K Number
K170313

Validate with FDA (Live)

Date Cleared
2017-07-27

(176 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ARIX Ankle Distal Tibia System is intended for fixation of complex intra- and extraarticular fractures and osteotomies of the distal tibia including distal tibia fractures in combination with diaphyseal fracture.

Device Description

The ARIX Ankle Distal Tibia System is rigid fixation consisting of plates and screws in various configurations, shapes and sizes. The plates are Medial Type and Anterolateral Type and T Type. The screws are Cortical Screw (28-SO-L Series, 35-SO-L Series) and Locking Screw (28L-SO-L Series, 35L-SO-L Series). The system is made of ASTM F 136, Titanium Alloy (Ti-6Al-4V). The plates vary essentially through different lengths and number of plate holes. The screws are self-tapping, which are applied with the reconstruction locking screws together. The Cortical Screws & locking screws are provided with diameter 2.8 mm to 3.5 mm and lengths from 8 mm to 80 mm. The system is provided as Gold color which is casted according to the anodizing technique while apply 5 V electric energy to the titanium. It also includes various manual surgical instruments, such as drill bits, counter sink, depth gauge, compression, drill guide, drill sleeve, guide pin, bender and screw driver handle. The ARIX Ankle Distal Tibia System is not provided sterile.

AI/ML Overview

This document, K170313, describes the ARIX Ankle Distal Tibia System, a bone plate and screw system intended for fixation of intra- and extra-articular fractures and osteotomies of the distal tibia.

Here's an analysis of the acceptance criteria and supporting studies based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Bench Test Standard)Reported Device Performance (Reference)
Plates: - Bending strength test per ASTM F382-14 - Bending fatigue test per ASTM F382-14The test results for bending strength and bending fatigue for the ARIX Ankle Distal Tibia System indicate equivalence to the predicate device. (Page 6)
Screws: - Driving torque test per ASTM F543-13 - Torsion test per ASTM F543-13 - Axial pull out test per ASTM F543-13The test results for driving torque, torsion, and axial pull out for the ARIX Ankle Distal Tibia System indicate equivalence to the predicate device. (Page 6)
Biocompatibility: Meet ASTM F136 for Titanium Alloy (Ti-6Al-4V)Device material (ASTM F 136, Titanium Alloy) is widely used for surgical implants with well-known biocompatibility. (Page 5)
Sterilization: Achieve a SAL of 10^-6 via autoclave, validated per ISO 17665-1: 2006Sterilization method validated per ISO 17665-1: 2006. (Page 5)

2. Sample Size for Test Set and Data Provenance

  • Test Set Sample Size: Not explicitly stated for specific tests. The document refers to "bench tests" performed using the "proposed device" and "predicate device." It doesn't specify the number of plates or screws tested for each mechanical test.
  • Data Provenance: The study is a non-clinical bench test conducted to verify design specifications and compliance with ASTM standards. The provenance is internal testing performed by the manufacturer or a contracted lab. There is no mention of country of origin for the data itself, but the manufacturer is Jeil Medical Corporation in Korea. This is a retrospective analysis in the sense that the testing was performed to support the 510(k) submission, not as a prospective clinical trial.

3. Number of Experts and Qualifications for Ground Truth

  • Not Applicable. This submission is for a medical device (bone fixation system) and relies on non-clinical bench testing to demonstrate substantial equivalence to a predicate device, not on diagnostic accuracy against a clinical ground truth established by experts.

4. Adjudication Method

  • Not Applicable. As no expert review or clinical diagnosis is involved in the acceptance criteria, there is no adjudication method.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No. The document explicitly states: "No clinical studies were considered necessary and performed." This means no MRMC study was conducted.

6. Standalone Performance Study

  • Yes, a form of standalone performance was done for the device itself. The bench tests (bending strength, fatigue, driving torque, torsion, axial pull-out) assess the mechanical performance of the ARIX Ankle Distal Tibia System independently, comparing it to established standards and a predicate device. This is "standalone" in the sense that the mechanical behavior of the device components is evaluated without human interaction during the test, beyond setting up the experiment.

7. Type of Ground Truth Used

  • Standardized Mechanical Performance Metrics / Predicate Device Performance. The "ground truth" for the acceptance criteria and performance evaluation is based on:
    • Compliance with recognized international standards (ASTM F382-14 for plates, ASTM F543-13 for screws).
    • Demonstration of equivalent mechanical performance to a legally marketed predicate device (Synthes LCP Distal Tibia Plates - K013248).

