor UseIndicationsfor use foroptionsThe FDX Console is aworkstation intended toassociate digital (CRand/or DR) images withpatient and examinformation, apply imageprocessing to facilitatediagnosis, display theimage, and output theresulting image andexam data for furtherdisplay, distribution, orarchiving. The FDXConsole is not for use inMammography.The Wireless/Wired FDR D-EVO flat panel detectorsystem is intended to capturefor display radiographicimages of human anatomy. Itis intended for use in generalprojection radiographicapplications includingpediatric and neonatal examswherever conventional film/screen or CR systems maybe used. The FDR D-EVO isnot intended formammography, fluoroscopy,tomography, and angiographyapplications.The FDR D-EVO can be usedwith Virtual Grid Software,which is optional imageprocessing software installedon Fujifilm's FDX Console.Virtual Grid Software can beused in lieu of an anti-scattergrid to improve imagecontrast in generalradiographic images byreducing the effects of scatterradiation.The Fuji CR Console is aCR modality workstationintended to associate digitalFCR images (exceptmammography images)with patient and examinformation, apply imageprocessing to facilitatediagnosis, display theimage, and output theresulting image and examdata for further display,distribution, or archiving.None.
None.
Image Processing Characteristics
SpatialFrequencyProcessingYesYesYes
TonalConversionYesYesYes
Dynamic RangeControlProcessing(DRC)YesYesYes
TomographicArtifactSuppressionProcessing(TAS)YesYesYes
GradationProcessingYesYesYes
Multi-ObjectiveFrequencyProcessingYesYesYes
Subject DevicePredicate DevicePredicate Device
FDX Console (DR-ID300CL) SoftwareFDX Console (DR-ID300CL) With Virtual GridSoftware Optionfor the FDR D-EVO FPDsystem (DR-ID600)(K153464 cleared4/8/2016)Fuji CR Console (FlashIIP)(K041990 cleared8/6/2004)
(MFP)Flexible Noise
Control (FNC)YesYesYes
Grid PatternRemoval (GPR)YesYesYes
ImageAutomaticCompositionProcessing(ImageStitching)YesYesYes
CoRrectionFilter (CRF)YesYesNo
ImageprocessingwithoutnormalizationYesYesNo
TomosynthesisOptionYesYesNo
Dual EnergySubtraction(DES) OptionYesYesNo
Virtual GridOptionYesYesNo
ImprovedVirtual GridSoftware (VG2)YesYesNo
Miscellaneous
WorkstationSoftwareVersion10.09.0A11
Transferimages toPACSYesYesYes
Enter Patientinfo manuallyYesYesYes
Enter Patientinfo from RISYesYesYes
Select ExamYesYesYes
MinimumBasicComputerConfiguration(as provided inUS)Model: HP ProDesk600 G2 Series SmallForm Factor PCProcessor: Intel Corei5-6500 with 3.2GHzRAM: 4GB (4x1GB)Hard Drive: 500GBCD/DVD: SlimSuperMulti DVD driveDisplay: 17 " or 21"color (touchscreen)Model: Mini TowerProcessor: Core 2 DuoRAM: 2GBHard Drive: 80GBCD/DVD: DVD driveDisplay: 17 " or 21" color(touchscreen optional)Monitor.Accessories: Keyboard,Mouse, Barcode scannerModel:DELL OptiplexGX260Processor: Pentium 4,3.06GHzHard Drive: 40GBRAM: 512MBDisplay: 17 " or 21" color(touchscreen optional)Monitor.
Subject DeviceFDX Console (DR-ID300CL) SoftwarePredicate DeviceFDX Console (DR-ID300CL) With Virtual GridSoftware Optionfor the FDR D-EVO FPDsystem (DR-ID600)(K153464 cleared4/8/2016)Predicate DeviceFuji CR Console (Flash IIP)(K041990 cleared8/6/2004)
optional) Monitor.Accessories:Keyboard, Mouse,Barcode scanner
OperatingSystemWindows 10/Windows7 Professional 32-bitWindows Vista/Windows 7Professional 32-bitWindows 2000Professional
ImageTransferSame as PredicatesStandard networkconnectivity via DICOMprotocol & via Fuji DMSNetworkStandard networkconnectivity via DICOMprotocol & via Fuji DMSNetwork
RemoteOperationCapability ofConsoleSmartController andRemote DesktopNoneNone

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the Fujifilm logo. The logo consists of the word "FUJIFILM" in bold, black letters, with a red square above the "I". Below the word is the tagline "Value from Innovation" in a smaller font. The logo is simple and modern, and the tagline suggests that Fujifilm is a company that values innovation.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the Fujifilm logo. The logo is black, except for a small red square above the "I" in Fujifilm. Below the logo, the words "Value from Innovation" are written in a smaller, black font.

V. SUMMARY OF STUDIES

Non-clinical Performance Data: The FDX Console conforms to the following voluntary standards:

StandardTitle
DICOM, V3.0 2007Digital Imaging and Communications inMedicine (DICOM)
IEC 62304: 2006Medical Device Software – Software LifeCycle Processes
IEC 62366: 2014Medical devices – Application of usabilityengineering to medical devices

As required by the risk analysis, all verification and validation activities related to the FDX Console were performed and the results were satisfactory. The verification and validation results are provided in the submission.

Clinical Performance Data: Clinical studies were not required as there is no change in image processing. The substantial equivalence has been demonstrated by nonclinical studies.

VI. CONCLUSION

Based upon the supporting data summarized above, we concluded the FDX Console (DR-ID300CL) Software is as safe and effective as the legally marketed devices K041990 andK153464 and does not raise different questions of safety and effectiveness than the predicate devices.

N/A