8. Sample Size for Training Set

  • Not Applicable. This is a medical device (bone fixation system) and not an AI/ML diagnostic algorithm. Therefore, there is no "training set" in the context of machine learning. The device design and manufacturing process are informed by established engineering principles and material science, not by training on a dataset.

9. How the Ground Truth for the Training Set Was Established

  • Not Applicable. As there is no training set, there is no ground truth established for it.

{0}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three heads facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Jeil Medical Corporation Sejin Ryu RA Specialist #702. Kolon Science Vallev 2nd 55 Digital-ro 34. Guro-gu Seoul, 152-728 Korea

July 27, 2017

Re: K170313

Trade/Device Name: ARIX Ankle Distal Tibia System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: June 30, 2017 Received: July 3, 2017

Dear Sejin Ryu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for JEIL MEDICAL CORPORATION, followed by their address and contact information. The address is #702, Kolon Science Valley 2nd, 811 Guro-Dong, Guro-Gu, Seoul, 152-728 Korea. The telephone number is +82 2 850 3500, and the fax number is +82 2 850 3536. The logo features the company name in a stylized font, with a graphic of three colored dots above the letters.

Indications for Use

510(k) Number (if known): K170313

Device Name: ARIX Ankle Distal Tibia System

Indications for Use:

The ARIX Ankle Distal Tibia System is intended for fixation of complex intra- and extraarticular fractures and osteotomies of the distal tibia including distal tibia fractures in combination with diaphyseal fracture.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for JEIL MEDICAL CORPORATION. The logo features a colorful abstract design to the left of the company name, with "JEIL" in a bold orange font and "MEDICAL CORPORATION" in a smaller gray font. Below the logo is the text "#702, Kolon Science Valley 2nd, 811".

Gu. Seoul Tel : +82 2 850 3500 / Fax : +82 2 8

510(k) Summary

[As required by 21 CFR 807.92]

1. Date Prepared [21 CFR 807.92(a)(1)]

26 January 2017

2. Submitter's Information [21 CFR807.92(a)(1)]

  • . Name of Sponsor: Jeil Medical Corporation
    • . #702, Kolon Science Valley 2nd Address: 811, Guro-dong, Guro-gu Seoul, 152-728, Korea
  • Contact Name: Sejin RYU / RA Specialist
    • Telephone No. : +82 2 850 3500 —
    • Fax No. : +82 2 850 3536 —
    • Email Address : rsj@jeilmed.co.kr
  • Registration Number: 3004049923
  • . Name of Manufacturer: Same as Sponsor – Address: Same as Sponsor

3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]

• Trade Name:ARIX Ankle Distal Tibia System
• Common Name:Bone Plate and Screw
• Classification Name:Plate, Fixation, Bone / Screw, Fixation, Bone
• Classification Description:Single/multiple component metallic bone fixationappliances and accessoriesSmooth or threaded metallic bone fixation fastener
• Classification Panel:Orthopedic
• Classification Regulation:21 CFR 888.303021 CFR 888.3040

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo and contact information for JEIL MEDICAL CORPORATION. The logo features a stylized smiling face with green, yellow, orange, and gray colors. Below the logo is the company's address: #702, Kolon Science Valley 2nd, 811 Guro-Dong, Guro-Gu, Seoul, 152-728 Korea. The phone number is +82 2 850 3500, and the fax number is +82 2 850 3536.

  • Product Code: HRS / HWC ●
  • Device Class: ll

4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]

The identified predicate devices within this submission are shown as follows:

- Primary Predicate Devices

• 510(k) Number:K013248
• Device Name:Synthes LCP Distal Tibia Plates

- Additional Predicate Devices

• 510(k) Number:K163044
• Device Name:Ortholoc 3Di Ankle Fracture Plating System
  • 510(k) Number: K103705 ● Device Name: Low Profile Screws

- Reference Predicate Devices

• 510(k) Number:K112812
• Device Name:LeForte System
  • 510(k) Number: K151468 ● Device Name: ARIX Wrist System ●
  • 510(k) Number: K131311 ● Device Name: ARIX Foot System
  • 510(k) Number: K132876 ● Device Name: ARIX Hand Locking System

There are no significant differences between the Model ARIX Ankle Distal Tibia System and the predicate devices (K013248, K163044, K103705) that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, materials, and operational principles as internal fixation components.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image contains the logo for JEIL MEDICAL CORPORATION, along with their address and contact information. The address is #702, Kolon Science Valley 2nd, 811 Guro-Dong, Guro-Gu, Seoul, 152-728 Korea. The telephone number is +82 2 850 3500, and the fax number is +82 2 850 3536. The logo features the company name in orange and gray, with a stylized graphic of people in blue, green, orange, and gray.

5. Description of the Device [21 CFR 807.92(a)(4)]

The ARIX Ankle Distal Tibia System is rigid fixation consisting of plates and screws in various configurations, shapes and sizes, as follows:

PlateCortical ScrewLocking Screw
Type/ConfigurationMedial Type
AnterolateralType28-SO-L Series28L-SO-L Series
T Type35-SO-L Series35L-SO-L Series
MaterialASTM F 136,Titanium Alloy(Ti-6Al-4V)ASTM F 136,Titanium Alloy(Ti-6Al-4V)ASTM F 136,Titanium Alloy(Ti-6Al-4V)

The ARIX Ankle Distal Tibia System is made of Titanium Alloy (Ti-6AL-4V), which meet ASTM F136. Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications, which are widely used for surgical implants with well-known biocompatibility.

The plates vary essentially through different lengths and number of plate holes. The screws are self-tapping, which are applied with the reconstruction locking screws together. The Cortical Screws & locking screws are provided with diameter 2.8 mm to 3.5 mm and lengths from 8 mm to 80 mm.

ARIX Ankle Distal Tibia System is provided as Gold color which is casted according to the anodizing technique while apply 5 V electric energy to the titanium.

It also includes various manual surqical instruments, such as drill bits, counter sink, depth gauge, compression, drill guide, drill sleeve, guide pin, bender and screw driver handle.

The ARIX Ankle Distal Tibia System not provided sterile. It is required to be sterilized via autoclave method to reach a SAL of 10° by the hospital prior to surgery. The sterilization method is presented in the instruction, which was validated per ISO 17665-1: 2006 Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices.

6. Indication for Use [21 CFR 807.92(a)(5)]

The ARIX Ankle Distal Tibia System is intended for fixation of complex intra- and extraarticular fractures and osteotomies of the distal tibia including distal tibia fractures in combination with diaphyseal fracture.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo and contact information for JEIL MEDICAL CORPORATION. The logo features a stylized design with green, gray, and orange elements, along with the company name in bold letters. Below the logo is the address: #702, Kolon Science Valley 2nd, 811 Guro-Dong, Guro-Gu, Seoul, 152-728 Korea. The telephone number is +82 2 850 3500, and the fax number is +82 2 850 3536.

7. Technological Characteristics [21 CFR 807.92(a)(6)]

ARIX Ankle Distal Tibia System, Bone Plates: Based on a technical feature comparison, the subject device was found to be similar to predicate devices with regard to design and materials. The subject plates also have a locking feature, similar to the design used in the predicate device (K013248).

ARIX Ankle Distal Tibia System, Bone Screws: They share similar head, neck, and thread designs as the screws that are currently cleared under the predicate device (K013248).

Non-Clinical Test Summary:

Bench tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

  • . ASTM F 382-14, Standard Specification and Test Method for Metallic Bone Plates
  • ASTM F 543-13, Standard Specification and Test Methods for Metallic Medical Bone Screws

The following tests were performed with the predicate device:

  • . Plates
    • । Bending strength test per ASTM F382
    • Bending fatigue test per ASTM F382 —
  • Screw
    • Driving torque test per ASTM F543 —
    • Torsion test per ASTM F543 —
    • Axial pull out test per ASTM F543 ।

The results of this testing indicate that the ARIX Ankle Distal Tibia System is equivalent to predicate device.

Clinical Test Summary

No clinical studies were considered necessary and performed.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo and contact information for JEIL MEDICAL CORPORATION. The logo features a stylized graphic with the company name beside it. Below the logo is the address: #702, Kolon Science Valley 2nd, 811 Guro-Dong, Guro-Gu, Seoul, 152-728 Korea. The telephone number is +82 2 850 3500 and the fax number is +82 2 850 3536.

8. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]

When compared to the predicate device (K013248) the ARIX Ankle Distal Tibia System presented in this submission has the same:

  • Intended Use
  • . Technological characteristics
  • . Operating principle
  • . Design features
  • . Performance
  • Biocompatibility
  • Materials
  • Method of sterilization and sterility assurance level .

9. Conclusion [21 CFR 807.92(b)(3)]

In all respects, the ARIX Ankle Distal Tibia System is the equivalent of currently marketed devices. This device is made of the same materials and has similar dimensions and characteristics. This device is manufactured from titanium that is used generally in this kind of bone plate/screw system. This device, ARIX Ankle Distal Tibia System, is substantially equivalent in design, material, and function to the predicate device.

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.